VLSI and Embedded Systems Conference, 5-9 Jan 2014, IIT Bombay, India ( VLSIDES14 ) Session: B-2 Embedded Platform, Venue: VMCC -21, Session Time: 4:30 pm to 6:30 pm An Embedded System Design for a Synchronous Demodulation Based Noninvasive Bioimpedance Sensor P. C . Pandey. IIT Bombay - PowerPoint PPT Presentation
Popular Tags:
30
VLSI and Embedded Systems Conference, 5-9 Jan 2014, IIT Bombay, India (VLSIDES14) Session: B-2 Embedded Platform, Venue: VMCC-21, Session Time: 4:30 pm to 6:30 pm An Embedded System Design for a Synchronous Demodulation Based Noninvasive Bioimpedance Sensor P. C. Pandey IIT Bombay 07/Jan/2014 <[email protected]> http://www.ee.iitb.ac.in/~pcpandey
Transcript
Page 1: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

VLSI and Embedded Systems Conference, 5-9 Jan 2014, IIT Bombay, India (VLSIDES14)Session: B-2 Embedded Platform, Venue: VMCC-21, Session Time: 4:30 pm to 6:30 pm

An Embedded System Design for a Synchronous Demodulation Based Noninvasive Bioimpedance Sensor

P. C. Pandey

IIT Bombay

07/Jan/2014<[email protected]> http://www.ee.iitb.ac.in/~pcpandey

Page 2: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

VLSI and Embedded Systems Conference, 5-9 Jan 2014, IIT Bombay, India (VLSIDES14)Session: B-2 Embedded Platform, Venue: VMCC-21, Session Time: 4:30 pm to 6:30 pm

P. C. Pandey: An embedded system design for a synchronous demodulation based noninvasive bioimpedance sensor (invited talk)

Abstract: A long-duration noninvasive monitoring of bioimpedance has the potential of serving as a low-cost diagnostic tool and monitoring device in several medical applications, e.g. impedance cardiography (sensing of variation in thoracic impedance to estimate cardiac output and some other hemodynamic parameters), pneumography (sensing of respiratory parameters), plethysmography for sensing peripheral blood circulation, glottography (for sensing movement of vocal chords), etc. These instruments pass an alternating current of high frequency and low amplitude through a pair of appropriately placed pair of electrodes, an amplifier to sense the resulting amplitude modulated voltage across the same or another pair of appropriately placed electrodes, a demodulator to detect the impedance signal, and signal processing for obtaining the desired parameters. An embedded system design approach is used to develop a body-worn device to be used for monitoring the clinically important physiological parameters during critical care, for ambulatory recording for early diagnosis of cardiovascular disorders and for post-operative care, for monitoring of physiological parameters for use in sports medicine, and as a low-cost diagnostic aid. It senses the basal value and time-varying component of the impedance waveform, with settable excitation frequency and with very low noise and demodulation related distortions. A microcontroller and an impedance converter chip are used for stable sinusoidal source with programmable frequency control and a digital synchronous demodulation. A voltage-to-current converter with balanced outputs is designed using two operational trans-conductance amplifiers for current excitation. The sensed voltage is added with a sinusoidal voltage obtained from the excitation source and with digitally controlled amplitude and polarity to increase its modulation index before digital synchronous demodulation and for baseline correction of the sensed impedance signal. Two digital potentiometers have been used to provide independent control over current excitation and baseline correction. Synchronous digital demodulation in the impedance converter chip gives real and imaginary part of the impedance. An isolated RS232 interface is provided to set the parameters and to acquire the sensed impedance signal.

Dr. P. C. Pandey, Professor, Electrical Engineering, IIT BombayEE Dept, IIT Bombay, Powai Mumbai 400076, India<[email protected]> http://www.ee.iitb.ac.in/~pcpandey

Page 3: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

3/30

Outline

1. Introduction

2. Design Approach

3. Hardware & Software

4. Test & Results

5. Summary

ReferenceHitendra Sahu: “Sensing of impedance cardiogram using synchronous demodulation”, M. Tech. dissertation, Biomedical Engineering, Indian Institute of Technology Bombay, June 2013.

Page 4: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

4/30

Noninvasive Monitoring of Bioimpedance oLow-cost diagnostic tool

oMonitoring device

Some Applications

oImpedance cardiography: sensing of variation in thoracic impedance to estimate cardiac output & some other hemodynamic parameters

oPneumography: sensing of respiratory parameters

oPlethysmography: sensing of peripheral blood circulation

oGlottography: sensing movement of vocal chords during speech production

Page 5: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

5/30

Instrumentation for Bioimpedance Sensing

o Passing an alternating current of high frequency and low amplitude through a pair of appropriately placed pair of electrodes

o Amplifier to sense the resulting amplitude modulated voltage across the same or another pair of appropriately placed electrodes

o Demodulator to detect the impedance signal

o Signal processing for obtaining the desired parameters

Page 6: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

6/30

Impedance detector

ECG extraction circuit

Current source

Differentiator

Zo(t)

dZ/dt

ECG

ICG blocks• AC excitation current • Voltage sensing amp.• Demodulator• Baseline correction• ECG extractor

Example: Impedance Cardiograph

Operation• Excitation current: 20 - 100 kHz, < 5 mA• Amplitude demodulation of the sensed voltage: Z(t) with basal impedance

(20 − 200 Ω) & time-varying component (< 0.2 Ω)• ICG: − dZ/dt, processed with ECG as the reference.

Page 7: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

7/30

Objective

To develop a body-worn bioimpedance sensing device for

• Monitoring the clinically important physiological parameters during critical care (multi-channel signal acquistion & processing)

• Ambulatory recording for early diagnosis of cardiovascular disorders and for post-operative care (recording in the presence of motion artifacts)

• Monitoring of physiological parameters for use in sports medicine (recording in the presence of external interference, strong respiratory and motion artifacts)

• Low-cost diagnosis (low distortion & high sensitivity)

Page 8: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

8/30

Design Approach

• Digital synchronous demodulation for noise and interference rejection • Circuit for increasing the modulation index of the waveform to increase the sensitivity and dynamic range

Basic Blocks• Microcontroller “Microchip PIC24FJ64GB04”• Impedance converter chip “Analog Devices AD5933”• V-to-I convertor and amplitude control • Voltage sensing amplifier and baseline correction• PC-based GUI with isolated serial communication for setting

parameters and data acquisition

Page 9: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

9/30

Impedance converter AD5933

Features• Excitation voltage generator & digital synch. demodulator• Programmable voltage with a settable frequency up to 100 kHz• Impedance measurement range from 1 kΩ to 10 MΩ• Internal system clock • DC rejection, error averaging, phase measurement• Accuracy: ± 0.5%.• I2C interface with a data rate of 100 kHz

Adaptations needed for bioimpedance sensing • Measurement using current excitation• Time-varying measurement• Dynamic range extension and sensitivity selection

Page 10: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

10/30

Functional block diagram of AD5933

Page 11: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

11/30

Design using the impedance converter chip with on-chip sinusoidal source & DFT for synchronous digital demodulation

Page 12: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

12/30

Impedance converter circuit

Page 13: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

13/30

• Digital pot. AD8400 (U3, U7) used for controlling the amplitudes the excitation current and baseline correction voltage.

• Total resistance 1 K with 8 bit resolution.• Wiper position changed via SPI interface.• Supply range : 2.6 – 5.5 V.

Page 14: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

14/30

B

E

-V

I

C+VRB

180

VDD

3

1

8

7

2

4

OPA 861

ix

VSS

Vin

RC

180

RIQ

180VSS

RE

180

Vo

V-to-I converter• Voltage generated from

AD5933 is converted to a current

• Operational trans-conductance amplifier (OTA) based V – I converter used

• “TI OPA861” from used as OTA which offers :• gm is 90 mA/V at a

quiescent current of 4.7 mA

• +5 V single supply operation

• Bandwidth 80 MHz

Page 15: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

15/30

V – I converter with balanced current outputs

• Complimentary balanced current output generated by using two OTA chips

• Quiescent current set byR1 and R3

connected atbase terminals

• Output current can is controlled by changing R2 or

quiescent current

Page 16: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

16/30

Voltage sensing amplifier

• Instr. amp. INA155 for amplifying the sensed voltage• BW: 5.5 MHz• Gain: 10 – 50• Slew rate 6.5 V/µs• Supply: 2.6 – 5.5 V• High pass filter

cut-off : 16 kHz

2

34

7

6+

10kR14R15

1

8

R13100k

C15 100p

C14 100p 100kR12

E2

E1

U6

A5V

V_IN1_BL

AGND

C16

AGND

0.1µ

A1V6

A1V6

5

A1V6

5k

A1V6

R9

R8

10k

10k

R11

R10

1M

1M

Page 17: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

17/30

Baseline correction

• Subtracting a sinusoidal reference voltage from the sensed voltage

• Amplitude and polarity of the correction voltage digitally controlled by varying digital pot (U7) ratio between 0.25 to 0.75

• Baseline correction output tracked by microcontroller using ADC.

• Potentiometer ratio is controlled digitally via SPI interface

where ,

where

K= ),β, and α are potentiometer

ratios from U3 and U7

respectively

Page 18: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

18/30

Demodulation

• Samples the sensed signal at 1.04 MHz.

• 1024-point DFT is calculated at the excitation frequency.

• The DFT algorithm returns a real (R) and imaginary (I) data values.

• Data values are transferred to microcontroller over an I2C.

• Impedance values send to PC over UART or USB.

Page 19: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

19/30

Microcontroller

• 44-pin PIC24F64GB004

used• Supply range : 3.0

– 3.6 V• 16 MHz clock • 64 KB program

memory, 8 KB RAM • Single channel 10

bit ADCs• UART module• USB module• SPI module• I2C module

Page 20: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

20/30

Power supply features• Separate analog & digital supplies of 3.3 V & 5 V.

• Analog reference of 1.6 V generated by MCP6021.

• LDO MCP1802 used as voltage regulator IC.

• Input to the LDO from a DC-DC converter LM2622.

• Input to the DC-DC converter: 3.6-5.5 V.

• Li-ion charge control IC MCP73833 used for battery charging.

• Total current consumption ~60 mA.

• Low battery indication.

• Provision for powering through USB.

Page 21: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

21/30

Power supply ckt

Page 22: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

22/30

AssemblyTwo-layer PCB (102 mm x 64 mm) with SMD components

Page 23: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

23/30

Signal acquisition interface

LabWindows CVI software for signal acquisition using RS232

Page 24: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

24/30

Test & Results

• Exc.: 65.5 kHz, 0.9 mA

• Lin. range: up to 400 Ω Plot between Voltage sensing amplifier output and test resistances

A) Voltage sensing amplifier: output linearity

B) Interference

Significant only over a b.w. of 3 kHz

Page 25: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

25/30

C) Automatic Sensitivity AdjustmentVoltage sensing amplifier output vs test resistances

for excitation current of 0.6 − 1.5 mA, set by varying β

10 20 30 40 50 60 70 80 90 100 110 1200

200

400

600

800

1000

1200

1400

1600

1800

2000

B=0.08

B=0.04

B=0.12

B=0.16

B=0.20

B=0.24

B=0.28

b=0.32

B=0.36

B=0.40

B=0.44

B=0.48

B=0.51

B=0.55

B=0.60

B=0.63

B=0.67

B=0.70

Test resistance (Ω)

Vo

ltag

e o

utp

ut

(mV

)

Page 26: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

26/30

D) Validation using thoracic impedance simulator

Excitation: 0.6 mA, 65.56 kHzSimulator settings:R = 49 Ω, ∆R = 0.5Ω , f = 1 HzSampling freq.: 200 Hz

Excitation: 0.6 mA, 65.56 kHzSimulator settings:R = 20 Ω, ∆R = 0.8 Ω, f = 0.1 HzSampling freq.: 10 Hz

0 5 10 15 20 25 30 35 4020.6

20.7

20.8

20.9

21

21.1

21.2

21.3

21.4

21.5

Time (s)

Impe

danc

e (o

hm)

Figure 5.3 Impedance Vs Time

Page 27: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

27/30

Excitation: 0.6 mA, 65.56 kHzSimulator settings:R = 30 Ω, ∆R = 0.8 Ω, f = 0.1 HzSampling freq. : 200 Hz

Excitation: 0.6 mA, 65.56 kHzSimulator settings :R = 19 Ω, ∆R = 0.5 Ω , f = 5 HzSampling freq.: 200 Hz

Page 28: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

28/30

SummaryDeveloped

A bioimpedance sensor using an impedance converter chip using digital synchronous demodulation

Further work

• Median filtering for further carrier ripple rejection without smearing transitions

• Adaptation for for specific applications

• Integration with the signal processing software

• Field testing

Page 29: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

29/30

References[1] R. P. Patterson, "Fundamentals of impedance cardiography," IEEE Eng.

Med. Biol. Mag., vol. 8, no. 1, pp. 35-38, 1989.

[2] L. E. Baker, "Applications of impedance technique to the respiratory system," IEEE Eng. Med. Biol. Mag., vol. 8, no. 1, pp. 50–52, 1989.

[3] L. E. Baker, "Principles of impedance technique," IEEE Eng. Med. Biol. Mag., vol. 8, no. 1, pp. 11–15, 1989.

[4] H. H. Woltjer, H. J. Bogaard, and P. M. J. M. de Vries, “The technique of impedance cardiography,” Euro. Heart J., vol. 18, no. 9, pp. 1396–1403, 1997.

[5] M. D. Desai, “Development of an impedance cardiograph,” M. Tech. dissertation, Biomedical Engineering,, IIT Bombay, 2012.

[6] H. Sahu: “Sensing of impedance cardiogram using synchronous demodulation”, M. Tech. dissertation, Biomedical Engineering, IIT Bombay, June 2013.

Page 30: VLSI and Embedded Systems Conference, 5-9 Jan 2014,  IIT  Bombay, India ( VLSIDES14 )

P. C

. Pan

dey,

EE

Dep

t., II

T B

omba

y

30/30


Recommended