Indication of Ultra-Wide�
SystemImmediate placement at the extraction socket
Immediate replacement of the failed implant
Delayed placement in the healed mature bone
12 US Ultra-Wide� System
13 US Ultra-Wide� fixture
15 US Ultra-Wide� abutment
16 Surgical Procedures
17 Drilling sequence
20 Insertion depth guide
Indication of Ultra-Wide� System
4 GS Ultra-Wide� System
5 GS Ultra-Wide� fixture
7 GS Ultra-Wide� abutment
8 SS Ultra-Wide� System
9 SS Ultra-Wide� fixture
1 SS Ultra-Wide� abutment
21 Surgical KIT
21 Ultra KIT
22 Instrument dimension indication
24 Instruction for Use
Table of Contents
SS Ultra-Wide�
1-Stage Internal connection
GS Ultra-Wide�
Internal connection
US Ultra-Wide�
External connection
Caution!!! When inserting Ultra-Wide fixture, maintain more than 1.5mm survival alveolar bone in order to prevent bone loss.
GS Ultra-Wide�
System
05
Features of GS Ultra-Wide£ Fixture
Ultra-Wide£ Systemwww.osstem.com
Connection
GS Ultra-Wide fixtures that have a 11 internal connection are
compatible with GS standard abutments.
Neck
Bevel neck design of fixture provides more bone volume which is
beneficial to minimizing bone resorption and support for
aesthetic gingival contour.
Body & Thread
0.8 pitch, 45° angled and 0.45 thread depth provides an
enhanced self-tapping ability.
Surface
RBM surface has good biocompatibility which induces secure
osseointegration. (Surface roughness Ra: 1.2~1.8㎛)
Apex
The thread which is existing until apex of fixture and 4 rounded cutting
edges provide progressive initial installation ability and good initial
stability in immediate placement case.
Hex 2.5mm
M2.0 Screw
∅3.35
11。
GS Ultra-Wide£ Fixture
Ultra-Wide£
System
�GS Ultra-Wide� fixture
�GS Ultra-Wide� abutment
06 Total Dental Solution Provider 07
Connection Standard
L D ∅6.0 ∅7.0
7 GS2W6007R02 GS2W7007R02
8.5 GS2W6008R02 GS2W7008R02
10 GS2W6010R02 GS2W7010R02
11.5 GS2W6011R02 GS2W7011R02
13 GS2W6013R02 GS2W7013R02
∅4.25
∅4.2
∅6.0
∅4.25
∅7.0
4
3
2
1
∅5.2
L
Order Code of GS Ultra - Wide£ Fixture
Fixture Only
- Fixture : Product Code (ex : GS2W6010R02)
Pre-Mounted Fixture
- Fixture + Simple Mount + Cover Screw : B + Product Code (ex : BGS2W6010R02)
∅6.0 fixture ∅7.0 fixture Pre-Mounted fixture
Rigid Transfer Angled Freeform ST
Freeform Convertible Combination
cylinder
Combination Angled
cylinder
GoldCast Convertible GoldCast
cylinder
Plastic cylinder
1. GS Ultra-Wide fixtures are wholly compatible with GS standard abutment components.
2. Prosthetic procedure is the same as GS prosthetic procedure.
(Refer to " Prosthetic procedure for GS Implant System")
The actual length of GS Ultra-Wide fixture is shorter by 0.5mm than the labeled length for easy control of
fixture's insertion depth. (Except for 7mm)
Important
Line-up
Ultra-Wide£ Systemwww.osstem.com
GS Ultra-Wide£ Fixture GS Ultra-Wide£ Abutment
AbutmentsType
Cement
retained
restorations
Screw
retained
restorations
09
SS Ultra-Wide�
System
Ultra-Wide£
System
�SS Ultra-Wide� fixture
�SS Ultra-Wide� abutment
Features of SS Ultra-Wide£ Fixture
Ultra-Wide£ Systemwww.osstem.com
Connection
SS Ultra-Wide fixtures that have a 8�internal connection are
compatible with SS wide abutments.
Trans-gingival section
Machined surface facilitates soft tissue management and plaque
control.
Neck
Expanded thread of fixture provides an enhanced initial stability.
Body & Thread
0.8 pitch, 45�angle and 0.45 thread depth provides an enhanced
self-tapping ability.
Surface
RBM surface has good biocompatibility which induces secure
osseointegration. (Surface roughness Ra: 1.2~1.8㎛)
Apex
The thread which is existing until apex of fixture and 4 rounded
cutting edges provide progressive initial installation ability and
good initial stability in immediate placement case.
SS Ultra-Wide£ Fixture
Octa 2.9mm
M2.0 Screw
10 Total Dental Solution Provider 11
Platform Wide
G/H 2.0
L D ∅6.0 ∅7.0
7 SS2WB62007A SS2WB72007A
8.5 SS2WB62008A SS2WB72008A
10 SS2WB62010A SS2WB72010A
11.5 SS2WB62011A SS2WB72011A
13 SS2WB62013A SS2WB72013A
Order Code of SS Ultra - Wide£ Fixture
Fixture Only
- Fixture : Product Code (ex: SS2WB62007A)
Pre-Mounted Fixture
- Fixture + Simple Mount + Cover Screw : A + Product Code (ex : ASS2WB62007A)
1. SS Ultra-Wide fixtures are wholly compatible with SS wide abutment components.
2. Prosthetic procedure is the same as SS prosthetic procedure.
(Refer to " Prosthetic procedure for SS Implant System")
Important
Line-up
Ultra-Wide£ Systemwww.osstem.com
SS Ultra-Wide£ Fixture SS Ultra-Wide£ Abutment
∅6.0 fixture ∅7.0 fixture Pre-Mounted fixture
Solid Excellent
Solid
Com Octa
Com Octa
Plus
Hanaro Com Octa
Angled
Com Octa
Gold
Octa Octa gold
cylinder
Octa plastic
cylinder
AbutmentsType
Cement
retained
restorations
Screw
retained
restorations
∅4.2
2.0
∅6.0
L
∅6.0
∅5.2
∅7.0
∅6.0
12 Total Dental Solution Provider 13
US Ultra-Wide�
System
Ultra-Wide£
System
�US Ultra-Wide� fixture
�US Ultra-Wide� abutment
Ultra-Wide£ Systemwww.osstem.com
US Ultra-Wide£ Fixture
Features of US Ultra-Wide£ fixture
Connection
US Ultra-Wide fixtures that have a 3.4mm external hex
connection are compatible with US wide abutments.
Neck
Platform switching effect of fixture minimizes bone resorption
and offers an aesthetic result.
Body & Thread
0.8 pitch, 45�angle and 0.45 thread depth provides a good self
tapping ability. Expanded thread of fixture provides an enhanced
initial stability.
Surface
RBM surface has good biocompatibility which induces secure
osseointegration. (Surface roughness Ra: 1.2~1.8㎛)
Apex
The thread which is existing until apex of fixture and 4 rounded
cutting edges provide progressive initial installation ability and
good initial stability in immediate placement case.
Hex 3.4mm
M2.5 Screw
14 Total Dental Solution Provider 15
Ultra-Wide£ Systemwww.osstem.com
US Ultra-Wide£ Fixture US Ultra-Wide£ Abutment
Line-up
Platform Wide
L D ∅6.0 ∅7.0
7 BFW6007A BFW7007A
8.5 BFW6008A BFW7008A
10 BFW6010A BFW7010A
11.5 BFW6011A BFW7011A
13 BFW6013A BFW7013A
∅5.1
∅4.2
∅6.0
∅6.0
∅7.0
∅5.2
L
Order Code of US Ultra - Wide£ Fixture
Fixture Only
- Fixture : Product Code (ex : BFW6007A)
Pre-Mounted Fixture
- Fixture + Simple Mount + Cover Screw : A + Product Code (ex : ABFW6007A)
∅6.0 fixture ∅7.0 fixture Pre-Mounted fixture
1. US Ultra-Wide fixtures are wholly compatible with US wide abutment components.
2. Prosthetic procedure is the same as US prosthetic procedure.
(Refer to "Prosthetic procedure for US Implant System")
Important
Cement Angled Safe
UCLA Esthetic-low Esthetic-low
gold cylinder
Esthetic-low
plastic cylinder
ER
AbutmentsType
Cement
retained
restorations
Combination
retained
restorations
Screw
retained
restorations
16 Total Dental Solution Provider 17
SurgicalProcedures
Ultra-Wide£
System
GS, SS and US Ultra-Wide fixture use the same drilling sequence
and surgical instruments which is included in Ultra Kit, and it
provides a convenient implant placement.
Ultra-Wide£ Systemwww.osstem.com
Drilling sequence for Ultra-Wide£ Fixture
Drilling sequence with Trephine drill
Conventional drilling sequence
∅4.2/∅5.0 ∅4.6/∅5.5
∅4.6 ∅4.6/∅5.5
Tap∅6.0
Tap∅6.0 ∅6.0
∅6.0
∅7.0
Tap∅7.0∅5.5/∅6.5 ∅7.0
∅5.2/∅6.0 ∅5.5/∅6.5 Tap∅7.0
Existing surgical
procedure application
※ When bone is especially hard(D1), the Surgical Tap must be used.
18 Total Dental Solution Provider 19
Ultra-Wide£ Systemwww.osstem.com
Surgical procedure with Trephine drill Surgical procedure with Trephine drill
∅2.0 drill
2.0/3.0 Pilot drill
∅3.0 drill
∅4.1(∅4.3) drill
∅4.6 drill
∅4.6/∅5.5
Direct drill
∅5.5 Trialpin
∅6.0 Ultra-Wide fixture
3.0/4.1 Pilot drill
10mm
1mm
∅4.2/∅5.0
Trephine drill
∅4.6/∅5.5
Direct drill
∅5.5 Trial pin
∅6.0 Ultra-Wide fixture
10mm
1mm
� Conventional drilling sequence
¤“6.0 Ultra-Wide£
fixture, Length : 10mm
� Delayed placement in the healed mature bone
¤“6.0 Ultra-Wide£
fixture, Length : 10mm
�Surgical tap is recommended in case of hard bone.
�Verify the depth with Trial pin after final drilling.
�Recommended rpm of Trephine drill : 800~1000rpm
Recommended rpm of drilling : 800~1500rpm
Recommended rpm of tapping : below 20rpm
∅4.2/∅5.0
Trephine drill
∅4.6/∅5.5
Direct drill ∅6.0 Ultra-Wide fixture
US GS SS
US GS SS
US GS SS
∅6.0 Ultra-Wide fixture
US GS SS
�
∅4.2/∅5.0
Trephine drill
∅4.0 failed
fixture
∅4.6/∅5.5
Direct drill
�
� Immediate replacement of the failed implant
¤“6.0 Ultra-Wide£
fixture, Length : 10mm
� Immediate placement at the extraction socket
¤“6.0 Ultra-Wide£
fixture, Length : 10mm
Removing the septal bone
with trephine drill
Removing failed fixture
with trephine drill
20 Total Dental Solution Provider 21
Ultra-Wide£ Systemwww.osstem.com
Insertion Depth Guide Surgical KIT
0.5~1.0
GS Ultra-Wide£ SS Ultra-Wide£ US Ultra-Wide£
Length = base on 10mm
10
0
-1
�GS Ultra-Wide£ ; 0.5~1mm Sub-Crestal
�SS Ultra-Wide£ ; Correspond with Bone Level
�US Ultra-Wide£ ; 0.25mm Supra-Crestal
� Ultra KIT(OUK)
* Lower case component - Open wrench(ASOW) for Simple mount
6.5 Direct drill : 3D6513FNLC SideCut drill :OSLMS20M
Mount driver :ASMDS1.2 Hex hand driver:
AHD12SH5.5 Direct drill : 3D5513FNLC
3 cutter twist drill : 3D4613FNLC
Surgical Tap : OAST60LC5.5 Trial pin :UWFTP55
Surgical Tap : OAST70LC
4.2/5.0 Trephine drill :TD42S
5.2/6.0 Trephine drill :TD52S6.5 Trial pin :UWFTP65
0.25
2322 Total Dental Solution Provider
� Drill & Surgical tap � Trephine drill
� Trial pin
Drill ¤“4.6
∅4
.6×
13
∅5
.5×
13
∅6
.5×
13
U/W
∅6
.0
U/W
∅7
.0
Direct drill ¤“5.5 Direct drill ¤“6.5 Tap ¤“6.0 Tap ¤“7.0
� Direct drill - 2-step drill has two functions - pilot drill & twist drill
Enables final drilling directly
without pilot drilling
Enhancement of initial stability in the
extract socket by decreasing the dead
space at the apex area
13
10
7
-1∅4.6
∅5.5
∅5.5
6
0
33.3
8.5
11.511.5
∅4.6 ∅5.5
∅5.5 ∅6.5
∅6.5
�Twin window provides clear view and copious
irrigation at target site and facilitates the
removal of bone in Trephine drill.
�You can experience a sufficient cutting
capability which is induced by 10 cutting edges
and chip groove design without spattering.
�Use for the removal of failed fixture
�Use for the removal of septal bone
�Use for the initial drilling
�It is simple to confirm 3mm drilling depth under the
extraction socket.
�Measurement of the socket's internal diameter after
removal of failed fixture.
�Use for purpose of parallel pin(¤“2.0, ¤“3.0).
Drilling Trial pin
11.5
8.5
7
10
13
6
0
21 8
6
4
2
0
11.5
8.5
6
3
0
7
10
13
31
∅4
.2 /
∅5
.0
∅5
.2 /
∅6
.0
Trial pin
¤“5.5
Trial pin
¤“6.5
∅3.0
Inner Dia. ∅4.2 ∅5.2
Outer Dia. ∅5.0 ∅6.0
5mm 3mm
Ultra-Wide£ Systemwww.osstem.com
Instrument Dimension Indication Instrument Dimension Indication
∅2.0
24 Total Dental Solution Provider 25
Surgical Procedure How to Maintain Surgical KIT
Cautions Cautions
The operator must check the following items before starting the surgery.
Patient's condition
�Sufficient bone volume and quality prior to treatment �Detailed health status
�Smoking and/or drinking habits �Masticatory pattern and habit
�Status of oral hygiene �Psychological state
�Patient's understanding of implant surgery
Treatment Plan
Discuss treatment options with your patient. Include the pros and cons of each treatment as well as the
procedure involved.
Medical Diagnosis
Check the patient’s health history before implant surgery. Discuss the treatment plan and health history with the
patient’s primary care physician before performing surgery.
Pay Particular Attention to the Following during an Implant Procedure
�Improper upper/lower posterior height �Improper lower anterior width
�Extremely poor bone quality �Congenital or acquired heart conditions
�Ischemic heart patient (angina, myocardial infarction) �High blood pressure
�Patient s distrust of implant treatment
①During surgery, be sure to keep used tools in saline or distilled
water.
② Rinse tools with distilled or running water to remove all blood and
debris.
③ Alcohol wash all tools used in surgery
Caution: Do NOT use hydrogen peroxide
*Exposure to hydrogen peroxide may cause discoloration to laser
markings and/or TiN coating
④ Remove all moisture with a dry cloth or warming fan
⑤ Replace dried tools into the Kit case
(Refer to the color-coding for easy placement.)
⑥ Autoclave Kit for 15 minutes at 132, then store at room
temperature.
Caution
�
�
�
�
�
Precautions Immediately after surgery separate, wash and store all tools.
Re-sterilize Kits prior to surgery (132 for 15 minutes)
Hanaro Kits have a 1 year product warranty (after opening); all drills and drivers are recommended for up to 50 uses.