Date post: | 18-Feb-2017 |
Category: |
Education |
Upload: | navni-garg |
View: | 727 times |
Download: | 5 times |
ULTRASOUND PHYSICSTRANSDUCERS
TRANSDUCER JELLY
SEMINAR BY
DR NAVNI GARG
CHARACTERISTICS OF SOUND
• A sound beam is similar to x-ray beam in that both are waves transmitting energy but important difference is that x-rays pass through a vacuum where as sound require a material medium ( solid , liquid , gas ) for transmission, they will not pass through the vacuum.
• Sound must be generated mechanically by vibrating body matter
HISTORY OF ULTRASOUND
• Piezoelectricity discovered by Pierre and Jacques Curie in 1880 using natural quartz• SONAR was first used in 1940s war
time• Diagnostic medical applications in
use since late 1950’s
• A : uniform distribution of molecules in a medium
• B: movement of the piston to the right produces a zone of compression
• C: withdrawl of the piston to left produces a zone of rarefraction
• D: alternate movement of the piston to right and left establishes a longitudinal wave in the medium
COMPRESSION WAVE
PROPERTIES OF SOUND WAVE
• Ultrasound obeys the wave equation : u = vλ where v = frequency
( Hz , number of cycles / sec. ) u = velocity of sound ( meter / sec.) λ = the wavelength ( which is distance between two successive
compression. meter.)
FREQUENCY
• Frequency refers to the number of cycles of compressions and rarefactions in a sound wave per second, with one cycle per second being 1 hertz. • Medically used ultrasound involves 1-10
MHz frequencies .(1 -10 million Hz /sec.).
WAVELENGTH
• The wavelength is the distance traveled by sound in one cycle, or the distance between two identical points in the wave cycle i.e. the distance from a point of peak compression to the next point of peak compression.
• It is inversely proportional to the frequency.
Wavelength is one of the main factors affecting axial resolution of an ultrasound image
• Smaller wavelength• Higher frequency• Higher resolution• Lesser penetration• Therefore, higher frequency
probes (5 to 10 MHz) provide better resolution but can be applied only for superficial structures and in children.
• Higher wavelength• Lesser frequency• Less resolution• Deeper penetration• Lower frequency probes (2
to 5MHz) provide better penetration albeit lower resolution and can be used to image deeper structures.
PROPAGATION VELOCITY
• The propagation velocity is the velocity at which sound travels through a particular medium
• Dependant on the compressibility and density of the medium.
• The average velocity of sound in soft tissues such as the chest wall and heart is 1540 metres/second.
COMPRESSIBILITY
• The velocity of sound is inversely related to the compressibility of the conducting material. That means less compressibility of material , the more rapidly transmits the sound.
• Sound waves move slowly in the gas because the molecules are far apart and are easily compressed.
• Solids > liquids > gases
DENSITY
• Dense materials have large molecules with large inertia : difficult to move or stop once in motion
• Propagation of sound requires rhythmic starting and stopping of particles
• Density is inversely related to velocity
AMPLITUDE/INTENSITY
• It is a measure of the degree of change within a medium, caused by the passage of a sound wave and relates to the severity of the disturbance
• Determined by the length of oscillation of particle
• Greater amplitude = more intense sound
• Sound intensity is measured in decibel (dB).• Ultrasonic intensities are expressed
in power / unit area (watts/cm2)
TRANSDUCER
• Transducer is the device which generates ultrasound wave .
• Transducers are used to convert an electric signal into ultrasonic energy that can be transmitted into tissue , and to convert ultrasonic energy reflected back from the tissue into an electric signal.
COMPOSITION OF TRANSDUCER
• The most important component is a thin (0.5 mm) piezoelectric crystal element located near the face of the transducer .
• The piezoelectric crystal consist of lead zirconate titanate or PZT.
• The front and back faces of the crystal are coated with a thin conducting film to ensure good contact with the two electrodes that will supply the electric field used to strain the crystal.
• Crystal is made up of numerous dipoles arranged in a geometric pattern.
• Dipole is a polarized molecule, one end positive and other end negative .
• The positive and negative ends arranged so that an electric field will cause them to realign thus changing the dimensions of the crystal.
• No current flows through the crystal• Plating electrodes behave as capacitors
and it is the voltage between them that produces the electric field which causes change in crystal shape
• When the high frequency voltage pulse is applied across the crystal , the crystal vibrates like a cymbal that has been struck a sharp blow and generates sound waves.
• The backing block must stop the crystal vibration within a microsecond because the transducer must be ready immediately to receive reflected waves (echoes) from tissue interface.
• As the sound pulse passes through the body ,echoes reflect back towards the transducer from each tissue interface. These echoes carry energy and they transmit their energy to the transducer , causing a physical compression of the crystal element . This compression forces the tiny dipoles to change their orientation , which induces a voltage between the electrodes
• The voltage is amplified and serves as the ultrasonic signal for display on television monitor.
• Compression force and associated voltage are responsible for the name piezoelectricity which means “ pressure “ electricity.
• Naturally occurring materials possess piezoelectric properties : Quartz
• Man made material ( ferroelectrics ) : Barium titanate lead zirconate titanate
• Curie Temperature : is the temperature at which polarization is lost.
• Heating the piezoelectric crystal above the Curie temperature reduces it to a useless piece of ceramic so transducer should never be autoclaved.
Resonant frequency :• The thickness of piezoelectric crystal determines
its natural frequency called its resonant frequency.
• The crystal is designed so that its thickness is equal to exactly half the wavelength of the ultrasound to be produced by the transducers.
• Thickness = wavelength/2
Transducer Q Factor :• Two characteristics :- purity of sound & the length of
time that the sound persists.• A high Q transducer produces a nearly pure sound
made up of narrow range of frequencies.• A low Q transducer produces whole spectrum of sound
covering wider range of frequencies.• The interval between initiation of the wave and
complete cessation of vibration is called the “ ring down time “.
• High Q : useful for doppler USG transducers because it furnishes narrow range of sound frequencies
• Low Q : useful for organ imaging because it can furnish short ultrasound pulses and will respond to a broad range of returning frequencies
• The Q factor can be controlled by altering the characteristic of the damping block.
• Damping block consist of powered rubber and tungsten blended with an epoxy resin.
• Ratio of tungsten to resin is chosen to satisfy the impedance requirements
• Rubber is added to increase the attenuation of sound in the backing block.
RECEPTION OF ULTRASOUND
• 1. Reflection :• Both ultrasound and light obey the law of
reflection , the angle of incidence and the angle of reflection are equal.
• The factor that determines the percent of the incident beam undergoing reflection is a property , peculiar to various tissues , called acoustic impendence
• Acoustic impendence Z = p u rayls• where p is density , u is velocity
of sound in cm/sec.• The velocity of sound in all soft tissue is
virtually same 1540 m/sec.• So , Z α p. example air and bone.
• As sound waves pass from one tissue to another , the amount of reflection is determined by the difference in the impedances of the two tissues .
• At a particular angle of incidence known as the critical angle , total reflection occurs at the skin
REFRACTION
• This occurs when an ultrasound beam passes, at an angle other than 90 degrees, from one tissue into another with change in velocity.
• It increase with the increasing angle of incidence .
• It passes deeper into the body where it gives rise to artifacts.
• If angle of incidence is less than 3 degrees, very little refraction seen.
ABSORPTION
• Due to friction among molecules in their back –forth movement , reduction in intensity of the ultrasound beam occurs as it traverse matter. Friction results in degradation of part of molecules kinetic energy to heat.
• The greater the frequency , the greater the attenuation coefficient. This means high frequency beam shows less penetration than a low frequency beam.
• Attenuation in soft tissue is 1 dB/cm/MHz
TYPES OF ELECTRONIC SCANNING
LINEAR ARRAY PHASED ARRAY
TYPES OF TRANSDUCERS
The ultrasound transducers differ in construction according to
• Piezoelectric crystal arrangement• Aperture ( footprint )• Operating frequency ( which is directly related
to the penetration depth )
SECTOR TRANSDUCER
• Crystal arrangement : phased array• Footprint size : small• Operating frequency : 1-5 MHz• Ultrasound beam shape : sector, almost
triangular• Use : small acoustic windows ,mainly ECHO,
gynecological ultrasound, upper body ultrasound
LINEAR TRANSDUCER
• Crystal arrangement : linear• Footprint size: usually big ( small for hockey
transducers )• Operating frequency : 3-12 MHz• Ultrasound beam shape : rectangular• Use : USG of superficial structures e.g.
obstetrics ultrasound , breast,thyroid,vascular ultrasound
CONVEX TRANSDUCER
• Crystal arrangement : curvilinear • Footprint size : big ( small for the micro convex
transducers )• Operating frequency : 1-5 MHz• Ultrasound beam shape • Use : useful in all USG types except ECHO,
typically abdominal ,pelvic and lung ( micro convex transducer )
TRANSDUCER JELLY/COUPLING AGENT
• Air and other gases impede sound waves• At tissue-air interface, more than 99.9% of the
beam is reflected so none is available for further imaging
• Jelly acts as a special aqueous conductive medium for the sound waves
• Prevents the formation of bubbles between the transducer and the patient’s skin
• Acts as a lubricant
PROPERTIES
• Non allergenic• Odourless• Non staining• Harmless• Neutral ph• Easily removable with tissue or towel
USG GEL INGREDIENTS• Water• Carbomer : synthetic high molecular weight polymer of
acrylic acid cross linked with allyl sucrose and containing 50-68% of carboxylic acid groups. Neutralized with alkali hydroxide to make it water soluble.
• EDTA• Propylene glycol : organic oil compound that doesnot irritate
the skin and helps retain moisture• Glycerine and trolamine : neutral colorless gel that absorbs
moisture from air• Colorant : occasionally used, usually blue color
Thank you