Women’s health deserves the bestPhilips MammoDiagnost DR Specifi cation
Superb image quality as well as effi cient, convenient
workfl ow characterize this system. The intuitive,
streamlined workfl ow at the Eleva workspot and
the system’s ergonomic design make working with it
comfortable and effi cient for the radiographer, and
ensures a quick and easy examination for the patient.
The excellent image quality resulting from our UNIQUE
image processing software and the high image resolution
of the MammoDiagnost DR provide a safe diagnosis
for both radiologist and the patient.
4522_962_27161.indd 14522_962_27161.indd 1 11/9/07 3:35:50 PM11/9/07 3:35:50 PM
2
4522_962_27161.indd 24522_962_27161.indd 2 11/9/07 3:36:25 PM11/9/07 3:36:25 PM
3
Content
04 Console
06 Gantry
08 Detector
09 Tube
10 Generator
10 Ambient conditions
10 Conformity
11 Dimensions
4522_962_27161.indd 34522_962_27161.indd 3 11/9/07 3:36:47 PM11/9/07 3:36:47 PM
44
Eleva workspot
Dimensions Height: 2000 mm / 78.7 "
Width: 750 mm / 29.5 "
Depth: 550 mm / 21.7 "
Glass shield 2000 mm x 750 mm / 78.7 " x 29.5 ", meets IEC Protective Shielding requirements
Monitor 19 inch colour touch screen monitor, adjustable in height and angle
Software handling Simultaneous handling of acquisition, processing and export of images
Internal disk space 200 GB, suffi cient to store at least 700 examinations
Automatic deletion Automatic deletion of the oldest images after a confi gurable time
Worklist Support of DICOM worklist management, worklist fi lter enables quick
organization
Exposure An audible and visual signal on the Eleva workspot indicates exposures
Image acquisition After exposure it takes < 16 seconds until the processed image is displayed
on the Eleva workspot monitor
Image processing UNIQUE image processing for excellent image quality and with optimized
visibility of all clinical relevant details
Archiving Image is automatically sent to PACS and/or printer as confi gured. Images
for presentation and / or images for processing can be archived as confi gured.
Alternatively the user can manually send the images to PACS and/or printer
Export status for archiving
and printing
Displayed for each image
IHE Support of the IHE-Mammography Image Integration Profi le
HIPAA HIPAA security profi le conformance including audit trail and time
synchronization
Login Individual login for each user visible after login, fast change of the user
possible
Security Update of the virus pattern signature fi les by license
DICOM grayscale standard Support for export of images
Eleva workspot
4522_962_27161.indd 44522_962_27161.indd 4 11/9/07 3:36:53 PM11/9/07 3:36:53 PM
Eleva workspot
Measurements and annotations Ruler, angle measurements and annotations are possible
Image sizing All images of one examination are displayed with the same magnifi cation factor
Zoom functions – zoom to view-port
– zoom to true-size
– 1:1 zooming
Reject analysis Customizable list of statistical items, appropriate graphical display
Worklist Display of patient, study and examination information
Search fi lter Search function for worklist fi ltering is provided
5
4522_962_27161.indd 54522_962_27161.indd 5 11/9/07 3:36:54 PM11/9/07 3:36:54 PM
6
Gantry
Dimensions Height: 2030 mm / 79.9 "
Width: 1570 mm / 61.8 "
Depth: 1175 mm / 46.3 "
Weight 400 kg
SID 65 cm / 25.6 "
Movements All movements are motor driven
Speed of vertical movement Confi gurable between 40 – 80 mm per/s / 1.6 " – 3.2 " per/s
Rotation of C-arm – 180 degree to + 180 degree, isocentric
Speed of rotation Appr. 20 degree / second
Height adjustment 650 – 1500 mm / 25.6 " – 59.1 " above the fl oor to the object
Wheelchair accessibility Yes
Location of control panels On both sides of the tube head and on both sides of the compression unit
Foot switches Mirrowed foot switches for elevation and compression
Optimized compression Display when the most sensible compression is applied
Grid Motorized linear grid, ratio 5 : 1, 31 lines/cm, 79 lines/inch
Magnifi cation unit Magnifi cation factor of 1.5
Automatic retraction of the grid and
automatic selection of small focus
Collimation Automatic collimation according to the paddle size
Paddles
standard:
optional:
18 x 24, 24 x 30
18 x 24 fl exible, 18 x 24 high edge, 24 x 30 high edge, 24 x 30 fl exible,
axillary paddle, magnifi cation spot paddle (all dimensions in cm)
Automatic compression Motorized with max. compression force of 200 N
Manual compression Max. compression force of 200 N
Max. distance breast support to paddle 320 mm / 12.6 "
Decompression Automatic decompression after each exposure to 150 mm / 5.9 "
above object table
Display Display at the foot of the gantry indicating compression force,
thickness, angulation and patient name
Gantry
4522_962_27161.indd 64522_962_27161.indd 6 11/9/07 3:36:59 PM11/9/07 3:36:59 PM
7
4522_962_27161.indd 74522_962_27161.indd 7 11/9/07 3:36:59 PM11/9/07 3:36:59 PM
8
Detector
Detector material Amorphous Selenium
Field of view 23,9 x 30,5 cm / 0.9 " x 1.2 "
Pixel size 85 µm
Pixel matrix 3584 x 2816
Image depth 13 bit
Refresh time between exposures < 30 seconds
Patient throughput 15 patients / hour
Spatial Resolution > 7 lp / mm
DQE
DQE @ 1 lp / mm > 0.5
DQE @ 5.8 lp / mm > 0.15
MTF > 40% at 5.8 lp / mm
Ghost image factor < 0.3
Detector
4522_962_27161.indd 84522_962_27161.indd 8 11/9/07 3:37:02 PM11/9/07 3:37:02 PM
9
Tube
Anode target / fi lter combination Mo / Mo, Mo / Rh, W / Rh
Focus 0,3 x 0,3 and 0,1 x 0,1
Max. tube current:
Molybdenum focal spot
Tungsten focal spot
large focal spot: 150 mA at 25 kV
small focal spot: 28 mA at 25 kV
large focal spot: 188 mA at 25 kV
small focal spot: 34 mA at 25 kV
Total heat storage capacity 1.8 MJ
Anode heat storage capacity 120 kJ
Anode speed 8800 rpm
Filter – 30 µm Molybdenum
– 25 µm Rhodium
– 50 µm Rhodium
Tube
4522_962_27161.indd 94522_962_27161.indd 9 11/9/07 3:37:21 PM11/9/07 3:37:21 PM
10
Generator
Generator Integrated into the gantry
mAs range 2 – 600 mAs
kV range 23 – 35 kV in 1 kV increments
Mains voltage: 1-phase 208 V, 230 V, 240 V, 277 V ±10%
Mains voltage: 2-phase 208 V, 230 V, 240 V, 277 V, 440 V ±10%
Exposure technique AEC with pre-pulse technique or manual
Dose optimization 4 pre-defi ned programs available, according to
compression automatic selection of target / fi lter
combination and kV value
Exposure times 10 ms to 4.0 s for large focus
60 ms to 6.0 s for small focus
(tolerance ±10%)
Conformity
IHE, DICOM, QA EUREF, MQSA
Ambient conditions
Temperature
Operation
Storage
+18 ºC to +30 ºC
– 10 ºC to +55 ºC
Rel. humidity
Operation
Storage
20% to 75%
10% to 95%
Air pressure
Operation & Storage 700 hPa to 1060 hPa
Generator
4522_962_27161.indd 104522_962_27161.indd 10 11/9/07 3:37:44 PM11/9/07 3:37:44 PM
11
836,4 mm / 32.9"
1220
,4 m
m /
48"
1500
mm
/ 59
.1"
2310
mm
/ 90
.9"
680 mm / 26.8"
Dimensions
1247,7 mm / 49.1"
1220
,4 m
m /
48"
836,4 mm / 32.9"49,5 mm / 1.9" 361,8 mm / 14.2"
2031
mm
/ 80
"
4522_962_27161.indd 114522_962_27161.indd 11 11/9/07 3:38:00 PM11/9/07 3:38:00 PM
Philips Medical Systems is part of
Royal Philips Electronics
Interested?
Would you like to know more about our imaginative
products? Please do not hesitate to contact us.
We would be glad to hear from you.
On the web
www.medical.philips.com
Via email
By fax
+31 40 27 64 887
By mail
Philips Medical Systems
Global Information Center
P.O. Box 1286
5602 BG Eindhoven
The Netherlands
By phone
Asia
Tel: +852 2821 5888
Europe, Middle East, Africa
Tel: +49 7031 463 2254
Latin America
Tel: +55 11 2125 0764
North America
Tel: +1 800 229 6417
© 2007 Koninklijke Philips Electronics N.V.
All rights are reserved.
Philips Medical Systems DMC GmbH reserves the right to make changes in specifi cations and/or to discontinue any product at any time without notice
or obligation and will not be liable for any consequences resulting from the use of this publication.
Printed in The Netherlands.
4522 962 27161/714 * NOV 2007