+ All Categories
Home > Documents > VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for...

VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for...

Date post: 20-Jul-2020
Category:
Upload: others
View: 1 times
Download: 0 times
Share this document with a friend
21
VBM-200 MINI LAMINAR FLOW METAL – SERIES ONE Installation and Operations Manual
Transcript
Page 1: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW

METAL – SERIES ONE

Installation and Operations Manual

Page 2: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 2 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

Page 3: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 3 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

MODELS COVERED

VBM-200 (2’) Series One (as shown above)

TABLE OF CONTENTS

MODELS COVERED ...................................................................................................................................................... 3

1. APPLICATION RANGE ....................................................................................................................................... 4

2. SHIPPING, UNPACKING, INSTALLATION, TECHNICAL PARAMETERS ............................................................ 5

3. PACKING LIST .................................................................................................................................................. 6

4. PERFORMANCE INDEX .................................................................................................................................... 7

5. STRUCTURE ..................................................................................................................................................... 8

6. AIR FLOW DIAGRAM ...................................................................................................................................... 11

7. OPERATION .................................................................................................................................................... 11

8. MAINTENANCE ............................................................................................................................................... 12

9. SERVICE WORK .............................................................................................................................................. 14

10. SPARE PARTS LIST ...................................................................................................................................... 18

11. NOTES .......................................................................................................................................................... 18

12. ELECTRICAL WIRING DIAGRAM .................................................................................................................. 19

13. SERVICE/MAINTENANCE NOTES ................................................................................................................ 20

14. WARRANTY .................................................................................................................................................. 21

Page 4: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 4 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

Thank you for choosing a Design Filtration Microzone VBM-200 Mini Laminar Flow. With proper care and

maintenance, your cabinet will provide many years of service.

This unit is designed to provide a particle-free work environment by filtering the air through a high efficiency

particulate air filter; a HEPA filter. The cabinet is designed to operate under constant positive pressure in order

to maintain the cleanest work environment possible within the work space itself. It is also suitable for PCR

work with the interlocking mode and UV timer.

Please be sure to read this manual before operating your cabinet, as it contains important information to assist

you in using your cabinet safely and to get the most out of its unique features.

Important information should be filled in below and this manual kept on hand at all times for quick and easy

reference, as well as a field reference for certifiers or service personnel.

1. APPLICATION RANGE

The VBM-200 Mini Laminar Flow creates a local high cleanliness air environment (ISO-5) by vertical air flow. It

is widely used for the nutritional industry, microbiological research, medicine and health, science research, the

semiconductor industry, precision instruments, meters, electronic components, optical instruments, the

chemical industry, smelting purification, magnetic materials, and many other departments. It is very helpful to

increase yield, precision, stability and reliability of products. It is also suitable for PCR work.

Mini Laminar Flow Information:

Model Number: ____________________________

Serial Number: ______________________________

Date of Manufacture: _________________________

Date of initial installation: _______________________

For replacement parts, warranty information or any other

inquiries, please contact Design Filtration Microzone at:

Toll Free: 1-888-636-8609 or www.dfmzgroup.com

Page 5: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 5 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

2. SHIPPING, UNPACKING, INSTALLATION, TECHNICAL PARAMETERS

2.1 STORAGE CONDITIONS

Temperature of the room should be <40ºC

Relative humidity of the room should be ≤75%

Good ventilation inside the warehouse

No acidic, alkaline gases or any other corrosive gases are allowed in storage warehouse

Maximum Storage Period or stand-by time: Maximum storage period is one year. (If storage period is

for a full year, then completing a comprehensive performance test for the machine is necessary. The

qualified machine may then be used normally.)

2.2 SHIPPING

Design Filtration Microzone assembles and factory tests each unit prior to shipping. On occasion, damage can

occur during the transportation of your new Mini Laminar Flow. You should inspect your unit for damage that

may have been caused and report it immediately to your transportation company.

All shipments are F.O.B. Design Filtration Microzone. All claims for damage due to shipping should be filed with

the carrier. Design Filtration Microzone and its dealers are not responsible for shipping damages.

Your Mini Laminar Flow has been factory tested and set for appropriate airflows, however all HEPA filtered

equipment should be certified by a qualified technician prior to use.

If you are re-shipping this unit, please ensure work tray and window sash are well taped closed before

transportation.

2.3 UNPACKING

Necessary tools for unpacking: electric drill with hexagon head M8.

2.4 INSTALLATION

Prior to Installation

Before installing your new VBM-200 Mini Laminar Flow please note:

Inspect the crate carefully for any exterior damage.

If the unit was transported during extreme temperatures, place the crate in heated/cooled area for 24

hours to allow all the components to come to room temperature.

Prepare and clean the area where the cabinet is to be installed.

During and after unpacking inspect the interior and exterior of the hood itself to be sure that it is free

from damage before completing the installation.

Use the packing list to be certain that all components are present, including all the necessary

hardware.

Page 6: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 6 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

Location

It is important that your new VBM-200 Mini Laminar Flow is installed correctly and in the best location possible.

It should be located in such a way that the airflows within the cabinet will not be disturbed by overhead

ventilation systems, doors that open and close frequently, or heavy traffic areas. Being located too close to any

of these things will have a negative impact on the airflows within the cabinet.

2.5 TECHNICAL PARAMETERS

The main technical parameters are shown below:

Parameters MODEL VBM-400

External Size (W x D x H) 21.75” (550) x 18.25” (460) x 27.63” (700 mm)

Internal Size (W x D x H) 18.75” (480) x 13.25” (340) x14.5” (370 mm)

Voltage 115V +/-10%

Frequency 50Hz/60Hz +/-1Hz

Consumption ≤100 W

Airflow Velocity 80-100fpm (0.3~0.5m/s)

UV Lamp 8 W

Fluorescent Lamp 8 W

HEPA Filter 99.99% efficiency at 0.3 µm

Noise ≤ 65 dB (A)

Working Pattern Vertical Positive Pressure

3. PACKING LIST

Name Quantity

Main Body 1

RVV Power cord 1

Fuse (5A) 1

User manual 1

Test Report 1

Certificate of Quality 1

Warranty Card 1

Page 7: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 7 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

4. PERFORMANCE INDEX

4.1 VIBRATION AMPLITUDE

The net vibration amplitude between 10Hz and 10KHz is less than 5µ (rms).

4.2 ILLUMINATION

Average illumination is not less than 650lx. Each illumination measured value is not less than 430lx.

4.3 ELECTRICAL CHARACTERISTICS

Maintain no breakdown in 5s when the voltage increases to 1390V alternating voltage in 5s.

Ground resistance ≤0.1Ω

Page 8: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 8 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

5. STRUCTURE

5.1 PARTS IDENTIFICATION

1. Control panel 4. Side glass window

2. Fluorescent lamp 5. Work table

3. Front glass window 6. UV lamp

5.2 CONSTRUCTION MATERIAL

1. The main body is made of cold-rolled steel with anti-bacteria powder coating. It enhances the

structural strength and allows the whole unit to be more stable.

2. The working area is made of 304 stainless steel, which is easily cleanable and corrosion-resistant.

3. The control panel has touch technology, which has an attractive appearance and can be operated

extremely easily.

4. Suitable for PCR work.

Page 9: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 9 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

5.3 FRONT WINDOW DRIVING SYSTEMS

Front window driving system consists of two constant springs and glass window.

5.4 HEPA FILTRATION SYSTEM

The HEPA filtering system is the most important part of any system and ensures the final performance of this

equipment.

The system consists of a motor/impeller and a HEPA filter assembly. The main function of this system is to

continually supply positively pressurized HEPA Filtered Air combined with the correct airflow velocity to achieve

the ISO-5 conditions required.

5.5 ULTRAVIOLET SOURCE

Make sure to turn off the fluorescent light and supply fan before lowering window to perform decontamination.

The UV lamp is located inside of the working zone and is fixed at the top-inside of the stainless steel, so that

the maximum lighted working area is ensured.

CAUTION: Make sure to turn off UV light before opening window after decontamination is finished.

5.6 LIGHTING SOURCE

Lighting is provided by a straight tube fluorescent lamp on early versions and LED lights on new units, and

ensures the average illumination in the working area is more than adequate for your needs.

Page 10: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 10 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

5.7 CONTROL PANEL (SERIES ONE UNITS)

1. Fan control

2. Ultraviolet light

3. Fluorescent light

4. Speed control up/down

5. Speed light bar

CAUTION: MAKE SURE UV LIGHT IS TURNED OFF BEFORE OPENING THE WINDOW!

5.8 FUSE

This equipment is equipped with a main power fuse and null line fuse, which are located next to the power cord

exports on the cabinet side. In terms of specific models of fuses, please see Section 12.

Page 11: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 11 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

6. AIR FLOW DIAGRAM

7. OPERATION

7.1 WORKING ENVIRONMENT

1. Indoor use only

2. Environment temperature: 15ºC~35ºC

3. Relative humidity: ≤75%

4. Pressure range: 70kPa~106kPa

7.2 OPERATING PROCEDURE

A. Plug in AC 115V +/-10%, 50Hz/60Hz +/-1Hz alternating current power supply.

B. Press appropriate functions keys as desired to operate (refer to Section 5.7).

C. Before use, turn off supply blower, decrease the glass window to its lowest position and turn on the UV

lamp. Then disinfect the working zone for at least 30 minutes.

WHEN UV LIGHT IS OPERATIONAL, IN ORDER TO PROTECT EYES AND SKIN AND TO AVOID THE

DAMAGE CAUSED BY INADVERTENT EXPOSURE, MAKE SURE WINDOW IS CLOSED AT ALL TIMES!

IF THE UV LIGHT IS AN IMPORTANT PART OF YOUR DECONTAMINATION PROCEDURES THEN WE

RECOMMEND CHANGING THE UV LIGHTS AT EACH CERTIFICATION PERIOD, TYPICALLY YEARLY.

Room Air

Pre-filtered Air

HEPA Filtered Air

Post-Process Air

Protected Area

Post-Process Area

Page 12: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 12 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

D. Open the glass window to the appropriate position and turn on the blower. Before starting work

operate the blower for 30 minutes to purge the work zone.

E. After use, turn off blower, close the glass window to the lowest position and turn on the UV lamp if

required.

8. MAINTENANCE

As the length of operation times and the quality of the back ground air will vary we recommend preparing a

detailed record of operating time for future reference. It is recommended to have any HEPA filtered equipment

certified at least once a year by a qualified person.

8.1 CABINET SURFACE CLEANING

A. Working zone surface cleaning

Use a cotton cloth soaked in concentrated soap solution to wipe the whole surface. Then use another

cotton cloth or towel soaked in hot or warm water to clean out any remaining soap foam. Finally, wipe

the surface by using a dry cloth or towel quickly.

A contaminated or marked work surface can be cleaned with medical alcohol or other disinfectant.

B. Outside surface and glass window cleaning

Any kind of non-abrasive household cleaner and a soft cotton cloth can be used to wipe the surface,

including alcohol-based cleaners.

8.2 COMPREHENSIVE MAINTENANCE CYCLE

We suggest that the comprehensive maintenance cycle should be one year or 1000 working hours.

8.3 MAINTENANCE METHOD

1. Daily or weekly maintenance

a. Working zone disinfection and cleaning (see Section 8.1 A)

b. Working zone surrounding surface and glass window cleaning (see Section 8.1 B)

c. Check whether the equipment’s functions are unusual.

d. Record this maintenance.

2. Monthly maintenance

a. Outside surface cleaning (see Section 8.1 B)

b. Wipe the equipment’s working table and working zone inside surface (except airflow grid in

working zone) by using 70% medical alcohol or 1:100 dilution household bleach (0.05%

sodium hypochlorite) first. Then use sterile water to wipe them again to remove residual

chlorine.

c. Check the pre-filter for dirty appearance.

NO DISINFECTANT CAN DAMAGE THE 304 STAINLESS STEEL INTERIOR.

Page 13: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 13 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

d. Check whether the equipment’s functions are unusual.

e. Record this maintenance.

3. Annual maintenance

a. Check whether the tightness of the front glass window’s driving system has remained the

same. If window springs have difficultly holding window open, then change the window

springs.

b. Check the fluorescent lamp and change the UV if this function is being used.

c. In order to ensure the safety of the cabinet we recommend certification by a qualified person.

d. Check the pre-filter for dirty appearance and change if necessary.

e. Record this maintenance.

8.4 COMMON FAULT TESTS

Before diagnosing a fault, please check:

1. Is the power connected correctly?

2. Is there any obvious breakage of the power cord?

3. Is the fuse tube functional?

Fault Items to check

UV lamp is not bright or lighting as usual

Is the lamp base connected correctly?

Is the lamp functional? Change if required.

Is the fluorescent light turned off?

Fluorescent lamp is not bright or lighting as usual.

Is the lamp base connected correctly?

Is the lamp functional? Change if required.

Is ballast working correctly? Change is required.

A button is ineffective or not functioning as usual. Is there any obvious breakage of the button?

Change if required.

Equipment is not powered.

Is the power cord connected correctly?

Is there any obvious breakage of the power cord?

Is the fuse tube functional?

1. Servicing of the above electrical components should be completed by a qualified person under the

safe conditions (power is cut off). Any other components should not be removed, as any resulting

consequence will be the responsibility of the user.

2. If the equipment’s malfunction is not listed above and the operator cannot solve the issue

immediately, please contact Design Filtration Microzone Inc. For your safety, do not attempt to repair

the equipment by yourself.

3. The maintenance of this equipment can only be completed by trained and certified workers.

4. If you need to order any component, you can contact our technical service department. Please specify

the model and serial number of the cabinet.

Page 14: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 14 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

9. SERVICE WORK

Please note this should be performed by a qualified person.

9.1 CHANGING THE SUPPLY HEPA FILTER.

1. Disconnect power from power source. Remove top access panel by unscrewing the two Philips screws

and releasing the two handles.

2. Unscrew the qty-4 16mm bolts holding the motor/blower assembly on top of the HEPA filter and lift up

out of the way.

3. Gently remove the old HEPA filter. Compare the new HEPA filter to ensure it is the same size and

model as the old filter.

Page 15: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 15 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

4. Lower the new HEPA filter back into position, noting any directional flow arrows and making sure the

HEPA filter is located in between the qty-4 locating brackets. Lower motor/blower housing gently back

down on to the HEPA filter, locating it squarely on top of the filter. Install qty-4 16mm bolts ensuring at

least ½” is screwed in to place and then lower nuts to torque down the hosing. Do not over torque

these nuts.

5. Install top panel.

PLEASE NOTE:

HEPA filters are fragile and must be handled with extreme care to avoid punctures or other damage which will

have a direct impact on the effective operation of your equipment. Your hood should be fully certified by a

qualified professional following a HEPA filter replacement to be certain the airflows are correct and that there

are no leaks in the filter.

9. 2 CHANGING THE MOTOR/BLOWER IMPELLAR

1. Disconnect power from power source. Remove top access panel by unscrewing the two Philips screws

and releasing the two handles.

2. Unscrew the qty-4 16mm bolts holding the motor/blower assembly on top of the HEPA filter and lift up

out of the way.

Page 16: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 16 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

3. Disconnect wiring while noting colours and remove on to bench.

4. Disconnect perforated supply screens and qty-4 screws hold impeller in place. Note if the new motor is

the same type and voltage as the old unit and assemble in reverse order as above.

5. If changing the motor capacitor, please make sure this is the same type and size as old unit.

Page 17: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 17 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

6. Lower motor/blower housing gently back down on to the HEPA filter, locating it squarely on top of the

filter. Install qty-4 16mm bolts ensuring at least ½” is screwed in to place and then lower nuts to

torque down the hosing. Do not over torque these nuts.

7. Install top panel.

PLEASE NOTE:

HEPA filters are fragile and must be handled with extreme care to avoid punctures or other damage which will

have a direct impact on the effective operation of your equipment. Your hood should be fully certified by a

qualified professional following a HEPA filter replacement to be certain the airflows are correct and that there

are no leaks in the filter.

Page 18: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 18 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

10. SPARE PARTS LIST

Name Quantity Part Number

Pre-filter pad 1 VBM-PRE

HEPA Filter 1 20PR0018

Motor Impeller 1 VBM-MI-190-65w

Capacitor 1 VBM-MC-6uf

Window spring kit 1 VBM-SS-2

Fluorescent bulb 1 20PR0019

UV light bulb 1 20PR0020

Control panel 1 VBM-CTLP-2

UV and Fluorescent Ballast 1 VBM-LB-T58W

11. NOTES

All repairs should be performed by qualified certifiers only. If you do not have a certifier that you work with,

please contact DFMZ and we can help you find a qualified certifier in your area.

Please make sure that all electrical requirements are met before connecting and using this equipment. This

equipment requires a three prong outlet. DO NOT remove the third prong in order to use an incompatible

outlet. If necessary, contact a qualified electrician to install a properly grounded three prong outlet.

Maximum weight of objects on the work surface should not exceed 23 kg or 50 lbs.

If possible, try to avoid the use of equipment that could shake or vibrate the hood (such as a centrifuge or

spiral oscillator). The vibrations given off by this type of equipment can lead to shedding of particles from the

HEPA filter, resulting in unclean work area or exposure to the operator.

NEVER USE OPEN FLAME INSIDE CABINET – open flames will disrupt the airflow and cause damage to the

filter. If high temperature sterilization is required, this should be done outside the cabinet.

HEPA filters have a limited service life up to several years but they CANNOT be cleaned. Debris accumulates on

the face of the filter, leading to pressure drops. Contact your certifier for service/HEPA filter replacement.

All sealed openings on the cabinet should not be tampered with. If a situation arises where access is required,

please contact a qualified certifier or DFMZ for assistance.

Storage period for this equipment is one year. Following any extended storage, the equipment should be

carefully cleaned, inspected and fully certified before first use.

DFMZ Group is not responsible for any damage or injury resulting from incorrect use of this equipment.

Page 19: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 19 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

12. ELECTRICAL WIRING DIAGRAM

Page 20: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 20 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

13. SERVICE/MAINTENANCE NOTES

Page 21: VBM-200 MINI LAMINAR FLOW - Design Filtration Microzone · 2.3 UNPACKING Necessary tools for unpacking: electric drill with hexagon head M8. 2.4 INSTALLATION Prior to Installation

VBM-200 MINI LAMINAR FLOW Installation and Operations Manual Revision: 10-May-17

Page 21 of 21

86 Harry Douglas Dr., Ottawa, ON, K2C 2C7

T: 613-831-8318, TF: 1-888-636-8609 DESIGN FILTRATION MICROZONE INC.

www.dfmzgroup.com

14. WARRANTY

Design Filtration Microzone warrants the equipment to be free of defects in materials and workmanship from

the date of invoice. The VBM Mini Laminar Flow shall be warranted against defects for a period of two years.

Any units or parts found to be defective during this period will be replaced or repaired at our discretion. The

buyer agrees to assume all transportation charges prepaid for return of unit to factory for repair. Design

Filtration Microzone will not accept charges for removal or re-installation.

Design Filtration Microzone will not accept any returned items without prior written consent. Design Filtration

Microzone is not responsible for repair or replacement by misuse or abuse by incorrect operating procedures of

its equipment. Pre-filters and carbon filters carry no warranty against loading. Any modifications or changes

made to equipment by customer without written consent of Design Filtration Microzone will void the warranty.

Model number and serial number located on your VBM Mini Laminar Flow must be provided to verify invoice

date. Failure to provide will void warranty.

FOR FURTHER INFORMATION, PLEASE CONTACT US DIRECTLY.

Design Filtration Microzone Inc., 86 Harry Douglas Drive, Ottawa, ON. K2S 2C7

T: 1-888-636-8609 (Canada & USA), T: 1-613-831-8318, F: 1-613-831-8321

[email protected]


Recommended