+ All Categories
Home > Documents > Verification of RIS data using DICOM data for the purpose ... · Verification of RIS data using...

Verification of RIS data using DICOM data for the purpose ... · Verification of RIS data using...

Date post: 30-Jun-2018
Category:
Upload: doanquynh
View: 221 times
Download: 0 times
Share this document with a friend
17
Verification of RIS data using DICOM data for the purpose of patient dose audit IPEM MPEC, Oct 2012 Charnock P 1 , Couch T 1 , Dunn AF 1 , Fazakerley J 1 , Graham S 1 , Jones R 1 , Moores BM 1 , Murphy J 1 , Wilde R 1 , 1 IRS Limited, Liverpool
Transcript

Verification of RIS data using DICOM data for the purpose of

patient dose audit

IPEM MPEC, Oct 2012

Charnock P1, Couch T1, Dunn AF1, Fazakerley J1, Graham S1, Jones R1, Moores BM1, Murphy J1, Wilde R1,

1IRS Limited, Liverpool

Why use RIS data?

• Patient dose audits are used to ensure that doses are ALARP.

• Previously, these have been calculated using a small, hand-written sample (≈20 records).

• RIS contains computerised data for all examinations, it’s readily available and easy to import into excel.

Methodology for Patient Dose Audit

• Data may contain kV, mAs and DAP for each examination in each room

• FFD (&/or FSD) has to be assumed based on technique and standard patient*

• kV & mAs combined with results of level B physics tests to calculate ESD including backscatter*

• *It would be relatively simple to incorporate this data to improve accuracy

Why can’t we rely on RIS data?

• Large number of ‘zero’ or blank values in kV/mAs/DAP.

• Impossible dose values frequently occur i.e. mAs of >500 or kV >150 are unlikely.

• DAP not given any units, could be µGym2 or Gycm2 for example

• If they are wrong, what are the real values?

DICOM!

Characteristics of DICOM Data

• Generated by the system so no chance of human error.

• Contains much more data about image acquisition parameters.

But...

• It is difficult to gather dose data quickly en masse.

• Study labelling errors remain.

Types of Error in RIS Data

• Digits typed in the wrong order i.e. 789 instead of 897.

• Wrong digits entered e.g. 228 instead of 558

• Omission of digits i.e. 119 instead of 1119.

• Insertion of digits i.e. 13690 instead of 1369.

• Errors where the user appears to have entered the ‘default’ parameters

• Transpose of data, e.g. kV value in the mAs field (& vice versa)

0%

2%

4%

6%

8%

10%

12%

20

0

35

0

50

0

65

0

80

0

95

0

11

00

12

50

14

00

15

50

17

00

18

50

20

00

21

50

23

00

24

50

35

00

50

00

Re

lati

ve F

req

ue

ncy

Dose Length Product - mGy.cm

Correct Data

Errors

Distribution of Abdomen/Pelvis Examination DLPs

Methodology in ‘cleansing’ data

• Full RIS data arrives, including all modalities

• Take out only modality that is being audited

• Remove age < 16

• Remove all data were there is < 10 records for an examination within a room

• Remove all zero or blank data

This is the data we would like to report on

Methodology in ‘calibrating’ data

• For a room that we have equivalent DICOM data, take a random sample of about 1,500 records from the 4 most common examination in that room & calculate ESD

• Using accession number, find equivalent DICOM records, record and calculate ESD

• Compare statistical profile

XR Abdomen XR C Spine

RIS DICOM RIS DICOM

N 169 169 183 183

Mean 4.204 4.104 0.891 0.891

Std Dev 5.335 3.506 0.657 0.659

MIN 0.191 0.500 0.434 0.434

1st Quartile 1.850 1.850 0.546 0.546

Median 3.298 3.319 0.613 0.613

3rd Quartile 4.893 5.153 1.239 1.239

MAX 59.460 20.858 6.321 6.321

Example

XR Abdomen XR C Spine

RIS DICOM RIS DICOM

N 165 165 181 181

Mean 3.695 3.765 0.833 0.832

Std Dev 2.630 2.715 0.359 0.354

MIN 0.500 0.500 0.434 0.434

1st Quartile 1.850 1.834 0.546 0.546

Median 3.259 3.167 0.613 0.613

3rd Quartile 4.683 4.833 1.239 1.239

MAX 14.007 14.007 2.305 2.259

Example

Limitations

• DICOM only available where site has DR rooms – would need a different method of calibration for non-DR sites

• Currently collecting DICOM is still manual – takes about a day to collect c.300 records

Conclusions

• RIS data can be used for patient dose audit

• Recommend it is calibrated regularly against known data set such as DICOM

• Results of calibration can be fed back to departments, number of records removed can be set as KPI

Thanks

• Fellow authors

• Local radiographers

• Local RIS/PACS managers

[email protected]

0%

2%

4%

6%

8%

10%

12%

14%

0 250 500 750 1000 1250 1500 1750 2000 2250

Re

lati

ve F

req

ue

ncy

RIS Dose Length Product (mGy.cm)

CT Chest Dose Distribution

Double distribution – graphs courtesy of Tom Couch

0.0%

0.5%

1.0%

1.5%

2.0%

2.5%

3.0%

3.5%

4.0%

4.5%

5.0%

2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5

Re

lati

ve F

req

ue

ncy

RIS Dose Length Product - Ln(mGy.cm)

CT Chest Dose Distribution

Double distribution – graphs courtesy of Tom Couch

0.0%

0.5%

1.0%

1.5%

2.0%

2.5%

3.0%

3.5%

4.0%

4.5%

5.0%

2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5

Re

lati

ve F

req

ue

ncy

RIS Dose Length Product - Ln(mGy.cm)

CT Chest Dose Distribution by Protocol

Chest -CE

Chest -CE Lung Nodule Other

Double distribution – graphs courtesy of Tom Couch


Recommended