+ All Categories
Home > Documents > Volume Estimation From Multiplanar 2D Ultrasound Images

Volume Estimation From Multiplanar 2D Ultrasound Images

Date post: 03-Jun-2018
Category:
Upload: harshoi-krishanna
View: 226 times
Download: 0 times
Share this document with a friend

of 12

Transcript
  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    1/12

    ELSEVIER

    Ultrasound in Med. & Biol., Vol. 22, No. 5. pp. 561-572, 1996

    Copyright

    0 1996 World Federation for Ultrasound in Medicine & Biology

    Printed in the USA. All rights reserved

    0301.S629/96 515.00 + .OO

    PII: SO301-5629( 96)00022-l

    *Original Contribution

    VOLUME ESTIMATION FROM MULTIPLANAR 2D ULTRASOUND

    IMAGES USING A REMOTE ELECTROMAGNETIC

    POSITION AND ORIENTATION SENSOR

    S. W.

    HUGHES,

    T. J.

    DARcY,*

    D. J.

    MAXWELL,*

    W.

    CHIU,

    A.

    MILNER, *

    J. E.

    SAUNDERS?

    and R. J.

    SHEPPARD~

    - Departments of Medical Physics, *Obstetrics & Gynaecology and Paediatrics, Guys and

    St. Thomas Hospital Trust, London UK; and Department of Physics, Kings College.

    University of London, Strand, London UK

    (Received 6 June 1995; in $nal form 21 November 1995)

    Abstract-A system is described for calculat ing volume from a sequence of multiplanar 2D ultrasound

    images. Ultrasound images are captured using a video digit ising card (Hauppauge Win/TV card) installed

    in a personal computer, and regions of interest transformed into 3D space using position and orientation

    data obtained from an electromagnet ic device (Polbemus, Fastrak). The accuracy of the system was assessed

    by scanning 10 water fll led balloons (13-141 mL), 10 kidneys (147-200 mL) and 16 fetal livers (S-37

    mL) in water using an Acuson 128XP/lO (5 MHz curvilinear probe). Volume was calculated using the

    ellipsoid, planimetry, tetrahedral and ray tracing methods and compared with the actual volume measured

    by weighing (balloons) and water displacement (kidneys and livers). The mean percentage error for the

    ray tracing method was 0.9 2 2.4 , 2.7 2 2.3 , 6.6 & 5.4 for balloons, kidneys and livers, respectively.

    So far the system has been used clin ical ly to scan fetal livers and lungs, neonate brain ventricles and adult

    prostate glands.

    Key Words: Three dimensional ultrasound, Multiplanar image registration, Volume measurement.

    INTRODUCTION

    The measurement of volume is becoming increasingly

    important in medicine as abnormal volume is often

    indicative of abnormal function or pathology. In the

    past, attempts have been made to estimate the volume

    of an organ or structure from plain X-rays by measur-

    ing, for instance, the height, width and depth and as-

    suming that the object approximates to a geometric

    shape such as an ellipsoid (Dodge et al. 1960; McLach-

    lan et al. 1968; Austin 1971 et al. 1977; Levine et al.

    1979). This has obvious drawbacks as it is impossible

    to accurately describe the shape of an organ from just

    three dimensions.

    With the advent of tomographic imaging modalities

    such as computed tomography (CT), magnetic resonance

    imaging (MRI) and positron emission tomography

    (PET) it has become possible to obtain a number of

    cross-sectional images through an object to enable more

    Address co rrespondence to: Stephen W. Hughes, M.Sc., Medi-

    cal Physics Department, St. Thomas Hospital, Lambeth Wing, Lon-

    don SE1 7EH UK.

    accurate determinations of volume. CT, MR and PET

    scanners produce parallel images of known separation

    making volume calculations fairly straightforward. Ultra-

    sound images are also tomographic in nature, but are

    usually acquired by hand and so tend to be multiplanar,

    i.e.,

    non-parallel, unevenly spaced and maybe sheared.

    Images must be registered in some way to a common

    coordinate system before accurate volume calculations

    can be made. There are two essential requirements for

    calculating volume: ( 1) delineation of a region interest

    (ROI) on each image; and (2) knowledge of the position

    and orientation of each image with respect to a common

    coordinate system.

    REGISTRATION OF IMAGES

    Various techniques have been developed for regis-

    tering ultrasound images. The ultrasound transducer

    can be attached to a mechanism to ensure that parallel

    images of known separation are acquired. e.g., for

    scanning the fetus (Klein et al. 1993), blood vessels

    (Blankenhorn et al. 1983; Picot et al. 1993; Pretorius

    et al. 1992), heart (Matsumoto et al. 1981; Raichlen

    561

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    2/12

    ih Llltmsound in Medicme and Biology

    Volume 37. Numhrr 5. I996

    et al. I986 ), prostate ( Hastak et al. 1982) and maxi llo-

    facial anatomy (Hell et al. 1993).

    Motorised pull-back systems for use with intra-

    cavity transducers have been reported for imaging the

    prostate t Sehgal et al . 1993 ), female urethra

    ( Kirschener-Hermanns et al. 1994) and transesopha-

    geal imaging of the heart (Wollschlager et al. 1989).

    In motorised pull-back systems, the distance between

    image planes is found by mult iply ing the pull-back

    velocity by the time interval between image acquisi-

    tions. In manual pull-back systems, image separation

    is measured using a scale beside the catheter. Intravas-

    cular catheters can also be used to acquire a 3D data

    set, either using the manual pull-back method (Ennis

    et al. 1993: Rosenheld et al. 1991, 1992). or rotation

    about the central axis ( Kok-Hwee et al. 1994).

    Ghosh et al. ( 1982) and Pini et al. ( 1989) describe

    a system in which the transducer is held in a rig allowing

    only central axis rotation of the transducer. Fine et al.

    ( 1991 ) describe a similar technique using a hand held

    transducer. Knowledge of the angle between image

    planes enables image registration to be carried out.

    Any system in which the ultrasound probe is

    attached to a mechanical device obviously restricts free-

    dom of movement in some way. This problem can be

    overcome to some extent by connecting the transducer

    to an articulated arm that records the position and orienta-

    tion as the transducer is scanned across the body (Sawada

    et al. 1983: Teicholz et al. 1974). Such arms usually

    have variable potentiometers placed at each join t to mea-

    sure limb angle.

    To allow more or less complete freedom of move-

    ment. remote systems for measuring the position and

    orientat ion of an ultrasound transducer are required.

    Brinkley et al. ( 1978), King et al. ( 1990), Moritz et al.

    ( 1983) and Levine et al. ( 1989) describe an acoustic

    ranging system. This utilises three spark gaps attached

    to a small plat form connected to the transducer, and four

    microphones placed on a stationary L shaped bar.

    The distance between the microphones and spark gaps

    is calculated from the time-of- flight of the sound. To

    avoid significant errors, the air temperature must be con-

    tinually monitored as the velocity of sound in air changes

    at a rate of about 2 per 10C and there must be a clear

    path between the spark gaps and microphones.

    Electromagnet ic systems for image registration

    have been reported by Gardener et al. ( 1991), Lees et

    al. ( 1991 a.b, 1993 ). Kel ly et al. ( 1992) and Hodges et

    al. ( 1994 ) In all of these systems, an electromagnet ic

    signal is emitted from a stationary transmitter and de-

    tected by a mobile receiver. Both the transmitter and

    receiver are constructed from three mutually orthogo-

    nal coils of copper wire. The distance and orientat ion

    of the receiver are calculated from the relative strength

    of the signals received in each sensing coil . The advan

    tage of electromagnetic systems is that the receiver i\

    small and ligh t and so can easily be attached to an

    ultrasound transducer. A major disadvantage is that

    nearby metal and stray electromagnetic signals can

    produce errors in the measurements.

    MATERIALS AND METHODS

    Hardware

    The system that we have developed tits onto a

    trolley (Fig. 1) , and comprises an IBM compatible PC

    (486 DX2-66, 8 MB memory, 200 MB hard disc)

    (ProData 486, Data Products and Services, Inc. Love-

    land, CO, USA ) . Ultrasound images are captured using

    a video digitising card (Win/TV, Hauppauge Com-

    puter Works. Inc., Hauppauge, NY, USA, distributed

    in the UK by ODT, London). The image data is stored

    in a 1 MB frame buffer in YUV 4: 1: 1 format and then

    transferred to the hard disc in various standard image

    formats (in this case BMP, Microsoft Windows device

    independent bitmap). The video signal is digitised with

    a resolution of 7 bits, resulting in 128 grey levels. The

    acquisition rate depends on image size. For 328 X 228

    pixe l images, the acquisition rate is two frames per

    second directly to the hard disc.

    The position and orientat ion of the ultrasound

    transducer is measured using a Polhemus 3Space Fas-

    trak system (Polhemus, Inc., Colchester, VT, USA.

    distributed in the UK by Virtual Presence, London).

    The Fastrak system comprises a transmitter and re-

    ceiver, both of which connect to a system electronics

    unit (SEU ). The SEU is connected to the PC via a

    RS-232 serial link (Fig. 2 ) . The transmitter emits a

    8013 Hz electromagnet ic signal that is detected by

    the receiver. In a metal-free environment, the static

    accuracy of the system is specified as 0.8-mm RMS

    and 0.15 (2.6 18 X 10 - radians) RMS, and the resolu-

    tion, 0.005 mm per cm of range and 0.025 (2.181 x

    lo- radians). These values apply when the receiver

    is within 76 cm of the centre of the transmitter.

    We found that direct placement of the receiver

    onto the Acuson l28XP transducer used in this experi-

    ment caused skewing of the outlines. To overcome this

    problem the receiver was attached to a perspex bar,

    which was attached to a moulded plastic shell clipped

    onto the transducer case (Fig. 3). The receiver is small

    (2.8 X 2.3 X 1.5 cm) and light ( 17 g) and so does

    not interfere with normal scanning.

    Software

    The SEU is controlled by sequences of ASCII

    characters transmitted from the PC via the serial link.

    Each value is transmitted in binary (4 bytes/number),

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    3/12

    Volume estimation from multiplanar 2D ultrasound images 0 S. W. HUGHES et a/. 563

    Fig. 1. The main components of the syste m fi t onto a trolley that can easily be transported between hospital

    departments and ultrasound machines. The Acuson 128XP scanner is seen in the centre with the PC and Fastrak

    on the trolley to the lef t and the wooden transmit ter stand to the right of the examination couch.

    at a baud rate of 9600 bits/s; therefore 41 ms are

    required to transmit each set of position and orientation

    data (seven numbers in total). Position and orientation

    data (POD) are acquired immediately after each image

    acquisition.

    The Win/TV card was programmed using a soft-

    ware development kit produced by Hauppauge.The pro-

    gram describedhere was written in C using the Microsoft

    Visual C+ + programming environment. Programswere

    developed to outline regions of interest (ROIs), trans-

    form the outline points

    to

    real space,connect the trans-

    formed outline points into a seriesof triangles. and apply

    receiver

    GlL

    ransmitter

    /

    ultrasound transducer

    C

    video signal from

    US machine

    Fig. 2. The Fastrak transmitter and receiver are attached to

    a sys tems electronic unit (SEU) , which in turn is connected

    to a personal computer (PC) via an RS-232 serial link. The

    video signal from any ultrasound machine is connected to a

    video capture board within the PC, which enables images to

    be displayed in Microsoft Windows and subsequent ly stored

    on disc.

    Fig. 3. The Fast& receiver is attached to the ultrasound

    transducer by means of a moulded plastic shell and perspex

    bar. The receiver is held away from the transducer to reduce

    electromagnetic interference from components within the

    transducer.

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    4/12

    564 Ultrasound in Medicine and Biology Volume 22, Number 5, 1996

    the volume calculations. In order to transform the outline

    points to real space, the position and orientation of the

    Fastrak receiver with respect to the centre of the front

    face of the ultrasound ransducer must be recorded along

    with information such as image calibration factors, etc.

    These data are stored as a separateposition information

    file. Analysis of the results was carried out using a Unix

    graphics supercomputer (Titan Kubota Pacific Inc.,

    Santa Clara, CA, USA).

    Algorithms

    extending above the origin and the +z axis extending

    out of the page toward the viewer. The 3D coordinate

    of any point within an image is calculated by summing

    the vectors between the centre of the receiver and a

    point on the image, as shown in Fig. 4 where:

    p = position vector of image point;

    f = position vector of Fastrak receiver;

    v1 = vector projecting from Fastrak receiver to a point

    above the transducer centre;

    v2 = vector projecting from the end of v, to the trans-

    ducer centre;

    Conversion

    of

    image coordinates to Fastrak coor-

    dinates.

    Orientation quarternions are used to calculate

    the direction cosines (Cooke et al. 1992) of three vec-

    tors that project from the centre of the receiver (Fig.

    4). Quarternions were used as only four numbers are

    required to encode the sameorientation information as

    six direction cosines, therefore, the data can be trans-

    ferred more quickly. Also quarternions are a more ro-

    bust way of encoding directional information, espe-

    cially compared to Euler angleswhere there is an ambi-

    guity between yaw and roll at high pitch angles.

    v,, vY = vectors pointing in the same direction as the

    image x and y axes;

    t,, t, = image coordinates of the centre of the trans-

    ducer face;

    c,, c, = calibration factors in the x and y direction;

    Ax, Ay = x,y coordinates of image point relative to

    the centre of the transducer face;

    Ax = t, - x, Ay = t, - y. where x and y represent

    the image coordinates of

    a point.

    The line-of-sight (LOS) vector projects from the

    front face of the receiver, the line-of-hear (LOH) vec-

    tor is directed out of the right hand side at right angles

    to the LOS, and the line-of-plumb (LOP) vector is

    directed downwards at right angles to both the LOS

    and LOH vectors (Fig. 4). The coordinate system is

    the sameas that generally used for graphs with the +x

    axis extending to the right of the origin, the +y axis

    If v, is in the LOS direction, v2 in the LOP direc-

    tion, v, in the LOH direction and v, in the LOS direc-

    tion, the coordinates of an image point in Fastrak space

    are given by:

    PX = fX + 10 Iv,1 + lopxlv*l

    + los,Ayc, + loh,Axc,

    py=fy+losy Iv11 +lop, Iv21

    + los,Ayc, + loh,Axc,

    Pz = fz + 10 Iv1 + lop, lvzl

    + los,Ayc, + loh,Axc,

    LOH

    traasducer case

    Fastrak sensor

    Perspex rod

    . centre of transducer face

    ROI points

    Fig. 4. The Fast& receiver is attached o the ultrasound

    transducer y means f a perspexbar. The receiver contains

    three orthogonalcopper coils that define he x, y and z axes

    (LOS, LOH and LOP, respectively). In this arrangement,

    four vectors (v, , v2, v, andv,) are summedo enable egion

    of interest (ROI) points to be transformed rom image to

    Fast& coordinates.Any other configurationof the vectors

    could be used,but the error is likely to increasewith the

    numberof vectors used.

    The above equations are converted into matrix

    form as,follows.

    Let:

    m, = loh,c,

    ml0 = los,c,

    mzo= f, + 10~ Iv11 + lop, Iv21

    mm = lohyc,

    ml, = los,c,

    ml = fy + los, 1~11 lop, IhI

    mo2 = loh,c,

    ml2 = los,c,

    m22 = L + los, Iv11 + lop, lb1

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    5/12

    Volume estimation from multiplanar 2D ultrasound images

    0 S. W. HUGHES

    et al.

    565

    and

    Pi = (AX, AY, 1).

    Then

    fi = pims

    where:

    f, = Fastrak coordinate of ith image point;

    pi = coordinate of ith point relative to the centre of

    the transducer face;

    m, = transformation matrix for slice s.

    A ruler was used to measure the distances of the

    centre of the receiver above the image plane ( 1 1 I) and

    between the point in the image plane directly below the

    receiver and the centre of the ultrasound transducer

    face ( 1 Z I) . The position and orientation of the receiver

    does not matter so long as the lengths and directions

    of the vectors are known. The position of the receiver

    relative to the ultrasound transducer face can be de-

    scribed by more than two vectors i f necessary, but this

    may increase the error in the volume estimate. This is

    because each vector has a certain error in the position

    of the start point and projection angle. The uncertainty

    will increase with the number of vectors placed end

    to end. For the receiver attachment shown in Fig. 3,

    I v, 1 = 17.9 cm and I v2 1 = 2.8 cm. The centre of the

    receiver was assumed to be where the manufacturer

    claims it is.

    The centre of the transducer face is marked on

    one of the images and this point assumed to be at the

    end of the LOS vector projecting to the centre of the

    transducer face. (If the LOS vector is offset to one

    side of the centre of the transducer face, then the equiv-

    alent point must be marked on the image.) This point

    becomes a new origin on the image and is essential

    for calculating the 3D coordinate of an image point.

    The x and y axis calibration factors are calculated using

    scale markers on the ultrasound image. The x,y,z Fas-

    trak coordinate of a point in the image is calculated

    by multiplying the x,y image coordinates (relative to

    the transducer origin) by the unit receiver vectors that

    happen to be aligned with the x and y axes of the

    image (in the example shown in Fig. 4, the LOH vector

    for the x axis and the LOS vector for the y axis).

    If the x axis of the image is antiparallel to the

    positive direction of the corresponding receiver axis,

    the object will be reflected about a line parallel to the

    y axis that passes through the centre of the transducer

    face. However, as all points retain their relative posi-

    tion,

    the calculated volume will not be affected. If, for

    some reason,

    the absolute positions of the ROI points

    are required, the image x axis may have to be reversed

    (most machines have a facility for doing this).

    Volume calculations.

    Volume was estimated using

    the ellipsoid, planimetry and tetrahedral methods, and

    a new ray tracing method. These methods were chosen

    to cover the types of algorithm that might be used to

    calculate volume. The latter two methods require the

    surface of the object to be represented as a triangular

    mesh and the former two require just the basic ROI

    points. The ellipsoid method is the most commonly

    used method to estimate volume, especially for hearts

    and prostate glands in which the volume of the organ

    is assumed to be equal to that of a prolate ellipsoid of

    the same length, height and width as the organ. The

    main advantage of the ellipsoid technique is its sim-

    plicity. The main disadvantage is that significant errors

    that can result if an organ is not roughly ellipsoidal in

    shape. The algorithm implemented on our system is a

    simulation of the usual ellipsoid technique. The long

    axis is taken as the distance between the centroids of

    the first and last outlines and the product of the two

    minor axes is calculated by dividing the largest outline

    area by 2~.

    Planimetry involves multiplying the area of each

    ROI by the local slice thickness (Watanabe 1982).

    The accuracy of this technique depends on image

    planes being close together with only a slight differ-

    ence in orientation between adjacent planes. If the sur-

    face of the ROI is modelled by a triangular mesh,

    the volume can be calculated be generating tetrahedra

    using the triangle vertices and outline centroids (Cook

    et al. 1980). We have developed a fourth ray tracing

    technique that enables the volume of a triangulated

    object of any shape to be calculated. A grid is con-

    structed in the x,y plane equal in extent to the maxi-

    mum x and y sizes of the triangulated object. Rays are

    cast through the object from each grid node, and the

    volume calculated by multiplying the distance between

    intercepts on each side of the object by the area of

    each grid element.

    Balloons.

    Ten balloons (ranging in size from

    13.6- 141.6 mL) were filled with water and the necks

    tied of f with surgical suture. Water filled balloons were

    chosen for this experiment because, unlike organ phan-

    toms, their volume can be accurately measured by

    weighing (assuming a density of 1 g/mL for water).

    The balloons were attached to a hollow perspex cylin-

    der by means of a rubber band and placed in a tank

    of tap water at 26C. The tank was placed on an exami-

    nation couch (Fig. 1) in approximately the same posi-

    tion as a patient undergoing an ultrasound scan. The

    balloons were scanned using an Acuson 128XP/lO

    with a ~-MHZ curvilinear probe. The transducer was

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    6/12

    566 Ultrasound in Medicine and Biology Volume 22, Number 5. 1996

    moved continuously across the balloons. The operator

    was an experienced ultrasonographer who is currently

    using the system in a clinical trial designed to assess

    the benefits of ultrasonic volume measurements n fetal

    medicine. The Fastrak transmitter was attached to a

    purpose-built wooden stand (Fig. 1) and suspended

    above and to the side of the tank. During scanning,

    the transducer was oriented so that there was a clear

    line of sight between the transmitter and receiver. The

    receiver and transmitter were between 0.5 and 1 m

    apart during scanning. Twenty images with a 324 x

    224 matrix were acquired over a period of 10 s (two

    images per second). The x,y dimensions of each pixel

    was 0.353 and 0.376 mm, respectively. After scanning,

    the balloons were emptied and weighed (Sartorius

    MC 1, Sartorius AG, Gottingen, Germany ) to an accu-

    racy of kO.01 g. The volume of the water in the bal-

    loons was assumed o be the total weight of the balloon

    minus the weight of the rubber and suture, divided by

    the density of water.

    of the difference in the speed of sound in tissue at

    37C ( 1540 m/s) and in water at 26C ( 1495 m/s;

    Bamber and Hill 1979). This discrepancy results in

    overestimation of distance in the direction of sound

    propagation by 2.92 ; therefore, the calculated vol-

    umes need to be reduced by a factor of 0.9708.

    The mean percentage error was calculated by di-

    viding the measuredvolume by the ( signed) difference

    between the measuredand calculated volumes. The F-

    test was used to assess he significance of the differ-

    ences between the variances and in cases where the

    variances were not significantly different, one-way

    analysis of variance was used to assesshe significance

    of differences between the means.

    The image and data files were transferred to the

    Titan computer via a local network and the edges of

    the balloons (Fig. 5 ) outlined using a mouse and trans-

    formed to Fastrak coordinates (Fig. 6). After transfor-

    mation, a triangulation algorithm (Hughes and Brueton

    1994) was used to tessellate he surface of each balloon

    (Fig. 7 ) . Prior to tessellation and transformation, the

    outline points are interpolated into a number of evenly

    spaced points to improve triangulation. The surface is

    displayed as a sequence of alternating light and dark

    triangular facets and can be rotated about the x,y,z axes

    in the display to check that triangulation has proceeded

    correctly. The volume of each balloon was estimated

    using the four algorithms described above.

    The volumes were corrected to take into account

    Kidneys. Ten cadaveric pigs kidneys (obtained

    from a local supermarket), ranging in volume from

    147-200 mL, were placed in turn on a perspex plinth

    in a 26-L perspex tank filled with normal saline (0.9

    NaCl solution), and scannedusing an Acuson 128XP/

    3. The transducer was placed in a derrick placed at

    right angles to the long axis of the tank, which enabled

    the transducer to be set at various angles. The kidneys

    were scanned with radial, parallel and slanted scans o

    simulate how an organ might be scanned n the clinic.

    The transducer was positioned by hand, but was sta-

    tionary during each mage acquisition. Between 15 and

    20 images were captured of each kidney, and the im-

    ages were analysed in the same way as the balloons.

    Immediately after scanning, each kidney was placed

    in a 2000-mL graduated cylinder to measure the vol-

    ume of water displaced. The accuracy of the water

    displacement technique was assessedy measuring the

    volume of five accurately machine perspex cylinders

    ranging in volume from 30-260 mL. (The experiment

    Fig. 5. Ultrasound mage hroughone of the water filled balloonsscannedn a plasticbowl full of tap water. The

    middle of the balloon wall wasoutlined on successivemages.

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    7/12

    Fig. 6.

    Volume estimation from multiplanar 2D ultrasound images

    0 S. W. HUGHES etnl. 567

    A balloon outline transformed using the position and orientation data obtained from the Fast&.

    was carried out without the observers knowing the

    volume of the cylinders.) The velocity of sound was

    assumed o be 1540 m/s within the kidneys. Errors in

    the path lengths of the A scan lines caused by differ-

    ences between the velocities of sound in water and

    tissue were ignored.

    Fetal livers. Sixteen cadaveric fetal livers (8-

    37.5 mL), fixed in formaldehyde, were scanned in

    a tank of distilled water. The same scanner, scan

    techniques and volume calculation methods were

    used as for the kidneys. The actual volume of the

    kidneys was assessed y placing the livers in a 500-

    mL graduated cylinder and noting the rise in water

    level. The accuracy of this method was assessedby

    measuring the volume of seven perspex blocks ( 3-

    71 mL).

    Scanner test

    The Acuson scanner was tested using the Cardiff

    resolution test object (Gammex-RMI Ltd, Nottingham,

    UK), which contains parallel wires of known separa-

    tion embedded in tissue equivalent gel. The tests in-

    volve measuringdistancesbetween wires and assessing

    the minimum wire separation that can be resolved. The

    range was set to 140 mm, with a medium depth focus.

    The mean vertical large scale linear calliper accuracy

    was 0.16 & 0.1 mm and the mean horizontal large

    scale inear calliper accuracy 0.6 2 1.64 mm. The axial

    Fig. 7. Transformed ROI points are connected to form a series of triangles. Alternate dark/light shading makes

    it easier for an observer to check that the triangulation algorithm has been performed correctly.

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    8/12

    568 Ultrasound in Medicine and Biology

    LOS

    Volume 22, Number 5, 1996

    d = 2r sin(W2) = 2.61 cm

    If 1 m and n designate x,y and z

    direction cosines, the angle

    between vl and v2 may be calculated

    from cos 8 = l,l, + m,m, + n,n2

    centre of wheel

    Fig. 8. A perspex wheel was constructed to test the accuracy of the Fastrak system. The receiver was attached to

    the wheel as shown. The wheel was moved in 15 increments and the angle and distance moved were calculated.

    resolution was

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    9/12

    Volume estimation from multiplanar 2D ultrasound images l S. W. HUGHES et al.

    569

    RESUTS

    Balloons

    Figure 9(a) shows a plot of the measured volume

    vs. that calculated by the four methods described

    above. It is apparent that the ellipsoid method has more

    scatter than the other three methods. Figure 9(b)

    shows that the 95 range for the ellipsoid method is

    much greater than for the other three methods. There

    is little difference in the 95 limits of agreement for

    the other three methods. The error bars indicate

    the mean +2 standard deviations of the differences

    between the measured and calculated volumes for the

    four methods (Bland and Altman 1986). Analysis of

    variance showed no significant difference (p = 0.49)

    between the planimetry, tetrahedral and ray tracing

    methods. However, there was a significant difference

    between the ellipsoid technique and the other three

    techniques (p = 0.018).

    Kidneys

    For the ray tracing method, there was no signifi-

    cant differences (p > 0.4) between the variances of the

    three scan techniques. However, there was a significant

    difference (p = 0.02) between the mean volumes pro-

    duced by the radial, parallel and slanted scan tech-

    niques. For radial scanning, there was no significant

    difference between the mean volumes produced by the

    four volume estimation methods (p = 0.56)) however,

    there was a highly significant difference between the

    variances of the ellipsoid and the other three methods

    (p < 0.001). The mean percentage error for the ray

    tracing method using the radial scan technique was

    2.7 k 2.3 . The accuracy and precision of the water

    displacement technique was -0.93 2 1.82 .

    Fetal livers

    As for the kidneys, for ray tracing there was no

    significant difference in the variance between the three

    different scanning techniques (p > 0.36) although

    there was a significant difference between the means

    (p < 0.001). For the radial scanning technique there

    was no significant difference between the mean and

    variance of the planimetry, tetrahedral and ray tracing

    volume estimation methods; however, there was a sig-

    nificant difference (p < 0.04) between variance of

    these three methods and the ellipsoid method. The

    mean percentage error for the radial scanning tech-

    nique and ray tracing volume calculation method was

    6.6 2 5.4 . The accuracy and precision of the water

    displacement technique was 7.8 2 9.8 .

    Balloons, kidneys and livers grouped together

    Figure 9(a) shows a plot of measured volume

    versus calculated volume for the balloons, kidneys and

    fetal livers. Figure 10(b) shows a Bland-Altman plot

    showing the variance of the absolute error about the

    mean error. An interesting point to note is that although

    the percentage error decreases with volume, or at least

    in the order liver, balloons and kidneys, the absolute

    error in terms of milliliters increases. Table 1 summa-

    0 20 40 60 60 100 120

    140 160

    Measured volume (ml)

    a

    10

    o

    i

    00

    0

    8 . , .bQ

    _ , . . . . . . 1 . . . O . .

    . . Q g ;

    t

    -1020

    ; 1,

    0 20 40 60 80 100 120 140

    160 180

    Average of measured and calculated volume (ml)

    b

    Fig. 9. (a) Plot o f measured

    balloon volume verses he volume calculatedby the ellipsoid el), planimetry ( pla),

    tetrahedral (tet) and ray (ray) tracing methods.The line of identity is shown. (b) Bland-Altman plot of the

    difference between he calculatedvolumesand tire actual volume against he averageof the two. The 95 range

    barsare designated y the first letter of the method hat they refer to. The middle of eachbar representshe mean

    difference and the two outer bars plusor minus wo standarddeviations.

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    10/12

    Ultrasound in Medicine and Biology

    Volume 22, Number 5, 1996

    25C

    s

    E

    200

    E

    _3

    P

    150

    8

    l-l

    P

    100

    t

    F

    0

    50 100 150 200 250

    Measured volume (ml)

    a

    -10 i +

    0 50 100

    150 200 250

    Average of the measured and calculated volumes (ml)

    b

    Fig. 10. (a) Plot of calculatedversusmeasured olume for the balloons,kidneys and fetal livers. The ray tracing

    techniquewas used o calculate all volumes.The balloonswere scanned ree-handand the kidneys and livers

    scannedwith radial techniquewith the transducer eld in a rig. The line of identity is shown. (b) Bland-Altman

    plot asdescribed or Fig. 9. The 95 rangebarserror barsare designated y the first letter of the scanned bject.

    rises the results for each group and for all groups taken

    as a whole. There appears to be a correlation (Y =

    0.99, p = 0.008) between the mean balloon, kidney

    and liver volumes and the standard deviations of the

    absolute errors. Table 2 shows the results for ray trac-

    ing volumes for kidneys and livers for the three scan-

    ning techniques. Note that the for some reason the

    slanted scansunderestimate volume.

    Fastrak accuracy

    Over a complete rotation of 360, the means and

    standard deviations of the LOS, LOH and LOP vectors

    were 15.03 + 0.7, 15.04 t 0.7 and 1.09 + 0.7,

    respectively, and the mean chord length was 2.715 ?

    0.22 cm. In reality the transducer would probably not

    be moved more than ?30 from the vertical. For a 60

    excursion over 10.4 cm, the angle was measured as

    15.18 5 0.06 ( 1.2 + 0.4 ) and the chord length 2.609

    + 0.019 cm (0.05 + 0.73 ). Over a distance of 10

    cm, a 0.18 error in measuring angle would result in

    Table 1. The meanand standarddeviation of the

    percentage rror, for the ellipsoid, planimetry,

    tetrahedraland ray tracing methods or the balloons,

    kidneys and ivers.

    Ellipsoid

    Planimetry Tetrahedral Ray trace

    Balloons

    13.4 t 20.9 2.1 * 2.1 0.9 2 1.9

    0.9 2 2.4

    Kidneys 5.3 10.6 5.9 t 2.1

    4.3 f 2.2 2.7 ir 2.3

    Livers

    2.9 ?I 8.7 6.5 k 5.2 1.3 + 5.2

    5.2 2 4.7

    The kidneys and livers were scanned radially (therefore all the

    scans used to produce data for this table were m ultiplanar).

    a position error of 0.3 mm. To assesshe noise level,

    the standard deviation of the x,y and z coordinates and

    LOS, LOH and LOP vectors were calculated. The

    mean standard deviation of the x,y and z coordinates

    was 0.012 mm, and the mean angular standard devia-

    tion 0.028, which over a distance of 10 cm will result

    in a position error of 0.05 mm.

    DISCUSSION

    The results show that the system has the potential

    to provide accurate estimates of organ volume using a

    variety of volume calculation algorithms. Our results

    are comparable to those obtained by other workers on

    balloon phantoms (Brinkley et al. 1978; Eaton et al.

    1979; Geirsson et al. 1982; Hastak et al. 1982; Hodges

    et al. 1994).

    In general, there was not much to choose between

    the planimetry, tetrahedral and ray tracing methods of

    estimating volume, which were much superior to the

    ellipsoid method. The radial, parallel and slanted scan

    methods produced results with similar variance, but

    with significantly different means. More work needs

    Table 2. The meanand standard eviationsof the

    percentage error for the kidney and liver ray trace volumes

    for the three scan echniques.

    Kidneys

    Livers

    Radial

    2.7 2 2.3

    6.6 F 5.4

    Parallel

    2.1 z 2.7

    4.6 2 5.5

    Slanted

    -0.9 2 2.8

    -2.8 lr 6.8

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    11/12

    Volume estimation from multiplanar 2D ultrasound images 0 S. W.

    HUGHES et al.

    571

    to be done to assess the importance of these differences

    and to see whether they can be reduced.

    A number of factors influence the overall error in

    the result, for instance the size and quality of the im-

    ages, the scanning technique (e.g., orientat ion and

    number of slices), errors in the Fastrak data, and errors

    in the volume algorithms. Interference from metal is

    a problem with electromagnet ic systems; however, our

    results show that for small excursions of the receiver

    the accuracy is adequate even in the presence meta l

    (for instance in the examination couch). The accuracy

    of the Fastrak is similar to that quoted by the manufac-

    turer over the kind of excursions used for scanning an

    organ. Errors in measuring relative position and angle

    affect the error in measuring volume, but not errors

    in absolute position. The percentage error seems to

    decrease with volume while the limits of agreement

    appear to increase. Clin ically, in some cases the per-

    centage error is likely to be of most interest, but in

    other cases, e.g., cardiology, the actual difference will

    be important (as this will have a direct bearing on the

    precision of estimating stroke volume and, therefore,

    cardiac output). As in other imaging modalities, the

    critical factor inf luencing the error is likely to be the

    accuracy of delineating in uivo ROIs. This will depend

    on the quality of the machine used and the ability of

    the ultrasonographer.

    Further work needs to be done on assessing: inter-

    and intraobserver ROI tracing errors; the accuracy re-

    quired in positioning the Fastrak receiver relative to the

    ultrasound transducer ; and environmental influences on

    the Fastrak. Tissue inhomogeneities in the path of the

    ultrasound beam are likely to cause geometric distortions

    due to refraction and time-of-fight errors.

    In recent years, dedicated 3D ultrasound scanners

    have been developed. These systems have various ad-

    vantages over add on systems, such as the one de-

    scribed in this article. For instance, dedicated 3D sys-

    tems are less cumbersome to use. They are excellent

    for small objects, but cannot be used over extended

    distances as is required for vascular imaging. There-

    fore, 2D systems with an added 3D capability are likely

    to complement dedicated 3D systems.

    Currently, the vast majority of scanners are 2D,

    and this is likely to remain the case for several years

    to come. The system that we describe also has the

    advantage of being relative ly cheap (the Polhemus de-

    vice costs just under &5000 and the video card &300),

    can be connected to any scanner, and is easily trans-

    portable enabling sharing between scanners. Another

    advantage of external localisation systems is the abil ity

    to scan an organ from different directions and to com-

    bine the images.

    The software to out line and transform ROIs and

    to calculate volume has now been implemented on a

    PC. So far the system has been used to scan fetal livers

    and lungs, adult prostate glands and neonate brain ven-

    tricles. We believe that these results are encouraging

    and demonstrate that volume estimates within 10 are

    obtainable in the clinic.

    Acknowledge ments-The authors would like to thank Terry Clark

    for scannin g the balloo ns. Roy Reed of the Medical Physic s Work-

    shop for crafting the transducer attachment, Chris Bunton of the

    Medical Physi cs Workshop for making the wheel, Agnes Aning-

    amoah for handling the organs, George Gore of the Carpentry De-

    partment for making the transmitter stand, Sue Chin and Julie Simp-

    son of the Public Health Department for advice on the statistical

    analysis, Acuson for the lone of the 128XP/3 and Tommys Cam-

    paign a nd BT for funding this work.

    REFERENCES

    Aust in JHM, Gooding C A. Roentgenograph ic measurement of skull

    size in children . Radiology 197 1;99:641-646.

    Bamber JC, Hill CR. Ultrasonic attenuation and propagation speed

    in mammalian tissues as a function of temperature. Ultrasound

    Med Biol 1979;5:149-157.

    Bland JM, Altman DG. Statistical methods for assessing agreement

    between two methods of clinical measurement. Lancet

    1986;1:307-310.

    Blankenhorn DH, Chin HP, Strikwerda S, Bamberger J, Hestenes

    JD. Common carotid artery contours reconstructed in three di-

    mensions from parallel ultrasonic images. Radiology 1983:

    148533-537.

    Brinkley JF, Moritz WE, Baker DW. Ultrasonic three-dim ensional

    imaging and volume from a series of arbitrary sector scan s.

    Ultrasound Med Bio l 1978;4:317-327.

    Cook LT, Cook P N, Lee KR. Batnitzky S, Wong BYS. An algorithm

    for volume estimatio n based on polyhedral approximation. IEEE

    Trans Biomed Eng BME 1980;27:493-500.

    Cooke JM, Zyda MJ, Pratt DR. McGhee RB. NPSN ET: Flight sim u-

    lation dynamic m odelling using quartemions. Presence 1992;

    1:404-420.

    Dodge HT, Sandler H, Ballew DW, Lord JD. The use of biplane

    angiocardiography for the measurement of left ventricular vol-

    ume in man. Am Heart J 1960;60:762-776.

    Eaton LW, Maughan WL, Shoukas AA, Weiss JL. Accurate volume

    determination in the isolated ejecting canine left ventricle by

    two-dim ensional echocardiography. Circulat ion 1979; 60:320-

    326.

    Enn is BM, Zientek DM, Ruggie NT , Billhar dt RA, Klein LW. Char-

    acterization of a sapheno us vein graft aneurysm by intravascular

    ultrasound and computerized three-dim ensional reconstructio n.

    Cathet Cardiovasc Diagn 1993;28:328-33 I.

    Fine D. Perring S, Herbetko J, Hacking CN, Fleming JS. Three-

    dimensional imaging of the gallbladder and dilated biliary tree:

    Reconstruction from real-time B-scans. Br J Radio1 1991:

    64: 1056- 1057.

    Gardener JE, Lees WR, Gillams A. Volume imaging with US. Radi-

    ology 1991;18l(P):l33.

    Geirsson RT, Christie AD, Pate1 N. Ultrasound volume measure-

    ments comparing a prolate ellipsoid method with a parallel plani-

    metric area method against a known volume. J Clin Ultrasound

    1982; 10:329-332.

    Ghosh A, Nanda NC. Maurer G. Three-d imension al reconstru ction

    of echocardiographic images using the rotation method. Ultra-

    sound Med Biol 1982;8:655-661.

    Gohari P . Berkowitz RL, Hobbins JC. Predica tion of intrauterine

    growth retardation by determination of total intrauterine volume.

    Am J Obstet Gynecol 1977; 127:255-260.

    Hastak SM, Gammelgaard J, Holm HH. Transr ectal ultraso nic vol-

    ume determination of the prostate-a preoperative and postoper-

    ative study. J Urol 1982; 127:l 115-I 118.

  • 8/11/2019 Volume Estimation From Multiplanar 2D Ultrasound Images

    12/12

    571

    Ultrasound in Medicine and Biology Volum e 22, Number 5, 1996

    Hell B, Walter FA, Schreiber S, Blase H, Bielke G. Three-dimen-

    siona l ultrasonography in maxill ofacia l surgery. A new diagnos-

    tic tool. Int J Oral Maxillofac Surg 1993;22:173-177.

    Hodges TC, Detmer PR, Bums DH, Beach KW , Strandness DE Jr.

    Ultraso nic three-dim ensional reconstruc tion: In vitro and in viva

    volume and area measuremen t. Ultrasound Med Biol 1994:

    20:7 199729.

    Hughes SW, Brueton RN. A technique for measuring the surface

    area of articular cartilage in acetabular fractures. Br J Radio1

    1994;67:584-588.

    Kelly IMG, Gardener JE, Lees WR. S-dim ensiona l fetal ultrasound.

    Lancet 1992;339:1062-1064.

    King DL, King DL Jr, YiCi Shao M. Three-dimensional spatial

    registration and interactive display of positi on and orientation of

    real-time ultrasound image s. J Ultrasound Med 1990;9:525-532.

    Kirschener-Herm anns R, Klein HM, Muller U, Schafer W, Jakse G.

    Intraurethral ultrasound in women with stress incontinence. Br

    J Ural 1994;74:315-318.

    Klein H , Marqaurt M, Polz H, Schul ler G. 3D reconstru ction of fetal

    images based on ultrasound. In: computer assisted radiology.

    Lemke HU, Inamura K, Jaffe CC, Felix R, ed. Berlin: Proceed-

    ings of the International Symposium , 1993534-538.

    Kok-Hwee Ng. Evans JL, Vonesh MJ, Meyers SN. Mills TA. Arte-

    rial imagin g with a new forward-viewing intravascular ultrasound

    catheter, II. Three-dimensional reconstruction and display of

    data. Circulat ion 1994; 89:718-723.

    Lees WR, Chong WK, Lawrence R, Gardener J. Intravascular ultra-

    sound with 3D reconstru ction: A new way to image vascular

    disease. Br J Radio1 1991a;64:655.

    Lees WR, Gardener JE, Brett AD, Davey MK. A ssess ment of 3D

    US image acquisiti on. Radiology 1993; 189( P):341.

    Lees WR, Gardener JE, Gillams A. Three-dimensional US of the

    fetus. Radiology 1991b; 181 (P):132.

    Levine RA, Handschum acher MD, Sanfi lippo AJ, Hagege AA, Har-

    rigan P. Three-dimensional echocardiographic reconstruction of

    the mitral valve, w ith implications for the diagnosis of mitral

    valve prolapse. Circula tion 1989; 80:589-598.

    Levine SC, Filly RA, Creasy R K. Identif ication of fetal growth

    retardation by ultraso nic estima tion of total intrauterine volume.

    .I Clin Ultrasound 1979;7:21-26.

    Matsumoto M, Inoue M, Tamura S, Tanaka K, Abe H. Three-dimen-

    sional echocardiography for spatial visualisation and volume cal-

    culation of cardiac structures. J Clin Ultrasound 1981;9: 157-

    165.

    McLachlan MSF, Williams ED, Fortt RW, Doyle FH. Estimation

    of pituitary gland dime nsions from radiographs of the sella tur--

    cica . Br J Radio1 1968;41:323-330.

    Moritz WE, Pearlm an AS, MC Cabe DH. et al. An ultraso nic tech-

    nique for imaging the ventricle in three dimensions and calculat-

    ing its volume. IEEE Trans Biomed Eng BME 1983:30:48X

    492.

    Pico t PA. Rickey DW, Mitchell R. Rankin RN. Fenster A. Three-

    dimensional Doppler imaging. Ultrasound Med Biol 1993:

    19:95-104.

    Pin i R, Monnini E, Masotti L, Greppi B, Gerofolini M. Echocardio -

    graphic computed tomography-of the heart: Preliminary results.

    J Am Co11 Cardiol 1989:49:896.

    Pretorius DH, Nelson TR, Jaffe JS. 3-dimensional sonographic anal-

    ysis based on color flow Doppler and gray scale image data: a

    preliminary report. J Ultrasound Med 1992; 11:225-232.

    Raichle n JS, Trivedi SS, Herman GT, St. John Sutton MG, R eichek

    N. Dynamic three-dim ensional reconstruc tion of the left ventricle

    from two-dimensional echocardiograms. J Am Co11 Cardiol

    1986;8:364-370.

    Rosenfield K, Kauffman J, Pieczek AM, Langevin RE, Palefski PE.

    Human coronary and peripheral arteries: On-line three-dimen-

    sional reconstruction from two-dimensional intravascular US

    scans . Radiology 1992; 184:823-832.

    Rosenfield K, Los&do DW, Ramaswamy K, Pastore JO, Langevin

    RE. Three-d imensio nal reconstructio n of human coronary and

    peripheral arteries from image s recorded during two-dim ensional

    intravascular ultrasound e xamination. Circula tion 1991: 84:

    1938-1956.

    Sawada H, Fujii J, Kato K, Onoe M, Kuno Y. Three dimensional

    reconstruction of the left ventricle from multiple cross sectional

    echocardiog rams. Value for measuring left ventricular volume.

    Br Heart J 1983; 50:438-442.

    Sehgal CM, Broderick GA, Gomiak RJ, Arger PH. Three-dimen-

    sional imaging of prostates with transrectal sonography. Radiol-

    ogy 1993; 189(P):342.

    Stroud KA. Engineering mathematics. London: Macmillan, 1984.

    Teicho lz LE, Cohen MV, Sonnen blick EH, Gorlin R. Study of left

    ventricular geometry and function by B-scan ultrasonography in

    patients with and without asynergy. N Engl J Med 1974;

    291:1220-1226.

    Watanabe Y. A method for volume estima tion by using vector areas

    and centroids of serial cross-sections. IEEE Trans Biomed Eng

    BME 1982;29:202-205.

    Wollschlager H, Zeiher AM, Klein H, Kasper W, Wollschlager S.

    Transeso phageal echo computer tomography: A new method for

    dynamic 3-D imaging of the heart. J Am Co11 Cardiol 1989;

    13:68A.


Recommended