Counting of Somatic Cells in Raw Milk
NucleoCounter®SCC-100TM
www.chemometec.com
NucleoCounter SCC-100
Stand-alone fluorescence microscope with built in computer for image analysis
Disposable sampling and measuring device, preloaded with nuclei staining dye
www.chemometec.com
Easy operation
1. Lyse 2. Load 3. Count
www.chemometec.com
Fast Result
+ = 30 seconds
= 30 seconds
------------------
max 1 minute
www.chemometec.com
Base technology of ChemoMetec
Fluorescence Microscopy
www.chemometec.com
Integrated Fluorescence Microscope
Lenses: magnification <1X.
Excitation filter: sorts out green light.
Emission filter: sorts out long wave length including red light.
LensesCCD
Emission Filter
LED’s
Excitation Filter
fixture
www.chemometec.com
NucleoCassette
www.chemometec.com
Propidium iodide (PI)
•Immobilized in the interior of the NucleoCassette, 2.8 ug in each
•A fluorescence dye, which intercalates with DNA of cell nuclei
•Cell membrane permeable
www.chemometec.com
1 dye molecule per 4-5 bp
Genome between 2.5 x 109 – 4.0 x 109 bp. Propidium Iodide
DNA base pair (bp):
A: Adenine
T: Thymine
G: Guanine
C: Cytosine
C27H34N42+.2I- Mr 668.4
www.chemometec.com
Reagent C-100• Disrupts the plasma membranes of the mammalian cells, rendering the nuclei
susceptible to staining with PI
www.chemometec.com
Total cell count
Viable Non-viable
Free nucleus
PI
PI-stained nucleus
Lysis buffer
(Reagents C)
www.chemometec.com
Sensitive parts,
moving parts and
“wear and tear” parts
have been taken out of the instrument
and integrated in the disposable NucleoCassette
- Pump- Tubing’s- Mixer- Precision optics- Chemicals- Calibration- Waste container
The NucleoCassette
www.chemometec.com
Calibrated SCC-cassettes
W
T (
co
de
)
Dots for thickness code (T)
TAF0.01TAreaTWLVolume
www.chemometec.com
Highly reliable LEDs
Thermal drift in optical intensity is approximately 0.2 %/°C
Optical intesity versus number of Number of analysis for the NucleoCounter SCC-100
0.90
0.95
1.00
1.05
1.10
0 250000 500000 750000 1000000
Number of analysis
Optica
l inte
nsi
ty (
a.u
.) v
Test conditions:t = 60°C, If(LED) = 100mASCC-100 Operating conditions max:t = 40°C, If(LED) = 30mA
www.chemometec.com
Key Advantages•Simple Operation
–Ready in 5 seconds–5 minutes operator training–No cleaning
•Calibration Free–LED's, CCD and fixed optics assure stable factory calibration–Individually calibrated cassettes assure long term stability
•Maintenance Free–Robust and reliable parts used in design–2 years product guarantee
•Operator Independent–Objective cell counting
•Safe sample handling–PI immobilized inside NucleoCassette
www.chemometec.com
Good reproducibility•Results from different NucleoCounters can be directly compared:
–All NucleoCounters are calibrated to respond as the same Master NucleoCounter at ChemoMetec.
–The initial calibration will maintain throughout the lifetime of a NucleoCounter.
www.chemometec.com
Repeatability•Repeatability error dominated by count statistics (Poisson distribution) •CV = 1/ x 100%
CV% - NucleoCounter SCC-100
0.0%
2.0%
4.0%
6.0%
8.0%
10.0%
12.0%
14.0%
16.0%
18.0%
0 200 400 600 800 1000 1200
SCC
Nucle
oC
ounte
r SC
C-1
00 C
V%
NucleoCounter SCC-100 Poisson
www.chemometec.com
Repeatability
•Repeatability*:
• - CV < 8% at 150,000 cells/ml• - CV < 6% at 300,000 cells/ml• - CV < 4% at 500,000 cells/ml• - CV < 3% at 1,000,000 cells/ml
•CV = Repeatability error Coefficient of Variance
www.chemometec.com
Reproducibility•Comparison of two instruments from different production series - Factory Settings!
NucleoCounter SCC-100 B vs A
y = 1.0047x - 0.0285
0
200
400
600
800
1000
1200
0 200 400 600 800 1000 1200
NucleoCounter SCC-100 Inst. A
Nucle
oC
ounte
r SC
C-1
00 I
nst.
B
www.chemometec.com
Reproducibility
•Only significant method contribution is repeatability–In test more than 95% of the intra-laboratory reproducibility explained by repeatability
www.chemometec.com
Routine Milk Samples
www.chemometec.com
Milk Application• SCC Stability
– The inherent stability of the NucleoCounter SCC-100 offers tools to monitor and assure stability of SCC results obtained by routine instruments
• Pilot Samples– The NucleoCounter SCC-100 allows fast and reliable determination of SCC pilot samples
www.chemometec.com
Somatic Cell Counting• NucleoCounter SCC-100
– Stand-alone fluorescence microscope– Maintenance free– Calibration free
• NucleoCassette– Pre-loaded with nuclei staining dye– Analysed volume 1µl (working factor 1,000)– Calibrated measurement chamber (Dot code)
• Cell Counting– Operator independent– Reproducibility - Appr. Equal to Repeatability!– Repeatability Error - Appr. Poisson statistics
• (≈5% @ 500,000/ml)– Sample throughput – About 100 samples/hour– SCC - 10,000/ml to 1,000,000/ml