Cell 12 Magnet. Data acquisition program. Magnet control. Data Acquisition computer. Magnet/Power supply status monitoring computer. unit: mm. 35.1 T, 32 mm bore Available cryostats: He3 (0.3 – 60 K) or VTI (1.4 – 300 K). System D Rotator Probe. Side A. Side B. Before mounting. - PowerPoint PPT Presentation
Popular Tags:
13
unit: mm Cell 12 Magnet T, 32 mm bore able cryostats: He3 (0.3 – 60 K) or VTI (1.4 – 300 K) Magnet/Power supply status monitoring computer Data Acquisition computer Data acquisition program Magnet control
Transcript
Page 1: unit: mm

unit: mm

Cell 12 Magnet

• 35.1 T, 32 mm bore• Available cryostats: He3 (0.3 – 60 K) or VTI (1.4 – 300 K)

Magnet/Power supply status monitoring computer

Data Acquisition computer

Data acquisition program

Magnet control

Page 2: unit: mm

System D Rotator Probe

Sample holder : side A

Sample holder mounted on the probe

Side ASide B

Before mounting

top view

• Sr2RuO4

• Magnetic torque • Capacitance between the

two parallel plates

side view

Page 3: unit: mm

1.5 mm

• SrMnBi2

• Electrical transport• Resistance (Rxx)

• Hall sensor (THS 118: GaAs based)• Hall Resistance (Rxy)

• Ba2Fe2As2

• Surface conductivity• TDO (Tunnel Diode Oscillator)

resonance freq. shift

• PdCoO2

• Magnetic torque • Resistance change of

piezo-resistive device

0.25 mm

Sample holder : side B11.4 mm

Page 4: unit: mm

C. Bergemann et al., Physica B 294, 371 (2001)

Torque: Capacitance signal, T= 30 mK

Sr2RuO4 Triplet superconductor (Tc = 1.4 K)

Page 5: unit: mm

BaFe2As2

D. E. Graf et al., PRB 85, 134503 (2012) J. G. Analytis et al., PRB 80, 064507 (2009)

TDO frequency shift

30 mK

350 mK

Parent compound of 122 pnictide superconductors

Page 6: unit: mm

SrMnBi2

K. Wang et al., PRB 85 041101(R) (2012)J. Park et al., PRL 107 126402 (2011)

CaMnBi2

SrMnBi2

Dirac Fermion compound

Page 7: unit: mm

PdCoO2 Metallic triangular lattice compound

C. W. Hicks et al., PRL 109 116401 (2012)

Torque signal (piezo-cantilever) T=0.7 K

Page 8: unit: mm

Measurement techniques

Resistance measurements by instrumentation (Rxx, Rxy, piezo-cantilever)principle: apply current (current source) and measure voltage (voltmeter)

• Conventional lock-in amplifier technique• Oscillating current applied, phase sensitive detection of corresponding voltage

• DC resistance measurement technique• DC current with switching polarities to remove offset

• AC resistance bridge• Resistance determined by nulling or measuring unbalanced signal

Q: what is advantage/disadvantage of the lock-in technique compared to the DC technique?

Capacitance measurement for magnetic torque• AH capacitance bridge: automatic balancing• GR capacitance bridge: manual balancing

Q: what other physical properties can be measured by the capacitance measurement device?

TDO resonance frequency measurementTypical capacitance bridge

Page 9: unit: mm

Exercises 1: Resistance measurement on the standard resistor

Lock-in technique• configure resistance measurement setup with 1 lock-in amplifier (SR 830)• Change parameters (current, frequency, time constant, ..) and check the voltage

readings

DC technique• Configure measurement setup with 1 DC current source (Keithley 6221) and 1

DC nanovoltmeter (Keithly 2182)• Apply constant DC current ( < 10 mA) and monitor the voltage• Setup the Delta mode and test with different parameters (current, delay time,..)

Exercises 2: Capacitance measurement on the standard capacitor

(1) AH bridge : hook up coax cables to the decade capacitor box and press buttons(2) GR bridge • configure GR bridge and 1 lockin-amplifer for capacitance measurement • balance the GR bridge to find the capacitance value (should be similar to obtained from (1)• find the conversion factor between lock-in signal and capacitance

(hints given in the appendix)

Page 10: unit: mm

Exercises 3: Balancing the piezo-cantilever

Demonstration 1: TDO measurement

Exercises 4: Setting up instruments for the real samples

inside cryostat

• Build a measurement circuit with a given bridge circuit box

• Balance the bridge

(1) Setup instruments for the Hall sensor (DC resistance delta mode) and piezo-cantilever (lock-in technique). Refer to the breakout box diagram in the next page.

NOTE: Do not apply currents until further notice.(2) Instructor will setup AC resistance bridge for the transport measurement(3) Connect cables to the AH bridge for the capacitance measurement(4) Using the data acquisition program, collect the data at zero field as a function of time

Page 11: unit: mm

Exercise/Demonstration: Data analysis

Exercises 5: Data acquisition

(1) Collect data while sweeping magnetic field under different instrumentation setup(2) Collect data at different temperatures(3) Collect data at different angles

(1) Plot the data using data analysis programQ1) Plot the temperature vs. magnetic field. Why does temperature reading change with field? Any way to circumvent the problem?Q2) What is the oscillation frequency of the quantum oscillations? Q3) Capacitance torque signal from Sr2RuO4 sample look different from the text book (saw-tooth not sinusoidal). Why?

(2) Perform FFT with different settings (field range, different background, FFT windows, …)

NHMFL breakout box

variableresistor

can be used as pseudo-current source setup when combined with lock-in oscillator

(answer given in the appendix)

Page 12: unit: mm

• Configure Delta (picture above) : set I-high, I-low, Delay. ,,,,• Press Delta to arm• Press TRIG to start delta mode operation

• Oscillator out to the current terminals with a large resistor connected in series• Connect voltage terminals to voltage input (A-B mode)

DC resistance measurement (delta mode)

osc. outvoltage input

sample

Current limiting resistor (R) I = V/(Rsample+R) ~ V/R (if R >> Rsample)

Resistance measurement (Lock-in technique)Appendix I

Page 13: unit: mm

Appendix IITorque interaction analysis in Sr2RuO4 by Naoki Kikugawa

Raw data(before de-torque)FFT: 28 – 34.5T

De-torquedFFT: 28 – 34.5 T

Sr2RuO4

T = 1.4 KB // ~ c axisData taken at Cell 12 (May 21, 2014)

Remember: The oscillation is periodic in 1/B not in 1/H