Operator’s Manual Model 3150 WristOx 2 ® Pulse Oximeter Wrist-Worn Pulse Oximeter with Bluetooth ® Wireless Technology 0123 English
Transcript
Page 1: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Operator’s Manual

Model 3150WristOx2

® Pulse Oximeter

Wrist-Worn Pulse Oximeter with Bluetooth® Wireless

Technology

0123 English

Page 2: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Nonin® reserves the right to make changes and improvements to this manual and the products it describes at any time, without notice or obligation.

Nonin Medical, Inc.13700 1st Avenue North

Plymouth, MN 55441-5443, USA

+ 1 (763) 553-9968800-356-8874 (USA and Canada)

Fax: + 1 (763) 553-7807E-mail: [email protected]

Nonin Medical B.V. Prins Hendriklaan 26

1075 BD Amsterdam, Netherlands

+31 (0)13 - 79 99 040 (Europe)Fax: +31 (0)13 - 79 99 042E-mail: [email protected]

nonin.com

MPS, Medical Product Service GmbH Borngasse 20

D-35619 Braunfels, Germany

References to “Nonin” in this manual imply Nonin Medical, Inc.

Nonin, WristOx2, PureLight, nVISION, FlexiWrap, and Flexi-Form are registered trademarks or trademarks of Nonin Medical, Inc. The Bluetooth word mark and logo are owned by the Bluetooth SIG, Inc. and any use of such marks by Nonin Medical, Inc. is under license. Other trademarks and trade names are those of their respective owners.

© 2014 Nonin Medical, Inc.9584-001-03

CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

Consult Instructions for Use.

0123

EC REP

Page 3: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Contents

Indications for Use.................................................................................................... 1

Contraindications.................................................................................................... 1Warnings ................................................................................................................ 1Cautions ................................................................................................................. 2Declaration of Conformity with FCC and Canadian Ministry of Health Rules for

Electromagnetic Compatibility ........................................................................ 3NCC........................................................................................................................ 3Federal Communications Commission (FCC) Notice ............................................ 4

Point-to-Point Communications .......................................................................... 5

Guide to Symbols...................................................................................................... 6

Displays, Controls, and Indicators .......................................................................... 7

Introduction ............................................................................................................. 10

Unpacking the WristOx2, Model 3150................................................................... 10Standard Kit ...................................................................................................... 10Starter Kit .......................................................................................................... 10

Batteries ............................................................................................................... 11Bluetooth Technology........................................................................................... 11

Operation Modes..................................................................................................... 12

Cable .................................................................................................................... 12Standby ................................................................................................................ 12On......................................................................................................................... 12

Spot Check Mode ............................................................................................. 12Sensor Activation Mode .................................................................................... 13Programmed Mode ........................................................................................... 13Memory Volume (MVI) Display Mode ............................................................... 14

Using the WristOx2, Model 3150............................................................................. 15

Installing Batteries ................................................................................................ 15Attaching the Wristband ....................................................................................... 16

Wristband Description....................................................................................... 16Attaching the Sensor ............................................................................................ 19Patient Application................................................................................................ 20Verifying Operation............................................................................................... 23

Startup Sequence and Self-Test....................................................................... 23Activation Switch .................................................................................................. 24

Activate Bluetooth Radio................................................................................... 24Activate Device ................................................................................................. 24

Error Codes .......................................................................................................... 24

Troubleshooting...................................................................................................... 25

Care and Maintenance ............................................................................................ 27

Cleaning the Device ............................................................................................. 27Cleaning the Sensor ............................................................................................. 27

i

Page 4: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Contents (continued)

Cleaning the Wristband.........................................................................................27Storing...................................................................................................................27

Memory and Data .....................................................................................................28

nVISION Software ....................................................................................................29

nVISION Settings ..................................................................................................29Accessing nVISION Settings .............................................................................29

Cable Connection..................................................................................................30USB Driver Installation (XP SP3).......................................................................31USB Driver Installation (Vista) ...........................................................................32USB Driver Installation (Windows 7)..................................................................32USB Driver Installation (Windows 8)..................................................................33

Bluetooth Connection............................................................................................33Bluetooth Security..............................................................................................34

Connecting the Device into a Medical System......................................................34

Parts and Accessories ............................................................................................35

Sensors .................................................................................................................35

Service, Support, and Warranty .............................................................................37

Service and Support..............................................................................................37Warranty................................................................................................................37

Technical Information..............................................................................................38

Manufacturer’s Declaration ...................................................................................38Equipment Response Time...................................................................................42Testing Summary ..................................................................................................43

SpO2 Accuracy Testing .....................................................................................43Pulse Rate Motion Testing.................................................................................43Low Perfusion Testing .......................................................................................43

Principles of Operation..........................................................................................43Specifications ........................................................................................................44

Oximeter Specifications .....................................................................................44System Specifications........................................................................................45Transmitter.........................................................................................................46

ii

Page 5: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

iii

Figures Figure 1. Front Display (Startup Screen).................................................................... 7Figure 2. Comparison of Full and Partial Display ..................................................... 13Figure 3. Memory Volume Display Mode ................................................................. 14Figure 4. Remove Battery Door................................................................................ 15Figure 5. Insert Batteries .......................................................................................... 15Figure 6. Wristband.................................................................................................. 16Figure 7. Thread Short Segment.............................................................................. 17Figure 8. Secure Long Segment .............................................................................. 17Figure 9. Device with Wristband Attached (Front and Back Views) ......................... 18Figure 10. Attach Sensor.......................................................................................... 19Figure 11. Thread and Tighten Wristband................................................................ 20Figure 12. Fasten Wristband.................................................................................... 21Figure 13. Using the Rectangle Fastener................................................................. 21Figure 14. Using the Square Fastener ..................................................................... 22Figure 15. Apply Sensor to Patient........................................................................... 22Figure 16. nVISION Settings Window ...................................................................... 30

Page 6: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

iv

TablesTable 1. Labeling Symbols......................................................................................... 6Table 2. Error Codes................................................................................................ 24Table 3. Electromagnetic Emissions........................................................................ 38Table 4. Electromagnetic Immunity.......................................................................... 39Table 5. Guidance and Manufacturer’s Declaration—Electromagnetic Immunity.... 40Table 6. Recommended Separation Distances ....................................................... 41

Page 7: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Indications for Use

Indications for Use

The Nonin WristOx2®, Model 3150 Pulse Oximeter is a small, wrist-worn device indicated for

use in measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. It is intended for spot-checking and/or data collection and recording of adult and pediatric patients, during both no motion and motion conditions, and for patients who are well or poorly perfused. The intended use environments are hospitals, medical facilities, ambulatory, subacute, and sleep study environments, and mobile units.

Contraindications

Do not use this device in a Magnetic Resonance (MR) environment or in the presence of flammable anesthetics or gases.

This device is not defibrillation proof per IEC 60601-1.

Warnings

This device is intended only as an adjunct device in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms.

Check the pulse oximeter sensor application site every 6-8 hours to determine the positioning of the sensor and the circulation and skin sensitivity of the patient. Patient sensitivity varies depending on medical status or skin condition.

Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.

Carefully route patient cables and connections to reduce the possibility of patient entanglement, strangulation, or injury to the patient.

To avoid patient injury, use only Nonin-branded PureLight® pulse oximeter sensors. These sensors are manufactured to meet the accuracy specifications for Nonin pulse oximeters. Using other manufacturers’ sensors can result in improper pulse oximeter performance.

To prevent improper performance and/or patient injury, verify compatibility of the monitor, sensor(s), and accessories before use.

No modifications to this device are allowed as it may affect device performance.

The USB cable must be unplugged from the device before replacing batteries.

Before changing the batteries, make sure the device is off and the sensor is not applied to a digit.

This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed carefully to verify normal operation.

The use of accessories, sensors, and cables other than those listed in this manual may result in increased electromagnetic emission and/or decreased immunity of this device.

Do not use the device when alarms are required.

Do not use a damaged sensor. If the sensor is damaged in any way, discontinue use immediately and replace the sensor.

1

Page 8: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Indications for Use

This equipment complies with International IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into service according to the EMC information specified in this manual.

Only use Nonin-branded sensors with a length of 1 meter or less. Accuracy may degrade if sensor cable is over 1 meter in length. Using the sensor cable adapter does not affect accuracy.

Cautions

If this device fails to respond as described, refer to “Troubleshooting” or discontinue use until the situation has been corrected. Contact Nonin Technical Service.

This device has motion tolerant software that minimizes the likelihood of motion artifact being misinterpreted as good pulse quality. In some circumstances, however, the device may still interpret motion as good pulse quality.

Do not place liquids on top of this device.

When setting the clock in Programmed Mode using nVISION software, verify all set times and dates are valid.

Do not place the WristOx2, Model 3150, in liquid or clean it with agents containing ammonium chloride or isopropyl alcohol. Refer to the “Care and Maintenance” section of this operator’s manual.

Use a detergent that is safe for skin and washable surfaces. Most detergents can be high sudsing, so use sparingly. Wipe with a damp, detergent free cloth to remove residue.

After cleaning the single-patient use wristband, it should only be applied to the same patient; do not apply it to a different patient.

Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries.

In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.

This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:

Do not perform any testing or maintenance on this device while it is being used to monitor a patient.

This device is a precision electronic instrument and must be repaired by qualified technical professionals. Field repair of this device is not possible. Except to replace batteries, do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.

Warnings (Continued)

- excessive ambient light- excessive motion- electrosurgical interference- blood flow restrictors (arterial

catheters, blood pressure cuffs, infusion lines, etc.)

- moisture in the sensor

- improperly applied sensor- incorrect sensor type- poor pulse quality- venous pulsations- anemia or low hemoglobin

concentrations

- cardiogreen and other intravascular dyes

- carboxyhemoglobin- methemoglobin- dysfunctional hemoglobin- artificial nails or fingernail

polish

2

Page 9: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Indications for Use

Declaration of Conformity with FCC and Canadian Ministry of Health Rules for Electromagnetic Compatibility

• Nonin Medical, Inc., of 13700 1st Avenue North, Plymouth, Minnesota, 55441, declares under its sole responsibility that Model 3150, WristOx2 Pulse Oximeter, to which this declaration relates, complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

• Ministry of Health (Canada), Safety Code 6: standards include a substantial safety margin designed to ensure the safety of all persons, regardless of age and health. The exposure standard for wireless mobile phones employs a unit of measurement known as the Specific Absorption Rate, or SAR. The SAR limit set by the FCC is 1.6 W/kg.

NCC

警語低功率電波輻射性電機管理辦法第十二條經型式認證合格之低功率射頻電機,非經許可,公

司、商號或使用者均不得擅自變更頻率、加大功率或變更原設計之特性及功能。第十四條低功率

射頻電機之使用不得影響飛航安全及干擾合法通信;經發現有干擾現象時,應立即停用,並改善

至無干擾時方得繼續使用。前項合法通信,指依電信規定作業之無線電信。低功率射頻電機須忍

受合法通信或工業、科學及醫療用電波輻射性電機設備之干擾

在 5.25-5.35 秭赫頻帶?操作之無線資訊傳輸設備,限於室?使用。

Verify all visible indicators appear during the start-up (initialization) sequence. If any indicator does not appear, do not use the device. Contact Nonin Technical Service for assistance.

Portable and mobile RF communications equipment can affect medical electrical equipment.

Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These actions may cause the batteries to leak.

To avoid the risk of confusing or misinterpreting patient data when transmitting data via Bluetooth, verify the device is paired with the correct display unit.

The pulse oximeter may not work when circulation is reduced. Warm or rub the finger or reposition the sensor.

A functional tester cannot be used to assess the accuracy of the oximeter or sensor.

Do not fasten the device too tightly around the patient’s wrist. Inaccurate readings and patient discomfort could result.

If the WristOx2, Model 3150 is being used with wireless communication, use the device within its designated range of approximately 100 meters (spherical radius). Moving outside this range may cause missing or lost data.

Failure of a network data coupling (serial cable/connectors/wireless connections) will result in loss of data transfer.

Radios and cell phones or similar devices can affect the equipment and must be kept at least 2 meters (6.5 feet) away from equipment.

Cautions (Continued)

3

Page 10: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Indications for Use

Federal Communications Commission (FCC) Notice

This device has been tested and found to comply with the limits for a class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This device generates, uses, and can radiate radio frequency energy. If not installed and used in accordance with the instructions, it may cause harmful interference to radio or television reception, which can be determined by turning the device off and on. The user is encouraged to try to correct the interference by one or more of the following measures:

• Reorient or relocate the receiving antenna.

• Increase the distance between the device and the receiver.

• Connect the device to an outlet on a circuit different from the outlet where the receiver is connected

• Consult the dealer or an experienced radio/TV technician for assistance.

• RF Exposure: For body worn operation, to maintain compliance with FCC RF exposure guidelines, use only accessories that contain no metallic components. Use of other accessories may violate FCC RF exposure guidelines and should be avoided.

• The WristOx2, Model 3150, is designed and manufactured not to exceed the emission limits for exposure to radio frequency (RF) energy set by the United States FCC. These limits are part of comprehensive guidelines and establish permitted levels of RF energy for the general population. The guidelines are based on the safety standards previously set by both U.S. and international standards bodies. This device has been shown to be compliant for localized specific absorption rate (SAR) for uncontrolled environment/general population exposure limits specified in ANSI/IEEE Std. C95.1-2005.

• The FCC requires the user to be notified that any changes or modifications to this device that are not expressly approved by Nonin Medical, Inc. may void the user’s authority to operate the device.

Nonin’s use of Bluetooth wireless technology allows SpO2, pulse rate, and plethysmographic data to be transmitted through a Bluetooth radio to a compatible Bluetooth-enabled device. Nonin’s system removes the connection from the sensor cable to the display device, giving patients increased ability to move freely—without being hindered by cables. Nonin’s patient module uses a Bluetooth radio with a range of about 100 meters (328 feet) (spherical radius).

100 m

100 m

4

Page 11: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Indications for Use

Point-to-Point Communications

The WristOx2, Model 3150, features point-to-point communications, allowing one master device (the display device) to be paired to one slave device (the patient module). Once connected, neither device is detectable by any other Bluetooth-enabled device, which reduces the risk of interference and preserves data integrity.

CAUTION: If the WristOx2, Model 3150 is being used with wireless communication, use the device within its designated range of approximately 100 meters (spherical radius). Moving outside this range may cause missing or lost data.

!

5

Page 12: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Guide to Symbols

Guide to Symbols

This chapter describes the symbols that are found in this manual and on the WristOx2, Model 3150. Detailed information about display symbols can be found in “Displays, Controls, and Indicators.”

Table 1: Labeling Symbols

Symbol Description

Caution!

Consult Instructions for Use.

Follow Instructions for Use.

Authorized Representative in the European Community.

CE Marking indicating conformance to EC directive No. 93/42/EEC concerning medical devices.

Type BF-Applied Part (patient isolation from electrical shock)

No alarms

Indicates separate collection for electrical and electronic equipment (WEEE).

Continua Certified™ signifies that this product has been tested and proven to be interoperable with other products that carry the Continua Certified symbol.

Bluetooth® figure mark

Non-ionizing electromagnetic radiation. Equipment includes RF transmitters. Interference may occur in the vicinity of equipment marked with this symbol.

UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1.

IP33 Protected against spraying water and against access to hazardous parts with a tool, per IEC 60529.

Manufacturer

!

EC REP

0123

®

6

Page 13: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Displays, Controls, and Indicators

Displays, Controls, and Indicators

Figure 1: Front Display (Startup Screen)

%SpO2 Display

This 3-digit display, located in the upper left corner of the LCD, shows percent blood oxygen saturation (%SpO2). The range is from 0 to 100 %.

This display also shows the month, year, and hour (24-hour clock format) during startup.

Pulse Rate Display

This 3-digit display, located below the %SpO2 display, shows the pulse rate in beats per minute (BPM). The range is from 18 to 321 BPM.

This display also shows the day and minute during startup.

Activation Switch Sensor Port

7

Page 14: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Displays, Controls, and Indicators

Activation Switch

This switch is located next to the sensor port. Pressing this switch activates the Bluetooth radio for 3 minutes.

It can also be used to turn the device on when it is in Standby mode. See “Activation Switch” section for more information.

Sensor Fault Indicator

This indicator displays if the device determines a sensor fault exists (e.g., sensor disconnect, misalignment, or incompatibility with the device). It also displays when the finger is removed from the sensor.

Pulse Strength Indicator

A pulse strength indicator displays when the device is recording data. The number bars on the display depends on the pulse strength as determined by the oximeter.

Full and Partial Display Mode – This heart-shaped indicator is followed by up to nine curved bars and displays next to the pulse rate.

Memory Volume (MVI) Display Mode – This indicator consists of up to nine curved bars and displays next to the minutes of stored data. For more information, see “Memory Volume (MVI) Display Mode” on page 14.

Poor Pulse Signal Indicator

This indicator displays when the pulse signal is inadequate or the device does not sense a pulse. It may also display if there is excessive motion at the sensor site.

Full and Partial Display Mode

Memory Volume Display Mode

8

Page 15: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Displays, Controls, and Indicators

Battery Indicator

This indicator shows remaining battery life as either full, half, low, and critical (as shown at left).

Replace the batteries when device reaches low state.

When the battery reaches critical state:

• All indicators clear from the display except for the blinking critical battery indicator.

• The current session closes.

• The Bluetooth radio shuts down.

• The clock settings are lost.

• The device reverts to Spot Check mode.

Bluetooth Indicator

This indicator displays when the Bluetooth radio is on. It appears as either the Bluetooth logo or the Bluetooth logo with animated bars.

This indicator displays for the first 2 minutes the device is on. If a master device does not connect to the device in those 2 minutes, the Bluetooth radio shuts down and the icon no longer displays.

When the device is connected to a master device, the indicator displays with animated bars.

If the Bluetooth radio is on when the device enters Standby mode or connects to the USB interface cable, the Bluetooth indicator appears on the LCD while the Bluetooth radio shuts down. It will be the only indicator on the LCD and will display for up to 10 seconds.

SmartPoint Indicator

This indicator displays during the startup sequence.

Full

Half

Critical

Low

Bluetooth indicator

Bluetooth indicator with animated bars

9

Page 16: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Introduction

Introduction

The Bluetooth-enabled WristOx2, Model 3150, is a small, wrist-worn device that displays, measures, and stores patient SpO2 and pulse rate data. The device includes a Bluetooth radio with a range (spherical radius) of approximately 100 meters (328 feet).

The device ships ready to use in Spot Check turn on mode. In Spot Check turn on mode, inserting a finger in the sensor automatically turns the device on. Approximately 10 seconds after the finger is removed, the device enters Standby mode.

Advanced memory and programming features are available with Nonin’s nVISION® software (version 6.3 or greater). See the “nVISION Software” section to learn more about using the device with nVISION.

NOTE: If using the WristOx2, Model 3150 with 3rd party software, please disregard nVISION information.

Unpacking the WristOx2, Model 3150

The WristOx2, Model 3150, standard or starter kit includes the items listed below. Once the shipping carton is unpacked, verify these items were received. Contact the carrier immediately if the shipping carton is damaged.

Standard Kit

• Model 3150, WristOx2 Pulse Oximeter

• Model 8000SM-WO2, reusable soft sensor

• 1 wristband

• 2 AAA (1.5 volt) alkaline batteries

• Operator’s manual (CD)

• USB driver software (on operator’s manual CD) – required to use the PC USB interface cable

Starter Kit

A starter kit is required to configure the device and download data to a PC. The starter kit consists of the standard kit, plus:

• 3 wristbands

• nVISION SpO2 data management software (CD)

• Model 3150SC, PC USB interface cable

10

Page 17: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Introduction

Batteries

The device uses 2 AAA alkaline batteries.

With new alkaline batteries, battery life is approximately 48 hours (minimum) when not connected to a Bluetooth device. When connected to a Bluetooth device, battery life will vary depending on class of operation. See “Specifications” for detailed battery life information.

The battery indicator shows one of four states: full, half, low, and critical. Replace the batteries when device reaches low state. A low battery has a minimum of 10 minutes before it reaches critical state. Actual battery life depends on Bluetooth radio use. In critical battery mode:

• The battery indicator blinks.

• The device no longer monitors or records patient data.

• The clock settings are lost.

• The device reverts to Spot Check mode.

When batteries are removed in low battery mode, the device maintains the time and date for up to 30 seconds. After battery replacement, check the device’s screen during startup to ensure date and time are set. Use nVISION software to synchronize the clock and change the operation mode (see “Accessing nVISION Settings” on page 29).

Remove the batteries and disconnect the sensor if the device is to be stored for more than 1 month. In storage, battery life is approximately 9 months.

NOTES:• This device contains non-volatile memory. Removing or replacing batteries does not affect

the data stored in memory. Stored data remains in memory until overwritten by newer data or cleared from memory with nVISION software (version 6.3 or greater).

• If batteries are replaced while recording data, the session will terminate and some data from the session may not be saved. The terminated session will be time stamped with the current date/time the next time the device turns on.

• If clock settings are lost, the date and time restarts at 01:01:10:00:00.

Bluetooth Technology

Bluetooth technology allows wireless connections between electronic communications and computing devices. The technology is based on a radio link that offers fast and reliable data transmissions. Bluetooth uses a license-free, globally available frequency range in the ISM band—intended to ensure communication compatibility worldwide.

Nonin’s use of Bluetooth wireless technology allows SpO2 and pulse rate data to be transmitted through a Bluetooth radio to a compatible Bluetooth-enabled device. Nonin’s wireless system removes the cable connection from the device, giving patients increased ability to move freely.

To make efficient use of battery life, Nonin’s WristOx2, Model 3150, uses an automatically switchable Class 1/Class 2 Bluetooth radio with a maximum range (spherical radius) of about 100 meters (328 feet). Obstacles and other conditions may affect range, and class of operation and connection mode will impact battery life. See “Specifications” for detailed battery life information.

11

Page 18: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Operation Modes

12

Operation Modes

The WristOx2, Model 3150, has three states: Cable, Standby, and On.

Cable

The device is in Cable mode when it is connected to a PC using the USB interface cable. While in Cable mode, the device does not collect or save data and the Bluetooth radio is off.

NOTE: To save battery life, the Model 3150 will automatically shut off after 60 minutes when it is connected to a PC using the USB interface cable.

Standby

When the device is in Standby mode, the screen is blank and the device appears to be off. In Standby, it is ready for a signal that will turn the device on (e.g., pressing activation switch, inserting finger in sensor [Spot Check mode], connecting sensor [Sensor Activation mode], or programmed start time [Programmed mode]). While in Standby mode, the device does not collect or save data and the Bluetooth radio is off.

On

When the device is on, it can collect and save data. The device features three turn on modes:

• Spot Check mode

• Sensor Activation mode

• Programmed mode

The device is delivered in Spot Check mode. nVISION software (version 6.3 or greater) is needed to access the device settings and change Spot Check mode to Sensor Activation or Programmed mode (see “nVISION Software”). nVISION software (version 6.4 or greater) is needed to access memory volume (MVI) display mode.

The device recalls the active settings when the device is shut off and turned on again.

Spot Check Mode

Spot Check mode is the default turn on operation mode.

The device automatically turns on when a finger is inserted into the sensor. It enters Standby mode 10 seconds after the finger is removed. If the sensor is disconnected, the device enters Standby mode immediately.

In this mode, the sensor can be left connected to the device.

NOTE: If the device determines that a sensor fault exists (a sensor failure, misalignment, or incompatibility with the device) or if a pulse oximeter sensor signal cannot be detected, the device enters Standby mode after 3 minutes.

Page 19: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Operation Modes

Sensor Activation Mode

Sensor Activation mode may be selected through nVISION software. In this mode, the device turns on when the activation switch is pressed or when the sensor is disconnected and reconnected. This mode is useful when using a sensor that is not easily removed from the sensor site (e.g., disposable or wrap sensor).

NOTE: The sensor does not need to be applied to a finger to turn the device on.

If the sensor is not used for at least 10 minutes or if an inadequate pulse signal is detected, the device automatically enters Standby mode. To turn the device on again, press the activation switch or disconnect and reconnect the sensor.

This mode allows for Full or Partial display (see figure 2 for display comparison). When using Partial display, the SpO2 and pulse rate readings do not display. The user will only see the battery indicator and the animated pulse strength indicator.

Figure 2: Comparison of Full and Partial Display

Programmed Mode

Programmed mode may be selected and setup through nVISION software. With the software, the user can program the device to start and stop for up to three sessions. Once programmed, the next start time displays on the LCD every 30 seconds in HH:MM format.

A sensor must be connected for Programmed mode to function.

If the programmed device is in Standby mode and the activation switch is pressed, the user activates the Bluetooth radio and the device for 3 minutes. During this time, the user is able to take and store measurements. After 3 minutes, the device returns to Standby mode.

This mode allows for Full or Partial display (see figure 2 above for display comparison). When using Partial display, the SpO2 and pulse rate readings do not display. The user will only see the battery indicator and the animated pulse strength indicator.

NOTE: A programmed device reverts to Spot Check mode if the clock is not set or if the clock settings are lost when replacing the batteries.

CAUTION: When setting the clock in Programmed Mode using nVISION software, verify all set times and dates are valid.

Full Display Partial Display

!

13

Page 20: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Operation Modes

Memory Volume (MVI) Display Mode

NOTE: When the device is in Memory Volume display mode, the %SpO2 and pulse rate readings do not display on the screen.

MVI display mode is selected using nVISION software (version 6.4 or greater) or it can be enabled using an OEM command (refer to the Model 3150 OEM Specification and Technical Information for details). MVI display mode functions with all operating modes (spot check, sensor activation, and programmed).

Memory Volume display mode is used to quickly see how many hours and minutes of valid data are stored in the device’s memory.

In Memory Volume display mode, the display screen (figure 3) only shows:

• The volume of data (in hours and minutes) stored in memory

• hours: display range of 0 – 199

• minutes: display range of 0 – 59

• The battery indicator

• The pulse strength indicator

When the animated pulse strength indicator displays, the device is recording data. The number next to the indicator are the minutes of stored data, not the pulse rate.

Figure 3: Memory Volume Display Mode

The example in figure 3 shows a device with 10 hours and 56 minutes of stored data.

Pulse Strength Indicator

Hours

Minutes

Battery Indicator

14

Page 21: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Using the WristOx2, Model 3150

Installing Batteries

1. Open the battery compartment by sliding the battery door off the back of the device (figure 4).

Figure 4: Remove Battery Door

2. Insert 2 new AAA alkaline batteries (figure 5). Battery orientation is shown inside the battery compartment.

Figure 5: Insert Batteries

3. Replace battery door by sliding it back into place.

WARNING: Do not use the device when alarms are required.

WARNING: The USB cable must be unplugged from the device before replacing batteries.

WARNING: Before changing the batteries, make sure the device is off and the sensor is not applied to a digit.

15

Page 22: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

4. Inserting batteries does not turn the device on. In Spot Check mode, the device turns on when a finger is inserted in the sensor.

NOTE: When batteries are removed in low battery mode, the device maintains the time and date for up to 30 seconds. After battery replacement, check the device’s screen during startup to ensure date and time are set. If the battery level is at or below the critical level, clock settings are lost and the device reverts to Spot Check mode. Use nVISION software to synchronize the clock and change the operation mode (see “Accessing nVISION Settings” on page 29).

Attaching the Wristband

The WristOx2, Model 3150, is designed to be applied to the patient’s wrist using a wristband.

This section contains instructions for attaching the wristband to the device. See the “Patient Application” section for instructions on how to apply the device to the patient.

Wristband Description

The adjustable wristband has a long segment, a short segment, and a plastic ring (figure 6). The wristband uses hook and loop fasteners to secure the wristband to the device and to the patient.

The long segment has two fasteners to accommodate a wide range of wrist sizes.

Figures 7 and 8 demonstrate how to attach the wristband to the device. Figure 9 shows front and back views of the attached wristband.

Figure 6: Wristband

Square Fastener

Long Segment

Short Segment

Plastic Ring

Rectangle Fastener

16

Page 23: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Figure 7: Thread Short Segment

Figure 8: Secure Long Segment

17

Page 24: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Figure 9: Device with Wristband Attached (Front and Back Views)

Front View

Fasteners Sensor Connector

Back View

18

Page 25: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Attaching the Sensor

The sensor can be connected to the device before or after applying the device to the patient.

The following steps apply to these Nonin sensors:

• 8000SS-WO2, 8000SM-WO2, 8000SL-WO2

• 8000AA-WO2

• 8000J-WO2

NOTE: Refer to the sensor Instructions for Use for appropriate sensor sizing.

If using another Nonin-branded sensor, use sensor adapter cable 3150I (see “Parts and Accessories”).

1. Insert the sensor connector into the sensor port at the top of the device (figure 10). The Nonin logo on the sensor connector should face the front of the device.

2. Push the connector until it clicks into place.

3. The device is ready to use.

Figure 10: Attach Sensor

WARNING: Only use Nonin-branded sensors with a length of 1 meter or less. Accuracy may degrade if sensor cable is over 1 meter in length. Using the sensor cable adapter does not affect accuracy.

19

Page 26: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Patient Application

The WristOx2, Model 3150, is usually worn on the back of a patient’s wrist.

NOTE: The wristband can be used to secure the device to an alternate location (e.g., the upper arm or a bed rail).

NOTE: Ensure the wristband fits comfortably on the patient’s arm. Do not over-tighten the wrist band.

1. Verify the wristband has been attached properly to the device (figure 9). If the wristband has not been attached to the device, see “Attaching the Wristband.”

2. Place the device on the patient’s wrist.

3. Thread the rounded end of the wristband through the plastic ring. Pull the strap through the plastic ring until the device fits comfortably on the wrist (figure 11).

Figure 11: Thread and Tighten Wristband

4. Fold the wristband back over the plastic ring (figure 12) and attach the fastener to the wristband (figure 13 or figure 14). Wrist circumference will determine which fastener is used.

NOTE: When using the rectangle fastener, the end of the wristband can be shortened. To do so, fold the end of the wristband so the square fastener attaches onto the wristband (figure 13).

20

Page 27: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Figure 12: Fasten Wristband

Figure 13: Using the Rectangle Fastener

21

Page 28: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Figure 14: Using the Square Fastener

5. Attach the sensor if it is not already connected (see “Attaching the Sensor”).

6. Apply the sensor to the patient (figure 15). Refer to the sensor Instructions for Use for proper sensor application sites and sensor application cautions and warnings.

Figure 15: Apply Sensor to Patient

7. When in Spot Check mode, inserting a finger in the sensor automatically turns the device on. When the finger is removed, the device enters Standby mode in approximately 10 seconds.

NOTE: Depending on the sensor and ambient light conditions, it may take up to 3 minutes for the device to enter Standby mode.

8. If the device does not turn on, verify battery orientation, operation mode, and sensor connection. Refer to “Troubleshooting” for additional information.

22

Page 29: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Verifying Operation

When the WristOx2, Model 3150, first turns on, it performs a startup sequence and self-test. It occurs:

• When a sensor is applied to a patient (Spot Check mode).

• When a sensor is attached to the device (Sensor Activation mode).

• At a programmed start time when a sensor is attached to the device (Programmed mode).

• After the activation switch is pressed while the device is in Standby mode.

• After the device disconnects from nVISION (Bluetooth connection only).

Verify all indicators display during the startup sequence. Indicators appear in the following order for 1 second each.

Startup Sequence and Self-Test

If the time is not set, the device displays 01:01:10:00:00.

If any indicator does not display, do not use the device. Contact Nonin Technical Service for assistance.

1. r and the software revision level:

2. All display icons:

3. Date/time using 24-hour clock format (MM:DD:YY:HH:MM) (example shows 23 April 2010 at 5:57 p.m.):

Month and Day Year (MM:DD) (YY)

Hour and Minutes (HH:MM)

23

Page 30: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Using the WristOx2, Model 3150

Activation Switch

The activation switch is located next to the sensor port at the top of the WristOx2, Model 3150. It is primarily used to:

• Activate the Bluetooth radio when the device is either on or in Standby.

• Activate the device when it is in Sensor Activation mode so the user does not need to disconnect and reconnect the sensor.

It will also activate the device when it is in Spot Check and Programmed modes.

Activate Bluetooth Radio

When the device’s Bluetooth radio is on, a master device can connect to it. If a connection is not made, the Bluetooth radio shuts down.

Pressing the activation switch turns the Bluetooth radio on for 3 minutes. The device will remain on until the Bluetooth radio shuts down. For example, if in Sensor Activation mode, unplugging the sensor will not put the device in Standby.

Activate Device

When in Sensor Activation mode, the device enters Standby mode after 10 minutes without a signal. Pressing the activation switch allows the user to turn the device on without disconnecting and reconnecting the sensor.

Error Codes

This device includes error codes that indicate problems with the unit. When an error occurs, the device displays the letters “Er” and a two-digit code (table 2).

Some error codes may be corrected by the user. See “Troubleshooting” for more information.

Table 2: Error Codes

Error Code Description

01 Configuration sector error

02 Patient data pointer error

03Main memory pointer error (Device memory is intact; however, the most recent session may be missing from the device.)

04 Data format 13 stored packet pointer error

05Main data format 13 pointer error (Device memory is intact; however, the most recent stored measurement may be missing from the device.)

24

Page 31: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Troubleshooting

Troubleshooting

Problem Possible Cause Possible Solution

Device will not activate.

Batteries inserted wrong. Check batteries.

Batteries are depleted. Replace batteries.

Sensor is disconnected. Reconnect sensor.

Device is in Sensor Activation mode and has timed out.

Press the activation switch.

Disconnect and then reconnect the sensor.

Device is in Programmed mode.

Use nVISION software to select Spot Check or Sensor Activation mode.

%SpO2 and pulse rate do not display.

Device set in Partial Display mode.

Use nVISION software to select Full Display mode. Reconnect sensor.

Poor pulse signal indicator displays.

Excessive patient motion. Reduce patient motion.

Poor pulse signal indicator displays and pulse strength indicator shows two bars or less.

Inadequate pulse signal.

Reposition or replace sensor, or place sensor on a different finger.

Remove and reconnect sensor.

Hands are cold. Warm sensor application site.

No pulse display on pulse strength bar graph indicator.

Sensor applied incorrectly.Refer to sensor Instructions for Use for proper sensor application.

Device needs repair. Contact Nonin Technical Service.

Possible interference from blood flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.).

Reduce or eliminate restriction.

Reduced circulation due to excess pressure from sensor.

Check sensor alignment, reposition sensor, verify correct sensor size.

Excessive ambient light.Shield sensor from light source. Check sensor alignment.

Sensor applied to polished or artificial nail.

Remove fingernail polish or an artificial nail.

Sensor Light-Emitting Diode (LED) is not lit.

Contact Nonin Technical Service.

25

Page 32: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Troubleshooting

If these solutions do not correct the problem, please contact Nonin Technical Service at (800) 356-8874 (USA and Canada), + 1 (763) 553-9968, or +31 (0)13 - 79 99 040 (Europe).

Er 01 displays on LCD.Device configuration memory failure.

Device reverts to default settings (Spot-Check mode, 4-second sample rate). Use nVISION software to change settings. If error code continues, contact Nonin Technical Service.

Er 02 or 04 displays on LCD.

Device memory failure. Contact Nonin Technical Service.

Er 03 or 05 displays on LCD.

Device failure. Device memory intact, but device may have lost most recent session or stored data.

If error code continues, contact Nonin Technical Service.

Dashes continually display on LCD.

Sensor malfunction.Replace sensor with a Nonin-branded sensor.

Device does not record in Programmed mode.

Data collection start and stop times are set incorrectly.

Use nVISION software to program correct start and stop times.

Clock settings are lost after replacing batteries.

Use nVISION software to reset clock.

Devices will not pair.

Device is out of range.Verify device is in range while being paired (approximately 100 meters [328 feet] spherical radius).

Bluetooth radio has timed out. Press activation switch to turn on Bluetooth radio.

%SpO2 indicator and the heart in the pulse strength indicator do not display.

Device has been set to Memory Volume (MVI) display mode.

Use nVISION software to configure the device to full or partial display mode.

Problem Possible Cause Possible Solution

26

Page 33: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Care and Maintenance

Care and Maintenance

The device requires no calibration or maintenance other than battery replacement. The device’s expected service life is 5 years.

Cleaning the Device

Wipe the device with a soft cloth dampened with a 10% bleach/90% water solution (household bleach [containing less than 10% sodium hypochlorite]). Do not use undiluted bleach or any cleaning solution other than those recommended here, as permanent damage could result. Dry with a soft cloth, or allow to air dry.

Clean once per week or more frequently if handled by multiple users.

Cleaning the Sensor

Refer to the sensor Instructions for Use for cleaning information.

Cleaning the Wristband

The wristband is designed for single-patient use. If it needs to be cleaned, hand wash with a mild detergent (see note) in cool water (30 °C/86 °F). Allow to air dry.

Do not machine wash or dry. The wristband will shrink if placed in a dryer.

NOTES:• Mild detergents, such as hand and dish washing liquid detergents, dissolve dirt and grease.

To clean washable surfaces, use in a solution of warm water.• Replace the wristband if the hook and loop fastener no longer secures the wristband to the

device or to the patient.

Storing

Store the device within the stated environmental specifications. See “Specifications” for additional information.

Remove the batteries and disconnect the sensor if it is to be stored for more than 1 month.

CAUTION: Do not place the WristOx2, Model 3150, in liquid or clean it with agents containing ammonium chloride or isopropyl alcohol.

CAUTION: Use a detergent that is safe for skin and washable surfaces. Most detergents can be high sudsing, so use sparingly. Wipe with a damp, detergent-free cloth to remove residue.

CAUTION: After cleaning the single-patient use wristband, it should only be applied to the same patient; do not apply it to a different patient.

!

!

!

27

Page 34: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Memory and Data

Memory and Data

The WristOx2, Model 3150 measures, collects, and stores up to 1,080 hours of SpO2 and pulse rate data with a 4-second data collection rate. Data collected at a 1 or 2-second rate reduces memory capacity to 270 or 540 hours, respectively.

When the memory is full, the device overwrites the oldest existing data with the new data. Each time the device is turned on, data are automatically stored in memory. Data collection of less than 1 minute is not retained in memory.

Each time the device turns on, the current oximeter time and date (if the clock is set properly) are stored in memory to allow quick differentiation of recording sessions. Patient SpO2 and pulse rate are stored every 4 seconds (default), or every 1 or 2 seconds if programmed using nVISION software. The oxygen saturation values are stored in 1% increments in the range of 0 to 100%.

This device contains non-volatile memory. Removing or replacing batteries does not affect the data stored in memory. Stored data remains in memory until overwritten by newer data or cleared from memory with nVISION software.

NOTE: Downloading data in memory does not clear memory. To clear memory, see “nVISION Settings.”

28

Page 35: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

nVISION Software

nVISION Software

Nonin’s nVISION software (version 6.3 or greater) works with Microsoft Windows® operating systems. It allows users to transfer recorded patient data from the device to a PC and then analyze, report, and archive the data. The software is required to access the device’s additional modes of operation and advanced features.

nVISION Settings

The following WristOx2, Model 3150, settings are programmed using nVISION:

• Date and time – 24-hour clock format

• Display options – allows clinicians to choose the best display option for each patient:

• Full display shows %SpO2 and pulse rate data

• Partial display shows pulse strength indicator, but not %SpO2 and pulse rate data

• MVI (memory volume) display shows pulse strength indicator and volume (hours and minutes) of data stored in memory. %SpO2 and pulse rate readings do not display on the screen.

• Patient data storage (sample) rate – 1, 2, or 4 seconds

• Operation Modes – Sensor Activation, Spot Checking, and Programmed (see “Activation Options”)

• Patient ID – up to 50 alphanumeric characters

• Bluetooth Radio – disable at startup

• Synchronize device time/date to the PC time/date

• Download and save patient data to a PC

• Clear device memory

To access nVISION settings, connect the device to a PC using either the PC USB interface cable or a Bluetooth connection.

Accessing nVISION Settings

1. Connect the device to a PC using the USB interface cable (see “Cable Connection”) or Bluetooth (see “Bluetooth Connection”).

NOTE: If using Windows 2000, the WristOx2, Model 3150 will only connect to a PC with a Bluetooth connection. Windows 2000 does not function with the USB interface cable.

2. Open nVISION.

3. Click the Data Capture icon, or select New Data Capture from the File drop down menu.

4. Select 3150 from the list of oximeters.

5. Click Settings.

6. “Enter Wrist Oximeter Settings” window opens (figure 16). Update or change settings as needed.

29

Page 36: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

nVISION Software

7. Click OK.

8. For more information, see nVISION Help. .

Figure 16: nVISION Settings Window

Cable Connection

NOTE: To save battery life, the Model 3150 will automatically shut off after 60 minutes when it is connected to a PC using the USB interface cable.

To connect the device to a PC, use the PC USB interface cable found in the starter kit. Once connected to a PC, the device settings may be accessed and data can be downloaded using nVISION software.

The USB driver software for the cable needs to be installed before the device can connect to the PC. The software is located in the USB Driver folder on the Operator’s Manual CD.

1. Install USB driver if needed. See appropriate “USB Driver Installation” section for more information.

2. Connect the cable to the USB port on the PC.

3. Connect the cable to the device’s sensor port.

4. When the device is ready to use with nVISION, these indicators display on the LCD:

• CP

• Battery indicator

30

Page 37: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

nVISION Software

5. For more information about nVISION, refer to nVISION Help.

NOTE: Disconnect the USB interface cable from the device when the data transfer or device configuration is complete. Leaving the cable connected will reduce battery life.

USB Driver Installation (XP SP3)

1. The USB driver software is on the Model 3150 Operator’s Manual CD. Insert the CD into the PC’s CD/DVD drive.

2. Connect the Model 3150SC USB cable to the sensor port on the Model 3150 and a USB port on the PC.

3. The Found New Hardware wizard opens and asks if Windows should connect to Windows Update to search for software. Select No, not this time and click Next.

4. Select Install from a list or specific location (Advanced) and click Next.

5. When asked to choose search and installation options:

a. Select Search for the best driver in these locations.

b. Unselect Search removable media

c. Select Include this location in the search:

d. Browse to the USB Driver folder on the Operator’s Manual CD and click OK.

e. Click Next.

6. If the Hardware Installation/Windows logo testing window appears, click Continue Anyway to continue the installation.

7. When the wizard is done with the software installation, click Finish.

8. Look up the communications (comm or COM) port for the device:

a. Click Start / Settings / Control Panel.

b. Select System. System Properties window opens.

c. On the Hardware tab, select Device Manager.

d. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).” Make a note of the COM#. It is needed to set up the Model 3150 with nVISION software.

31

Page 38: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

nVISION Software

USB Driver Installation (Vista)

1. Connect the Model 3150SC USB cable to the sensor port on the Model 3150 and a USB port on the PC.

2. In the Found New Hardware window, select Locate and install driver software (recommended).

3. Windows needs your permission to continue pop up appears. Continue with installation.

4. The USB driver software is on the Model 3150 Operator’s Manual CD. Insert the CD into the PC’s CD/DVD drive and click Next.

5. Update Driver Software - Model 3150 window opens. Choose Browse my computer for driver software.

6. Browse to the USB Driver folder on the Operator’s Manual CD and click OK.

7. Click Next.

8. In the Windows Security pop-up window, select Install this driver software anyway.

9. Driver software installs. When Windows has successfully updated the driver software, click Close.

10. Open the Device Manager by clicking Start / Control Panel / Device Manager.

11. In the Device Manager window, look up the communications (comm or COM) port for the device. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).” Make a note of the COM#. It is needed to set up the Model 3150 with nVISION software.

USB Driver Installation (Windows 7)

1. The USB driver software is on the Model 3150 Operator’s Manual CD. Insert the CD into the PC’s CD/DVD drive.

2. Connect the Model 31501SC USB cable to the sensor port on the Model 3150 and a USB port on the PC.

3. Open the Device Manager by clicking Start / Control Panel / System and then selecting Device Manager.

4. Expand Other devices.

5. Right click Model 3150 and select Update Driver Software...

6. Update Driver Software - Model 3150 window opens. Choose Browse my computer for driver software.

7. Browse to the USB Driver folder on the Operator’s Manual CD and click OK.

8. Click Next.

9. In the Windows Security pop-up window, select Install this driver software anyway.

10. Driver software installs. When Windows has successfully updated the driver software, click Close.

11. In the Device Manager window, look up the communications (comm or COM) port for the device. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).” Make a note of the COM#. It is needed to set up the Model 3150 with nVISION software.

32

Page 39: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

nVISION Software

USB Driver Installation (Windows 8)

1. The USB driver software is on the Model 3150 Operator’s Manual CD. Insert the CD into the PC’s CD/DVD drive.

2. Connect the Model 3150SC USB cable to the sensor port on the Model 3150 and a USB port on the PC.

3. Open the Device Manager by right clicking in the bottom left corner of the screen and then click Device Manager. Device Manager window opens.

4. If needed, expand Other devices.

5. Right click Model 3150 and select Update Driver Software...

6. Update Driver Software - Model 3150 window opens. Choose Browse my computer for driver software.

7. Browse to the USB Driver folder on the Operator’s Manual CD and click Next. Verify that “Include subfolders” is checked.

8. In the Windows Security pop-up window, check “Always trust software from Nonin Medical, Inc.” and then click Install.

9. Driver software installs. When Windows has successfully updated the driver software, click Close.

10. In the Device Manager window, look up the communications (comm or COM) port for the device. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).” Make a note of the COM#. It is needed to set up the Model 3150 with nVISION software.

Bluetooth Connection

NOTE: Etched onto the device is the word “pin” followed by a 6-digit number. This is the device’s unique identification number, also known as the Bluetooth Passkey or PIN Code. This number is used when pairing the device to the host system. Refer to the host system’s operator’s manual for additional information.

Before a Bluetooth master device can connect with the WristOx2, Model 3150 (slave device), the devices must be paired. Once paired, the WristOx2, Model 3150, will automatically connect with the last paired master device when turned on or activated.

1. To connect the WristOx2, Model 3150, to a PC or another device using Bluetooth, see Nonin’s online Bluetooth Connection Tutorial:

http://www.nonin.com/training/products/3150/bluetooth_connection_tutorial/

2. When nVISION connects to the WristOx2, Model 3150, the device stops recording patient data and the following indicators display on the LCD:

• CP

• Battery indicator

• Bluetooth icon with animated bars

3. For more information about nVISION, refer to nVISION Help.

33

Page 40: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

nVISION Software

Bluetooth Security

The Bluetooth radio contained in the 3150 is compliant to version 2.0 of the Bluetooth Specification. It supports the Serial Port Protocol (SPP) and the Health Device Profile (HDP) with security mode 2 (service level enforced). The supported encryption key size is up to 128 bits and encryption is enforced on all outgoing and incoming data channels. While the 3150 is in a Bluetooth connection, it will be unavailable for other connections.

Connecting the Device into a Medical System

Incorporating the device into a medical system requires the integrator to identify, analyze, and evaluate the risks to patient, operators, and third parties. Subsequent changes to the medical system after device integration could introduce new risks and will require additional analysis. Changes to the medical system that must be evaluated include:

• Changing the system configuration

• Adding devices to or disconnecting devices from the system

• Updating or upgrading equipment connected to the system

Issues resulting from user-initiated system changes may include corruption or loss of data.

NOTES:• When using the sensor port to connect the device to other equipment, follow each device’s

cleaning instructions.• Verify all equipment connected to the device is suitable for the patient’s environment.

Bluetooth Profiles Supported: Serial Port Profile (SPP), Health Device Profile (HDP)

Security Mode: Mode 2 (service-level enforced security)

Authentication and Encryption: Enforced on all data channels (outgoing and incoming)

Encryption Key Size: Up to 128 bits

CAUTION: Failure of a network data coupling (serial cable/connectors/wireless connections) will result in loss of data transfer.!

34

Page 41: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Parts and Accessories

Parts and Accessories

For more information about Nonin parts, accessories, and sensors, contact your distributor, or contact Nonin at (800) 356-8874 (USA and Canada), +1 (763) 553-9968, or +31 (0)13 - 79 99 040 (Europe). This information is also available on Nonin’s website: www.nonin.com.

Sensors

Model Number Description

3100CC Carrying Case

3150 Manual CD with Operator’s Manual and USB Driver Software

3150SC PC USB Interface Cable

nVISIONnVISION Software (version 6.3 or greater). Used with Microsoft Windows operating systems.

3150ISensor Adapter Cable. Used to connect 1-meter, 9-pin connector sensors to the WristOx2, Model 3150. For compatible 1-meter sensors, see below, contact Nonin or your distributor, or visit www.nonin.com.

3150WB Wristband

3100WBE Wristband Extender, 5 in. (13 cm)

WARNING: Only use Nonin-branded sensors with a length of 1 meter or less. Accuracy may degrade if sensor cable is over 1 meter in length. Using the sensor cable adapter does not affect accuracy.

Model Number Description

Reusable Pulse Oximeter Sensors – 12 inch (0.3 meter) length

8000AA-WO2 Adult Articulated Finger Clip Sensor

8000J-WO2 Adult Flex Sensor

8000SS-WO2 Soft Sensor Small

8000SM-WO2 Soft Sensor Medium

8000SL-WO2 Soft Sensor Large

35

Page 42: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Parts and Accessories

Model Number Description

Optional Pulse Oximeter Sensors (use with Adapter Cable 3150I)

Reusable – 1 meter length

8000AA Adult Articulated Finger Clip Sensor

8000AP Pediatric Finger Clip Sensor

8000Q2 Ear Clip Sensor

8000R Reflectance Sensor

8000H Reflectance Sensor Holder

8000SS Soft Sensor (small)

8000SM Soft Sensor (medium)

8000SL Soft Sensor (large)

8000J / 8000JFW Adult Flex Reusable Sensor / FlexiWrap® Single-Use Sensor Wrap

Disposable – 1 meter length

6000 Series Disposable Sensors

6000CA Adult

6000CP Pediatric

7000 Series Flexi-Form® III Single-Patient Use Sensors

7000A Adult

7000P Pediatric

6500MA Adult/Pediatric

6500SA Adult/Pediatric

36

Page 43: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Service, Support, and Warranty

Service, Support, and Warranty

Service and SupportFor information about the device and accessories, contact your local sales representative or distributor. For the sales representative or distributor in your area, contact Nonin.

A return authorization number is required before returning any product to Nonin. To obtain this return authorization number, contact Nonin’s Technical Service Department at:

Nonin Medical, Inc.13700 1st Avenue North

Plymouth, Minnesota 55441-5443 USA

(800) 356-8874 (USA and Canada) + 1 (763) 553-9968

Fax: + 1 (763) 553-7807E-mail: [email protected]

Nonin Medical B.V. Prins Hendriklaan 26

1075 BD Amsterdam, Netherlands

+31 (0)13 - 79 99 040 (Europe)Fax: +31 (0)13 - 79 99 042

E-mail: [email protected]

nonin.com

WarrantyNONIN MEDICAL, INCORPORATED, (Nonin) warrants to the purchaser the Model 3150, WristOx2 Pulse Oximeter for 3 years from the date of purchase. Nonin shall repair or replace any WristOx2, Model 3150, found to be defective in accordance with this warranty, free of charge, for which Nonin has been notified by the purchaser by serial number that there is a defect, provided said notification occurs within the applicable warranty period. This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any WristOx2, Model 3150, delivered to the purchaser that is found to be defective in any manner, whether such remedies be in contract, tort, or by law.

This warranty excludes cost of delivery to and from Nonin. All repaired units shall be received by the purchaser at Nonin’s place of business. Nonin reserves the right to charge a fee for a warranty repair request on any unit found to be within specifications.

The WristOx2, Model 3150, is a precision electronic instrument and must be repaired by knowledgeable and specially trained Nonin personnel only. Accordingly, any sign or evidence of opening the WristOx2, Model 3150, field service by non-Nonin personnel, tampering, or any kind of misuse of the WristOx2, Model 3150, shall void the warranty.

All non-warranty work shall be performed according to Nonin standard rates and charges in effect at the time of delivery to Nonin.

DISCLAIMER/EXCLUSIVITY OF WARRANTY:

THE WARRANTIES IN THIS MANUAL ARE EXCLUSIVE, AND NO OTHER WARRANTIES OF ANY KIND, WHETHER STATUTORY, WRITTEN, ORAL, OR IMPLIED, SHALL APPLY.

37

Page 44: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Technical Information

NOTE: This product complies with ISO 10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.

Manufacturer’s Declaration

Refer to the following table for specific information regarding this device’s compliance to IEC 60601-1-2.

CAUTION: A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.

CAUTION: Portable and mobile RF communications equipment can affect medical electrical equipment.

Table 3: Electromagnetic Emissions

Emissions Test Compliance Electromagnetic Environment—Guidance

This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is used in such an environment.

RF Emissions

CISPR 11

Group 2 This device must emit electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.

RF Emissions

CISPR 11

Class B This device is suitable for use in all establishments, including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.Harmonic Emissions

IEC 61000-3-2

N/A

Voltage Fluctuations/Flicker Emissions

IEC 61000-3-3

N/A

!

!

38

Page 45: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Table 4: Electromagnetic Immunity

Immunity Test IEC 60601 Test LevelCompliance

LevelElectromagnetic

Environment—Guidance

This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is used in such an environment.

Electrostatic Discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, relative humidity should be at least 30%.

Electrical Fast Transient/Burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

N/A Mains power quality should be that of a typical commercial or hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

N/A Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions, and voltage variations on power supply input lines

IEC 61000-4-11

±5% UT (>95% dip in UT) for 0.5 cycle

±40% UT (60% dip in UT) for 5 cycles

±70% UT (30% dip in UT) for 25 cycles

<5% UT (>95% dip in UT) for 5 seconds

N/A Mains power quality should be that of a typical commercial or hospital environment.

Power Frequency (50/60 Hz) Magnetic Field

IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE: UT is the AC mains voltage before application of the test level.

39

Page 46: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Table 5: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity

Immunity TestIEC 60601 Test

LevelCompliance

LevelElectromagnetic Environment—Guidance

This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is used in such an environment.

Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended Separation Distance

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

3 Vrms

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to 2.5 GHz

3 V/m

3 V/m

80 MHz to 800 MHz

800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency range.b

Interference may occur in the vicinity of equipment marked with the symbol:

NOTES:

1. At 80 MHz and 800 MHz, the higher frequency range applies.

2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocat-ing the device.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

d 1.17 P=

d 1.17 P=

d 2.33 P=

40

Page 47: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Table 6: Recommended Separation Distances

This table details the recommended separation distances between portable and mobile RF communications equipment and this device

This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Users of this device can help prevent electromagnetic interference by maintaining a minimum

distance between portable and mobile RF communication equipment (transmitters) and the device as recommended below, according to maximum output power of the communications equipment.

Separation Distance According to Frequency of Transmitter

Rated Maximum Output Power of Transmitter

W

150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz

0.01 0.12 0.12 0.23

0.1 0.37 0.37 0.73

1 1.2 1.2 2.3

10 3.7 3.7 7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTES:

1. At 80 MHz and 800 MHz, the higher frequency range applies.

2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

d 1.17 P= d 1.17 P= d 2.33 P=

41

Page 48: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Equipment Response Time

If the signal from the sensor is inadequate, the last measured SpO2 and pulse rate values freeze for 10 seconds and are then replaced with dashes.

Example - SpO2 Exponential Averaging

SpO2 decreases 0.75% per second (7.5% over 10 seconds)

Pulse Rate = 75 BPM

Specific to this example:

• The response of the 4-beat average is 1.5 seconds.

SpO2 Values Average Latency

Standard/Fast Averaged SpO2 4 beat exponential 2 beats

Pulse Rate Values Response Latency

Standard/Fast Averaged Pulse Rate 4 beat exponential 2 beats

Equipment Delays Delay

Display Update Delay 1.5 seconds

60

65

70

75

80

85

90

95

100

0.0

8.0

16.0

24.0

32.0

40.0

48.0

56.0

64.0

72.0

80.0

Time in seconds

Sp

O2

SaO2 Reference 4 Beat Average

42

Page 49: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Testing Summary

SpO2 accuracy and low perfusion testing was conducted by Nonin Medical, Inc., as described below.

SpO2 Accuracy Testing

During motion and no-motion conditions at an independent research laboratory, SpO2 accuracy testing is conducted during induced hypoxia studies on healthy, male and female, non-smoking, light- to dark-skinned subjects that are 18 years of age and older. The measured arterial hemoglobin saturation value (SpO2) of the sensors is compared to arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory co-oximeter. The accuracy of the sensors in comparison to the co-oximeter samples measured over the SpO2 range of 70 – 100%. Accuracy data is calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment—Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.

Pulse Rate Motion Testing

This test measures pulse rate oximeter accuracy with motion artifact simulation introduced by a pulse oximeter tester. This test determines whether the oximeter meets the criteria of ISO 80601-2-61 for pulse rate during simulated movement, tremor, and spike motions.

Low Perfusion Testing

This test uses an SpO2 Simulator to provide a simulated pulse rate, with adjustable amplitude settings at various SpO2 levels for the oximeter to read. The oximeter must maintain accuracy in accordance with ISO 80601-2-61 for heart rate and SpO2 at the lowest obtainable pulse amplitude (0.3% modulation).

Principles of Operation

Pulse oximetry is a non-invasive method that passes red and infrared light through perfused tissue and detects the fluctuating signals caused by arterial pulses. Well-oxygenated blood is bright red, while poorly oxygenated blood is dark red. The pulse oximeter determines functional oxygen saturation of arterial hemoglobin (SpO2) from this color difference by measuring the ratio of absorbed red and infrared light as volume fluctuates with each pulse.

43

Page 50: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Specifications

Oximeter Specifications

Oxygen Saturation Display Range: 0 to 100 % SpO2

Pulse Rate Display Range: 18 to 321 beats per minute (BPM)

Displays:

Numeric: 3-digit LCD

Pulse Strength: Pulse Strength Bar Graph

Accuracy – Sensors: Declared accuracy data for compatible sensors can be found in Nonin’s Sensor Accuracy document.

Measurement Wavelengths and Output Powera:

a.This information is especially useful for clinicians performing photodynamic therapy.

Red: 660 nanometers @ 0.8 mW max. avg.

Infrared: 910 nanometers @ 1.2 mW max. avg.

44

Page 51: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

System Specifications

Temperature:

Operating: -5 °C to 40 °C (23 °F to 104 °F)

Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)

Device temperature will not exceed 41°C as measured during a controlled environment test.

Humidity:

Operating: 10 % to 95 % noncondensing

Storage/Transportation: 10 % to 95 % noncondensing

Operating Altitude: Up to 12,192 meters (40,000 feet)

Operating Hyperbaric Pressure: Up to 4 atmospheres

Power Requirements: Two AAA (1.5V) alkaline batteries

Battery Life:

Storage (with Memory Volume display mode off): Approx. 9 months

Operating without Bluetooth: Approx. 48 hours (minimum) of continuous operation

Operating with Bluetooth radio – Class 1a:

a. When operating with Bluetooth, typical battery life may vary depending on proximity to host connection and configuration of host-to-device communications. Times provided are minimum times for common configurations.

Approx. 8 hours minimum

Operating with Bluetooth radio – Class 2: Approx. 24 hours minimum

Dimensions (without sensor or wristband): 51 mm x 73 mm x 19 mm (H x W x D)

(2.0 in. x 2.9 in. x 0.75 in.)

Weight (with batteries and wristband): 70.0 g (2.5 oz)

Memory:

Type: Non-volatile

Capacity: up to 1,080 hours (4 sec. data storage rate)up to 540 hours (2 sec. data storage rate)up to 270 hours (1 sec. data storage rate)

Classification per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1 / UL60601-1:

Type of Protection: Internally powered (battery power)

Degree of Protection: Type BF-Applied Part

Mode of Operation: Continuous

Enclosure Degree of Ingress Protection: IP33

This product complies with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing.

45

Page 52: WristOx 2 Pulse Oximeter - DocStock - Welch Allyn · Nonin, WristOx2, PureLight, nVISION, FlexiWrap, ... WristOx2 Pulse Oximeter, to which this declaration relates, complies with

Technical Information

Transmitter

Bluetooth Compliance: Version 2.0

Operating Frequency: 2.4 to 2.4835 GHz

Output Power: < 20 dBm

Operating Range: 100-meter (328-foot) radius indoors

Network Topology: Point-to-Point

Operation: Slave

Antenna Type: Internal

Modulation Type: Frequency Shift Keying

Frequency Hopping Spread Spectrum

Band Width: 1 MHz

46