05/13/2015 Products Solutions Services Welcome to Heartbeat Technology! Directive 17 Compliant, TUV Certified! Technology Day – Calgary – May 2015 Heartbeat Technology Slide 1 Jason Riegert
Transcript
Page 1: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Products Solutions Services

Welcome to Heartbeat Technology!

Directive 17 Compliant, TUV Certified! Technology Day – Calgary – May 2015

Heartbeat Technology

Slide 1 Jason Riegert

Page 2: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Addressing Your Metering Challenges Measurement & Compliance Manager:

“How do I ensure our AER D17 requirements are met?”

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Maintenance Manager:

“Is my meter working properly?”

Process Engineer:

“Is my process running as designed?”

Availability

Quality/Compliance

Safety

Slide 2 Jason Riegert

Page 3: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Why Check Flow Meters?

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 3 Jason Riegert

Page 4: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Traditional Meters May Lose Accuracy

Turbine meters: • Over spinning damage • Worn or broken blades

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

PD Meters: • Debris in line • Entrained gases • Slippage due to viscosity

Differential Pressure: • Worn edges of element • Element deformed from

excess pressure

Slide 4 Jason Riegert

Page 5: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Intelligent Field Devices – Today!

Why keep sending the meter back for re-calibration? - Most meters returned for re-calibration do not require adjustment In fact only Endress+Hauser Vortex meters have a lifetime Calibration!

Coriolis Vortex Magmeter Ultrasonic

Slide 5 Jason Riegert

Page 6: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Heartbeat Technology™ – The Universal Solution

For all Endress+Hauser flow measurement technologies • Heartbeat Technology is at the heart of the new Proline

flowmeter platform, ensuring reliable measurement operation – whatever the flow measurement technology

Coriolis Vortex Magmeter Ultrasonic

Slide 6 Jason Riegert

Page 7: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ - The Elements

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Heartbeat Verification

Heartbeat Diagnostics

Heartbeat Monitoring

• ------------------ • ------------------ • ------------------ • ------------------ • ------------------

Pass / Fail

Heartbeat Technology™

Slide 7 Jason Riegert

Page 8: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Monitoring Verification Diagnostics

At a glance: What Heartbeat Technology can achieve

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Reliable self-monitoring • Continuous, built-in

diagnostic functionality • Broadest test coverage • Quick, concise guidance

in case of errors (NE 107)

Compliant testing • On-demand, effortless

procedure • Easy handling,

no process interruption • Electronic quality reporting

Condition-based maintenance • Permanent reading of

process-related variables • Impact assessment • External trend analysis

Slide 8 Jason Riegert

Page 9: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Diagnostics™ - Standard with Proline

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Heartbeat Diagnostics

Check Failure OK

Flowmeter health information

• Process safety • Product quality • Operation &

Maintenance

supports

Slide 9 Jason Riegert

Page 10: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Diagnostics™ FieldCare DTM – Conforms to Namur NE 107

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 10 Jason Riegert

Page 11: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Why do YOU get YOUR Gas meters verified?

D-17 Gas Meters • 2.5 • 2.5.2.1 • Exception #9 • Internal metering diagnostics may be used to determine if the structural

integrity of the primary measurement element is within acceptable operating parameters and checked at the same required intervals as an internal inspection. Then internal inspection is not required until an alarm or error is generated by the device or as recommended by the manufacturer. The operator must maintain documentation on the diagnostic capability of the measurement system and make it available to the ERCB on request. An initial baseline diagnostic profile must be performed and documented during the commissioning process.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 11 Jason Riegert

Page 12: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Why do YOU get YOUR Liquid meters verified?

D-17 Liquid Meters • 2.6 • 2.6.1 • Exceptions • 1) If a meter used to measure fluids at flow-line conditions is a type

that uses no internal moving parts (e.g., orifice meter, vortex meter, cone meter, Coriolis meter, ultrasonic meter), it does not require proving, provided that all the following conditions are met:

• The flow through the meter must be continuous (not intermittent) or the meter must qualify for bench proving or be a Coriolis-type meter with sufficient structural integrity internal diagnostics (see below) and maintained within the rates specified by the meter manufacturer as providing accurate measurement.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 12 Jason Riegert

Page 13: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Why do YOU get YOUR MARP Meters verified?

Heavy Thermal Oil

• 12.3.4 • All steam measurement devices must be calibrated/proved/verified on an

annual basis or as otherwise stated in Section 2.5 or above.

• 12.3.5 • All water measurement devices must be calibrated/proved/verified on an

annual basis or as otherwise stated in Sections 2.6 and 2.10.

• 12.3.7 • All solvent injection measurement devices must be

calibrated/proved/verified on an annual basis in accordance with Section 2.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 13 Jason Riegert

Page 14: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Now really why do YOU get YOUR meters verified?

• These meters are how royalties are calculated and resources are conserved or consumed.

• If you’re over reporting this is a financial loss!

• If you’re under reporting you might have to go back one day and make adjustments which could become a financial loss!

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 14 Jason Riegert

Page 15: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Terminology: Accredited, Calibration, Proving, Verification • Accreditation demonstrates technical competence for a defined scope and

operation of a laboratory quality management system through 3rd party audit compliance and traceable calibration. Result: Quality Calibration

• Calibration is a performance comparison of a meter under test and a

standard under laboratory conditions. Result: Calibration Factor

• Proving is a performance comparison of a meter under test and a standard

under actual process conditions to account for installation and actual process fluid effects. Result: Meter Factor

• Verification is a functional test comparing a meter under test to a known

standard to determine quality of measurement which may be traceable to laboratory calibration. Used to extend Calibration intervals. Result: Pass / Fail indicating that the Calibration is still valid.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 15 Jason Riegert

Page 16: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Endress+Hauser Portfolio of testing methods

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

On-site In Laboratory

On-site

Heartbeat Verification

Fieldcheck (not for new Proline)

Mobile Bench Prover

Fixed Bench Prover

(Edmonton 2015/2016)

Metrological performance assessment

(adjustment via calibration factor, certificate)

• Primary measurand () • Reference conditions

(fluid, p, T)

Flow-meter functional assessment

(comparison still vs. traceable references)

• Internal parameters • Operating conditions

Verification Calibration

Slide 16 Jason Riegert

Page 17: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Uncontested benefits: • Quick and reliable testing • Flowmeter remains in place • Flowmeter calibration data manually

inputted • Traceability of all components • Good proof-test coverage Improvement areas:

• Cumbersome equipment • Manual reports • Mandatory trip to the field • Disconnected output,

interruption of measurement • Potential handling errors

Flowjack

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Flowjack: First signal simulator for entire flow portfolio

Slide 17 Jason Riegert

Page 18: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Uncontested benefits: • Quick and reliable testing • Flowmeter remains in place • Flowmeter calibration data

automatically downloaded • Automated report • Traceability of all components • Excellent proof-test coverage Improvement areas:

• Cumbersome equipment • Mandatory trip to the field • Disconnected output,

interruption of measurement • Potential handling errors

Fieldcheck TM

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Fieldcheck: Recognized method (signal simulation)

Slide 18 Jason Riegert

Page 19: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Products Solutions Services

Lets Perform a Heartbeat Verification Now!

Slide 19 Jason Riegert

Page 20: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Heartbeat Verification to succeed Fieldcheck

Similar functionality, now embedded in the device electronics • Permanently available, no mandatory accessories • Accessible via all available interfaces • Testing of the entire signal chain without service interrupt • Simplistic handling, clear / automated procedure • Verification results stored in the device (8 most recent datasets) • FieldCare© PC can be connected, so the verification DTM can upload

results, generate a report, and manage the entire verification history

Heartbeat Technology

Slide 20 Jason Riegert

Page 21: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification vs. Fieldcheck

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Heartbeat Verification Fieldcheck • Built-in, with remote options

• Embedded test functionality

• App. 10 minutes

• Total test coverage* TTC > 95% acc. to IEC 61508

• External

• Simulation and sensor checking

• 30 to 60 minutes

• Proof-test coverage 90% or better

*Cumulative rating of safe failure across entire device electronics

Slide 21 Jason Riegert

Page 22: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Heartbeat Verification™ - Independently Certified by TUV

Detection by Heartbeat Verification™

Continuously covered by Heartbeat Diagnostics™

Verification Proof-Test λDU -Dangerous Undetected Failure λDD -Dangerous Detected Failure λSD -Safe Detected Failure λSU -Safe Undetected Failure SFF – Safe Failure Fraction including Verification is upto 98%.

Slide 22 Jason Riegert

Page 23: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification™ - TUV Certified, Directive 17 Compliant!

Due to our Safety-by-Design philosophy Endress + Hauser flow meters that have Heartbeat Technology™ are independently certified by TUV and comply with the requirements of AER Directive 17.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 23 Jason Riegert

Page 25: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat addresses your most serious challenges

Availability Quality Safety

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 25 Jason Riegert

Page 26: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Challenges in the Process Industry …

Maintenance Costs: High costs and constraints

for preventive maintenance

D17 Compliance: Keep up trust in the

integrity of measurements

Erosion inside the measuring tube of a

Coriolis flowmeter

?

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 26 Jason Riegert

Page 27: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

What Heartbeat Technology can do …

Heartbeat Diagnostics: Greatly reduced risk of failure

Heartbeat Verification: Clarity and evidence without

process interruption

Heartbeat Monitoring of process-related

parameters

PDF

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 27 Jason Riegert

Page 28: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

What‘s in it for you …

Full flexibility to schedule proof-tests,

enables cost savings

Proof of functional reliability, no discussion about the

verification process Informed decisions about flowmeter maintenance

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 28 Jason Riegert

Page 29: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology – what makes it unique

Uniform handling Seamless integration Reliability

For all new Proline flowmeters

Audited and attested: Traceable, redundant

references inside the device

Large variety of seamless integration

options

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 29 Jason Riegert

Page 30: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ – reliable self-monitoring

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

• Ex-factory status is stored invariably in the device • Each single test is passed only if the current parameter value remains within

tight, unchangeable tolerances supporting the device specification

Long-term stability • Measuring electronics • Test electronics

Redundant internal references • Dual frequency generators

(Promass, Prowirl) • Multiple voltage references (Promag)

Traceable factory calibration according to ISO 17025 on accredited facilities

• Flow measurement (, ) • Electronic modules

IEC 61508 audits completed

ISO 9001 conformity certification, audits completed on Proline 200

Slide 30 Jason Riegert

Page 31: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification – onsite access

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

SIL loop integrated via 4..20mA HART

Control cabinet

FXA 291 Asset management tool, e.g. FieldCare

Communication interface

Slide 31 Jason Riegert

Page 32: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification – access via HART modem

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

SIL loop integrated via 4..20mA HART

Control cabinet

Asset management tool, e.g. FieldCare

HART Modem

Slide 32 Jason Riegert

Page 33: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification – access via Foundation Fieldbus

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Honeywell DCS environment

FF function block MDO • Start Verification

FF function block DI • PASS: Value 1 • FAIL: Value 0

Slide 33 Jason Riegert

Page 34: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification – access via Foundation Fieldbus

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Honeywell DCS environment

Asset management tool • Field Device Manager • FieldCare

Slide 34 Jason Riegert

Page 35: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ fundamentals

Long-term stability Redundant internal references Traceable factory calibration

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Electronic modules • Designed to minimize risk of failure • 100% testing against traceable

references

Completed device • Accredited calibration • Primary measurand (, )

Slide 35 Jason Riegert

Page 36: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Total Test Coverage from sensor to outputs

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Power Supply

Temperature- sensors

Exciter coil

Pick-up coil

ISEM, CPU

I/O Modul

Example: Proline Promass F 200

Flowmeter development according to IEC 61508 Continuous self-monitoring based on NE 107

Diagnostic messages securely stored in the event logbook

Slide 36 Jason Riegert

Page 37: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Proline Promass: Overview of test groups

Syst

em

(AM

S. P

LC)

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

I/O module • 4…20 mA output loop-back test Sensor electronic module • Zero-point tracking • Reference clock • Reference temperature Sensor • Inlet / Outlet pick-up coil • Measuring / Carrier tube temperatures • Pick-up coil symmetry • Frequency lateral / torsion mode Sensor Integrity (HBSI) • Electromechanical integrity of Promass

Proline Promass 100 / 300 / 500

Continuous

On demand

Slide 37 Jason Riegert

Page 38: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Traceable internal references for measurement + testing

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

High-precision frequency generators • Continuous synchronous-run test in the device reliably detects ambient

and process influences potentially causing drift in the electronics • Traceable test of each module against accredited reference standard prior

to traceable factory calibration of the entire device

20 MHz Quartz

20 MHz Quartz

Impulse count in given time window <0.01%

ISEM

Slide 38 Jason Riegert

Page 39: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Current output loop-back test

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

• Drift and aging of electronics • Internal connectivity problems • External influences (load resistance)

4...20 mA I/O module

CPU

1% + 100μA

Proline Promass 100 example

Slide 39 Jason Riegert

Page 40: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Verification – electronic Pass / Fail report

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Promass F 200

Slide 40 Jason Riegert

Page 41: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Verification DTM for FieldCare© asset management tool

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 41 Jason Riegert

Page 42: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Products Solutions Services

Lets look at our Heartbeat Verification Report!

Slide 42 Jason Riegert

Page 43: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Proline Prowirl 200: Summary of test groups

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Main Electronic Module

• Supply Voltage

I/O module • 4…20 mA output loop-back test Sensor electronic module • Reference frequency drift • DSC sensor reference • Temperature measuring path • DSC sensor measuring path Sensor • DSC Sensor • Temperature Sensor

Proline Prowirl 200

On demand

Only Prowirl C (Canada) has inspection ports for shedder bar inspection

Slide 43 Jason Riegert

Page 44: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

DSC sensor: Gap capacity difference test

• Capacitive sensor to count the vortex frequency

• Warning level yields a “Fail” in the test, but still allows to measure

• Small impact on measurement accuracy, strong indicator for device functionality

Level dC in pF

Normal 0 to < 4.6

Warning 4.6 to 6.8

Failure > 6.8

dC = | C2 – C1 | C1 C2

Slide 44 Jason Riegert

Page 45: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Proline Promag: Summary of test groups

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Main Electronic Module

• Supply Voltage

I/O module • 4…20 mA output loop-back test Sensor electronic module • Reference voltage • Linearity of electrode measuring circuit • Offset of electrode measuring circuit Sensor • Coil current shot time • Coil hold voltage • Coil current

Proline Promag 100/200/400

On demand

Slide 45 Jason Riegert

Page 46: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Proline Promass condition monitoring applications

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Interpretation requires sound understanding of the measuring point and the process influence on the oscillation system of the specific Coriolis device type

Types of process impact Relevant parameters

Entrained gas, cavitation, as well as empty tube

• Tube damping • Damping fluctuation • Frequency fluctuation

Coating build-up, corrosion, erosion

• Tube damping • Signal asymmetry • Resonance frequency

Slide 46 Jason Riegert

Page 47: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Monitoring for entrained gas

Entrained gas is a challenge in Coriolis measurement in general. In can therefore be useful to monitor its impact below the limit values of Heartbeat Diagnostics

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Tube damping =

Base value

Tolerated e.g. up to factor 30

Tube damping is also a parameter of continuous diagnostics, and leads to a

warning above a threshold of 200,000 A/m. Damping fluctuation and frequency

fluctuation are powerful indicators of cavitation and entrained gas as well.

Diagnostic threshold Critical peaks: factor 100 to 1000 Exciter current

Amplitude

Slide 47 Jason Riegert

Page 48: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Monitoring: Coating build-up in the tube

If the fluid’s density and viscosity are stable, coating build-up can be detected by decreasing frequency, and concurrent increase in tube damping or signal asymmetry.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Tube damping Signal asymmetry Resonance frequency

Base value

e.g. +10%

e.g. - 1%

The density measurement can be tracked in addition, if the

fluid is unchanged.

e.g. +10%

Slide 48 Jason Riegert

Page 49: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Monitoring: Corrosion

Corrosion may occur in manifold forms, so condition monitoring requires a variety of parameters, particularly here.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Base values

Gradual, steady drift

Tube damping Signal asymmetry Resonance frequency

The density measurement can be tracked in addition, if the

fluid is unchanged.

Slide 49 Jason Riegert

Page 50: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Monitoring: Abrasion

Abrasion particularly impacts the curved sections inside the tube. Promass 100 offers the HBSI test, Promass 200 does not!

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Signal asymmetry (Inlet / outlet amplitude ratio) HBSI: Heartbeat Sensor Integrity

Base values

HBSI: Warning at 2%, failure at 4%

Tolerance 2% / 4%

Tolerance 25% Promass I 100 • Straight / Titanium

mono-tube, suitable in abrasion use cases

• Only Promass I offers HBSI as a continuous monitoring parameter!

Slide 50 Jason Riegert

Page 51: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Heartbeat Technology – Technology beating reliably!

Availability Quality Safety

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 51 Jason Riegert

Page 52: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Products Solutions Services

Questions?

We have included some of your questions and answers in the next slides

Slide 52 Jason Riegert

Page 53: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Questions & Answers #1 Question • On slide 21 the Presenter

mentions that the Test Coverage is across the entire device electronics, does that mean only the electronics are tested by Heartbeat Verification.

Answer • TUV certifies the device’s ability

to identify a fault and all the software and tests are carried out in the electronics. This means while we are testing the sensor integrity it is the electronics that tests and verifies the sensor and ultimately generates a fault. TUV is interested in the ability for the device to identify and annunciate a fault safely meaning the fault does not go undetected.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 53 Jason Riegert

Page 54: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Questions & Answers #2 Question • On slide 8 the Presenter

mentioned that the only 4 wire devices would have Heartbeat Monitoring®.

Answer • All Proline 3 devices (2 & 4 wire)

come with Heartbeat Diagnostics® which is a nearly full test coverage of the complete device! Optionally Heartbeat Verification® is an on-demand test that takes the test coverage up to 98%. While Heartbeat Monitoring® allows you to monitor in to your DCS, continuously in real-time device parameters that normally would only be tested on-demand (ie. Heartbeat Verification®) and this can only be done with the 4 wire devices. This is just a matter of power consumption or availability.

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 54 Jason Riegert

Page 55: Welcome to Heartbeat Technology! - Primary Target Mediaendress.primarytargetmedia.com/techdays/heartbeat...•External trend analysis Slide 8 Jason Riegert 05/13/2015 Heartbeat Diagnostics

05/13/2015

Questions & Answers #3 Question • Why do you call the

presentation TUV Certified, Directive 17 Compliant!

Answer • Heartbeat Verification® is an integral

part of the overall proof test for these Endress+Hauser SIL rated devices.

• TUV is an independent 3rd party that has tested and certified these devices to have a high availability and safe failure meaning if a fault was to occur the device identifies it before a dangerous situation occurs.

• The SIL device and non SIL device uses all the same components.

• These devices are Directive 17 compliant even though AER doesn’t endorse, register or certify devices so you have Certification and Compliance with Endress+Hauser!

Heartbeat Technology™ - TUV Certified, Directive 17 Compliant!

Slide 55 Jason Riegert