Date post: | 27-Mar-2015 |
Category: |
Documents |
Upload: | riley-frazier |
View: | 215 times |
Download: | 0 times |
X-POLSolar X-ray Polarimeter
Rudawy, P., Sylwester, J., Zawistowski, T., Siarkowski, M., Bąkała, J., Podgórski, P., Kowaliński, M. and Gburek, S.
Space Research Centre, Polish Academy of Sciencesul. Kopernika 11, Wroclaw , POLAND
PURPOSE
Understanding impulsive particle acceleration and plasma heating processes in the solar flares through X-ray polarimetry
•High energy- and time-resolution polarimetric observations •Active regions•Solar flares (non-stationary processes in flares; energy release and energy transport processes in solar flares)•Impulsive acceleration of the non-thermal electrons•Beam diagnostics•Energy deposition in the solar chromosphere •Temperature and differential emission measure studies (plasma diagnostics)•Chemical composition of the coronal plasma;
Space Research Centre, Polish Academy of Sciences
INSTRUMENT
Space Research Centre, Polish Academy of Sciences
Detectors: 4 CCD back-iluminated fully depletedEnergy range: soft X-rays 1.86Å (Fe XXV), 3.20Å (CaXIX), 4.0Å (ArXVII)Energy resolution: 104:1 Sensitivity: >C GOES class eventsOrbit: Sun synchronous preferred, altitude: >500 and <800 km Pointing: 2-axes stabilization, detector axis pointed to +/-2ºLifetime: >3 years
The overall size of the instrument is approx.: 32 cm x 26 cm x 26 cm Mass approx. 4 kg
INSTRUMENT
Space Research Centre, Polish Academy of Sciences
TWO-AXES ADJUSTABLE FRAMEWITH THREE
DETECTOR MODULES
TOP VIEW
UV-FILTERS &
ENTRANCECAPILLARY MESH
PINHOLESOLAR FLARE
LOCATOR
Space Research Centre, Polish Academy of Sciences
TWO-AXES ADJUSTABLE FRAMEWITH THREE
DETECTOR MODULESBENTCRISTAL
CCD
COOLER
ENTRANCECAPILLARY MESH
INSTRUMENT
Space Research Centre, Polish Academy of Sciences
INSTRUMENT
w x+y z
ArXVII He-like triplet
ENTRANCECAPILLARY MESH
TWO-AXES ADJUSTABLE FRAMEWITH THREE DETECTOR MODULES
CCDSOLAR FLARE
LOCATOR
MOTOR 1
MOTOR 2
TWO-AXESSUSPENSION
HEAT-BELT
TWO-AXES ADJUSTABLE
FRAMEWITH THREE DETECTOR MODULES
OUR HERITAGE
• RF15-I X-ray spectrophotometer- Imager(Interball-Tail, 1995-2000) PI
• RESIK Bragg Bent crystal spectrometer 3-6 Å(Coronas-F, 2001-2003) PI in collab. NRL, RAL, IZMIRAN
• Diogeness Bragg scanning spectrometer(Coronas-F, 2001) PI
• Solar Photometer in X-rays (SphinX)(Coronas-Foton, 2009) PI, in collab. U. Palermo, AI CAS
• ChemiX(InterHelioProbe, in prep. 2018) PI, in collab. IZMIRAN
Space Research Centre, Polish Academy of Sciences
TWO BRAGG SPECTROMETERS CONSTRUCTED AT SRCWITH INTERNATIONAL COOPERATION
RESIK - bent crystal spectrometerDiogeness - flat scanning crystal
spectrometer
Space Research Centre, Polish Academy of Sciences
New X-Ray emission lines and continuum emitted by active solar corona successfully measured for
the first time in the wavelength range 3.3 – 6.1 Å.
RESIK RESULTS
Absolute abundance of argon determined for the first time for
the Sun using direct spectroscopic tools. Similar results for S, Cl, K
(series of ApJ papers)
Space Research Centre, Polish Academy of Sciences
SphinXSRC CONCEPT, DESIGN & MANUFACTURE
Measures the X-ray emission of the Sun in the 0.8 – 15 keV band with:– Time resolution ~10 μ s– Sensitivity 100 x better than GOES (the standard for 30+ years)– Energy resolution 3x RHESSI (NASA)
Principal aim: to record quiet-Sun level of emission
Space Research Centre, Polish Academy of Sciences
ChemiXSRC CONCEPT & DESIGN
Space Research Centre, Polish Academy of Sciences
Basic Instrument parametersMass: ~5 -6 kg
Power consumption: ~10 WSize: 30cm x 30cm x 120cm
Telemetry: up to 12 kB/s Specific instrument Characteristics
Wavelength coverage: 1Å –7 Å, Qartz and SiSpectral resolution : better than 0.0005 Å
Detectors: 3 CCD back illuminated & particle PINSensitivity: 1.5 –2 x RESIK, S/N: 10 xRESIK
Collimators: multi-grid, slit, Electronics: FPGA based
X-POLSolar X-ray Polarimeter
Space Research Centre, Polish Academy of Sciencesul. Kopernika 11, Wroclaw , POLAND
http://www.cbk.pan.wroc.pl/