What are the dose quantities in CT ?
IAEA/RCA Kampala
Kampala
12345678
12
3
45
6
7
8
CT Principles
Kampala
Kampala
Reminder : CT terminology • Axial scanning
– 1 rotation of beam followed by stepwise couch movement
– Increment – distance moved by couch• Helical/spiral scanning
– Continuous table movement & beam rotation– Pitch – table movement relative to nominal
slice width
Kampala
CT Slice
Kampala
CT Slice
Kampala
CT Slice
Kampala
Dose Profile
Kampala
Dose profile for a series of slices
Kampala
Dose Profiles in Phantoms / PatientsEffects of scatter
Broadened profilePeak kerma dependent on beam widthLong scatter tails for wide beams
Kampala
Which Dosemeter ?
Kampala
Integrated Profile
Kampala
Integrated Profile
Kampala
CTDose Index
• Measure of dose for a single axial slice (mGy)– no account of Helical/MDCT gaps or overlaps
• CTDI = area under dose profile (mGy.mm) nominal slice width, T (mm)
Kampala
kerma
T
CTDI(mGy
)
Z-axis (mm)
CTDI = height of rectangle if area
under profile was all contained
within nominal slice width
CTDI100 – What is it ?• Analagous to
Multiple Scan Average Dose.
• Originally defined for single slice scanners.
Kampala
• In air : CTDI100,Air
• In air : CTDI100,Air
• Sometimes normalise to 1 mAs:
• nCTDI100,Air
• Just like in radiography!
Kampala
In Perspex / PMMA
Kampala
CT dosimetry in PerspexThe computed tomography air kerma index is also measured inside PMMA head and body phantoms
The quantity CTDI100,c is measured at the centre of the standard CT dosimetry phantom and CTDI100,p is the average of values measured at four positions around the periphery of the same phantom.
Kampala
CTDI100
• CTDI can be measured in air or in phantoms
• CTDI in air can be used as a measure of tube output
• For dosimetry purposes CTDI is measured in perspex phantoms• 16cm diameter ‘head’• 32cm diameter ‘body’
Kampala
Patient Size.• CTDI is a measure of scanner output.• CTDI does take patient size into account• CTDI is NOT patient dose• Important for paediatrics, small adults
and possibly bariatrics.• You can correct to patient size
– To be mentioned later.
Kampala
Phantoms for CT dosimetry
• Cylindrical PMMA phantoms with holes for pencil chamber
• 32 cm body phantom• 16 cm head phantom
Kampala
Centre Periphery – 1cm depth (mean of 4 positions)
C
P1
P2
P3
P4
32 cm
‘Head’ ‘Body’
16 cm
140 mm
CTDI Phantoms
Kampala
40
‘Head’
‘Body’
40
40 40
40
20
20 20
20
10
Periphery : centre 1:1 Periphery : centre 2:1
Typical CTDI100 ValuesCTDI100,p values quoted in mGy
Kampala
CTDIW• Weighted average CTDI: represents the
average dose for contiguous irradiation • CTDIw = 1/3 CTDIcentre + 2/3 CTDIperiphery • nCTDIw = CTDIw/mAs
CTDIC CTDIP
Kampala
CTDIvol
Pitch = 1 CTDIvol = CTDIw
Pitch = 2 CTDIvol = CTDIw
2
Pitch = 0.5 CTDIvol = 2 x CTDIw
Kampala
CTDIvol
• CTDIW is the average dose in the scan plane
• Volume CTDI considers non-contiguous scanning
• The CTDIvol can be considered to be the average CTDIW in the scan volume
• CTDIvol = CTDIW Pitch
Kampala
Dose Length Product
• CTDIvol is a measure of absorbed dose, energy absorbed per unit mass
• Radiation risk is based on total energy absorbed
• Estimate using dose length product (DLP)• DLP = CTDIvol x scan length• Units are mGy cm
Kampala
Dose Quantities in CT
Kampala
Dose Summary Sheet - Commercial CT Scanner
Huda W , Mettler F A Radiology 2011;258:236-242
Kampala
CTDI on the Console
• The CTDI value displayed will depend on the phantom that it is based on.
• Body : 32 cm; Head 16 cm.• CTDI for 16cm reference twice that of
CTDI for 32cm reference.• Paediatric protocol MAY reference to 16
cm phantom• Implementation manufacturer
dependent.Kampala
CTDI MEASUREMENT IN PRACTICE
Kampala
Equipment for CT measurements
• CT ionization chamber and electrometer calibrated for CT beam qualities
• Chamber support stand• CT head & body phantoms
Kampala
Methodology for CT measurements in air
• Position chamber in stand so beyond couch
• Adjust couch position so chamber along axis of scanner and beam centred at central point of chamber
• Record dose measurements for single axial scan at selected parameters (no couch movement)
• Record all data
Kampala
Methodology for CT measurements in phantoms
• Position phantom on couch (use head rest if appropriate)
• Position chamber in central position in phantom• Adjust couch position so chamber centred as
before• Record dose measurements for single axial
scan at selected parameters (no couch movement)
• Repeat for peripheral positions (north, south, east & west)
Kampala