UNIVERSIDADE ESTADUAL DE CAMPINAS
FACULDADE DE ODONTOLOGIA DE PIRACICABA
RENATA PEREIRA
AVALIAÇÃO DA RESISTÊNCIA DE UNIÃO, NANOINFILTRAÇÃO E
ADAPTAÇÃO MARGINAL DE SISTEMAS RESINOSOS BULK-FILL
SUBMETIDOS A ENVELHECIMENTO TERMOMECÂNICO
EVALUATION OF BOND STRENGTH, NANOLEAKAGE AND
MARGINAL ADAPTATION OF BULK-FILL RESIN COMPOSITES
SUBMITTED TO THERMOMECHANICAL AGING
Piracicaba
2017
RENATA PEREIRA
AVALIAÇÃO DA RESISTÊNCIA DE UNIÃO, NANOINFILTRAÇÃO E
ADAPTAÇÃO MARGINAL DE SISTEMAS RESINOSOS BULK-FILL
SUBMETIDOS A ENVELHECIMENTO TERMOMECÂNICO
EVALUATION OF BOND STRENGTH, NANOLEAKAGE AND
MARGINAL ADAPTATION OF BULK-FILL RESIN COMPOSITES
SUBMITTED TO THERMOMECHANICAL AGING
Dissertação apresentada à Faculdade de Odontologia de
Piracicaba da Universidade Estadual de Campinas como parte
dos requisitos exigidos para a obtenção do título de Mestra em
Clínica Odontológica, na Área de Dentística.
Dissertation presented to the Piracicaba Dental School of the
University of Campinas in partial fulfillment of the requirements
for the Degree of Master in Dental Clinic in Operative Dentistry
Area.
Orientador: Prof. Dr. Flávio Henrique Baggio Aguiar
ESTE EXEMPLAR CORRESPONDE À VERSÃO FINAL
DA DISSERTAÇÃO DEFENDIDA PELA ALUNA RENATA
PEREIRA E ORIENTADA PELO PROF. DR. FLÁVIO
HENRIQUE BAGGIO AGUIAR.
Piracicaba
2017
DEDICATÓRIA
Dedico este trabalho a quem tem a maior parte da minha inspiração e da minha
força, àquela que tem o melhor consolo e cura para todas as minhas aflições e certamente
quem mais torce pelas minhas vitórias: minha mãe, Lucimara.
AGRADECIMENTOS
A Deus, por proteger e iluminar meus caminhos e escolhas, e pelas pessoas
maravilhosas que colocou em minha trajetória.
À Faculdade de Odontologia de Piracicaba – UNICAMP e todos os professores
com quem eu tive prazer em ter aprendido tanto em todos esses anos de crescimento
profissional, em especial ao meu orientador, professor Dr. Flávio Henrique Baggio Aguiar,
por ter me acolhido na Dentística desde a iniciação científica. Sou muito grata pelas
oportunidades que me deu, por confiar no meu trabalho e pela dedicação, apoio e paciência
que tem comigo. Não tenho dúvidas de que o senhor é o melhor orientador que eu poderia ter.
Obrigada por todo o conhecimento transmitido!
À professora Dra. Maria Cecília Caldas Giorgi, por ter me ajudado a dar os
“primeiros passos” como pesquisadora e ter me transmitido conhecimentos e valores
essenciais para que eu seguisse essa trajetória.
Às professoras Dra. Débora Alves Nunes Leite Lima, Dra. Vanessa Cavalli
Gobbo e Dra. Lúcia Trazzi Prieto, pela competência com que avaliaram meu trabalho no
exame de qualificação.
À Microscopia Eletrônica da ESALQ- USP, em nome do Prof. Dr. Francisco
Tanaka e do técnico Renato Salaroli, por permitir a realização das análises microscópicas.
Aos técnicos de laboratório de Odontopediatria e Materiais Dentários da FOP –
UNICAMP, Marcelo Corrêa Maistro e Marcos Blanco Cangiani, respectivamente, pelo
auxílio no uso dos equipamentos de ciclagens mecânica e térmica e ajuda com o software
ImageJ.
A CAPES e a FAPESP, pela concessão das bolsas de estudo, que me permitiram
desenvolver este trabalho.
De forma especial agradeço aos meus pais, Lucimara e João Carlos, por serem
meu porto seguro, por sempre me apoiarem e estarem presentes em todos os momentos da
minha vida: desde a escolha do curso de odontologia, até a decisão de ingressar no curso de
mestrado, e em todos os outros momentos, menos ou mais importantes. Sem vocês nada disso
seria possível. Nenhuma palavra escrita aqui é capaz de demonstrar suficientemente o quanto
sou grata por ter vocês como pais. Vocês são minha grande inspiração como dentistas e como
pessoas!
À minha irmã, Ana Paula, por ser minha melhor amiga e confidente. Tenho
muito orgulho da mulher maravilhosa que você se tornou. Só tenho a agradecer pela nossa
sintonia única e por dividir essa vida com você.
Ao meu irmão, Matheus, por ser o melhor irmão do mundo e uma inspiração pra
mim. Admiro muito sua inteligência, sua determinação e todos os valores que norteiam as
escolhas da sua vida!
Aos meus tios e primos queridos da família, sempre presentes em minha vida. Em
especial à minha tia, Cíntia, minha segunda mãe, por sempre torcer por mim e se preocupar
comigo em todas as fases da minha vida, e à Ana Clara, nossa pequena, que tanto nos dá
alegrias.
Ao meu melhor amigo e namorado, Thalles, por toda a paciência,
companheirismo e amor. Obrigada por caminhar ao meu lado, pelos sonhos divididos e por
tornar a minha vida tão mais leve. Espero poder comemorar ainda muitas conquistas ao seu
lado! Aproveito para estender o meu agradecimento a toda família Del Tedesco Almeida
Leme, por sempre terem me acolhido tão bem e me fazerem sentir parte da família.
A todos os meus amigos, pelo companheirismo, especialmente: Bruna e Ana
Eliza, por dividirem não só a casa comigo, mas todas as alegrias e aflições do dia-a-dia;
Patrícia, pela nossa sintonia e por estar sempre presente na minha vida e Maria Giulia, que
me prova que verdadeiras amizades continuam a crescer mesmo a longas distâncias. Obrigada
por me ensinarem o verdadeiro significado de amizade! Estendo meus agradecimentos aos
amigos conquistados durante o curso do mestrado na Dentística: Rodrigo, Maicon, Josué,
Mariana e Mayara; aos amigos do doutorado: Jéssica, Mari, Waldemir, Michele, Suelem,
Thayla, Maria Jordão e à Marcela, que está começando o mestrado agora. Vocês tornaram
esses dois anos bem mais leves.
Por fim, a todos que, direta ou indiretamente, contribuíram para a realização desse
trabalho, muito obrigada!
RESUMO
Este estudo teve como objetivo avaliar a resistência de união (RU) por
microtração, nanoinfiltração (NI) e adaptação marginal (AM) de três compósitos
fotopolimerizáveis bulk-fill, comparando-os com um nanocompósito convencional, após
serem submetidos ou não a ciclagens mecânica e térmica (CMT). Noventa e seis terceiros
molares humanos hígidos foram selecionados e divididos em 8 grupos (n=12), de acordo com
o tipo de compósito: Filtek Z350 XT – 3M ESPE (Z350), Tetric N-Ceram Bulk Fill – Ivoclar
Vivadent (TET), Filtek Bulk Fill Posterior Restorative - 3M ESPE (FBF) e SonicFill – KAVO
Kerr (SF) e envelhecimento: Com CMT e sem CMT. As amostras foram restauradas em uma
cavidade classe I (5mm no sentido mésio-distal, 4mm vestíbulo-lingual e 4 mm de
profundidade), envelhecidas ou não e submetidas a RU (n=7) por meio de palitos obtidos da
região restaurada. Para NI, dois palitos de cada dente foram selecionados para se avaliar a
infiltração por íons de nitrato de prata. Para AM (n=5), os dentes foram seccionados paralela e
perpendicularmente à superfície oclusal e a partir de impressões das superfícies obtidas, foram
confeccionadas réplicas em resina epóxica. O padrão de fratura, NI e a AM das paredes
circundantes e de fundo da cavidade foram avaliados através de microscópio eletrônico de
varredura. As análises quantitativas de NI e AM foram realizadas com o auxílio do software
ImageJ e os dados obtidos de todos os testes foram submetidos à análise estatística ANOVA
dois fatores e teste de Tukey (ɑ = 0.05). Os resultados mostraram que TET apresentou
maiores valores de RU quando comparado a SF. Z350 e FBF apresentaram valores de RU
estatisticamente semelhantes aos demais. Os padrões de fratura mais comumente ocorridos
foram coesiva em resina ou adesiva para todos os compósitos, exceto para FBF não
submetido a CMT, que apresentou maior porcentagem de falhas mistas. Quando submetidos a
CMT, a taxa de falhas adesivas aumentou para todos os compósitos, exceto para TET, que
teve sua taxa de falhas mistas aumentada. Quanto à NI, os maiores valores foram obtidos por
TET, diferindo significativamente dos demais, que apresentaram resultados semelhantes entre
si. Quando não submetidos a CMT, os menores valores de formação de fenda nas paredes de
fundo foram obtidos por TET e SF, diferindo estatisticamente dos demais, que não diferiram
entre si. Quando submetidos a CMT, FBF apresentou os maiores valores de formação de
fenda nas paredes de fundo, seguido por Z350 e SF. TET não diferiu estatisticamente de FBF
e Z350. As maiores porcentagens de formação de fenda nas paredes circundantes foram
obtidas por Z350, independente de ter sido submetido ou não a CMT. Quando não submetidos
a CMT, os demais compósitos apresentaram os menores valores de formação de fenda e não
diferiram estatisticamente entre si. Quando submetidos a CMT, TET e SF apresentaram os
menores valores. FBF não diferiu significativamente dos demais. O envelhecimento das
amostras por meio de CMT afetou somente os resultados de AM. Concluiu-se que os
compósitos bulk-fill testados apresentaram desempenho semelhante ao nanocompósito
convencional e se mantiveram estáveis diante dos desafios do envelhecimento.
Palavras-chave: Resinas compostas. Resistência à tração. Infiltração dentária. Adaptação
marginal dentária.
ABSTRACT
This study aimed to evaluate the microtensile bond strength (BS), nanoleakage
(NL) and marginal adaptation (MA) of three light-cured bulk-fill composites, comparing them
with a conventional nanocomposite, after being submitted or not to mechanical and thermal
cycling (MTC). Ninety-six healthy human third molars were selected and divided into 8
groups (n=12), according to the composite type: Filtek Z350 XT – 3M ESPE (Z350), Tetric
N-Ceram Bulk Fill – Ivoclar Vivadent (TET), Filtek Bulk Fill Posterior Restorative - 3M
ESPE (FBF) and SonicFill – KAVO Kerr (SF) and aging: With MTC and without MTC. The
samples were restored in a class I cavity (5mm in the mesio-distal width, 4mm in bucco-
lingual and 4mm of depth) aged or not and submitted to BS (n=7) through beams obtained
from the restored area. For NL, two beams of each tooth were selected to assess the
infiltration of silver nitrate ions. For MA (n=5), the teeth were sectioned parallel and
perpendicularly to the occlusal surface and replicas were made in epoxy resin from
impressions of the obtained surfaces. The fracture pattern, NL and MA of the external and
internal cavity walls were evaluated using scanning electron microscope. The quantitative
analysis of NL and MA were performed with the aid of ImageJ software and data obtained
from all tests were submitted to two-way ANOVA statistical analysis and Tukey’s test (ɑ =
0.05). The results showed that TET presented higher values of BS when compared to SF.
Z350 and FBF presented BS values statically similar to the others. The most commonly
occurred fracture patterns were cohesive in composite or adhesive for all composites, except
for FBF not submitted to MTC, which presented a higher percentage of mixed failures. When
submitted to MTC, the adhesive failures rate increased for all composites, except for TET,
which had its mixed failure rate increased. Regarding NL, the highest values were obtained by
TET, differing significantly from the others, which presented similar results among
themselves. When not submitted to MTC, the lowest values of internal gap formation were
obtained by TET and SF, differing statistically from the others, which did not differ from each
other. When submitted to MTC, FBF presented the highest internal gap formation values,
followed by Z350 and SF. TET did not differ statistically from FBF and Z350. The highest
percentages of external gap formation were obtained by Z350, regardless of being submitted
or not to MTC. When not submitted to MTC, the other composites presented the lowest gap
formation values and did not differ statistically from each other. When submitted to MTC,
TET and SF presented the lowest values. FBF did not differ significantly from the others. The
aging of the samples by MTC affected only MA results. It was concluded that bulk-fill
composites tested presented similar performance to the conventional nanocomposite and
remained stable in face of aging challenges.
Key Words: Composite resins. Tensile strength. Dental leakage. Dental marginal adaptation.
SUMÁRIO
1 INTRODUÇÃO 13
2 ARTIGO: Evaluation of bond strength, nanoleakage and marginal adaptation of bulk-fill
resin composites submitted to thermomechanical aging 16
3 CONCLUSÃO 36
REFERÊNCIAS 37
APÊNDICE 1 – Delineamento experimental e metodologia ilustrada 41
ANEXOS
Anexo 1 – Certificado do Comitê de Ética em Pesquisa 53
Anexo 2 – Certificados de submissão do artigo ao periódico Journal of Dentistry 54
13
1 INTRODUÇÃO
A procura pelo material restaurador ideal tem sido constante na Odontologia.
Espera-se que este material tenha propriedades físico-químicas satisfatórias, proporcionando
maior durabilidade, seja biocompatível, tenha capacidade de união aos tecidos dentais,
propiciando um preparo cavitário mais conservador, além de apresentar estética aceitável.
Desde sua introdução no mercado, os compósitos odontológicos popularizaram-se e tornaram-
se objeto de inúmeras pesquisas (Leprince et al., 2010). Assim, tem sido o material de escolha
para restaurações diretas anteriores e posteriores, uma vez que consegue atender a esses
requisitos de forma satisfatória (Rueggeberg, 2002; Ferracane, 2011).
Os compósitos resinosos dentais, em sua maioria, são ativados através da luz.
Nesse sistema o iniciador consiste em uma molécula fotossensível, como a canforoquinona, e
uma amina iniciadora. Quando ocorre exposição à luz, mais precisamente na região azul do
espectro (comprimento de onda de aproximadamente 468nm), esses dois componentes reagem
entre si, formando radicais livres, que iniciam a polimerização por adição (Annusavice, 2013).
Os compósitos fotopolimerizáveis, no entanto, possuem suas desvantagens, tais como a
contração e tensão de polimerização e a subpolimerização dos compósitos (Tirtha et al., 1982;
Rueggeberg, 1999; Stansbury, 2000).
Durante a polimerização, os monômeros se aproximam para estabelecer ligações
covalentes com objetivo principal de produzir ligações cruzadas para formar uma rede
polimérica com altas propriedades mecânicas (Stansbury, 2012). Essa aproximação resulta em
contração volumétrica do material, que varia de 1 a 5% para as resinas compostas
convencionais (Stansbury, 1990; Labella et al., 1999; Yamazaki et al., 2013; Fronza, 2015). A
real preocupação em relação à contração de polimerização, que é inerente aos compósitos
resinosos, é a tensão criada no material. Essa tensão é produto da limitação da contração livre
do polímero, e é dependente de vários fatores, incluindo a configuração e natureza das
moléculas monoméricas, a aquisição de rigidez do material durante a polimerização, a
velocidade da reação e as limitações externas impostas pela adesão do compósito ao dente. O
fato é que essa tensão interna não pode ser adequadamente aliviada, seja por mudanças na
estrutura molecular do compósito ou por deformação em suas superfícies livres (Ferracane &
Hilton, 2016). Sendo assim, essas tensões são transmitidas para a interface adesiva dente-
restauração, criando trincas ou fraturas dentais onde e quando a tensão localizada exceder a
força de adesão ou a força da estrutura dental residual adjacente. Podem ainda ser observadas
deflexão de cúspides e formação de fendas entre a restauração e o dente, o que pode ocasionar
14
sensibilidade pós-operatória, descoloração marginal e infiltração bacteriana com posterior
desenvolvimento de cárie secundária. Além disso, essas tensões podem aumentar com o
tempo, causando prejuízos tardios às margens da cavidade (Ferracane & Mitchem, 2003;
Calheiros et al., 2004; Fronza, 2015; Ferracane & Hilton, 2016).
A fim de minimizar esses danos e também devido à limitada profundidade de
penetração da luz dos aparelhos fotoativadores, convencionou-se que os compósitos devem
ser aplicados em incrementos horizontais ou oblíquos com até 2 mm de espessura (Lutz et al.,
1986; Pollack, 1987; Hilton & Ferracane, 1999; Yap, 2000; Leprince et al., 2014). Assim
seria possível assegurar um maior grau de conversão monomérico do compósito e um controle
da tensão de polimerização, através da diminuição do volume de material e do fator C (Borges
et al., 2014). Entretanto, a necessidade de fotoativação em múltiplos incrementos pode
consumir muito tempo operatório (Francis et al., 2015). Além disso, esse conceito de técnica
incremental já foi questionado em alguns estudos, que sugerem que a contração de
polimerização de cada incremento faz com que as paredes da cavidade inclinem-se, causando
deformação da cavidade e consequentemente maior tensão de contração (Versluis et al., 1996;
Rees et al., 2004; Bicalho et al., 2014). Independente da controvérsia, muitos outros fatores
estão envolvidos no desenvolvimento da tensão de contração, como composição e conteúdo
inorgânico, módulo de elasticidade, viscosidade, grau de conversão e profundidade de
polimerização do compósito (Benetti et al., 2015; Francis et al., 2015).
Ao longo dos anos os compósitos tem passado por diversas mudanças em suas
formulações para melhorar suas propriedades e facilitar o procedimento restaurador -
diminuição do tamanho das partículas, quantidade das mesmas, forma ou tratamento de suas
superfícies, mudanças na estrutura monomérica ou química e modificação das dinâmicas de
reação de polimerização (Ilie & Hickel, 2011). Recentemente, uma nova geração de
compósitos dentais, denominada bulk-fill, foi introduzida no mercado com a promessa de
poder ser polimerizada em profundidades de até 5mm, sem a necessidade de estender o tempo
de fotoativação e sem comprometer a qualidade marginal da restauração (Moorthy et al.,
2012; Furness et al., 2014; Campos et al., 2014; AlQahtani et al., 2015). Ou seja, as cavidades
podem ser preenchidas e fotoativadas em um incremento, simplificando o procedimento e
economizando tempo para o operador e o paciente (Francis et al., 2015). É sabido, contudo,
que à medida que a espessura de incremento aumenta, menos fótons de luz atingem a
superfície de fundo do compósito (AlQahtani et al., 2015).
A fim de neutralizar esse efeito, fabricantes utilizaram diferentes estratégias na
formulação desses materiais, incluindo o uso de fotoativadores mais reativos, capazes de
15
absorver energia de irradiação mesmo em camadas mais profundas; diminuição na quantidade
de partículas, tornando o compósito mais translúcido e favorecendo a transmissão de luz; uso
de monômeros singulares que funcionam como liberadores de tensão; e a incorporação de
diferentes tipos de partículas, tais como partículas pré-polimerizadas e segmentos de fibra de
vidro (AlQahtani et al., 2015; Fronza et al., 2015). Todas essas modificações foram essenciais
para que essa nova classe de compósitos, que inclui os tipos fluido e de alta viscosidade,
apresentasse maior profundidade de polimerização, contração de polimerização reduzida e
baixa tensão de contração (Lassila et al., 2012; Leprince et al., 2014; Rosatto et al., 2015).
Além disso, o preenchimento da cavidade através de aplicação única de
incremento evita a incorporação de ar e a contaminação entre as camadas de compósitos,
levando a restaurações mais compactas (Par et al., 2015). Espera-se também que estes
materiais tenham um menor módulo de elasticidade nas camadas mais profundas, o que
favoreceria a dissipação de forças durante a mastigação ou mudanças bruscas de temperaturas
(Soares et al., 2014), e consequentemente geraria menos problemas na interface adesiva.
Alguns autores reportaram desempenho favorável para os compósitos bulk-fill
(Czasch & Ilie, 2013; Ilie & Stark, 2014; Francis et al., 2015; Benetti et al., 2015). Poucos
estudos são encontrados na literatura, no entanto, que simulem, de forma mais fiel possível, o
cenário clínico de envelhecimento das restaurações bulk-fill e analisem a interferência deste
envelhecimento nas propriedades dos materiais. Assim sendo, o objetivo geral no presente
estudo foi avaliar três sistemas resinosos bulk-fill quanto às suas propriedades físicas em
comparação com um compósito convencional. Os objetivos específicos neste estudo in vitro
foram avaliar a resistência à união, a nanoinfiltração marginal e a adaptação marginal dos
materiais às paredes circundantes e de fundo da cavidade logo após o procedimento
restaurador e após envelhecimento mecânico e térmico.
16
2 ARTIGO: Evaluation of bond strength, nanoleakage and marginal adaptation of bulk-
fill resin composites submitted to thermomechanical aging
Artigo submetido ao periódico Journal of Dentistry (Anexo 2)
Renata Pereira, Débora Alves Nunes Leite Lima, Maria Cecília Caldas Giorgi, Giselle
Maria Marchi, Flávio Henrique Baggio Aguiar
ABSTRACT
Objectives: To evaluate the bond strength, nanoleakage and marginal adaptation of three
bulk-fill systems compared with a conventional composite, submitted or not to mechanical
and thermal aging. Methods: Ninety-six human molars were selected and restored in standard
class I cavity, with depth of 4 mm. The samples were divided in 8 groups (n=12), according to
composite: Filtek Z350 XT (Z350), Tetric N-Ceram Bulk Fill (TET), Filtek Bulk Fill
Posterior Restorative (FBF) and SonicFill (SF) and aging: Mechanical and thermal cycling or
No cycling – Control. The teeth (n=7) were submitted to microtensile bond strength (BS)
through beams obtained from the restored area, of which two from each tooth were selected to
assess the infiltration of silver nitrate ions. For the analysis of marginal adaptation, teeth (n=5)
were sectioned parallel and perpendicularly to the occlusal surface and replicas of the
obtained surfaces were made in epoxy resin. Fracture pattern, nanoleakage and marginal
adaptation were evaluated using scanning electron microscope. Quantitative analysis of
nanoleakage and marginal adaptation were performed with ImageJ software. Data obtained
were submitted to two-way ANOVA and Tukey’s test (ɑ = 0.05). Results: TET presented the
highest values of BS and nanoinfiltration and appopriate marginal adaptation. SF presented
the lowest BS results, low nanoinfiltration and the best marginal adaptation. Z350 and FBF
presented intermediate BS results, low nanoinfiltration and minor marginal adaptation. Aging
by mechanical and thermal cycling affected only marginal adaptation. Conclusions: Bulk-fill
composites tested may perform as well as a conventional nanocomposite and may remain
stable when aged.
Clinical Significance: Bulk fill composites present compatible physical properties with
conventional composites and stability when aged under simulated clinical conditions.
Keywords: Composite resins. Tensile strength. Dental leakage. Dental marginal adaptation.
17
INTRODUCTION
Over the years, composites have undergone several changes in their formulations
to improve their properties and facilitate the restorative procedure - reduction of particles, the
amount of the same, shape or treatment of their surfaces, changes in monomeric or chemical
structure and modification of polymerization reaction dynamics [1]. Recently, a new
generation of dental composites denominated bulk-fill was launched in the market with the
promise of being polymerized in 4 mm depth or more, without having to extend the light
exposure time and without compromising the marginal quality of the restoration [2,3]. In
other words, the cavities can be filled and photoactivated in a single increment, simplifying
the process and saving time to the operator and the patient [4,5]. It is known, however, that as
the increment thickness increases, fewer photons of light reach the bottom surface of the
composite [3].
In order to counteract this effect, manufacturers have used different strategies in
the formulation of these materials, including using more reactive photoinitiators, capable of
absorbing irradiation energy even in deeper layers; decreasing the amount of fillers,
improving composite translucency and favoring light transmission; using of singular
monomers that act as a stress-reliever; and the incorporation of different types of fillers, such
as prepolymerized particles and fiberglass rod segments [3,5]. All these modifications were
essential for this new class of composites, which includes flowable and high-viscosity types,
to present greater depth of cure, slower polymerization reaction, reduced polymerization
shrinkage and low shrinkage stress [5,6,7].
Furthermore, bulk filling reduces void incorporation and contamination between
the composite layers, leading to more compact fillings [8]. It is also expected these materials
to have lower modulus of elasticity in the deeper layers, which would favor the forces
dissipation during mastication or abrupt changes in temperature [9], and consequently would
decrease troubles in the adhesive interface.
Some authors reported favorable performance for bulk-fill composites
[4,10,11,12]. Few studies are found in literature, however, that, more reliably as possible,
simulate a clinical scenario of aging of bulk-fill restorations and analyze the interference of
this aging in the materials properties. Thus, the purpose of this study was to evaluate the bond
strength, nanoleakage and marginal adaptation of three bulk-fill systems compared with a
conventional composite, submitted or not to mechanical and thermal aging. The null
hypotheses tested were: (1) there would be no significant difference in physical properties
18
among the composite resin restorative materials; (2) the mechanical and thermal cycling
would not interfere in the tested properties.
MATERIALS AND METHODS
Teeth preparation:
Ninety-six freshly extracted healthy human third molars were selected, with the
approval by the Ethics Committee in Research of the Piracicaba Dental School, University of
Campinas (register number 53245216.10). The cleaning of the teeth’s outer surfaces was
performed by root scaling, followed by blasting with sodium bicarbonate and water (Figure
1A). After cleaning, the teeth were stored in distillated water at 37ºC until cavity preparation.
Considering that the teeth would undergo mechanical cycling and that the
periodontal ligament is an important structure for the stress distribution generated by load
application over teeth [13], it was chosen to simulate the periodontal ligament with polyether
based material (Impregum Soft, 3M ESPE, St. Paul, MN, USA). This material was chosen,
since mechanical response of a soft tissue to external stress is non-linear and viscous, similar
to the characteristics of elastomeric materials used in impression procedures [14]. Besides the
periodontal ligament simulation, roots were embedded in polystyrene resin, thus reproducing
the movement of the tooth in the alveolus and making the experiment closer to a real clinical
situation.
Thereunto, root surfaces were dipped into melted wax (Epoxiglass, Diadema, SP,
Brazil) up to 2.0 mm below the cement-enamel junction, resulting in a 0.2 to 0.3 mm thick
wax layer (Figure 1B). The wax-covered roots were submitted to dual impression, using
polyvinyl siloxane material with light and heavy consistencies materials (Express XT, 3M
ESPE, St. Paul, MN, USA) (Figure 1C). After that, the wax was removed (Figure D), the
polyether based material was inserted in the silicone impression and the tooth was introduced
under digital pressure (Figure 1E). After polymerization, the material excesses were removed
with a scalpel blade and the roots were embedded in polystyrene resin, using a PVC pipe
(Tigre tubos e conexões S/A, Joinville, SC, Brazil) as a die and a fragment of 0.40" steel rod
wire (Morelli Ortodontia, Sorocaba, SP, Brazil) with sticky wax to stabilize the dental crown
to the pipe, so that the roots were positioned laterally inside the tube and the occlusal surface
was parallel to the base of the pipe (Figures 1F-G).
19
Tooth cusps were abraded in a polishing machine (Politriz, AROTEC, São Paulo,
SP, Brazil) with 400-, 800- and 1200-grit sandpaper (3M ESPE 411Q, Sumaré, SP, Brazil),
sequentially, to obtain a flat enamel occlusal surface (Figure 1H). Class I cavities were
prepared using #56L carbide burs (KG Sorensen, Barueri, SP, Brazil) at high-speed, under
water cooling, in a custom made apparatus that allowed the standardization of the cavity
dimensions in 5mm in the mesio-distal width, 4mm in bucco-lingual and 4mm of depth
(Figure 1I). All cavosurface margins of the cavity were surrounded by enamel.
Figure 1 - Illustrative scheme of teeth preparation
A – cleaned teeth; B – root dipping in melted wax; C – dual impression
of the root; D – removal of wax; E – production of periodontal ligament
with polyether; F - tooth positioning inside PVC pipe with orthodontic
wire and wax; G - tooth inclusion in polystyrene resin; H – cusp
planning; I – cavity preparation.
Restorative procedure:
Enamel and dentin surfaces were etched with 35% phosphoric acid (Ultra-Echt,
Ultradent Products Inc., South Jordan, UT, USA) and treated with Optibond FL adhesive
system (Kerr, Orange, CA, USA), following manufacturer instructions, as follows: a layer of
OptiBond FL Prime was applied over enamel and dentin surfaces with light scrubbing motion
for 15 seconds. The surfaces were air dried for approximately 5 seconds. Thereafter, a layer of
OptiBond FL Adhesive was applied over enamel and dentin uniformly creating a thin coating.
20
The restorative procedure was performed by inserting the composites, A2 shade, in single
increment in case of bulk-fill composites, or in oblique increments of 2 mm in case of
conventional composite, into the dental cavity. SonicFill composite was placed by own sonic-
activation condensing device. Both adhesive and composites increments were light-cured by
polywave light emitting diode curing unit (VALO, Ultradent Products Inc., South Jordan, UT,
USA) in Standard mode: 1000 mW / cm² for 10 and 20 seconds, respectively. Finishing and
polishing of the occlusal surface were carried out in a polishing machine (Politriz, AROTEC,
São Paulo, SP, Brazil) with 600-, 1200- and 2000-grit silicon carbide abrasive papers (Norton
Abrasivos, Vinhedo, SP, Brazil) and felt disks containing 3, 1 and 0.5 μm diamond pastes
(Arotec Ind. Com., São Paulo, SP, Brazil). Specimens were sonicated (Ultrasound Ultrason
1440 D- Odontobrás Ind. E Com. Med. Odont. Ltda, Rio Preto, SP, Brazil) for 10 min among
each abrasive paper and paste granulation in order to remove the polishing debris.
Experimental groups (n = 12) were distributed according to the composites and
accomplishment or not of mechanical and thermal cycling. Composites details used in this
experiment are presented in Table 1.
Table 1: Composites evaluated and respective manufacturer information
Material brand
name
(abbreviation)
Manufacturer Matrix
composition Filler type
Filler
loading
(weight%/
volume%)
Shade
Filtek Z350 XT
(Z350)
3M ESPE,
Dental Products,
St. Paul, MN,
USA
Bis-GMA,
UDMA,
TEGDMA,
PEGDMA, Bis-
EMA
Silica, zirconia and
aggregated zirconia/silica
filler
78.5/63.3 A2E
Tetric N-Ceram
Bulk Fill (TET)
Ivoclar
Vivadent, AG,
Schaan,
Liechtenstein
Bis-GMA, Bis-
EMA, UDMA
Barium aluminium
silicate glass, “isofiller”
(cured dimethacrylates,
glass filler and ytterbium
fluoride), ytterbium
fluoride and spherical
mixed oxide
76-77/61
IVA –
Corres
pondin
g to A2
or A3
Filtek Bulk Fill
Posterior
Restorative
(FBF)
3M ESPE,
Dental Products,
St. Paul, MN,
USA
AUDMA, AFM,
UDMA e 1,12-
dodecane-DMA
Silica, zirconia,
aggregated zirconia/silica
filler and ytterbium
trifluoride
76.5/58.4 A2
SonicFill (SF)
Kerr Co,
Orange, CA,
USA
Bis-GMA,
TEGDMA,
EBPDMA
Silica, silicon dioxide,
barium glass, oxides 83.5/66 A2
Abbreviations: AFM: addition-fragmentation monomers; AUDMA: aromatic urethane dimethacrylate; Bis-
EMA: ethoxylated bisphenol-A dimethacrylate; Bis-GMA: bisphenol-A diglycidyl ether dimethacrylate;
EBPDMA: ethoxylated bisphenol-A dimethacrylate; PEGDMA: polyethylene glycol dimethacrylate; TEGDMA:
triethyleneglycol dimethacrylate; UDMA: urethane dimethacrylate. Data were provided by manufacturers.
21
Mechanical and thermal cycling:
After the restorative procedure, finishing and polishing, half of the samples were
submitted to 200,000 mechanical cycles with 86 N load at 2 Hz frequency in a mechanical
cycling machine (ER-37000, ERIOS, São Paulo, SP, Brazil), with the charge being applied at
the center of the restoration. They were, then, submitted to 550 thermal cycles (temperature
ranging from 5 ± 1 ° C, 37 ± 1 ° C and 55 ± 1 ° C in water chambers, for 30 seconds each
bath, with an interval of 15 seconds) in a thermal cycling device (MSCT-3e, ElQuip, São
Carlos, SP, Brazil) to simulate aging in similar conditions to the oral cavity.
Microtensile bond strength (µTBs) and Nanoleakage tests (n= 7 teeth):
Seven teeth of each group were used for microtensile and nanoleakage tests. First,
the dental crowns were separated from the root portion and were fixed in acrylic plates with
sticky wax (ASFER Ind., São Caetano do Sul, SP, Brazil). The set was stabilized to a
metallographic precision cutter (Isomet 1000, BUEHLER, Lake Bluff, IL, USA), in which a
diamond blade (Isomet Diamond Wafering Blades, Buehler Ltd., Lake Buff, IL, USA)
performed serial sections with water-cooling at 250 rpm in mesiodistal and buccolingual
directions to obtain approximately 10 beams of the restored area, from the pulpal interface,
with a cross section area of 0.7 mm². The beams were attached to a Universal Testing
Machine (EZ Test L, Shimadzu, Japan) with cyanoacrylate-based glue (Super Bond gel,
Loctite, Henkel, São Paulo, SP, Brazil) by their extremities and the microtensile bond strength
test was conducted with ± 500 kgf load cell at a crosshead speed of 0.5 mm/min until failure.
The cross-sectional area at the side of fracture was measured with a digital caliper (Mitutoyo
Corporation, Tokyo, Japan), and bond strengths were reported in MegaPascal (MPa). Bonded
beams with premature failures were not included in the tooth mean, due to their low
frequency in the experiment.
After observing the data’s normality and equality of the variances, results were
submitted to two-way statistical analysis ANOVA and Tukey test (α=0.05) and the pattern of
fracture was evaluated in SEM (LEO 435 VP, LEO Electron Microscopy Ltd, Cambridge,
UK) in: cohesive in dentin, cohesive in composite, adhesive or mixed, considering as
cohesive and adhesive fractures when at least 70% of the total area was composed by the
same pattern. Mixed fracture was stated when there was more than one pattern and none
prevailed.
22
For the nanoleakage analysis, two beams of each tooth, totaling 14 beams per
group, were selected and immersed in an aqueous solution of ammoniacal silver nitrate for 24
h, followed by 8 hours in a photo-developer solution, under fluorescent lamp, in order to
permit reduction of the diammine silver ions to metallic silver grains [15]. They were, then,
washed in distillated water and embedded in polystyrene resin to favor the polishment, which
was performed with 600-,1200-, 2000-grit silicon carbide abrasive papers and diamond pastes
(3-, 1-, and 0.25-μm granulations). Specimens were sonicated for 10 min among each abrasive
paper and paste granulation in order to remove the polishing debris. Samples were
demineralized and deproteinized with 85% phosphoric acid and 2% sodium hypochlorite,
respectively, and thereat, dehydrated in serial ethanol solutions (25%, 50%, 75%, 90%, and
100% for 10 min in each concentration). They were, then, coated with carbon and analyzed
using SEM operated at 20 kV under high vacuum power, yielding backscattered electron
images. Images were recorded to assess quantitatively the infiltrated area with the support of
ImageJ software (ImageJ, National Institute of Health, Bethesda, USA). The SEM image
scale bar was used for distance calibration, the lengths of infiltrated tags and segments were
obtained in millimeters and then converted to percentage of the total length of the interface.
After observing the data’s normality and equality of the variances, results were submitted to
two-way statistical analysis ANOVA and Tukey test (α=0.05).
Marginal adaptation test (n= 5 teeth):
For this test, 5 teeth of each group were used. The teeth were removed from the
polystyrene resin with the aid of a dental surgery forceps #17 (Millennium, GOLGRAN, São
Caetano do Sul, SP, Brazil) and were fixed to acrylic plates with hot melt adhesive (Rhamos e
Brito, Guarulhos, SP, Brazil). Then, the dental crowns were sectioned up to 1,5 mm below the
occlusal surface using a diamond blade with water-cooling at 300 rpm in a metallographic
precision cutter in order to obtain a flat occlusal surface, which was polished with 600-,1200-,
2000-grit silicon carbide abrasive papers and diamond pastes (3-, 1-, and 0.25-μm
granulations). The samples were sonicated for 10 min to remove polishing debris. After,
impressions of the occlusal polished surfaces were taken using a polyvinyl siloxane material
with light and heavy consistencies materials and were poured with epoxy resin (Epoxicure
Resin, Buehler, Lake Bluff, IL, USA). The samples were, then, fixed to acrylic plates and
sectioned again, longitudinally, in the bucco-lingual direction, right in the middle of the
restoration, by the diamond blade. One of the halves was randomly selected and fixed to
23
acrylic plate to favor the polishment, which was performed using the same protocol applied
for occlusal surface. Impressions of the polished, half-restored surfaces were taken and
poured with epoxy resin as well. The polymeric replicas of the surfaces were sputter-coated
with gold in a vacuum evaporator (SCD 050, Bal-Tec AG, Balzers, Liechtenstein) in order to
record images in SEM of the presence of interfacial gaps in the external and internal cavity
walls. The extension of the gaps was quantitatively evaluated using ImageJ software. The
SEM image scale bar was used for distance calibration, the lengths of debonded segments
were obtained in millimeters and then converted to percentage of the total length of the
interface. It is important to note that only the length of the debonded segments was evaluated.
The width of each gap was not considered in this study. After observing the data’s normality
and equality of the variances, results were submitted to two-way statistical analysis ANOVA
and Tukey test (α=0.05).
RESULTS
Microtensile bond strength test:
Analysis showed statistical significance for the factor “composite” (ρ = 0,045),
but not for performed treatment (p = 0,6462), neither for interaction between factors (p =
0,9834). The results are presented in Table 2, which illustrates that TET presented higher
values of bond strength when compared to SF. Z350 and FBF did not differ significantly from
the other groups. Furthermore, the results obtained by the aged groups were not statistically
different from the groups which were not submitted to mechanical and thermal cycling.
Table 2 - Mean (standard deviation) bond strength (MPa) of composites at
different treatments
Composite Treatment
Tukey None
Mechanical and thermal cycling
Filtek Z350 XT 32,31 (6,11) 30,44 (7,81) ab Tetric N-Ceram 37,04 (8,03) 36,82 (11,09) a Filtek Bulk Fill 32,04 (8,35) 31,99 (7,42) ab SonicFill 28,73 (10,46) 26,78 (6,44) b
Tukey A A
Means followed by distinct letters (uppercase in horizontal and lowercase in vertical)
differ from each other (p≤0.05).
24
Fracture pattern:
The observed failure patterns are shown in Figure 2. The main failure occurred
were of cohesive in composite or adhesive types for all composites, except for not aged FBF,
which presented higher percentage of mixed failures. In general, when submitted to
mechanical and thermal cycling, the adhesive failures rate increased for all composites, except
for TET, which had its mixed failures rate increased. In the main, the cohesive in dentin
failures have presented low rates for all tested groups.
Figure 2 – Failure pattern (%) of composites at different treatments
Nanoleakage Analysis:
Analysis showed statistical significance for the factor “composite” (ρ = 0,0013),
but not for performed treatment (p = 0,1137), neither for interaction between factors (p =
0,9625). The results are presented in Table 3, which illustrates that the highest infiltration
values were obtained by TET. Z350, FBF and SF presented the lowest values of infiltration
and did not differ significantly from each other. Furthermore, the results obtained by the aged
groups were not statistically different from the groups which were not submitted to
mechanical and thermal cycling. Figure 3 shows a representative SEM micrograph of
nanoleakage.
11,53 16,66 12,9
31,57 37,5
21,05 25 10,52
36,53
41,66
25,8
24,56 20,83 50
15,9 44,73
13,46 4,16
3,22
8,77 12,5
10,52
20,45
7,89
38,46 37,5
58,06
35,08 29,16 18,42
38,63 36,84
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
NotagedZ350
AgedZ350
NotagedTET
AgedTET
NotagedFBF
AgedFBF
Notaged
SF
AgedSF
Cohesive in composite
Cohesive in dentin
Adhesive
Mixed
25
Table 3 - Mean (standard deviation) nanoleakage (% length of entire
interface) of composites at different treatments
Composite Treatment
Tukey None
Mechanical and thermal cycling
Filtek Z350 XT 16,69 (7,63) 14,15 (6,28) b
Tetric N-Ceram 24,26 (9,44) 21,56 (7,59) a
Filtek Bulk Fill 15,6 (2,93) 14,13 (5,35) b SonicFill 16,31 (7,65) 12,35 (3,57) b
Tukey A A
Means followed by distinct letters (uppercase in horizontal and lowercase in vertical)
differ from each other (p≤0.05).
Figure 3 - Representative scanning electron micrographs (SEM) of nanoleakage interfacial
expression
Representative image of nanoleakage presented by Z350, when submitted to mechanical and thermal cycling.
Image on the right side represent magnification (1500x) of the left side image (x500). C - composite; D –
dentin; Arrows – silver nitrate infiltration.
Marginal adaptation:
The percentages of internal marginal disadaptation (total gap formation length
as a function of total length of the internal walls interface) are presented in Table 4.
Analysis showed statistical significance for the factor “composite” (ρ = 0,0001), for
performed treatment (p < 0,0001), as well as for interaction between factors (p = 0,0033).
Table 5 shows the percentages of external marginal disadaptation (total gap formation
length as a function of total length of the external walls interface). Analysis showed
26
statistical significance for the factor “composite” (ρ < 0,0001), for performed treatment (p =
0,0023), as well as for interaction between factors (p = 0,0042).
Table 4 illustrates that, when not submitted to mechanical and thermal cycling,
Z350 and FBF were the composites which had the highest incidence of internal gap
formation, while the lowest internal gap formation was obtained by TET and SF. When
submitted to mechanical and thermal cycling, FBF presented the highest values of internal
gap formation, while SF presented the lowest ones. Z350 showed intermediate incidence and
TET did not differ significantly from FBF and Z350. All the composites presented
significantly increased gap formation values when aged by mechanical and thermal cycling.
Table 5, in turn, shows that Z350 was the composite which presented the highest
external debonded segments percentages, regardless of being submitted or not mechanical and
thermal cycling. When not aged by cycling, the other composites presented the lowest values
of gap formation and did not differ significantly from each other. When submitted to
mechanical and thermal cycling, TET and SF presented the lowest values. FBF did not differ
significantly from the others. In the case of composites adaptation to external walls, unlike the
adaptation to internal walls, Z350 and SF presented significantly decreased marginal gap
formation values when aged by mechanical and thermal cycling. TET and FBF did not present
statistical difference between the aged composites and not aged ones.
Figures 4 and 5 show representatives SEM micrographs of marginal
disadaptations. Regarding the internal marginal disadaptation, not aged Z350 presented gaps
mainly in buccal and/or lingual walls (BLW), followed by internal angles (IA). Not aged TET
and FBF showed gaps along the pulpal wall (PW), in IA and BLW, all in the same proportion.
In all of the others groups were observed gaps mainly along the PW, followed by gaps in IA.
Table 4 - Mean (standard deviation) of internal marginal
disadaptation (% length of entire internal interface) of
composites at different treatments
Composite Treatment
None Mechanical and thermal cycling
Filtek Z350 XT 12,23 (3,31) Ba 38,78 (5,61) Ab Tetric N-Ceram 5,98 (2,06) Bb 48,91 (7,82) Aab Filtek Bulk Fill 12,36 (1,83) Ba 59,27 (15,19) Aa SonicFill 5,38 (0,92) Bb 27,47 (9,44) Ac
Means followed by distinct letters (uppercase in horizontal and
lowercase in vertical) differ from each other (p≤0.05).
27
Table 5 - Mean (standard deviation) of external marginal
disadaptation (% length of entire external interface) of
composites at different treatments
Composite Treatment
None Mechanical and thermal cycling
Filtek Z350 XT 16,61 (3,94) Aa 8,55 (2,77) Ba Tetric N-Ceram 2,21 (2,91) Ab 1,32 (1,67) Ab Filtek Bulk Fill 4,47 (2,85) Ab 5,72 (3,67) Aab SonicFill 6,21 (2,06) Ab 2,40 (0,64) Bb
Means followed by distinct letters (uppercase in horizontal and
lowercase in vertical) differ from each other (p≤0.05).
Figure 4 - Representative scanning electron micrographs (SEM) of composite marginal
adaptation to internal walls
On the top: Representative image of FBF not submitted to mechanical and thermal cycling. On the
bottom: Representative image of FBF submitted to mechanical and thermal cycling. Images on the
right side represent magnification (100x) of the left side images (x60). C - composite; D – dentin; E –
enamel. Arrows – gaps located at buccal and/or lingual wall (BLW), pulpal wall (PW) and internal
angle (IA).
28
Figure 5 - Representative scanning electron micrographs (SEM) of composite marginal
adaptation to external walls
Representative image of TET not submitted to mechanical and thermal cycling. Images on the right side
represent magnification (70x) of the left side images (x46). C - composite; E – enamel. Arrow – gaps located
at external wall.
DISCUSSION
The first null hypothesis, that there would be no significant difference in
properties among the composite resin restorative materials, was rejected. µTBs evaluation
showed that TET presented higher values of bond strength when compared to SF, while Z350
and FBF did not differ significantly from the other materials. TET has relatively increased
translucency [16], which can play a role on its good depth of cure by reducing light scattering
and improving the incident light transmission [11,17]. According to manufacturer
information, this material also has a pre-polymer shrinkage stress reliever, the photo initiator
Ivocerin® (polymerisation booster), which is able to polymerize the material in higher depths,
and a composition of organic matrix (Bis-GMA, Bis-EMA and UDMA), which exhibit low
polymerisation shrinkage by volume. The presence of pre-polymerized particles and the
increased filler content of TET can also, to a certain extent, reduce the polymerization
shrinkage by increasing the filler-to-monomer ratio [18,19]. Combining the benefits of TET
with the remarkable performance of a gold-standard total-etch adhesive system may have
provided superior resistance to challenging stresses, which explains the good marginal
adaptation, as well as µTBs results. Evidence of this is given by the fracture pattern obtained
by TET, which was predominantly cohesive in composite, mainly when the composite was
29
not submitted to mechanical and thermal cycling. Namely, TET’s bond strength was sufficient
to avoid high rates of adhesive failures.
The highest values of nanoleakage obtained by TET, although it presented great
bond strength and marginal sealing, can be explained by its filler content. TET contains
prepolymerized fillers that are substantially organic, which means that the inorganic filler
content is consistently lower [10]. This may have increased water sorption by the composite,
which reached the adhesive and nanometric spaces located inside the hybrid layer, making the
adhesive polymers act, even after polymerization, as semi-permeable membranes, increasing
silver uptake [20]. It was reported that even in the absence of gaps, silver ions could penetrate
the hybrid layer, in microporosity zones [21]. A higher nanoleakage, in short-therm, may not
mean bad performance of the restoration, but in long-term can compromise the stability of
composite-dentin bond, due to penetration of bacterial products or oral fluids along the
interface and consequent hydrolytic breakdown of either adhesive or collagen within the
hybrid layer [22].
Few studies exist regarding the microtensile bond strength of SF, and although it
has the highest filler content of all the materials used in this study and it is known that an
increased filler content may reduce the polymerization shrinkage [17], the high inorganic
filler content may also represent a challenge for the composite curing. Light attenuation
within the composite is the result of complicated scattering and absorption process by
constituents, such as monomer, pigments, and fillers [17,23]. Thus, the high filler content may
reduce light transmission and impair the curing process, mainly in bulk-fill systems, which
are polymerized in higher depths. Furthermore, previous studies [12,24,25] found for SF a
depth of cure ranging from 3.43 to 3.67mm. Considering that the cavity tested in this study
had 4mm of depth, the SF low values of bond strength may be explained by a possible
inefficient polymerization in the cavity bottom. Nevertheless, SF is placed by a sonic-
activation condensing device, which provides flow consistency during application. Peutzfeldt
and Asmussen, in 2004 [26], showed that the degree of fluidity, when applying the composite,
influences the marginal adaptation; increased fluidity of the composite makes it adhere better
to the walls of the cavity. The fluidity consistence of SF while it is applied, combined with its
increased filler content and low modulus of elasticity, feature of low-viscosity composites
[27], may have provided for this material low nanoinfiltration values and the best results of
marginal adaptation.
Z350 and FBF presented intermediate bond strength results, low nanoinfiltration
values, but relatively minor marginal adaptation. According to manufacturer information, both
30
composites present a composition of organic matrix and filler content, which act in
combination, to lower polymerization shrinkage. Furthermore, they have increased inorganic
filler loading, which can also reduce the polymerization shrinkage by increasing the filler-to-
monomer ratio [18,19]. These benefits may have provided for both composites intermediate
bond strength results, as well as low nanoinfiltration values. The highest values of marginal
disadaptation can be explained for FBF, as the composite presents a combination of singular
monomers, such as 1, 12-Dodecanediol dimethacrylate and AFM. The first one is defined by
the manufacturer as a high-modulus resin with good flexibility and impact resistance. A
hidrophobic monomer with high elastic modulus, in this case, may favor the polymerization
shrinkage [6], thus impairing the marginal adaptation. AFM, in turn, is defined by the
manufacturer as a class of compounds called addition-fragmentation monomers, whose
structure and action means are not well informed by the manufacturer data. As already
mentioned by Leprince et al., 2014 [6], since specific material composition is unknown, the
impact of each specific compound on the final material properties is difficult to predict.
The highest values of marginal disadaptation can be explained for Z350, since,
unlike the bulk-fill composites, this one was placed by incremental build-up technique.
Versluis et al. in 1996 [28], assessed the developing stress fields for different incremental
filling techniques using finite element analysis, concluding that the incremental filling
technique increased the deformation of the restored tooth and led to a more highly stressed
tooth-composite structure. Abbas et al. in 2003 [29], also showed, in cuspal deflection
measurements using premolars, that multiple increments induced greater cuspal movement
than a single increment. Despite the controversy over the advantages of single incremental
filling technique, it is worthy to notice that in this study, FBF, even being applied in single
increment, presented similar properties performances to Z350, applied by incremental
technique. It is still necessary to consider that Z350 presents PEGDMA and TEGDMA in its
composition, which, even in small quantities, present low molecular weight. This result in
high number of double bonds per unit weight and create a high degree of crosslinking,
creating a rigid resin with a relatively high shrinkage [30]. Also, the majority of TEGDMA
monomers elute within few hours, which may contribute to microgap formations [31].
In order to simulate the restoration maintenance in the oral cavity and evaluate its
interference in the physical properties of the composites, in this study the samples were
submitted to mechanical and thermal cycling. According to ISO standards (TR 11405), 500
thermal cycles in water with temperatures between 5 ° C and 55 ° C are considered as an
adequate test for dental materials aging, with 10,000 cycles corresponding to a year of in vivo
31
approach [32]. Leloup et al., 2001 [33], on the other hand, analyzing literature data on
adherence tests of dentin-bonding systems, reported that short-term thermocycling (500×,
5 °C/55 °C) does not significantly alter the ultimate bond strength values. In this study, even
submitting the composites to mechanical loading, besides thermal cycling, the only property
affected was marginal adaptation. Accordingly, the second null hypothesis must be rejected.
The simulation of periodontal ligament may play a role in this situation, once the periodontal
ligament is an important structure for the stress distribution generated by load application over
teeth [13]. When load is applied, periodontal fibers are compressed. As the tooth is forced
within its alveolus, the resistance of these fibers against tooth displacement is progressively
increased, until the maximum load resistance is achieved, so the periodontal membrane gets
rigid, the load is transferred to the bone support and the stress is distributed to bone in all root
surfaces [14]. A greater influence of periodontal ligament simulation was already noted in
previous study involving mechanical load [14,34].
Regarding the marginal adaptation analysis, when composites were aged, the
incidence of gaps formation was much higher in the internal than external walls. When
photoactivation is performed with the light tip positioned as closer as possible to the
composite, high irradiance photoactivation initiates a multitude of growth centers of polymers
with higher cross-linking density on the top surface [35]. The bottom surface, along the pulpal
wall, in turn, is reached by low irradiant energy, which causes a decrease in the degree of
conversion and cross-linking density [36,37]. As already discussed by Soares et al. in 2014
[9], the high polymerization rate on the top surface and the lower polymerization rate on the
bottom surface may lead to higher bond strength around the external walls, where the light
curing tip is nearer, and lower bond strength on the internal walls. It is possible that better
polymerization on the top creates a stress flow of polymerization shrinkage from internal to
external walls which opens a gap in the pulpal interface [9]. Similar outcomes have been
already reported by Campos et al. [2] and Fronza et al. [5]. In the face of mechanical and
thermal challenges, the tendency is that the gap in the pulpal interface gets wider. It is
possible that this increase in pulpal gap width may cause certain stress relief and, the external
interface, that presented previously debonded areas, gets better accommodated by composites.
This would explain the external marginal disadaptation results obtained by aged Z350 and SF.
Although the application of in vitro study results in clinical practice is limited,
considering that standardizations common to laboratory practice are not expected in clinical
general practice, the results of this study prove conceptions from previous researches, which
point out bulk-fill composites systems as promising restorative alternatives.
32
CONCLUSIONS
It can be concluded that investigated bulk-fill composites (TET, FBF, SF)
demonstrated, in general, lower incidence of gap formation and similar bond strength and
nanoleakage values when compared to conventional composite (Z350), regardless of being
submitted or not to mechanical and thermal cycling. Thus, it is possible to affirm that bulk-fill
composites tested may perform as well as a conventional nanocomposite and may present a
good behavior in face of aging challenges.
ACKNOWLEDGEMENTS
This work was supported by the State of São Paulo Research Foundation
(FAPESP) (No. 2015/23201-7). The authors are grateful to Dr. Francisco André Ossamu
Tanaka and Mr. Renato Salaroli (NAP/MEPA—ESALQ/USP) for scanning electron
microscopy support, and to Mr. Marcos Blanco Cangiani and Mr. Marcelo Corrêa Maistro for
their technical assistance. The authors declare no conflicts of interest with respect to the
authorship and/or publication of this article.
REFERENCES
[1] Ilie N, Hickel R. Resin composite restorative materials. Aust Dent J, 2011 Jun; 56:1:59-
66.
[2] Campos EA, Ardu S, Lefever D, Jassé FF, Bortolotto T, Krejci I. Marginal adaptation of
class II cavities restored with bulk-fill composites. J Dent, 2014 May; 42(5):575-81.
[3] AlQahtani MQ, Michaud PL, Sullivan B, Labrie D, AlShaafi MM, Price RB. Effect of
high irradiance on depth of cure of a conventional and a bulk fill resin-based composite. Oper
Dent, 2015 Nov-Dec; 40(6):662-72.
[4] Francis AV, Braxton A, Ahmad W, Tantbirojn D, Simon J, Versluis A. Cuspal Flexure
and Extent of Cure of a Bulk-fill flowable base composite. Oper Dent, 2015 Sep-Oct;
40(5):515-23.
[5] Fronza BM, Rueggeberg FA, Braga RR, Mogilevych B, Soares LE, Martin
AA, Ambrosano G, Giannini M. Monomer conversion, microhardness, internal
33
marginal adaptation, and shrinkage stress of bulk-fill resin composites. Dent Mater, 2015
Dec; 31(12):1542-51.
[6] Leprince JG, Palin WM, Vanacker J, Sabbagh J, Devaux J, Leloup G. Physico-mechanical
characteristics of commercially available bulk-fill composites. J Dent, 2014 Aug; 42(8):993-
1000.
[7] Rosatto CM, Bicalho AA, Veríssimo C, Bragança GF, Rodrigues MP, Tantbirojn D,
Versluis A, Soares CJ. Mechanical properties, shrinkage stress, cuspal strain and fracture
resistance of molars restored with bulk-fill composites and incremental filling technique. J
Dent, 2015 Dec; 43(12):1519-28.
[8] Par M, Gamulin O, Marovic D, Klaric E, Tarle Z. Raman spectroscopic assessment of
degree of conversion of bulk-fill resin composites--changes at 24 hours post cure. Oper Dent,
2015 May-Jun; 40(3):E92-101.
[9] Soares GP, Ambrosano GM, Lima DA, Marchi GM, Correr-Sobrinho L, Lovadino JR,
Aguiar FH. Effect of light polymerization time, mode, and thermal and mechanical load
cycling on microleakage in resin composite restorations. Lasers Med Sci, 2014 Mar;
29(2):545-50.
[10] Czasch P, Ilie N. In vitro comparison of mechanical properties and degree of cure of bulk
fill composites. Clin Oral Investig, 2013 Jan; 17:227-235.
[11] Ilie N, Stark K. Curing behaviour of high-viscosity bulk-fill composites. J Dent, 2014
Aug; 42(8):977-85.
[12] Benetti AR, Havndrup-Pedersen, Honore D, Pedersen MK, Pallesen U. Bulk-fill Resin
composites: polymerization contraction, depth of cure, and gap formation. Oper Dent, 2015
Mar-Apr; 40(2):190-200.
[13] Rosentritt M, Fürer C, Behr M, Lang R, Handel G. Comparison of in vitro fracture
strength of metallic and tooth-coloured posts and cores. J Oral Rehabil, 2000 Jul; 27(7):595-
601.
[14] Soares CJ, Pizi EC, Fonseca RB, Martins LR. Influence of root embedment material and
periodontal ligament simulation on fracture resistance tests. Braz Oral Res, 2005 Jan-Mar;
19(1):11-6.
[15] Tay FR, Pashley DH, Yoshiyama M. Two modes of nanoleakage expression in single-
step adhesives. J Dent Res, 2002 Jul; 81(7):472-6.
[16] Kim EH, Jung KH, Son SA, Hur B, Kwon YH, Park JK. Effect of resin thickness on the
microhardness and optical properties of bulk-fill resin composites. Restor Dent Endod, 2015
May; 40(2):128-35.
34
[17] Son SA, Park JK, Seo DG, Ko CC, Kwon YH. How light attenuation and filler content
affect the microhardness and polymerization shrinkage and translucency of bulk-fill
composites? Clin Oral Investig, 2016 Jul; 31.
[18] Braga RR, Ballester RY, Ferracane JL. Factors involved in the development of
polymerization shrinkage stress in resin-composites: a systematic review. Dent Mater, 2005
Oct; 21(10):962-70.
[19] Kleverlaan CJ, Feilzer AJ. Polymerization shrinkage and contraction stress of dental
resin composites. Dent Mater, 2005 Dec; 21(12):1150-7.
[20] Tay FR, King NM, Chan KM, Pashley DH. How can nanoleakage occur in self-etching
adhesive systems that demineralize and infiltrate simultaneously? J Adhes Dent, 2002 Winter;
4(4):255-69.
[21] Duarte S Jr, Perdigão J, Lopes MM. Effect of dentin conditioning time on nanoleakage.
Oper Dent, 2006 Jul-Aug; 31(4):500-11.
[22] Eick JD, Gwinnett AJ, Pashley DH, Robinson SJ. Current concepts on adhesion to
dentin. Crit Rev Oral Biol Med, 1997; 8(3):306–35.
[23] Emami N, Sjödahl M, Söderholm KJ. How filler properties, filler fraction, sample
thickness and light source affect light attenuation in particulate filled resin composites. Dent
Mater, 2005 Aug; 21(8):721–730.
[24] Garcia D, Yaman P, Dennison J, Neiva GF. Polymerization shrinkage and depth of cure
of bulk-fill flowable composite resins. Oper Dent, 2014 Jul-Aug; 39(4):441-448.
[25] Ibarra ET, Lien W, Casey J, Dixon SA, Vandewalle KS. Physical properties of
a new sonically placed composite resin restorative material. Gen Dent, 2015 May-Jun;
63(3):51-6.
[26] Peutzfeldt A, Asmussen E. Determinants of in vitro gap formation of resin composites. J
Dent, 2004 Feb; 32(2):109-15.
[27] Orlowski M, Tarczydlo B, Chalas R. Evaluation of marginal integrity of four bulk-fill
dental composite materials: in vitro study. Scientific World Journal, 2015; 2015:701262.
[28] Versluis A, Douglas WH, Cross M, Sakaguchi RL. Does an incremental filling technique
reduce polymerization shrinkage stresses? J Dent Res, 1996 Mar; 75(3):871-8.
[29] Abbas G, Fleming GJ, Harrington E, Shortall AC, Burke FJ. Cuspal movement and
microleakage in premolar teeth restored with a packable composite cured in bulk or
increments. J Dent, 2003 Aug; 31(6):437-44.
35
[30] Sharma RD, Sharma J, Rani A. Comparative evaluation of marginal adaptation between
nanocomposites and microhybrid composites exposed to two ligh cure units. Indian J Dent
Res, 2011 May-Jun; 22(3):495.
[31] Nalcaci A, Uluosoy N, Atakol O. Time-based elution of TEGDMA and BisGMA from
resin composite cured with LED, QTH and high -intensity QTH lights. Oper Dent, 2006 Mar-
Apr; 31(2):197-203.
[32] Xie C, Han Y, Zhao XY, Wang ZY, He HM. Microtensile bond strength of one- and
two-step self-etching adhesives on sclerotic dentin: the effects of thermocycling. Oper Dent,
2010 Sep-Oct; 35(5):547–55.
[33] Leloup G, D’Hoore W, Bouter D, Degrange M, Vreven J. Meta-
analytical review of factors involved in dentin adherence. J Dent Res, 2001 Jul; 80(7):1605-
14.
[34] Rees JS. An investigation into the importance of the periodontal ligament and alveolar
bone as supporting structures in finite element studies. J Oral Rehabil, 2001 May; 28(5):425-
32.
[35] Schneider LF, Pfeifer CS, Consani S, Prahl SA, Ferracane JL. Influence of photoinitiator
type on the rate of polymerization, degree of conversion, hardness and yellowing of dental
resin composites. Dent Mater, 2008 Sep; 24(9):1169-77.
[36] Aguiar FHB, Lazzari CR, Lima DANL, Ambrosano GMB, Lovadino JR. Effect of light
curing tip distance and resin shade on microhardness of a hybrid resin composite. Braz Oral
Res, 2005 Oct-Dec; 19(4):302-6.
[37] Calheiros FC, Daronch M, Rueggeberg FA, Braga RR. Influence of irradiant energy on
degree of conversion, polymerization rate and shrinkage stress in an experimental resin
composite system. Dent Mater, 2008 Sep; 24(9):1164-8.
36
3 CONCLUSÃO
Pôde-se concluir que os compósitos bulk-fill testados (TET, FBF e SF)
demonstraram, em geral, menor incidência de formação de fendas e valores de resistência de
união e nanoinfiltração similares quando comparados ao compósito convencional (Z350),
independentemente de serem submetidos ou não a ciclagem mecânica e térmica. Assim, é
possível afirmar que os compósitos bulk-fill testados podem funcionar tão bem quanto
nanocompósitos convencionais e podem apresentar um bom comportamento diante dos
desafios do envelhecimento.
37
* De acordo com as normas da UNICAMP/FOP, baseadas na padronização do International Committee of
Medical Journal Editors - Vancouver Group. Abreviatura dos periódicos em conformidade com o PubMed.
REFÊRENCIAS*
1. AlQahtani MQ, Michaud PL, Sullivan B, Labrie D, AlShaafi MM, Price RB. Effect of
high irradiance on depth of cure of a conventional and a bulk fill resin-based
composite. Oper Dent, 2015 Nov-Dec; 40(6):662-72.
2. Annusavice KJ. Philips Science of Dental Materials. St. Louis: Elsevier; 2013.
3. Benetti AR, Havndrup-Pedersen C, Honoré D, Pedersen MK, Pallesen U. Bulk-fill
resin composites: polymerization contraction, depth of cure, and gap formation. Oper
Dent, 2015 Mar-Apr; 40(2): 190-200.
4. Bicalho AA, Valdívia AD, Barreto BC, Tantbirojn D, Versluis A, Soares CJ.
Incremental filling technique and composite material--part II: shrinkage and shrinkage
stresses. Oper Dent, 2014 Mar-Apr; 39(2): E83-92.
5. Borges AL, Borges AB, Xavier TA, Bottino MC, Platt JA. Impact of quantity of resin,
C-factor, and geometry on resin composite polymerization shrinkage stress in Class V
restorations. Oper Dent, 2014 Mar-Apr; 39(2):144-51.
6. Calheiros FC, Braga RR, Kawano Y, Ballester RY. Relationship between contraction
stress and degree of conversion in restorative composites. Dent Mater, 2004 Dec;
20(10):939-46.
7. Campos EA, Ardu S, Lefever D, Jassé FF, Bortolotto T, Krejci I. Marginal adaptation
of class II cavities restored with bulk-fill composites. J Dent, 2014 May; 42(5): 575-
81.
8. Czasch P, Ilie N. In vitro comparison of mechanical properties and degree of cure of
bulk fill composites. Clin Oral Investig, 2013 Jan; 17:227-235.
9. Ferracane JL, Hilton TJ. Polymerization stress – Is it clinically meaningful? Dent
Mater, 2016 Jan; 32(1):1-10.
10. Ferracane JL, Mitchem JC. Relationship between composite contraction stress and
leakage in Class V cavities. Am J Dent, 2003 Aug; 16(4):239-43.
11. Ferracane JL. Resin Composite – state of the art. Dent Mater, 2011 Jan; 27(1): 29-38.
12. Francis AV, Braxton A, Ahmad W, Tantbirojn D, Simon J, Versluis A. Cuspal
Flexure and Extent of Cure of a Bulk-fill flowable base composite. Oper Dent. 2015
Sep-Oct; 40(5):515-23.
38
13. Fronza BM. Avaliação de propriedades físico-químicas de compósitos resinosos bulk-
fill [dissertação]. Piracicaba: Faculdade de Odontologia de Piracicaba, Universidade
Estadual de Campinas; 2015.
14. Fronza BM, Rueggeberg FA, Braga RR, Mogilevych B, Soares LE, Martin
AA, Ambrosano G, Giannini M. Monomer conversion, microhardness, internal
marginal adaptation, and shrinkage stress of bulk-fill resin composites. Dent Mater,
2015 Dec; 31(12):1542-51.
15. Furness A, Tadros MY, Looney SW, Rueggeberg FA. Effect of bulk/incremental fill
on internal gap formation on bulk-fill composites. J Dent, 2014 Apr; 42(4):439-49.
16. Hilton TJ, Ferracane JL. Cavity preparation factors and microleakage of Class II
composite restorations filled at intraoral temperatures. Am J Dent, 1999 Jun;
12(3):123-30.
17. Ilie N, Hickel R. Resin composite restorative materials. Aust Dent J, 2011 Jun; 56
Suppl 1:59-66.
18. Ilie N, Stark K. Curing behaviour of high-viscosity bulk-fill composites. J Dent, 2014
Aug; 42(8):977-85.
19. Labella R, Lambrechts P, Van Meerbeek B, Vanherle G. Polymerization shrinkage
and elasticity of flowable composites and filled adhesives. Dent Mater, 1999 Mar;
15(2):128-37.
20. Lassila LV, Nagas E, Vallittu PK, Garoushi S. Translucency of flowable bulk-
filling composites of various thicknesses. Chin J Dent Res, 2012; 15(1): 31-5.
21. Leprince JG, Palin WM, Mullier T, Devaux J, Vreven J, Leloup G. Investigating filler
morphology and mechanical properties of new low-shrinkage resin composite types. J
Oral Rehab 2010 May; 37(5):364-376.
22. Leprince JG, Palin WM, Vanacker J, Sabbagh J, Devaux J, Leloup G. Physico-
mechanical characteristics of commercially available bulk-fill composites. J Dent,
2014 Aug; 42(8): 993-1000.
23. Lutz E, Krejci I, Oldenburg TR. Elimination of polymerization stresses at the margins
of posterior composite resin restorations: a new restorative technique. Quintessence
Int, 1986 Dec; 17(12):777-84.
24. Moorthy A, Hogg CH, Dowling AH, Grufferty BF, Benetti AR, Fleming GJ. Cuspal
deflection and microleakage in premolar teeth restored with bulk-fill flowable resin-
based composite basematerials. J Dent, 2012 Jun; 40(6): 500-5.
39
25. Par M, Gamulin O, Marovic D, Klaric E, Tarle Z. Raman spectroscopic assessment of
degree of conversion of bulk-fill resin composites--changes at 24 hours post cure.
Oper Dent, 2015 May-Jun; 40(3):E92-101.
26. Pollack BF. Class II composites: 1987 thoughts and techniques. NY State Dent J, 1987
May; 53(5):25-7.
27. Rees JS, Jagger DC, Williams DR, Brown G, Duguid W. A reappraisal of
the incremental packing technique for light cured composite resins. J Oral
Rehabil, 2004 Jan; 31(1): 81-4.
28. Rosatto CM, Bicalho AA, Veríssimo C, Bragança GF, Rodrigues MP, Tantbirojn D,
Versluis A, Soares CJ. Mechanical properties, shrinkage stress, cuspal strain and
fracture resistance of molars restored with bulk-fill composites and incremental filling
technique. J Dent, 2015 Dec; 43(12):1519-28.
29. Rueggeberg FA. Contemporary issues in photocuring. Compend Contin Educ Dent
Suppl, 1999; (25):S4-15; quiz S73.
30. Rueggeberg FA. From vulcanite to vinyl, a history of resins in restorative dentistry. J
Prosthet Dent, 2002 Apr; 87(4):364-79.
31. Soares GP, Ambrosano GM, Lima DA, Marchi GM, Correr-Sobrinho L, Lovadino JR,
Aguiar FH. Effect of light polymerization time, mode, and thermal and mechanical
load cycling on microleakage in resin composite restorations. Lasers Med Sci. 2014
Mar; 29(2):545-50.
32. Stansbury JW. Curing dental resins and composites by photopolymerization. J Esthet
Dent, 2000; 12(6):300-308.
33. Stansbury JW. Cyclopolymerizable monomers for use in dental resin composites. J
Dent Res, 1990 Mar; 69(3):844-8.
34. Stansbury JW. Dimethacrylate network formation and polymer property evolution as
determined by the selection of monomers and curing conditions. Dent Mater, 2012
Jan; 28(1):13-22.
35. Tirtha R, Fan PL, Dennison JB, Powers JM. In vitro depth of cure of photoactivated
composites. J Dent Res, 1982 Oct; 61(10):1184-7.
36. Versluis A, Douglas WH, Cross M, Sakaguchi RL. Does an incremental filling
technique reduce polymerization shrinkage stresses? J Dent Res, 1996 Mar;
75(3):871-8.
37. Yamazaki LC, De Vito Moraes AG, Barros M, Lewis S, Francci C, Stansbury JW,
Pfeifer CS. Polymerization Development of “low-shrink” resin composites: Reaction
40
kinetics, polymerization stress and quality of network. Dent Mater, 2013 Sep;
29(9):169-79.
38. Yap AU. Effectiveness of polymerization in composite restoratives claiming bulk
placement: impact of cavity depth and exposure time. Oper Dent, 2000 Mar-Apr;
25(2):113-20.
41
APÊNDICE 1 - Delineamento experimental e metodologia ilustrada
DELINEAMENTO EXPERIMENTAL:
Unidades Experimentais: 96 terceiros molares humanos
Fator em estudo:
- Compósito em quatro níveis:
Filtek Z350 XT (3M - ESPE)
Tetric N-Ceram Bulk Fill (Ivoclar - Vivadent)
Filtek Bulk Fill Posterior Restorative (3M - ESPE)
SonicFill (KAVO - Kerr)
- Tratamento realizado:
Ciclagens mecânica e térmica
(Nenhuma Ciclagem) Controle
Forma de designar o tratamento às unidades experimentais: Por processo aleatório
Variável resposta:
Resistência à microtração em MPa
Padrão de fratura – análise qualitativa
Nanoinfiltração marginal – análise quantitativa
Formação de fendas – análise quantitativa.
42
METODOLOGIA ILUSTRADA:
Preparo dos dentes:
Para a realização deste estudo, foram selecionados 96 terceiros molares humanos
hígidos recém-extraídos, com aprovação do Comitê de Ética em Pesquisa da FOP –
UNICAMP afiliado à Comissão de Ética em Pesquisa – CONEP (53245216.10). A limpeza
das superfícies externas dos dentes foi realizada com cureta periodontal 5-6 (Duflex,
SSWhite, Rio de Janeiro, RJ, Brasil) através de raspagem radicular, seguida de jateamento
com bicarbonato de sódio e água (Figura 1A). Após limpeza, os dentes foram armazenados
em água destilada em estufa à temperatura de 37ºC até o início da realização do preparo
cavitário. Considerando que os dentes seriam submetidos à ciclagem mecânica, foi simulado o
ligamento periodontal e o dente foi incluído em resina de poliestireno para reproduzir a
movimentação da raiz no alvéolo. Para isso, o dente foi demarcado com caneta para
retroprojetor a 2 mm da junção amelocementária e a porção radicular foi mergulhada em um
recipiente com cera nº 7 derretida (Figura 1B), permitindo então que a raiz ficasse coberta por
uma fina espessura de cera (aproximadamente 0,2mm) até a demarcação. A raiz coberta de
cera foi moldada pela técnica de dupla impressão, através de material de moldagem à base de
silicone de adição nas consistências densa e leve (Express XT, 3M ESPE, St. Paul, MN, USA)
(Figuras 1C-D). A cera coberta sobre a porção radicular foi removida (Figura 1E). Material de
moldagem à base de poliéter (Impregum Soft, 3M ESPE, St. Paul, MN, USA) foi manipulado
e inserido na moldagem de silicone, no espaço correspondente ao ligamento periodontal, e o
dente foi introduzido sob leve pressão digital (Figura 1F). Após a polimerização, os excessos
foram removidos com lâmina de bisturi, de forma que o “ligamento periodontal” cobrisse a
raiz até a marcação (Figura 1G). Após, o dente foi fixado com cera pegajosa, através da coroa,
a um fragmento de 4 cm de fio vareta de aço 0,40’’ (Morelli Ortodontia, Sorocaba, SP, Brasil)
(Figura 1H). O conjunto foi apoiado sobre um cilindro de PVC (Tigre tubos e conexões S/A,
Joinville, SC, Brasil) de 2,1 cm de diâmetro interno e, aproximadamente, 2 cm de altura, de
forma que as raízes ficassem posicionadas lateralmente no interior do tubo e a face oclusal do
dente ficasse paralela a base do tubo (Figura 1I). Resina de poliestireno autopolimerizável foi
manipulada na proporção de 2% (100g de resina: 2g de catalisador) e vertida no interior do
cilindro de PVC a fim de facilitar a padronização dos preparos cavitário (Figura 1J). Após sua
polimerização, o conjunto foi removido do tubo e o fio desacoplado do dente, e as faces
oclusais foram planificadas em uma máquina de polimento (Politriz, AROTEC, São Paulo,
SP, Brasil) com lixas de carbeto de silício de granulações 400-, 800- e 1200- (3M ESPE
43
411Q, Sumaré, SP, Brasil), sequencialmente (Figura 1K), a fim de se obter uma superfície
plana de esmalte sem exposição dentinária (Figura 1L). As cavidades Classe I foram
preparadas usando brocas do tipo Carbide burs #56L (KG Sorensen, Barueri, Brasil) em alta
rotação, sob irrigação constante, em um aparato personalizado que permitiu a padronização
das dimensões da cavidade em 5mm no sentido mesio-distal, 4mm no sentido vestíbulo-
lingual e 4mm de profundidade com o ângulo cavosuperficial em esmalte (Figuras 1M-N). A
broca foi substituída após a confecção de cinco cavidades.
Figura 1 – Preparo dos dentes
A
B C
D
E F
G
H I
H
J
H
K
H
L
H
M
H
N
H
A – Terceiros molares limpos. B – Raízes cobertas por fina espessura de cera. C/D – Moldagem
da raiz pela técnica da dupla impressão. E – Remoção da cera. F – Poliéter inserido no molde.
G – Raiz coberta por fina espessura de poliéter, simulando ligamento periodontal. H – Dente
acoplado a fio ortodôntico. I – Dente posicionado no interior do tubo de PVC. J – inserção de
resina de poliestireno. K – Planificação das cúspides em politriz. L – Superfície oclusal plana.
M – Preparo cavitário. N – Cavidade preparada.
A
H
D
H
G
H
J
H
M
H
44
Procedimento restaurador:
Previamente ao procedimento restaurador, as superfícies de esmalte e dentina
foram condicionadas com ácido fosfórico a 35% (Ultra-Echt, Ultradent Products Inc., South
Jordan, UT, USA) por 20 segundos (Figura 2A), lavadas também por 20 segundos (Figura
2B), de acordo com recomendações do fabricante, e secas com bolinha de algodão hidrófila
(Figura 2C). Após este passo, foram tratadas com uma camada do primer Optibond FL (Kerr,
Orange, CA, USA) com auxílio de um pincel descartável (Microbrush, KG), com leves
movimentos de fricção, por 15 segundos (Figura 2D), secas com leve jato de ar por 5
segundos. Em seguida, receberam uma camada do adesivo Optibond FL (Kerr, Orange, CA,
USA) com auxílio de um pincel descartável (Microbrush, KG) (Figura 2E) e foram
fotoativadas durante 10 segundos por aparelho fotoativador LED de 3ª geração (VALO,
Ultradent Products Inc., South Jordan, UT, USA) (Figura 2F). O procedimento restaurador
foi realizado através da inserção dos compósitos, em cor A2 - ou IVA para Tetric N-Ceram
Bulk Fill, em incremento único (no caso dos compósitos bulk-fill) ou através da técnica
incremental oblíqua, com incrementos de 2 mm (no caso do compósito Filtek Z350 XT) na
cavidade do elemento dental, utilizando-se espátula Suprafill #1 (SS White Duflex), com
exceção do compósito SonicFill, que vem acompanhado de aplicador próprio (Figura 2G). O
incremento de compósito foi então fotoativado pelo aparelho fotoativador LED de 3ª geração
(VALO) em modo Standard: 1000 mW/cm² durante 20 segundos (20 J/cm2) (Figura 2H). Os
grupos experimentais (n = 12) estão apresentados no quadro 1.
Quadro1: Grupos experimentais
Grupos Compósitos Tratamento
1 Filtek Z350 XT Ciclagens mecânica e térmica
2 Filtek Z350 XT -
3 Tetric N-Ceram Ciclagens mecânica e térmica
4 Tetric N-Ceram -
5 Filtek Bulk Fill Ciclagens mecânica e térmica
6 Filtek Bulk Fill -
7 SonicFill Ciclagens mecânica e térmica
8 SonicFill -
45
Figura 2 – Procedimento restaurador
Acabamento e polimento:
Após o término da restauração, foram realizados acabamento e polimento da
superfície oclusal com lixas de carbeto de silício números 600, 1200 e 2000 e polimento com
pasta diamantada de granulações 3, 0,5 e 0,25 μm (AROTEC Ind. Com- São Paulo, Brasil)
com discos de feltro em politriz e refrigeração à água. Entre cada granulação de lixa e de
pasta, as amostras foram levadas a cuba de ultrassom (Ultrasound Ultrason 1440 D-
Odontobrás Ind. E Com. Med. Odont. Ltda, Rio Preto, SP, Brasil) por 10 minutos para
remoção de detritos (Figura 4). A figura 3 apresenta imagens do procedimento de acabamento
e polimento.
Figura 3 – Acabamento e polimento das restaurações
A – Condicionamento ácido do esmalte e da dentina. B – Lavagem das superfícies. C –
Secagem com bolinha de algodão. D – Aplicação do primer. E – Aplicação do adesivo. F –
Fotoativação. G – Inserção do compósito. H – Fotoativação.
H E F
G
A B C D
46
Figura 4 – Cuba de ultrassom
Ciclagem mecânica e térmica:
Finalizada a restauração, acabamento e polimento, as amostras dos grupos 1, 3, 5
e 7 foram submetidas a 200.000 ciclos mecânicos com carga de 86 N à frequência de 2 Hz, na
máquina de ciclagem mecânica ER-37000 (ERIOS, São Paulo, SP, Brasil) (Figura 5A) com a
carga sendo aplicada no centro da restauração (Figura 5B) . Após, foram submetidas a 550
ciclos térmicos (temperatura variando entre 5 ± 1°C , 37 ± 1°C e 55 ± 1°C em câmaras de
água, durante 30 segundos em cada cuba, com um intervalo de 15 segundos) em máquina de
ciclagem térmica (MSCT-3e, ElQuip, São Carlos, SP, Brazil) para simular o envelhecimento
em condições aproximadas às da cavidade bucal (Figura 5C).
Figura 5 – Ciclagens mecânica e térmica
A – Máquina de Ciclagem mecânica. B – Carga aplicada no centro da restauração. C
- Máquina de ciclagem térmica.
C
A
B
47
Obtenção dos espécimes para o ensaio de microtração:
7 dentes de cada grupo foram utilizados para o ensaio de microtração e
nanoinfiltração. Para o ensaio de microtração, as coroas dentais foram separadas da porção
radicular na altura do limite amelocementário, através de seccionamento perpendicular do
elemento dentário em relação a seu longo eixo, utilizando disco diamantado dupla face (KG
Sorensen Ind. E Com. Ltda, Barueri, SP, Brasil), sob refrigeração constante (Figura 6A). Em
seguida, as coroas foram fixadas em placas acrílicas com cera pegajosa (New Wax, Thechnew
Com.e Ind.Ltda, Rio de Janeiro, RJ, Brasil) (Figura 6B). O conjunto foi fixado na cortadeira
metalográfica de precisão (Isomet 1000, BUEHLER Ltda. Lake Buff, IL, USA) (Figura 6C),
na qual um disco diamantado de alta concentração (Isomet Diamond Wafering Blades,
Buehler Ltd., Lake Buff, IL, USA) foi adaptado e, girando em baixa velocidade, sob irrigação
constante com água destilada, realizou cortes seriados no sentido mésio-distal e vestíbulo-
lingual, para se obter aproximadamente 10 palitos de 0,7 x 0,7 mm da interface pulpar
(Figuras 6D-E).
Figura 6 – Obtenção dos espécimes para o ensaio de resistência à microtração
A B
C
D
E
A – Dente demarcado para secção de coroa. B – Coroa fixada à placa acrílica. C - cortadeira
metalográfica de precisão. D – Secção seriada da coroa dental. E – Palitos de 0,7mm².
48
Ensaio mecânico de resistência à microtração:
Os palitos foram fixados ao dispositivo de microtração (Geraldeli) com adesivo à
base de cianoacrilato (Super Bonder Gel, Loctite, Henkel Ltda., Itapevi, SP, Brasil) pelas suas
extremidades (Figura 7A), de modo a posicioná-los paralelamente ao carregamento de tração.
Foram, então, levados à máquina de ensaio universal (EZ Test L Shimadzu, Japão) (Figura
7B) e o ensaio foi conduzido com célula de carga de 500 kgf, na velocidade de 0,5 mm/min,
até a ruptura (Figuras 7C-D). A força necessária para causar a ruptura dos corpos de prova,
em quilograma-força (kgf), foi anotada e as dimensões da interface adesiva dos espécimes
foram aferidas com paquímetro digital (Mitutoyo Corporation, Tokio, Japão). A resistência à
fratura em MegaPascal (MPa) foi calculada de acordo com a fórmula matemática: R= F(kgf)
x 9,8/A (R= resistência de união em MPa, F= força em quilograma-força (kgf) e A=área em
mm². Após a observação da normalidade dos dados e igualdade das variantes, os resultados
foram submetidos à análise estatística ANOVA dois fatores e Teste de Tukey (α=0.05) e o
padrão de fratura foi avaliado em Microscópio Eletrônico de Varredura (MEV) (LEO 435 VP,
LEO Electron Microscopy Ltd, Cambridge, UK) em: coesiva em dentina, coesiva em resina,
adesiva ou mista, considerando como fraturas coesiva e adesiva quando pelo menos 70% da
área total era composta pelo mesmo padrão. A fratura mista foi estabelecida quando existiam
mais de um padrão e nenhum prevalecia.
Figura 7 – Ensaio mecânico de resistência à microtração
A – Palito fixado ao dispositivo de microtração. B – Máquina de ensaio
universal. C – Ensaio de resistência à microtração. D – Ruptura do palito.
A
B
C
D
49
Análise de nanoinfiltração:
Para a análise da nanoinfiltração, dois palitos de cada dente, totalizando 14 palitos
por grupo, foram selecionados aleatoriamente e imersos por 24 horas em solução de nitrato de
prata amoniacal, contendo 10 gramas de cristais de nitrato, 10 mL de água deionizada e gotas
de hidróxido de amônio a 28% (Figura 8A). Após, os palitos foram imersos em solução
reveladora por 8 horas sob luminária fluorescente (Figura 8B); foram lavados em água
destilada (Figura 8C) e incluídos em resina de poliestireno (Figuras 8D-E). Após
embutimento, os palitos foram desgastados em politriz com lixas d’ água 600-, 1200- e 2000-,
respectivamente (Figuras 8F-H), e polidos com discos de feltro e pastas diamantadas em
granulação decrescente 3, 0,5 e 0,25μm (Figuras 8I-K). Entre cada granulação de lixa e de
pasta, as amostras foram levadas a cuba de ultrassom por 10 minutos para remoção de
detritos. As amostras foram secas com papel absorvente e, a seguir, receberam aplicação de
uma solução de ácido fosfórico a 85% por 30 segundos para desmineralização, seguida de
lavagem com água destilada. Para a desproteinização, foi utilizada solução de hipoclorito de
sódio a 10% por 10 minutos. Em seguida, foi realizada a lavagem com água destilada e a
secagem a temperatura ambiente. Posteriormente, as amostras foram desidratadas em álcool
etílico em concentrações crescentes (25%, 50%, 75%, 90% e 100%) por 10 minutos em cada
concentração (Figura 8L). As amostras foram cobertas com carbono (BalTec SCD 050-
SputterCoater) para serem observadas em MEV, operando em alto vácuo numa potência de 20
KV, no qual foram obtidas imagens em elétrons retroespalhados. As imagens foram gravadas
para se avaliar quantitativamente a área infiltrada com o auxílio do Software ImageJ (ImageJ,
National Institute of Health, Bethesda, USA). A barra de escala das imagens do MEV foi
utilizada para calibração da distância, os comprimentos dos tags infiltrados e segmentos foram
obtidos em milímetros e convertidos para porcentagem do comprimento total da interface.
Após a observação da normalidade dos dados e igualdade das variantes, os resultados foram
submetidos à análise estatística ANOVA dois fatores e Teste de Tukey (α=0.05).
50
Figura 8 – Preparo dos espécimes para análise de nanoinfiltração
A B C
D E F
G H I
J K L
A – Palitos imersos em solução de nitrato de prata amoniacal. B – Palitos imersos em solução reveladora. C
– Lavagem em água destilada. D – Embutimento dos palitos em resina de poliestireno. E – Palitos
incluídos em resina de poliestireno. F/G/H – Desgaste dos palitos em lixas d’água. I/J – Polimento dos
palitos em discos de feltro e pastas diamantadas. K – Palitos incluídos e polidos. L – desidratação das
amostras em álcool etílico.
51
Análise de adaptação marginal:
Para este teste, 5 dentes de cada grupo foram utilizados. Os dentes foram
removidos da resina de poliestireno com auxílio de um fórceps cirúrgico dental #17
(Millennium, GOLGRAN, São Caetano do Sul, SP, Brasil) e foram fixados a placa acrílica
com cola quente (Rhamos e Brito, Guarulhos, SP, Brazil) (Figura 9A). Após, as coroas
dentais foram seccionadas em até 1,5 mm abaixo da face oclusal utilizando-se disco
diamantado, sob refrigeração constante a água a 300 rpm, em uma cortadeira metalográfica de
precisão (Figura 9B) a fim de se obter superfícies oclusais planas, que foram desgastadas em
politriz com lixas d’ água 600-, 1200- e 2000-, respectivamente, e polidas com discos de
feltro e pastas diamantadas em granulação decrescente 3, 0,5 e 0,25μm (Figuras 9C-E). Entre
cada granulação de lixa e de pasta, as amostras foram levadas a cuba de ultrassom por 10
minutos para remoção de detritos. Em seguida, foram realizadas impressões da superfície
oclusal com material de baixa viscosidade de polivinilsiloxano (Express XT, 3M ESPE, St.
Paul, MN, EUA) (Figuras 9F-H) para se confeccionar réplicas dos dentes em resina epóxica
(EpoxicureResin, Buehler, Lake, IL, EUA) (Figura 9P). As amostras foram, então, fixadas
novamente a placas acrílicas (Figura 9I) e seccionadas longitudinalmente, na direção
vestíbulo-lingual, exatamente ao centro da restauração, por disco diamantado em cortadeira
metalográfica de precisão (Figuras 9J-K). Uma das metades obtidas foi selecionada
aleatoriamente e fixada à placa acrílica para facilitar o processo de polimento (Figura 9L), que
foi realizado pelo mesmo protocolo utilizado na superfície oclusal (Figuras 9M-N). As novas
superfícies obtidas foram moldadas (Figura 9O) para se confeccionar também réplicas em
resina epóxica (Figura 9P). Estas réplicas foram cobertas com uma delgada camada de ouro
(Balzers-SCD 050 Sputter Coater, Liechtenstein) e foram utilizadas para avaliação em MEV
da adaptação marginal nas paredes circundantes e de fundo. A extensão das fendas foi
quantitativamente analisada utilizando o Software ImageJ. A barra de escala das imagens do
MEV foi utilizada para calibração da distância, os comprimentos dos segmentos de fenda
foram obtidos em milímetros e convertidos para porcentagem do comprimento total da
interface. Após a observação da normalidade dos dados e igualdade das variantes, os
resultados foram submetidos à análise estatística ANOVA dois fatores e Teste de Tukey
(α=0.05).
52
Figura 9 – Obtenção dos espécimes para análise de adaptação marginal
A B
DC E
F G H
I J K
L M N
O P
A – Dente removido da resina de poliestireno e fixado à placa acrílica. B – Secção da coroa em cortadeira
metalográfica de precisão. C/D – Acabamento e polimento da amostra. E – Superfície oclusal plana e
polida. F/G/H – Moldagem da superfície oclusal. I – Dente fixado à placa acrílica. J – Secção longitudinal
em cortadeira metalográfica de precisão. K – Dente seccionado. L – Metade do dente fixada à placa
acrílica. M/N – Acabamento e polimento da amostra. O – Moldagem da superfície longitudinal. P – resina
epóxica vertida em molde para confecção de réplicas.
53
ANEXO 1 - Certificado do Comitê de Ética em Pesquisa
54
ANEXO 2 - Certificados de submissão do artigo ao periódico Journal of Dentistry