A Multi-Wavelength Interferometry System for the Orion Laser Facility S.Patankar 1(a), , E.T.Gumbrell 2 ,T.S.Robinson 1 ,H.F.Lowe 1 , S.Giltrap 1 , C.J.Price 1 , N.H.Stuart 1 , P.Kemshall 2 ,J.Fyrth 2 ,J. Luis 2 , J. W.Skidmore 2 , R.A.Smith 1 1 Blackett Laboratory, Imperial College, London, SW7 2AZ, UK 2 AWE, Aldermaston, Reading, Berkshire, RG7 4PR, UK We report on the design and testing of a multi-wavelength interferometry system for the Orion laser facility based upon the use of self-path matching Wollaston prisms. The use of UV corrected achromatic optics allows for both easy alignment with an eye safe light source and small (~ mm)offsets to the focal lengths between different operational wavelengths. Inteferograms are demonstrated at wavelengths corresponding to first, second and fourth harmonics of a 1054nm Nd:glass probe beam. Example data confirms the broadband achromatic capability of the imaging system with operation from the UV (263nm) to visible (527nm) and demonstrates that features as small as 5µm can be resolved for object sizes of 15mm by 10mm. Results are also shown for an off harmonic wavelength that will underpin a future capability. The primary optics package is accommodated inside the footprint of a Ten Inch Manipulator (TIM) to allow the system to be deployed from a multitude of viewing angles inside the 4m diameter Orion target chamber. I. INTRODUCTION The recently commissioned Orion 1 laser facility at the UK Atomic Weapons Establishment (AWE), uses ten ~kJ class long pulse lasers and two independent sub-picosecond petawatt class short pulse lasers to underpin a wide range of plasma physics experiments. A comprehensive array of diagnostic capabilities 2,3 exist to cover the full spectrum of radiation and particles emitted from intense laser matter interactions ranging from Hohlraums to shocks. The optical diagnostics suite consisting of a multi-channel pyrometry system, a Velocity Interferometer System for Any Reflector (VISAR), Gated Optical Imagers (GOI) and short pulse interferometry provides key experimental tools for studying plasma properties. This paper describes the optical setup and testing of a robust and flexible (a) Author to whom correspondences should be address. Electronic Mail - [email protected]
Transcript
Page 1: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

A Multi-Wavelength Interferometry System for the Orion Laser Facility

S.Patankar1(a),, E.T.Gumbrell2,T.S.Robinson1,H.F.Lowe1, S.Giltrap1, C.J.Price1, N.H.Stuart1, P.Kemshall2,J.Fyrth2,J. Luis2, J. W.Skidmore2, R.A.Smith1

1Blackett Laboratory, Imperial College, London, SW7 2AZ, UK

2AWE, Aldermaston, Reading, Berkshire, RG7 4PR, UK

We report on the design and testing of a multi-wavelength interferometry system for the Orion laser facility based upon the use of self-path matching Wollaston prisms. The use of UV corrected achromatic optics allows for both easy alignment with an eye safe light source and small (~ mm)offsets to the focal lengths between different operational wavelengths. Inteferograms are demonstrated at wavelengths corresponding to first, second and fourth harmonics of a 1054nm Nd:glass probe beam. Example data confirms the broadband achromatic capability of the imaging system with operation from the UV (263nm) to visible (527nm) and demonstrates that features as small as 5µm can be resolved for object sizes of 15mm by 10mm. Results are also shown for an off harmonic wavelength that will underpin a future capability. The primary optics package is accommodated inside the footprint of a Ten Inch Manipulator (TIM) to allow the system to be deployed from a multitude of viewing angles inside the 4m diameter Orion target chamber.

I. INTRODUCTION

The recently commissioned Orion1 laser facility at the UK Atomic Weapons Establishment (AWE), uses ten ~kJ class long pulse lasers and two independent sub-picosecond petawatt class short pulse lasers to underpin a wide range of plasma physics experiments. A comprehensive array of diagnostic capabilities2,3 exist to cover the full spectrum of radiation and particles emitted from intense laser matter interactions ranging from Hohlraums to shocks. The optical diagnostics suite consisting of a multi-channel pyrometry system, a Velocity Interferometer System for Any Reflector (VISAR), Gated Optical Imagers (GOI) and short pulse interferometry provides key experimental tools for studying plasma properties. This paper describes the optical setup and testing of a robust and flexible interferometry system devised for use at the Orion laser facility to probe two dimensionalplasma electron density profiles.

Optical interferometry using short pulse lasers is a commonly applied technique for diagnosing plasma electron densities in under dense conditions4. Alaser pulse passing through the plasma acquires a phase shift which is proportional to the density-length product of the free electron density. When compared to a reference fringe pattern for light which has not traversed the plasma, it is possible to extract total imparted optical phase and consequently determine a 2D projection of the electron density-length product. The use of short pulse lasers with sub-picosecond pulse duration also offers excellent temporal resolution and extends this technique to the study of rapidly evolving hydrodynamic motion. For a typical 500fs pulse from a Nd:glass laser, the motion blur from hydrodynamics at 1000km/s would only be 0.5µm which is typically less the optical resolution of the imaging system.Two optical probe beams on Orionare specified to deliver 400fs 263nm (4ω) pulses with a nominal energy of 50mJ. Although the system we describe here is designed specifically for probing at the fourth harmonic, its capability to operate effectively at different wavelengths and configurations is demonstrated. Additional probing wavelengths are not currently delivered to target but they are available in the target chamber for the 4th harmonic generation process and can be used

(a) Author to whom correspondences should be address. Electronic Mail - [email protected]

Page 2: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

Focal region saturated

Front surface expansion

150 µmLaser

Fused silicaslide

for active imaging in the future. Whilst shorter wavelengths are able to probe denser plasmas (the critical electron density Ncr= 1.6 × 1022 cm-3 at 263nm), longer wavelengths offer greater sensitivity in terms of fringe displacement. Interactions of ultrashort laser pulses with solid targets can produce strong harmonic self-emission from the generated plasma. Experimental evidence from AWE's HELEN5 laser system for example has indicated that the generated emission results in a weaker self-emission background when probing at the fourth harmonic. Figure 1 shows a 4ω backlit optical image of a fused silica plate irradiated at ~1020 Wcm-2 from the HELEN shortpulse system and probed transversely 6.6ps later. It is clear that only the region immediately in the vicinity of the laser focus is saturated with self-emission and important early time plasma dynamics such as front surface expansion is clearly visible. This is in stark contrast to the second harmonic image which was completely saturated by self-emission.

FIG 1:Fourth harmonic transverse shadowgraphy of a short pulse laser irradiation of a fused silica plate at 1020 W cm-2 . Whilst self-emission in the fourth harmonic band saturates the imaging systemclose to the laser focus, early time plasma dynamics such as the expansion of the front surface of the target is clearly visible.A reduction in background light from plasma self-emission is a key advantage of probing with the fourth harmonic.

A range of optical schemes have been realised for plasma interferometry including Mach-Zehnder, Michelson and Normarski configurations4. All methodologies rely on interference between light pulses which have traversed the plasma and those which have not. This is achieved either by splitting the probe pulse and recombining it with a reference copy, or by overlapping spatial regions in the probe beam which have travelled the plasma with those which have not.

The contrast of the interferogram (the ratio of the intensities of the light and dark zones of the fringe pattern) is dependent upon the relative intensity of the two beams being interfered and the spatial and temporal coherence of the two pulses of light. The latter places rigid constraints on the alignment of optics, particularly for use with short duration probe pulses. For example, the expected ~400fs probe pulse on Orion would require path length matching to better than 20µm should the pulse be split and then recombined.

Given the operational and engineering constraints on Orion (e.g. a low shot rate, potentially high radiation Class 4 laser environment with carefully controlled access), a robust and flexible optical system based on the use of Wollaston prisms was devised to accommodate a range of laser probing scenarios. Although interferometry is the primary purpose of the system, shadowgraphy4 and Schlieren imaging4 modes are also possible to implement with minor modifications to the optical package. Polarised light interferometry has been demonstrated on laser experiments previously 6 and offers unique advantages from an operational perspective. The lack of a reference beam removes the

Page 3: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

OP L1WP CP

IP1 L2 L3 IP2

UV

F2F3Z

F1

L

need for complex in-vacuo engineering along with the condition to have it path matched to inteferometric accuracy with the probe. This allows the optical system to be stable and compact enough to fit inside a TIM.

II. DESCRIPTION OF OPTICAL SETUP

The optical configuration for the interferometry package is based around the use of UV corrected triplet lenses which greatly simplifies the alignment procedure for the fourth harmonic probing system. UV pulses are only available at Orion on a single shot basis, and as a result, pre-shot alignment is performed using an off wavelengthlow power CW 532nm Nd:YAG laser. The need for additional correcting optics to account for the large focal length offsets between UV and visible is avoided by using custom manufactured UVachromats which require very minor adjustments across a broad wavelength range. Additionally, any wavelength from the UV to NIR can be imaged with the appropriate minor correction to lens positionswhich removes the need to rebuild the system with different optics when radically changing the mode of operation.

FIG2: A schematic of the optical configuration of the Wollaston prism based interferometry system. OP - Object plane; L1 - Lens 1; WP - Wollaston Prism; CP - Cube Polariser; L2 - Lens 2; L3 - Lens 3; U - Object distance; V - Image distance; L - Distance from the Wollaston to IP1; F2 - Focal length of lens 2 ;F3 - focal length of lens 3; IP1 & IP2 - Image planes. Blue lines show image rays forming a double image at IP1 and IP2.

The layout of the optical system is detailed in figure 2. The object plane is imaged using a 180mmfocal length F/8UV triplet (Edmund optics 64-840) which is located at a distance of 216mm (U in the figure 2) from the object plane. An image of the object plane is formed at a distance of 1080mm (V) behind L1, which gives a linear magnification of 5. For typical laser interaction experiments, the target is relatively "small" (few mm scale) which demands the use of high magnification optics to resolve key physical processes. It is also possible to change the magnification by changing the focal lengths of the telescope between L2 and L3 which relay the image outside the TIM vacuum chamber. The Wollaston prism is orientated at 45° to the incoming polarisation (vertical) such that two orthogonally polarised beams of equal intensity are generated. The cube polariser (CP) is orientated at 45° to the orthogonally polarised beams to ensure the transmitted beams have the correct polarisation state for optimal fringe contrast. This optic is wavelength dependent and will have to be changed depending on the wavelength configuration. The two orthogonally polarised beams diverge at an angle ε, which is set by the geometry of the prism (and is a fixed quantity for a specific prism set). At the first image plane (IP1), the centres of the two images are spatially separated by an amount given by the relation εZ, where Z is the distance between the centre of the Wollaston and the first image plane (IP1).The fringe spacing is given by the relation5 (λ/ε)/(Z/L). Fringes are visible in the region where the two images overlap. It is crucial that the overlapped region contains

Page 4: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

L2

CP

WP

L1

TIM Payload frame

one image of the plasma and one "empty" region to allow unambiguous interpretation of the data. This necessitates the use of Wollaston prisms with sufficient divergence to separate the images adequately, and this will be influenced by the precise detail of a given experimental configuration, e.g. target size and region of interest. Prisms with divergence angles of 0.5 o and 2owere tested and shown to suffice for even quite large laser targets up to scales of 15 mm. Fringe orientation can be chosen arbitrarily if a half wave plate is used to adjust the probe beam polarisation before passing through the target.Fringe spacing can be adjusted by changing the position of the Wollaston prism (effectively adjusting the ratio of Z/L); however, this is limited to a range where the two images are still sufficiently separated such that unambiguous data can be obtained. For a given image separation, fringe spacing can also be changed by switching to a different probing wavelength, however, the advantages of shorter wavelength probing (e.g. less harmonic selfemission) would potentially be lost.

Whilst the magnified image at position IP1 can be used to accommodate a detector, it lies only 1280mm from OrionTarget Chamber Centre (TCC) which means that any active camera would require vacuum compatibility (and potentially EMP and radiation hardness) to operate inside the chamber, which would add significantly to the complexity of the system.An additional pair of achromatic relay lenses (L2 & L3) allows the detector along with UV and optical filters to be placedon a small platform outside the TIM vacuum chamber and also allows the magnification to be altered. The lens L2 (200mm F/4 B.Halle7 UV triplet) is positioned at its focal length from IP1 in order to project the image rays to infinity. These rays are then collected and refocused using lens L3 (300mm F/5 B.Halle UV triplet) placed at its focal length from the front surface of a large format CCD detector. The combination of L2 and L3 provides an additional magnification of 1.5 from IP1 which increases the total image magnification to 7.5. It is also possible to change magnification externally to 2.5 by changing L3 to a 100mm F/2 UV quintuplet lens6. This offers additional flexibility depending on target size and required field of view. It should also be noted that the UV achromats do not have anti-reflection coatings at 263nm, however, due to the limited temporal width of the probe pulses, multiple reflections were not found to cause any unwanted secondary fringes during testing.Figure 3 shows a 3D representation of the TIM based optics package with the interferometry optics placed in a line along the system axis.

In addition to interferometry, the optical package can easily be modified for Schlieren imaging by placing a stop at the probe ray focus of L1 and removing the Wollaston prism. Shadowgraphy is also

FIG.3: 3D layout of the TIM based optical probing payload frame along with the key optics which are labelled as per figure 1. The first lens also hosts a removable debris shield and a ring light to aid focusing.

Page 5: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

possible to implement by removing both, the Wollaston prism and Schlieren stop. Modified commercial digital single lens reflex (DSLR) cameras are used as time integrated detectors for the probe beam light.High temporal resolution is provided by the limited duration of the probe pulse and its greater intensity compared to self-emission. A Canon 350D DSLR with a 24mm by 16mm chip size and 6.4µm pixel size was used for detecting the UV light. A large Nikon D810 detector with a 36mm by 24mm chip size and 4.8µm pixel size is also available for use in the visible and IR. If self-emission from the initiallaser-target interaction is relatively high, then a time gated imaging system (GOI) can be used as the detector, albeit with significantly lower spatial resolution and dynamic range.

The availability of two independent probe beams on Orion along with the flexibility of multiple wavelengths and imaging configurations offers a compelling capability for a wide range of experiments. Advanced probing scenarios such as two colour inteferometry8 and Faraday rotation9imaging of magnetic fields are key future considerations included in the design of this system.

III. Active System Testing

Optical testing and metrology of the complete interferometry and imaging system was carried out at Imperial College London due to the limited availability of access toOrion. An OPCPA Nd:Glass hybrid laser system10 with equivalent specifications to the Orion probe beam was used for interferometry and a 1 TW (500mJ, 500fs, 1054nm) drive laser was used to create test laser plasmas that acted as a short lived, rapidly evolving phase object. The probe beams were generated using two cascaded KDP SHG crystals to provide co-propagating 1ω, 2ω and 4ω pulses which were separated using appropriate filters and dichroic mirrors. Although the plasma conditions probed were not comparable with those created by the very high energy drive lasers at Orion, the aim was to verify the fidelity and flexibility of the probing system under a range of dynamic scenarios.

Test interferograms obtained using the system described above along with the 0.5oWollaston prism are shown below in figure 3. The images have been cropped to the centre of the field of view. Lineouts showing the fringe contrast are also attached for different test wavelengths. Table I shows the maximum fringe contrast ratio obtained at each wavelength. The contrast obtained at 1ω and 2ω are very good with very clear transitions from bright to dark zones, the 4ω fringes are a factor of two lower in contrast as difficulties were encountered in driving the nonlinear conversion process used to create the DUV probe light into saturation. This resulted in a modulated spatial profile which lead to unequal intensities in the two arms of the interferometer system and consequently, reduced fringe contrast. This issue was isolated solely to the quality of the available probe beam and is not a fundamental limitation of the optical components. Fringe patterns are shown in three typically used configurations although it should be noted that the orientation can be changed arbitrarily by using half wave plates.

Page 6: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

FIG 3. Typical interferograms obtained at centre of the image for 1ω (a), 2ω (b) and 4ω (c) probe pulses. Corresponding lineouts are taken in the direction indicated by the dashed white line. Excellent fringe contrast is obtained at the first and second harmonic owing to better beam quality. We noticed internal defects inside UV interference filters which resulted in a localised reduction in fringe contrast. Difficulties were also encountered in obtaining a smooth UV beam profile which resulted in non-equal intensities in the interferometer arms. These issues are not attributed as an optical fault with the imaging system.

Table I: Measured fringe contrast at image centre on a typical interferogram for the first, second and fourth harmonic probe beams.

Wavelength Peak Fringe Contrast1054nm 58.9%527nm 71.8%263nm 35%

A. Experimental Examples

Experimental validation was performed at both the second and fourth harmonic using two different scale length targets with the low (0.5°) and high(2°) divergence Wollaston prisms. Detailed interpretations of the experiments are notthe subject of this paper and will be addressed in a separate publication. The probing direction was set orthogonal to the drive beam. For the UV probing test, a

Page 7: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

A B

9 µm

DriveLaser

laser interaction with a small scale solid density target was performed.A 50µm diameter silicon oil droplet11 attached to 9µm diameter Carbon fibre was irradiated with 150GW of power using F/20 focusing. Figure 4a shows the pre-shot UV interferogram of the micro-droplet attached to a stalk. Given the small scale length of the target, only a modest image separation is required to isolate the two images while ensuring a uniform field of fringes. The imaging configuration was setup as per figure 2 with the 0.5° Wollaston placed at a distance of 850mm from IP1 (L = 50mm). This gives an image separation of ~ 7.4mm which even at a magnification of 5 gives enough isolation (IS ~ 1.3mm) between the two images. The resultant fringe spacing is ~ 2µm which is well resolved by the detector (Canon 350D). It should be noted that the relatively poor fringe contrast seen in the images was due to inhomogenities in the interference filter and non-uniformities in the spatial profile over the field of view. Figure 4b shows the deflection in the fringe pattern 3ns after the drive laser hitting the oil droplet. It should be noted that the UV images are not exactly in focus as adjustments were not made from the 532nm alignment laser. This highlights the achromatic capability of the imaging system as useable data can be obtained without a UV alignment beam.

FIG4: 263nm short pulse interferometry of an oil droplet attached to a 9µm diameter carbon wire. A) Background interferogram showing a 50µm oil droplet attached to the wire. B) Fringe pattern deflection observed 3ns after irradiation with a 100mJ 650fs laser pulse. The relatively low fringe contrast is due to difficulties encountered in uniform UV illumination and not an optical defect. The bright spot in the corner of the second image is scattered light from the drive laser.

Page 8: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

Low density gaseous targets are transparent at optical wavelengths and therefore ideally suited for testing a number of probing scenarios. Noble gas clusters from pulsed gas jets have been shown to be very efficient at absorbing short pulse laser light12 and make interesting targets for experiments in areas including laboratory astrophysics13 and x-ray generation14.To verify the imaging system capabilities at the second harmonic wavelength, argon gas clusters from a large aperture (12mm) gas nozzle were irradiated with 1 TW pulses using F/20 focusing. The measured average atomic density of the cluster gas produced by this system is ~ 1019 cm-3and the measured mean cluster size is ~5nm. The interaction was imaged perpendicular to the drive laser using simultaneous interferometry and Schlieren imaging.Given the size of the gas jet, the imaging system was modified from the configuration shown in figure 2 by changing L1 to 300mm F/5 UV triplet and the Wollaston divergence to 2°. The magnification was lowered to 2 and a larger area detector (Nikon D810) was used to increase the field of view for the interferometry. The additional image relay shown in the design was also not used as it is specific for TIM compatibility and consequently images were taken at the IP1 position. A 4" beamsplitter was used after L1 to split the probe beam for performing Schlieren imaging independently. A 500µm wire was placed at the focus of L1 to block the probe rays and perform dark field Schlieren. To generate sufficient image separation for the long plasma, the 2° Wollaston was placed 55cm from the detector giving a fringe separation of ~66µm and image separation greater than 1 cm. The finite size of the probe pulse limits the region with fringes to approximately half the interferogram and accounting for optical apertures of the probe beam in Orion, sets a limit of ~15mm on the region that can probed.

FIG 5: Schlieren image of Argon clusters (50 bar backing pressure density, >1019 cm-3, cluster size ~ 5nm) irradiated with 1TW 1054nm pulse and F/20 focusing. Transverse probing was performed using 2ω 500fs pulses 16ns after the laser-cluster interaction. Here a well-defined thin -shelled blast wave is launched into the surrounding gas, and is preceded by a radiative precursor.

FIG 6 : Interferogram taken under identical conditions (but different shot) as the Schlieren image in the previous figure along with an inset figure from the white box. Left hand side image shows the large field of view afforded by a 36mm by 24mm detector. This allows the entire 12mm bore gas jet to be imaged. Right ) Close-up of the boxed region shows high resolution interferometricdata can be obtained at large field of view.

Page 9: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

Figure 5 shows a dark field Schlieren image of a blastwave formed 16ns after the laser-cluster interaction. The faint halo surrounding the interaction region is due to a radiative precursor, where plasma self-emission at soft x-ray wavelengths (XUV) ionises the cluster medium, creating an electron density gradient that is visualised by the Schlieren technique15. Figure 6 shows the interferogram obtained under identical conditions, this can be unfolded using an Abel inversion technique to extract the plasma electron density4. It is worth highlighting the large field of view possible with a 36mm by 24mm detector which can encompass large targets such as a 12 mm bore gas jet. The close up image inset demonstrates that fine details in the turbulent blast wave region can be well resolved.

B. OFF HARMONIC CAPABILITY

In short pulse driven experiments (currently 1054nm and 527nm drive wavelengths are available at Orion), harmonic emission and optical transition radiation from the laser-target interaction can be a powerful source of background light in addition to plasma self-emission. Since the Orion probe beam is also a harmonic of the drive beam, it is not possible to filter out this light for discrimination purposes using bandpass or interference filters. This makes the currentscheme for the short pulseimaging probing system using open shutter cameras potentially incompatible with short pulse drive experiments that create very large self-emission backgrounds. Two solutions exist for this problem, one of which would be to use a gated detector (such as GOIs) which would allow the initial bright background from the interaction to be filtered in the time, though at the expense of spatial resolution and dynamic range. The other option would be use an off-harmonic wavelength for probing, in which case the current methodology will be applicable with the simple addition of filters to remove harmonic wavelengths. Whilst Orion has access to GOIs, they are significantly lower resolution devices than CCD camerasand have a smaller field of view. The use of Raman-shifted probe16 beams at 630nm has been demonstrated as a method to observe small scale structure close to irradiated targets and would be a useful probing capability for Orion. Imaging performance at 630nm was assessedby selecting a narrow spectrum of a laser driven supercontinuum17source using interference filters. Fringes obtained under an identical setup to that used in figure 3are shown in figure 7.Whilst the fringe quality is limited by the spatial coherence of the generated supercontinuum light, the preliminary results highlight the possibility of using this imaging system over a very wide wavelength range.

Page 10: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

FIG. 7: Fringes obtained using a supercontinuum source which is spectrally apertured at 630nm. Whilst fringe quality is limited by the spatial coherence and phase aberrations of the broadband light, the possibility of using Raman shifted probe wavelengths is demonstrated.

IV. Conclusion

We report on the development and testing of a multi-wavelength TIM compatible imaging system for short pulse probing on the ORION laser facility. Multi-harmonic capability together with operation over a wide wavelength range without the need for extensive system realignment is achieved using custom manufactured highly corrected UV-triplets. Interferometry is achieved using a self-path matched configuration based on Wollaston prisms. This allows for a compact linear optical configuration which is easy to align and fits within a restricted engineering package suitable for use in a demanding experimental environment. Data is presented which highlights promising performance at second and fourth harmonic wavelengths of anNd:glass probe beam. Preliminary measurements also highlights the possibility of using Raman shifted probe wavelengths to perform off-harmonic interferometry.

ACKNOWLEDGEMENTS

This work was performed for and funded by the Atomic Weapons Establishment (AWE), Aldermaston, UK. N.H.Stuart, T.S.Robinson, S.Giltrap and H.F.Lowe are funded by AWE/EPSRC CASE Studentships. S. Patankar is funded by EPSRC KTS fellowship RSRO P43480.

1. Hopps, Nicholas, et al. Plasma Physics and Controlled Fusion 57.6 (2015): 064002.2. Bentley, C. D., et al. Review of Scientific Instruments 83.10 (2012): 10D732.3. Oades, Kevin, et al. Review of scientific instruments 75.10 (2004): 4222-4224.4. Hutchinson, Ian H. Principles of plasma diagnostics. Cambridge university press, 2005.5. Norman, Michael J., et al. .Applied optics 41.18 (2002): 3497-3505.6. Benattar, R., C. Popovics, and R. Sigel.Review of Scientific Instruments 50.12 (1979): 1583-

1586.

Page 11: spiral.imperial.ac.uk · Web viewPolarised light interferometry has been demonstrated on laser experiments previously6 and offers unique advantages from an operational perspective.

7. Custom UV optics manufacturer, Bernhard Halle Nachfl. GmbH, Hubertusstraße, 10D - 12163 Berlin, Germany

8. Moosman, B., et al. " Review of scientific instruments 70.1 (1999): 672-676.9. Stamper, J. A., et al,.Physical Review Letters, 40(18), 1177.10. Bigourd, D., et al. Applied Physics B 113.4 (2013): 627-633.11. Kurt. J. Lesker 704 vacuum pump oil was used for its low vapour pressure. 12. Ditmire, T., et al. Physical review letters 78.16 (1997): 3121.13. Moore, A. S., Symes, D. R., & Smith, R. A. (2005). Physics of plasmas, 12(5), 52707-52707.14. Ditmire, T., et al. " Applied physics letters 71.2 (1997): 166-168.15. Ditmire, T., et al. The Astrophysical Journal Supplement Series 127.2 (2000): 299.16. RS Adrain et al, Optics Communications, 15 (2), pages 290-292, (1975 )17. Alfano, Robert R. "The supercontinuum laser source." (1989): 458.