+ All Categories
Home > Documents > XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by...

XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by...

Date post: 23-May-2018
Category:
Upload: hangoc
View: 213 times
Download: 0 times
Share this document with a friend
32
2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 9868090 Fax: (315) 9868091 www.ankom.com Operator’s Manual Rev E 4/2/12
Transcript
Page 1: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

          

2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 986‐8090 Fax: (315) 986‐8091   www.ankom.com 

Operator’s Manual

Rev E 4/2/12 

Page 2: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 

 

 

 

 

 

 

 

 

This page intentionally left blank    

Page 3: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

Table of Contents 

     Introduction .................................................................................................................................................................. 5 

Warranty ....................................................................................................................................................................... 5 

Filter Bags ..................................................................................................................................................................... 5 

Operating Environment ................................................................................................................................................ 5 

Contact Information ..................................................................................................................................................... 6 

Instrument Description ................................................................................................................................................. 7 

Safety Precautions ........................................................................................................................................................ 8 

Instrument Installation ................................................................................................................................................. 9 

Fat Extraction Process ................................................................................................................................................. 10 

Periodic Maintenance ................................................................................................................................................. 19 

Troubleshooting .......................................................................................................................................................... 21 

Appendix A – AOCS Official Fat Extraction Procedure ................................................................................................ 23 

Appendix B – Parts & Assemblies ............................................................................................................................... 25 

Appendix C – Electrical Diagram ................................................................................................................................. 28 

 

  

   

Page 4: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 

 

 

 

 

 

 

 

 

This page intentionally left blank 

Page 5: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 5 

 

Introduction 

ANKOM Technology designs, manufactures, and markets instruments and support products used by analytical laboratories around the world in the environmental, agricultural, biomass, and food industries. ANKOM Technology can provide you with products for determining or monitoring dietary fiber, detergent fiber, fat, digestibility, microbial fermentation (anaerobic or aerobic) and more.

Committed to Total Customer Satisfaction, ANKOM designs every product based on a thorough assessment of customer needs.

Congratulations on your purchase of the ANKOMXT15 Extraction System. We are confident that this product will effectively serve your needs.

The ANKOMXT15 Extraction System was designed to automate the process of fat extraction. This reduces technician variation, increasing precision. By carefully following the operating instructions in this manual you will understand the details of sample and filter bag handling as well as the instrument controls, helping you to achieve the best possible results.

 Please review the entire contents of this manual before you begin operating this product.

Warranty 

ANKOM Technology warrants the ANKOMXT15 Extraction System against any defects due to faulty workmanship or material for one year after the original date of purchase. This warranty does not include damage to the instrument resulting from neglect or misuse. During the warranty period, should any failure result from defects in workmanship or materials ANKOM Technology will, at its discretion, repair or replace the instrument free of charge.

Extended warranties are available upon request.

Filter Bags 

Use only ANKOM Technology filter bags (part # XT4) in your ANKOMXT15 Extraction System. Use of any other filter mechanism will void the warranty. Filter bags can be purchased from ANKOM Technology or from your local authorized ANKOM distributor.

Operating Environment 

Your ANKOMXT15 Extraction System is designed to operate within the following environments:

Ambient Temperature Range: 15°−35°C  Power: 100V−120V ~ 50/60Hz 10A / 220V−240V ~ 50/60Hz 5A  

   

Page 6: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 6       Rev E 4/2/12 

 

Contact Information 

At ANKOM Technology we are committed to your total satisfaction and therefore always available to help you get the most from your ANKOM products. We are also very interested in any comments or suggestions you may have to help us improve.

For any questions or suggestions regarding your instrument, please contact us at:

Telephone: (315) 986-8090 Fax: (315) 986-8091 Email: [email protected] www.ankom.com        

  

Page 7: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 7 

 

Instrument Description 

General Description   

The ANKOMXT15 Extraction System is designed to extract Crude Fat using conventional solvents, typically petroleum ether. The compounds extracted are primarily triacylglycerols together with a small quantity of related lipids, traditionally termed “Crude Fat.” The analysis is achieved by measuring the loss of mass due to the extraction of fat/oil from the sample encapsulated in a filter bag. The quantitative isolation of the sample is accomplished by surrounding the sample in a sealed filter bag with a filtering capacity in the 2 to 3 micron range. The filter bag has sufficient porosity to permit rapid solvent passage and is composed of polymeric material that is resistant to the higher temperatures and solvents used in the instrument.

The quantitative isolation achieved by the filter bag permits the ANKOMXT15 Extraction System to automatically process samples in batches of up to 15. With the solvent reservoir filled, simply place filter bags with sample in the Extraction Vessel, lock the vessel in place, select the extraction time and temperature, and press “Start.” The instrument will automatically fill the Extraction Vessel with solvent, extract the fat from the sample, and recycle the solvent. 

Below is a detailed view of the ANKOMXT15 Extraction System.

 

Keypad 

Display 

Power Switch (not visible) 

Safety Valve Handle 

Pressure Gauge 

Extraction Vessel 

Solvent Fill Port 

Main Handle 

Solvent Sight Level Reflux Water Flow Control 

Page 8: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 8       Rev E 4/2/12 

 

Safety Precautions 

 

Flammable Hazard – Static electricity is to be avoided as it is a spark hazard. Hot Surfaces – Do not touch the extraction vessel during operation. The surface can exceed 70°C (158°F). Failure to observe this caution may result in injury. Hazardous Voltages – Do not operate the instrument with the electrical compartment open. Hazardous voltages are present during operation. Failure to observe this caution may result in electrical shock or electrocution. Hazardous Materials – The ANKOMXT15 Extraction System comes equipped with a charcoal filter which traps any residual vapor in the vent system. By properly maintaining the charcoal filter, volatiles will not escape the vent system. We recommend replacement of the filter every 3 months or after 100 extraction runs. Trace amounts of vapor may also be present when samples are removed. It is recommended that the ANKOMXT15 Extraction System be placed in a well ventilated environment, away from any heat or ignition sources. Organic solvents may be flammable. Follow safe laboratory practices according to both local and federal regulations when installing this instrument and when handling any organic solvents.

  WARNING: Attempts to override safety features or to use this instrument in a manner not specified by ANKOM Technology voids the warranty and may result in serious injury or even death.

This system is designed to meet and/or exceed the applicable standards of CE, CSA, NRTL and OSHA.

 

 

 Please review the entire contents of this manual before you begin operating this instrument.

    

Page 9: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 9 

 

Instrument Installation 

Site Requirements 

To install and operate the ANKOMXT15 Extraction System you will need the following:

Adjustable wrench Chilled or tap water supply not to exceed 23°C Adequate power (see “Operating Environment” section) Drain

Instrument Installation Procedure 

To install the ANKOMXT15 Extraction System, follow the procedure detailed below.

1. Remove the instrument from the shipping container and place in an area that is free of dust and excessive moisture, and is well ventilated (away from any heat or ignition sources).

 

Because the ANKOMXT15 Extraction System is a closed system, there is very limited exposure to solvents. When used in conjunction with a charcoal filter on the vent line the instrument produces less than 1ml of volatiles. It is the user’s responsibility to determine whether venting of this instrument requires placement in a hood. Please follow both local and federal regulations when installing this instrument.

2. Connect the tube from the top fitting to the water supply and the one from the bottom fitting to the drain.

 Ensure that there are no kinks or bends in any of the tubing connected to the instrument.

3. Connect the vent flask to the Vent Flask Connector on the instrument and run the tubing to the inline charcoal filter. Periodically check the vent line to verify that it is free from any blockage or kinks. Generally, there should not be any significant solvent in the vent flask. Any solvent visible in the flask may indicate a problem with the water supply to the instrument.

  Change the Charcoal Filter every 3 months or after 100 extraction runs.

4. Plug the power cord into a grounded receptacle.

Charcoal Filter 

Water Supply

Drain

Vent Flask Connector 

Page 10: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 10       Rev E 4/2/12 

 

Fat Extraction Process 

Crude Fat Calculation 

Crude Fat contained within a food or feed sample can be calculated using the following formula:

  % Crude Fat  =  100 x (W2 – W3)         W1 

  Where:  W1  =  Original weight of sample     W2  =  Weight of pre‐dried sample and filter bag    W3  =  Weight of dried sample and filter bag after extraction

For meat samples with > 15% fat and plant samples with > 20% fat:

  % Crude Fat  =  100 x ((W2 – W4) – W3)         W1 

  Where:  W4  =  Weight of the weigh tin 

Total Fat Calculation 

If hydrolysis is done prior to fat extraction, then Total Fat contained within a food or feed sample can be calculated using the following formula:

  % Total Fat  =  100 x (W2 – (W3 + (C1 – C2)))         W1   

  Where:  W1  =  Original weight of sample     W2  =  Weight of dried sample, filter bag, and Diatomaceous Earth (DE) after hydrolysis     W3  =  Weight of dry extracted sample, filter bag, and DE    C1  =  Blank filter bag dry weight after hydrolysis    C2  =  Blank filter bag weight after extraction

Fat Extraction Support Items 

The following support items are needed to perform the fat extraction procedure:

Item  Recommended Product 

Electronic Balance with four-place readout  ANKOM #TB Balance Hardware

ANKOM #TBS Balance Software

Filter Bags  ANKOM #XT4 

Bag Holder (used for adding sample to an empty filter bag) ANKOM #101.2 

Heat Sealer for sealing the filter bags  ANKOM #1915 (120V), #1920 (220V)

Solvent Resistant Marker  ANKOM #F08

Desiccant Pouch or Weigh Tin Desiccator for drying ANKOM #X45, #X49 

Oven for drying (capable of maintaining 102°C±2°) ANKOM #RD (120V), #RDI (220V)

Ventilated oven trays to allow for sufficient oven air flow -----------------

Weigh Tins (used with samples of > 20% fat) -----------------

Sample -----------------

Spoon  -----------------

Page 11: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 11 

 

Solvent preparation procedure 

To fill the solvent reservoir follow the procedure detailed below.

1. Unscrew the Fill Cap.

2. Pour solvent into the Fill Port up to the dark blue level mark on the Solvent Sight Level. Do not add solvent up to the top of the Solvent Sight Level tube or it will start flowing into the Vent Flask.

 

If using Diethyl Ether or a solvent mixture that includes Diethyl Ether (such as the Mojonnier Mixture), you will need the O-ring Kit for High Penetration Solvents (part # X95) to replace the standard O-rings on the top of the Extraction Vessel and the Valve Kit for High Penetration Solvents (part # X55).

 

 

 

Warning: Do NOT use Acetone in this instrument because it will cause damage to the valves.

 

   

Page 12: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 12       Rev E 4/2/12 

 

Sample Preparation Procedure for Crude Fat – for LOW fat samples 

To prepare meat samples with ≤ 15% fat by weight or plant samples with ≤ 20% fat by weight for fat extraction, follow the procedure detailed below.

1. Using a Solvent Resistant Marker, number all of the empty filter bags you will use during the fat extraction procedure. Although it is not necessary in the determination of fat, it is recommended that you weigh and record the weights of all the filter bags prior to filling them in order to calculate the moisture value for the samples.

      % Moisture = 100 x ((Filter Bag Weight + Sample Weight) ‐ Weight after drying)                                                                       Sample Weight

2. Place an empty filter bag in the Bag Holder in an open position.

3. Tare the weight of the empty filter bag and the holder together.

4. Add 1.5 – 2g of sample to the filter bag. Keep all particles away from the sealing area of the filter bag.

5. Record the weight of the sample (W1) and tare.

6. Set the Heat Sealer dial to 6. (The setting may vary from sealer to sealer.)

7. Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of the filter bag (as shown to the right). If the seal is not strong, re-seal the bag.

8. Repeat steps 1 – 7 for all filter bags that will be used in the ANKOMXT15 Extraction System. (Up to 15 bags can be processed during one procedure.)

9. Pre-dry all samples at 102°C±2° for 3 hours to remove moisture.

10 Remove the samples from the oven and place them in a Desiccant Pouch.

11 Allow the samples to cool to room temperature. This should take about 10 – 15 minutes.

12 Re-weigh each bag (W2) immediately after removing from the Desiccant Pouch.

  The samples are now ready for the fat extraction procedure.

Seal 

Page 13: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 13 

 

Sample Preparation Procedure for Crude Fat – for HIGH fat samples 

To prepare meat samples with > 15% fat by weight or plant samples with > 20% fat by weight for fat extraction, follow the procedure detailed below.

1. Using a Solvent Resistant Marker, number all of the weigh tins and empty filter bags you will use during the fat extraction procedure. Use one weigh tin per filter bag. Although it is not necessary in the determination of fat, it is recommended that you weigh and record the weights of all the filter bags prior to filling them in order to calculate the moisture value for the samples.

    % Moisture = 100 x ((Filter Bag Weight + Sample Weight) ‐ (Weight after drying ‐ Tin Weight))                                                                       Sample Weight

2. Dry all weigh tins in an oven at 102°C±2° for 15 minutes.

3. Remove weigh tins from the oven and place them in a Weigh Tin Desiccator or in a Desiccant Pouch.

4. Allow weigh tins to cool to room temperature.

5. Weigh and record tin weights (W4).

6. Place an empty filter bag in the Bag Holder in an open position.

7. Tare the weight of the empty filter bag and the holder together.

8. Add 1.5 – 2g of sample to the filter bag. Keep all particles away from the sealing area of the filter bag.

9. Record the weight of the sample (W1) and tare.

10. Set the Heat Sealer dial to 6. (The setting may vary from sealer to sealer.)

Page 14: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 14       Rev E 4/2/12 

 

11. Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of the filter bag (as shown to the right). If the seal is not strong, re-seal the bag.

12. Repeat steps 1 – 11 for all filter bags that will be used in the ANKOMXT15 Extraction System. (Up to 15 bags can be processed during one procedure.)

13. Place the sealed filter bags in their corresponding weigh tins for pre-drying.

14. Pre-dry all samples at 102°C±2° for 3 hours to remove moisture.

15. Remove the weigh tins and filter bags from the oven and place them together in a Weigh Tin Desiccator.

 

A Weigh Tin Desiccator is recommended because it will hold between 10 and 14 weigh tins with filter bags, depending on the size of the weigh tins. A Desiccant Pouch will hold about 4.

16. Allow the samples to cool to room temperature. This should take about 10 – 15 minutes.

17. Re-weigh and record the weights of each weigh tin (W4) and filter bag (W2) immediately after removing from the Weigh Tin Desiccator.

  The samples are now ready for the fat extraction procedure.

 

   

Seal 

Page 15: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 15 

 

Fat Extraction step‐by‐step procedure using the ANKOMXT15 Extraction System 

To perform fat extraction on prepared samples with solvent in the reservoir, follow the procedure detailed below.

1. On the Display you should see “Remove/Insert Samples Enter.”

2. Press down the Safety Valve Handle to release the Safety Valve Pin and rotate the Main Handle upward until it stops.

3. Use the Extraction Vessel Handle to pull the Extraction Vessel out.

4. Slide the Extraction Vessel toward you far enough to remove the white Teflon Insert or to put the Bag Holder directly into Teflon Insert while it remains in the Extraction Vessel. 

 

 It is not necessary to remove the white Teflon Insert from the Extraction Vessel in order to put the Bag Holder in place.

5. Put the Bag Holder (with up to 15 bags) into the Teflon Insert. 

Teflon Insert 

Page 16: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 16       Rev E 4/2/12 

 

6. If you removed the Teflon Insert, put it back into the Extraction Vessel, aligning the Teflon Insert slot with the Extraction Vessel pin. 

IMPORTANT  The Teflon Insert O-ring must be in place before running the instrument. The Extraction Vessel O-ring must be in place before running the instrument. The Teflon Insert slot must be aligned with the Extraction Vessel pin before running

the instrument.

7. Slide the Extraction Vessel into the instrument ensuring the Extraction Vessel O-ring stays in place. 

8. Pull the Main Handle downward, closing and locking the Extraction Vessel in place. You will hear a click noise when the Safety Valve Pin locks in place.  

IMPORTANT Before running the instrument, verify that the Extraction Vessel is locked in place by attempting to rotate the Main Handle upward. If it is not locked, follow step #7 above.

9. Press ENTER on the Keypad.

10. On the Display you should see “Select time 60 ^…Enter.” (The number may not be “60.”)

11. Using the arrow keys on the Keypad, select the Extraction time. (The recommended time is 60 minutes.)

12. Press ENTER on the Keypad.

13. On the Display you should see “Heat Temp 90 ^…Enter.”

14. Using the arrow keys on the Keypad, select the Temperature. (The recommended maximum temperature is 90°C.)

15. Press ENTER on the Keypad.

16. On the Display you should see “Close Vessel Water On Start.”

17. Verify that the water supply is on.

18. Press START on the Keypad. The instrument will run through its automatic operation, showing its status on the Display.

Teflon Insert O‐ring

Extraction Vessel O‐ring

Teflon Insert slot and Extraction Vessel pin 

Page 17: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 17 

 

19. When the instrument has completed its automatic operation, the Display will read: “Remove/Insert Samples Enter.”

 At this time, the only light shining on the Display should be the green power light.

20. When the Pressure Gauge reads zero, remove the Extraction Vessel by pushing down on the Safety Valve Handle to release the locking pin and then rotating the Main Handle to the open position.

21. Use the Extraction Vessel Handle to remove the Extraction Vessel.

 

Warning: The Extraction Vessel interior and top will be HOT when the extraction process is complete.

22. Remove the Teflon Insert from the Extraction Vessel.

23. Clean the fat from the bottom of the vessel using paper towel.

IMPORTANT If too much fat accumulates in the bottom of the Extraction Vessel, an overtemp fault will occur.

24. Remove the Bag Holder from the Teflon Insert.

25. Remove the filter bags from the Bag Holder.

26. Dry the samples in an oven at 102°C±2° for 30 minutes.

27. Remove the samples from the oven and place them in a Desiccant Pouch.

28. Allow the samples to cool to room temperature. This should take about 10 – 15 minutes.

29. Re-weigh each filter bag (W3) immediately after removing from the Desiccant Pouch.

 

   

Page 18: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 18       Rev E 4/2/12 

 

 

 

 

 

 

 

 

 

 

This page intentionally left blank   

Page 19: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 19 

 

Periodic Maintenance 

Cleaning the Extraction Vessel 

Some compounds extracted during the fat extraction procedure can create a black residue in the bottom of the Extraction Vessel. The vessel can most easily be cleaned by using a drill with a long shafted stiff wire brush. This treatment will remove a lot of the black coating that accumulates on the bottom of the vessel. When the vessel bottom is clean, put some solvent on a cloth or paper towel and wipe down the inside of the vessel in an effort to remove any oil residue and particles that have been loosened from the vessel bottom.

IMPORTANT  Do not use soap and/or water to clean the vessel. When cleaning the vessel, do not damage the level sensor tip. The tip may be cleaned

with window cleaner or alcohol.

Cleaning the exterior of the instrument 

Wipe the outside of the instrument with window cleaner or a damp sponge as necessary.  

  Do not use Acetone to clean the instrument.

Cleaning the solvent filter 

If filter bags are not sealed properly and sample is lost from the bag, it will be recovered in the solvent filter in the exhaust line.

To clean the solvent filter, follow the procedure detailed below.

1. Remove the rear cabinet of the instrument.

Rear View of instrument with cabinet back removed 

2. Unscrew the compression fitting.

3. Unscrew the filter body.

4. Unscrew and clean the metal filter screen.

5. Reassemble the filter and compression fitting. Make sure the valve body is tight and the Teflon washer has not come out of the filter body.

 

   

Compression Fitting 

Filter body 

Page 20: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 20       Rev E 4/2/12 

 

Cleaning the water filter 

To clean the water filter (located on the water supply line), follow the procedure detailed below.

1. Unscrew the filter from the instrument.

2. Unscrew the filter clamp from the water supply line.

3. Use compressed air to blow backwards through the filter (blow opposite the arrow on the filter).

4. Reconnect the filter and fittings.

5. Turn on the water and check for leaks.

Replacing the Fuses 

To replace the fuses in the ANKOMXT15 Extraction System, follow the procedure detailed below.

1. Turn off the instrument power and unplug the Power Cord from the outlet.

2. Remove the two screws on the Top Cover.

3. Swing the Top Cover down exposing the electrical compartment.

4. Locate the Fuse Holders on the terminal strip. There are two fuses within the instrument; one for each pole (100 – 120V 10A or 220 – 240V 5A).

5. Press the tabs to lift the Fuse Holders up.

6. Replace the fuses and close the Fuse Holders completely.

7. Close the Top Cover and secure it with the two screws that you previously removed.

Replacing the Charcoal Filter 

The Charcoal Filter should be replaced every 3 months, or after 100 extraction operations using the instrument. To replace the filter, remove the old one from the vent line and install the new one.

Fuse Holders 

Tab to open Fuse Holder 

Fuse  

Filter 

Page 21: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 21 

 

Troubleshooting 

The ANKOM technology web site has the most current troubleshooting information. Therefore, if you have any questions about the operation of your ANKOMXT15 Extraction System, please visit our web site at www.ankom.com.     

Page 22: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 22       Rev E 4/2/12 

 

 

 

 

 

 

 

 

 

 

This page intentionally left blank   

Page 23: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 23 

 

Appendix A – AOCS Official Fat Extraction Procedure 

ANKOM Technology Method – Version 1/30/09  AOCS Official Procedure Am 5‐04 

Rapid Determination of Oil/Fat Utilizing High Temperature Solvent Extraction 

Definition 

This method determines crude fat by extracting with petroleum ether. The compounds extracted are predominantly triacylglycerides. Small amounts of other lipids along with minor compounds having some solubility in petroleum ether are also extracted.

Scope 

This method is applicable to solid processed foods with 0-100% fat content.

Apparatus 

1. Analytical Balance—capable of weighing 0.1 mg. 2. Oven—capable of maintaining a temperature of 102 ± 2°C

(ANKOMRD Dryer, ANKOM Technology). 3. Extraction instrument—capable of extracting at 90 ± 2°C.

(XT10, XT15, XT20, ANKOM Technology). 4. Filter Bags—constructed from chemically inert and heat

resistant filter media, capable of being heat sealed closed and able to retain 1 micron particles while permitting solution penetration (XT4, ANKOM Technology).

5. Heat sealer—sufficient for sealing the filter bags closed to ensure complete closure (1915, ANKOM Technology).

6. Desiccant Pouch—collapsible sealable pouch with desiccant inside that enables the removal of air from around the filter bags (MoistureStop weigh pouch, ANKOM Technology).

7. Marking pen—Solvent and acid resistant (F08, ANKOM Technology).

Reagents 

1. Petroleum Ether (B.P. 35-65º C). Depending on the method used, other solvents such as hexanes and ethyl ether may be used. CAUTION: Most fat solvents are extremely flammable. Avoid static electricity and follow safe laboratory practices according to your local regulations when handling solvents.

Precision 

Results from the collaborative study (see Table 1) indicate the precision (Sr, RSDr, r) that the analyst should use as a benchmark for evaluation replication in the same laboratory.

Procedure (see the Fat Extraction Process section of the Operator’s Manual for more detail) 

1. Place a labeled filter bag on the balance and zero the balance. 2. Weigh 1.5 – 2g of sample into the bag and record the weight

(W1). 3. Heat seal the filter bag closed within 4mm of the top to

encapsulate the sample. 4. It is good practice to include a blank bag periodically,

especially when changing solvents, in order to measure any effect on the filter bag.

5. If a meat sample has an expected fat value of > 15% or a plant sample has an expected fat value of > 20%, then place the filter bag with sample in a tared weigh tin (W4) during the oven drying procedure in order to capture fat that may migrate out of the bag during heating.

6. Place samples in the drying oven for 3 hrs (see Numbered Note 1).

7. Cool the dried samples in the Desiccant Pouch, weigh and record the weight (W2).

8. Place filter bags with sample into the bag holder and place in the extractor.

9. Select the extraction time and proceed according to the extraction instruments instructions.

10. When the extraction process is complete place the samples in the oven for 15 – 30 minutes (see Numbered Note 2).

11. Cool in the Desiccant Pouch and weigh (W3).

Calculations 

Meat samples ≤ 15% fat and plant samples ≤ 20% fat  Meat samples > 15% fat and plant samples > 20% fat 

% Crude Fat  =  100 x (W2 – W3)    % Crude Fat  =  100 x ((W2 – W4) ‐ W3) 

    W1          W1 

Where:  W1  =  Original weight of sample  Where:  W4  =  Weight of the weigh tin 

W2  =  Weight of pre‐dried sample with filter bag       

W3  =  Weight of dry sample with filter bag after extraction 

     

 

   

Page 24: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 24       Rev E 4/2/12 

 

 

 Numbered Notes 

1. Pre-drying time of the samples in the filter bags can be greatly reduced by using a microwave oven. However, microwave ovens vary in energy output and samples vary in water and fat composition. You will need to experiment with your microwave oven to determine the proper conditions for use. Contact ANKOM Technology for further information.

2. Some samples with large amounts of carbohydrates (starches and cellulose) may bind small quantities of solvent and require additional post drying. To remove bound solvent, dry at 100°C for 1 – 3 hrs or at 125°C for 30 minutes.

Sample ID oat

mea

l

bro

wni

e m

ix

soy

bean

A

can

ola

soy

bean

mea

l

cor

n A

pou

ltry

star

ter

cat

tle

feed

pig

sta

rter

alf

alfa

cat

foo

d

coo

kies

bkf

t cer

eal

tort

illa

chi

ps

Number of laboratories 12 12 11 9 12 12 11 10 11 11 12 11 12 12Number of replicates 24 24 22 18 24 24 22 20 22 22 24 22 24 24

Filter Bag Average, % Crude Fat 5.8 8.7 20.9 39.0 1.6 3.3 3.3 3.2 5.6 2.4 6.3 22.7 2.3 19.9

Certified Labs Averagea 5.7 8.7 21.1 39.7 1.6 3.6 3.5 3.0 5.5 2.2 6.2 23.1 2.3 20.0AOCS Am 2-93 Avg. for Oilseeds 21.9 44.8 3.6

Sr 0.36 0.20 0.35 0.23 0.14 0.31 0.24 0.18 0.20 0.39 0.27 0.20 0.26 0.39

SR 0.54 0.31 0.63 0.68 0.27 0.42 0.42 0.20 0.28 0.50 0.30 0.20 0.36 0.48

RSDr, % 6.2 2.3 1.7 0.6 8.5 9.5 7.3 5.6 3.6 16.1 4.2 0.9 11.4 2.0

RSDR, % 9.4 3.5 3.0 1.7 16.3 12.7 12.6 6.1 5.0 20.7 4.7 0.9 15.7 2.4

r 0.99 0.56 0.98 0.65 0.39 0.88 0.68 0.51 0.56 1.08 0.75 0.56 0.72 1.09R 1.52 0.86 1.76 1.90 0.75 1.18 1.16 0.55 0.78 1.39 0.83 0.56 1.00 1.35

HORRAT 3.05 1.21 1.19 0.76 4.38 3.87 3.77 1.83 1.61 5.92 1.56 0.35 4.45 0.95

Sample ID dog

foo

d

cra

cker

s

turk

ey

ham

bee

f gr

ound

chi

cken

bre

ast

soy

bean

B

saf

flow

er

pot

ato

chip

s

hot

dog

sau

sage

cor

n B

che

ese

curl

s

cor

n si

lage

Number of laboratories 12 10 12 9 11 11 11 9 11 12 11 12 12 12Number of replicates 24 20 24 18 22 22 22 18 22 24 22 24 24 24

Filter Bag Average, % Crude Fat 6.8 23.8 3.2 11.6 23.8 2.8 19.4 22.5 32.0 39.5 25.7 3.4 30.6 2.3

Certified Labs Averagea 6.9 24.0 3.2 11.3 23.5 2.7 19.7 23.2 32.0 39.0 25.0 3.7 30.5 2.3AOCS Am 2-93 Avg. for Oilseeds 20.8 24.7

Sr 0.35 0.23 0.21 0.30 0.24 0.33 0.38 0.53 0.48 0.35 0.34 0.39 0.48 0.23

SR 0.35 0.23 0.34 0.30 0.36 0.33 0.62 0.83 0.52 0.59 0.51 0.41 0.69 0.51

RSDr, % 5.23 0.96 6.57 2.59 1.01 11.89 1.97 2.36 1.49 0.89 1.33 11.48 1.59 9.87

RSDR, % 5.23 0.96 10.84 2.59 1.49 11.89 3.19 3.69 1.61 1.49 1.98 11.93 2.27 22.45

r 0.99 0.64 0.58 0.84 0.67 0.94 1.07 1.49 1.34 0.98 0.96 1.10 1.36 0.63R 0.99 0.64 0.96 0.84 0.99 0.94 1.73 2.33 1.45 1.65 1.43 1.14 1.94 1.44

HORRAT 1.74 0.39 3.22 0.94 0.6 3.47 1.25 1.48 0.68 0.65 0.81 3.59 0.95 6.36

aAOCS Official Methods Ba 3-38, AOAC 920.39 or equivalent

Table 1. A summary of the statistical analysis of the international collaborative study of the Filter Bag Technique for the proposed Official AOCS Method, “Rapid Determination of Oil/Fat Utilizing High Temperature Solvent Extraction” is presented in this table. Included in the summary is a comparison of the Filter Bag Technique with the results of analysis by AOCS certified laboratories using official AOCS or AOAC methods

Page 25: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 25 

 

Appendix B – Parts & Assemblies 

X84 – Extraction Vessel Assembly          X87 – O‐ring Kit for Pet Ether         X95 – O‐ring kit for High 

Penetration Solvents 

   X86 –  Teflon Insert Assembly with Wave Spring Bag Holder 

     

983.7 – Wave Spring Bag Holder  X96 – Thermister Assembly  X99 – Level Sensor Assembly 

     

X72 – Safety Valve Assembly  X91 – Valve Plunger Assembly  5604 – Pressure Gauge 

    

X89 – Splat Controller  X90 – Splat Display  146.1 – Solvent Filter Element 

      

  

 

Page 26: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 26       Rev E 4/2/12 

 

   

X83 – Charcoal Filter  X97 – Water Filter  X98 – RFI Filter 

   

206 – Fuse 10A  5578 – Fuse 5A       X55 – Valve Kit for High Penetration Solvents 

   

460 – 120V Relay  461 – 220V Relay  462 – 24V Relay 

 

97 – 20A Relay  963 – Power Supply  5653 – On/Off Switch 

         

 

 

  

 

Page 27: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 27 

 

  

 

 

24 – XT Burkert Valve 5/16” orifice  414 – 24V Burkert Valve 1/8” orifice  Z132 – Valve Seals (1 per valve) 

     

101.2 – Bag Holder  X49 – Weigh Tin Desiccator  X45 – Desiccant Pouch 

   

5605 – Vent Flask  SH – High Fat Check Sample  SL – Low Fat Check Sample 

     

TCH1 –   Chiller Assembly 120V with Bypass and Alarm 

    TCH2 – Chiller Assembly 220V with Bypass and Alarm 

      X93 – By‐pass Assembly for Chiller 

   

     

Page 28: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 28       Rev E 4/2/12 

 

Appendix C – Electrical Diagram 

 

Page 29: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

  Operator’s Manual        

   

Rev E 4/2/12      pg. 29 

 

Appendix C – Electrical Diagram (continued)  

 

Page 30: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 Operator’s Manual  

       

 

pg. 30       Rev E 4/2/12 

 

Appendix C – Electrical Diagram (continued) 

 

Page 31: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

   

 

 

 

 

 

 

 

 

 

 

This page intentionally left blank   

Page 32: XT15 - Operator's Manual - Rev E - 4-2-12€™s Manual Rev E 4/2/12 ... The analysis is achieved by measuring the loss of mass due to the ... C2 = Blank filter bag weight after extraction

 

 

Automation saves time and money! 

ANKOM Technology is an international company with products that include… 

 

TDF Dietary Fiber Analyzer 

Automates AOAC 991.43/AACC 32.07.01 IDF/SDF and TDF values Faster, Technician-free Filtering  Computer controlled operation  Reduced per assay costs 

 

A2000 Fiber Analyzer 

Crude Fiber (AOCS Ba 6a-05), ADF, NDF Automatically adds solutions and rinses Batch process - up to 24 samples at one time 

 

XT15 Fat Extractor 

Official Method AOCS Am 5-04 Fully automatic Solvent recovery at 97% or greater Batch process - up to 15 samples at one time 

 

RF Gas Production System 

High sensitivity pressure measurement Applications include: Ruminant Nutrition, Human Digestion, Yeast Activity,

Beer/Wine Fermentation, Biomass-to-Energy analysis (e.g., Ethanol), Biodegradability, Soil respiration, BOD, etc.

Wireless Computer control and data storage 

 

Chemicals 

A wide variety of chemicals used for many different lab operations Pre-mixed solutions available 

Please visit our web site at www.ankom.com for more information.  

2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 986‐8090 Fax: (315) 986‐8091   www.ankom.com 

  


Recommended