+ All Categories
Home > Documents > Tutorial HPLC

Tutorial HPLC

Date post: 02-Apr-2018
Category:
Upload: mauricio-camarena
View: 236 times
Download: 0 times
Share this document with a friend

of 34

Transcript
  • 7/27/2019 Tutorial HPLC

    1/34

    www.waters.com

    HPLC - High Performance Liquid ChromatographyWhat Is HPLC (High Performance Liquid Chromatography)?

    Brief History and DefinitionLiquid chromatography was defined in the early 1900s by the work of the Russian botanist,

    Mikhail S. Tswett. His pioneering studies focused on separating compounds [leaf pigments],

    extracted from plants using a solvent, in a column packed with particles.

    Tswett filled an open glass column with particles. Two specific materials that he found useful

    were powdered chalk [calcium carbonate] and alumina. He poured his sample [solvent extract of

    homogenized plant leaves] into the column and allowed it to pass into the particle bed. This was

    followed by pure solvent. As the sample passed down through the column by gravity, different

    colored bands could be seen separating because some components were moving faster than

    others. He related these separated, different-colored bands to the different compounds that

    were originally contained in the sample. He had created an analytical separation of these

    compounds based on the differing strength of each compounds chemical attraction to theparticles. The compounds that were more strongly attracted to the particlesslowed down, while

    other compounds more strongly attracted to the solvent moved faster. This process can be

    described as follows: the compounds contained in the sample distribute, or partition differently

    between the moving solvent, called the mobile phase, and the particles, called the stationary

    phase. This causes each compound to move at a different speed, thus creating a separation of

    the compounds.

    Tswett coined the name chromatography[from the Greek words chroma, meaning color,

    and graph, meaning writingliterally, color writing] to describe his colorful experiment.

    [Curiously, the Russian name Tswett means color.] Today, liquid chromatography, in its various

    forms, has become one of the most powerful tools in analytical chemistry.

    Figure A: Tswett's Experiment

  • 7/27/2019 Tutorial HPLC

    2/34

    Liquid Chromatography (LC) Techniques

    Liquid chromatography can be performed using planar [Techniques 1 and 2] or column

    techniques [Technique 3]. Column liquid chromatography is the most powerful and has the

    highest capacity for sample. In all cases, the sample first must be dissolved in a liquid that is

    then transported either onto, or into, the chromatographic device.

    Technique 1. The sample is spotted onto, and then flows through, a thin layer ofchromatographic particles [stationary phase] fixed onto the surface of a glass plate [Figure B].

    The bottom edge of the plate is placed in a solvent. Flow is created by capillary action as the

    solvent [mobile phase] diffuses into the dry particle layer and moves up the glass plate. This

    technique is called thin-layer chromatography or TLC.

    Figure B: Thin-layer Chromatography

    Note that the blacksample is a mixture of FD&C yellow, red and blue food dyes that has been

    chromatographically separated.

    Technique 2. In Figure C, samples are spotted onto paper [stationary phase]. Solvent [mobile

    phase] is then added to the center of the spot to create an outward radial flow. This is a form of

    paper chromatography. [Classic paper chromatography is performed in a manner similar to that

    of TLC with linear flow.] In the upper image, the same black FD&C dye sample is applied to the

    paper.

  • 7/27/2019 Tutorial HPLC

    3/34

    Figure C: Paper Chromatography

    Notice the difference in separation power for this particular paper when compared to the TLC

    plate. The green ring indicates that the paper cannot separate the yellow and blue dyes from

    each other, but it could separate those dyes from the red dyes. In the bottom image, a green

    sample, made up of the same yellow and blue dyes, is applied to the paper. As you would

    predict, the paper cannot separate the two dyes. In the middle, a purple sample, made up of red

    and blue dyes, was applied to the paper. They are well separated.

    Technique 3. In this, the most powerful approach, the sample passes through a column or a

    cartridge device containing appropriate particles [stationary phase]. These particles are called

    the chromatographic packing material. Solvent [mobile phase] flows through the device. In

    solid-phase extraction [SPE], the sample is loaded onto the cartridge and the solvent stream

    carries the sample through the device. As in Tswetts experiment, the compounds in the sample

    are then separated by traveling at different individual speeds through the device. Here

    the blacksample is loaded onto a cartridge. Different solvents are used in each step to create

    the separation.

  • 7/27/2019 Tutorial HPLC

    4/34

    Figure D-1: Column Chromatography Solid-Phase Extraction [SPE]

    When the cartridge format is utilized, there are several ways to achieve flow. Gravity or vacuum

    can be used for columns that are not designed to withstand pressure. Typically, the particles in

    this case are larger in diameter [> 50 microns] so that there is less resistance to flow. Open

    glass columns [Tswetts experiment] are an example of this. In addition, small plastic columns,

    typically in the shape of syringe barrels, can be filled with packing-material particles and used toperform sample preparation. This is called solid-phase extraction [SPE]. Here, the

    chromatographic device, called a cartridge, is used, usually with vacuum-assisted flow, to clean

    up a very complex sample before it is analyzed further.

    Smaller particle sizes [

  • 7/27/2019 Tutorial HPLC

    5/34

    High performance liquid chromatography is now one of the most powerful tools in analytical

    chemistry. It has the ability to separate, identify, and quantitate the compounds that are present

    in any sample that can be dissolved in a liquid. Today, compounds in trace concentrations as low

    asparts per trillion [ppt] may easily be identified. HPLC can be, and has been, applied to just

    about any sample, such as pharmaceuticals, food, nutraceuticals, cosmetics, environmental

    matrices, forensic samples, and industrial chemicals.

    Figure D-2: HPLC Column

    What Is UltraPerformance Liquid Chromatography (UPLC Technology)?

    In 2004, further advances in instrumentation and column technology were made to achieve very

    significant increases in resolution, speed, and sensitivity in liquid chromatography. Columns with

    smaller particles [1.7 micron] and instrumentation with specialized capabilities designed to

    deliver mobile phase at 15,000 psi [1,000 bar] were needed to achieve a new level ofperformance. A new system had to be holistically created to perform ultra-performance liquid

    chromatography, now known as UPLC technology.

    Basic research is being conducted today by scientists working with columns containing even

    smaller 1-micron-diameter particles and instrumentation capable of performing at 100,000 psi

    [6,800 bar]. This provides a glimpse of what we may expect in the future.

    How Does High Performance Liquid Chromatography Work?

    The components of a basic high-performance liquid chromatography [HPLC] system are shown in

    the simple diagram in Figure E.

    A reservoir holds the solvent [called the mobile phase, because it moves]. A high-pressure pump

    [solvent delivery system or solvent manager] is used to generate and meter a specified flow rate

    of mobile phase, typically milliliters per minute. An injector [sample manager or autosampler] is

    able to introduce [inject] the sample into the continuously flowing mobile phase stream that

    carries the sample into the HPLC column. The column contains the chromatographic packing

    material needed to effect the separation. This packing material is called the stationary phase

    because it is held in place by the column hardware. A detector is needed to see the separated

    compound bands as they elute from the HPLC column [most compounds have no color, so we

    cannot see them with our eyes]. The mobile phase exits the detector and can be sent to waste,

    or collected, as desired. When the mobile phase contains a separated compound band, HPLC

    provides the ability to collect this fraction of the eluate containing that purified compound for

    further study. This is called preparative chromatography [discussed in the section on HPLC

    Scale].

    Note that high-pressure tubing and fittings are used to interconnect the pump, injector, column,

    and detector components to form the conduit for the mobile phase, sample, and separated

    compound bands.

  • 7/27/2019 Tutorial HPLC

    6/34

  • 7/27/2019 Tutorial HPLC

    7/34

    Figure F: A Typical HPLC [Waters Alliance] System

    HPLC Operation

    A simple way to understand how we achieve the separation of the compounds contained in a

    sample is to view the diagram in Figure G.

    Mobile phase enters the column from the left, passes through the particle bed, and exits at the

    right. Flow direction is represented by green arrows. First, consider the top image; it representsthe column at time zero [the moment of injection], when the sample enters the column and

    begins to form a band. The sample shown here, a mixture of yellow, red, and blue dyes, appears

    at the inlet of the column as a single black band. [In reality, this sample could be anything that

    can be dissolved in a solvent; typically the compounds would be colorless and the column wall

    opaque, so we would need a detector to see the separated compounds as they elute.]

    After a few minutes [lower image], during which mobile phase flows continuously and steadily

    past the packing material particles, we can see that the individual dyes have moved in separate

    bands at different speeds. This is because there is a competition between the mobile phase and

    the stationary phase for attracting each of the dyes or analytes. Notice that the yellow dye band

    moves the fastest and is about to exit the column. The yellow dye likes [is attracted to] the

    mobile phase more than the other dyes. Therefore, it moves at a fasterspeed, closer to that ofthe mobile phase. The blue dye band likes the packing material more than the mobile phase. Its

    stronger attraction to the particles causes it to move significantlyslower. In other words, it is the

    most retained compound in this sample mixture. The red dye band has an intermediate

    attraction for the mobile phase and therefore moves at an intermediate speed through the

    column. Since each dye band moves at different speed, we are able to separate it

    chromatographically.

  • 7/27/2019 Tutorial HPLC

    8/34

    Figure G: Understanding How a Chromatographic Column Works Bands

    What Is a Detector?

    As the separated dye bands leave the column, they pass immediately into the detector. The

    detector contains a flow cell that sees [detects] each separated compound band against a

    background of mobile phase [see Figure H]. [In reality, solutions of many compounds at typical

    HPLC analytical concentrations are colorless.] An appropriate detector has the ability to sense

    the presence of a compound and send its corresponding electrical signal to a computer data

    station. A choice is made among many different types of detectors, depending upon the

    characteristics and concentrations of the compounds that need to be separated and analyzed, as

    discussed earlier.

    What Is a Chromatogram?

    A chromatogram is a representation of the separation that has chemically [chromatographically]

    occurred in the HPLC system. A series of peaks rising from a baseline is drawn on a time axis.

    Each peak represents the detector response for a different compound. The chromatogram is

    plotted by the computer data station [see Figure H].

    Figure H: How Peaks Are Created

  • 7/27/2019 Tutorial HPLC

    9/34

    In Figure H, the yellow band has completely passed through the detector flow cell; the electrical

    signal generated has been sent to the computer data station. The resulting chromatogram has

    begun to appear on screen. Note that the chromatogram begins when the sample was first

    injected and starts as a straight line set near the bottom of the screen. This is called the

    baseline; it represents pure mobile phase passing through the flow cell over time. As the yellow

    analyte band passes through the flow cell, a stronger signal is sent to the computer. The line

    curves, first upward, and then downward, in proportion to the concentration of the yellow dye inthe sample band. This creates a peak in the chromatogram. After the yellow band passes

    completely out of the detector cell, the signal level returns to the baseline; the flow cell now has,

    once again, only pure mobile phase in it. Since the yellow band moves fastest, eluting first from

    the column, it is the first peak drawn.

    A little while later, the red band reaches the flow cell. The signal rises up from the baseline as

    the red band first enters the cell, and the peak representing the red band begins to be drawn. In

    this diagram, the red band has not fully passed through the flow cell. The diagram shows what

    the red band and red peak would look like if we stopped the process at this moment. Since most

    of the red band has passed through the cell, most of the peak has been drawn, as shown by the

    solid line. If we could restart, the red band would completely pass through the flow cell and the

    red peak would be completed [dotted line]. The blue band, the most strongly retained, travels atthe slowest rate and elutes after the red band. The dotted line shows you how the completed

    chromatogram would appear if we had let the run continue to its conclusion. It is interesting to

    note that the width of the blue peak will be the broadest because the width of the blue analyte

    band, while narrowest on the column, becomes the widest as it elutes from the column. This is

    because it moves more slowly through the chromatographic packing material bed and requires

    more time [and mobile phase volume] to be eluted completely. Since mobile phase is

    continuously flowing at a fixed rate, this means that the blue band widens and is more dilute.

    Since the detector responds in proportion to the concentration of the band, the blue peak is

    lower in height, but larger in width.

    Identifying and Quantitating Compounds

    In Figure H, three dye compounds are represented by three peaks separated in time in the

    chromatogram. Each elutes at a specific location, measured by the elapsed time between the

    moment of injection [time zero] and the time when the peak maximum elutes. By comparing

    each peaks retention time [tR] with that of injected reference standards in the same

    chromatographic system [same mobile and stationary phase], a chromatographer may be able

    to identify each compound.

  • 7/27/2019 Tutorial HPLC

    10/34

    Figure I-1: Identification

    In the chromatogram shown in Figure I-1, the chromatographer knew that, under these LC

    system conditions, the analyte, acrylamide, would be separated and elute from the column at

    2.85 minutes [retention time]. Whenever a new sample, which happened to contain acrylamide,

    was injected into the LC system under the same conditions, a peak would be present at 2.85

    minutes [see Sample B in Figure I-2].

    [For a better understanding of why some compounds move more slowly [are better retained]

    than others, please review the HPLC Separation Modes section on page 28].

    Once identity is established, the next piece of important information is how much of each

    compound was present in the sample. The chromatogram and the related data from the detector

    help us calculate the concentration of each compound. The detector basically responds to theconcentration of the compound band as it passes through the flow cell. The more concentrated it

    is, the stronger the signal; this is seen as a greater peak height above the baseline.

  • 7/27/2019 Tutorial HPLC

    11/34

    Figure I-2: Identification and Quantitation

    In Figure I-2, chromatograms for Samples A and B, on the same time scale, are stacked one

    above the other. The same volume of sample was injected in both runs. Both chromatograms

    display a peak at a retention time [tR] of 2.85 minutes, indicating that each sample contains

    acrylamide. However, Sample A displays a much bigger peak for acrylamide. The area under a

    peak [peak area count] is a measure of the concentration of the compound it represents. This

    area value is integrated and calculated automatically by the computer data station. In this

    example, the peak for acrylamide in Sample A has 10 times the area of that for Sample B. Using

    reference standards, it can be determined that Sample A contains 10 picograms of acrylamide,

    which is ten times the amount in Sample B [1 picogram]. Note there is another peak [not

    identified] that elutes at 1.8 minutes in both samples. Since the area counts for this peak in both

    samples are about the same, this unknown compound may have the same concentration in both

    samples.

    Isocratic and Gradient LC System Operation

    Two basic elution modes are used in HPLC. The first is called isocratic elution. In this mode, the

    mobile phase, either a pure solvent or a mixture, remains the same throughout the run. A

    typical system is outlined in Figure J-1.

  • 7/27/2019 Tutorial HPLC

    12/34

    Figure J-1: Isocratic LC System

    The second type is called gradient elution, wherein, as its name implies, the mobile phase

    composition changes during the separation. This mode is useful for samples that contain

    compounds that span a wide range of chromatographic polarity [see section on HPLC Separation

    Modes]. As the separation proceeds, the elution strength of the mobile phase is increased to

    elute the more strongly retained sample components.

    Figure J-2: High-Pressure-Gradient System

    In the simplest case, shown in Figure J-2, there are two bottles of solvents and two pumps. The

    speed of each pump is managed by the gradient controller to deliver more or less of each

  • 7/27/2019 Tutorial HPLC

    13/34

    solvent over the course of the separation. The two streams are combined in the mixer to create

    the actual mobile phase composition that is delivered to the column over time. At the beginning,

    the mobile phase contains a higher proportion of the weaker solvent [Solvent A]. Over time, the

    proportion of the stronger solvent [Solvent B] is increased, according to a predetermined

    timetable. Note that in Figure J-2, the mixer is downstream of the pumps; thus the gradient is

    created under high pressure. Other HPLC systems are designed to mix multiple streams of

    solvents under low pressure, ahead of a single pump. A gradient proportioning valve selectsfrom the four solvent bottles, changing the strength of the mobile phase over time [see Figure J-

    3].

    Figure J-3: Low-Pressure-Gradient System

    HPLC Scale [Analytical, Preparative, and Process]

    We have discussed how HPLC provides analytical data that can be used both to identify and to

    quantify compounds present in a sample. However, HPLC can also be used to purify and collect

    desired amounts of each compound, using a fraction collector downstream of the detector flow

    cell. This process is called preparative chromatography [see Figure K].

    In preparative chromatography, the scientist is able to collect the individual analytes as they

    elute from the column [e.g., in this example: yellow, then red, then blue].

  • 7/27/2019 Tutorial HPLC

    14/34

    Figure K: HPLC System for Purification: Preparative Chromatography

    The fraction collector selectively collects the eluate that now contains a purified analyte, for aspecified length of time. The vessels are moved so that each collects only a single analyte peak.

    A scientist determines goals for purity level and amount. Coupled with knowledge of the

    complexity of the sample and the nature and concentration of the desired analytes relative to

    that of the matrix constituents, these goals, in turn, determine the amount of sample that needs

    to be processed and the required capacity of the HPLC system. In general, as the sample size

    increases, the size of the HPLC column will become larger and the pump will need higher

    volume-flow-rate capacity. Determining the capacity of an HPLC system is called selecting the

    HPLC scale. Table A lists various HPLC scales and their chromatographic objectives.

    Table A: Chromatography Scale

    The ability to maximize selectivity with a specific combination of HPLC stationary and mobile

    phasesachieving the largest possible separation between two sample components of interest

    is critical in determining the requirements for scaling up a separation [see discussion on HPLC

    Separation Modes]. Capacity then becomes a matter of scaling the column volume [Vc] to the

    amount of sample to be injected and choosing an appropriate particle size [determines pressure

    and efficiency; see discussion of Separation Power]. Column volume, a function of bed length [L]

    and internal diameter [i.d.], determines the amount of packing material [particles] that can be

    contained (see Figure L).

  • 7/27/2019 Tutorial HPLC

    15/34

    Figure L: HPLC Column Dimensions

    In general, HPLC columns range from 20 mm to 500 mm in length [L] and 1 mm to 100 mm in

    internal diameter [i.d.]. As the scale of chromatography increases, so do column dimensions,

    especially the cross-sectional area. To optimize throughput, mobile phase flow rates must

    increase in proportion to cross-sectional area. If a smaller particle size is desirable for more

    separation power, pumps must then be designed to sustain higher mobile-phase-volume flow

    rates at high backpressure. Table B presents some simple guidelines on selecting the column i.d.

    and particle size range recommended for each scale of chromatography.

    For example, a semi-preparative-scale application [red X] would use a column with an internal

    diameter of 1040 mm containing 515 micron particles. Column length could then be calculated

    based on how much purified compound needs to be processed during each run and on how much

    separation power is required.

    Table B: Chromatography Scale vs. Column Diameter and Particle Size

  • 7/27/2019 Tutorial HPLC

    16/34

    HPLC Column Hardware

    A column tube and fittings must contain the chromatographic packing material [stationary

    phase] that is used to effect a separation. It must withstand backpressure created both during

    manufacture and in use. Also, it must provide a well-controlled [leak-free, minimum-volume,

    and zero-dead-volume] flow path for the sample at its inlet, and analyte bands at its outlet, and

    be chemically inert relative to the separation system [sample, mobile, and stationary phases].

    Most columns are constructed of stainless steel for highest pressure resistance. PEEK[an

    engineered plastic] and glass, while less pressure tolerant, may be used when inert surfaces are

    required for special chemical or biological applications. [Figure M-1].

    Figure M-1: Column Hardware Examples

    A glass column wall offers a visual advantage. In the photo in Figure M-2, flow has been stopped

    while the sample bands are still in the column. You can see that the three dyes in the injected

    sample mixture have already separated in the bed; the yellow analyte, traveling fastest, is just

    about to exit the column.

    Figure M-2: A Look Inside a Column

    Separation Performance Resolution

    The degree to which two compounds are separated is called chromatographic resolution [RS].

    Two principal factors that determine the overall separation power or resolution that can be

  • 7/27/2019 Tutorial HPLC

    17/34

    achieved by an HPLC column are: mechanical separation power, created by the column length,

    particle size, and packed-bed uniformity, and chemical separation power, created by the

    physicochemical competition for compounds between the packing material and the mobile phase.

    Efficiency is a measure of mechanical separation power, while selectivity is a measure of

    chemical separation power.

    Mechanical Separation Power EfficiencyIf a column bed is stable and uniformly packed, its mechanical separation power is determined

    by the column length and the particle size. Mechanical separation power, also called efficiency, is

    often measured and compared by a plate number [symbol = N]. Smaller-particle

    chromatographic beds have higher efficiency and higher backpressure. For a given particle size,

    more mechanical separation power is gained by increasing column length. However, the trade-

    offs are longer chromatographic run times, greater solvent consumption, and higher

    backpressure. Shorter column lengths minimize all these variables but also reduce mechanical

    separation power, as shown in Figure N.

    Figure N: Column Length and Mechanical Separating Power [Same Particle Size]

    Figure O: Particle Size and Mechanical Separating Power [Same Column Length]

    For a given particle chemistry, mobile phase, and flow rate, as shown in Figure O, a column of

    the same length and i.d., but with a smaller particle size, will deliver more mechanical

    separation power in the same time. However, its backpressure will be much higher.

  • 7/27/2019 Tutorial HPLC

    18/34

  • 7/27/2019 Tutorial HPLC

    19/34

    Another way to think of this is by the familiar analogy: oil [non-polar] and water [polar] dont

    mix. Unlike in magnetism where opposite poles attract each other, chromatographic separations

    based on polarity depend upon the stronger attraction between likes and the weaker attraction

    between opposites. Remember,Like attracts likein polarity-based chromatography.

    Figure Q: Proper Combination of Mobile and Stationary Phases Effects Separation Based on Polarity

    To design a chromatographic separation system [see Figure Q], we create competition for the

    various compounds contained in the sample by choosing a mobile phase and a stationary phase

    with different polarities. Then, compounds in the sample that are similar in polarity to the

    stationary phase [column packing material] will be delayed because they are more stronglyattracted to the particles. Compounds whose polarity is similar to that of the mobile phase will

    be preferentially attracted to it and move faster.

    In this way, based upon differences in the relative attraction of each compound for each phase,

    a separation is created by changing the speeds of the analytes.

    Figures R-1, R-2, and R-3 display typical chromatographic polarity ranges for mobile phases,

    stationary phases, and sample analytes, respectively. Lets consider each in turn to see how a

    chromatographer chooses the appropriate phases to develop the attraction competition needed

    to achieve a polarity-based HPLC separation.

    Figure R-1: Mobile Phase Chromatographic Polarity Spectrum

    A scale, such as that shown in Figure R-1, upon which some common solvents are placed in

    order of relative chromatographic polarity is called an eluotropic series. Mobile phase molecules

    that compete effectively with analyte molecules for the attractive stationary phase sites displace

    these analytes, causing them to move faster through the column [weakly retained]. Water is at

    the polar end of mobile-phase-solvent scale, while hexane, an aliphatic hydrocarbon, is at the

    non-polar end. In between, single solvents, as well as miscible-solvent mixtures [blended in

    proportions appropriate to meet specific separation requirements], can be placed in order of

  • 7/27/2019 Tutorial HPLC

    20/34

    elution strength. Which end of the scale represents the strongest mobile phase depends upon

    the nature of the stationary phase surface where the competition for the analyte molecules

    occurs.

    Figure R-2: Stationary Phase Particle Chromatographic Polarity Spectrum

    Silica has an active, hydrophilic [water-loving] surface containing acidic silanol [silicon-

    containing analog of alcohol] functional groups. Consequently, it falls at the polar end of the

    stationary-phase scale shown in Figure R-2. The activity or polarity of the silica surface may be

    modified selectively by chemically bonding to it less polar functional groups [bonded phase].

    Examples shown here include, in order of decreasing polarity, cyanopropylsilyl- [CN], n-

    octylsilyl- [C8], and n-octadecylsilyl- [C18, ODS] moieties on silica. The latter is a hydrophobic

    [water-hating], very non-polar packing.

    Figure R-3: Compound/Analyte Chromatographic Polarity Spectrum

    Figure R-3 repeats the chromatographic polarity spectrum of our sample [shown in Figure P].

    After considering the polarity of both phases, then, for a given stationary phase, a

    chromatographer must choose a mobile phase in which the analytes of interest are retained, but

    not so strongly that they cannot be eluted. Among solvents of similar strength, the

    chromatographer considers which phase combination may best exploit the more subtle

    differences in analyte polarity and solubility to maximize the selectivity of the chromatographic

    system. Like attracts like, but, as you probably can imagine from the discussion so far, creating

    a separation based upon polarity involves knowledge of the sample and experience with various

    kinds of analytes and retention modes. To summarize, the chromatographer will choose the best

    combination of a mobile phase and particle stationary phase with appropriately opposite

    polarities. Then, as the sample analytes move through the column, the rule like attracts like will

    determine which analytes slow down and which proceed at a faster speed.

  • 7/27/2019 Tutorial HPLC

    21/34

    Normal-Phase HPLC

    In his separations of plant extracts, Tswett was successful using a polar stationary phase [chalk

    in a glass column; see Figure A] with a much less polar [non-polar] mobile phase. This classical

    mode of chromatography became known as normal phase.

    Figure S-1: Normal-Phase Chromatography

    Figure S-1 represents a normal-phase chromatographic separation of our three-dye test mixture.

    The stationary phase is polar and retains the polar yellow dye most strongly. The relatively non-

    polar blue dye is won in the retention competition by the mobile phase, a non-polar solvent, and

    elutes quickly. Since the blue dye is most like the mobile phase [both are non-polar], it moves

    faster. It is typical for normal-phase chromatography on silica that the mobile phase is 100%

    organic; no water is used.

    Reversed-Phase HPLC

    The term reversed-phase describes the chromatography mode that is just the opposite of normal

    phase, namely the use of a polar mobile phase and a non-polar [hydrophobic] stationary phase.

    Figure S-2 illustrates the black three-dye mixture being separated using such a protocol.

    Figure S-2: Reversed-Phase Chromatography

    Now the most strongly retained compound is the more non-polar blue dye, as its attraction to

    the non-polar stationary phase is greatest. The polar yellow dye, being weakly retained, is won

    in competition by the polar, aqueous mobile phase, moves the fastest through the bed, and

    elutes earliest like attracts like.

    Today, because it is more reproducible and has broad applicability, reversed-phase

    chromatography is used for approximately 75% of all HPLC methods. Most of these protocols use

  • 7/27/2019 Tutorial HPLC

    22/34

    as the mobile phase an aqueous blend of water with a miscible, polar organic solvent, such as

    acetonitrile or methanol. This typically ensures the proper interaction of analytes with the non-

    polar, hydrophobic particle surface. A C18bonded silica [sometimes called ODS] is the most

    popular type of reversed-phase HPLC packing.

    Table C presents a summary of the phase characteristics for the two principal HPLC separation

    modes based upon polarity. Remember, for these polarity-based modes,like attracts like.

    Table C: Phase Characteristics for Separations Based on Polarity

    Hydrophilic-Interaction Chromatography [HILIC]

    HILIC may be viewed as a variant of normal-phase chromatography. In normal-phase

    chromatography, the mobile phase is 100% organic. Only traces of water are present in the

    mobile phase and in the pores of the polar packing particles. Polar analytes bind strongly to the

    polar stationary phase and may not elute.

    Adding some water [< 20%] to the organic mobile phase [typically an aprotic solvent like

    acetonitrile] makes it possible to separate and elute polar compounds that are strongly retained

    in the normal-phase mode [or weakly retained in the reversed-phase mode]. Water, a very polar

    solvent, competes effectively with polar analytes for the stationary phase. HILIC may be run in

    either isocratic or gradient elution modes. Polar compounds that are initially attracted to thepolar packing material particles can be eluted as the polarity [strength] of the mobile phase is

    increased [by adding more water]. Analytes are eluted in order of

    increasinghydrophilicity[chromatographic polarity relative to water]. Buffers or salts may be

    added to the mobile phase to keep ionizable analytes in a single form.

    Hydrophobic-Interaction Chromatography [HIC]

    HIC is a type of reversed-phase chromatography that is used to separate large biomolecules,

    such as proteins. It is usually desirable to maintain these molecules intact in an aqueous

    solution, avoiding contact with organic solvents or surfaces that might denature them. HIC takes

    advantage of the hydrophobic interaction of large molecules with a moderately hydrophobic

    stationary phase, e.g., butyl-bonded [C4], rather than octadecyl-bonded [C18], silica. Initially,

    higher salt concentrations in water will encourage the proteins to be retained [salted out] on thepacking. Gradient separations are typically run by decreasing salt concentration. In this way,

    biomolecules are eluted in order of increasing hydrophobicity.

    Separations Based on Charge: Ion-Exchange Chromatography [IEC]

    For separations based on polarity, like is attracted to like and opposites may be repelled. In ion-

    exchange chromatography and other separations based upon electrical charge, the rule is

    reversed. Likes may repel, while opposites are attracted to each other. Stationary phases for

    ion-exchange separations are characterized by the nature and strength of the acidic or basic

  • 7/27/2019 Tutorial HPLC

    23/34

    functions on their surfaces and the types of ions that they attract and retain. Cation exchange is

    used to retain and separate positively charged ions on a negative surface.

    Conversely, anion exchange is used to retain and separate negatively charged ions on

    apositive surface [see Figure T]. With each type of ion exchange, there are at least two general

    approaches for separation and elution.

    Figure T: Ion-Exchange Chromatography

    Strong ion exchangers bear functional groups [e.g., quaternary amines or sulfonic acids] that

    are always ionized. They are typically used to retain and separate weakions. These weak ions

    may be eluted by displacement with a mobile phase containing ions that are more strongly

    attracted to the stationary phase sites. Alternately, weak ions may be retained on the column,

    then neutralizedby in situchanging the pH of the mobile phase, causing them to lose their

    attraction and elute.

    Weakion exchangers [e.g., with secondary-amine or carboxylic-acid functions] may be

    neutralized above or below a certain pH value and lose their ability to retain ions by charge.

    When charged, they are used to retain and separate strong ions. If these ions cannot be eluted

    by displacement, then the stationary phase exchange sites may be neutralized, shutting offthe

    ionic attraction, and permitting elution of the charged analytes.

  • 7/27/2019 Tutorial HPLC

    24/34

    Table D: Ion-Exchange Guidelines

    When weak ion exchangers are neutralized, they may retain and separate species

    by hydrophobic[reversed-phase] or hydrophilic[normal-phase] interactions; in these cases,elution strength is determined by the polarity of the mobile phase [Figure R-1]. Thus, weak ion

    exchangers may be used for mixed-mode separations [separations based on both polarity and

    charge].

    Table D outlines guidelines for the principal categories of ion exchange. For example, to retain

    a strongly basicanalyte [always positively charged], use a weak-cation-exchange stationary

    phase particle at pH > 7; this assures a negativelycharged particle surface. To release or elute

    the strong base, lower the pH of the mobile phase below 3; this removes the surface charge

    and shuts offthe ion-exchange retention mechanism.

    Note that a pKa is the pH value at which 50% of the functional group is ionized and 50% is

    neutral. To assure an essentially neutral, or a fully charged, analyte or particle surface, the pH

    must be adjusted to a value at least 2 units beyond the pKa, as appropriate [indicated in Table

    D].

    Do not use a strong-cation exchanger to retain a strong base; both remain charged and strongly

    attracted to each other, making the base nearly impossible to elute. It can only be removed by

    swamping the strong cation exchanger with a competing base that exhibits even stronger

    retention and displaces the compound of interest by winning the competition for the active

  • 7/27/2019 Tutorial HPLC

    25/34

    exchange sites. This approach is rarely practical, or safe, in HPLC and SPE. [Very strong acids

    and bases are dangerous to work with, and they may be corrosive to materials of construction

    used in HPLC fluidics!]

    Separations Based on Size: Size-Exclusion Chromatography [SEC]

    Gel-Permeation Chromatography [GPC]

    In the 1950s, Porath and Flodin discovered that biomolecules could be separated based on their

    size, rather than on their charge or polarity, by passing, or filtering, them through a controlled-

    porosity, hydrophilic dextran polymer. This process was termed gel filtration. Later, an

    analogous scheme was used to separate synthetic oligomers and polymers using organic-

    polymer packings with specific pore-size ranges. This process was called gel-permeation

    chromatography [GPC]. Similar separations done using controlled-porosity silica packings were

    called size-exclusion chromatography [SEC]. Introduced in 1963, the first commercial HPLC

    instruments were designed for GPC applications [see Reference 3].

    All of these techniques are typically done on stationary phases that have been synthesized with

    a pore-size distribution over a range that permits the analytes of interest to enter, or to be

    excluded from, more or less of the pore volume of the packing. Smaller molecules penetrate

    more of the pores on their passage through the bed. Larger molecules may only penetrate pores

    above a certain size so they spend less time in the bed. The biggest molecules may be totally

    excluded from pores and pass only between the particles, eluting very quickly in a small volume.

    Mobile phases are chosen for two reasons: first, they are good solvents for the analytes; and,

    second, they may prevent any interactions [based on polarity or charge] between the analytes

    and the stationary phase surface. In this way, the larger molecules elute first, while the smaller

    molecules travel slower [because they move into and out of more of the pores] and elute later,

    in decreasing order of their size in solution. Hence the simple rule: Big ones come out first.

    Since it is possible to correlate the molecular weight of a polymer with its size in solution, GPC

    revolutionized measurement of the molecular-weight distribution of polymers that, in turn,

    determines the physical characteristics that may enhance, or detract from, polymer processing,

    quality, and performance [how to tell goodfrombadpolymer].

    Conclusion

    We hope you have enjoyed this brief introduction to HPLC. We encourage you to read the

    references below and to study the Appendix on HPLC Nomenclature.

    Appendix: HPLC Nomenclature

    *Indicates a definition adapted from: L.S. Ettre, Nomenclature for Chromatography, Pure Appl.Chem. 65: 819-872 [1993], 1993 IUPAC; an updated version of this comprehensive report is

    available in the Orange Book, Chapter 9: Separations [1997] at:

    .

    Alumina

    A porous, particulate form of aluminum oxide [Al203] used as a stationary phase in normal-

    phase adsorption chromatography. Alumina has a highly active basic surface; the pH of a 10%

    aqueous slurry is about 10. It is successively washed with strong acid to make neutral and acidic

  • 7/27/2019 Tutorial HPLC

    26/34

    grades [slurry pH 7.5 and 4, resp.]. Alumina is more hygroscopic than silica. Its activity is

    measured according to the Brockmann scale for water content; e.g., Activity Grade I contains

    1% H2O.

    H. Brockmann and H. Schodder, Ber. 74: 73 (1941).

    Baseline*

    The portion of the chromatogram recording the detector response when only the mobile phase

    emerges from the column.

    Cartridge

    A type of column, without endfittings, that consists simply of an open tube wherein the packing

    material is retained by a frit at either end. SPE cartridges may be operated in parallel on a

    vacuum-manifold. HPLC cartridges are placed into a cartridge holder that has fluid connections

    built into each end. Cartridge columns are easy to change, less expensive, and more convenient

    than conventional columns with integral endfittings.

    Chromatogram*

    A graphical or other presentation of detector response or other quantity used as a measure of

    the concentration of the analyte in the effluent versus effluent volume or time. In planar

    chromatography [e.g., thin-layer chromatography or paper

    chromatography], chromatogram may refer to the paper or layer containing the separated

    zones.

    Chromatography*

    A dynamic physicochemical method of separation in which the components to be separated are

    distributed between two phases, one of which is stationary [thestationary phase] while the other

    [the mobile phase] moves relative to the stationary phase.

    Column Volume* [Vc]

    The geometric volume of the part of the tube that contains the packing [internal cross-sectional

    area of the tube multiplied by the packed bed length, L]. Theinterparticle volume of the column,

    also called the interstitial volume, is the volume occupied by the mobile phase between the

    particles in the packed bed. The void volume [V0] is the total volume occupied by the mobile

    phase, i.e. the sum of the interstitial volume and the intraparticle volume [also calledpore

    volume].

    Detector* [see Sensitivity]

    A device that indicates a change in the composition of the eluent by measuring physical or

    chemical properties [e.g., UV/visible light absorbance, differential refractive index, fluorescence,

    or conductivity]. If the detectors response is linear with respect to sample concentration, then,

    by suitable calibration with standards, the amount of a component may be quantitated. Often, it

    may be beneficial to use two different types of detectors in series. In this way, more

    corroboratory or specific information may be obtained about the sample analytes. Some

    detectors [e.g.,electrochemical, mass spectrometric] are destructive; i.e., they effect a chemical

    change in the sample components. If a detector of this type is paired with a non-destructive

    detector, it is usually placed second in the flow path.

    Display

    A device that records the electrical response of a detector on a computer screen in the form of a

  • 7/27/2019 Tutorial HPLC

    27/34

    chromatogram. Advanced data recording systems also perform calculations using sophisticated

    algorithms, e.g., to integrate peak areas, subtract baselines, match spectra, quantitate

    components, and identify unknowns by comparison to standard libraries.

    Efficiency [H, see Plate Number, Resolution, Sensitivity, Speed]

    A measure of a columns ability to resist the dispersion of a sample band as it passes through

    the packed bed. An efficient column minimizes band dispersion orbandspreading. Higherefficiency is important for effective separation, greater sensitivity, and/or identification of similar

    components in a complex sample mixture.

    Nobelists Martin and Synge, by analogy to distillation, introduced the concept ofplate height[H,

    or H.E.T.P., height equivalent to a theoretical plate] as a measure of chromatographic efficiency

    and as a means to compare column performance. Presaging HPLC and UPLC technology, they

    recognized that a homogeneous bed packed with the smallest possible particle size [requiring

    higher pressure] was key to maximum efficiency. The relation between column and separation

    system parameters that affect bandspreading

    was later described in an equation by van Deemter.

    Chromatographers often refer to a quantity that they can calculate easily and directly from

    measurements made on a chromatogram, namelyplate number[N], as efficiency. Plate height is

    then determined from the ratio of the length of the column bed to N [H = L/N; methods of

    calculating N from a chromatogram are shown in Figure U]. It is important to note that

    calculation of N or H using these methods is correct only for isocratic conditions and cannot be

    used for gradient separations.

    A.J.P. Martin and R.M. Synge, Biochem. J. 35: 1358-1368 [1941]

    J.J. van Deemter, F. J. Zuiderweg and A. Klinkenberg, Chem. Eng. Sci. 5: 271-289 [1956]

    Eluate

    The portion of the eluentthat emerges from the column outlet containing analytes in solution. In

    analytical HPLC, the eluate is examined by the detector for the concentration or mass of analytes

    therein. In preparative HPLC, the eluate is collected continuously in aliquots at uniform time or

    volume intervals, or discontinuously only when a detector indicates the presence of a peak of

    interest. These fractions are subsequently processed to obtain purified compounds.

    Eluent

    The mobile phase [see Elution Chromatography].

    Eluotropic Series

    A list of solvents ordered by elution strength with reference to specified analytes on a standard

    sorbent. Such a series is useful when developing both isocratic and gradient elution methods.

    Trappe coined this term after showing that a sequence of solvents of increasing polarity could

    separate lipid fractions on alumina. Later, Snyder measured and tabulated solvent strength

    parameters for a large list of solvents on several normal-phase LC sorbents. Neher created a

    very usefulnomogram by which equi-eluotropic

    [constant elution strength] mixtures of normal-phase solvents could be chosen to optimize the

    selectivity of TLC separations.

    A typical normal-phase eluotropic series would start at the weak end with non-polar aliphatic

    hydrocarbons, e.g., pentane or hexane, then progress successively to benzene [an aromatic

  • 7/27/2019 Tutorial HPLC

    28/34

    hydrocarbon], dichloromethane [a chlorinated hydrocarbon], diethyl ether, ethyl acetate [an

    ester], acetone [a ketone], and, finally, methanol [an alcohol] at the strong end [see Figure R-

    1].

    W. Trappe, Biochem. Z. 305: 150 [1940]

    L. R. Snyder, Principles of Adsorption Chromatography, Marcel Dekker [1968], pp. 192-197

    R. Neher in G.B. Marini-Bettlo, ed., Thin-Layer Chromatography, Elsevier [1964] pp. 75-86.

    Elute* [verb]

    To chromatograph by elution chromatography. The process of elution may be stopped while all

    the sample components are still on the chromatographic bed [planar thin-layer or paper

    chromatography] or continued until the components have left the chromatographic bed [column

    chromatography].

    Note: The term elute is preferred to develop [a term used in planar chromatography], to avoid

    confusion with the practice of method development, whereby a separation system [the

    combination of mobile and stationary phases] is optimized for a particular separation.

    Elution Chromatography*

    A procedure for chromatographic separation in which the mobile phase is continuously passed

    through the chromatographic bed. In HPLC, once the detector baseline has stabilized and the

    separation system has reached equilibrium, a finite slug of sample is introduced into the flowing

    mobile phase stream. Elution continues until all analytes of interest have passed through the

    detector.

    Elution Strength

    A measure of the affinity of a solvent relative to that of the analyte for the stationary phase. A

    weak solvent cannot displace the analyte, causing it to be strongly retained on the stationary

    phase. A strong solvent may totally displace all the analyte molecules and carry them through

    the column unretained. To achieve a proper balance of effective separation and reasonable

    elution volume, solvents are often blended to set up an appropriate competition between the

    phases, thereby optimizing both selectivity and

    separation time for a given set of analytes [see Selectivity].

    Dipole moment, dielectric constant, hydrogen bonding, molecular size and shape, and surface

    tension may give some indication of elution strength. Elution strength is also determined by the

    separation mode. An eluotropic series of solvents may be ordered by increasing strength in one

    direction under adsorption or normal-phaseconditions; that order may be nearly opposite

    under reversed-phase partitionconditions [see Figure R-1].

    Fluorescence Detector

    Fluorescence detectors excite a sample with a specified wavelength of light. This causes certain

    compounds to fluoresce and emit light at a higher wavelength. A sensor, set to a

    specific emission wavelength and masked so as not to be blinded by the excitation source,

    collects only the emitted light. Often analytes that do not natively fluoresce may be derivatized

    to take advantage of the high sensitivity and selectivity of this form of

    detection, e.g., AccQTag derivatization of amino acids.

    Flow Rate*

    The volume of mobile phase passing through the column in unit time. In HPLC systems, the flow

  • 7/27/2019 Tutorial HPLC

    29/34

    rate is set by the controller for the solvent delivery system [pump]. Flow rate accuracy can be

    checked by timed collection and measurement of the effluent at the column outlet. Since a

    solvents density varies with temperature, any calibration or flow rate measurement must take

    this variable into account. Most accurate determinations are made, when possible, by weight,

    not volume.

    Uniformity [precision] and reproducibilityof flow rate is important to many LC techniques,especially in separations where retention times are key to analyte identification, or in gel-

    permeation chromatographywhere calibration and correlation of retention times are critical to

    accurate molecular-weight-distribution measurements of polymers.

    Often, separation conditions are compared by means oflinear velocity, not flow rate. The linear

    velocity is calculated by dividing the flow rate by the cross-sectional area of the column. While

    flow rate is expressed in volume/time [e.g., mL/min], linear velocity is measured in length/time

    [e.g., mm/sec].

    Gel-Permeation Chromatography*

    Separation based mainly upon exclusion effects due to differences in molecular size and/or

    shape. Gelpermeation chromatographyand gel filtration chromatographydescribe the process

    when the stationary phase is a swollen gel. Both are forms ofsize-exclusion chromatography.

    Porath and Flodin first described gel-filtration using dextran gels and aqueous mobile phases for

    the size-based separation of biomolecules. Moore applied similar principles to the separation of

    organic polymers by size in solution using

    organic-solvent mobile phases on porous polystyrene-divinylbenzene polymer gels.

    J. Porath, P. Flodin, Nature 183: 1657-1659 [1959]

    J.C. Moore, U.S. Patent3,326,875 [filed Jan. 1963; issued June 1967]

    Gradient

    The change over time in the relative concentrations of two [or more] miscible solvent

    components that form a mobile phase of increasing elution strength. A step gradientis typically

    used in solid-phase extraction; in each step, the eluent composition is changed abruptly from a

    weaker mobile phase to a stronger mobile phase. It is even possible, by drying the SPE sorbent

    bed in between steps, to change from one solvent to another immiscible solvent.

    A continuous gradient is typically generated by a low- or high-pressure mixing system [see

    Figures J-2 and J-3] according to a pre-determined curve [linear or non-linear] representing the

    concentration of the stronger solvent B in the initial solvent A over a fixed time period. A holdat

    a fixed isocraticsolvent composition can be programmed at any time point within a continuous

    gradient. At the end of a separation, the gradient program can also be set to return to the initial

    mobile phase composition to re-equilibrate the column in preparation for the injection of the

    next sample. Sophisticated HPLC systems can blend as many as four or more solvents [or

    solvent mixtures] into a continuous gradient.

    Injector [Autosampler, Sample Manager]

    A mechanism for accurately and precisely introducing [injecting] a discrete, predetermined

    volume of a sample solution into the flowing mobile phase stream. The injector can be a simple

    manual device, or a sophisticated autosampler that can be programmed for unattended

    injections of many samples from an array of individual vials or wells in a predetermined

  • 7/27/2019 Tutorial HPLC

    30/34

    sequence. Sample compartments in these systems may even be temperature controlled to

    maintain sample integrity over many hours of operation.

    Most modern injectors incorporate some form of syringe-filled sample loop that can be switched

    on- or offline by means of a multi-port valve. A well-designed, minimal-internal-volume injection

    system is situated as close to the column inlet as possible and minimizes the spreading of the

    sample band. Between sample injections, it is also capable of being flushed to waste by mobilephase, or a wash solvent, to prevent carryover[contamination of the present sample by a

    previous one].

    Samples are best prepared for injection, if possible, by dissolving them in the mobile phase into

    which they will be injected; this may prevent issues with separation and/or detection. If another

    solvent must be used, it is desirable that its elution strength be equal to or less than that of the

    mobile phase. It is often wise to mix a bit of a sample solution with the mobile phase offline to

    test for precipitation or miscibility issues that might compromise a successful separation.

    Inlet

    The end of the column bed where the mobile phase stream and sample enter. A porous, inert frit

    retains the packing material and protects the sorbent bed inlet from particulate contamination.

    Good HPLC practice dictates that samples and mobile phases should be particulate-free; this

    becomes imperative for small-particle columns whose inlets are much more easily plugged. If

    the column bed inlet becomes clogged and exhibits higher-than-normal backpressure,

    sometimes, reversing the flow direction while directing the effluent to waste may dislodge and

    flush out sample debris that sits atop the frit. If the

    debris has penetrated the frit and is lodged in the inlet end of the bed itself, then the column has

    most likely reached the end of its useful life.

    Ion-Exchange Chromatography* [see section: Separations Based on Charge]

    This separation mode is based mainly on differences in the ion-exchange affinities of the sample

    components. Separation of primarily inorganic ionic species in water or buffered aqueous mobile

    phases on small particle, superficially porous, high-efficiency, ion-exchange columns followed by

    conductometric or electrochemical detection is referred to as ion chromatography [IC].

    Isocratic Elution*

    A procedure in which the composition of the mobile phase remains constant during the elution

    process.

    Liquid Chromatography* [LC]

    A separation technique in which the mobile phase is a liquid. Liquid chromatography can be

    carried out either in a column or on a plane [TLC or paper chromatography]. Modern liquid

    chromatography utilizing smaller particles and higher inlet pressure was termed high-

    performance (or high-pressure) liquid chromatography[HPLC] in 1970. In 2004, ultra-

    performance liquid chromatographydramatically raised the performance of LC to a new plateau

    [see UPLC Technology].

    Mobile Phase* [see Eluate, Eluent]

    A fluid that percolates, in a definite direction, through the length of the stationary-phase sorbent

    bed. The mobile phase may be a liquid [liquid chromatography] or a gas [gas chromatography]

    or a supercritical fluid [supercritical-fluid chromatography]. In gas chromatography the

    expression carrier gas may be used for the mobile phase. In elution chromatography, the mobile

  • 7/27/2019 Tutorial HPLC

    31/34

    phase may also be called the eluent, while the word eluate is defined as the portion of the

    mobile phase that has passed through the sorbent bed and contains the compounds of interest

    in solution.

    Normal-Phase Chromatography*

    An elution procedure in which the stationary phase is more polar than the mobile phase. This

    term is used in liquid chromatography to emphasize the contrast toreversed-phasechromatography.

    Peak* [see Plate Number]

    The portion of a differential chromatogram recording the detector response while a single

    component is eluted from the column. If separation is incomplete, two or more components may

    be eluted as one unresolvedpeak. Peaks eluted under optimal conditions from a well-packed,

    efficient column, operated in a system that minimizes bandspreading, approach the shape of a

    Gaussian distribution. Quantitation is usually done by measuring thepeak area [enclosed by the

    baseline and the peak curve]. Less often, peak height [the distance measured from the peak

    apex to the baseline] may be used for quantitation. This procedure requires that both the peak

    width and the peak shape remain constant.

    Plate Number* [N, see Efficiency]

    A number indicative of column performance [mechanical separation power or efficiency, also

    calledplate count, number of theoretical plates, or theoretical plate number]. It relates the

    magnitude of a peaks retention to its width [variance orbandspread]. In order to calculate a

    plate count, it is assumed that a peak can be represented by a Gaussian distribution [a

    statistical bell curve]. At the inflection points [60.7% of peak height], the width of a Gaussian

    curve is twice the standard deviation [] about its mean [located at the peak apex]. As shown in

    Figure U, a Gaussian curves peak width measured at other fractions of peak height can be

    expressed in precisely defined multiples of . Peak retention [retention volume, VR, or retention

    time, tR] and peak width must be expressed in the same units, because N is a dimensionless

    number. Note that the 5 sigma method of calculating N is a more stringent measure of column

    homogeneity and performance, as it is more severely affected by peak asymmetry. Computerdata stations can automatically delineate each resolved peak and calculate its corresponding

    plate number.

    Preparative Chromatography

    The process of using liquid chromatography to isolate a compound in a quantity and at a purity

    level sufficient for further experiments or uses. For pharmaceutical or biotechnological

    purification processes, columns several feet in diameter can be used for multiple kilograms of

    material. For isolating just a few micrograms of a valuable natural product, an analytical HPLC

    column is sufficient. Both are preparative chromatographic approaches, differing only in scale

    [see section on HPLC Scale and Table A].

    Resolution* [Rs, see Selectivity]

    The separation of two peaks, expressed as the difference in their corresponding retention times,

    divided by their average peak width at the baseline. Rs = 1.25 indicates that two peaks of equal

    width are just separated at the baseline. When Rs = 0.6, the only visual indication of the

    presence of two peaks on a chromatogram is a small notch near the peak apex. Higher efficiency

    columns produce narrower peaks and improve resolution for difficult separations;

    however, resolution increases by only the square root of N. The most powerful method of

    increasing resolution is to increase selectivityby altering the mobile/stationary phase

  • 7/27/2019 Tutorial HPLC

    32/34

    combination used for the chromatographic separation [see section on Chemical Separation

    Power].

    Retention Factor* [k]

    A measure of the time the sample component resides in the stationary phase relative to the time

    it resides in the mobile phase; it expresses how much longer a sample component is retarded by

    the stationary phase than it would take to travel through the column with the velocity of themobile phase. Mathematically, it is the ratio of the adjusted retention time [volume] and the

    hold-up time [volume]: k = tR'/tM [see Retention Time and Selectivity].

    Note: In the past, this term has also been expressed as partition ratio, capacity ratio,capacity

    factor, or mass distribution ratio and symbolized by k'.

    Retention Time* [tR]

    The time between the start of elution [typically, in HPLC, the moment of injection or sample

    introduction] and the emergence of the peak maximum. The adjusted retention time, tR', is

    calculated by subtracting from tR the hold-up time [tM, the time from injection to the elution of

    the peak maximum of a totally unretained analyte].

    Reversed-Phase Chromatography*

    An elution procedure used in liquid chromatography in which the mobile phase is significantly

    more polar than the stationary phase, e.g. a microporous silica-based material with alkyl chains

    chemically bonded to its accessible surface. Note: Avoid the incorrect term reverse phase. [See

    Reference 4 for some novel ideas on the mechanism of reversed-phase separations.]

    Selectivity [Separation Factor, ]

    A term used to describe the magnitude of the difference between the relative thermodynamic

    affinities of a pair of analytes for the specified mobile and stationary phases that comprise the

    separation system. The proper term is separation factor[]. It equals the ratio of retention

    factors, k2/k1 [see Retention Factor]; by definition, is always 1. If = 1, then both peaks

    co-elute, and no separation is obtained. It is important in preparative chromatography to

    maximize for highest sample loadability and throughput. [see section on Chemical Separation

    Power]

    Sensitivity* [S]

    The signal output per unit concentration or unit mass of a substance in the mobile phase

    entering the detector, e.g., the slope of a linear calibration curve [see Detector]. For

    concentration-sensitive detectors [e.g., UV/VIS absorbance], sensitivity is the ratio of peak

    height to analyte concentration in the peak. For mass-flow-sensitive detectors, it is the ratio of

    peak height to unit mass. If sensitivity is to be a unique performance characteristic, it must

    depend only on the chemical measurement process, not upon scale factors.

    The ability to detect [qualify] or measure [quantify] an analyte is governed by many

    instrumental and chemical factors. Well-resolved peaks [maximum selectivity] eluting from high

    efficiency columns [narrow peak width with good symmetry for maximum peak height] as well

    as good detector sensitivity and specificity are ideal. Both the separation system interference

    and electronic component noise should also be minimized to achieve maximum sensitivity.

    Solid-Phase Extraction [SPE]

    A sample preparation technique that uses LC principles to isolate, enrich, and/or purify analytes

  • 7/27/2019 Tutorial HPLC

    33/34

    from a complex matrix applied to a miniature chromatographic bed.Offline SPE is done

    [manually or via automation] with larger particles in individual plastic cartridges or in micro-

    elution plate wells, using low positive pressure or vacuum to assist flow. Online SPE is done with

    smaller particles in miniature HPLC columns using higher pressures and a valve to switch the

    SPE column online with the primary HPLC column, or offline to waste, as appropriate.

    SPE methods use step gradients [see Gradient] to accomplish bed conditioning, sample loading,washing, and elution steps. Samples are loaded typically under conditions where the kof

    important analytes is as high as possible, so that they are fully retained during loading and

    washing steps. Elution is then done by switching to a much stronger solvent mixture [see Elution

    Strength]. The goal is to remove matrix interferences and to isolate the analyte in a solution,

    and at a concentration, suitable for subsequent analysis.

    Speed [see Efficiency, Flow Rate, Resolution]

    A benefit of operating LC separations at higher linear velocities using smaller-volume, smaller-

    particle analytical columns, or larger-volume, larger-particle preparative columns. Order-of-

    magnitude advances in LC speed came in 1972 [with the use of 10 m particles and pumps

    capable of delivering accurate mobile-phase flow at 6000 psi], in 1976 [with 75-m preparative

    columns operated at a flow rate of 500 mL/min], and in 2004 [with the introduction of UPLCtechnology1.7 m-particle columns operated at 15,000 psi].

    High-speed analytical LC systems must not only accommodate higher pressures throughout the

    fluidics; injector cycle time must be short; gradient mixers must be capable of rapid turnaround

    between samples; detector sensors must rapidly respond to tiny changes in eluate composition;

    and data systems must collect the dozens of points each second required to plot and to

    quantitate narrow peaks accurately.

    Together, higher resolution, higher speed, and higher efficiency typically deliver

    higher throughput. More samples can be analyzed in a workday. Larger quantities of compound

    can be purified per run or per process period.

    See #3 on list of References for Further Reading above.

    Stationary Phase*

    One of the two phases forming a chromatographic system. It may be a solid, a gel, or a liquid. If

    a liquid, it may be distributed on a solid. This solid may or may not contribute to the separation

    process. The liquid may also be chemically bonded to the solid [bonded phase] or immobilized

    onto it [immobilized phase].

    The expression chromatographic bedor sorbentmay be used as a general term to denote any of

    the different forms in which the stationary phase is used.

    The use of the term liquid phase to denote the mobile phase in LC is discouraged. This avoids

    confusion with gas chromatography where the stationary phase is called a liquid phase [most

    often a liquid coated on a solid support].

    Open-column liquid-liquid partition chromatography [LLC] did not translate well to HPLC. It was

    supplanted by the use of bonded-phase packings. LLC proved incompatible with modern

    detectors because of problems with bleed of the stationary-phase-liquid coating off its solid

    support, thereby contaminating the immiscible liquid mobile phase.

  • 7/27/2019 Tutorial HPLC

    34/34

    UPLC Technology

    The use of a high-efficiency LC system holistically designed to accommodate sub-2 m particles

    and very high operating pressure is termed ultra-performance liquid chromatography[UPLC

    technology]. The major benefits of this technology are significant improvements in resolution

    over HPLC, and/or faster run times while maintaining the resolution seen in an existing HPLC

    separation.

    For more information, visit:


Recommended