+ All Categories
Home > Documents > Wetstrilization.ppt

Wetstrilization.ppt

Date post: 14-Apr-2018
Category:
Upload: antonygamalpharma
View: 214 times
Download: 0 times
Share this document with a friend

of 22

Transcript
  • 7/27/2019 Wetstrilization.ppt

    1/22

    2013-10-22 Moist Heat Sterilization 1

    VALIDATION OF

    MOIST HEAT STEILIZATION

    JM Tech.

    Do-Young Ahn

  • 7/27/2019 Wetstrilization.ppt

    2/22

    2013-10-22 Moist Heat Sterilization 2

    Definition

    SterilizationThe act or process, physical or chemical, that destroys or eliminates

    all viable microbes including resistant bacterial spores from a fluid or a

    solid.Examples of sterilization methods are : steam treatment at 121, dry

    heat at 230, flushing with a sterilizing solution such as hydrogen

    peroxide (H2O2) or ozone (O3), irradiation, and filtration.

    SterilityThe reduction of anticipated levels of contamination in a load to the

    point where the probability of survival is less than 10-6.

  • 7/27/2019 Wetstrilization.ppt

    3/22

    2013-10-22 Moist Heat Sterilization 3

    Definition

    D-valueThe time in minutes required for a one-log or 90% reduction of a

    specific microbial population under specified lethal conditions. For

    steam sterilization it is determined at a constant temperature

    z-valueThe number of degree of temperature change necessary to change the

    D-value by a factor of 10.

  • 7/27/2019 Wetstrilization.ppt

    4/22

    2013-10-22 Moist Heat Sterilization 4

    Definition

    F value(lethal rate, instantaneous Fo)The F value is a measurement of sterilization effectiveness. F(T,z) is

    defined as the equivalent time at temperature T delivered to a container

    or unit of product for the purpose of sterilization, calculated using aspecific value of z.

    Fo value(accumulated Fo)The term "Fo " is defined as the number of equivalent minutes of

    steam sterilization at temperature 121.1C delivered to a container orunit of product calculated using a z-value of 10C.

    Fo = 10^((121-T)/z)*t

  • 7/27/2019 Wetstrilization.ppt

    5/22

    2013-10-22 Moist Heat Sterilization 5

    Methodology

    Overkill SterilizationProvides a minimum 12 log reduction of a resistant BI w/ a known

    D-value of not less than 1 minute.

    Required minimal information on the bioburden

    Bioburden/Bioindicator Sterilization

    Provides a probability of survival of less than 1 in 106 for the

    bioburden as demonstrated using a resistant BI w/ a known D-value.

    BI may not be inactivated

    Requires information on the numbers and heat resistance of the BI.

    Requires ongoing monitoring or control over bioburden.

  • 7/27/2019 Wetstrilization.ppt

    6/22

    2013-10-22 Moist Heat Sterilization 6

    Methodology

    Bioburden Sterilization

    Provides a probability of survival of less than 1 in 106 for the most

    resistance bioburden expected in the load.

    Requires information on the numbers and heat resistance of the BI.

    Requires ongoing monitoring or control over bioburden.

  • 7/27/2019 Wetstrilization.ppt

    7/22

    2013-10-22 Moist Heat Sterilization 7

    Sterilizer Cycle

    Gravity DisplacementDifference of density

    Density of air at 20 = 1.2 g/

    Density of steam at 100 = 0.6 g/

    Effectiveness of air elimination depends on the rate of steam supply

    Air pocket : too rapidly

    Diffusion into the steam : too slowly, more difficult to remove

    Specially designed steam trap permitting the passage of largevolume of air

  • 7/27/2019 Wetstrilization.ppt

    8/22

    2013-10-22 Moist Heat Sterilization 8

    Sterilizer Cycle

    Prevacuum cycleA more effective method

    By means of a mechanical vacuum pump or a steam eductor

    Vacuum as low as 15~20 mmHg, apply for 8~10 min.

    Pulsing cycleA series of alternating steam pulses followed by vacuum excursions

    Air-steam mixtureTerminal sterilization of large volume parenterals

    Air injection required to compensate the great expansion of air or

    nitrogen in the head space above the liquid

    Well mixed chamber : fan, raining effect by external pump w/ cooling

  • 7/27/2019 Wetstrilization.ppt

    9/22

    2013-10-22 Moist Heat Sterilization 9

    Cycle Development

    Consider factors into accountNature of the load : porous materials, heat sensitivity of the products

    Type of the sterilizer

    Employed containers and closures

    Heat stable product : overkill approach

    Heat liable product : bioburden approach

    Bioburden studies : number of microorganisms

    D-value studies : only highly resistant spore formers,BIER(biological indicator evaluator resistometer)

    Inoculate the spore into the actual solutions

    For solid materials, precut strips

  • 7/27/2019 Wetstrilization.ppt

    10/22

    2013-10-22 Moist Heat Sterilization 10

    Preparing for Validation

    Temperature sensing devices :

    T type thermocouples(copper-constantan) encased in flexible sheaths

    Premium grades of wire having 0.1 accuracy

    Temperature standards

    RTD traceable to the National Bureau of Standards , IPR, HTR

    Calibration of thermocouples

    At two temperatures : 0, 130

    Correction factors

    Stability : 0.03

    Accuracy : 0.5

  • 7/27/2019 Wetstrilization.ppt

    11/22

    2013-10-22 Moist Heat Sterilization 11

    Preparing for Validation

    Autoclave

    Validation nozzle and adaptor

    Data logger : digital output and multi-channel deviceBIs or biological challenges

    Loads

  • 7/27/2019 Wetstrilization.ppt

    12/22

    2013-10-22 Moist Heat Sterilization 12

    Validation Protocol

    Protocol should include

    Objectives of the validation

    Responsibilities of validation personnel and operatingdepartment personnel

    Identification and description of the sterilizer and its

    process control

    Identification of SOPs :equipmentCalibration of instrument : SOPs and/or description

    Identification and calibration of the temperature monitoring

    equipment

  • 7/27/2019 Wetstrilization.ppt

    13/22

    2013-10-22 Moist Heat Sterilization 13

    Validation Protocol

    A description of the following studies

    Bioburden determination studies(if applicable)

    Empty chamber heat distribution studies

    Container mapping studies(if applicable)Loaded chamber heat penetration studies

    Microbiological challenge studies

    Evaluation of drug product cooling water(if applicable)

    Integrity testing of vent filter

    Acceptance criteria

    References

    Review and approval

  • 7/27/2019 Wetstrilization.ppt

    14/22

    2013-10-22 Moist Heat Sterilization 14

    Heat Distribution Studies

    To demonstrate the temperature uniformity and stability of the

    sterilizing medium throughout the sterilizer

    Conduct on both the empty and loaded chamber with max. and

    min. load configurations

    Acceptance criteria : Less than 1of the mean temperature

    Conduct 3 runs to obtain consistent results

    Distribution of the thermocouples : geometricalrepresentatives, exhaust drain, adjacent to the control sensor

    At least 10 probes, normally 15~20 probes

  • 7/27/2019 Wetstrilization.ppt

    15/22

    2013-10-22 Moist Heat Sterilization 15

    Heat Distribution Studies

    At loaded chamber heat distribution test, the thermocouples

    should be positioned in the same locations used for empty

    chamber heat distribution

    Avoid contacting solid surfaces

    Do not place within any containers

    Data should be obtained at regular intervals

  • 7/27/2019 Wetstrilization.ppt

    16/22

    2013-10-22 Moist Heat Sterilization 16

    Container Mapping

    To determine the coolest point within the liquid filled

    container

    Temperature mapping should be conducted on all the different

    container types, sizes and fill volume to be validated

    The number of the thermocouples used depends on the

    container volume

    Possible to use a single thermocouple at different positions,

    and can be conducted in a smaller autoclave or retort

    Penetration thermocouples should be positioned at the cold

    spothaving lowest temperature or Fo

  • 7/27/2019 Wetstrilization.ppt

    17/22

    2013-10-22 Moist Heat Sterilization 17

    Heat Penetration Studies

    To determine the coolest point(s) within the specified load

    and configuration, and to assure that these points be

    consistently exposed to sufficient heat lethality

    Prior to conduct heat penetration studies, determine max.

    and min. load configurations

    Probed container at the cold spot should be distributed

    uniformly throughout the load

    Penetration thermocouple are positioned at points within

    the process equipment suspected to be the most difficult

    for steam heat penetration

  • 7/27/2019 Wetstrilization.ppt

    18/22

    2013-10-22 Moist Heat Sterilization 18

    Heat Penetration Studies

    Lethal rate can be determined from the temperature data

    by the following formula :

    L = log-1

    (To-Tb)/z = 10^((To-Tb)/z)A summation over time of the lethal rate at a series of

    temperature(accumulated lethality)

    Fo = 10^((121-T)/z)*t

    Regard to product stability

  • 7/27/2019 Wetstrilization.ppt

    19/22

    2013-10-22 Moist Heat Sterilization 19

    Microbial Challenge Studies

    Biological challenges are employed during heat penetration

    studies in order to demonstrate the degree of process

    lethality provided by the sterilization cycle

    Microorganism frequently utilizedOverkill :Bacillus stearothermophilus and Clostridium sporogenes

    Bioburden : Calibrated BIs from environmental and process

    isolates such asE. coli

    Type of BI :Spore strips or spore suspension into the suspending medium

    Microbial challenge studies are conducted concurrently with

    the heat penetration studies

  • 7/27/2019 Wetstrilization.ppt

    20/22

    2013-10-22 Moist Heat Sterilization 20

    Validation Report

    Common elements of all reports :

    Identification of the task report by number

    Reference to protocol

    A brief summary of the range of operational conditionsexperienced and how they were controlled

    A procedure for maintaining control within the approved range

    A summary and analysis of the experimental results

    A brief description of any deviation

    Conclusion

    Review and approval

    Cycle development reports are not usually a part of the

    validation report

  • 7/27/2019 Wetstrilization.ppt

    21/22

    2013-10-22 Moist Heat Sterilization 21

    Maintenance of Validation

    A routine calibration program for all instruments critical to the

    operation of the sterilizer and its support system

    A preventative maintenance program including periodic

    operational rechecks and comparison to OQ record

    Routine monitoring of bioburden and periodic BI

    challenges(optionally)

    Operating records and equipment logs

    Process and equipment change control procedures including

    review to establish whether additional validations are required

    On-going validation

  • 7/27/2019 Wetstrilization.ppt

    22/22

    2013-10-22 Moist Heat Sterilization 22

    Controversial Issues

    Incubation of the sterility test : 7 days vs. 14 days

    USP provide information concerning critical parameters for

    Parameteric Release

    Reduction extent and frequency of revalidation

    Verification of D-value of BIs

    Use of alternative toB. stearothermophilus as a BI