+ All Categories
Home > Documents > Vital Signs Monitor UConn BME 4900

Vital Signs Monitor UConn BME 4900

Date post: 06-Jan-2016
Category:
Upload: purity
View: 23 times
Download: 0 times
Share this document with a friend
Description:
Vital Signs Monitor UConn BME 4900. Vital Signs Monitor. Purpose As the population ages, many people are required by their doctors to take vital signs on a daily basis. Developed for the average person to use at home. Vital signs transmitted to a computer and ultimately to a doctor. - PowerPoint PPT Presentation
Popular Tags:
17
Vital Signs Monitor Vital Signs Monitor UConn BME 4900 UConn BME 4900
Transcript
Page 1: Vital Signs Monitor UConn BME 4900

Vital Signs MonitorVital Signs MonitorUConn BME 4900UConn BME 4900

Page 2: Vital Signs Monitor UConn BME 4900

Vital Signs MonitorVital Signs Monitor Purpose

As the population ages, many people are required by their doctors to take vital signs on a daily basis.

Developed for the average person to use at home.

Vital signs transmitted to a computer and ultimately to a doctor.

Page 3: Vital Signs Monitor UConn BME 4900

MeasurementsMeasurements

What can it measure? Electrocardiogram

(ECG) Three leads (Right

arm, left arm, right leg)

Page 4: Vital Signs Monitor UConn BME 4900

What is an ECG?What is an ECG?

Records a time waveform of heart electrical activity.

Used to diagnose heart problems Arrhythmia Myocardial infarction Conduction blocks (bundle branch block)

Page 5: Vital Signs Monitor UConn BME 4900

ECG TraceECG Trace

Page 6: Vital Signs Monitor UConn BME 4900

Cardiac Electrical ActivityCardiac Electrical Activity

Page 7: Vital Signs Monitor UConn BME 4900

The ECG WaveformThe ECG Waveform

Page 8: Vital Signs Monitor UConn BME 4900

Measuring Electrical ActivityMeasuring Electrical Activity 3 leads

2 arm 1 leg

As cardiac muscle depolarizes it creates a potential on the skin

Potential measured as a vector Difference between right and left

electrode measured with a differential amplifier.

Page 9: Vital Signs Monitor UConn BME 4900

Measuring Electrical ActivityMeasuring Electrical Activity

Circuit consists of: 3 op amp differential amplifier Low pass filter (Cutoff: 0.05 Hz) High pass filter (Cutoff: 160 Hz) Gain Amplifier

Page 10: Vital Signs Monitor UConn BME 4900

Differential AmplifierDifferential Amplifier

What does the buffer do? Provides a high input impedance which means low current flow

What is the purpose of C2 and C3? Remove DC offset Offset would be amplified by 1000x

10 mV at 1000x gain = ~10V

Feedback Amplifier

Buffer

Differential Amplifier

Page 11: Vital Signs Monitor UConn BME 4900

Differential AmplifierDifferential Amplifier

V1 & V2 are in the microvolt range

We need it to be in the milivolt range so Rf/R1 is set to ~1,000

Page 12: Vital Signs Monitor UConn BME 4900

11stst Order LPF Order LPF

The ECG is known to be a low frequency signal

Usually less than 100Hz

A LPF can be used to remove the high frequency noise in the signal

Page 13: Vital Signs Monitor UConn BME 4900

High Pass FilterHigh Pass Filter ECG frequency range ~0.05 Hz – 150 Hz Why do we need to eliminate frequencies <0.05

Hz? Avoid distortion of the ST wave

Page 14: Vital Signs Monitor UConn BME 4900

MicroprocessorMicroprocessor Provides Analog to Digital (A/D) conversion

Waveform y(t) sampled at a fixed rate Δt Voltage read every Δt and converted to a number If resolution is 8 bits

Gives us 28 = 256 counts over 5V range ~19.5 mV per count

Page 15: Vital Signs Monitor UConn BME 4900

Patient IsolationPatient Isolation

Bluetooth is wireless technology Bluetooth provides separation of the patient from

the equipment and line source Very important to protect patient from electrical

shock

Page 16: Vital Signs Monitor UConn BME 4900

Labview SoftwareLabview Software

Page 17: Vital Signs Monitor UConn BME 4900

Labview RequirementsLabview Requirements

Receive serial data from bluetooth Display raw and filtered EKG Calculate heart rate


Recommended