Date post: | 03-Jun-2018 |
Category: |
Documents |
Upload: | chris-jones |
View: | 216 times |
Download: | 0 times |
of 156
8/11/2019 Vertex 70vs Manual
1/156
User Manual
1007253
VERTEX 70v
8/11/2019 Vertex 70vs Manual
2/156
1stedition 2007, publication date November 2007
2007 BRUKER OPTIK GmbH, Rudolf Plank Str. 27, D-76275 Ettlingen, www.brukeroptics.com
All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any
means including printing, photocopying, microfilm, electronic systems etc. without our prior written per-
mission. Brand names, registered trademarks etc. used in this manual, even if not explicitly marked as
such, are not to be considered unprotected by trademarks law. They are the property of their respective
owner.
The following publication has been worked out with utmost care. However, Bruker Optik GmbH does
not accept any liability for the correctness of the information. Bruker Optik GmbH reserves the right to
make changes to the products described in this manual without notice.
This manual is the original documentation for the VERTEX 70v spectrometer.
8/11/2019 Vertex 70vs Manual
3/156
VERTEX 70v User Manual iii
.
.
.
.
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TABLEOFCONTENTS
1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Warning Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Delivery Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Connecting VERTEX 70v to the Power Supply . . . . . . . . . . . . . . . . . . . 11Connecting VERTEX 70v to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Connecting VERTEX 70v to the Vacuum Pump . . . . . . . . . . . . . . . . . . 13Connecting VERTEX 70v to the Purge Gas Line. . . . . . . . . . . . . . . . . . 15
4 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19External Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Optical Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Switching VERTEX 70v On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31QuickLock. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Automatic Accessory Recognition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Performing a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Evacuating and Venting the Spectrometer. . . . . . . . . . . . . . . . . . . . . . . 36Optimizing the Vacuum Operation of the Spectrometer. . . . . . . . . . . . . 40
Purging the Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Exchanging the Beamsplitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Exchanging the Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Cooling an MCT Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6 Maintenance and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Evacuating the MCT Detector Dewar. . . . . . . . . . . . . . . . . . . . . . . . . . . 58Replacing the Laser Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61Replacing a defective IR Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Replacing the Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8/11/2019 Vertex 70vs Manual
4/156
iv VERTEX 70v User Manual
Ta b l e o f C o n te n ts
Replacing the Sample Compartment Windows . . . . . . . . . . . . . . . . . . . 69Cleaning the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73Maintaining the Vacuum Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75Diagnostic Means. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76Problem - Possible Cause - Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
A Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
B Consumable Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
C Default Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
D Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
E Connecting VERTEX 70v to PC . . . . . . . . . . . . . . . . . . . . . . 115General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115Possible Connection Topologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116Selecting Network Addresses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119Assigning Network Addresses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120Checking the Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
F Electronics and Power Supply . . . . . . . . . . . . . . . . . . . . . . . 125Electronics Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125Power Supply Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
G Firmware Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131Updating the Firmware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132Restoring a previous Firmware Version . . . . . . . . . . . . . . . . . . . . . . . . 134Backing up the current Firmware Version . . . . . . . . . . . . . . . . . . . . . . 135
H Sample Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137Sample Preparation Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
I Service Addresses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
8/11/2019 Vertex 70vs Manual
5/156
8/11/2019 Vertex 70vs Manual
6/156
2 VERTEX 70v User Manual
1 S A F E T Y
Warning Labels
Correct Usage
The spectrometer and its components should only be used according to the instructions
described in the manual or advised by a Bruker engineer. In case of accessories or
components made by other manufacturers and used in connection with the spectrome-
ter, Bruker does not assume any liability for safe operation and proper functioning.
. . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
W A R N I N G L A B E L S
When operating the spectrometer you have to observe a number of safety instructions
which are highlighted by various warning labels. This section describes the warning
labels and explains their meaning. All warning labels on the spectrometer must always
be kept legible. Immediately replace a worn or damaged label.
The following warning labels indicate different dangerous situations which may be
caused by improper use of the spectrometer.
Caut ion - Genera l Hazard
This warning symbol indicates general hazard. Observe the safety
instructions and follow the precautions described to avoid personal injury
and/or property damage.
Caut ion - E lectr ica l Shock
This warning symbol indicates electrical hazard. The symbol is located near
live parts or on enclosures behind which are live parts that represent an
accidental contact hazard. Never touch these parts. Before removing the
corresponding compartment covers and beginning any maintenance or
repair work, first turn off the mains switch and unplug the main power cable.
Ensure that all live parts do not come into contact with a conductive sub-
stance or liquid. Non-observance of these safety instructions can cause
severe personal injury and/or property damage.
Caut ion - Hot Sur face
This warning symbol indicates components and surfaces which can
become very hot during spectrometer operation. Do not touch these com-
ponents and surfaces. Risk of skin burn! Be careful when operating nearhot components and/or surfaces.
Caut ion - Laser Radiat ion
This warning symbol indicates the existence of laser radiation. Never look
directly into the laser beam or use any kind of optical instruments to do so.
Otherwise permanent eye damage can be the result.
8/11/2019 Vertex 70vs Manual
7/156
8/11/2019 Vertex 70vs Manual
8/156
4 VERTEX 70v User Manual
1 S A F E T Y
Safety Instructions
Waste Disposal
Dispose all waste produced (chemicals, infectious and radioactively contaminated sub-
stances etc.) according to the prevailing laboratory regulations. Detergents and clean-
ing agents must be disposed according to the local waste regulations.
. . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
S A F E T Y I N S T R U C T I O N S
The following chapters describe all relevant safety aspects of the spectrometer opera-
tion. Depending on the degree of hazard the safety instructions are classified as fol-
lows:
Danger indicates that death, severe personal injury or substantial property damage
WILL result if proper precautions are not taken.
Warning indicates that death, severe personal injury or substantial property damage
CAN result if proper precautions are not taken.
Caution indicates that minor personal injury or property damage CAN result if proper
precautions are not taken.
Note draws your attention to particularly important information on the product,
e.g. product operation or to a special part of the manual.
The safety instructions Danger, Warningand Caution stand out by the corresponding
warning labels.
Caut ion - Corrosive Substance
This warning symbol indicates the possible presence of corrosive sub-
stances. When working with corrosive substances, always observe the lab-
oratory safety regulations and take protective measures (e.g. wearing
protective masks and gloves). Failure to do so may cause severe personalinjury or even death.
8/11/2019 Vertex 70vs Manual
9/156
VERTEX 70v User Manual 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G E N E R A L 2
VERTEX 70v is an evacuable, fully digital FT-IR spectrometer for demanding R&D
applications. The spectrometer is equipped with a number of features such as AAR
(Automatic Accessory Recognition) ACR (Automatic Component Recognition) and Per-
formanceGuard that facilitate performing spectroscopic measurements and ensure reli-
able measurement results. The function AAR identifies automatically the accessory
installed in the sample compartment, performs several tests and loads automatically the
corresponding experiment file including the pre-defined measurement parameters. The
feature ACR recognizes automatically the currently installed optical components like
source, detector and beamsplitter. These components are electronically coded so that
the spectrometer firmware can recognize them. This information is passed on to the
application software OPUS. The purpose of ACR is to enable the user to select the right
optics parameters in OPUS. In addition, the spectrometer components are monitored
permanently to ensure that they operate within the specification range. This feature is
called Performance Guard. Its purpose is to facilitate fault diagnostics and maintenance.
The data acquisition is based on a free running delta-sigma, dual-channel A/D converter
with 24-bit dynamic range. The A/D converter is integrated into the detector preamplifier
electronics. The DigiTect technology ensures a signal transmission free from interfer-
ences and guarantees the highest signal-to-noise ratio.
VERTEX 70v can be controlled by any data system (PC workstation, notebook etc.) on
which the operating system Microsoft Windows and the spectroscopic software OPUS isinstalled. The Ethernet connection provides the possibility to control the spectrometer
also via your intranet or the internet.
The standard spectrometer configuration is designed for data acquisition in the mid IR
region. Optionally, VERTEX 70v can be equipped with additional optical components to
cover the whole spectral range - starting in the far infrared or THz region at 10cm -1 up to
the ultraviolet region at 28,000cm-1. Due to the pre-aligned optical components and the
permanently aligned RockSolid interferometer, the spectral range can be changed eas-
ily. If you work with the advanced spectrometer configuration (i.e. two detector positions
and two source positions are available inside the spectrometer) you can select them
using the software. Removable vacuum-tight covers provide access to the detector and
beamsplitter if you want to exchange these components.
VERTEX 70v has five IR-beam outlet ports (on the right, front and left side) and two IR-
beam inlet ports (on the right and rear side) allowing the connection of a multitude of
optional accessories and/or components like:
TGA-coupling
PMA 50 (Polarization Modulation Accessory for VCD and PM-IRRAS)
HYPERION 1000/2000 IR microscope and HYPERION 3000 imagingmicroscope with FPA detector (Focal Plane Array detector system)
IMAC module (Imaging Accessory with FPA detector)
8/11/2019 Vertex 70vs Manual
10/156
8/11/2019 Vertex 70vs Manual
11/156
VERTEX 70v User Manual 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .I N S T A L L A T I O N 3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
G E N E R A L I N F O R M A T I O N
Unpacking and initial installation of VERTEX 70v is done by Bruker service engineers.
The operating company has to provide an installation site that meets the site require-
ments described in this chapter. (See also the technical document Installation Require-
ments for VERTEX 70vprovided by Bruker Optik GmbH in advance.)
This chapter contains a list of the standard as well as the optional spectrometer compo-
nents and describes the procedures for connecting the spectrometer:
to the power supply, to a PC,
to the vacuum pump and
to the purge gas supply line, if necessary.
For detailed information about how to install the computer, refer to the PC manual.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
D E L I V E R Y S C O P E
The basic instrument of VERTEX 70v allows upgrading with additional components and/or accessories. The delivery scope depends on the spectrometer configuration you
have ordered.
Standard Components
The basic instrument includes the following items:
VERTEX 70v spectrometer (including the user manual)
Power cord
PC compatible data system (if desired, the PC can also be provided by the
customer) Data cable (Cat5, crossover cable for 10Base-T Ethernet standard)
Purge gas hose (OD: 6mm, length: approx. 5m)
Tool kit (slot-head screw driver, cross-head screwdriver and hex keys of severalsizes, sample preparation tools, 3x spare fuses, IR sensor card, metallic cap
shown in fig. 28)
Software package OPUS/IR (including the OPUS Reference Manual)
8/11/2019 Vertex 70vs Manual
12/156
8 VERTEX 70v User Manual
3 I N S T A L L A T I O N
Delivery Scope
For installing the vacuum pump, the following items are included:
Vacuum pump (including the user manual)
Noise reduction hood
Vibration absorber
2x flexible metal hoses 4x hose clamps
4x sealing rings
Opt ional Components
Depending on the ordered spectrometer configuration, the delivery scope can also
include following optional components:
Optional spectrometer components (e.g. optional detectors) and/or accessories
Optional OPUS software packages (e.g OPUS/STEP) including the corresponding
manuals
Inspect ing the Packaging
After the receipt of the spectrometer, inspect the packaging for damages. If there are
any signs of damage, contact your local shipping representative before opening the
shipping box.
Warning: Do not put a spectrometer into operation that shows signs of
damage. Failure to do so may result in severe personal injuries
and/or property damage.
Transpor tat ion
Due to its weight (about 105kg), VERTEX 70v has to be carried by at least four persons
using the supplied transport handles. For transportation purposes, attach these handles
properly to the right and left spectrometer side as shown in figure 1 using 12 screws
(M5 x 16). Tighten the screws using a hex kex (size 4mm). After having transported the
instrument to the desired place, you can remove the transport handles again. Alterna-
tively, you can transport the instrument with a fork lifter.
Warning: Due to the high instrument weight, improper transportation can
lead to personal injuries and/or spectrometer damage.
8/11/2019 Vertex 70vs Manual
13/156
VERTEX 70v User Manual 9
.
.
.
.
.
I N S T A L L A T I O N
Site Requirements
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
S I T E R E Q U I R E M E N T S
Space Requirements
The spectrometer dimensions are 85cm (w) x 71cm (d) x 32cm (h). (For exact spec-
trometer dimensions refer to appendix D.) At the rear side, the spectrometer requires a
clearance of at least 25cm (10). The spectrometer should be placed on a stable and
horizontal base. Note that the basic instrument has a weight of about 105kg.
When preparing the installation location for the spectrometer, take into consideration
that the mains power supply connection is easily accessible at any time. The mains
power supply can be interrupted, for example, either by disconnecting the safety plug or
switching off the mains switch on the spectrometer rear side or disconnecting the pri-
mary power receptacle.
Environmenta l Requi rements
To ensure optimum spectrometer performance and long-term reliability the following
environmental conditions are essential:
Temperature Range: 18 - 35 C (64 - 95 F)
In case the vacuum pump is operated with installed noise reduction hood ensure the
ambient temperature does not exceed 32C (90F).
Humidity (non-condensing): 80% (relative humidity)
Temperature variations can impair the results of long-term measurements. Therefore,
the temperature variations should be less than 1C per hour and should not exceed 2C
per day for this type of measurement.
Figure 1: Installing the Transport Handles
Transport
Handles
8/11/2019 Vertex 70vs Manual
14/156
10 VERTEX 70v User Manual
3 I N S T A L L A T I O N
Site Requirements
Vibrat ion
Ideally, the spectrometer should not be installed near vibration sources (e.g. ventilation
hoods, air conditioners, motors, elevator etc.) or in rooms with intense floor vibration.
Power Supply
The spectrometer power supply unit automatically adapts to the most common power
sources.
Valid voltage range: 100 V AC to 240 V AC
Valid frequency range: 50 to 60 Hz
VERTEX 70v is an instrument of the protection class I.
Caution: To avoid personal injury and spectrometer damage, connect the
spectrometer only to a socket outlet with earthing contact.
To provide for good data quality and a long spectrometer service life, ensure that the fol-
lowing site requirements are met:
Do not install the spectrometer near sources of potential inductive electricalinterference (e.g. pumps, switching motors, microwave ovens etc.), sources of
high energy pulses, and sources that might cause magnetic or radio frequency
interference.
Do not place devices such as large electric motors, heaters, welding equipment,radio transmitting equipment, units emitting pulsed NMRs, or high powered lasers
in close vicinity to the spectrometer. These devices can interfere with thespectrometer and cause spectrometer malfunction. Ensure that these types of
devices are not connected to the same electrical circuit as the spectrometer.
If a reliable mains power supply is a problem at your site (caused by brownouts,power surges, frequent thunderstorms, for example), take precautions to ensure
an uninterruptible power supply.
8/11/2019 Vertex 70vs Manual
15/156
VERTEX 70v User Manual 11
.
.
.
.
.
I N S T A L L A T I O N
Connecting VERTEX 70v to the Power Supply
C O NN E CT I NG V E RT E X 7 0 V TO T H E PO W ER
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
S U P P L Y
Power CordBefore connecting the power cord, make sure that the spectrometer is switched off, i.e.
the mains switch (B in figure 2) is in the O position. Connect the supplied power cord
to the primary power receptacle (C in figure 2) as well as to the mains socket outlet.
The power cord length should not exceed 3m. Depending on the local conditions, the
original power cord may need to be exchanged for a power cord that complies with the
standards of the country in question. The power cord must have approbation of at least
your local authority, UL for US, CSA for Canada or VDE for Europe. The spectrometer
power supply unit automatically adapts to the local voltage and frequency range. (See
section Site Requirements.)
Component
A Ethernet port
B Mains switch
CPrimary power receptacle (for connecting the
power cord)
Figure 2: Spectrometer Rear Side - Connections for Power Supply and PC
A
B
C
8/11/2019 Vertex 70vs Manual
16/156
12 VERTEX 70v User Manual
3 I N S T A L L A T I O N
Connecting VERTEX 70v to a PC
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C O NN E CT I NG VE R TE X 7 0 V T O A PC
Data Cable
The data cable included in the spectrometer delivery scope is a CAT5 crossover cable(labelled Cross-over) with two RJ-45 plugs. This cable is only used for the direct con-
nection of VERTEX 70v to a computer. If you intend to connect the spectrometer to a
network, a different type of cable (i.e. non-crossover, CAT 5 cable for the 10Base-T Eth-
ernet standard) is required. (See appendix E.) The data cable length should not exceed
100m (without repeater).
Connect one end of the data cable to the Ethernet port (ETH) (A figure 2) and the other
end of the data cable to the RJ-45 socket of the computer network interface card. (For
detailed information refer to the computer manual.)
After having set up the data cable connection, turn on the spectrometer using the mainsswitch. After a few seconds, the spectrometer beeps once and starts a self test. After
the initialization has been completed successfully, the STATUS LED (figure 10) turns
from red to green. Now switch on the computer and the monitor. (For information on
how to install the computer and how to set up signal and power cable connections for
the computer, monitor etc. refer to the computer manual.)
Computer Setup
VERTEX 70v and the delivered PCare already configured for the stand-alone opera-
tion. The spectrometer IP address is factory-set to 10.10.0.1. In case you have not pur-
chased the computer together with the VERTEX 70v spectrometer, you have to assign
an appropriate IP address to the computer to which you want to connect the spectrome-
ter. For detailed information about how to assign an IP-address to the computer refer to
appendix E.
8/11/2019 Vertex 70vs Manual
17/156
VERTEX 70v User Manual 13
.
.
.
.
.
I N S T A L L A T I O N
Connecting VERTEX 70v to the Vacuum Pump
C O NN E CT I NG V ER T EX 7 0 V TO T H E VA C UU M
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
P U M P
The attachment flange (NW25 flange) for connecting the vacuum pump is at the spec-
trometer rear side. Figure 3shows the valve block with removed cover.
The vent openings are covered by a plug made from sintered-powder metal which is air-
permeable (i.e. the spectrometer can be vented with the plugs installed on the vent
opening). (See figure 8.) The plug functions like a filter preventing particles from enter-ing the spectrometer together with the influent air.
Note: For detailed information about the vacuum pump refer to the user
manual provided by the vacuum pump manufacturer.
Attachment Flange for the Vacuum Pump
Opening for venting
the optical bench(Note:When purging the
spectrometer this port is
used as purge gas inlet for
the optical bench.
Valve for evacuatingthe optical bench
Valve for evacuating
the sample compart-
ment
Figure 3: Valve Block (Spectrometer rear Side)
Opening for venting
the sample compart-
ment(Note:When purging the
spectrometer this port is
used as purge gas inlet for
the sample compartment.
Valve for venting the
sample compartment
Valve for venting the
optical bench
8/11/2019 Vertex 70vs Manual
18/156
14 VERTEX 70v User Manual
3 I N S T A L L A T I O N
Connecting VERTEX 70v to the Vacuum Pump
Insta l la t ion Procedure
Remove the valve block cover shown in figure 4by loosening the two Allen screwsusing a hex key (size 3mm) and pulling off the cover.
Install the supplied sealing ring at the attachment flange. See figure 5.
Press the supplied flexible metal hose against the attachment flange (figure 6a)and attach the hose to the flange using the supplied hose clamp (figure 6b).
Secure the hose clamp by fastening the wing screw.
Allen Screws Valve Block Cover
Figure 4: Removing the Valve Block Cover
Figure 5: Connecting VERTEX 70v to Vacuum Pump - Step 1
Attachment Flange Sealing Ring
FlexibleMetal Hose
8/11/2019 Vertex 70vs Manual
19/156
VERTEX 70v User Manual 15
.
.
.
.
.
I N S T A L L A T I O N
Connecting VERTEX 70v to the Purge Gas Line
During operation, the vacuum pump generates vibrations. In order to prevent these
vibrations from being transferred to the spectrometer via the flexible metal hose, the
supplied vibration absorber has to be installed between the vacuum pump and the
spectrometer. The procedure for connecting the flexible metal hose to the vacuum
pump and to the vibration absorber is identical to the procedure described above.
Note: Make sure that the vibrating metal hoses do not come into contact
with the table on which the spectrometer is placed.
During the operation, the vacuum pump produces an increased noise level. In order toreduce the noise level install the supplied noise reduction hood over the vacuum pump.
For information about the noise reduction hood installation refer to instructions provided
by the pump manufacturer.
C O NN E CT I NG V E RT E X 7 0 V TO T H E PU R GE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
G A S L I N E
As an alternative to the vacuum operation, VERTEX 70v can be purged with either dry
air or dry nitrogen gas. The spectrometer has two purge gas inlets; one for purging thesample compartment and the other for purging the optical bench. The purge gas inlets
are at the spectrometer rear side. See figure 7.
Figure 6: Connecting VERTEX 70v to Vacuum Pump -Step 2
Hose ClampWing Screw
8/11/2019 Vertex 70vs Manual
20/156
16 VERTEX 70v User Manual
3 I N S T A L L A T I O N
Connecting VERTEX 70v to the Purge Gas Line
For detailed information about the required purge gas supply conditions refer to chapter
Operation, section Purging the Spectrometer.
Insta l la t ion Procedure
Note: In case the spectrometer is evacuated, first vent it before starting
the installation procedure. Otherwise, a warning message regarding
unstable pressure conditions inside the spectrometer will appear.
To connect the spectrometer to the purge gas supply you need a stiff hose with anouter diameter of 6mm. Remove the plug (made from sintered-powder metal) from
the purge gas inlet by pressing the lock ring inwards (figure 8) and pulling out the
plug. Connect one end of the hose to your supply line for dry air or dry nitrogen
gas and insert the other end of the hose into the purge gas inlet for either the
sample compartment or optical bench.
If you want to purge both the sample compartment and the optical bench, youneed aT-shape connecting hose with two hose ends leading to the spectrometer.
After having connected the main end of the hose to the supply line, insertone of
the other two hose ends into the purge gas inlet for the sample compartment and
the other hose end into the purge gas inlet for the optical bench.
Figure 7: Purge Gas Inlets
Purge gas inlet for
optical bench.(Note:In case of vacuum
operation - vent opening for
venting the optical bench.)
Purge gas inlet for
sample compartment.(Note:In case of vacuum
operation - vent opening for
venting the sample com-
partment.)
8/11/2019 Vertex 70vs Manual
21/156
VERTEX 70v User Manual 17
.
.
.
.
.
I N S T A L L A T I O N
Connecting VERTEX 70v to the Purge Gas Line
Figure 8: Purge Gas Inlet with removed Plug
Plug
Lock Ring
8/11/2019 Vertex 70vs Manual
22/156
18 VERTEX 70v User Manual
3 I N S T A L L A T I O N
Connecting VERTEX 70v to the Purge Gas Line
8/11/2019 Vertex 70vs Manual
23/156
VERTEX 70v User Manual 19
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .O V E R V I E W 4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
G E N E R A L I N F O R M A T I O N
This chapter describes all relevant external and internal spectrometer components.
Note: The local indications r ightand leftassume that the operator stands in
front of the spectrometer. The indications forwardand backwardrefer
to the spectrometer front side and rear side, respectively.
Compartment
A Power Supply Connector
B Status Indicator Board
C Electronics Compartment
D Interferometer Compartment
E Detector Compartment
F Sample Compartment
G Laser
H Vacuum Pump Connection Port
I Beam Direction Control Compartment
A
B
C
D
E
F
GH
I
Figure 9: General Overview
8/11/2019 Vertex 70vs Manual
24/156
20 VERTEX 70v User Manual
4 O V E R V I E W
External Components
The detector compartment, the interferometer compartment and the beam direction
control compartment are not separated from each other but form one compartment. All
spectrometer compartments are accessible by removing the corresponding cover.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
E X T E R N A L C O M P O N E N T S
Status Indicator Board
The status indicator board is in the left rear corner of the spectrometer, more precisely,
on the electronic compartment cover. (See figure 10.) The color of the LEDs gives a
general indication of the operating status of the corresponding spectrometer compo-
nent. Moreover, the color of the Vacuum LED indicates the current pressure situation
inside the spectrometer compartments (i.e it shows whether a certain compartment is
being evacuated/vented just now or is already evacuated/vented). In case one of these
LEDs lights up red indicating a spectrometer problem refer to chapter Troubleshooting.This chapter shows possible causes of a problem and provides solutions.
V A C U U M
The color of VACUUM LED depends on the current pressure situation inside the individ-
ual spectrometer compartments. The following table explains the meaning of the differ-
ent LED colors:
LED is off. Sample compartment and optical bench are vented.
LED flashes green. Sample compartment and optical bench are being either evacuated or vented.
LED lights up green. Sample compartment and optical bench are evacuated. The ultimate vacuum isachieved.
LED flashes yellow. Sample compartment is being either evacuated or vented. (In case the sample
compartment is already vented, it flashes yellow also when the optical bench is
being vented.)
LED lights up yellow. Sample compartment is vented.
LED lights up red. When the spectrometer is being evacuated, but a certain threshold pressure value
is not reached within a certain period of time (i.e. the ultimate vacuum is not
achieved). A red VACUUM LED indicates a problem. See chapter Troubleshooting,
section Problem - Possible Cause - Solution, subsection Spectrometer problem
indicated by spectrometer status indicator.
Figure 10: Status Indicator Board
8/11/2019 Vertex 70vs Manual
25/156
VERTEX 70v User Manual 21
.
.
.
.
.
O V E R V I E W
External Components
LA S E R
The LASERLED lights green when the laser is in operation and the laser signal is OK.
The LASERLED lights up red if the laser power is too weak, the laser beam is blocked
or if the laser module is defective or out of alignment. (See chapter Troubleshooting,
section Problem - Possible Cause - Solution, subsection Spectrometer problem indi-
cated by spectrometer status indicator.) This control lamp also lights up red during the
spectrometer initialization phase. After the initialization is completed successfully, this
LED turns to green.
S TA TU S
A green STATUSLED indicates that the spectrometer is in proper operating condition.
The STATUSLED lights up red in case of a spectrometer malfunction or during the ini-
tialization phase. After the initialization is completed successfully, this LED turns to
green. (See chapter Troubleshooting, section Problem - Possible Cause - Solution, sub-
section Spectrometer problem indicated by spectrometer status indicator.)
Sample Compartment
Normally, you gain access to the sample compartment from the spectrometer top side
by removing the blue cover using the handle. See figure 11a. In exceptional cases, if
your measurement accessory requires access from the spectrometer front side (e.g. for
exchanging the sample), you can remove the blue front cover by loosening the six Allen
screws using a hex key size 3mm. See figure 11b.
Note: When performing measurements under vacuum condition do not
forget to reinstall the sample compartment front cover.
The sample compartment dimensions are 25.5cm (w) x 27cm (d) x 16cm (h). For more
information about the sample compartment interior refer to chapter Operation, section
QuickLock.
Figure 11: a) Sample Compartment Top Cover b) Sample Compartment Front Cover
Allen ScrewsSample Compartment Cover Handle
8/11/2019 Vertex 70vs Manual
26/156
22 VERTEX 70v User Manual
4 O V E R V I E W
External Components
IR Beam Por ts
VERTEX 70v has seven IR beam ports (five outlet ports and two inlet ports) allowing
the adaptation of external accessories and/or components (e.g. microscope, TG-IR cou-
pling or external light source). The IR beam ports are at the front and rear side as well
as at the left and right hand side of the spectrometer. For the exact dimensions of the IRbeam port positions refer to appendix D.
IR Beam Ports
A Outlet port for focussed beam (e.g. for connecting a bolometer)
BOutlet port for parallel beam (e.g. for connecting a fiber optic coupling
module)C Inlet port for connecting a light emission source (e.g. Hg source)
D Outlet port for parallel beam
EOutlet port for parallel beam (e.g. for connecting a microscope, PMA50,
external sample compartment XSA)
FInlet port for connecting a light emission source (e.g. FT-Raman mod-
ule, water-cooled, high-power MIR source)
GOutlet port for parallel beam (e.g. for connecting a microscope) or
focussed beam (e.g. for connecting a bolometer)
Figure 12: c) Right Side d) Left Side
Figure 12: a) Front Side b) Rear Side
A B
D E F G
C
8/11/2019 Vertex 70vs Manual
27/156
VERTEX 70v User Manual 23
.
.
.
.
.
O V E R V I E W
External Components
The IR beam ports are vacuum-tight sealed by circular covers. To remove a cover
loosen the six Allen screws using a hex key size 3mm. See figure 13.
Note: External accessories are installed by the Bruker service technicians.
Spectrometer Rear S ide
Allen Screws
Figure 13: Removing an IR Beam Port Cover
Figure 14: Spectrometer Rear View
External Beam
Port
Vent Opening/
Purge Gas Inlet
Electronics
Panel
Mains Switch
CAN BUS
Port
Primary Power
Receptacle
Attachment Flange
for Vacuum Pump
8/11/2019 Vertex 70vs Manual
28/156
24 VERTEX 70v User Manual
4 O V E R V I E W
External Components
E X TE R N A L B E A M P OR T
The inlet port is used for connecting a light source (e.g. Hg-Source) or an emission
sample.
V E N T OP E N IN G / P U R GE GA S IN LE T
Depending on whether you evacuate or purge the spectrometer, these two ports serve
different purposes. In case of evacuating the spectrometer these ports serve as vent
openings, whereas, when purging the spectrometer the purge gas supply lines are con-
nected to these ports. (For detailed information about installing the purge gas connec-
tion refer to chapter Installation.)
E LE C TR ON IC S P A N E L
On the electronics panel are a number of ports (e.g. Ethernet port), the reset button as
well as LEDs indicating, for example, the status of the interferometer. For a detailed
description of the electronics panel refer to appendix F.
M A IN S S WITC H A N D P R IM A R Y P OWE R R E C E P TA C LE
The mains switch is used to turn the spectrometer on and off. The power supply socket
is used to connect the power cord to the spectrometer.
C A N B U S P OR T
The CAN bus port is primarily used to connect external automated units to the spec-
trometer. For more information refer to appendix F.
A TTA C H M E N T FLA N GE FOR V A C U U M P U M P
The vacuum pump can be connected to this attachment flange (NW25) using the sup-
plied sealing ring, flexible metal hose and hose clamp. (For detailed information about
how to connect the vacuum pump to the spectrometer refer to chapter Installation.)
8/11/2019 Vertex 70vs Manual
29/156
VERTEX 70v User Manual 25
.
.
.
.
.
O V E R V I E W
Internal Components
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
I N T E R N A L C O M P O N E N T S
The following figure identifies only the most important internal components and their
location inside the spectrometer.
Component
A RockSolid interferometer (permanently aligned)
B DigiTect Detectors
C Sample holder for transmission measurements (exchangeablefor other optional accessories with QuickLock baseplate)
D HeNe laser
E Two beamsplitters storage positions (optional)
F Beamsplitter (operation position)
G Optional NIR source (operating position)
H MIR source (operating position)
I QuickLock mechanism for accessories (including connectors)
Figure 15: Internal Spectrometer Components
B
D
E
A
F
G
H
IC
8/11/2019 Vertex 70vs Manual
30/156
26 VERTEX 70v User Manual
4 O V E R V I E W
Internal Components
Light Source
The basic instrument is equipped with a MIR source (H in figure 15). The MIR light
source is a globar (i.e. an U-shaped silicon carbide piece) that emits mid-infrared light.
Apart from the standard air-cooled MIR source, the following optional sources are avail-
able:
VIS/NIR source (tungsten halogen lamp), installed in the spectrometer (G infigure 15), air-cooled
FIR source (mercury lamp), connected externally to the spectrometer, water-cooled
UV/VIS/NIR source (tungsten lamp), connected externally to the spectrometer,water-cooled
UV source (deuterium lamp), connected externally to the spectrometer, air-cooled
High power MIR source (globar), connected externally to the spectrometer, water-cooled
All external sources can be connected to one of the two inlet ports (C in figure 12b or F
in figure 12c). For the FIR source (mercury lamp), the preferred connection port is the
inlet port at the spectrometer rear side, C in figure 12b.
Detector
The basic spectrometer configuration is equipped with a DigiTect DLaTGS detector with
integrated preamplifier. This detector package contains an analog-to-digital-converter
that converts the analog signal from the detector directly into a digital signal. This digitalsignal is transmitted to the data processing electronics unit of the spectrometer. The
standard detector is a pyroelectric DLaTGS detector which covers a spectral range from
12,000 to 250cm-1, operates at room temperature and has a sensitivity of D*>4x108cm
Hz1/2W-1.
Apart from the standard detector, there is a large number of optional detectors. All
detectors are mounted on dovetail slides which allow an easy exchange. The following
optional detectors are available:
DetectorSpectral
Range (cm-1)Sensitivity
Operating
Temperature
Mid-Infrared
DLaTGS with KBrwindow
12,000 - 250 D*>4x108cm Hz1/2W-1 Temperature-sta-bilized
DLaTGS with CsI
window
12,000 - 160 D*>4x108cm Hz1/2W-1 Room temperature
MCT narrow band,
with BaF2window
CAUTION - HARMFUL!
12,000 - 850 D*:>4x1010cm Hz1/2 W-1 Liquid N2cooled
8/11/2019 Vertex 70vs Manual
31/156
VERTEX 70v User Manual 27
.
.
.
.
.
O V E R V I E W
Internal Components
Warning: Some detectors are equipped with windows of which the material
is harmful or (very) toxic. During normal spectrometer operation,
these materials do not pose a health risk. However, should these
windows break caused by mechanical impact, be extremely
careful. Avoid generating dust. These materials are harmful or
toxic if swallowed or inhaled. Also avoid skin and eye contact.
MCT mid band, withZnSe window
CAUTION - TOXIC!
12,000 - 600 D*:>2x1010cm Hz1/2 W-1 Liquid N2cooled
MCT broad band,with KRS-5 windowCAUTION - TOXIC!
12,000 - 420 D*:>5x109cm Hz1/2W-1 Liquid N2cooled
Photovoltaic MCT,
with BaF2 window
CAUTION - HARMFUL!
12,000 - 850 D*:>2x1010cm Hz1/2 W-1 Liquid N2cooled
MCT/InSb Sandwich,
with ZnSe windowCAUTION - TOXIC!
10,000 - 600 D*:>2x1010cm Hz1/2W-1 (MCT)
D*:>1.5x1011cm Hz1/2W-1(InSb)
Liquid N2cooled
Near-Infrared
InSb 10,000 - 1,850 D*:>1.5x1011cm Hz1/2 W-1 Liquid N2cooled
InSb with cold filter 10,000 - 3,100 D*>5x1011cm Hz1/2 W-1 Liquid N2cooled
Ge Detector (Raman) 11,750 - 5,900 NEP
8/11/2019 Vertex 70vs Manual
32/156
28 VERTEX 70v User Manual
4 O V E R V I E W
Internal Components
Beamspl i t ter
The standard KBr beamsplitter covers a spectral range from 8000 to 350cm-1. Apart
from the standard beamsplitter, there are also optional beamsplitters. They allow data
acquisition in wavelength ranges other than MIR (standard) when used in conjunction
with the appropriate light source and detector. Note that the combination of light source,
detector, beamsplitter and sample compartment window material defines the IR mea-
surement range. The following optional beamsplitters are available:
Caution: The beamsplitter material CaF2is harmful if inhaled or swallowed.
Avoid also skin and eye contact.
Laser VERTEX 70v is equipped with a HeNe laser (D in fig. 15) It emits red light with a wave-
length of 633nm. The rated power output is 1mW. The laser controls the position of the
moving interferometer mirror (also called scanner) and is used to determine the data
sampling positions. The monochromatic beam produced by the HeNe laser is modu-
lated by the interferometer to generate a sinusoidal signal. For information about how to
replace a defective laser module, refer to chapter Maintenance and Repair.
In ter ferometer VERTEX 70v is equipped with a high stability interferometer with ROCKSOLID perma-
nent alignment. The ROCKSOLID interferometer incorporates dual retroreflecting cube
corner mirrors in pendulum arrangement. The high throughput design ensures the high-
est possible signal-to-noise ratio.
Beamsplitter Spectral Range (cm-1)Color Coding of theBeamsplitter Handle
Mid-Infrared
KBr (standard) 7,500 - 370 red
KBr (broad band) 10,000 - 400 red
Csl 5,000 - 210 red
Near-InfraredCaF2CAUTION - HARMFUL!
15,500 - 1,200 black
Visible & UV
Quartz VIS/UV 25,000 - 9,000 white
Far-Infrared
Multilayer (far IR) 680 - 30 nickel-plated
Mylar 25m 120 - 20 nickel-plated
Mylar 50m 50 - 10 nickel-plated
Solid state 600 - 30 ** limited to a spectral resolution of 0.5cm-1
nickel-plated
Al ignm ent Tool
Glass For alignment purposes only! nickel-plated
8/11/2019 Vertex 70vs Manual
33/156
VERTEX 70v User Manual 29
.
.
.
.
.
O V E R V I E W
Optical Path
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
O P T I C A L P A T H
The beam path shown in figure 16ist the beam path of the standard spectrometer con-
figuration.
D1 Standard detector D2 Optional detector
BMS Beamsplitter
APT Aperture wheel
OPF Optical filter wheel
IN1 ... IN2 Beam inlet port 1 ... 2
OUT1 ... OUT5 Beam outlet port 1 ... 5
Figure 16: VERTEX 70v - Optical Path
8/11/2019 Vertex 70vs Manual
34/156
30 VERTEX 70v User Manual
4 O V E R V I E W
Optical Path
8/11/2019 Vertex 70vs Manual
35/156
VERTEX 70v User Manual 31
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OPERATION 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
G E N E R A L I N F O R M A T I O N
After the spectrometer has been installed and connected to the power supply, the PC,
and the vacuum pump, the spectrometer is ready for operation. VERTEX 70v is com-
pletely computer-controlled, i.e. operating the spectrometer (e.g. selecting the corre-
sponding optical components) performing a measurement and evacuating/venting the
spectrometer is done using the spectroscopic software OPUS.
This chapter describes mainly the spectrometer related aspects of the operation. For
detailed information about the OPUS software refer to the OPUS Reference Manual.
The OPUS manual Getting Started explains step by step how to perform the first mea-
surement after the spectrometer has been set up.
The standard spectrometer configuration is designed for measurements in the mid infra-
red region. Optionally, the spectral region can be expanded by substituting the installed
MIR components (source, detector, beamsplitter and sample compartment windows, if
available) for the corresponding optical components that allow measurements in the far
or near infrared as well as in the visible or ultraviolet region. (For information about the
replacement procedure of these optional components refer to the corresponding sec-
tions in this chapter and in chapter Maintenance and Repair.)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
S W IT C HI N G V E RT E X 7 0 V O N AN D OF F
Genera l Informat ion
The spectrometer is turned on and off using the mains switch at the spectrometer rear
side (figure 14). After having switched on the spectrometer, it starts booting. The boot
process takes about 30 seconds. As soon as this process is completed successfully, the
STATUS LED (figure 10) turns from red to green.After having switched on the spectrometer wait at least ten minutes before starting the
first measurement. This allows for the electronics and the light source to stabilize ther-
mally.
Caution: After having switched the spectrometer off, wait at least 30 sec-
onds before switching the spectrometer on again. This measure
avoids peaks in the initial current which could lead to fuse blowing
and/or damaging the power switch.
8/11/2019 Vertex 70vs Manual
36/156
32 VERTEX 70v User Manual
5 O P E R A T I O N
Switching VERTEX 70v On and Off
Switch-on Procedure
To put the spectrometer into operation again, proceed as follows:
1 Switch on the PC.
2 Switch on the spectrometer. The spectrometer begins to start up.
Note: After the spectrometer initialization is completed successfully, the
STATUS LED turns to green. Now the spectrometer is ready for
operation again.
3 Connect the the vacuum pump to the power supply.
Note: For information about how to operate the vacuum pump refer to the
supplied user manual of the vacuum pump manufacturer.
Switch-of f Procedure
Ideally, the spectrometer should uninterruptedly be kept under vacuum, even during
times of nonuse. If, however, the circumstances require a switching-off of the vacuum
pump and/ or the spectrometer the following procedure is recommended:
1 Evacuate the optical bench.
2 As soon as the final pressure is reached, switch off the spectrometer.
Note: The evacuation will take about 5 minutes. In the electroless
spectrometer state, all valves (for venting as well as for evacuatingthe spectrometer) are closed.
3 Disconnect the vacuum pump from the power supply.
In this state, the spectrometer interior is isolated from the laboratory environment and
the optical spectrometer components are protected against air humidity and they are no
longer current-carrying.
8/11/2019 Vertex 70vs Manual
37/156
VERTEX 70v User Manual 33
.
.
.
.
.
O P E R A T I O N
QuickLock
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Q U I C K L O C K
The sample compartment is equipped with a locking mechanism, called QuickLock, for
positioning and locking different measurement accessories. Therefore, you can use
only accessories that are mounted on a QuickLock baseplate. The QuickLock mecha-
nism enables a solid lock even for heavy and bulky accessories and allows a quick,
easy and reproducible positioning of the measurement accessories in the sample com-
partment.
When you insert and lock the accessory, all connections (purge gas connection and
electrical connection)are established and the accessory is automatically recognized by
the application software OPUS. This software feature is called AAR - Automatic Acces-
sory Recognition. In addition, the recommended measurement parameters are selected
automatically, provided that you have already stored the parameters for the accessory
in question. (See OPUS Reference Manual.)
The QuickLock mechanism also allows purging the sample compartment with dry air or
nitrogen gas. The purge gas enters the sample compartment via the gas diffusor(figure 18).
Purge gas
connection port
QuickLock
locking deviceElectronic
connectors
Figure 17: a) Sample Compartment - QuickLock Holder b) QuickLock Release Button
8/11/2019 Vertex 70vs Manual
38/156
34 VERTEX 70v User Manual
5 O P E R A T I O N
QuickLock
To inser t an accessory wi th QuickLock baseplate:
1 Hold the accessory with the QuickLockbaseplate front edge slightly tilted
upwards. Then, gently push the electrical connectors of the baseplate
against their counterpart of the QuickLock holder. Put the baseplate down.
Ensure that the baseplate is horizontally aligned to the QuickLock holder.
2 Gently press the front edge of the baseplate downward until it snaps into
place. To facilitate the insertion of the accessory, press the release button
outside the sample compartment. (See figure 17b.)
To remove an accessory wi th QuickLock baseplate:
1 Press the QuickLock release button outside the sample compartment. (See
figure 17b.)
2 While pressing the QuickLock release button, lift the front edge of the
QuickLock baseplate until the baseplate snaps free.
3 Carefully lift the accessory off the QuickLock holder to avoid damages to the
electrical connectors at the baseplate rear side.
Figure 18: Accessory with QuickLock Baseplate
Purge gas
diffusorElectronic connectors
for AAR and CAN bus
8/11/2019 Vertex 70vs Manual
39/156
VERTEX 70v User Manual 35
.
.
.
.
.
O P E R A T I O N
Automatic Accessory Recognition
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A U T O M A T I C A C C E S S O R Y R E C O G N I T I O N
As soon as an accessory is locked into the QuickLock holder, the OPUS/AAR software
(Automatic Accessory Recognition) starts and recognizes automatically the accessory
in question, provided you have activated the AAR function in the OPUS software. (For
information about how to activate the AAR function refer to the OPUS reference man-
ual, OPUS manual partAutomatic Accessory Recognition).
The OPUS/AAR software identifies the accessory, performs several tests, adapts the
measurement parameters and opens the Measurementdialog window to start a mea-
surement. If the automatic accessory recognition has been completed successfully,
OPUS displays a corresponding message.
Each time you start OPUS, the AAR program checks whether an accessory is installed
into the sample compartment. If AAR detects an accessory, the corresponding dialog
box is displayed. It also appears when the accessory is substituted by another one.
Note: When installing a new accessory for the first time, it is not yetregistered so that the OPUS/AAR software can not recognize it. In
this case, you first have to register the new accessory in question.
(See OPUS Reference Manual.)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
P E R F O R M I N G A M E A S U R E M E N T
The measurement procedure described in the following refers exclusively to measure-
ments under vacuum conditions. In case you want to perform a measurement not under
vacuum ignore the steps regarding evacuating and venting the spectrometer.
Specify the measurement parameters in the OPUS programme. To do this, selectin the OPUS Measuremenu theAdvanced Measurementfunction and select or
enter the corresponding parameter values. (The standard parameter values are
listed in appendix C.)
Evacuate the spectrometer as described in the following section. (Wait until theultimate vacuum is achieved.)
Acquire a background spectrum without the sample in the sample compartment byclicking in OPUS on the Background Single Channelbutton. (See figure 19.)
Vent the sample compartment as described in the following section.
Put the sample in the sample compartment. (For information about how to install aQuickLock accessory into the sample compartment refer to the section QuickLock
in this chapter. For information about sample preparation refer to appendix G.)
Evacuate the sample compartment again. (Wait until the ultimate vacuum isachieved.)
Acquire a sample spectrum by clicking in OPUS on the Sample Single Channelbutton (figure 19) and calculate the ratio (transmittance spectrum).
Note: Use the same parameter values for the background and the sample
measurement. Ensure that both measurements are performed under
identical ambient conditions.
8/11/2019 Vertex 70vs Manual
40/156
36 VERTEX 70v User Manual
5 O P E R A T I O N
Evacuating and Venting the Spectrometer
For detailed information about OPUS functions for data acquisition, manipulation and
evaluation refer to the OPUS Reference Manual.
E V A C U A T I N G A N D V E N T I N G T H E
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
S P E C T R O M E T E R
VERTEX 70v is primarily designed for vacuum operation, but it can be purged as well.To activate the vacuum mode, select in the OPUS Measuremenu the Optic Setup and
Servicefunction. Click on the Devices/Optionstab and make sure that the Purge Mode
check box is not ticked off. See figure 20.
Figure 19: OPUS Measurement Dialog Window
Figure 20: Activating the Vacuum Mode
With this checkboxbeing deactivated,
the vacuum mode is
activated.
8/11/2019 Vertex 70vs Manual
41/156
VERTEX 70v User Manual 37
.
.
.
.
.
O P E R A T I O N
Evacuating and Venting the Spectrometer
The flaps and the venting and evacuating valves are controlled automatically via the
OPUS software. So evacuating and venting the sample compartment and/or optical
bench is done using the OPUS software. The corresponding buttons are at the Basic
page of the Measurementdialog window. See figure 21.
Let us assume the following initial situation: both the sample compartment and the opti-
cal bench are vented. In this case, it is not possible to evacuate only the sample com-
partment. (The evacuation of only the sample compartment is not possible as in this
case the pressure difference between the sample compartment and the optical bench
would damage the flaps, i.e. the flaps are not designed for such an operation condition.)
So, clicking on either button effects the evacuation of both compartments. The evacua-
tion process is indicated by the message Sample / Optics Evacuatingthat appears in
the fields below the buttons. The progress of the evacuation is shown by the perma-
nently updated pressure readings in the lower fields. See figure 22.
Note: After you have clicked on a button, the labeling of this button
changes immediately showing the action that can be performed next
(i.e. Evacuate...turns to Vent...and versa vice).
Current state in the
individual compart-ments including the
current pressurereading
Command that can be
executed next byclicking on this button.
Figure 21: Optical bench and sample compartment are vented.
8/11/2019 Vertex 70vs Manual
42/156
38 VERTEX 70v User Manual
5 O P E R A T I O N
Evacuating and Venting the Spectrometer
As soon as the evacuation process is completed, the message Sample / Optics Evacu-
atedappears in the lower fields. See figure 23.
Note: If the sample compartment is evacuated you can not open it.
Note: To prevent OPUS from starting a measurement while the
spectrometer is being evacuated or vented proceed as follows: Click
in the Measurementdialog window on the Opt ictab and select in the
Opt ical bench readydrop-down list the option Pressur e stable. See
figure 24.
Figure 22: Optical bench and sample compartment are being evacuated.
Figure 23: Both compartments are evacuated.
8/11/2019 Vertex 70vs Manual
43/156
VERTEX 70v User Manual 39
.
.
.
.
.
O P E R A T I O N
Evacuating and Venting the Spectrometer
When both compartments are evacuated you can vent the sample compartment sepa-
rately (for example, if you want to open the sample compartment in order to exchange
the sample) by clicking on the Vent Samplebutton.
Note: When both compartments are evacuated, venting only the optical
bench is not possible as the pressure ratio inside the spectrometer
would damage the flaps. For safety reasons, the instrument does not
perform this operation. In this case, clicking on the Vent Op tics
button effects the ventilation of the sample compartment as well. This
precaution prevents the instrument from being operated wrongly.
Figure 24: Defining the Measurement Start Precondition
Figure 25: Sample compartment is vented and optical bench is evacuated.
8/11/2019 Vertex 70vs Manual
44/156
40 VERTEX 70v User Manual
5 O P E R A T I O N
Optimizing the Vacuum Operation of the Spectrometer
O P T I M I Z I N G T H E V A C U U M O P E R A T I O N O F
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
T H E S P E C T R O M E T E R
Genera l Informat ionTo get optimum measurement results under vacuum conditions, there are some aspects
that need to be taken into consideration:
The thermal conditions in an evacuated optics bench and in a purged optics bench
are completely different, i.e., under vacuum there is no thermal conduction at all
due to the lack of the purge gas. This aspect has consequences on the
reproducibility of the measurement results.
Water molecules are very polar. Due to this property, they tend to stick at the inner
wall of the optics compartment. For this reason, it takes time to get the water
vapor pumped off completely.
The purpose of the following advice is to help you in achieving optimum measurement
results.
Reproducibi l i ty of the Resul ts
After having evacuated the spectrometer, it is highly recommended that you allow the
spectrometer to stabilize long enough.An optimally stabilized spectrometer is able to
achieve an extreme high 100%-line stability in the sub-%-level with the standard optical
components designed for MIR measurements. (Note: A precondition is that the roomtemperature does not vary by more than 1C per hour and 2C per day. Typically, this
condition can be fulfilled in an air-conditioned environment.)
R e c o mme n d a t io n s :
For demanding experiments, a stabilization period of at least 4 hours is
recommended. After this period, the maximum instrument stability is achieved.
For non demanding experiments, a stabilization time of 0.5 hour is sufficient.
During a long-term experiment, it is recommended to repeat the background
measurement in regular interval, at least every hour. Ideally, the spectrometer should be kept under vacuum overnight.
8/11/2019 Vertex 70vs Manual
45/156
VERTEX 70v User Manual 41
.
.
.
.
.
O P E R A T I O N
Optimizing the Vacuum Operation of the Spectrometer
Residual Water Vapor
Longer evacuation times will further reduce the residual water vapor concentration
inside the spectrometer:
Note: Besides the necessity of a water vapor concentration being as low aspossible, there is another aspect regarding water vapor you have to
take into consideration: The water vapor line intensity in the sample
spectrum does not depend on the absolute residual water vapor
concentration in the spectrometer but on the different water vapor
concentrations during the background and the sample measurement.
Therefore, it is of crucial importance that the residual water vapor
concentration is (nearly) identical during both the background
measurement and the sample measurement.
E v a c u a t io n T ime
As mentioned above, water molecules are very polar. Due to this property, they tend to
stick at the inner wall of the optics compartment, even under vacuum. For this reason, a
long evacuation time is recommended. Ideally, the evacuation of the spectrometer
should not be interrupted overnight. This action will further reduce the residual water
vapor content.
E v a c u a t io n P r o c e d u r e
Before acquiring a background spectrum, simulate a sample exchange in the same way
as you will do it later for the real sample measurement:
1 Vent the sample compartment.
2 Afterwards, evacuate the sample compartment for about 5 to 10 minutes.
(An evacuation time longer than 10 minutes is not necessary because after
that period, the final pressure of < 0.2hPa (< 0.2mbar) will be achieved.)
Note: As soon as the pressure falls below < 1hPa, the message Sample
Evacuated, including the current pressure value, is displayed in the
Measuredialog window (figure 26). The achievement of the final
pressure is also indicated by the VACUUM LED at the spectrometer
top side, i.e. this LED lights green.
8/11/2019 Vertex 70vs Manual
46/156
42 VERTEX 70v User Manual
5 O P E R A T I O N
Optimizing the Vacuum Operation of the Spectrometer
Important Note:The evacuation times before the background measurement
and before the sample measurement have to be more or less
identical. To ensure reproducible evacuation times, specifyin OPUS a Delay before Measurement. See the figure 27.
3 Acquire a single channel background spectrum.
4 Afterwards, vent the sample compartment and place the sample in the
sample compartment.
5 Evacuate sample compartment for about 5 to 10 minutes.
6 Acquire a single channel sample spectrum.
Figure 26: OPUS dialog window Measurement- page Basic
Current state inside the
individual compartments,
including the current pres-
sure reading
8/11/2019 Vertex 70vs Manual
47/156
VERTEX 70v User Manual 43
.
.
.
.
.
O P E R A T I O N
Purging the Spectrometer
Note: Take into account that the intensity of the water vapor band in the
sample spectrum does not depend on the absolute residual water
vapor concentration but results from a water vapor concentration
difference during the background and the sample measurement.
With the above described operation conditions and a spectral resolution of 4cm-1, typi-
cally a residual water vapor band intensity in the range of significantly less than 0.1%T
can be achieved.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
P U R G I N G T H E S P E C T R O M E T E R
Genera l Informat ion
Purging the spectrometer is not necessarily required, especially when you perform
measurements under vacuum conditions. However, if the spectrometer is not evacu-
ated, purging is recommended, especially when you frequently open the compartmentcovers (e.g. due to a detector or beamsplitter replacement or a sample substitution) or if
the ambient air humidity content is too high because this measure reduces the level of
water vapor, CO2or other components of the ambient air inside the spectrometer.
Note: Water vapor, CO2and other atmospheric contaminants cause
unwanted absorption. Therefore, open the sample compartment, the
detector compartment and/or the interferometer compartment only if
necessary in order to prevent water vapor, CO2 or other contaminants
from entering the above mentioned compartments.
Figure 27: OPUS Measurement dialog
Specifying themeasurement
delay time
8/11/2019 Vertex 70vs Manual
48/156
44 VERTEX 70v User Manual
5 O P E R A T I O N
Purging the Spectrometer
Purge the spectrometer, for example, with dry air or low pressure nitrogen gas. Provide
the following purge gas conditions:
Dry (dew point < -40C corresponds to a degree of dryness of 128ppm humidity),
oil-free and dust-free air or nitrogen gas
Maximum pressure of 2 bar (29 psi) Initial purge gas flow rate should not exceed 500 liters/hour
Sustained purge gas flow rate should not exceed 200 liters/hour
Danger: Do not use flammable gases for purging the spectrometer. Some
spectrometer components become very hot during operation. If
flammable gases come in contact with hot components there will
be the risk of fire and/or explosion!
For information about how to connect the spectrometer to a purge gas supply line, refer
to chapter Installation.
If you want to purge an enclosed accessory (e.g. micro ATR unit) you have to cover the
opening, which is also intended for evacuating and venting the sample compartment
(see figure 28), using the supplied cap in order to ensure a sufficient purge of the
accessory. Put the cap over opening and screw on the cap.
Attention: If you want to perform measurements under vacuum conditions
do not forget to remove this cap again! Otherwise, the evacuation
of sample compartment via the small purge gas inlet in the Quick-
Lock clamping device (figure 28) will take too long causing a red
VACUUM LED after a certain period of time.
This opening is intended for evacu-
ating, venting and purging the sam-ple compartment.
Purge gas inlet for purging an enclosed accessory
mounted on a QuickLock baseplate
Figure 28: Sample Compartment
Cap with screw thread
8/11/2019 Vertex 70vs Manual
49/156
VERTEX 70v User Manual 45
.
.
.
.
.
O P E R A T I O N
Purging the Spectrometer
Control l ing the F laps
VERTEX 70v is primarily designed for vacuum operation, but it can be purged as well.
To activate the purge mode, select in the OPUS Measuremenu the Optic Setup and
Servicefunction. Click on the Devices/Optionstab and make sure that the Purge Mode
check box is activated. See figure 29.
This operating mode allows you to control (open and close) the flaps in order to purge
either the sample compartment or the optical bench or both. The flaps are controlled viathe OPUS software. The corresponding buttons are at the Basicpage of the Measure-
mentdialog window. See the following figure.
Figure 29: Activating the Purge Mode
Purge mode is
activated.
Current state of the
flaps
Next possible action
that can be performedby clicking on this
button.
Figure 30: Controlling the Flaps
8/11/2019 Vertex 70vs Manual
50/156
46 VERTEX 70v User Manual
5 O P E R A T I O N
Exchanging the Beamsplitter
The flaps can be opened or closed only if the pressure difference between the sample
compartment and the optical bench is below the threshold value of 5 hPa.
Specia l Case
Besides the normal purge mode in which the optical bench and/or the sample compart-
ment are only purged, the following special case is also possible: the vented sample
compartment is purged while the optical bench is evacuated. For the realization of this
special case, the spectrometer needs to be equipped with windows mounted on either
the sample compartment walls or the flaps which are closed in this case. To realize this
special case, proceed as follows:
Make sure that the purge mode is deactivated in OPUS. (See figure 20.)
Evacuate the optical bench and the sample compartment. (See chapter Opera-
tion, section Evacuating and Venting the Spectrometer.)
Afterwards, vent the sample compartment again. (In this condition, the flaps are
closed.)
Connect a hose to the purge gas inlet for the sample compartment. (The purge
gas inlet is at the spectrometer rear side. See chapter Installation, section Con-
necting VERTEX 70v to the Purge Gas Line, figure 7.)
Now start the purge gas supply.
Note: The flaps isolate the sample compartment hermetically from the
optical bench.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E X C H A N G I N G T H E B E A M S P L I T T E R
Genera l Informat ion
The standard version and the optional version of VERTEX 70v differ from each other
with regard to the interferometer compartment cover design (see fig. 33a and 33b) and
the availability of the beamsplitter storage positions (see fig. 31) inside the interferome-
ter compartment.
Feature Standard Version Optional Version (S239/V)
Interferometer compartment
cover design
A beamsplitter exchange
requires the removal of the
complete interferometer com-
partment cover.
Wing-shaped cover provides
for easy access to the beam-
splitter.
Availability of the beamsplitter
storage positions inside the
interferometer compartment
No Yes
8/11/2019 Vertex 70vs Manual
51/156
VERTEX 70v User Manual 47
.
.
.
.
.
O P E R A T I O N
Exchanging the Beamsplitter
The standard spectrometer version is equipped with a MIR beamsplitter (made of KBr).
If your measurement requires a different spectral range, you can exchange the beam-
splitter manually. For a list with all available beamsplitters (including their spectral
ranges) refer to chapter Overview, section Internal Components.
Note: Make sure that the spectral ranges of the installed optical
components (source, beamsplitter, detector and samplecompartment windows, if installed) correspond with each other!
Changing the beamsplitter is easy because a precise locking mechanism automatically
fixes the beamsplitter at its pre-aligned position, as soon as you move the release lever
(figure 34)into the locked position (i.e. backward).
All beamsplitters for VERTEX 70v are electronically coded enabling the spectrometer
firmwareto auto-detect the beamsplitter type. The information about the component is
passed on to the OPUS software. This feature is called ACR (Automatic Component
Recognition)1. Its purpose is to prevent you from selecting a wrong component in
OPUS when you set up a measurement. (Note: A wrongly selected component is indi-cated in OPUS by a red colored entry field of the corresponding drop-down list. See
also the OPUS Reference Manual.)
1. ACR is restricted only to the optical components beamsplitter, detector and source.
Figure 31: VERTEX 70v - Beamsplitter Installation Positions (only In case of option S239/v)
Beamsplitter(Operating position)
Two additional Beamsplitters
(Storage position)
Knob
Wing-shaped
cover
8/11/2019 Vertex 70vs Manual
52/156
48 VERTEX 70v User Manual
5 O P E R A T I O N
Exchanging the Beamsplitter
Handl ing Instruct ions
Caution: The beamsplitter is a very delicate component. Handle it with
utmost care and observe the following handling instructions to
ensure a long service life.
Do not touch the beamsplitter surface as this will damage the surface and, as a
consequence, the beamsplitter may become useless. Hold the beamsplitter using
always the handle (figure 32).
Some beamsplitter materials are hygroscopic. Never expose them to humidity or
water vapor. Store the beamsplitter either in a dry and sealed container (e.g. in the
beamsplitter storage box) or inside the spectrometer (storage position, figure 31).
Do not try to loosen or fasten the screws as this will impair the optical quality of the
beamsplitter and lead to malfunctions.
Do not try to clean the beamsplitter. Do not expose the beamsplitter (especially beamsplitters made of KBr) to temper-
ature changes.
Figure 32: Beamsplitter
Handle
8/11/2019 Vertex 70vs Manual
53/156
VERTEX 70v User Manual 49
.
.
.
.
.
O P E R A T I O N
Exchanging the Beamsplitter
Exchange Procedure
1 The procedure for gaining access to the beamsplitter depends on the
interferometer compartment cover design. Standard cover(fig. 33a): Take
off the complete cover. Optional cover (fig. 33b): Turn the knob counter-
clockwise until the stop and rotate the wing-shaped cover aside as shown in
figure 31.
Caution: Class 2 laser radiation. When the cover is removed do not stare
into the laser beam.
2 Move the release leverinto the unlocked position (i.e. backward).
3 Carefully pull the beamsplitter straight upwards without catching an edge.
4 Take the other beamsplitter out of the storage position and insert it with the
electrical contacts facing to the front side (figure 34).
5 Push down the beamsplitter completely until you feel resistance.
6 Move the release lever into the locked position, i.e. forward. (See figure 34.)
Note: A beep indicates that the beamsplitter has been recognized by the
electronics. After a few seconds the spectrometer will start scanning.
7 Insert the beamsplitter, you have taken out of the operating position, either
into the storage position holder (figure 31) or store it in the intended box.
8 Standard cover: Place the cover on the interferometer compartment again.
Optional cover: Rotate the wing-shaped cover over the openings and
secure it by turning the knob clockwise.
9 Check whether a signal is detected and the optics works correctly. (For
detailed information refer to the OPUS reference manual.)
Figure 33: a) Standard Cover b) Wing-shaped Cover (in case of option S239/V)
8/11/2019 Vertex 70vs Manual
54/156
50 VERTEX 70v User Manual
5 O P E R A T I O N
Exchanging the Detector
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
E X C H A N G I N G T H E D E T E C T O R
Genera l Informat ion
The basic spectrometer configuration is equipped with a DigiTect DLaTGS detector. Ifyour measurement requires a different spectral range or another detector sensitivity you
can install another detector into the second detector position, which is an optional spec-
trometer feature, or exchange the installed DigiTect DLaTGS detector for another Digi-
Tect detector, such as a MCT with a higher sensitivity or a NIR detector. (For the list
with all available detectors including their spectral ranges refer to chapter Overview,
section Internal Components.)
Note: Make sure that the spectral ranges of the installed optical
components (source, beamsplitter, detector and sample
compartment windows, if installed) correspond with each other!
A removable cover provides access to the detector compartment. The dovetail detectormounting facilitates the exchange. A re-alignment is not necessary. All detectors for
VERTEX 70v are electronically coded, enabling the spectrometer firmware to auto-
detect the type of detector currently installed. This information is passed on to the
OPUS software. This feature is called ACR (Automatic Component Recognition)1. Its
purpose is to prevent you from selecting a wrong component in OPUS when you set up
a measurement. (Note: A wrongly selected component is indicated in OPUS by a red
colored entry field of the corresponding drop-down list. See also the OPUS Reference
Manual.)
1. ACR is restricted only to the optical components beamsplitter, detector and source.
Figure 34: Beamsplitter (Operating Position)
Release lever(in locked position)
Beamsplitter handle
8/11/2019 Vertex 70vs Manual
55/156
VERTEX 70v User Manual 51
.
.
.
.
.
O P E R A T I O N
Exchanging the Detector
Exchange Procedure
1 Take off the detector compartment cover (E in figure 9).
Caution: If there is a MCT detector installed in the detector compartment
(including a vacuum-tight closure at the filling hole in the detector
compartment cover) the detector compartment cover can not be
taken off. In this case, do not try to remove the cover forcibly as
this may cause a spectrometer damage! Therefore, first screw off
the sealing adapter mating part (figure 39) before you take off the
detector compartment cover. (See chapter Operat ion, section
Cool in g an MCT Detector, subsection Detector Com partment
Cover Preparat ion Procedure.)
2 Loosen the locking screw (allen screw) that secures the detector using a
hex key (size 6mm). See figure 35. Depending on which detector you wantto remove, the allen screw is on the left or right side of the detector.
3 Pull the detector straight upwards out of the dovetail guide.
Caution: Remove the detector carefully in order not to damage the detector
and/or the mirrors.
4 Insert the other detector precisely into the dovetail guide and push the
detector downwards until you feel a resistance.
Figure 35: Detector Compartment
MCT detector
DLaTGS
detector
Moveable mirror
Locking screw
of the detector
installed inposition 1
Locking screw of
the detector in-stalled in position 2
Fixed mirror
8/11/2019 Vertex 70vs Manual
56/156
52 VERTEX 70v User Manual
5 O P E R A T I O N
Cooling an MCT Detector
Note: A beep indicates that the detector has been recognized by the
electronics. The electrical connections are established automatically.
5 Fasten the allen screw slightly using a hex key(size 6mm).
6 Place the cover on the detector compartment. Make sure that the four
plastic pins in the corners at the bottom side of the detector compartment
cover engage into the corresponding hole of the spectrometer case.
Note: If there is a MCT detector in the detector compartment do not forget
to reinstall the vacuum-tight closure at the filling hole in the detector
compartment cover. For information about it refer to chapter
Operat ion, section Cool in g an MCT Detector, subsection Detector
Comp artment Cover Preparat ion Procedure.)
7 Check whether a signal is detected and the optics works correctly (For
detailed information refer to the OPUS reference manual).
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C O O L I N G A N M C T D E T E C T O R
Genera l Informat ion
To ensure operating ability, MCT detectors have to be cooled regularly with liquid nitro-
gen. The typical hold time depends on the detector. There are MCT-detectors with a
hold time of 8, 12 or 24 hours.
To fill the detector with liquid nitrogen you need neither to remove the detector from the
spectrometer nor even open the detector compartment. The supplied funnel facilitates
the filling in of the liquid nitrogen in the detector. See figure 40.
Detector Compartment Cover Preparat ion Procedure
In case the MCT detector has been delivered together with VERTEX 70v, the detector
compartment cover is already prepared for the funnel insertion. If you have ordered the
MCT detector at a later date you need to prepare the cover as described in the follow-ing.
In accordance with the number of detectors that can be installed in the VERTEX 70v
detector comp