What we did up to now: Summary of the measurements performed up to now with the gas setup built for the THGEMs coated with the CsI. Photocurrent measurements (Leszek’s chamber) with several gas mixtures (Ne, Ar, CH 4 , NeCH 4 X/100-X, ArCH 4 X/100-X, ArCO 2 70/30) at different pressures (i.e. atm pressure, 0.2 bar, 30 mbar) for a dummy THGEM coated with CsI and with a dummy THGEM as a drift (see Jarda’s presentation GDD 32 meeting); Quick characterization (drift and voltage scans) for a double structure THGEM (C7 type: d=0.4 mm, p=0.8 mm, t=0.4 mm, rim=0.01 mm) used as a single structure with 55 Fe source flushing the gas at atmospheric pressure and with several gas mixtures (NeCH 4 90/10, CF4, ArCH 4 X/100-X, ArCO 2 70/30); Optimization of the voltage on the double THGEM structure for few gas mixtures at atmospheric pressure with 55 Fe source; Drift scan with the UV light (Oriel Lamp) on the double THGEM used as a single and as a double structure just with the Ar/CO 2 gas mixture (70/30); Single Photon spectrum acquisition and estimation of the rate of the photons from the current measured onto the anode .
Transcript
Page 1: What we did up to now:

What we did up to now:Summary of the measurements performed up to now with the gas setup built for the

THGEMs coated with the CsI.

• Photocurrent measurements (Leszek’s chamber) with several gas mixtures (Ne, Ar, CH4, NeCH4 X/100-X, ArCH4 X/100-X, ArCO2 70/30) at different pressures (i.e. atm pressure, 0.2 bar, 30 mbar) for a dummy THGEM coated with CsI and with a dummy THGEM as a drift (see Jarda’s presentation GDD 32 meeting);

• Quick characterization (drift and voltage scans) for a double structure THGEM (C7 type: d=0.4 mm, p=0.8 mm, t=0.4 mm, rim=0.01 mm) used as a single structure with 55Fe source flushing the gas at atmospheric pressure and with several gas mixtures (NeCH4 90/10, CF4, ArCH4 X/100-X, ArCO2 70/30);

• Optimization of the voltage on the double THGEM structure for few gas mixtures at atmospheric pressure with 55Fe source;

• Drift scan with the UV light (Oriel Lamp) on the double THGEM used as a single and as a double structure just with the Ar/CO2 gas mixture (70/30);

• Single Photon spectrum acquisition and estimation of the rate of the photons from the current measured onto the anode.

Page 2: What we did up to now:

Same setup, several & different purposes

Gas mixer

Avantes – UV lamp

picoammeters

55Fe

Double Double THGEM THGEM C7C7

Leszek’s Leszek’s chamber chamber Dummy/dummyDummy/dummy

Quartz window of the double THGEM chamber (type C7)

Oriel Lamp

Three positions of the shutter: 1.completly open, 2.closed (some internal reflections),3.Open + filter 3.0 OD

N2 inlet

Hidden N2 exhaust

Setup for photocurrent measurements and optimization of performance for the double

THGEM structure with the iron source

Setu

p f

or

UV

lig

ht

meas.

wit

h

the d

ou

ble

TH

GE

M s

tru

ctu

re

Filter 0.5 Filter 0.5 ODOD

Narrow Filter

Filter 3.0 Filter 3.0 ODOD

Page 3: What we did up to now:

Some examples about performed meas.

THGEM: C7d=0.4 mmpitch =0.8 mmt=0.4 mmrim=0.01 mm

drift = wires

THGEM 2THGEM 1

anode

d=4.5 mmd=1.6 mmd=1.6 mm

200 400 600 800 1000 1200 1400 1600

0

1000

2000

3000Data: CK122_BModel: Exp2PMod1 Chi^2/DoF = 1461.68412R^2 = 0.99289 a 2793.52937 ±9.03815b -0.00231 ±6.9399E-6

Cou

nts

ADC channels

Energy Spectrum

-0.8 -0.6 -0.4 -0.2 0.0 0.2 0.4 0.6 0.8

-680

-660

-640

-620

-600

-580

-560

-540

-520

-500

-480

V=1 kV, atm pressure Ar/ CO2 70/ 30;

Oriel + 0.5 OD+ narrow filter

Cu

rren

t (p

A)

Drift Field (kV/ cm)

V=1 kV V=0.9 kV

Single Photon spectrum

Collection efficiency of electron from CsI

-0.5 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0-350

-300

-250

-200

-150

-100

-50

0

50

100

150

200

250

V1=1.36 kV, V

2=1.36 kV, E

induction=3 kV/ cm, E

drift=0.67 kV/ cm;

Atm pressure Ar/ CO2 70/ 30; source 55Fe big collimator.

Cu

rren

t (p

A)

Transfer Field (kV/ cm)

Anode Current Bottom 2 Current Top 2 Current Bottom 1 Current Drift Current Top 1 projection Current

Transfer scan of the double THGEM with Iron source