Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ
ROCK SOLID Performance
for Trace-Level Quantitation
Agilent Technologies
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
1
Brief History of Agilent 6400 QQQ LC/MS
2
6410
2006 6460 Agilent Jet Stream
2008
6430
2009
2010
6490 iFunnel
2011
6420
2012
6460 ESI
2014
6495 iFunnel
2015
6470 AJS
Agilent 6470 and 6495 Triple Quadrupole LC/MS
• It all starts with the source!
• Improved Robustness
• Improved performance at Intrument Detection Limit (IDL Specification)
• Up to 4000 MRM transitions per method with DynamicMRM
• Simultaneous Quantitation and ID Confirmation with TriggeredMRM
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
3
Higher Throughput
Higher Confidence
Higher Productivity
+
=
Agilent 6470 and 6495 Triple Quadrupole LC/MS
• It all starts with the source!
• Improved Robustness
• Improved performance at Intrument Detection Limit (IDL Specification)
• Up to 4000 MRM transitions per method with DynamicMRM
• Simultaneous Quantitation and ID Confirmation with TriggeredMRM
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
4
Higher Throughput
Higher Confidence
Higher Productivity
+
=
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
5
6470 - 6495 QQQ Technologies Enhanced Performance in a Smaller Space
Agilent Jet Stream Technology
• Thermal gradient focusing
• Efficient desolvation
• Creates an ion rich zone
• Up to 10x gains in sensitivity
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
6
Agilent Jet Stream Technology
Unique to 6495: Proven iFunnel Technology
6/27/2016
Agilent Restricted
7
Agilent Jet Stream Hexabore Capillary Dual Ion Funnel
• Thermal gradient focusing
• Efficient desolvation
• Creates an ion rich zone
• Six capillary inlets
• Samples x10 times more ion rich gas • Removes the gas but captures the
ions
• Removes neutral noise
Agilent 6470 and 6495 Triple Quadrupole LC/MS
• It all starts with the source!
• Improved Robustness
• Improved performance at Intrument Detection Limit (IDL Specification)
• Up to 4000 MRM transitions per method with DynamicMRM
• Simultaneous Quantitation and ID Confirmation with TriggeredMRM
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
8
Higher Throughput
Higher Confidence
Higher Productivity
+
=
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
9
6470 - 6495 QQQ Technologies Enhanced Performance in a Smaller Space
Agilent Jet Stream Technology
• Thermal gradient focusing
• Efficient desolvation
• Creates an ion rich zone
• Up to 10x gains in sensitivity
• An Ion Detector with High Energy Conversion Dynode and Low Noise
• Improved ion detection 3
• A Curved and Tapered Hexapole Collision Cell
• Effective ion collection 2
• Enhanced Q1 Ion Optics
• Improved ion transmission 1
Agilent 6470 and 6495 Triple Quadrupole LC/MS
• It all starts with the source!
• Improved Robustness
• Improved performance at Intrument Detection Limit (IDL Specification)
• Up to 4000 MRM transitions per method with DynamicMRM
• Simultaneous Quantitation and ID Confirmation with TriggeredMRM
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
10
Higher Throughput
Higher Confidence
Higher Productivity
+
=
Many Ways to Manipulate S/N
6/27/2016
Agilent 6495 QQQ LC/MS System
11
Variation in S/N measurements makes direct assessment difficult
Increase signal
• Increase the gain
• Narrow chromatographic peak width
• Increase scan averaging
Lower noise
• Select noise region
• Narrow the width of noise region
• Adjust baseline
• Apply peak smoothing & noise filtering
• Vary noise calculation algorithms: Peak-to-Peak, RMS, and ...
a. Peak-to-peak noise
b. RMS noise
c. Smoothing
d. Baseline noise filtering
Instrument Detection Limit (IDL) is Defined by Statistics
6/27/2016
Agilent 6495 QQQ LC/MS System
12
IDLLCMS = t x SD = t x (%RSD / 100) x amount measured
Based on a well-established statistical formula, follows regulatory guidelines
• The minimum amount of analyte that is detectable and distinguishable from background noise with a confidence level
IDL
• Student “t” value, for
• 99% confidence level
• n – 1 degree of freedom t
• Relative standard deviation / precision of peak area at the amount measured
• From n replicate injections %RSD
• Limited to 2 – 5 x times higher than the Detection Limit (DL)
Amount measured
• Theoretical fitting of %RSD is based on
ion statistics
• %RSD increases at lower injected
amount
%RSD vs. Injected Amount
Why Add IDL Specs for QQQ LC/MS?
6/27/2016
Agilent 6495 QQQ LC/MS System
13
S/N
Many factors impact S/N: - Lower noise - Increase signal
Significant variation in S/N measurements
A relatively high level is used for S/N measurement
Fails to estimate the true limits of detection and quantitation (LLOQ)
Not a good metric of sensitivity performance
IDL (%RSD)
Access sensitivity performance from area %RSD (precision) of replicate injections
Based on a well established statistical formula – follows IUPAC / EPA guidelines
An analytical low level is used for IDL measurement – determined using calibration curve
Accurate assessment of the true limits of detection and quantitation (LLOQ)
A better and more rigorous sensitivity performance metric
Improved Sensitivity and Precision – 6495 QQQ IDL
6/27/2016
Agilent 6495 QQQ LC/MS System
14
1 fg of reserpine used to measure IDL (+) 1 fg of chloramphenicol used to measure IDL (-)
IDL = t x (%RSD / 100) x Amount measured
= 2.821 x (7.2 / 100) x 1 fg
= 0.20 fg
IDL = t x (%RSD / 100) x Amount measured
= 2.821 x (9.7 / 100) x 1 fg
= 0.27 fg
6495 QQQ IDL Amount measured Replicates Area %RSD t (99%) IDL
Reserpine (+) 1 fg n = 10 7.2 2.821 0.20 fg
Chloramphenicol (-) 1 fg n = 10 9.7 2.821 0.27 fg
IDL for the Agilent 6495 QQQ LC/MS System
1 x10
5.4
5.6
5.8
6
6.2
6.4
6.6
6.8
7
7.2
7.4
7.6
7.8
3.6 3.7 3.8 3.9 4 4.1 4.2 4.3
… 1 x10
5.4
5.6
5.8
6
6.2
6.4
6.6
6.8
7
7.2
7.4
7.6
7.8
Acquisition Time (min) 1.45 1.55 1.65 1.75 1.85 1.95 2.05
Acquisition Time (min)
IDLLCMS = t x SD = t x (%RSD / 100) x amount measured
IDL Specifications for 6400 QQQ LC/MS Agilent Differentiator!
QQQ Model 6420A 6460C 6470A 6495A
Positive
Using Reserpine
60 fg
Measured @ 100 fg
12.5 fg
Measured @ 20 fg
4.0 fg
Measured @ 10 fg
0.75 fg
Measured @ 1 fg
Negative
Using Chloramphenicol
60 fg
Measured @ 100 fg
12.5 fg
Measured @ 20 fg
4.0 fg
Measured @ 10 fg
0.75 fg
Measured @ 1 fg
Agilent 6470 and 6495 Triple Quadrupole LC/MS
• It all starts with the source!
• Improved Robustness
• Improved performance at Intrument Detection Limit (IDL Specification)
• Up to 4000 MRM transitions per method with DynamicMRM
• Simultaneous Quantitation and ID Confirmation with TriggeredMRM
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
16
Higher Throughput
Higher Confidence
Higher Productivity
+
=
Dynamic MRM (dMRM)
6/27/2016
Agilent Restricted
17
Dynamic MRM, Fast MRM Speed and Polarity Switch
• Group MRMs in RT windows instead of
time segments
• 2x shorter cycle time supports narrow
UHPLC peaks
• Supported by very fast MRM speed (1 ms
dwell) and polarity switch (20 ms)
Dynamic MRM (dMRM): Up to 4000 MRM Transitions per Run in Complex Matrix!
6/27/2016
Agilent Restricted
18
Agilent 6470 and 6495 Triple Quadrupole LC/MS
• It all starts with the source!
• Improved Robustness
• Improved performance at Intrument Detection Limit (IDL Specification)
• Up to 4000 MRM transitions per method with DynamicMRM
• Simultaneous Quantitation and ID Confirmation with TriggeredMRM
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
19
Higher Throughput
Higher Confidence
Higher Productivity
+
=
Triggered MRM (tMRM)
6/27/2016
Agilent 6495 QQQ LC/MS System
20
• tMRM database accelerates
acquisition method setup
• tMRM library – Match Score
for confirmation
• tMRM is Fast and Sensitive
• UHPLC compatible
• Simultaneous Quantitation,
Screening and Confirmation
Triggered MRM, tMRM Database and Library
Match
Score
95.1 Trigger
Threshold
tMRM Product
Ion Spectrum
Apex
Enhanced Optimizer Workflow in MassHunter B.07 (to Better Support tMRM)
• Find 10 MRM transitions in Optimizer
• Injection volume flexibility
• Improved method editing
MassHunter
Optimizer B.07
Triggered MRM (tMRM)
6/27/2016
Agilent Restricted
21
Threshold
Secondary MRM
Transitions are “Triggered”
Primary cycle (below threshold)
Triggered cycle (above threshold)
Compound Precursor Product
Analyte 1 334.0 145.0
Analyte 1 334.0 117.0
Compound Precursor Product
Analyte 1 334.0 145.0
Analyte 1 334.0 117.0
Analyte 1 334.0 132.1
Analyte 1 334.0 105.1
Analyte 1 334.0 91.1
Analyte 1 334.0 76.0
Analyte 1 334.0 119.0
QTrap Approach
• Scan the entire finger print
• Not very fast and sensitive
QQQ tMRM Approach
• Focus on ten fingerprint features (10 MRMs)
• Maintains good MDL
• Fast – UHPLC compatible
tMRM Composite Product Ion Spectrum
6/27/2016
6400 QQQ LC/MS Overview
22
x103
8
7
6
5
4
3
2
1
0 80 100 120 140 160 180 200 220 240 260 280 300 320 340
334.2 171.1
145.0
132.1
117.0
105.1 91.1 76.0
tMRM Product Ion
Spectrum
x103 8
7
6
5
4
3
2
1
0 80
76.0
x103 8
7
6
5
4
3
2
1
0 80 100
91.1
x103 8
7
6
5
4
3
2
1
0 80 100 120
105.1
x103 8
7
6
5
4
3
2
1
0 80 100 120
117.0
x103 8
7
6
5
4
3
2
1
0 100 140
132.1
x103 8
7
6
5
4
3
2
1
0 80 100 120
119.0
x103 8
7
6
5
4
3
2
1
0 100 140 180
145.0 x103
8
7
6
5
4
3
2
1
0 100 140 180
147.0
x103 8
7
6
5
4
3
2
1
0 80 100 120 140 160 180
171.1
334 > 76 334 > 91 334 > 105 334 > 117 334 > 119
334 > 132 334 > 145 334 > 147 334 > 171
tMRM Library Searching – Match Score for Confirmation
6/27/2016
6400 QQQ LC/MS Overview
23
Library match score: 96.75
LC/MS Application
Kits
Method
Standards
On Site Trainin
g
LC Column
tMRM Database & Library
tMRM Application Kits For LC/MS Targeted Screening & Confirmation with QQQ
Pesticides
Test Mix: 254 compounds
DB: 700+ compounds
Library: 200+ compounds
Veterinary Drugs
Test Mix: 146 compounds
DB: 500+ compounds
Library: 100+ compounds
Forensic Toxicology
Test Mix: 139 compounds
DB: 2500+ compounds
Library: 100+ compounds
• Agilent unique data dependent acquisition for fast and sensitive compound
screening, quantitation and confirmation.
6/27/2016
6400 QQQ LC/MS Overview
24
Applications for 6470 and 6495 QQQ LC/MS Which one is needed?
6470 6495
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
25
Food Safety - Pestcides
Large Panels, High Throughput
Environmental – Water Analysis
PPCPs in surface water using direct injection
Pharmaceutical –
High-throughput analysis of drugs & metabolites in plasma
Environmental - Water Analysis
Ultra-trace level hormones (EDCs) in drinking water using direct injection
Clinical - Peptide Quantitation
Quantitation of peptides at sub-attomole level using nanoflow and standard flow chromatography
Food Safety - Pesticides
Tough matrices, ultra-trace levels
6/27/2016
26
Summary: 6470 and 6495 QQQ LC/MS Rock Solid Performance for Confident Quantitation and Highest Lab Productivity
Additional Robustness
Improved Sensitivity
Improved Precision
Faster Speed
Smaller Footprint
6470 Upgradeable
to the 6495
Ion Optics
20KV Detector
Curved Collision Cell
Less Maintenance Cleaning
Reliable Consistent Results
over Longer Time (24/7)
Streamlined Analytical
Workflow
Reproducible
High Quality Data
Higher Throughput
at UHPLC Speed
Bench space can be optimized
Protection of
Your Investment
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
27
Merci pour votre attention!
QUESTIONS??
Enhanced Q1 Ion Optics with Optimized Prefilter Improved transmission of precursor ions and additional system robustness
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
28
1. Reduced Octopole length 1
1. New prefilter 2
1
2
• New optimized MS 1 prefilter geometry for improved precursor ion transmission
– Improved peak area response and peak area %RSD, more sensitive and precise
• New optical lens elements for reduced the probability of contamination
– More reliable and robust performance
Curved and Tapered Hexapole Collision Cell Effective collection and transmission of product ions and smaller footprint
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
29
• Curved and Tapered Hexapole Assembly for efficient collection and transmission of
product ions
• Designed for consistent collision energies across all QQQ platforms
• A compact, smaller benchtop footprint
6460 QQQ LC/MS
- Linear Collision Cell
6470 QQQ LC/MS
- Curved Collision Cell
30% smaller
New Detector with High Energy Conversion Dynode More efficient detection and quantitation of ions with low noise characteristics
6/27/2016
Customer Facing Presentation - The New 6470 QQQ LC/MS
System
30
• Improved ion detection efficiency with High Energy Dynode (HED) voltage up to 20 kV
- Improved peak area response and peak area %RSD in positive & negative ion mode
- Improved sensitivity and precision for a wide mass range
• Low noise level at 20 kV
- Improved signal to noise
40 – 113% increase in
response to 20kV vs. 10kV
iFunnel Express iFunnel Express – Source and iFunnel Optimization
Method Generation Acquisition Optimization
• Automate the source/iFunnel
optimization process
Create
Methods
& Worklist
MH Source
and Funnel
Optimizer
Acquire Data
for Each
Parameter
MassHunter
Quant
Data
Files Integrate for
Peak Data
Excel File for
Optimized
Parameter
6/27/2016
Agilent 6495 QQQ LC/MS System
31