15
Ver 1.3 Introduction to NFC Tester RWC5010A RedwoodComm Ver 1.3 Contents Overview Key Features Test Setup GUI Architecture Test Jig Automated PC Software Demo 2
Transcript
Page 1: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Introduction to NFC Tester

RWC5010A

RedwoodComm

Ver 1.3

Contents

Overview

Key Features

Test Setup

GUI Architecture

Test Jig

Automated PC Software

Demo

2

Page 2: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Overview

RWC5010A NFC Tester A test equipment to test NFC Forum Devices

specified in NFC Forum Technical Specifications.

Provides very convenient working environment with full control over all system parameters.

Supports various kinds of functions such as power vs. time measurement, link analyzer and modulation characteristic measurement.

FDT and GT are measured in real time with Link analyzer.

Provides sensitivity test with variable VDC and load modulation amplitude

3

Ver 1.3

Key Features

Protocol supported LISTEN Test, POLL Test, and SNIFF Mode

Digital Protocol: NFC-A, NFC-B, NFC-F, NFC-DEP, ISO-DEP

All types of NFC tags supported: Type 1, 2, 3 and 4

User-controllable ASK modulation index

Link Analyzer Protocol Message logger and analyzer

Link timing measurement (GT/FDT/TR0)

Tests & Measurements Graphical Power vs. Time (built-in oscilloscope)

Modulation characteristic and load modulation measurements

Sensitivity Tests

Command-based Remote Control RJ45 and RS232C port

4

Page 3: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Tests & Measurements List

5

Polling DUT Listening DUT

Power CW ON/OFF power meas -

Frequency Carrier Freq meas Freq tuning

Link Timing GT, FDT, TR0 meas FDT, TR0 meas

Modulation Rise time, fall time,

overshoots (undershoots),

modulation index meas

Load modulation meas

Sensitivity Load modulation,

Position

Power,

Position

Ver 1.3

Specifications

Frequency

Range: 13.50MHz ~ 13.60MHz

Resolution: 1kHz

Accuracy: ±1ppm/year @ operating temperature

Output Level

Range: MAX. 20Vpp @ 50Ω

Resolution: 1mV

Accuracy: ±10mV

6

Page 4: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Calibration

Test & Measurement

Test Setup – LISTEN Test

7

Reference Poller Antenna (0,3,6)

RWC5010A

Listening DUT

TX

RX

Reference Listener Antenna (1,3,6)

VDC

Reference Poller Antenna (0,3,6)

RWC5010A

TX

RX

Ver 1.3

Calibration

Test & Measurement

Test Setup – POLL Test

8

Reference Listener Antenna (1,3,6)

RWC5010A

TX

RX

Reference Poller Antenna (0,3,6)

Load Modulation

Polling DUT

Reference Listener Antenna (1,3,6)

RWC5010A

TX

RX

VDC

CW using SG

Measure

Page 5: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Test Setup – SNIFF Mode

Test & Measurement

9

Polling DUT

Reference Listener Antenna (1,3,6)

RWC5010A

RX

Listening DUT

Ver 1.3

Pre-calibration

VDC

In LISTEN Test, the VDC can be set directly with the

pre-calibrated data according to the reference

Poller/Listener combinations, not by tuning TX level

to reach its target value.

Load Modulation

In POLL Test, the load modulation amplitude can be

set directly with the pre-calibrated data according to

the reference Listener, not by tuning TX level to reach

its target value.

10

Page 6: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

GUI Architecture

MAIN MENU

LISTEN TEST

The Tester operates as a reference Polling Device

POLL TEST

The Tester operates as a reference Listening Device

SNIFF

The Tester sniffs the communication between two NFC devices

GENERAL

General-purpose SigGen and Pow-Time

SETUP

System configuration

11

Ver 1.3

Sub Menu

LISTEN TEST

LINK / POW-TIME / SENSITIVITY / CONFIG

POLL TEST

LINK / POW-TIME / MEASURE / CONFIG

SNIFF

LINK / POW-TIME

GENERAL

SG / POW-TIME

SETUP

SYSTEM / REMOTE

12

Page 7: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

LISTEN TEST :: LINK

Link Analyzer

Protocol messages logging: PL and LP

GT/FDT(TR0) measurement

Unwanted messages filtering

13

Ver 1.3

Description on LINK Screen

Sub Menu

Link messages:

Black (Tester signal)

Blue (DUT signal)

Byte-format

data @ cursor

Run STATUS:

RUNNING

STOPPED

Run MODE:

SINGLE

CONT

PROTOCOL:

NFC-A/B/F

Status Indicator:

RMT – Remote enabled

AGC – AGC locked

TRG – Meas triggered

CAP – CAPs locked

ETH – Ethernet connected

EXT – External ref. detected

F – Function key enabled

Main Menu title

Current Protocol

FDT/GT : link timing (GT, FDT, TR0)

CMD : commands and responses

CONTENTS : information in commands and responses

14

MSG CLR:

Clear messages

FILTER:

Command mask

Page 8: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

LISTEN TEST :: POW-TIME

Power-Time Measurement

Measures Load modulation as Vpp

Similar to use of oscilloscopes (Zoom, Markers, etc.)

Running mode: Single or Continuous

Command-selective measurement

15

Ver 1.3

Description on POW-TIME Screen

SYNC:

Select Target CMD

(ALL, SENS, SDD, …)

TIMESPAN:

0~38ms

DISPLAY:

OFF / RESULT /

MSG / BOTH

Values on

Marker 1

Values on

Marker 2

Delta Marker

Hardware KEYs:

Zoom In

Zoom Out

Zoom Full

Zoom area

indicator

Marker 1 (Red):

Solid if selected

Dotted if not

Marker 2 (Blue):

Solid if selected

Dotted if not

16

Max/Min values

of window

Load modulation

(Vpeak-to-peak)

Horizontal/

Vertical DIV

Run STATUS:

RUNNING

STOPPED

Run MODE:

SINGLE

CONT

PROTOCOL:

NFC-A/B/F

SCALE:

AUTO/MANUAL

Page 9: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

LISTEN TEST :: SENSITIVITY

Receiver Sensitivity Test

Validates responsiveness of a Listening DUT

User-configurable X-axis: LEVEL (VDC) or TIME

Load modulation measurement when response exists

Various test conditions:

position on Test Jig, VDC level, modulation index

17

Ver 1.3

Description on SENSITIVITY Screen

18

Run STATUS:

RUNNING, STOPPED

X_AXIS:

VDC, TIME

REPEAT:

Number of try

VDC RANGE:

START, STOP, STEP

VDC TYPE:

NOMINAL, MIN, MAX, USER

VDC:

VDC value in Volt

DISPLAY:

BOTH / SENS / LOAD

Page 10: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

LISTEN TEST :: CONFIG

Parameter Configuration

Digital Protocol

Analog

19

Ver 1.3

POLL TEST :: LINK

Link Analyzer

Protocol messages logging: PL and LP

GT/FDT(TR0) measurement

Unwanted messages filtering

20

Page 11: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

POLL TEST :: POW-TIME

Power-Time Measurement

Similar to use of oscilloscopes (Zoom, Markers, etc.)

Running mode: Single or Continuous

Command-selective measurement

Measurement index view for modulation characteristic

21

Ver 1.3

POLL TEST :: MEASURE

Modulation Characteristic

Select a command as measurement target

Timing measurement according to the Spec

Power and Carrier Frequency measurement

Automatic calculation of UL and LL for each

22

Page 12: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Description on MEASURE Screen

SYNC:

Select Target CMD

(ALL, SENS, SDD, …)

RESET:

Reset Tester status

Carrier frequency

CW OFF Power

CW ON Power

Modulation

characteristic

measurements

depending

on protocol

Lower/upper limits

according to the Spec.

Measured values:

Blue if passed

Red if failed

23

Run STATUS:

RUNNING

STOPPED

Run MODE:

SINGLE

CONT

PROTOCOL:

NFC-A/B/F

Ver 1.3

POLL TEST :: SENSITIVITY

Load Modulation Sensitivity

Load modulation amplitude is pre-calibrated

according to the combination of reference antenna

Various test conditions:

Position on Test Jig

Load modulation amplitude

24

Page 13: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

POLL TEST :: CONFIG

Parameter Configuration

Digital Protocol

Analog

25

Ver 1.3

SNIFF :: LINK

Link Analyzer

Captures protocol messages between two NFC

devices as ‘sniffer’

GT/FDT(TR0) measurement

Ready to receive all type of NFC protocol (A/B/F)

26

Page 14: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

SNIFF :: POW-TIME

Power-Time Measurement

Similar to use of oscilloscopes (Zoom, Markers, etc.)

Running mode: Single or Continuous

Command-selective measurement

27

Ver 1.3

Test Jig

28

Page 15: Ver 1.3 Introduction to NFC Tester - nubicomnubilight.nubicom.co.kr/upload/datasheet/RWC5010A.pdf · 2014. 1. 9. · FDT and GT are measured in real time with Link analyzer. Provides

Ver 1.3

Automated PC Software

29

Ver 1.3

Demo

30


Recommended