Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP
Initial Studies in Proton ComputedTomography
L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang
Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064V. Bashkirov, R. W. M. Schulte, K. Shahnazi
Loma Linda University Medical Center, Loma Linda, CA 92354
• Proton Energy Loss in Matter
• Proton Tomography / Proton Transmission Radiography
• Proton Transmission Radiography Data
• Proton Transmission Radiography MC Study
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPComputed Tomography (CT)
X-ray tube
Detector array
• Based on X-ray absorption• Faithful reconstruction of
patient’s anatomy• Stacked 2D maps of linear
X-ray attenuation• Coupled linear equations• Invert Matrices and find
(hopefully) non-malignant structures
• Proton CT replaces X-ray absorption with proton energy loss
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP Radiography: X-rays vs. Protons
Attenuation of Photons, ZN(x) = Noe- µ x
Energy Loss of Protons, ρ∑∫ ∆≈=∆ l
dxdEdx
dxdEE ρ
NIST Data
dxdE
dldE ρ=
0.01
1
100
104
1 10 100 1000
X-Ray Absorption Coefficient
BoneMuscleH2OFat
µ
X-Ray Energy [keV]
[1/cm]
1
10
100
10 100 1000
Stopping Power for Protons
Bone MuscleH2OFat
dE/dl[MeV/cm]
Proton Energy E [MeV]
Low Contrast: ∆ρ∆ρ∆ρ∆ρ = 0.1 for tissue, 0.5 for bone
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP Advantages of Protons in Therapy
NIST Data
0
100
200
300
0
5
10
0 10 20 30 40 50
Proton Energy Loss in H2O
Proton Energy [MeV]
Energy Deposit in 1mm
[MeV/mm]
Water Depth [cm]
E = 130 MeV E = 250 MeV
• Relatively low entrance dose (plateau)
• Maximum dose at depth (Bragg peak)
• Rapid distal dose fall-off
• Energy modulation (Spread Bragg peak)
• RBE close to unity
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP Use of Proton Beam CT: Treatment Planning
Alderson Head Phantom
Range Uncertainties(measured with PTR)
> 5 mm> 10 mm> 15 mm
Schneider U. & Pedroni E. (1995), “Proton radiography as a tool for quality control in proton therapy,” Med Phys. 22, 353.
X-ray CT use in Proton Cancer Therapy can lead to large Uncertainties in Range Determination
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP
0
0.2
0.4
0.6
0.8
1Energy Loss in Water
Energy Loss
[MeV/mm]
150
200
250
0 5 10 15 20 25
Proton Energy
Proton Energy[MeV]
Depth in H2O [cm]
Low Contrast in Proton CT
ρρρρ
ρρρρ
1.01.01.01.01.11.11.11.11.51.51.51.52.02.02.02.0
Inclusion of 1cm depth at midpoint of 20cm H2O
132537.2158.92.0131737.7161.51.5131138.1163.61.1130938.2164.11.0
TOF[ps]
Range [cm]
Energy[MeV]
ρ
[g/cm2]
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP Proton CT Measurements
Requirements: • Proton location to few hundred um• Proton angle to a degree
• Average Proton Energy <E> to better than %
• Improve energy determination with statistics?
• Problem: Dose D = Absorbed Energy / MassVoxel with diameter d = 1mm 106 protons of 200 MeV = 7.2 [cGy]
• In order to minimize the dose, the final system needs to employ the best energy resolution!
NE
Eσσ =><
%1≈Eσ
52
2
~ D
AN D
d
dxdE
E
⋅∆
⋅=
ρσ
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP Development of Proton Beam Computed Tomography
• Exploratory Study in Proton Radiography– two x-y detector modules– Crude phantom in front
• Theoretical Study– GEANT4 MC simulation– influence of MCS and range straggling– importance of angular measurements– Optimization of energy
• Experimental Study in pCT – Three or four x-y Si planes– water phantom on turntable
Collaboration Loma Linda University Medical Center – UC Santa Cruz
Protonbeam
Simodule 2
Simodule 1
Water phantom
Turntable
Scatteringfoil
Simodule 3
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP Proton Energy Measurement with LET in Si
Simple 2D Silicon Strip Detector Telescope built
for Nanodosimetry (based on GLAST Design)
2 single-sided SSD
194um Pitch
400um thick
1.3us shaping time
Binary readout
Time-over-Threshold TOT
Large dynamic range
Measure particle energy via LET
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP GLAST Front-End Electronics ASIC
Binary Readout: •Low-power (~200uW/channel) •Peaking time ˜ 1.3 ms•Low noise (Noise occupancy <10-5)•Threshold set in every ASIC•Separate Masks for Trigger and Readout in every Channel•Self - Trigger = OR of one Si plane
(1536 channels)
Pulse Charge:Time – over-Threshold on the OR of every Si plane
Distinguish single tracks from two tracks in one strip
Electron Events
Photon Events
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPP
TOT ∝∝∝∝ charge ∝∝∝∝ LET
Charge ~ Time-Over-Threshold (TOT):
Digitization of Position and Energy Deposit with large Dynamic Range
PulseThreshold
Time-over-Threshold TOT
0
20
40
60
80
100
120
0 50 100 150 200
TOT Measurement vs Charge in MIP'sEffect of Threshold and Voltage
TOT SLACTOT LLUMC
Input Charge [fC]
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPProton Energy Measurement with LET
10
100
1 10 100 1000 104
TOT vs. Proton EnergyMeasurement vs. Expectation
TOT & Resolution measuredTOT expected
Proton Energy [MeV]
LLUMCSynchrotron P Beam
GLAST SLAC Test Beam
TOT Saturation
0
0.1
0.2
0.3
0.4
10 100 10000
0.1
0.2
0.3
0.4
Proton Energy [MeV]
TOT Saturation
Derive Energy Resolution from TOT vs. E Plot
Good agreement between measurement and MC simulations
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPExploratory Proton Radiography Set-up
Use Loma Linda University Medical Ctr 250 MeV Proton BeamDegraded down to 130 MeV by 10” Wax BlockObject is Aluminum pipe 5cm long, 3cm OD, 0.67cm ID Very large effects expected, but beam quite non-uniformSilicon detector telescope with 2 x-y modules
Beam from Synchrotron
30 cm
Object
Wax block
y x x y
Air
27.3 cm
Air
250MeV 130MeV 60+130MeV
Si Modules
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPImage !
Subdivide SSD area into pixels1. Strip x strip 194um x 194um2. 4 x 4 strips (0.8mm x 0.8mm)
Image is average energy in pixel
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPEnergy Resolution = Position Resolution
Hole “filled in”
“Fuzzy” Edges
Slice of 4x4 pixels
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPGEANT4 MC: Energy Reconstruction
NIST Data
Energy Loss in Si Energy Reconstructed
from Energy Loss in Si
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPMC: Loss of Resolution in Back
First Plane, 2cm behind Object
Second Plane, 30cm behind Object: Fuzzy
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPMultiple Scattering: Migration
Protons scatter OUT OF Target (not INTO).
Those have larger energy loss, larger angles, fill hole, dilute energy
Features:Washed out image in 2nd plane (30cm downstream)
Energy diluted at edges and in hole (Fuzzy edges,Hole filled partially)
Migration of events
are all explained by Multiple Coulomb Scattering MCS
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPMigration: MC
Dilution: Protons entering the Object in Front Facebut leaving it before the Rear Face
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPMC: Use Angular Information
Effect of Angular Cut: Energy more uniform
Less Migration
Sharp edges (RMS and Energy)
Hit Profile after angle cut
Hit Profile before angle cut
ΘΘΘΘ
≈≈≈≈ MCS angle
Imaging with MCS Angle?
Vertex2002 pCT: Hartmut F.-W. Sadrozinski , SCIPP
SCIPPSCIPPConclusions
Imaging with protons is working!
GEANT4 program describes the data well
(energy and position resolution, migration)
Issues:
• Energy needs Optimization depending on Target
• Improve Resolution with cut on exit angle √√√√
• Investigate independent Energy Measurement
• Dose – Contrast - Resolution Relationship
Next steps: pCT