ww
Edan USA, 9823 Paci� c Heights Blvd., Suite E, San Diego, CA 92121 www.edanusa.com
W E C A R E A B O U T Y O U R H E A L T H !
Call For More Info
888.850.4597
Quick, accurate and easy to use
Real-time data or USB data can be transferred to a PC through PC mange-ment software to review and print
Respironics Lo� o sidestream and CAPNOSTAT 5 ETCO2 mainstream measurement for intubated and non-intubated patients(M3B)
Powerful storage capacity: 72 hours trend review of all parameters, 30,000 sets NIBP review, 800 items alarm review and USB data storage
M3 Vital Signs Monitor
Edan M3 Vital Signs Monitor has made its mark in out-patient department and doctor’s o� ce for its accuracy, durability and cost-e� ectiveness by SpO2, NIBP and Fast TEMP monitoring. Its a� ordable price and multi-parameter functionality address vital signs monitoring needs.
PATIENT MONITORING
FEATURES• 5.7 inch high resolution display for easy reading• Lightweight, portable design and user-friendly interface for easy operation• SpO2, Pulse Rate, NIBP and Fast TEMP measurement• Nellcor OxiMax SpO2• Fast Temp thermometer to obtain temperature in approximately 15 seconds• Display numeric and waveform information simultaneously • Nurse call• Powerful storage capacity• Built-in Lithium-ion Battery for 8 hours working• Suitable for adult, pediatric and neonatal patients• PR measurement (form SpO2/ NIBP)• Trend table review and record• Trend graph review and record• USB data storage and review• Wired and wireless network capability
Con� guration• M3B (SpO2 + CO2)
Call For More Info
888.850.4597Edan USA, 9823 Paci� c Heights Blvd., Suite E, San Diego, CA 92121 www.edanusa.com
W E C A R E A B O U T Y O U R H E A L T H !
Classi� cationAnti-electroshock typeClass I equipmentand internal powered equipmentEMC typeClass AAnti-electroshock degreeSpO2, NIBP: BF Defi brillation type;TEMP: CF type.Ingress Protection IPX1 (W/O Temp module)IPX1 (with Temp module)
Speci� cationsSize and WeightSize200.8mm (L) x 241mm (H) x 189 mm (D)Weight3 kgDisplay5.7 inches,LCDColor TFT resolution: 640X480Power Supply100-240 VAC, 50/60HZPmax=70VA FUSE T 1.6ALBatteryType: Lithium-ionVoltage:14.8 V DC Capacitance:4,400 mAhWorking period Color TFT: 480minRechargeable period <360minRecorder (Optional)Record Width 48 mmPaper Speed 25mm/s
NIBPMethod OscillometricMode Manual, Auto, ContinuousMeasuring Interval in AUTO Mode1/2/3/4/5/10/15/30/60/90/120/240/480 MinContinuous 5min, interval is 5sMeasuring Type Systolic Pressure,Diastolic Pressure, Mean PressureMeasuring RangeAdult ModeSYS 40~270mmHgDIA 10~215mmHgMAP 20~235mmHgPediatric ModeSYS 40~200mmHgDIA 10~150mmHgMAP 20~165mmHg
Neonatal ModeSYS 40~135mmHgDIA 10~100mmHgMAP 20~110mmHgCuff Pressure measuring Range 0~290mmHgPressure Resolution 1mmHgMaximun mean error 5mmHgMaximum Standard deviation 8mmHgEntire Measuring Period 30~45s typical(depend on HR/motion disturbance)Dual Overpressure protectionAdult 297 3mmHgPediatric 240 3mmHgNeonatal 145 3mmHgPRMeasuring Range 40~240bpmResolution 1bpmAccuracy 3bpm or 3.5% of the maximumIEC 60601-2-30
SpO2 (EDAN)Measuring Range 0 ~ 100 %Alarm Range 0 ~ 100 %Resolution 1 %AccuracyAdult (including Pediatric)2% 70%~100% SpO2Undefi ned 0~70% SpO2Neonate 3% 70%~100% SpO2Undefi ned 0~70% SpO2Pulse RateMeasuring and Alarm Range 30 ~ 300bpmResolution 1bpmAccuracy 3bpmData update period 2sISO 9919
SpO2 (optional, by Nellcor OxiMax )Measuring Range 1 ~ 100 % Alarm Range 1 ~ 100 %Resolution 1 % AccuracyAdult and Low-perfusion2% 70%~100% SpO2
Undefi ned 0~70% SpO2Neonate 3% 70%~100% SpO2Undefi ned 0~70% SpO2Pulse RateMeasuring and Alarm Range 20~300bpmResolution 1bpmAccuracy 3bpmQuick TemperatureMeasuring Range 25 ~ 45Probe Type Oral/Axillary sensorRectal sensorResolution 0.1Accuracy Monitor mode: 0.1Typical measurement time <15sUpdate time 1s ~ 2sIEC 12470-4
Respironics CO2 (M3B only)CO2 (Mainstream and Sidestream, optional)By Philips Respronics CAPNOSTAT 5 & LoFloTechnologyRange: 0 ~ 150 mmHgAccuracy: 2% 0 ~ 40 mmHg5% 41 ~ 70 mmHg8% 71 ~ 100 mmHg10% 101 ~ 150 mmHgAwRR Accuracy: 1rpmConvenient design for intubated and non-intubatedapplicationsPossible to work at low sample fl ow rate: 50ml/minDetailed specifi cation refer to the user manual ofRespironicsISO 21647
PATIENT MONITORINGM3 Vital Signs Monitor Technical Speci� cations