Using Solid Phase Extraction for Multi-Cartridge Extractions
and Extract Clean-up Procedures
Objective
• Utilize the FMS 625/8270 Turbo Trace SPE system’s ability to perform automated multi-cartridge extractions for complex procedures
• For single cartridge extractions, utilize the second cartridge loop to automate SPE formatted clean-up procedures in-line with SPE extractions.
Advantages
• Reduced evaporation steps • Reduced manual transfers of extract • Simplified turn key procedures • Greater reproducibility
FMS, Inc. 625/8270 Turbo Trace SPE
Touch Screen Control
SD Card for storing and Transferring Methods
Sample Size 2ml to Unlimited
Automatic Bottle Rinse
FMS, Inc. 625/8270 Turbo Trace SPE
Primary Cartridge
Secondary Cartridge
Automatic Liquid Level Sensor
Real Time Graphical Display of Extraction Steps
Positive Pressure Solvent/Sample Pump
FMS, Inc. SuperVap w/direct to GC vial
Extraction Configurations • PCBs
– DVB Cartridge – NaSO4/Acidified Silica Gel clean-
up • Organochlorine Pesticides
– DVB Cartridge – NaSO4/Florisil clean-up
• Other – Any organic inline with NaSO4
• EPA 8270D – Mixed CAX/DVB cartridge – Coconut Charcoal cartridge
• Organophosphorus/nitrogen pesticides – C18 Cartridge – Coconut Charcoal Cartridge
• PAHs – C18 Cartridge – NaSO4/Silica Gel clean-up
Flow Diagram
Procedures • Traditional Cartridge
Conditioning for Extraction – Wet with Organic Solvent (MeCl,
EtAC, Hexane, etc) – Dry – Wet with MeOH – Wet with H2O
§ Load Sample § Dry Cartridge § Elute Sample
• Pre-wet clean-up Cartridge – Wet with Organic Solvent
(Hexane, Hexane/Ether, DCM/Hexane, etc)
EPA 8270D • Mixed bed Cartridge re-conditioned with
MeOH. • Both cartridges conditioned with a 1%
NH3OH solution • Cartridges independently dried with N2
for 1 minute each • Mixed bed Cartridge eluted with
additional 30 mls DCM and collected as Fraction #2
• Coconut charcoal eluted with 30 mls DCM and added to Fraction #2
• Cartridges independently purged of residual solvent via N2 stream.
• Cartridges wet with DCM & N2 Dried • Cartridges conditioned with MeOH • Cartridges conditioned with H2O • Samples (ph=2) passed across both
cartridges at ~15mmHG • Mixed bed cartridge partially dried with
N2 at 10 PSI • Sample bottles sprayed with 20mls
DCM & loaded across mixed bed cartridge and collected as Fraction #1
• Cartridge eluted with additional 10 mls DCM
• Cartridges purged with N2 passing all DCM to collection vials.
Phosphorus/Nitrogen Pesticides
• Cartridge eluted with additional 10 mls DCM
• Cartridges purged with N2 passing all DCM to collection vials.
• Coconut charcoal eluted with 20 mls DCM and added to Fraction
• Cartridges independently purged of residual solvent via N2 stream.
• Cartridges wet with DCM & N2 Dried • Cartridges conditioned with MeOH • Cartridges conditioned with H2O • Samples (ph=2) passed across both
cartridges at ~15mmHG • C18 cartridge partially dried with N2 at
10 PSI • Sample bottles sprayed with 20mls
DCM & loaded across C18 cartridge and collected
PAHs • Cartridge eluted with additional 5 mls
DCM • Clean-up cartridge eluted with 15mls
DCM • Fraction line purged with N2 passing all
DCM to collection vials.
• Wet Clean-up cartridge with DCM • Wet C18 cartridge with DCM and dry • C18 Cartridge conditioned with MeOH • C18 Cartridge conditioned with H2O • Samples (ph=2) passed across both
cartridges at ~15mmHG • C18 cartridge partially dried with N2 at
10 PSI • Sample bottles sprayed with 20mls
DCM & loaded across C18 cartridge in-line with clean-up cartridge.
TeCDD, PCBs & PBDEs • Cartridge eluted with additional 15mls
Hexane • Clean-up cartridge eluted with 20mls
Hexane • Fraction line purged with N2 passing all
Hexane to collection vials.
• Wet Clean-up cartridge with Hexane • Wet DVB cartridge with Hexane and dry • DVB Cartridge conditioned with MeOH • DVB Cartridge conditioned with H2O • Samples (ph=2) passed across both
cartridges at ~15mmHG • DVB cartridge dried with N2 at 10 PSI • Sample bottles sprayed with 20mls
Hexane & loaded across DVB cartridge in-line with clean-up cartridge.
Organochlorine Pesticides • Cartridge eluted with additional 15mls
Hexane • Clean-up cartridge eluted with 20mls
Ether/Hexane • Fraction line purged with N2 passing all
solvent to collection vials.
• Wet Clean-up cartridge with Hexane • Wet DVB cartridge with Hexane and dry • DVB Cartridge conditioned with MeOH • DVB Cartridge conditioned with H2O • Samples (ph=2) passed across both
cartridges at ~15mmHG • DVB cartridge partially dried with N2 at
10 PSI • Sample bottles sprayed with 20mls
Hexane & loaded across DVB cartridge in-line with clean-up cartridge.
Analytical Platforms
• Agilent 7890 w/µECD • Thermo Trace GC, DSQ • Thermo Trace GC w/PTV, Quantum TSQ
AB 8070 Results
OPP Results
PAH-Low Results
Analytes spiked at .05 µg/ml
PCDD/DFs Results
PBDE Results
OCP Results
Conclusions
• Utilization of automation for multi-cartridge procedures streamlines multi-step extraction processes into a single process.
• Eliminates extra evaporation steps, thus additional solvent consumption and reclamation.
• Allows laboratory staff to perform other duties rather than attend manual SPE manifolds.
Acknowledgements
Questions?