    Measurement Computing (508) 946-5100 1 [email protected] mccdaq.com

    WaveBook system with 8 strain inputs and 8 accelerometer inputs

    WaveBook/516EEthernet-Based Portable High-Speed

    Waveform Acquisition

    Also included is DaqCOM, an ActiveX/COM-based set of development tools that allowusers to easily develop their own customapplications in either Visual Basic, orC/C++. DaqCOM allows users to distributeWaveBooks throughout a facility of net-enabled PCs, allowing data acquired onany WaveBook attached on the network tobe viewed on other PCs on the network.

    DASYLab is also available for theWaveBook, allowing a user to simplyconnect icons to develop custom test andanalysis applications.

    The WaveBook/516E digitizer offers multi-channel waveform acquisition and analysisfor portable or laboratory applications. TheWaveBook includes 8 built-in channelsexpandable up to 72 channels of voltage,accelerometer, microphone, strain gage,thermocouple, position encoder, frequen-cy, high voltage, and other signal types.For applications beyond 72 channels,up to four WaveBooks can be combinedwithin one measurement system, for atotal capacity of 288 channels. You canalso add up to 854 thermocouples, withoutconsuming measurement bandwidth of theWaveBooks, using the WBK40 Series, andDBK90 signal conditioning options.

    WaveBooks are supported with a widevariety of software to address differentapplications and skill levels. IncludedWaveView software allows effortless set-up,time-domain waveform viewing, and real-time storage of acquired data to disk*. Alsoincluded is PostView, a post-acquisitionwaveform viewing application, allowingyou to visually scroll through multiplewaveforms on your PC screen. For userswho prefer to program, the WaveBookalso includes comprehensive drivers forDASYLab, LabVIEW, MATLAB, VisualC++, Visual C#, Visual Basic, and VisualBasic .NET.

    * WaveView supports up to 72 high-speed input channels. Applications with multiple WaveBooks, or withWBK40/41 thermocouple options, should use DASYLab or other programming applications supported bythe WaveBook, including C++, Visual Basic, or LabVIEW.

    Features 16-bit/1 MHz A/D

    1 s /channe l scanning of anycombination of channels Single and multichannel analog trigger-

    ing with programmable level & slope D i g it al T T L- le v el a nd p at t er n

    triggering Pulse trigger and external clock Programmable pre- and post-trigger

    sampling rates Sixteen digital inputs can be scanned

    synchronously with analog signals Operable from AC line, a 10 to 30 VDC

    source, such as a car battery, or optionalcompact rechargeable battery module

    Expandable up to 288 high-speedchannels

    SYNC connection allows multipleunits to sample synchronously

    Add up to 854 lower-speed thermo-couple channels

    DSP-based design provides real-timedigital calibration on all channels

    Signal Conditioning Options IEPE dynamic signal inputs Strain gages Programmable ltering Simultaneous sampling Quadrature encoder inputs Pulse/frequency measurements Thermocouples High-voltage measurements Vehicle bus network

    Software Includes WaveView for Out-of-the-Box

    setup, acquisition, & real-time display :- Scope mode for real-time waveform

    display- Logger mode for continuous stream-

    ing to disk Comprehensive drivers for DASYLab ,

    LabVIEW , MATLAB, Visual C++ ,Visual C# , Visual Basic , and VisualBasic .NET

    WaveCal software application for easyuser calibration

    Supported Operating Systems:Windows 7/Vista/XP SP2,

    32-bit or 64-bit

    The WaveBook/516E is ideal for measuring dynamic signals in portable and laboratory applications

  • 8/6/2019 Wavebook Data

    2/13

    Measurement Computing (508) 946-5100 2 [email protected] mccdaq.com

    Communicationsprocessor

    65536FIFO data

    buffer

    High-speed

    EPPinterface

    Differentialamplifier

    EEPROM

    100 MHz

    timebase

    16high-speed

    digitalinputs

    Expansionunit control

    BNC Inputs

    2

    3

    4

    5

    6

    7

    8

    Expansioninput select

    1

    HD-15Sconnector

    BNCinput

    Eightdifferential

    analoginputs

    Control&

    timing

    Unipolar/bipolar

    level shifter

    Buffers

    Buffers

    TTL trigger input

    Trigger select

    Expansionsignal input

    (from optional WBK10A, WBK14,

    WBK15, WBK16, WBK17, or WBK18)

    Digital signalprocessor

    (DSP)Scan

    sequencetable

    Real-timedigital

    correction

    +5Power supply

    5+1515

    DC/DCconverter

    Power from included AC adapter, optionalrechargeable batterymodule, 12V car battery, or any 10 to30 VDC source

    Power input

    1-MHz, 16-bit A/D converter

    circuitry

    Pulsediscriminator and analog

    trigger

    Ext.clock and

    frequencycounter

    PGAx1, 2, 5, 10

    Differentialamplifier

    16Pulseinput

    DAC

    Fixed 5-poleanti-alias filter

    Wide bandamplifier

    75

    16

    Expansion port 1 Expansion port 2 Expansion port 3

    To additional WBK40 series options

    10/100BaseTEthernet

    * Expandable with WBK30 option up to 128 Mbytes

    ExternaSYNC

    *

    WaveBook/516EGeneral Information

    A family of 8-channel WBK expansion op-tions provide volts, accelerometer, straingage, frequency, and quadrature encodermeasurements. Other WBK options provideisolated inputs as well as temperaturemeasurement capability. Most WBK optionsare fully programmable for gain/range, ltercut-off frequency, AC/DC coupling, etc.

    The WaveBooks design is optimized forexpansion, recon guration and portabil-ity. There are no bulky expansion chassisthat must be purchased in anticipation of future applications. The compact, all-metalchassis features a low-pro le package withfront panel inputs.

    The WaveBook provides sophisticatedmultichannel triggering, usually associatedwith larger, more expensive waveform re-corders. The DSP-based design also makesthe system more compact than others of similar performance.

    The WaveBook operates on a wide range of power sources, such as a standard AC line,an optional rechargeable battery supply, oreven a 12V car battery.

    All WaveBooks feature an input bufferampli er on each of its eight channels.This architecture isolates the input signalsfrom any multiplexing transients andgreatly improves frequency response. Italso provides far less sensitivity to input- signal source impedance.

    To achieve superior signal delity, excel-lent noise immunity, and greater accuracy,the WaveBook follows the input bufferamplifier with individual differentialand programmable gain ampli ers oneach channel. In addition, a 5-pole,20 kHz Butterworth anti-aliasing ltercan be software-enabled on a per- channelbasis to protect against aliasing of high-frequency signal components. For addi-tional gain and ltering, the WaveBookcan be factory con gured with a WBK11A,WBK12A, or WBK13A.

    The WaveBooks DSP automatically com-pensates for offset and gain errors in eachof its ampli ers, including those foundin the WBK expansion options on a per- reading, real-time basis. This compensationis based on calibration constants calculated

    during the systems calibration process. Theconstants are stored in the systems non-volatile memory. The DSP also permits theWaveBook to accept user-supplied calibra-tion constants, enabling it to automaticallyscale the input signal for gain and offsetindividually on every channel.

    The WaveBook offers both bipolar and uni-polar input ranges, which are per- channel

    programmable via the sequencer. Bipolarranges can extend from 0.05V to 10V.Unipolar ranges can span from 100 mVto 10V.

    The WaveBook also features a 16-bit high-speed TTL digital I/O port for recordingdiscrete TTL-level signals at the beginningof each scan, providing time correlationwith the analog inputs.

    WaveBook/516E Block Diagram

  • 8/6/2019 Wavebook Data

    3/13

    Measurement Computing (508) 946-5100 3 [email protected] mccdaq.com

    WaveBook/516ETriggering

    TriggeringThe WaveBook offers a variety of trigger

    modes. Selection of the optimal triggermode for your particular application re-quirements saves time and disk space byensuring that you capture only the dataof interest.

    A wide selection of programmable analogand digital trigger modes are available forstarting an acquisition. All trigger modes,along with the number of scans and thesample rate for pre- and post-trigger data,are software programmable prior to thestart of a scan sequence. The WaveBookalso supports digital pattern and pulsetriggering. Trigger latency (the maximumtime from the trigger to the rst readingof a scan group) and jitter (the variationof the latency from acquisition to acquisi-tion) depend on the speci c trigger sourceand type of acquisition. Descriptions of each trigger source and the various triggermodes follow.

    Software Trigger. A software trigger is is-sued by the PC, and causes the WaveBookto begin scanning the setup prede nedin the scan buffer. The trigger latency inthis mode is longer than in other triggermodes and is a direct function of the PCsperformance. It is typically 100 s or lessin post-trigger acquisitions.

    Digital TTL Trigger. The WaveBook ac-cepts a single TTL-level signal input to theDB25 digital I/O connector with rising- orfalling-edge trigger sensitivity selectedthrough software. Trigger latency in thismode is less than 300 ns for post-triggeracquisitions.

    Digital Pattern Trigger. In addition todigital trigger, the WaveBook supportsdigital pattern triggering. This allows data

    collection to start when a user-de ned16-bit digital pattern is matched on thedigital I/O connector. For example, it isuseful when trying to capture noise, vibra-tions, or some other physical disturbancewhich occurs at a particular point in a

    digitally sequenced process, from a PLC orrelay logic control system. Trigger latencyof the digital pattern trigger is less than300 ns for post-trigger acquisitions.

    Pulse Trigger. This high bandwidth inputenables the triggering and the correla-tion of lower speed waveforms with theoccurence of a high speed pulse. WithPulse Trigger, the user de nes a pulse byan amplitude between 5V and a pulsewidth from 300 ns to 0.8 sec.

    Analog Trigger Source. Analog sourcesare the most commonly used triggers. TheWaveBook supports both single-channelanalog triggers for quick captures, as well

    as multichannel analog triggering.

    Single-Channel Analog Trigger. Simplesingle-channel trigger performs ana-log comparison of channel one to aprogrammable 12-bit DAC value. You canalso select rising- or falling-edge criteria.Trigger latency is less than 500 ns.

    When a WBK11A, WBK12A, or WBK13Aoption is installed (providing simultaneoussample and hold or ltering), the triggerinput signal is ampli ed by the simulta-neous sample and hold ampli er before

    being compared against the analog triggerlevel. This can increase the trigger signalssensitivity by a factor of 100.

    Multichannel Analog Trigger. Multichan-nel triggering eliminates spurious data byletting you enter a more selective triggercondition to capture events of speci cinterest. In multichannel mode, any com-bination of up to 72 analog channels cancontribute to the trigger condition. Youcan individually program each channel tosatisfy its trigger criteria using one of eightstates from a combination of above/below

    level, rising/falling edge, and instantaneous/latched duration. In addition, you can alsode ne a hysteresis band for each channel,reducing false triggers when used withauto re-arm. Finally, you can join all triggerchannels together using ANY (logical

    OR condition) or ALL (logical ANDcondition) Boolean logic operands to forma single, uni ed trigger condition.

    The maximum latency possible in post-trigger acquisitions of the multichanneltrigger mode is 2 s per designated triggerchannel, plus 4 s. For example, if ve trig-ger channels are designated, the maximumlatency is 14 s. The minimum latency inthis mode is half the maximum rate; thus,the multichannel trigger jitter time is 1 sper trigger channel, plus 2 s.

    Acquisition ModesThe WaveBook lets you select one of several

    acquisition modes for collecting your pre-and post-trigger data. The system offersfour post-trigger modes and two pre-triggermodes. The WaveBook uses a combinationof internal memory, PC RAM, and your PCshard disk to seamlessly record all acquireddata during the acquisition process.

    Post-Trigger AcquisitionIn post-trigger acquisition, the WaveBookacquires data only after a trigger conditionhas been met. You have a choice of fourpost-trigger acquisition modes, each offer-ing bene ts suited to particular applications.

    In nite Linear Mode. In in nite linearmode, the system acquires data for an in -nite amount of time after a trigger occurs.This mode is most useful for chart-recorderreplacement applications that require longrecordings. Once started, the system keepsdigitizing until a stop command is issuedby your PC. The amount of data you canacquire depends on your PCs availablememory resources.

    In nite Circular Mode. In in nite circularmode, the system acquires data into a

    circular buffer inde nitely until it re-ceives a stop command from the PC.When the circular buffer is full, it over-writes previously acquired data; thus thebuffer always contains the most recentlyacquired data. This mode is most useful for

  • 8/6/2019 Wavebook Data

    4/13

    Measurement Computing (508) 946-5100 4 [email protected] mccdaq.com

    ChannelGain

    #2x1

    #4 #7 #2D #18 #9 #16x10 x5 x10 x10 x5

    WaveBook Scanning Example

    Channels can besampled arbitrarily

    Gain can be programmed for each channel dynamically

    Digital input lines,tach inputs, andencoders can also besampled before theanalog inputs in asequence 1 s

    t

    Unipolar or bipolar operation can beprogrammed for each channel dynamically

    UniUnipolar or

    bipolar

    Uni Bi Uni Uni Bi Bi

    Programmable from either aninternal or anexternal clock

    Scan group All channels within a scan group aremeasured at a fixed 1 s/channel

    x1

    t

    Expansion channelsare sampled at thesame rate asinternal channels

    WaveBook/516ETriggering

    The WaveBook provides maximum scanning exibility via a programmable channel/range sequencer. You can load the sequencer with any combination of channelsand associated ranges. Once loaded, the sequencer waits for the trigger condition tobe satis ed, after which it begins scanning and digitizing. When scanning begins,the WaveBook selects the appropriate channel and gain, digitizes the input signal,calibrates the reading, and transmits it to the PC via the on-board, 65 Kword FIFObuffer (or into optional internal memory, up to 128 Mbytes). This process is repeatedat a 1 s/sample rate until all channels within a scan group are completed.

    Upon completion of a scan group, the WaveBook can be con gured to proceedin one of several fashions. For example, the system can be programmed to beginthe next scan after a period of 1 s to 100s, programmable in 0.05 s increments.It can also be programmed to wait inde nitely until the trigger condition is nextsatis ed, after which it will again perform a speci ed number of scans, from 1 to100 million.

    The WaveBook also supports pre- triggering, and is capable of acquiring up to100 million scans prior to satisfaction of the trigger condition. The pre-triggerbuffer is circular, and thus always provides the most recently acquired readingsprior to the occurrence of the trigger. In addition, the system permits pre- triggerscan group repetition at one rate and post-trigger scan group repetition at anotherrate. For example, the system can be con gured to repeat scans 1,000 times persecond prior to the trigger, and then 100,000 times per second after receipt of the trigger.

    Channel-Scanning Flexibility applications in which le size is limited byPC resources, or an indeterminate numberof scans will occur before the stop conditionoccurs and only the last scans are required.A typical application is destructive testingin which acquisition is complete when thedevice fails, and only the nal failure modecharacteristics need to be captured.

    Finite Linear Mode. When operating innite linear mode, the system acquires data

    after receipt of the trigger until a speci ednumber of scans (from 1 to 100 million)are acquired. Finite linear mode is suitablefor applications in which the duration of the event is known.

    Re-arm. Finite linear mode permits thespeci cation of the re-arm condition.Under such a condition, after a speci ednumber of scans is acquired, the system au-tomatically prepares for a new acquisitionby re-arming and re-enabling the trigger,and then capturing a new nite numberof scans without the need for either useror computer intervention. This capabilityis useful in emulating a DSO (Digital Stor-age Oscilloscope), which typically offerscontinuous retrace. It is also bene cialfor unattended captures in which criticaltrigger events occur at indeterminate inter-

    vals, making manual re-arms awkward. Itis further useful for applications in whichtrigger events occur so quickly that it isdif cult to respond manually or undersoftware control.

    Finite Circular Mode. In this mode, theWaveBook acquires data into a circular buf-fer until a speci ed number of scans (from1 to 100 million) is acquired. When thecircular buffer becomes full, it writes overpreviously acquired data, and thus alwayscontains the most recently acquired data.This mode is useful for trigger delays; for

    example, the unit can be pre-con guredto record 100,000 scans (after the trigger)at 10 s intervals and to save only the last10,000 of the 100,000. In this example,the nal record would contain only dataacquired 900 ms after the trigger.

  • 8/6/2019 Wavebook Data

    5/13

    Measurement Computing (508) 946-5100 5 [email protected] mccdaq.com

    WaveBook/516ETriggering

    * PostView software does not support pre-trigger scan rates that differ from post** Pre-trigger linear mode and auto re-arm in the WaveBook hardware are not supported in WaveView*** Full 1M rate not available with external clock External clock counter and the high-speed digital inputs are mutually exclusive; adding the external clock

    counter to the scan list is equivalent to adding two analog channels Requires two locations in the scan sequencer The maximum scan rate when using SS&H is where n=number of channels in the scan list

    Pre-Trigger AcquisitionPre-trigger permits the acquisition of read-ings prior to the receipt of a trigger condi-tion. Pre-trigger can be used with any of the four trigger modes described above.

    When pre-trigger is selected, the rate atwhich a scan group repeats can differ beforeand after a trigger*. (Please note while thesample rates between pre- and post-triggerscan groups can differ, the time betweenconsecutive samples within a group is

    xed at 1 s.) For example, a scan group of 8 channels can be repeated at 1000 scans persecond prior to a trigger condition, and at100,000 scans per second after the trigger.The number of scans acquired during both

    time periods is also programmable.

    The WaveBook rst acquires a speci ednumber of pre-trigger scans at the pre- trigger scan rate, and then arms the trigger,guaranteeing that at least the speci ednumber of scans have been acquired beforethe trigger. The system continues to collectscans at the pre-trigger scan rate until thetrigger occurs. Once the trigger is detected,the system nishes collecting the currentpre-trigger scan, switches to the post-triggerscan rate, acquires readings using one of thefour post-trigger modes described above,

    and then stops the acquisition. The numberof pre-trigger and post-trigger scans areeach independently programmable from1 to 100 million.

    Pre-Trigger Circular Mode (requiresWBK30 installed). In this mode, data isplaced into a circular buffer until the buffer

    lls, after which the buffer writes over pre-viously acquired data. The system stays inthis mode, waiting for the pre-set triggercondition to be met, until the conditionoccurs or the acquisition is halted by the PC.Upon detection of the trigger condition,the system begins the post-acquisition

    readings. Since, for practical applications,the WaveBook is only limited by the PCsresources, extremely deep circular bufferscan be con gured for both the pre- andpost-trigger data.

    Pre-Trigger Linear Mode**. In this mode,the WaveBook continues collecting datawithout writing over old data. Since theamount of pre-trigger data can be nearlyin nite (if the trigger takes a very long timeto occur), this mode may require extensivePC resources to avoid data loss. Unlike thepre-trigger circular mode, which delays thetrigger arm condition until a prede nednumber of scans is collected, the pre-triggerlinear mode permits the WaveBook to begin

    recording post-trigger data as soon as thetrigger condition is met, regardless of thenumber of pre-trigger scans collected. Thismode is useful for applications in whichthe desired number of pre-trigger scans isindeterminate prior to the start of acquisi-tion and all data must be acquired.

    External Clock Input ***The WaveBook supports an external clockinput, allowing the scanning of data to bedependent upon an external pulse train.This feature is useful in rotating machine ormotion applications where data collection

    is dependent upon rotational speed or axialposition. In addition to allowing one scanper pulse, the WaveBooks external clockinput features a programmable dividercapable of reducing the incoming clock byup to 255. This is useful when the externalclock source is faster than the optimumsample rate for the data collection task.

    External Clock Timer . The WaveBook featuresan internal timer capable of reporting theperiod of the external clock input. This valuecan be read with each scan of the analogdata and is reset by the rising edge of theincoming clock. This is often bene cial in

    later analysis where physical phenomenaneeds to be correlated to speed.

    Multi-Unit SynchronizationMultiple WaveBook/516Es and WBK40/41scan be synchronized via the rear-panel SYNCports. Simply connect 2, 3, or 4 WaveBookstogether using SYNC cables (CA-74-1).WaveBook software establishes one of theWaveBooks as the master and the others asslaves . Master WaveBooks can run at thefull 1 MHz aggregate sampling rate; slaveWaveBooks must have 0.1 s of unassignedsampling time in the scan group (sometrigger modes are not supported in multi-WaveBook systems).

    Simultaneous Sampling The WaveBook samples each channel insequence, at a xed 1 s/channel rate. Forexample, when eight channels are scannedsequentially, the time between samplingthe rst channel and the eighth channel is7 s. For applications that require si-multaneous sampling (within 100 ns)of all channels, the optional WBK11Aeight-channel simultaneous sample andhold card and the WBK13A program-mable low-pass lter card with simulta-neous sample and hold are available. The

    WBK11A or WBK13A can be installed inthe WaveBook, or the WBK10A expan-sion chassis; they provide simultaneoussampling of all channels in a module.Even when multiple WBK11A or WBK13Acards are used within one system, allchannels with simultaneous sample andhold active are sampled within 100 ns of one another. Some WBK options includebuilt-in simultaneous sampling, includingthe WBK16/SSH and WBK18.

    1 MHz(n + 1)

  • 8/6/2019 Wavebook Data

    6/13

    Measurement Computing (508) 946-5100 6 [email protected] mccdaq.com

    DBK30A 1.9 Amps

    DBK34A 5.0 Amps

    TR-40U 3.3 Amps

    Input Maximum Output

    The WaveBook and its associated WBKmodules offer the exibility to be pow-ered either directly from a 10V to 30VDC source or via the included TR-40UAC power adapter. Options such as theWBK11A, WBK12A, and WBK13A signalconditioning cards install directly into theWaveBook or WBK10A and derive theirpower from those units.

    WaveBook/516ESystem Power Connection

    Table 2: Wavebook & Option Current Usage

    Qty.Units Amps Total

    WaveBook/516E x 2.10 =

    WBK10A, 8-channel expansion module x 0.30 =

    WBK11A, 8-channel simultaneous sample & hold card x 0.45 =

    WBK12A, 8-channel programmable low-pass filter card x 0.45 =

    WBK13A, 8-channel programmable low-pass filter card with SS&H x 0.50 =

    WBK15, 8-channel 5B signal conditioning module x 0.10 =

    WBK15, with 8 5B strain modules (max load) x 0.75 = WBK16, 8-channel strain gage module (no load) x 1.00 =

    WBK16/SSH, 8-channel simultaneous sample & hold module x 1.20 =

    WBK17, 8-channel counter/encoder module x 0.52 =

    WBK18, 8-channel dynamic signal conditioning module x 1.20 =

    WBK30, memory option for the WaveBook x 0.01 =

    WBK40 series, modules for WaveBook/516E, with no expansion x 0.60 =

    WBK40 series, modules for WaveBook/516E, with five DBK84s x 1.30 =

    DBK65, 8-channel transducer interface module x 0.83 =

    DBK70, vehichle network interface* x 0.50 =

    Max. Amps:

    Calculating the CurrentTable 2 provides the approximate required amperage for each component in yoursystem when using a 15V supply. Using this table, calculate the maximum amount of amperage your system will draw by multiplying the quantity of components used bythe amperage. Total the values in the last column to arrive at the maximum amperage;then verify that your power source has enough current capacity (see Table 1).

    * Typically draws power from diagnostic connector

    If you are operating each module with itsincluded TR-40U AC power adapters, thenno calculations are required; just connectthe power adapters. In certain applications,however, it may be advantageous or moreconvenient to operate all components froma single source such as a battery or UPS suchas the DBK34A. In those cases, the followingworktables provide the necessary informa-tion for calculating current requirementsfor your particular system.

    The table below shows the current draw avail-able from IOtechs DBK30A rechargeablebattery module, the DBK34A DC UPS, andthe included TR-40U AC power adapter.To minimize cabling, each module has apass-through connector allowing a numberof modules to be daisy chained, up to the5-ampere limit of the connector and CA-115power expansion cable. The exact number of daisy chained modules is dependent uponyour exact con guration.Table 1

  • 8/6/2019 Wavebook Data

    7/13

    Measurement Computing (508) 946-5100 7 [email protected] mccdaq.com

    The CA-116 optional cigarette-lighter power adapter provides

    DC power to the WaveBook inautomotive applications

    WaveBook/516EGeneral Information

    The DBK34A rechargeable lead-acid battery/UPS (uninterruptible power supply) module

    External Power ModulesDBK30AFor small, portable applications, the optional DBK30A recharge-able battery module provides up to 3.5 hours of operation at2.0A current draw. The DBK30A, which is housed in a ruggedmetal enclosure of the same footprint as the WaveBook, and canbe mounted under the system by using the built-in mountingbrackets. The DBK30A includes an AC charging supply.

    The DBK30A rechargeable battery/excitation module

    DBK34AThe DBK34A provides back-up power as an uninterruptiblepower supply (UPS) to the WaveBook. If DC power is inter-rupted during an acquisition, this module provides temporarypower so the measurement is not disturbed. The DBK34A ispowered from an external DC power supply; it does not includean AC charger.

    The WaveBook/516E includes a 10/100BaseT Ethernet interface

    The WaveBook/516E transfers acquired data to the PC via10/100BaseT Ethernet, allowing a continuous stream of a virtu-ally unlimited amount of data to be collected and stored in aPCs memory or hard drive.

    The 8-channel WaveBook/516E can be expanded up to 72 channelsusing 8-channel WBK expansion options. The WaveBook/516Ealso has 3 built-in parallel expansion ports, permitting connec-tion of up to three additional WBK40 Series units. A sync signalbetween all devices insures that multi-device systems acquiredata synchronously. In total, up to 288 channels of high-speedinput can be measured via one Ethernet link*. Also, additionalchannels are possible using an Ethernet expansion hub, allowingmultiple WaveBook/516E units to be attached to one PC.

    The WBK40 Series of options connect to one of the WaveBook/516Esparallel expansion ports. Since the WBK40 has its own 200 kHzA/D converter, it does not consume bandwidth from the Wave-Books 1 MHz A/D. The SYNC connection insures that both A/Dsmeasure synchronously.

    There are two advantages of seamlessly supporting multipleWaveBooks in one system. First is the ability to expand beyondthe 72 channel capacity of a single WaveBook. Second, if theper-channel sampling rate of a single WaveBook system is inad-equate, then additional WaveBooks can provide more bandwidthper channel.

    * The maximum continuous data transfer rate from a multiple WaveBook systemto the PC on a dedicated Ethernet link is 2 Mreadings/s

    Ethernet Features

  • 8/6/2019 Wavebook Data

    8/13

    Measurement Computing (508) 946-5100 8 [email protected] mccdaq.com

    WaveBook/516EGeneral Information

    SYNC

    Ethernet

    Parallel

    ACTIVE

    READY

    POWER

    DIGITALI/OEXTERNALCLOCK

    TTLTRIGGERPULSETRIGGERCH8CH7CH6CH5CH4CH3CH2CH1

    ANALOGTRIGGER ANALOGCOMMON

    MADEINU.S.A. WBK40 , thermocouple moduleP3

    WaveBook/516Ewith up to eight 8-channel WBK options

    Up to fifteenDBK84 TC expansionmodules

    PC

    WBK40 ExampleThe parallel expansion ports on aWaveBook/516E can also be used to attacha fourteen channel WBK40 thermocoupleinterface module. The WBK40 has an in-ternal A/D converter, so that it does notconsume valuable bandwidth from theWaveBook516Es 1 MHz A/D converter.The SYNC signal on the WaveBook at-taches to the SYNC input on the WBK40,insuring that both A/Ds are operatingsynchronously to one another. The WBK40can be expanded up to 244 thermocouplechannels using DBK84 14- channel TCinterface modules.

    In synchronous master/slave systems orwhen using a WBK40/41 module, it is pos-sible to run the slave units at a slower clockrate than that of the master device in thesystem. This con guration may be neces-sary when it is desirable to have a systemwhich has signals that need to be sampledat widely varying rates while remainingtime synchronous to the master unit.

    Note: The WBK40 and WBK41 are not currently supported in WaveView. We recommend using DASYLabwhen using these modules, which includes full support for the WBK40/41. The WBK40/41 are also supportedin LabVIEW , and DaqCOM. If it is necessary to use these modules in conjunction with a WaveView appli-cation, then it is necessary to launch a second, concurrent application to acquire data from the WBK40/41into separate data les.

    ACTIVEREADY

    POWER

    DIGITALI/OEXTERNAL CLOCK

    TTL TRIGGERPULSE TRIGGERCH8CH7CH6CH5CH4CH3CH2CH1

    ANALOG TRIGGER ANALOGCOMMON

    MADE IN U.S.A.

    WaveBook/516EEthernet

    Notebook PC

    Point-to-point Ethernet connection

    ACTIVREAD

    POWE

    DIGITALI/OEXTERNAL CLOCK

    TTL TRIGGERPULSE TRIGGERCH8CH7CH6CH5CH4CH3CH2CH1

    ANALOG TRIGGER ANALOGCOMMON

    MADE IN U.S.A.

    WaveBook/516E

    ACTIVEREADYPOWER

    DIGITALI/OEXTERNAL CLOCK

    TTL TRIGGERPULSE TRIGGERCH8CH7CH6CH5CH4CH3CH2CH1

    ANALOG TRIGGER ANALOGCOMMON

    MADE IN U.S.A.

    WaveBook/516E

    Enterprise Ethernet Network

    Desktop PCNotebook PC

    Enterprise-wide Ethernet connection

    The most common and highest-performance WaveBook/516Econnection is with dedicated, point-to-point Ethernet linkbetween the PC and the WaveBook/516E. Data transfer rates inthis con guration will accommodate continuous, 1 Mreading/stransfers from the WaveBook/516E to the PC.

    With an enterprise-wide Ethernet network connection, any numberof WaveBook/516Es can be connected to the network, which is sharedamongst a potentially large number of Ethernet-connected devices.In this mode, the data transfer rates from the WaveBook/516E willbe dependent on other network traf c at the time of data transfer.

    This connection method is not recommended when continuous,1 Mreading/s transfers from the WaveBook/516E are required. Toimprove the data transfer performance of WaveBook/516Es in thiscon guration, the WBK30 memory options may be installed intothe WaveBook.

  • 8/6/2019 Wavebook Data

    9/13

    Measurement Computing (508) 946-5100 9 [email protected] mccdaq.com

    Module-to-Module Connection for WaveBook Systems

    Assembling a WaveBook system is easywith our new packaging and module-

    to-module connection system. EveryWaveBook and WBK option is housedin an all-metal enclosure, and is encasedwith rugged molded bumpers on all cor-ners. The bumpers serve to protect theconnectors as well as to attach multiplemodules together. Within each bumperis a tab which can be rotated 90 to lockwith other modules attached to eitherthe top or bottom of each module.

    One handle is included with eachWaveBook, and additional handles canbe purchased for in-vehicle applications

    where a handle on both sides of thesystem is desirable for securing the systemto the vehicle. When multiple modulesare attached in a system, the handle canbe easily moved from the WaveBook toany other module in the system.

    For owners of existing WaveBook and/orWBK systems, the new bumpers can beeasily added to your existing hardware.Contact IOtech or your local IOtechrepresentative for details. Built-in connection tabs in every expansion

    module make assembling a system easy aboveillustrates how a WBK18 would attach to aWaveBook/516E

    An assembled system consisting of a WaveBook/516E plus one WBK18 accelerometer module plus oneWBK16 strain gage module

  • 8/6/2019 Wavebook Data

    10/13

    Measurement Computing (508) 946-5100 10 [email protected] mccdaq.com

    Enhanced features such as lter type, cutoff frequency,IEPE current source, and bridge con guration automaticallyappear when hardware is added to the WaveBook system.There are no switches to set and WaveView automaticallyupdates itself for new con gurations.Click to review system con guration and acquisition pa-rameters such as scan rate, pre- and post-trigger usage, andtrigger criteria.Scope Mode allows customizing and displaying of multipletraces; it supports cursors, re-scaling, and more.Store data to disk in real-time; the auto re-arm function, withautomatic le naming, supports back-to-back acquisitionof over one million captures without user intervention.Review acquired waveforms with a strip chart style displayvia PostView.Strain gage setup to calibrate WBK16 strain gage channels.

    WaveViewWaveView* is a Windows-based setup and acquisition application

    that allows you to con gure, display, and save data to disk withinminutes of taking the WaveBook Out-of-the-Box . WaveView providesa point-and-click interface that simpli es operation of the WaveBookby allowing setup of all hardware, including the eld-installable WBKoptions, without programming or connecting icons.

    Unlike the mere example programs that many suppliers providewith data acquisition hardware, WaveView is a full-featuredacquisition and display engine that provides all the functional-ity needed for many data-logging and display applications. Formore frequency-domain analysis applications, use DASYLab .

    WaveViewOut-of-the-Box Software

    WaveViews intuitive approach to hardware control simpli essystem setup by automatically querying the WaveBook upon

    connection to your PC. As WBK options are added for signalconditioning or increased system channel count, WaveViewschannel con guration spreadsheet automatically expands toaccommodate the additional channels. Speci c channel charac-teristics, such as gain, unipolar/bipolar, and channel labels, areautomatically updated, and any additional functionality (suchas low-pass ltering, lter cutoff, or excitation output), alsoautomatically appear in the channel-con guration spreadsheet.WaveView is also designed for easy operation with display andanalysis packages. WaveView provides data in formats compatiblewith a variety of sophisticated display and analysis packages,including MATLAB .

    Included WaveView is an Out-of-the-Box setup, acquisition, and real-time display program

    Designate an individual data channel as active or inactiveby clicking on a select channel.Reading column provides quick indication of sensor condi-tion before acquisition begins.Choose different ranges or gains for each channel, based onthe particular WBK options installed. Bipolar and unipolarscales can also be selected for each channel.Assign each channel a unique label, which will be auto-matically referenced throughout WaveView. Choose thedesired engineering units in which to display acquireddata, based on the installed WBK options. Parameters canalso be entered to perform mX+b scaling on each readingbefore displaying it.Use auto-zero to remove small offsets such as transducerdrift or pre-load conditions prior to acquisition.

    4

    6

    1 2 3 4 6

    7 8 9

    5

    10

    10

    98

    7

    5

    1

    2

    3

    11

    11

    Note: The WBK40 and WBK41 are notcurrently supported in WaveView. Werecommend using DASYLab whenusing these modules, which includesfull support for the WBK40/41. TheWBK40/41 are also supported inLabVIEW , and DaqCOM. If it isnecessary to use these modules inconjunction with a WaveView appli-cation, then it is necessary to launcha second, concurrent application toacquire data from the WBK40/41 intoseparate data les.

    * Supported Operating Systems: Windows 7/Vista/XP SP2, 32-bit or 64-bit

  • 8/6/2019 Wavebook Data

    11/13

    Measurement Computing (508) 946-5100 11 [email protected] mccdaq.com

    You can con gure the WaveBook via WaveViews scanand trigger con guration screen

    Supports continuous or single-shot capture and displaymodes.

    Provides zoom-in and zoom-out window control of thex-axis.

    Displays user label and cursor data values.

    Scroll through all data.

    Provides auto-scale for quick maximization of the y-axisdisplay.

    Automatically scales axis in user-de ned engineering units.

    Enables cursor for on-screen measurements.

    Scan and Trigger Con guration. The WaveBooks powerfulevent-capture capability is made available through the simple,

    ll-in-the-blank style boxes in WaveView. In addition to single-channel, manual, and external TTL, advanced triggering formultichannel, digital pattern, and pulse trigger is also available.When using multichannel trigger all channels can be combinedwith boolean AND & OR operators to begin the acquisitionat just the right time. In digital pattern mode, the WaveBooktriggers on a user-de ned bit pattern making it easy to associateanalog data with digital sequences.

    Unique to the WaveBook is the pulse trigger capability shownbelow. Typically available in systems many times its price, pulsetrigger allows capturing that elusive event by de ning the level of the signal and pulse width criteria. Now runt pulses or glitchesburied in apparently good signals can be easily captured.

    Scope Mode. Unlike a traditional scope with only two to fourchannels, WaveViews Scope Mode allows any eight channelsto be displayed. Furthermore, WaveView is not handicappedby the small memory limitations of DSOs. In fact, WaveViewdynamically and transparently allocates a PCs RAM prior tobeginning an acquisition. A simple point-and-click is all thatsnecessary to initiate multi-Msample acquisitions. In addition,because the data is already in the PCs RAM, a second point-and-click on the disk icon automatically saves this data to diskfor import into PostView, a post-acquisition waveform reviewpackage, or into analysis packages, such as MATLAB , DADiSPor Excel .

    1

    Allows selection of internal or external clock.Lets you individually set acquisition duration for both pre-and post-trigger data.

    Offers separate scan rates for both pre- and post-trigger data.Select the trigger mode you need: immediate, manual,multichannel, digital pattern, or pulse.

    5

    3

    2

    4

    1

    WaveViewOut-of-the-Box Software

    6

    7

    Scope Mode allows you to display any eight of the WaveBooks channels

    5

    6

    4

    3

    21 7

    3

    2

    1

    4

    2 3

    4

  • 8/6/2019 Wavebook Data

    12/13

    Measurement Computing (508) 946-5100 12 [email protected] mccdaq.com

    Logger Mode. For applications where PC RAM is insuf cientto record the entire test or where rapid back-to-back recordingsneed to be saved to disk, WaveView provides a Logger Mode. Itcomplements WaveViews Scope Mode by allowing continuous*recording data directly to disk.

    Logger Mode can auto-increment le names to provide the unat-tended capture of millions of back-to-back events, without userintervention. Acquired data can be stored in several data formatsfor direct import to packages such as Excel or PostView.

    PostViewPostView is a time-domain post- acquisition data viewing pack-

    age which is integrated and ready to use from within yourView package when you install it. PostView provides easy touse basic time-domain data viewing for IOtech data acquisi-tion Out-of-the-Box View packages.

    WaveCalThe WaveBook is also shipped with WaveCal, an application thatfacilitates periodic calibration of the system. Although all WaveBookcomponents are factory-calibrated to their rated accuracies priorto shipment, annual recalibration is recommended. WaveCalssimple on-screen instructions and direct access to the WaveBookscomponents, make recalibration fast and easy.

    Allows the collection of single-shot or auto-increment lenames for back-to-back acquisitions.File name, path, and auto-increment index can be enteredby simply lling in the blank.Auto increment creates multiple data les, one per trigger.Choose from an ever-growing list of data le formats foreasy importing into other software packages such as DADiSP,DASYLab, MATLAB, Snap-Master, ASCII/Excel Universal FileFormat (UFF) 58A (ASCII), Universal File Format (UFF) 58B(Binary), and .WAV.

    WaveViews Logger Mode can stream data direct todisk in a variety of formats

    WaveViewOut-of-the-Box Software

    PostView, a Windows -based, post-acquisition waveform viewing application

    4

    3

    2

    1

    4

    3

    2

    1

    * For acquisitions in excess of 100 million scans, use DASYLab or other available software

  • 8/6/2019 Wavebook Data

    13/13

    Measurement Computing (508) 946-5100 13 [email protected] mccdaq.com

    General Speci cationsWarm-up: 30 minutes to rated speci cationsEnvironment

    Operating: 0 to 50 C, 0 to 95% RH,non-condensing

    Storage: -20 to 70 CPower Consumption: 1.8A max @ 15 VDCInput Power Range: 10 to 30 VDCVibration: MIL STD 810EPC Communication: 10/100BaseT Ethernet (300 ft. max)Channel Capacity: 8 built-in voltage channels, ex-

    pandable to 72 channels with WBK options. Alsocan accommodate up to 3 additional WBK40 Seriesoptions (any combination). Maximum WBK41capacity is 854 TC input channels, 4 analog outputchannels, 272 digital I/O channels, and 6 counter/timer channels (see WBK40 on for details).

    Dimensions: 285 mm W x 220 mm D x 70 mm H(11 x 8.5 x 2.70)

    Weight: 1.9 kg (4.2 lbs)Handle: One carrying handle is included with each

    WaveBook

    Analog Inputs (18 to 28 C)*Channels: 8 differential, expandable up to 72

    differentialConnector: BNCResolution: 16 bitRanges: Unipolar/bipolar operation is software

    selectable via sequencerUnipolar*: 0 to +10V, 0 to +4V, 0 to +2VBipolar: 10, 5V, 2V, 1V

    Maximum Overvoltage: 35 VDCInput Bandwidth: DC to 500 kHzInput Impedance

    Single-Ended: 5M Ohm in parallel with 30 pFDifferential: 10M Ohm in parallel with 30 pF

    Accuracy2 to 10V: 0.012% of reading; 0.006% of range

    1V: 0.018% of reading; 0.008% of rangeInput Noise:


Recommended