+ All Categories
Home > Documents > Validation of a simple method for measuring cranial ... · Validation of a simple method for...

Validation of a simple method for measuring cranial ... · Validation of a simple method for...

Date post: 11-Mar-2020
Category:
Upload: others
View: 15 times
Download: 0 times
Share this document with a friend
14
Validation of a simple method for measuring cranial deformities (plagiocephalometry) Léon N.A. van Adrichem 1 Leo A. van Vlimmeren 2 Dominika Cadanová 1 Paul J.M. Helders 3 Raoul H.H. Engelbert 3 Han J.W. van Neck 1 Anton H.J. Koning 4 1 Department of Plastic and Reconstructive Surgery, Erasmus University Medical Center Rotterdam, Sophia Children's Hospital, Rotterdam 2 Department of Physical Therapy, Bernhoven Hospital, Veghel 3 Department of Pediatric Physical Therapy and Exercise Physiology, University Medical Center; Wilhelmina Children's Hospital, Utrecht 4 Erasmus Center for Bioinformatics, Erasmus University Medical Center, Rotterdam The Netherlands Submitted for publication Chapter 5 Chapter 5 Chapter 5 Chapter 5
Transcript

Validation of a simple method for measuring cranial deformities (plagiocephalometry)

Léon N.A. van Adrichem1

Leo A. van Vlimmeren2

Dominika Cadanová1

Paul J.M. Helders3

Raoul H.H. Engelbert3

Han J.W. van Neck1

Anton H.J. Koning4

1Department of Plastic and Reconstructive Surgery,

Erasmus University Medical Center Rotterdam,

Sophia Children's Hospital, Rotterdam 2Department of Physical Therapy,

Bernhoven Hospital, Veghel 3Department of Pediatric Physical Therapy and

Exercise Physiology, University Medical Center;

Wilhelmina Children's Hospital, Utrecht 4Erasmus Center for Bioinformatics,

Erasmus University Medical Center, Rotterdam

The Netherlands

Submitted for publication

Chapter 5 Chapter 5 Chapter 5 Chapter 5

Abstract

Context

Craniofacial measuring is essential for diagnosis or evaluation of growth and therapies. Skull

deformities in children are mainly caused by craniosynostosis or by external pressure in

positional skull deformations. Traditional anthropometry does not sufficiently analyse

craniofacial shape. In CT scanning radiation loads are considerable and both CT and MRI

scanning, due to their long acquisition time, require anaesthesia in children if an accurate

picture is needed. This makes CT and MRI unsuitable for long term follow up of paediatric

patients, unless there is a compelling reason to do so. Other non-invasive 3D surface scanners

still have limited practical use.

Van Vlimmeren et al. presented plagiocephalometry as a simple and versatile instrument to

quantify skull deformities and reliability was proven by high intrarater and interrater reliability,

but concurrent validity was not investigated.

Objective

To explore concurrent validity of plagiocephalometry and 3D-CT scanning, being the golden

standard of 3D monitoring, including correlations and clinical agreement between the scores of

both measurements.

Methods

At the Erasmus University Medical Center Rotterdam, Sophia Children’s Hospital

plagiocephalometry was compared to 3D-CT scanning in 21 children with craniosynostosis early

in life, in order to investigate concurrent validity of plagiocephalometry.

Results

The plagiocephalometry ring proved to fit closely to the skin with mean differences less than 1

mm (p<0.05). The shape of the plagiocephalometry ring was not significantly changed when

taken off the head (p>0.05). Finally no significant differences are shown between

measurements on the skull (CT-scan) and plagiocephalometry ring off the head (p>0.05).

Conclusions

The study supports concurrent validity of the measurements for all of the explorations:

plagiocephalometry fitting to the skin, retaining plagiocephalometry shape off the head and

correspondance of the actual asymmetry of the skull as acquired by plagiocephalometry and CT

scanning. Plagiocephalic measurements are in agreement with the measurements from 3D-CT

scanning, the present golden standard.

Although only 2 dimensional measurements are done by plagiocephalometry, the combination

of simplicity, reliability and validity make it a promising tool for daily practice.

68

Introduction Measurement of craniofacial structures is essential for the diagnosis or evaluation of growth

and subsequent intervention. Skull deformities in children are mainly caused by craniosynostosis

or by external pressure in positional skull deformations.

In traditional anthropometry, distances and angles are measured, but shape is not recorded.

Furthermore, on the skull, especially in the deformed skull, no clear landmarks are present.

Young children are lively, which complicates anthropometry. Plain X-rays of the skull do

visualize cranial sutures, but shape is not sufficiently recorded.1

Ideally, sequential, complete 3D craniofacial pictures should be obtained. In CT scanning,

radiation loads are considerable and both CT and MRI scanning, due to their long acquisition

time, require anaesthesia in children if an accurate picture is needed.2-4 This makes CT and MRI

unsuitable for long term follow up of paediatric patients, unless there is a compelling reason to

do so.

Non-invasive 3D surface scanning might be a good solution, but although numerous articles

using various forms of 3D surface data acquisition have been described since the late 70's,

these methods are of limited practical use. The acquisition time of these devices is in the range

of several seconds, not sufficiently short to capture a highly mobile child. Three-D

photogrammetry, however, is accurate, fast, non contact and non-invasive and is a promising

system to obtain sequential complete 3D surface data.5 Disadvantages are the price and the size

of this system, so acquisition can only be done in specialized centres.

Van Vlimmeren et al.6 recently presented a simple and versatile instrument to quantify skull

deformities: Plagiocephalometry (PCM). PCM is performed with a strip of thermoplastic

material, which is positioned around the infant’s head at the widest transverse circumference.

Landmarks of the ear and the nose bridge are traced. PCM turned out to be easy applicable,

non-invasive at low costs, while intrarater and interrater reliability were high as illustrated by

intraclass correlations and limits of agreement.6 Although the method is 2D, sufficient

information was obtained by PCM to quantify the skull deformities.

Present study was performed to explore concurrent validity of PCM and 3D-CT scanning, being

the golden standard of 3D monitoring, including correlations and clinical agreement between

the scores of both measurements.

The aim of this study was:

1. To investigate how closely the thermoplastic strip of PCM fits to the skin.

2. To investigate whether the thermoplastic strip of PCM retains its shape after taking it off

the infants head.

3. To determine to what extends the asymmetry, as acquired from the thermoplastic PCM ring,

corresponds with the actual asymmetry of the skull as acquired from 3D-CT scanning.

Chapter 5

69

Patients and methods

Patients

All measurements were performed at the Erasmus University Medical Center Rotterdam, Sophia

Children’s Hospital, from September 2004, till March 2005. All of 21 included patients were

scheduled for 3D-CT scanning of the head due to a serious suspicion of existing

craniosynostosis (age in months: mean 12.6 +/- 12.7; median 10,5; minimum 3.8; maximum

62,7). The parents agreed that PCM with a thermoplastic strip was performed simultaneously.

No children with positional skull deformations were included, because in these patients CT

scanning is redundant and not ethical due to anaesthesia.

Methods

Plagiocephalometry is performed with a strip of thermoplastic material (3.2 mm thick) of 18 mm

x 50 cm (Thermo extra-comfort, non-perfo by GeniMedical, the Netherlands), which is

positioned around the infant’s head at the widest transverse circumference. In less than two

minutes, the ring is fixed and the three landmarks (both ears and nose) are marked

perpendicular on the ring in a standardized manner. Both landmarks at the posterior edge of

the tragus, correspond best with the meatus acousticus externus. The third landmark is traced

off at the middle of the nose bridge. In this way it is possible to trace the exact positions of the

ears and the nose in relation to the transverse circumference and contours of the head.

Afterwards, the ring is removed from the head and a fourth landmark is marked representing

the middle of the posterior circumferential distance between the left and right ear, measured

with a measuring tape. Using a standard copying machine, the upper side of the ring is copied

on paper. Nine lines are drawn on the paper copy and measured to the nearest millimetre, by

which the degree of asymmetry can simply be determined by calculating the differences

between the lengths of the left and right lines. The clinically most important measures are

arranged in three parts:6

Part 1. Position of the ears, nose and local flattening of the skull.

Ear deviation (ED), anterior-sinistra – anterior-dextra (ASAD) and posterior-dextra – posterior-

sinistra (PSPD).

Part 2. Diameter difference: Oblique diameter difference (ODD). The oblique diameter left (ODL)

and oblique diameter right (ODR) lines are drawn from points located 40o either side of the

antero-posterior (AP) line. The ODD is calculated as ODL– ODR. The angle of 40o is chosen

because this has been used by other authors formerly, probably because the differences

between the diameters of the typical shape of the skull at these angles are the most

outstanding.7 The ratio between the ODL and the ODR is calculated as the longest/shortest

diameter x 100% and is called oblique diameter difference index (ODDI).

Part 3. Transversal shape and proportion of the skull.

The ratio between the sinistra-dextra (SD) and the anterior-posterior (AP) is calculated as SD/AP

x 100%, and is called the cranio proportional index (CPI) (Fig. 1-2).

70

Fig. 1.Fig. 1.Fig. 1.Fig. 1. Illustration plagiocephalometry: symmetric skull (boy 3 months old): a) photograph of the child with the PCM ring fitted and landmarks. b) paper copy of the same PCM ring with drawn and measured lines.

Fig. 2. Fig. 2. Fig. 2. Fig. 2. Illustration plagiocephalometry: asymmetry DP left occipital of the skull (boy 4 months old): a) Photograph of the child with the thermoplastic ring and landmarks. The digitally drawn lines are

made to illustrate the agreement with the paper copy and to explain the names of the lines. b) Paper copy of the same ring with drawn and measured lines. Legends: : : : AP: anterior-posterior; SD: sinistra-dextra; AS: anterior-sinistra; AD: anterior-dextra; PS: posterior-sinistra; PD: posterior-dextra; ED: ear deviation; ODL: oblique diameter left; ODR: oblique diameter right; ASAD = AS minus AD; PDPS = PD minus PS; ODD = ODL minus ODR. ODDI: oblique diameter difference index (ratio between ODL and ODR calculated as longest oblique diameter / shortest oblique diameter x 100%); CPI: cranio proportional index (ratio between SD and AP calculated as SD/AP x 100%).

Chapter 5

71

Medical Student (DC) was trained to measure PCM using a standardized protocol. Prior to CT

scanning the PCM ring was obtained. During CT scanning the PCM ring was in place. After CT

scanning the PCM ring was copied on paper, and finally all lines were drawn and the distances

were measured in millimetres.

Three-D-CT scanning was performed at the radiology department of the Erasmus University

Medical Center Rotterdam, Sophia Children’s Hospital. All of the children were under general

anaesthesia. The CT scanner equipment was Siemens Emotion 6; a slice thickness of 2.5 mm

with an interslice distance of 1.2 mm was used (Fig. 3). The 3D-CT data sets were imported into

“I-Space” at the Department of Bioinformatics of the Erasmus University Medical Center

Rotterdam. The “I-Space” is a CAVETM-like virtual reality system with 3D-processing tools.8

I-Space, equipped with 3D volume rendering software, uses eight projectors on three walls and

the floor, to create a true 3D image in a special viewing arena.9 Users can then interact with the

image using a 6 degrees-of-freedom pointing device, and investigate the 3D data wearing a

pair of glasses with polarizing lenses. The person interacting with the data is presented with the

correct perspective by means of wireless head tracking.

In I-Space the plane of the PCM ring was determined visually by rotating and clipping the

volume. In this plane the same ear and nose markers of the PCM method were placed on the

bone and visually checked for accuracy by looking at the volume from different angles. After

this, a screen dump was made and printed on A4 size paper, closely cropped on the skull. All

the measurements described in the PCM part were performed on the bone.

To investigate how close the PCM ring fits to the skin and the underlying bone, the distance

between the PCM ring and the skin, and the distance between the PCM ring and the bone were

measured at positions of 0o, 45o, 90o, 135o, 180o, 225o, 270 o, 315 o (Fig. 4). A difference of 1

mm between the skin and the PCM ring was defined as an acceptable difference due to the

interposing of hair, based on clinical findings in children of 1 year of age. At the 8 angles the

distance was measured with a ruler between the centre (crossing of the anterior-posterior and

sinistra-dextra lines) and the PCM ring on the head, the PCM ring off the head, the bone and

the skin. In a pilot study the method appeared to be reproducible.

Statistical analysis

All the obtained data were registered and statistically analysed by making use of the Statistical

Program for the Social Sciences version 12.0.1 (SPSS). The T-test has been used for determining

the significance of the obtained data.

72

FFFFig. 3.ig. 3.ig. 3.ig. 3. Example of a 3D-CT image with PCM ring on the head of an infant with trigonocephaly

Fig. 4.Fig. 4.Fig. 4.Fig. 4. Example of the technique used for establishing the distance between:

• the strip and the centre of the head at the various positions.

• the strip and the skin at the various positions.

Chapter 5

73

Results

The distance between PCM ring and skin

The mean distance between PCM ring and skin is summarized in Table 1. The zero-hypothesis is

that the difference is equal to 1 mm. All mean values are less than 1 mm. The results of the

one-sample t-test show that in all but the 900 positions, the values are statistically significant

(p<0.05) different from 1 mm. Taken into account the mean values, this indicates that the

distances are significantly smaller than 1 mm. At the 900 position, the value is not statistically

significant (p>0.05) which means that the distance between the PCM ring and the skin matches

our assumed value of 1 mm. In all positions the distance between the PCM ring and skin fulfils

the criterion of 1 mm or less.

Table 1. Table 1. Table 1. Table 1. Distance in millimetres between PCM ring and skin.

Test Value = 1Test Value = 1Test Value = 1Test Value = 1

AnglesAnglesAnglesAngles

NNNN

MeanMeanMeanMean

SDSDSDSD

Sig. (2Sig. (2Sig. (2Sig. (2----tailed)tailed)tailed)tailed)

CI 95%CI 95%CI 95%CI 95%

LowerLowerLowerLower UpperUpperUpperUpper

0° 21 0.48 0.60 0.00 0.21 0.76

45° 21 0.58 0.48 0.00 0.36 0.80

90° 21 0.82 0.61 0.20 0.54 1.10

135° 21 0.43 0.38 0.00 0.26 0.60

180° 21 0.53 0.44 0.00 0.34 0.73

225° 21 0.42 0.43 0.00 0.22 0.61

270° 21 0.73 0.49 0.02 0.51 0.96

315° 21 0.64 0.49 0.00 0.42 0.87

The distance between PCM ring and bone

The mean distance between PCM ring and skull is summarized in Table 2. The zero-hypothesis

was that the difference is equal to 4 mm, based on the clinical soft tissue-thickness of about 3

mm (measured during surgery in children of 1 year of age) combined with the expected distance

of maximum 1 mm between the PCM ring and the skin. All mean values are between 3.2 to 4.5

mm. The results of the One-Sample T-Test show that in all but the 450 positions the values are

not statistically significant (p>0.05) different from 4 mm. No statistical significance (p>0.05)

indicates that the distance between the PCM ring and the skull matches our assumed value of 4

mm. Taken into account the mean value at 450, this indicates that the distance is significantly

smaller than 4 mm (p<0.05). In all positions the distance between the PCM ring and skin fulfils

the criterion of 4 mm or less.

74

Table 2. Table 2. Table 2. Table 2. Distance in millimetres between PCM ring and skull.

Test Value = 4Test Value = 4Test Value = 4Test Value = 4

AnglesAnglesAnglesAngles

NNNN

MeanMeanMeanMean

SDSDSDSD

Sig. (2Sig. (2Sig. (2Sig. (2----tailed)tailed)tailed)tailed)

CI 95% CI 95% CI 95% CI 95%

LowerLowerLowerLower UpperUpperUpperUpper

0° 21 4.51 1.37 0.11 3.88 5.13

45° 21 3.21 0.92 0.00 2.79 3.63

90° 21 3.69 0.93 0.14 3.26 4.11

135° 21 3.84 0.75 0.34 3.50 4.18

180° 21 3.85 0.97 0.49 3.41 4.29

225° 21 3.81 0.70 0.23 3.49 4.13

270° 21 3.86 0.63 0.32 3.57 4.14

315° 21 3.61 0.87 0.05 3.22 4.01

PCM ring taken off the head compared to ring on head

Comparison of the distance between the centre (crossing of the anterior-posterior and sinistra-

dextra lines) and the PCM ring at the 8 angles, when the PCM ring was on the head and off the

head showed a statistically significant correlation for every angle (Table 3).

Table 3.Table 3.Table 3.Table 3. Comparison in millimetres of the PCM ring on the head and off the head.

AnglesAnglesAnglesAngles

MeanMeanMeanMean

NNNN

SDSDSDSD

ICCICCICCICC

Sig.Sig.Sig.Sig.

Pair 1 Ring on the head at 0° 76.72 21 8.92

Ring off of the head at 0° 76.70 21 8.80

1.00 0.00

Pair 2 Ring on the head at 45° 67.67 21 7.71

Ring off of the head at 45° 67.29 21 8.08 0.98 0.00

Pair 3 Ring on the head at 90° 62.92 21 5.01

Ring off of the head at 90° 62.95 21 5.05 0.95 0.00

Pair 4 Ring on the head at 135° 70.17 21 6.81

Ring off of the head at 135° 70.40 21 6.78 0.98 0.00

Pair 5 Ring on the head at 180° 76.72 21 8.92

Ring off of the head at 180° 76.68 21 8.81 1.00 0.00

Pair 6 Ring on the head at 225° 69.60 21 5.47

Ring off of the head at 225° 69.76 21 5.70 0.96 0.00

Pair 7 Ring on the head at 270° 63.02 21 4.48

Ring off of the head at 270° 63.50 21 4.28 0.94 0.00

Pair 8 Ring on the head at 315° 67.66 21 5.83

Ring off of the head at 315° 67.45 21 6.63 0.97 0.00

Chapter 5

75

The mean differences between PCM ring on the head and off the head are summarized in Table

4. The results of the paired-samples t-test show that in all positions the values are not

statistically significant (p>0.05). The 95% confidence intervals of difference are within 1.2 mm.

So only a negligible shape difference occurs when the PCM ring is taken off the head

Table 4.Table 4.Table 4.Table 4. Difference in millimetres between PCM ring on and off the head.

Paired DifferencesPaired DifferencesPaired DifferencesPaired Differences

AnglesAnglesAnglesAngles

MeanMeanMeanMean

SDSDSDSD

CI 95% of the CI 95% of the CI 95% of the CI 95% of the

DifferenceDifferenceDifferenceDifference

Sig. (2Sig. (2Sig. (2Sig. (2----tailed)tailed)tailed)tailed)

LowerLowerLowerLower UpperUpperUpperUpper

Pair 1 on head 0° - off head 0° 0.02 0.51 -0.21 0.25 0.85

Pair 2 on head 45° - off head 45° 0.39 1.59 -0.33 1.11 0.28

Pair 3 on head 90° - off head 90° -0.03 1.54 -0.73 0.67 0.93

Pair 4 on head 135° - off head 135° -0.24 1.52 -0.93 0.46 0.48

Pair 5 on head 180° - off head 180° 0.05 0.47 -0.17 0.26 0.66

Pair 6 on head 225° - off head 225° -0.16 1.53 -0.86 0.53 0.63

Pair 7 on head 270° - off head 270° -0.48 1.56 -1.19 0.24 0.18

Pair 8 on head 315° - off head 315° 0.21 1.83 -0.63 1.04 0.61

Measurements on the skull (CT scan) compared to the PCM ring off the head

Table 5 shows a statistically significant correlation between measurements on the skull (CT

scan) and PCM ring off the head.

Table 5.Table 5.Table 5.Table 5. Comparison of measurements on the skull and on the PCM ring off the head

(pairs 1-4 in millimetres; pairs 5-6 in percentages).

MeasurementsMeasurementsMeasurementsMeasurements MeanMeanMeanMean NNNN SDSDSDSD ICCICCICCICC Sig. (2Sig. (2Sig. (2Sig. (2----tailed)tailed)tailed)tailed)

Pair 1 ASAD strip 0.33 21 6.45

ASAD skull 0.46 21 6.55

0.95

0.00

Pair 2 PDPS strip 0.02 21 3.62

PDPS skull 0.08 21 3.38

0.95

0.00

Pair 3 ED strip 5.19 21 3.99

ED skull 5.06 21 3.71

0.96

0.00

Pair 4 ODD strip 0.81 21 5.93

ODD skull 0.53 21 5.69

0.95

0.00

Pair 5 ODDI strip 103.03 21 2.76

ODDI skull 103.38 21 2.86

0.86

0.00

Pair 6 CPI strip 83.93 21 11.24

CPI skull 82.91 21 12.52

0.98

0.00

76

The mean differences between measurements on the skull and on the PCM ring off the head are

summarized in Table 6. The results of the Paired-Samples T-Test show that in all measurements

the values are not statistically significant (p>0.05). The 95% confidence intervals of difference

are within 2.3 mm.

Table 6.Table 6.Table 6.Table 6. Difference between measurements in millimetres on the skull and on the PCM ring off the head

(pairs 1-4 in millimetres; pairs 5-6 in percentages).

Paired DifferencesPaired DifferencesPaired DifferencesPaired Differences

MeasurementsMeasurementsMeasurementsMeasurements

MeanMeanMeanMean

SDSDSDSD

CI 95% of the CI 95% of the CI 95% of the CI 95% of the

DifferenceDifferenceDifferenceDifference

Sig. (2Sig. (2Sig. (2Sig. (2----tailed)tailed)tailed)tailed)

LowerLowerLowerLower UpperUpperUpperUpper

Pair 1 ASAD strip – ASAD skull -0.12 2.00 -1.03 0.79 0.78

Pair 2 PDPS strip – PDPS skull -0.06 1.18 -0.59 0.48 0.83

Pair 3 ED strip – ED skull 0.13 1.06 -0.35 0.62 0.58

Pair 4 ODD strip – ODD skull 0.28 1.91 -0.59 1.15 0.51

Pair 5 ODDI strip – ODDI skull -0.35 1.47 -1.02 0.32 0.29

Pair 6 CPI strip – CPI skull 1.02 2.85 -0.28 2.32 0.12

Table 7 shows the corresponding limits of agreement for this study and for the intrarater

comparison of two PCM rings in the PCM reliability study.6 The limits of agreement between

skull and PCM off the head are comparable to the intrarater PCM study of van Vlimmeren et al.

It can be concluded that the measurements on the PCM ring off the head correspond to the

skull measurements on the CT scan.

Table 7. Table 7. Table 7. Table 7. Limits of agreement in this study and in PCM intrarater reliability study

Skull Skull Skull Skull –––– PCM off the head PCM off the head PCM off the head PCM off the head

(this study)(this study)(this study)(this study)

PCM PCM PCM PCM –––– PCM off the head PCM off the head PCM off the head PCM off the head

(intrarater PCM(intrarater PCM(intrarater PCM(intrarater PCM6666))))

ASAD

-4.12 till +3.87

-5.86 till +5.86

PDPS -2.42 till +2.30 -2.88 till +3.08

ED -1.99 till +2.26 -4.32 till +4.24

ODD -3.55 till +4.11 -3.37 till +3.41

ODDI -3.30 till +2.60 -2.69 till +2.81

CPI -4.69 till +6.72 -5.32 till +3.68

Chapter 5

77

Discussion

Plagiocephalometry (PCM) is an easy-to-apply, non-invasive measurement to assess skull shape

with good clinical use and low application costs. Van Vlimmeren et al.6 showed a statistically

acceptable intrarater and interrater reliability for the application in clinical care and research.

Our study proves that the measurements on the PCM ring yield a valid representation of the

skull shape.

The tape fits closely to the skin with mean differences less than 1 mm. Furthermore only a

negligible shape difference occurs when the PCM ring is taken off the head. The distances

between the PCM ring and the skull can be explained by the interposing soft tissues. No

statistically significant difference is shown between measurements on the skull (CT scan) and

the PCM ring off the head. Comparison of the measurements on the skull and on the PCM ring

off the head show corresponding limits of agreement with the intrarater comparison of two

PCM rings in the PCM reliability study.6 The fact that the CPI value shows slightly higher limits

of agreement might be caused by the supine position of the infant during CT scanning with

pressure at the back of the head. It can be concluded that plagiocephalometry is a validated

measuring method for skull deformities.

We have to realize that PCM is in fact a 2D measuring method. It only provides information on

the transverse plane. No information is gathered on the height of the skull. Most skull

deformities, however, can be recognized and analysed in this transverse plane. True 3D

measuring methods will provide more information, because more data are gathered. As

explained earlier no 3D system is available that is non-invasive, easy-to-apply and with low

application costs. The available 3D surface monitoring systems are promising, but still reserved

to specialised centres.5 This study proves that PCM actually reproduces skull shape; the

simplicity of the method makes it a promising tool for analysis of plagiocephaly in daily practice.

Conclusions

Present study was performed to explore concurrent validity of PCM and 3D-CT scanning, being

the golden standard of 3D monitoring, including correlations and clinical agreement between

the scores of both measurements. The study supports concurrent validity of the measurements

for all of the explorations: PCM fitting to the skin, retaining PCM shape off the head and

correspondence of the actual asymmetry of the skull as acquired by PCM and CT scanning.

Plagiocephalic measurements are in agreement with the measurements from 3D-CT scanning,

the present golden standard. Although only 2 dimensional measurements are done by

plagiocephalometry, the combination of simplicity, reliability and validity make it a promising

tool for daily practice.

78

References 1. Pollack IF, Losken HW, Fasick P. Diagnosis and management of posterior plagiocephaly. Pediatrics 1997;99:180-

185. 2. Binaghi S, Gudinchet F, Rilliet B. Three-dimensional spiral CT of craniofacial malformations in children. Pediatr

Radiol 2000;30:856-860. 3. Benson ML, Oliverio PJ, Yue NC, Zinreich SJ. Primary craniosynostosis: imaging features. Am J Roentgenol

1996;166:697-703. 4. Darling CF, Byrd SE, Allen ED, Radkowski MA, Wilczynski MA. Three-dimensional computed tomography

imaging in the evaluation of craniofacial abnormalities. J Natl Med Assoc 1994;86:676-680. 5. Riphage JM, van Neck JW, van Adrichem LNA. 30 Years of 3D surface imaging in medicine: A review of working

principles and implications for imaging the unsedated child. Submitted 2006. 6. van Vlimmeren LA, Takken T, van Adrichem LN, van der Graaf Y, Helders PJ, Engelbert RH. Plagiocephalometry:

a non-invasive method to quantify asymmetry of the skull; a reliability study. Eur J Pediatr 2006;165:149-157. 7. Hutchison BL, Hutchison LA, Thompson JM, Mitchell EA. Plagiocephaly and brachycephaly in the first two years

of life: a prospective cohort study. Pediatrics 2004;114:970-980. 8. Cruz-Neira C, Sandin DJ, DeFanti T. Surround-screen projection-based virtual reality: the design and

implementation of the CAVE. In SIGGRAPH ’93: Proceedings Association for Computing Machinery, August 1993.

9. Koning AHJ. Applications of volume rendering in the CAVE. In B. Enquist et al (eds.), Simulation and visualization on the grid: Paralleldatorcentrum, seventh annual conference, Stockholm, 2000, pp. 112-121.

Chapter 5

79


Recommended