+ All Categories
Home > Documents > Unit Extra PDC

Unit Extra PDC

Date post: 02-Apr-2018
Category:
Upload: nirmal-kumar-pandey
View: 218 times
Download: 0 times
Share this document with a friend

of 261

Transcript
  • 7/27/2019 Unit Extra PDC

    1/261

    Fundamentals of Instrumentation& Process Control

    Interactive Training Workshop

  • 7/27/2019 Unit Extra PDC

    2/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Fundamentals of Instrumentation & Control

    2

  • 7/27/2019 Unit Extra PDC

    3/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Introduction to Process Control

    A common misconception in process control is that it is allabout the controller that you can force a particularprocess response just by getting the right tuningparameters.In reality, the controller is just a partner. A process will

    respond to a controllers commands only in the mannerwhich it can. To understand process control you mustunderstand the other partners as well: sensors, finalcontrol elements and the process itself.All of these determine what type of response the controlleris capable of extracting out of the process. It is not theother way around.

    3

  • 7/27/2019 Unit Extra PDC

    4/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Outline of the Course

    4

  • 7/27/2019 Unit Extra PDC

    5/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Introduction to Process Control

    Objectives:Why do we need process control?What is a process?What is process control?

    What is open loop control?What is closed loop control?What are the modes of closed loop control?What are the basic elements of process control?

    5

  • 7/27/2019 Unit Extra PDC

    6/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Motivation for Automatic Process Control

    Safety First:people, environment, equipment

    The Profit Motive:meeting final product specsminimizing waste productionminimizing environmental impactminimizing energy usemaximizing overall production rate

    6

  • 7/27/2019 Unit Extra PDC

    7/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Loose Control Costs Money

    It takes more processing to remove impurities, so greatestprofit is to operate as close to the maximum impuritiesconstraint as possible without going over

    It takes more material to make a product thicker, so greatestprofit is to operate as close to the minimum thicknessconstraint as possible without going under

    Copyright 2007by Control Station, Inc.

    All Rights Reserved

    SPSP P V & S P ( % )

    65

    60

    55

    45

    40

    Time

    operating constraint

    m o r e p r o

    f i t

    poor control = large variation in PV

    set point far fromprofit constraint

    7

  • 7/27/2019 Unit Extra PDC

    8/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Tight Control is More Profitable

    A well controlled process has less variability in the measuredprocess variable (PV), so the process can be operated closeto the maximum profit constraint.

    Copyright 2007 by Control Station, Inc. All Rights Reserved m o r e p r o

    f i t

    SPSP

    Time

    65

    60

    50

    45

    40

    P V & S P ( % )

    operating constraint

    tight control = small variation in PV

    set point near profit constraint

    8

  • 7/27/2019 Unit Extra PDC

    9/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Terminology for Home Heating Control

    Control Objective

    Measured Process Variable (PV)Set Point (SP)Controller Output (CO)Manipulated VariableDisturbances (D)

    furnace

    temperaturesensor/transmitter

    set pointheat loss

    (disturbance)

    thermostatcontroller

    valve

    TTTC

    controlsignal

    fuel flow

    Copyright 2007 by Control Station, Inc. All Rights Reserved

    9

  • 7/27/2019 Unit Extra PDC

    10/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Process Control

    Terminology:The manipulated variable (MV) is a measure of resourcebeing fed into the process, for instance how much thermalenergy.A final control element (FCE) is the device that changes

    the value of the manipulated variable.The controller output (CO) is the signal from thecontroller to the final control element.The process variable (PV) is a measure of the processoutput that changes in response to changes in the

    manipulated variable.The set point (SP) is the value at which we wish tomaintain the process variable at.

    10

  • 7/27/2019 Unit Extra PDC

    11/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is a Process

    A process is broadly defined as an operationthat uses resources to transform inputs intooutputs.It is the resource that provides the energy into

    the process for the transformation to occur.

    11

  • 7/27/2019 Unit Extra PDC

    12/261

  • 7/27/2019 Unit Extra PDC

    13/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is a Process

    Each process exhibits a particular dynamic (timevarying) behavior that governs thetransformation.

    That is, how do changes in the resource or inputsover time affect the transformation.

    This dynamic behavior is determined by thephysical properties of the inputs, the resource,and the process itself.

    13

  • 7/27/2019 Unit Extra PDC

    14/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is a Process

    Can you identify some of the elements that willdetermine the dynamic properties of thisprocess?

    14

  • 7/27/2019 Unit Extra PDC

    15/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Process Control

    Process control is the act of controlling a finalcontrol element to change the manipulatedvariable to maintain the process variable at adesired set point.

    A corollary to our definition of process control is acontrollable process must behave in a predictablemanner.For a given change in the manipulated variable, theprocess variable must respond in a predictable and

    consistent manner.

    15

  • 7/27/2019 Unit Extra PDC

    16/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Process Control

    16

  • 7/27/2019 Unit Extra PDC

    17/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Process Control

    17

  • 7/27/2019 Unit Extra PDC

    18/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Section Assessment:Basic Terminology Assessment

    18

  • 7/27/2019 Unit Extra PDC

    19/261

  • 7/27/2019 Unit Extra PDC

    20/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Open Loop Control

    20

  • 7/27/2019 Unit Extra PDC

    21/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Open Loop Control

    Can you think of processes in which open loopcontrol is sufficient?

    21

  • 7/27/2019 Unit Extra PDC

    22/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Closed Loop Control

    In closed loop control the controller output isdetermined by difference between the processvariable and the set point. Closed loop control isalso called feedback or regulatory control.

    The output of a closed loop controller is a function of the error.Error is the deviation of the process variable fromthe set point and is defined asE = SP - PV.

    22

  • 7/27/2019 Unit Extra PDC

    23/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Closed Loop Control

    23

  • 7/27/2019 Unit Extra PDC

    24/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Closed Loop Control

    From the controllersperspective theprocess encompassesthe RTD, the steamcontrol valve, and

    signal processingof the PV andCO values.

    24

  • 7/27/2019 Unit Extra PDC

    25/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop Control

    Closed loop control can be, depending on thealgorithm that determines the controller output:ManualOn-Off PIDAdvanced PID (ratio, cascade, feedforward)or Model Based

    25

  • 7/27/2019 Unit Extra PDC

    26/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlManual Control

    In manual control an operator directlymanipulates the controller output to the finalcontrol element to maintain a set point.

    26

  • 7/27/2019 Unit Extra PDC

    27/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlOn-Off Control

    On-Off control provides a controller output of either on or off in response to error.

    27

  • 7/27/2019 Unit Extra PDC

    28/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlOn-Off Control Deadband

    Upon changing the direction of the controlleroutput, deadband is the value that must betraversed before the controller output will changeits direction again.

    28

  • 7/27/2019 Unit Extra PDC

    29/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlPID Control

    PID control provides a controller output thatmodulates from 0 to 100% in response to error.

    29

  • 7/27/2019 Unit Extra PDC

    30/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlTime Proportion Control

    Time proportion control is a variant of PID control thatmodulates the on-off time of a final control element that onlyhas two command positions.

    To achieve the effect of PIDcontrol the switching frequencyof the device is modulated inresponse to error. This is

    achieved by introducing theconcept of cycle time.

    Cycle Time is the time base of thesignal the final control element will

    receive from the controller. The PID

    controller determines the final signalto the controller by multiplying thecycle time by the output of the PID

    algorithm.

    30

  • 7/27/2019 Unit Extra PDC

    31/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlCascade Control

    Cascade control uses the output of a primary (master orouter) controller to manipulate the set point of a secondary(slave or inner) controller as if the slave controller were thefinal control element.

    31

  • 7/27/2019 Unit Extra PDC

    32/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are the Modes of Closed Loop ControlCascade Control

    32

  • 7/27/2019 Unit Extra PDC

    33/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Basic Elements of Process Control

    Controlling a process requires knowledge of four

    basic elements, the process itself, the sensor that measures the process value, the final control element that changes the manipulatedvariable, and the controller .

    33

  • 7/27/2019 Unit Extra PDC

    34/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Section Assessment:Basic Process Control Assessment

    34

  • 7/27/2019 Unit Extra PDC

    35/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Dynamic Process Behavior What It Is & Why We Care

    What a FOPDT Dynamic Model Represents

    Analyzing Step Test Plot Data to Determine FOPDTDynamic Model ParametersProcess Gain, Time Constant & Dead Time

    How To Compute Them From Plot DataHow to Use Them For Controller Design and Tuning

    How to Recognize Nonlinear Processes

    What We Will Learn in This Section

    Understanding Dynamic Process Behavior

    35

  • 7/27/2019 Unit Extra PDC

    36/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Dynamic Process Behavior and Controller Tuning

    Consider cruise control for a car vs a truckhow quickly can each accelerate or deceleratewhat is the effect of disturbances (wind, hills, etc.)

    Controller (gas flow) manipulations required to maintainset point velocity in spite of disturbances (wind, hills)are different for a car and truck because the dynamicbehavior of each "process" is different

    Dynamic behavior how the measured process variable(PV) responds over time to changes in the controlleroutput (CO) and disturbances (D)

    36

  • 7/27/2019 Unit Extra PDC

    37/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Graphical Modeling of Dynamic Process Data

    To learn about the dynamic behavior of a process, weanalyze measured process variable (PV) test data

    PV test data can be generated by suddenly changing thecontroller output (CO) signal

    The CO should be moved far and fast enough so that thedynamic behavior is clearly revealed as the PV responds

    The dynamic behavior of a process is different as operatinglevel changes (nonlinear behavior), so collect data atnormal operating conditions (design level of operation)

    37

  • 7/27/2019 Unit Extra PDC

    38/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 38

    Modeling Dynamic Process Behavior

    The best way to understand process data is through modeling

    Modeling means fitting a first order plus dead time (FOPDT)dynamic model to the process data:

    where:PV is the measured process variableCO is the controller output signal

    The FOPDT model is simple (low order and linear) so it onlyapproximates the behavior of real processes

    dPVp + PV = Kp CO(t p)

    dt

  • 7/27/2019 Unit Extra PDC

    39/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 39

    Modeling Dynamic Process Behavior

    When a first order plus dead time (FOPDT) modelis fit to dynamic process data

    The important parameters that result are:Steady State Process Gain, KpOverall Process Time Constant,Apparent Dead Time, p

    dPVp + PV = Kp CO(t p)

    dt

    p

  • 7/27/2019 Unit Extra PDC

    40/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 40

    The FOPDT Model is All Important

    FOPDT model parameters (Kp, and p ) are used in

    correlations to compute controller tuning valuesSign of Kp indicates the action of the controller

    (+ Kp reverse acting; Kp direct acting)

    Size of indicates the maximum desirable loop sample time(be sure sample time T 0.1 )

    Ratio of p / indicates whether model predictive controlsuch as a Smith predictor would show benefit

    (useful if p > )

    Model becomes part of the feed forward, Smith Predictor,decoupling and other model-based controllers

    p

    p

    p

    p

    p

  • 7/27/2019 Unit Extra PDC

    41/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 41

    Step Test Data and Dynamic Process Modeling

    Process starts at steady state in manual modeController output (CO) signal is stepped to new valueProcess variable (PV) signal must complete the response

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    Manual Mode Step Test on Gravity Drained Tanks

  • 7/27/2019 Unit Extra PDC

    42/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 42

    Process Gain (Kp) from Step Test Data

    Kp describes how far the measured PV travels inresponse to a change in the COA step test starts and ends at steady state, so Kp canbe computed directly from the plot

    where PV and CO are the total change from initial tofinal steady state

    A large process gain, Kp, means that each CO actionwill produce a large PV response

    Steady State Change in the Process Variable, PVKp

    Steady State Change in the Controller Output, CO

  • 7/27/2019 Unit Extra PDC

    43/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 43

    Process Gain (Kp) for Gravity-Drained Tanks

    Compute PV and CO as final minus initial steady state values

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    PV = (2.9 1.9) = 1.0 m

    CO = (60 50) = 10 %

  • 7/27/2019 Unit Extra PDC

    44/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 44

    Process Gain (Kp) for Gravity-Drained Tanks

    Kp has a size (0.1); a sign (+0.1), and units (m/%)

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    PV = 1.0 m

    CO = 10 %

    Kp = = = 0.1 PV 1.0 m mCO 10% %

  • 7/27/2019 Unit Extra PDC

    45/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Additional Notes on Process GainMeasuring the Process Gain

    Process gain as seen by a controller is theproduct of the gains of the sensor, the finalcontrol element and the process itself.

    The gain of a controller will be inverselyproportional to the process gain that it sees

    ElementControlFinalGainSensor GainProcessGainGainProcess x x

    GainProcess

    1GainController

    45

  • 7/27/2019 Unit Extra PDC

    46/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Additional Notes on Process GainConverting Units of Process Gain

    There is one important caveat in this process;the gain we have calculated has units of m/%.Real world controllers, unlike most softwaresimulations, have gain units specified as %/%.When calculating the gain for a real controller thechange in PV needs to be expressed in percent of span of the PV as this is how the controllercalculates error.

    Span PV SpanCO x

    CO PV Gain

    46

  • 7/27/2019 Unit Extra PDC

    47/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Assuming the process gain of the gravitydrained tanks is 0.1 m/% then, to convertto %/%

    Additional Notes on Process GainConverting Units of Process Gain

    %%

    0.1%100%1000

    %1.0PVCO

    %1.0m

    span spanm

    47

  • 7/27/2019 Unit Extra PDC

    48/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Additional Notes on Process Gain Values for Process Gain

    Process gain that the controller sees is influencedby two factors other than the process itself, thesize of the final control element and the span of the sensor.In the ideal world you would use the full span of both final control element and the sensor whichwould give a process gain of 1.0.

    48

  • 7/27/2019 Unit Extra PDC

    49/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Additional Notes on Process Gain Values for Process Gain

    As a rule of thumb:Process gains that are greater than 1 are a result of oversized final control elements.Process gains less than 1 are a result of sensor spansthat are too wide.

    49

  • 7/27/2019 Unit Extra PDC

    50/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Additional Notes on Process Gain Values for Process Gain

    The result of a final control element being too large (highgain) is:

    The controller gain will have to be made correspondingly smaller,smaller than the controller may accept.High gains in the final control element amplify imperfections(deadband, stiction), control errors become proportionately

    larger .If a sensor has too wide of a span:

    You may experience problems with the quality of themeasurement.The controller gain will have to be made correspondingly largermaking the controller more jumpy and amplifying signal noise.An over spanned sensor can hide an oversized final element.

    50

  • 7/27/2019 Unit Extra PDC

    51/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Additional Notes on Process Gain Values for Process Gain

    The general rule of thumb is theprocess gain for a self regulating

    process should be between0.5 and 2.0.

    51

  • 7/27/2019 Unit Extra PDC

    52/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 52

    Process Time Constant ( ) from Step Test Data

    Time Constant, , describes how fast the measured PVresponds to changes in the COMore specifically, how long it takes the PV to reach 63.2%of its total final change ( PV), starting from when it firstbegins to respond

    p

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    p

  • 7/27/2019 Unit Extra PDC

    53/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 53

    1) Locate , the time where the PV starts afirst clear response to the step change in CO

    Process Time Constant ( ) from Step Test Datap

    t PVstart

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    tPVstarttPVstart

  • 7/27/2019 Unit Extra PDC

    54/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 54

    Process Time Constant ( ) from Step Test Data

    2) Compute 63.2% of the total change in PV as:PV63.2 = PV inital + 0.632 PV

    p

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    63.2% of PV total

    PV 63.2 PV total

    tPVstarttPVstart

    f

  • 7/27/2019 Unit Extra PDC

    55/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 55

    Process Time Constant ( ) from Step Test Data

    3)4) Time Constant, , is then:

    p

    t 63.2 tPVstart p t 63.2 is the time when the PV reaches PV 63.2

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    P

    PV 63.2

    tPVstarttPVstart t63.2t63.2

    ( ) f d k

  • 7/27/2019 Unit Extra PDC

    56/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 56

    Time Constant ( ) for Gravity-Drained Tanksp

    Here, = 4.1 mintPVstart

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    tPVstart = 4.1tPVstart = 4.1

    Ti C ( ) f G i D i d T k

  • 7/27/2019 Unit Extra PDC

    57/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 57

    PV63.2 = PV inital + 0.632 PV

    = 1.9 + 0.632(1.0) = 2.5 m

    Time Constant ( ) for Gravity-Drained Tanksp

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    63.2% of PV total

    PV = 2.563.2 PV = 1.0 m

    4.1

    Ti C ( ) f G i D i d T k

  • 7/27/2019 Unit Extra PDC

    58/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 58

    Time Constant ( ) for Gravity-Drained Tanksp

    = t 63.2 tPVstart = 1.6 minp Time Constant,must be positive and have units of timep

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    P

    4.1 5.7

    PV = 2.563.2

    = 5.7 4.1 = 1.6 min= 5.7 4.1 = 1.6 min P

    P D d Ti ( ) f S T D

  • 7/27/2019 Unit Extra PDC

    59/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 59

    Process Dead-Time ( p) from Step Test Data

    Dead time, p, is how much delay occurs from the time whenthe CO step is made until when the measured PV shows afirst clear response.

    p is the sum of these effects:

    transportation lag, or the time it takes for material totravel from one point to another

    sample or instrument lag, or the time it takes to collect,analyze or process a measured PV sample

    higher order processes naturally appear slow to respondand this is treated as dead time

    Dead time, p, must be positive and have units of time

    D d Ti ( ) i h Kill f C l

  • 7/27/2019 Unit Extra PDC

    60/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 60

    Dead-Time ( p) is the Killer of Control

    Tight control grows increasingly difficult as p becomes large

    The process time constant is the clock of the process. Deadtime is large or small relative only to

    When dead time grows such that p > , model predictivecontrol strategies such as a Smith predictor may show benefit

    For important PVs, work to select, locate and maintaininstrumentation so as to avoid unnecessary dead time in a loop

    p

    p

    P D d Ti ( ) f St T t D t

  • 7/27/2019 Unit Extra PDC

    61/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 61

    Process Dead-Time ( p) from Step Test Data

    p = t PVstart tCOstep

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    ptCOsteptCOstep

    tPVstarttPVstart

    D d Ti ( ) f G it D i d T k

  • 7/27/2019 Unit Extra PDC

    62/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 62

    Dead-Time ( p) for Gravity Drained Tanks

    p = t PVstart tCOstep = 0.3 min

    Copyright 2007 by Control Station, Inc. All Rights ReservedCopyright 2007 by Control Station, Inc. All Rights Reserved

    = 4.1tCOstep= 3.8

    tCOstep= 3.8

    tPVstarttPVstart

    p = 4.1 3.8 = 0.3 min p = 4.1 3.8 = 0.3 min

    Th FOPDT M d l P t

  • 7/27/2019 Unit Extra PDC

    63/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 63

    The FOPDT Model Parameters

    Process Gain, Kp How Far PV travels

    Time Constant, How Fast PV responds

    Dead Time, p How Much Delay Before PV Responds

    For a change in CO:

    p

    In Summary

    H d O W k h

  • 7/27/2019 Unit Extra PDC

    64/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 64

    Hands-On Workshop

    Workshop #1Exploring Dynamics of Gravity-Drained Tanks

    Processes have Time Varying Behavior

  • 7/27/2019 Unit Extra PDC

    65/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 65

    Processes have Time Varying Behavior

    The CO to PV behavior described by an ideal FOPDT model is

    constant, but real processes change every day because:surfaces foul or corrodemechanical elements like seals or bearings wearfeedstock quality varies and catalyst activity decaysenvironmental conditions like heat and humidity change

    So the values of Kp, and p that best describe the dynamicbehavior of a process today may not be best tomorrow

    As a result, controller performance can degrade with time andperiodic retuning may be required

    p

  • 7/27/2019 Unit Extra PDC

    66/261

    Processes have Nonlinear Behavior

  • 7/27/2019 Unit Extra PDC

    67/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 67

    Processes have Nonlinear Behavior

    A FOPDT model response is constant as operating level changes

    Since the FOPDT model is used for controller design and tuning, aprocess should be modeled at a specific design level of operation!

    2) FOPDT model response isconstant as operating level changes

    1) with equal CO steps

    3) but behavior of real processeschanges with operating level

    Copyright 2007 by Control Station, Inc. All Rights Reserved

    What is a Nonlinear Process

  • 7/27/2019 Unit Extra PDC

    68/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is a Nonlinear Process

    While linear processes may be the design goal of processengineers, the reality is that most processes are nonlinearin nature due to nonlinearity in the final control element orthe process itself.

    Example: A heating process is nonlinear because the rateat which heat is transferred between two objects depends

    on the difference in temperature between the objects.Example: A valve that is linear in the middle of itsoperating range may become very nonlinear towards itslimits.

    68

    What is a Nonlinear Process

  • 7/27/2019 Unit Extra PDC

    69/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is a Nonlinear ProcessDealing with Nonlinearity Set Point Response

    69

    What is a Nonlinear Process

  • 7/27/2019 Unit Extra PDC

    70/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is a Nonlinear ProcessDealing with Nonlinearity

    The robustness of a controller is a measure therange of process values over which the controllerprovides stable operation.The more nonlinear a process is, the lessaggressive you must be in your tuning approachto maintain robustness.

    70

    What is Process Action

  • 7/27/2019 Unit Extra PDC

    71/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What is Process Action

    Process action is how the process variable changes withrespect to a change in the controller output. Process actionis either direct acting or reverse acting.The action of a process is defined by the sign of theprocess gain. A process with a positive gain is said to bedirect acting. A process with a negative gain is said to bereverse acting.

    71

  • 7/27/2019 Unit Extra PDC

    72/261

  • 7/27/2019 Unit Extra PDC

    73/261

    Fundamentals of Instrumentation& Process Control

    Interactive Training Workshop

    Introduction to Instrumentation

  • 7/27/2019 Unit Extra PDC

    74/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Introduction to Instrumentation

    The intent of this chapter is neither to teach you how toselect a particular instrument nor to familiarize you with all of the available types of instruments.

    The intent of this material is to provide an introduction tocommonly measured process variables, including the basic terminology and characteristics relevant to each variablesrole in a control loop.

    Detailed information and assistance on device selection istypically available directly from the instrumentationsupplier.

    74

    Introduction to Instrumentation

  • 7/27/2019 Unit Extra PDC

    75/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Introduction to Instrumentation

    Objectives:What are sensors and transducers?What are the standard instrument signals?What are smart transmitters?What is a low pass filter?

    What instrument properties affect a process?What is input aliasing?What is instrument noise?How do we measure temperature?How do we measure level?

    How do we measure level?How do we measure flow?

    75

    Introduction to Instrumentation

  • 7/27/2019 Unit Extra PDC

    76/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Introduction to Instrumentation

    You cannot control what youcannot measure

    When you can measure what you are speaking about,and express it in numbers, you know somethingabout it. When you cannot measure it, when youcannot express it in numbers, your knowledge is of a

    meagre and unsatisfactory kind.

    - L o r d K e lv i n

    76

    What are Sensors and Transducers

  • 7/27/2019 Unit Extra PDC

    77/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Sensors and Transducers

    A sensor is a device that has a characteristic that changesin a predictable way when exposed to the stimulus it wasdesigned to detect.A transducer is a device that converts one form of energyinto another.

    77

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    78/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Standard Instrumentation Signals

    Standard instrument signals for controllers toaccept as inputs from instrumentation andoutputs to final control elements are:

    pneumatic current loop0 to 10 volt

    78

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    79/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Standard Instrumentation SignalsPneumatic

    3 to 15 psigBefore 1960, pneumatic signals were used almostexclusively to transmit measurement and controlinformation.Today, it is still common to find 3 to 15 psig used as

    the final signal to a modulating valve.Most often an I/P (I to P) transducer is used.

    This converts a 4-20 mA signal (I) into a pressuresignal (P).

    79

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    80/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Standard Instrumentation SignalsPneumatic Scaling

    What would our pneumatic signal be if ourcontroller output is 40%?

    psig3 psig12OutputController % psigSignal x

    80

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    81/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Standard Instrumentation SignalsCurrent Loop

    4-20 milliampCurrent loops are the signal workhorses in ourprocesses.A DC milliamp current is transmitted through a pairof wires from a sensor to a controller or from a

    controller to its final control element.Current loops are used because of their immunity tonoise and the distances that the signal can betransmitted.

    81

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    82/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Standard Instrumentation SignalsCurrent Loop Scaling

    Output ScalingScale outputs for a one-to-one correspondence.Controller output is configured for 0% to correspondto a 4mA signal and 100% to correspond to a 20mAsignal.

    The final control element is calibrated so that 4mAcorresponds to its 0% position or speed and 20mAcorresponds to its 100% position or speed.

    82

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    83/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    What are Standard Instrumentation SignalsCurrent Loop Scaling

    Input ScalingScale inputs for a one-to-one correspondence as well.Example:

    If we were using a pressure transducer with a requiredoperating range of 0 psig to 100 psig we would calibratethe instrument such that 0 psig would correspond to 4mA

    output and 100 psig would correspond to a 20mA output.At the controller we would configure the input such that4mA would correspond to an internal value of 0 psig and10mA would correspond to an internal value of 100 psig.

    83

    What are Standard Instrumentation Signals

  • 7/27/2019 Unit Extra PDC

    84/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    g0 to 10 Volt

    0 to 10 volt is not commonly used in controlsystems because this signal is susceptible toinduced noise and the distance of the instrumentor final control element is limited due to voltagedrop.

    You may find 0-10 volt signals used in controlsystems providing the speed reference tovariable speed drives.

    84

    What are Smart Transmitters

  • 7/27/2019 Unit Extra PDC

    85/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    A smart transmitter is a digital device thatconverts the analog information from a sensorinto digital information, allowing the device tosimultaneously send and receive information andtransmit more than a single value.

    Smart transmitters, in general, have thefollowing common features:Digital CommunicationsConfigurationRe-Ranging

    Signal ConditioningSelf-Diagnosis

    85

    What are Smart Transmitters

  • 7/27/2019 Unit Extra PDC

    86/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Digital Communications

    Smart transmitters are capable of digital communicationswith both a configuration device and a process controller.Digital communications have the advantage of being free of bit errors, the ability to multiple process values anddiagnostic information, and the ability to receivecommands.

    Some smart transmitters use a shared channel for analogand digital data (Hart, Honeywell or Modbus over 4-20mA).Others use a dedicated communication bus (Profibus,Foundation Fieldbus, DeviceNet, Ethernet).

    86

    What are Smart Transmitters

  • 7/27/2019 Unit Extra PDC

    87/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Digital Communications

    Most smart instruments wired tomulti-channel input cards requireisolated inputs for the digitalcommunications to work.

    87

    What are Smart Transmitters

  • 7/27/2019 Unit Extra PDC

    88/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Configuration, Signal Conditioning, Self-Diagnosis

    ConfigurationSmart transmitters can be configured with ahandheld terminal and store the configurationsettings in non-volatile memory.

    Signal ConditioningSmart transmitters can perform noise filtering andcan provide different signal characterizations.

    Self-DiagnosisSmart transmitters also have self-diagnostic

    capability and can report malfunctions that mayindicate erroneous process values.

    88

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    89/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p

    The instruments range and span. The resolution of the measurement.The instruments accuracy and precision. The instruments dynamics

    89

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    90/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Range and Span

    The range of a sensor is the lowest and highestvalues it can measure within its specification.The span of a sensor is the high end of theRange minus the low end of the Range.Match Range to Expected Conditions

    Instruments should be selected with a range thatincludes all values a process will normally encounter,including expected disturbances and possible failures.

    90

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    91/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Measurement Resolution

    Resolution is the smallest amount of inputsignal change that the instrument can detectreliably.

    Resolution is really a function of the instrument spanand the controllers input capability.

    The resolution of a 16 bit conversion is:

    The resolution of a 12 bit conversion is:The bit error.

    535,65

    SpanInput

    095,4

    SpanInput

    91

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    92/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Accuracy

    Accuracy of a measurement describes how closethe measurement approaches the true value of the process variable.

    % error over a range x% over

    % of full scale% of span

    Absolute over a range x units over

    full scale

    span

    92

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    93/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Precision

    Precision is the reproducibility with whichrepeated measurements can be made underidentical conditions.

    This may be referred to as drift.

    93

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    94/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Accuracy vs. Precision

    Why is precision preferred over accuracy?

    94

    What Instrument Properties Affect a Process

  • 7/27/2019 Unit Extra PDC

    95/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Instrumentation Dynamics: Gain and Dead Time

    The gain of an instrument is often called sensitivity .

    The sensitivity of a sensor is the ratio of the outputsignal to the change in process variable.

    The dead time of an instrument is the time ittakes for an instrument to start reacting toprocess change.

    95

  • 7/27/2019 Unit Extra PDC

    96/261

    What is Input Aliasing

  • 7/27/2019 Unit Extra PDC

    97/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Input aliasing is a phenomenon that occurs fromdigital processing of a signal.When a signal is processed digitally it is sampledat discrete intervals of time. If the frequency atwhich a signal is sampled is not fast enough thedigital representation of that signal will not becorrect.Input aliasing is an important consideration indigital process control. Processor inputs that

    have configurable sample rates and PID loopupdate times must be set correctly.

    97

  • 7/27/2019 Unit Extra PDC

    98/261

    What is Input Aliasing

  • 7/27/2019 Unit Extra PDC

    99/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Correct Sampling Frequency

    Nyquist Frequency Theorem:To correctly sample a waveform it is necessary tosample at least twice as fast has the highestfrequency in the waveform.In the digital world this means sampling at 1/20th of

    the period of the waveform.Period = 1/frequency

    99

    What is Input Aliasingl

  • 7/27/2019 Unit Extra PDC

    100/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Correct Sampling Frequency

    2 Hz waveform sampled every 0.025 seconds

    100

  • 7/27/2019 Unit Extra PDC

    101/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Workshop Lab:

    Input Aliasing 1

    101

    What is Input AliasingD i i h C S li I l

  • 7/27/2019 Unit Extra PDC

    102/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Determining the Correct Sampling Interval

    Rules of ThumbSet the sample interval for an instrument at 1/10th to1/20th of the rise time (1/2 to 1/4th of the time constant).Set the sample interval to 1/10th to 1/20th of the processtime constant.Temperature instrumentation (RTDs and thermocouples inthermowells) typically have time constants of severalseconds or more. For these processes sampling intervalsof 1 second are usually sufficient.Pressure and flow instrumentation typically have timeconstants of to 1 second. For these processes sampling

    intervals of 0.1 second are usually sufficient.

    102

    What is Input AliasingD i i h C S li I l

  • 7/27/2019 Unit Extra PDC

    103/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Determining the Correct Sampling Interval

    103

  • 7/27/2019 Unit Extra PDC

    104/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Workshop Lab:

    Input Aliasing 2

    104

  • 7/27/2019 Unit Extra PDC

    105/261

    What is Instrument NoiseS

  • 7/27/2019 Unit Extra PDC

    106/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Sources

    Noise is generally a result of the technology usedto sense the process variable.

    Electrical signals used to transmit instrumentmeasurements are susceptible to having noiseinduced from other electrical devices.

    Noise can also be caused by wear and tear on themechanical elements of a sensor.

    Noise may be uncontrolled, random variations inthe process itself.

    Whatever the source, noise distorts themeasurement signal.

    106

    What is Instrument NoiseEffects of Noise

  • 7/27/2019 Unit Extra PDC

    107/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Effects of Noise

    Noise reduces the accuracy and precision of process measurements. Somewhere in the noiseis the true measurement, but where?

    Noise introduces more uncertainty into themeasurement.

    Noise also introduces errors in control systems.To a controller, fluctuations in the processvariable due to noise are indistinguishable fromfluctuations caused by real disturbances.

    Noise in a process variable will be reflected in theoutput of the controller

    107

    What is Instrument NoiseEliminating Noise

  • 7/27/2019 Unit Extra PDC

    108/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Eliminating Noise

    The most effective means of eliminating noise isto remove the source.

    Reduce electrically induced noise by following propergrounding techniques, using shielded cabling, andphysically separating the signal cabling form otherelectrical wiring.If worn mechanical elements in the sensor arecausing noise, then repair or replace the sensor.

    When these steps have been taken and excessivenoise is still a problem in the process variable alow pass filter may be appropriate.

    108

    What is Instrument NoiseLow Pass Filter

  • 7/27/2019 Unit Extra PDC

    109/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Low Pass Filter

    A low-pass filter allows the low frequencycomponents of a signal to pass while attenuatingthe higher frequency components.

    Raw (unfiltered) Signal

    Filtered Signal

    109

    What is Instrument NoiseLow Pass Filter

  • 7/27/2019 Unit Extra PDC

    110/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Low Pass Filter

    A low pass filter introduces a first order lag intothe measurement.

    Raw (unfiltered) Signal

    Filtered Signal

    110

    What is Instrument NoiseSelecting a Low Pass Filter

  • 7/27/2019 Unit Extra PDC

    111/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Selecting a Low Pass Filter

    By Cut-Off FrequencyCut-off frequency is defined as frequency abovewhich the filter provides -3dB of signal attenuation.

    An attenuation of 0 dB would mean the signal willpass with no reduction in amplitude while a largenegative dB would indicate a very small amplituderatio.

    Select a cut-off frequency that is above thefrequency of your process.

    InAmplitude

    OutAmplitude10log20nattenuatioof dB

    111

    What is Instrument NoiseSelecting a Low Pass Filter

  • 7/27/2019 Unit Extra PDC

    112/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Selecting a Low Pass Filter

    By Time ConstantSome filters are configured by selecting a timeconstant for the lag response of the filter.The relationship between the cut-off frequency andthe time constant of a low pass is approximately

    given by:

    ConstantsTime5

    1 FrequencyOff -Cut

    112

    What is Instrument NoiseSelecting a Low Pass Filter

  • 7/27/2019 Unit Extra PDC

    113/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Selecting a Low Pass Filter

    By ValueSome filters are selected by a value called alpha ( ),notably the derivative filter in a PID controller.

    is an averaging weighting term used in controllercalculations to impart a first order lag on themeasured variable.

    generally has values between 0 and 1. A filter witha = 0 would pass the signal through unfiltered. Afilter with a = 1 would filter everything allowingnothing to pass through the filter.

    113

    What is Instrument NoiseSelecting a Low Pass Filter

  • 7/27/2019 Unit Extra PDC

    114/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Selecting a Low Pass Filter

    When a filter is specified by cut-off frequency,the lower the frequency the greater the filteringeffect.When a filter is specified by time constant, thegreater the time constant the greater the filtering

    effect.When a filter is specified by , when =0 nofiltering is done, when = 1 no signal passesthrough the fitter.

    114

  • 7/27/2019 Unit Extra PDC

    115/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Workshop Lab:Noise Filtering

    115

    What is Temperature

  • 7/27/2019 Unit Extra PDC

    116/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Temperature is a measure of degree of thehotness or coldness of an object.Units of Temperature

    Two most common temperature scales areFahrenheit (F) and Celsius (C).

    The reference points are the freezing point and theboiling point of water.

    Water freezes at 32F and boils at 212F.The Celsius scale uses the same reference pointsonly it defines the freezing point of water as 0C andthe boiling point as 100C.

    116

    What is TemperatureUnit Conversion

  • 7/27/2019 Unit Extra PDC

    117/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Unit Conversion

    32C5

    9F

    x

    32-F95

    C x

    117

    What Temperature Instruments Do We Use

  • 7/27/2019 Unit Extra PDC

    118/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Temperature is one of the most common processvariables.Temperature is most commonly measured by

    Resistance Temperature Devices (RTD)Thermocouples

    Infrared (IR) (to a lesser degree) Thermistors (may also be found embedded in somecontrol equipment)

    118

    What Temperature Instruments Do We UseThermocouples

  • 7/27/2019 Unit Extra PDC

    119/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Thermocouples

    Thermocouples are fabricated from two electricalconductors made of two different metal alloys.

    Hot or sensing junctionCold or reference junctionGenerate an open-circuit voltage, called the Seebeck

    voltage that is proportional to the temperaturedifference between the sensing (hot) and reference(cold) junctions.

    119

    What Temperature Instruments Do We UseThermocouple Junctions

  • 7/27/2019 Unit Extra PDC

    120/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p

    There is a misconception of how thermocouples operate.

    The misconception is that the hot junction is the source of the output voltage. This is wrong. The voltage isgenerated across the length of the wire.Another misconception is that junction voltages aregenerated at the cold end between the special

    thermocouple wire and the copper circuit. Hence, a cold junction temperature measurement is required. Thisconcept is wrong. The cold end temperature is thereference point for measuring the temperature differenceacross the length of the thermocouple circuit.

    120

    What Temperature Instruments Do We UseThermocouple Junctions

  • 7/27/2019 Unit Extra PDC

    121/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p

    Hot junction voltage proportional to 210F.Extension wire voltage proportional to 18F.

    121

    What Temperature Instruments Do We UseThermocouple Junctions

  • 7/27/2019 Unit Extra PDC

    122/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p

    Use of the correct extension wire is critical inthermocouple applications.

    An incorrect extension will cause the temperaturedifferential across the extension leads to beintroduced as measurement error.

    Does this mean you cannot use copper terminalblocks for thermocouples?No, it is highly unlikely there will be a temperaturedifferential across the terminal block; no error will beintroduced.

    122

    What Temperature Instruments Do We UseThermocouple Linearization

  • 7/27/2019 Unit Extra PDC

    123/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p

    123

    What Temperature Instruments Do We UseThermocouple Gain

  • 7/27/2019 Unit Extra PDC

    124/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p

    124

    What Temperature Instruments Do We UseThermocouple Types

  • 7/27/2019 Unit Extra PDC

    125/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p yp

    125

    What Temperature Instruments Do We UseRTDs

  • 7/27/2019 Unit Extra PDC

    126/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    A resistance-temperature detector (RTD) is a temperature

    sensing device whose resistance increases withtemperature.An RTD consists of a wire coil or deposited film of puremetal whose resistance at various temperatures has beendocumented.RTDs are used when applications require accuracy, long-term stability, linearity and repeatability.100 Platinum is most common.

    126

  • 7/27/2019 Unit Extra PDC

    127/261

    What Temperature Instruments Do We UseRTD Lead Wire Resistance

  • 7/27/2019 Unit Extra PDC

    128/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The RTD is a resistive device, you must drive acurrent through the device and monitor theresulting voltage. Any resistance in the leadwires that connect your measurement system tothe RTD will add error to your readings.

    40 feet of 18 gauge 2conductor cable has6 resistance.For a platinum RTDwith = 0.00385,the resistance equals0.6 /(0.385 /C) = 1.6C of error.

    128

    What Temperature Instruments Do We UseRTD Lead Wire Resistance

  • 7/27/2019 Unit Extra PDC

    129/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Error is eliminated through the use of three wireRTDs.

    With three leads, two voltage measurements can bemade.When the voltages are subtracted, the result is

    the voltage drop that would have occurred throughRT alone.

    129

    What Temperature Instruments Do We UseRTD Self Heating Effect

  • 7/27/2019 Unit Extra PDC

    130/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The current used to excite an RTD causes the RTD to

    internally heat, which appears as an error. Self-heating istypically specified as the amount of power that will raisethe RTD temperature by 1 C, or 1 mW/ C.Self-heating can be minimized by using the smallestpossible excitation current, but this occurs at the expenseof lowering the measurable voltages and making the signalmore susceptible to noise from induced voltages.The amount of self-heating also depends heavily on themedium in which the RTD is immersed. An RTD can self-heat up to 100 times higher in still air than in moving water

    130

    What Temperature Instruments Do We UseThermistors

  • 7/27/2019 Unit Extra PDC

    131/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    A thermistor is similar to an RTD in that it is apassive resistance device.

    Thermistors are generally made of semiconductormaterials giving them much different characteristics.Thermistors do not have standardized electrical

    properties like thermocouples or RTDs.

    131

    What Temperature Instruments Do We UseInfrared

  • 7/27/2019 Unit Extra PDC

    132/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Objects radiate electromagnetic energy.The higher the temperature of an object the moreelectromagnetic radiation it emits.This radiation occurs within the infrared portion of the electromagnetic spectrum.

    132

    What Temperature Instruments Do We UseInfrared Emittance

  • 7/27/2019 Unit Extra PDC

    133/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The emittance of a real surface is the ratio of thethermal radiation emitted by the surface at agiven temperature to that of the ideal black bodyat the same temperature.

    By definition, a black body has an emittance of 1.

    Another way to think of emissivity is theefficiency at which an object radiates thermalenergy.Emittance is a decimal number that ranges

    between 0 and 1 or a percentage between 0%and 100%.

    133

    What Temperature Instruments Do We UseInfrared Emittance

  • 7/27/2019 Unit Extra PDC

    134/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    To function properly an infrared temperatureinstrument must take into account the emittanceof the surface being measured.Emittance values can often be found in referencetables, but such tables will not take into account

    local conditions of the surface.A more practical way to set the emittance is tomeasure the temperature with an RTD orthermocouple and set the instrument emissivity

    control so that both readings are the same.

    134

    What Temperature Instruments Do We UseInfrared Field of View

  • 7/27/2019 Unit Extra PDC

    135/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Infrared temperature instruments are like an optical

    system in that they have a field of view. The field of viewbasically defines the target size at a given distance.Field of view may be specified as:

    Angle and focal range (2.3 from 8" to 14')Distance to spot size ratio and focal range (25:1 from 8" to 14')Spot size at a distance (0.32" diameter spot at 8")All of these specifications are equivalent, at 8 inches our field of view will be a 0.32" diameter spot (8"/25), at 14 feet our field of view will be 6.72" (14' / 25).

    An infrared temperature sensor measures the averagetemperature of everything in its field of view. If thesurface whose temperature we are measuring does notcompletely fill the field of view we will get inaccurateresults.

    135

    What Temperature Instruments Do We UseInfrared Spectral Response

  • 7/27/2019 Unit Extra PDC

    136/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The range of frequencies that an infrared sensorcan measure is called its spectral response.Infrared temperature sensors do not measure theentire infrared region.

    Different frequencies (wavelengths) are selected for

    detection to provide certain advantages.If you wanted to use an infrared temperature sensor tomeasure the temperature of an object behind glass youwould need to select an instrument with a spectralresponse in the infrared region in which glass is"transparent.

    136

  • 7/27/2019 Unit Extra PDC

    137/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Section Assessment:Temperature

    137

    What is Pressure

  • 7/27/2019 Unit Extra PDC

    138/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Pressure is the ratio between a force acting on asurface and the area of that surface.Pressure is measured in units of force divided byarea.

    psi (pounds per square inch)

    bar, 1 bar = 14.5 PSI, common for pump ratingsinH20 (inches of water), 27.680 inH20 = 1 psi,common for vacuum systems and tank levelsmmHg (millimeters of mercury), 760 mmHg = 14.7psi, common for vacuum systems

    138

    What is Pressure Absolute, Gauge and Differential

  • 7/27/2019 Unit Extra PDC

    139/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Gauge pressure is defined relative to atmosphericconditions.

    The units of gauge pressure are psig, however gaugepressure is often denoted by psi as well.

    Absolute pressure is defined as the pressure

    relative to an absolute vacuum.The units of absolute pressure are psia.

    Differential pressure uses a reference point otherthan full vacuum or atmospheric pressure.

    139

    What is Pressure Absolute, Gauge and Differential

  • 7/27/2019 Unit Extra PDC

    140/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 140

  • 7/27/2019 Unit Extra PDC

    141/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Section Assessment:Pressure

    141

    Common Level Sensing Technologies

  • 7/27/2019 Unit Extra PDC

    142/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Point sensing level probes only sense tank level at a

    discrete level.Typically used for high-high or low-low level sensing to prevent plantpersonnel and/or process equipment from being exposed to harmfulconditions.Also used in pairs in processes in which we do not particularly carewhat the exact level in a tank is, only that it is between two points.

    Continuous level probes sense the tank level as a percentof span of the probes capabilities.

    Continuous level probes are typically used where we need some typeof inventory control, where we need to know with some degree of confidence what the particular level in a tank is.Contact and Non-Contact Types

    142

    Common Level Sensing TechnologiesNon-Contact: Ultrasonic

  • 7/27/2019 Unit Extra PDC

    143/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Ultrasonic makes use of sound waves.

    A transducer mounted in the top of a tank transmits soundwaves in bursts onto the surface of the material to bemeasured. Echoes are reflected back from the surface of the material to the transducer and the distance to thesurface is calculated from the burst-echo timing.

    The key points in applying an ultrasonic transducer are:The speed of sound varies with temperature.Heavy foam on the surface of the material interferes with the echo.An irregular material surface can cause false echoes resulting inirregular readings.Heavy vapor in the air space can distort the sound waves resulting infalse reading.

    143

    Common Level Sensing TechnologiesNon-Contact: Ultrasonic

  • 7/27/2019 Unit Extra PDC

    144/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 144

    Common Level Sensing TechnologiesNon-Contact: Radar/Microwave

  • 7/27/2019 Unit Extra PDC

    145/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Radar, or microwave level measurement,

    operates on similar principles to ultrasonic levelprobes but, instead of sound waves,electromagnetic waves in the 10GHz range areused.When properly selected, radar can overcomemany of the limitations of ultrasonic level probes.

    Be unaffected by temperature changes in the tank air space.See through heavy foam to detect the true material level.See through heavy vapor in the tank air space to detect truematerial level.

    145

    Common Level Sensing TechnologiesNon-Contact: Radar/Microwave

  • 7/27/2019 Unit Extra PDC

    146/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 146

    Common Level Sensing TechnologiesNon-Contact: Nuclear Level Measurement

  • 7/27/2019 Unit Extra PDC

    147/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Radiation from the source is detected on the

    other side of the tank. Its strength indicates thelevel of the fluid. Point, continuous, and interfacemeasurements can be made.As no penetration of the vessel is needed there are anumber of situations that cause nucleonic transmitters to

    be considered over other technologies.Nuclear level detection has some drawbacks. One is highcost, up to four times that of other technologies. Othersare the probable requirement for licenses, approvals, andperiodic inspections; and the difficulty and expense of disposing of spent radiation materials.

    147

    Common Level Sensing TechnologiesNon-Contact: Nuclear Level Measurement

  • 7/27/2019 Unit Extra PDC

    148/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 148

    Common Level Sensing TechnologiesContact: Hydrostatic Pressure

  • 7/27/2019 Unit Extra PDC

    149/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Measurement of level by pressure relies on hydrostatic

    principles.Pressure is a unit force over a unit area.

    A cubic foot (12"L x 12"W x 12"H) of water weighs 62.4796 pounds.The area that our cubic foot of water occupies is 144 square inches(12"L x 12"H), therefore our cubic foot of water exerts a force of 62.4796 pounds over 144 square inches, or 0.4339 psig for a 12"water column.1 inch H20 = 0.0316 psig

    It would not matter how many cubic feet of water wereplaced side by side, our pressure would still be 0.4399psig. Hydrostatic pressure is only dependent on the height

    of the fluid, not the area that it covers.

    149

    Common Level Sensing TechnologiesContact: Hydrostatic Pressure

  • 7/27/2019 Unit Extra PDC

    150/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    How would you handle measuring other fluids usingpressure?

    Compare the densities.For instance, chocolate weighs approximately 80 poundsper cubic foot.

    80 divided 62.5 times 0.0316 = 0.0404 psig.For a change in level of one inch in a liquid chocolate tankthe pressure measurement will change by 0.0404 psig.

    Level measurement by pressure requires a constant densityfor accurate measurements.If the head pressure in the tank can be other thanatmospheric, we must use a differential pressure sensor.

    150

    Common Level Sensing TechnologiesContact: Hydrostatic Pressure

  • 7/27/2019 Unit Extra PDC

    151/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 151

    Common Level Sensing TechnologiesContact: RF/Capacitance

  • 7/27/2019 Unit Extra PDC

    152/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    RF (radio frequency) Capacitance

    level sensors make use of electrical characteristics of acapacitor to infer the level in a vessel.

    As the material rises in the vessel,the capacitance changes.

    The level transducer measuresthis change, linearizes it and transmits the signal to theprocess control system.A point level probe will look for a specific change incapacitance to determine whether it is on or off.

    152

    Common Level Sensing TechnologiesContact: RF/Capacitance

  • 7/27/2019 Unit Extra PDC

    153/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Requires the material being measured to have a high

    dielectric constant.Level measurements are affected by changes in thedielectric of the material (moisture content).

    Proper selection requires informing the probe vendor of thematerial to be measure, especially in applications where

    you are measuring conductive materials or have anonmetallic tank.Point probe sensitivity can be increased by welding a plateon the sensor tip to increase the capacitance, and thereforethe sensitivity (gain).

    153

    Common Level Sensing TechnologiesContact: RF/Capacitance

  • 7/27/2019 Unit Extra PDC

    154/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 154

    Common Level Sensing TechnologiesContact: Guided Wave Radar

  • 7/27/2019 Unit Extra PDC

    155/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Guided wave radar is similar to the non-contact radar

    probes, only a rod or cable is immersed into the materiallike an RF capacitance probe.The rod or cable is used to guide the microwave along itslength, where the rod or cable meets the material to bemeasured a wave reflection is generated. The transit time

    of the wave is used to calculate level very precisely.Unlike an RF capacitance probe, a guided wave radar probecan measure extremely low dielectric material.

    155

    Common Level Sensing TechnologiesContact: Guided Wave Radar

  • 7/27/2019 Unit Extra PDC

    156/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 156

  • 7/27/2019 Unit Extra PDC

    157/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Section Assessment:Level

    157

    What is Flow

  • 7/27/2019 Unit Extra PDC

    158/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Flow is the motion characteristics of constrained

    fluids (liquids or gases).Fluid velocity or mass are typically not measureddirectly.Measurements are affected by the properties of

    the fluid, the flow stream, and the physicalinstallation.

    158

    Factors Affecting Flow Measurement

  • 7/27/2019 Unit Extra PDC

    159/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The critical factors affecting flow measurement

    are:Viscosity Fluid TypeReynolds Number

    Flow Irregularities

    159

    Factors Affecting Flow Measurement Viscosity

  • 7/27/2019 Unit Extra PDC

    160/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Dynamic or absolute viscosity ( ) is a measure of

    the resistance to a fluid to deformation undershear stress, or an internal property of a fluidthat offers resistance to flow.Commonly perceived as thickness or resistance

    to pouring.Water is very thin having a relatively low viscosity,while molasses is very thick having a relatively highviscosity.

    160

    l fl h f l f l f l f fl

    Factors Affecting Flow Measurement Viscosity

  • 7/27/2019 Unit Extra PDC

    161/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Visualize a flat sheet of glass on a film of oil on top of a flatsurface.

    A parallel force appliedto the sheet of glass willaccelerate it to a final velocitydependent only on theamount of force applied.The oil that is next to the sheetof glass will have a velocity

    close to that of the glass; whilethe oil that is next to the stationarysurface will have a velocity near zero.This internal distribution of velocities is due to the internalresistance of the fluid to shear stress forces, its viscosity.

    The viscosity of a fluid is the ratio between the per unitforce to accelerate the plate and the distribution of thevelocities within the fluid film

    161

    Factors Affecting Flow Measurement Viscosity

  • 7/27/2019 Unit Extra PDC

    162/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Effect of Temperature

    The dynamic viscosity of a fluid varies with itstemperature.In general, the viscosity of a liquid will decrease withincreasing temperature while the viscosity of a gaswill increase with increasing temperature.Viscosity measurements are therefore associatedwith a particular temperature

    162

    Factors Affecting Flow Measurement Viscosity

  • 7/27/2019 Unit Extra PDC

    163/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Viscosity is measured in units of poise or

    Pascalseconds or stokes.Poise or Pascalseconds are the units of dynamicviscosity.Stokes are the units of kinematic viscosity.

    Kinematic viscosity is the dynamic viscosity of a fluiddivided by the density of the fluid.

    163

    Factors Affecting Flow Measurement Viscosity Conversion Chart

  • 7/27/2019 Unit Extra PDC

    164/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    poiseP centipoisecP PascalsecondsPas milliPascalsecmPas stokesS centistokescS

    1 100 0.1 100 x 1/density x 100/density

    0.01 1 .001 1 x 0.01/density x 1/density

    10 1000 1 1000 x 10/density X000/density

    0.01 1 0.001 1 x 0.01/density x 1/density

    x density x 100 density x 0.1 density x 100 density 1 100

    x 0.01 density x density x 0.001 density x density 0.01 1

    164

    Factors Affecting Flow Measurement Viscosity of Common Fluids

  • 7/27/2019 Unit Extra PDC

    165/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    centiPoise (cP) Typical liquid Specific Gravity

    1 Water 1.0

    12.6 No. 4 fuel oil 0.82 - 0.95

    34.6 Vegetable oil 0.91 - 0.9588 SAE 10 oil 0.88 - 0.94

    352 SAE 30 oil 0.88 - 0.94

    820 Glycerine 1.26

    1561 SAE 50 oil 0.88 - 0.9417,640 SAE 70 oil 0.88 - 0.94

    165

    Factors Affecting Flow Measurement Viscosity

  • 7/27/2019 Unit Extra PDC

    166/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The fluid streams in most processes can include

    both high and low viscous fluids.The viscosity of the fluid under processconditions must be taken into account whenselecting a flow instrument for optimum

    performance.

    166

    Factors Affecting Flow MeasurementFluid Type

  • 7/27/2019 Unit Extra PDC

    167/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Newtonian Fluids

    When held at a constant temperature, the viscosity of a Non-Newtonian fluid will change with relation to thesize of the shear force, or it will change over timeunder a constant shear force.Water, glycerin, liquid sugar, and corn syrup areexamples.

    167

    Factors Affecting Flow MeasurementFluid Type

  • 7/27/2019 Unit Extra PDC

    168/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Non-Newtonian Fluids

    When held at a constant temperature, the viscosity of aNon-Newtonian fluid will change with relation to the sizeof the shear force, or it will change over time under aconstant shear force.

    Shear Thickening: Peanut Butter, Starch & WaterTime Thickening: Milk, Molasses

    168

    Factors Affecting Flow MeasurementReynolds Number

  • 7/27/2019 Unit Extra PDC

    169/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    The Reynolds number is the ratio of inertial

    forces to viscous forces of fluid flow within a pipeand is used to determine whether a flow will belaminar or turbulent.

    Dv Re 124

    cPinviscosityFluid

    3lb/ftindensityFluid

    ft/secinvelocityFluid inchesindiameter Pipe

    number Reynolds:Where

    v D Re

    169

    Factors Affecting Flow MeasurementLaminar Flow

    f

  • 7/27/2019 Unit Extra PDC

    170/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Laminar flow occurs at low Reynolds numbers,

    typically Re < 2000, where viscous forces aredominant.Laminar flow is characterizedby layers of flow traveling at

    different speeds with virtuallyno mixing between layers.The velocity of the flow is highest in the center of the pipe and lowest at the walls of the pipe.

    170

    Factors Affecting Flow MeasurementTurbulent Flow

    b l fl h h ld b

  • 7/27/2019 Unit Extra PDC

    171/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Turbulent flow occurs at high Reynolds numbers,

    typically Re > 4000, where inertial forces aredominant.Turbulent flow ischaracterized

    by irregular movementof the fluid in the pipe.There are no definite layers.The velocity of the flow is nearly uniform through

    the cross-section of the pipe.

    171

    Factors Affecting Flow MeasurementTransitional Flow

    T i i l fl i ll R ld

  • 7/27/2019 Unit Extra PDC

    172/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Transitional flow typically occurs at Reynolds

    numbers between 2000 and 4000.Flow in this region may be laminar, it may beturbulent, or it may exhibit characteristics of both.

    172

    Factors Affecting Flow MeasurementReynolds Number

  • 7/27/2019 Unit Extra PDC

    173/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Flow instruments must beselected to accurately

    measure laminar flow orturbulent flow

    173

    Factors Affecting Flow MeasurementFlow Irregularities

  • 7/27/2019 Unit Extra PDC

    174/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Flow irregularities arechanges in thevelocity profile of thefluid flow caused byinstallation of the flowinstrument.

    Normal velocityprofiles are

    established byinstalling a sufficientrun of straight piping.

    174

    Common Flowmeter Technologies

  • 7/27/2019 Unit Extra PDC

    175/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 175

    Common Flowmeter Technologies

    V l t i Fl

  • 7/27/2019 Unit Extra PDC

    176/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Volumetric Flow

    Magnetic flowmetersPositive displacement for viscous liquidsOrifice Plates

    Mass FlowCoriolis flowmeters

    176

    Common Flowmeter Technologies Volumetric Flow

    V l t i fl i ll i f d

  • 7/27/2019 Unit Extra PDC

    177/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Volumetric flow is usually inferred.

    Volumetric Flow = Fluid Velocity x AreaMeasured directly with positive displacement flowmeters.Measured indirectly by differential pressure

    Volumetric flow meters provide a signal that is inunits of volume per unit time.gpm (gallons per minute)cfm (cubic feet per minute)l/hr (liters per hour)

    177

    Common Flowmeter Technologies Volumetric Flow: Positive Displacement

    Oper te b c pt ring the fl id to be me s red in

  • 7/27/2019 Unit Extra PDC

    178/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Operate by capturing the fluid to be measured in

    rotating cavities of a known volume. The volumeof the cavity and the rate of rotation will give thevolumetric flow value.

    Unaffected by Reynolds number and work withlaminar, turbulent, and transitional flow.

    Low viscosity fluids will slip past the gearsdecreasing the accuracy of the flowmeter.

    178

    Common Flowmeter Technologies Volumetric Flow: Magnetic

    M ti fl t i f th

  • 7/27/2019 Unit Extra PDC

    179/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    kB DE

    kBD E D

    v Dvr

    velocity Area Pipe Flow

    44

    42

    22

    Magnetic flowmeters infer the

    velocity of the moving fluid bymeasuring a generatedvoltage.Magnetic flowmeters arebased on Faradays law of electromagnetic induction - a

    wire moving though amagnetic field will generate avoltage.Requires a conductive fluid.

    VelocityFluid

    Diameter Pipe

    DensityFluxMagnetic

    Constant

    v

    D

    B

    k

    kBDv E

    179

    Common Flowmeter Technologies Volumetric Flow: Orifice Plates

  • 7/27/2019 Unit Extra PDC

    180/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    An orifice plate is basically

    a thin plate with a hole inthe middle. It is usuallyplaced in a pipe in whichfluid flows.In practice, the orifice

    plate is installed in thepipe between two flanges.The orifice constricts theflow of liquid to produce adifferential pressure across

    the plate. Pressure taps oneither side of the plate areused to detect thedifference.

    180

    Common Flowmeter Technologies Volumetric Flow: Orifice Plates

    Orifice Plate type meters only work well when

    http://www.answers.com/main/ntquery?method=4&dsname=Wikipedia+Images&dekey=OrificePlate.JPG&gwp=8
  • 7/27/2019 Unit Extra PDC

    181/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Orifice Plate type meters only work well when

    supplied with a fully developed flow profile.This is achieved by a long upstream length (20 to 40diameters, depending on Reynolds number)

    Orifice plates are small and cheap to install, but

    impose a significant energy loss on the fluid dueto friction.

    181

    Common Flowmeter TechnologiesMass Flow

    Measured directly by measuring the inertial

  • 7/27/2019 Unit Extra PDC

    182/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Measured directly by measuring the inertial

    effects of the fluids moving mass.Mass flow meters provide a signal that is in unitsof mass per unit time.Typical units of mass flow are:

    lb/hr (pounds per hour)kg/s (kilograms per second)

    182

    Common Flowmeter TechnologiesMass Flow: Coriolis

    Coriolis flowmeters are based on the affects of the

  • 7/27/2019 Unit Extra PDC

    183/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Coriolis flowmeters are based on the affects of the

    Coriolis force.When observing motion from a rotating frame thetrajectory of motion will appear to be altered by aforce arising from the rotation.

    183

    Common Flowmeter TechnologiesMass Flow: Coriolis

  • 7/27/2019 Unit Extra PDC

    184/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 184

    Common Flowmeter TechnologiesMass Flow: Coriolis

  • 7/27/2019 Unit Extra PDC

    185/261

    Copyright 2007 Control Station, Inc. All Rights Reserved 185

    Common Flowmeter TechnologiesMass Flow: Coriolis

    Coriolis mass flowmeters

  • 7/27/2019 Unit Extra PDC

    186/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Coriolis mass flowmeterswork by applying avibrating force to a pairof curved tubes throughwhich fluid passes, ineffect creating a rotatingframe of reference.

    The Coriolis Effect of themass passing throughthe tubes creates a forceon the tubesperpendicular to both thedirection of vibration andthe direction of flow,causing a twist in thetubes.

    186

    Common Flowmeter TechnologiesFlowmeter Turndown

    A specification commonly found with flow sensors

  • 7/27/2019 Unit Extra PDC

    187/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    A specification commonly found with flow sensors

    is turndown or rangeability.Turndown is the ratio of maximum to minimum flowthat can be measured with the specified accuracy.

    For example, a flow meter with a full span of 0 to100 gallons per minute with a 10:1 turndowncould measure a flow as small as 10 gallons perminute.Trying to measure a smaller flow would result invalues outside of the manufacturers statedaccuracy claims.

    187

    Common Flowmeter TechnologiesInstallation and Calibration

    As with any instrument installation optimal flow

  • 7/27/2019 Unit Extra PDC

    188/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    As with any instrument installation, optimal flow

    meter performance and accuracy can only beachieved through proper installation andcalibration. Requirements will vary between typeand manufacturer by some generalconsiderations are:

    The length of straight piping required upstream anddownstream of the instrument. This is usually specified inpipe diameters.The properties of the fluid/gas to be measured.Mounting position of the flowmeter.

    Flowmeters must be kept full during operation.

    188

    Common Flowmeter TechnologiesInstallation and Calibration

    In general calibration of a flowmeter requires:

  • 7/27/2019 Unit Extra PDC

    189/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    In general, calibration of a flowmeter requires:

    Entering the sensor characterization constants intothe flow transmitter.Recording the zero flow conditions.

    Many flow meters must be zeroed only when full. Comparing the meter flow value to an independentlymeasured value for the flow stream.

    Measuring the volume or mass of a catch sample andcomparing the values.Use the internal flowmeter totalizer value for

    comparison.

    189

  • 7/27/2019 Unit Extra PDC

    190/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Section Assessment:Flow

    190

    Introduction to Final Control Elements

    Objectives:

  • 7/27/2019 Unit Extra PDC

    191/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    j

    What is a control valve?What is an actuator?What is a positioner?What is C v?What are valve characteristics?What is valve deadband?What is stiction?What are the types of valves

    What is a centrifugalpump?What is pump head?Why do we use pump headand not psi?What is a pump curve?

    What is a system curve?What is the systemoperating point?What is a positivedisplacement pump?

    How does a PD pump differfrom a centrifugal pump?

    191

    Introduction to Final Control Elements

    The intent of this chapter is neither to teach you how to

  • 7/27/2019 Unit Extra PDC

    192/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p ysize a valve or select a pump nor to familiarize you with all of the available types of valves and pumps.The intent of this chapter is to provide an introduction tosome of the commonly used pumps and valves, includingbasic terminology and characteristics relevant to their rolein a control loop.Detailed information and assistance on device selection istypically available from your instrumentation suppliers.

    192

    What is a Control Valve

    A control valve is an inline device in a flow

  • 7/27/2019 Unit Extra PDC

    193/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    A control valve is an inline device in a flow

    stream that receives commands from a controllerand manipulates the flow of a gas or fluid in oneof three ways:

    Interrupt flow ( Shut-Off Service )Divert flow to another path in the system ( Divert Service )Regulate the rate of flow ( Throttling Service )

    193

    What is a Control ValveShut-Off Service

  • 7/27/2019 Unit Extra PDC

    194/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Control valves forflow shut-off servicehave two positions.

    In the open positionflow is allowed to exitthe valve.In the closed positionflow is blocked fromexiting the valve.

    194

    What is a Control ValveDivert Service

  • 7/27/2019 Unit Extra PDC

    195/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    Control valves for divertservice also have twopositions, however flowis never blocked.

    In one position, flowis allowed from thecommon port to portA.In the other position,flow is allowed formthe common port toport B.

    195

    What is a Control ValveThrottling Service

    Control valves for throttling service have many

  • 7/27/2019 Unit Extra PDC

    196/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    g y

    positions.The position of the valve determines the rate of flowallowed through the valve.

    196

    What is an Actuator

    Actuators are pneumatic, electrical, or hydraulic

  • 7/27/2019 Unit Extra PDC

    197/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p , , y

    devices that provide the force and motion toopen and close a valve.

    197

    What is an Actuator

    Some actuators can place a valve at any position

  • 7/27/2019 Unit Extra PDC

    198/261

    Copyright 2007 Control Station, Inc. All Rights Reserved

    p y p

    between the on and off points.These actuators typically accepta 3-15 psi signal to move adiaphragm, which in turn movesa connected valve stem.

    A pneumatic positional actuatorwill fail into the pneumaticallyde-energized position.The interface to a control system for a positionalactuator is typically through an I/P transducer.

    198

    What is an Actuator

    Just as processes have a time constants and dead time,l h i d d d i hi h i

  • 7/27/2019 Unit Extra PDC


Recommended