www.elsevier.com/locate/jim
Journal of Immunological Methods 284 (2004) 107–118
Immuno-SLM—a combined sample handling
and analytical technique
Madalina Tudorachea, Mariusz Rakb, Piotr P. Wieczorekb,Jan Ake Jonssona, Jenny Emneusa,*
aDepartment of Analytical Chemistry, Lund University, P.O. Box 124, 221 00 Lund, Swedenb Institute of Chemistry, University of Opole, Oleska 48, 45-052, Poland
Received 3 February 2003; received in revised form 3 October 2003; accepted 22 October 2003
Abstract
Immuno-supported liquid membrane (immuno-SLM) extraction is a new technique that makes use of antibody (Ab)–
antigen interactions as the ‘‘extraction force’’ to drive the mass transfer in a selective way. In immuno-SLM, anti-analyte (Ag)
Abs are introduced into the acceptor phase of the SLM unit to trap the Ag that passes from the flowing donor through the SLM
into the stagnant acceptor. The amount of immuno-extracted analyte (AbAg) is quantified by connecting the immuno-SLM unit
on-line with a non-competitive heterogeneous fluorescence flow immunoassay (FFIA) that makes use of a fluorescein-labeled
analyte tracer that titrates the residual excess of Ab present in the acceptor. A restricted access (RA) column is used for the
separation of the two tracer fractions (Ag* and AbAg*) formed, and the eluted AbAg* fraction is measured downstream by a
fluorescence detector.
Factors influencing the optimum immuno-SLM extraction parameters, i.e., donor flow rate, extraction time and type of Ab,
were investigated for immuno extraction of the model analyte atrazine. Immuno-SLM coupled to FFIA (immuno-SLM–FFIA)
and FFIA alone were compared in terms of the assay sensitivities obtained and the sample matrix influence. The concentration
at the mid-point of the calibration curve (IC50) was 16.0F 1.4 and 36F 16 Ag/l, the limit of detection (LOD) was 2.0F 1.1 and
20F 10 Ag/l, and the dynamic range was 2–100 and 20–500 Ag/l atrazine for immuno-SLM–FFIA and FFIA, respectively.
The matrix influence on the FFIA was significant in orange juice and surface water, whereas the influence was minor for
immuno-SLM–FFIA with recoveries between 104% and 115% for 5 Ag/l atrazine in tap water, orange juice and river water.
D 2003 Elsevier B.V. All rights reserved.
Keywords: Immuno extraction; Immunoassay; Supported liquid membrane extraction; Fluorescein; Restricted access; Atrazine; Tap water;
Orange juice; River water
0022-1759/$ - see front matter D 2003 Elsevier B.V. All rights reserved.
doi:10.1016/j.jim.2003.10.014
Abbreviations: SLM, Supported liquid membrane; FFIA, Fluorescence flow immunoassay; SPE, Solid phase extraction; LLE, Liquid–
liquid extraction; LC, Liquid chromatography; PTFE, Polytetrafluoroethylene; RA, Restricted access; LOD, Limit of detection; IC50, Inhibition
concentration at 50%; PBS, Phosphate-buffered saline; EDF, Fluorescein thiocarbamyl ethylene diamine; FITC, Fluorescein isothiocyanate.
* Corresponding author. Tel.: +46-46-222-48-20; fax: +46-46-222-45-44.
E-mail address: [email protected] (J. Emneus).
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118108
1. Introduction
The trace-level determination of pollutants in
complex environmental matrices (surface water, soil,
foodstuffs, etc.) is laborious and time-consuming
because sample pre-treatment procedures involve
many steps (Pichon et al., 1998). Many extractions,
based on, e.g., solid phase extraction (SPE) (Battista
et al., 1988; Yook et al., 1994; Sun et al., 1998) and
liquid– liquid extraction (LLE) (Muir and Baker,
1978; Lee and Stokker, 1986; Yrieix et al., 1996),
have been developed and are widely used in this
field. SPE is a modern extraction technique which
involves analyte (Ag) trapping by a solid sorbent
following elution of the analyte with an organic
solvent. In this way, separation of the analyte from
a complex matrix is achieved. Some drawbacks with
the SPE technique can be insufficient clean up of the
sample, insufficient retention of very polar com-
pounds, limited selectivity and high costs of dispos-
able sorbent materials. The SPE selectivity was found
to be greatly enhanced by using antibody–antigen
interactions in the extraction mechanisms, where an
immuno sorbent was used as a selective solid phase
extractor (Rivasseau and Hennion, 1999; Pou et al.,
1994). Immuno sorbents have been used for the
extraction of many different compounds from various
environmental matrices (surface water, wastewater,
soil, sediment, plant and foodstuff) (Thomas et al.,
1994; Pichon et al., 1998).
LLE is a classical technique for sample prepara-
tion of liquid samples. It provides a large potential
for tuning the extraction by pH adjustments, select-
ing solvents with specific properties and/or incorpo-
rating different reagents. LLE results in simultaneous
enrichment and clean up of samples, but involves
high consumption of organic solvent, and is difficult
to automate and to connect on-line with analytical
instruments. In order to eliminate the drawbacks of
LLE, a technique based on initial extraction of an
analyte from an aqueous sample (donor) into an
immobilized organic phase (organic membrane) fol-
lowed by re-extraction into a second aqueous phase
(acceptor) was developed, called supported liquid
membrane (SLM) extraction (Jonsson and Mathias-
son, 1999; Jonsson and Mathiasson, 2000). The
advantage of SLM extraction compared to other
LLE techniques consists of its capacity for simulta-
neous selective clean-up, concentration and extrac-
tion of the analyte.
In the present paper, a new combined extraction
and analytical technique called immuno-SLM–FIIA,
previously presented with preliminary results for 4-
nitrophenol (Thordarson et al., 2000), is characterized
for extraction and detection of the model analyte
atrazine. The technique is based on the selective
extraction of the analyte from an aqueous donor phase
over an organic liquid membrane (the SLM) into an
aqueous acceptor phase, containing soluble antibod-
ies. The amount of immuno-extracted analyte in the
acceptor is then quantified on-line by a fluorescence
flow immunoassay (FFIA). To verify the mechanism
of immuno-SLM extraction, the immuno-SLM–FFIA
results obtained have been compared with results
obtained with the FFIA alone, i.e., excluding the
immuno extraction step. Some theoretical aspects of
immuno-SLM are discussed and its selectivity and
applicability to measure the model analyte atrazine in
tap and river waters as well as in orange juice are
demonstrated.
2. Materials and methods
2.1. Materials and solutions
2-Chloro-4-ethylamino-6-isopropylamino-1,3,5-
triazine (atrazine) of 99.2% purity was purchased
from the Institute of Organic Industrial Chemistry
(Warsaw, Poland). A stock solution of atrazine (5
mg/l) was prepared in water (by stirring for 1
week) purified with a Milli-Q/RO4 unit (Milli-
pore, Milford, MA, USA) and kept at room
temperature.
Phosphate-buffered saline (PBS) was prepared by
mixing a basic solution (0.15 M Na2HPO4 and 0.15
M NaCl) with an acidic solution (0.15 M NaH2PO4
and 0.15 M NaCl) until reaching a pH of 7.4. All
buffer reagents were purchased from Merck (Darm-
stadt, Germany). Di-n-hexyl ether (97%, Sigma-
Aldrich, Steinheim, Germany) was used to impregnate
the polymeric membrane (PTFE). 0.5 M H2SO4 (95–
97% Merck, Darmstadt, Germany) was used as the
regeneration solution of the organic membrane (to
avoid any memory effect of analyte not extracted from
the organic membrane).
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118 109
Three different antibodies were obtained from two
different sources: Ab I (poly-IgG-anti-atrazine, from
rabbit, lot A9) and Ab II (poly-IgG-anti-atrazine, from
rabbit, lot ARINA P) were kindly provided by Dr.
Sergei Eremin (Lomonosov Moscow State University,
Russia) and Ab III (affinity purified anti-atrazine,
from sheep) was kindly provided by Dr. Ram Abu-
knesha (Kings College, London, UK). The antibody
stock solutions (1 mg/ml), prepared in 0.15 M PBS
(pH 7.4), were kept at + 4 jC and were diluted daily
with 0.15 M PBS (pH 7.4). A s-triazine hapten
derivative iPr/Cl/(CH2)5COOH, kindly provided by
Dr. Sergei Eremin, was used to prepare a fluorescein-
labeled tracer.
Fluorescein thiocarbamyl ethylene diamine (EDF)
was synthesized from fluorescein isothiocyanate
(FITC, isomer I, lot 18H2603, Sigma, St. Louis,
MO, USA) as described by Pourfazaneh et al.
(1980). The carboxylic group of the iPr/Cl/(CH2)5COOH hapten was reacted with EDF, forming a
fluorescent tracer, as described by Onnerfjord et al.
(1998a). The tracer stock solution in methanol was
kept at � 20 jC and the working tracer solutions were
prepared daily by diluting with 0.15 M PBS (pH 7.4).
The concentration of the tracer stock solution (5 AM)
was estimated spectrophotometrically at 492 nm,
assuming that the absorptivity in sodium bicarbonate
buffer (0.05 M, pH 9) was the same as for fluorescein
(8.78� 104 M� 1 cm� 1) Onnerfjord et al. (1998a). All
vials containing tracer were covered with aluminium
foil to protect the tracer against light.
Samples were tap water (Chemical Center, Lund
University, Lund, Sweden), river water (Hoje River, 2
km south of Lund, Sweden) and orange juice with
fruit parts (Monte Carlo, Arla, Stockholm, Sweden).
To remove particle matter from the samples, tap and
river waters were filtered (Millipore, 0.45 Am), where-
Fig. 1. The immuno-SLM–FFIA system: (1) SLM unit; (2, 5, 6) automatic
mixing coil; (8) restricted access column (C8); (9) fluorescence detector.
as the juice samples were first centrifuged and then
filtered. The pH of all samples was adjusted with
NaOH solution to pH 7.4.
2.2. System set-up and method
2.2.1. SLM unit
The SLM unit (Chemical Center Workshop, Lund
University, Lund, Sweden) consisted of two blocks of
inert material (one of PEEK and one of PTFE), each
with a machined groove (2.5� 0.1� 40 mm). When
the blocks were clamped together, a porous membrane
(PTFE, TE 35, Schleicher and Schuell, Dassel, Ger-
many, length 7.10 cm, width 0.60 cm with a 180-Am-
thick supporting polyester backing, 0.2 Am i.d. pore
size and 60–80% porosity) impregnated with the
organic solvent di-n-hexyl ether was held between
them. The membrane separated two identical channels
serving as donor and acceptor, respectively, with a
volume of approximately 10 Al each. The SLM
membrane was prepared by immersing the porous
polymer support membrane in the organic solvent
for 30 min (immobilization time), after which the
membrane was placed in the SLM unit as described
above. By pumping 3 ml water through each channel
(acceptor and donor), the excess of organic solvent
was removed from the SLM membrane surfaces,
making the unit ready for analyte extraction.
2.2.2. Instrumentation
The immuno-SLM–FFIA system is shown in
Fig. 1 and contained the following components:
SLM unit (1), a peristaltic pump (Minipuls 3,
Gilson, Villiers-le-Bel, France) (3), three syringe
pumps (P/N 50300 with six-port valve, Kloehn,
Las Vegas, NV, USA) (2, 5, 6), a 50-Al loop placed
on a manual six-port injection valve (Rheodyne,
syringe pumps; (3) peristaltic pump; (4) manual injection valve; (7)
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118110
California, USA) (4), a 250-Al mixing coil (1 mm
i.d. and 32 cm length) (7), a restricted access (RA)
column (PEEK, 2 mm i.d. and 2 cm length, packed
with LiChrospher 60, RP-8 ADS, 25 Am porosity
(Merck)) (8), and a fluorescence detector (L-7480,
Hitachi, Tokyo, Japan) (9), set at 515 nm excitation
wavelength and 490 nm emission wavelength. The
different parts of the flow set-up were connected
using PEEK tubing (0.25 mm i.d.) and finger tight
screw fittings.
2.2.3. Immuno-SLM–FFIA procedure
Fig. 1 shows the immuno-SLM system used. The
Ab solution (in 0.15M PBS pH 7.4) was dispensed into
the acceptor channel of the SLM unit (1) by a syringe
pump (2) after which the flow was stopped. The
sample, containing analyte (Ag = atrazine), was pump-
ed for 15 min by a peristaltic pump (3) with a constant
flow rate through the donor channel of the SLM unit
(1). The Ag molecules diffused from the donor over
the SLM membrane to the acceptor, in which they
formed antibody–analyte complexes (AbAg). Since
the Ab was present in excess, both free Ab and AbAg
complex were found in the acceptor after extraction.
The donor flow was then stopped and the acceptor
content (Ab and AbAg) was dispensed into the loop
of a manual injection valve (4) by a second syringe
pump (5). The injection valve (4) was switched and the
loop content was dispensed and mixed with a tracer
(Ag*) solution in excess, pumped with a third syringe
pump (6), in a mixing coil (7). Pumps (5) and (6) were
stopped for 5 min to allow efficient incubation between
the residual excess of Ab with excess of Ag*. After
incubation, the content of the mixing coil, now contain-
ing Ag +AbAg +AbAg* +Ag*, was passed through a
RA column (8) where the Ag and Ag* fractions were
trapped and theAbAgandAbAg* fractionswere eluted.
The AbAg* fraction was detected by a fluorescence
detector (9) where the analytical signal obtained was
inversely proportional to the Ag concentration in the
sample.
2.2.4. FFIA procedure
The FIIA procedure was identical to the immuno-
SLM–FFIA procedure, excluding the SLM extraction
step. Suitable concentrations of Ab and Ag were pre-
mixed and incubated for 15 min and then introduced
into the acceptor by the syringe pump (2), while the
donor channel contained pure buffer solution. The
subsequent assay steps were identical to those de-
scribed above for the immuno-SLM–FFIA procedure.
3. Results and discussion
Considering the basic principle of SLM extraction,
involving antigen–antibody interactions as the extrac-
tion driving force, the immuno-SLM technique was
introduced (Thordarson et al., 2000) in order to
improve SLM extraction selectivity. The SLM is a
nonporous membrane(Jonsson and Mathiasson,
2000), which contains an organic solvent (e.g., di-n-
hexyl ether, n-undecane or a combination of both
solvents) (Trocewicz, 1996; Chimuka et al., 1997;
Megersa et al., 2000) immobilized in the pores of the
support material (PTFE) by capillary forces. In
immuno-SLM, a continuous flow of sample, contain-
ing the Ag, is passed through the donor channel while
a stagnant solution of anti-analyte antibodies is pres-
ent in the acceptor. Uncharged analyte molecules are
transported by diffusion from the donor through the
SLM into the acceptor where the formation of strong
AbAg complexes takes place, thus preventing the Ag
from re-entering the membrane. Since the analyte is
found as AbAg in the acceptor and as Ag in the donor,
the mass transfer of the diffusing species (Ag) will be
unaffected by the total concentration of extracted Ag
in the acceptor channel, allowing a high degree of
analyte enrichment.
An on-line non-competitive FFIA system was de-
veloped to quantify the degree of immuno-SLM ex-
traction. The acceptor content, after extraction
(AbAg + Ab + Ag), was mixed with excess of a
fluorescein-labeled tracer to titrate the unbound
residual Ab. The resulting reagent mixture (AbAg +
AbAg* +Ag* +Ag) was then introduced into a restrict-
ed access column to separate the free and bound
fractions, as described in the experimental section.
3.1. Immuno-SLM–FFIA
The extraction of analytes from the first aqueous
phase (the donor) to the second aqueous phase (the
acceptor) through the SLM membrane can be de-
scribed by two quantitative parameters: the extraction
efficiency (E) and the enrichment factor (Ee). E is
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118 111
defined as the fraction of analyte that is transported
from the donor phase through the SLM membrane to
the acceptor phase, see Eq. (1). Ee is defined as the
ratio between the analyte concentration in the acceptor
and donor, respectively, as seen in Eq. (2) (Knutsson
et al., 1996).
E ¼ VACA=VSCS ð1Þ
Ee ¼ CA=CS ¼ EVS=VA ð2Þ
where VA and VS are the acceptor channel volume
and the total volume of extracted sample, respec-
tively, and CA and CS are the analyte concentra-
tion in the acceptor and in the extracted sample,
respectively.
The rate of mass transfer over the membrane is
proportional to the difference between the analyte
concentration in the donor and acceptor (DC), which
can be described by Eq. (3).
DC ¼ CS � aACA ð3Þ
It is assumed that the analyte is in its extractable
form in the donor, and aA represents the fraction of
analyte that is in its extractable form in the acceptor
(Jonsson et al., 1993).
The analyte concentration in the acceptor increases
as the extraction proceeds and thus DC approaches
zero. When the enrichment of the analyte in the
acceptor reaches the maximum level (Ee =Ee(max)),
then DC = 0, mass transfer stops, and Eq. (3) can be
re-written to Eq. (4) (Chimuka et al., 1998).
EeðmaxÞ ¼ ðCA=CSÞmax ¼ 1=aA ð4Þ
The equilibrium reaction between the antibody and
the analyte, taking place in the acceptor, and the affinity
constant (K) that characterizes the corresponding equi-
librium may be simply described by the law of mass
action, given by Eq. (5):
K ¼ ½AbAg�=½Ab�½Ag� ð5Þ
where [AbAg] is the concentration of the immuno
complex, [Ab] is the concentration of free antibody
and [Ag] is the concentration of extractable analyte
in the acceptor, and the fraction of extractable
analyte in the acceptor channel (aA) can thus be
written as Eq. (6):
aA ¼ ½Ag�=ð½Ag� þ ½AbAg�Þ ð6Þ
The maximum enrichment factor (Ee(max)) for
immuno-SLM extraction is obtained by combining
Eqs. (4), (5) and (6), giving Eq. (7) (Thordarson et al.,
2000):
EeðmaxÞ ¼ 1þ K½Ab� ð7Þ
Eq. (7) shows that the affinity of the antibody for
the antigen (K) as well as the antibody concentration
in the acceptor influence the value of Ee(max) of the
immuno-SLM extraction. To obtain a high degree of
immuno extraction, an antibody with a high affinity
for the analyte must therefore be used.
Apparent extraction efficiency (Eapp) and apparent
enrichment factor (Eeapp) are now introduced to point
out that the experimental values obtained using the
above equations with the present system are not true
values since the sensitivity of the FFIA used directly
influences the extraction parameters, i.e., any change
in antibody concentration in the acceptor will be
followed by a change in the FFIA sensitivity and a
change in antibody affinity will change the FFIA
sensitivity due to the complex interplay between
different antibody affinity ratios for analyte versus
tracer (Wild, 1994; Diamandos and Chistopoulos,
1996). These apparent values should thus be looked
upon as comparative constants describing the whole
immuno-SLM–FIIA system.
3.1.1. Mechanism of the immuno-SLM process
In order to test the mechanism of the immuno-SLM
process, an antibody–antigen system that we knew
worked well in the simple FFIA format was used. To
minimize analyte trapping by the ‘‘conventional’’
SLM process (Chimuka et al., 1997), the pH of the
donor and the acceptor should be the same and the
analyte be neutral at this pH. For this purpose the s-
triazines, and, in particular, atrazine was found to be
the ideal analyte being a weak base (pKa = 1.68) and
thus a pH of 7.4 could be used in both acceptor and
donor.
Three different experiments were performed as
shown in Fig. 2. In Experiment I, atrazine concen-
Fig. 2. Visualizing the immuno-SLM process: (o) Experiment I—immuno-SLM–extraction of analyte from a continuous flowing donor into an
antibody containing acceptor; (5) Experiment II—pre-incubation of antibodies and analyte off-line, then injection into the system (Fig. 1),
briefly passing over the acceptor, and with a stagnant donor flow (no analyte present in donor); (x) Experiment III—pre-incubation of antibody
and analyte offline as in Experiment II, but then kept in the acceptor for 15 min and with a continuously pumped donor (no analyte present in
donor). Conditions: SLM: PTFE support (7.1 cm length and 0.6 cm width) impregnated with di-n-hexyl ether, [Ag*]w = 5 nM iPr/Cl/(CH2)5CO-
EDF I, [Ab III]0.25 = 1.3 Ag/ml, donor flow rate = 300 Al min� 1 and extraction time = 15 min. B*= signal of the antibody bound tracer and
B0.25* = signal at 25% binding of the tracer.
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118112
trations in the range 0–1000 Ag/l were extracted for
15 min from a continuous donor flow into the
stagnant Ab-containing acceptor, i.e., the immuno-
SLM procedure. In Experiment II, the Ab was pre-
incubated off-line with atrazine and then introduced
into the acceptor while the donor flow, containing no
analyte, was kept stagnant. In this experiment, the
acceptor content was immediately transferred into the
FFIA system, before any free atrazine molecules
could diffuse from the acceptor to the donor. Thus,
no analyte extraction was performed and an experi-
ment corresponding to a simple FFIA was attempted
with E = 0. In Experiment III, the Ab was pre-
incubated offline with the analyte and introduced
into the acceptor channel, as in Experiment II, but
then kept there for 15 min. The donor flow, con-
taining only buffer solution, was continuously
pumped through the channel, thus testing the degree
of extraction of the analyte from the acceptor to the
donor. By comparing Experiments I and II in Fig. 2,
it can be seen that extraction and enrichment of
atrazine was achieved with Experiment I, since a
shift of the calibration curve to lower concentrations
was observed and the sensitivity for the analysis was
improved (limit of detection LOD10% = 2 Ag/l and
IC50 = 16 Ag/l) compared with Experiment II
(LOD10% = 20 Ag/l and IC50 = 36 Ag/l). If Experi-
ments II and III are compared, a shift of the
calibration curve to even higher concentrations and
thus lower sensitivity was observed for Experiment
III (LOD10% = 40 Ag/l and IC50 = 74 Ag/l), indicatingthat about half of the atrazine was lost from the
acceptor by diffusion through the membrane to the
donor. Here, the continuous pumping of the donor
creates an additional driving force for the diffusion
of the analyte from the acceptor into the donor.
However, due to the high partition coefficient
(Kp>1) of atrazine between the organic and aqueous
phases it is very likely that some of the analyte was
lost to the membrane phase also in Experiment II.
The experiments illustrated in Fig. 2 suggest that the
diffusion of atrazine through the SLM membrane
was a reversible process and that the presence of
antibodies in the acceptor resulted in extraction and
enrichment of the analyte in the acceptor.
An apparent low dose hook effect was seen in both
Experiments II and III, i.e., when no immuno extrac-
tion was performed. Based on the literature on this
M. Tudorache et al. / Journal of Immunolo
topic (Bachas et al., 1984; Barbarakis et al., 1993), we
have no reasonable scientific explanation for this
effect in our experiments based on e.g., the type of
tracer label, size of analyte, size of antibody). There
was, however, a significant difference between Ex-
periment I (no low dose hook effect) and Experiments
II and III (with a low dose hook effect). The analyte
concentrations given in Fig. 2 for Experiment I are the
concentrations in the donor, whereas the ones given
for Experiments II and III are the analyte concentra-
tions in the acceptor, since no extraction was per-
formed. This means that the analyte concentration in
the acceptor after extraction/enrichment was always
higher in Experiment I than the analyte concentrations
present in the acceptor during Experiments II and III.
3.1.2. Optimization of immuno-SLM extraction
The theory of traditional SLM extraction describes
the quantitative parameters (E and Ee) of the extrac-
tion as depending on donor flow rate and extraction
time, as well as on the polarity of the organic
membrane (Jonsson et al., 1993). Thus, high efficien-
cy of immuno-SLM extraction should be obtained by
optimizing conditions such as donor flow rate, extrac-
tion time, stability of organic membrane and the type
and concentration of the antibody in the acceptor.
These factors will be discussed in the following
sections.
Fig. 3. Dependence of the apparent enrichment factor (Eeapp) and appare
[atrazine] = 5 Ag/l, otherwise, as in Fig. 2.
3.1.2.1. Donor flow rate, extraction time and stability
of organic liquid membrane. The SLM theory states
that the enrichment factor (Ee) increases and the
extraction efficiency (E) decreases with increasing
donor flow rate (Jonsson and Mathiasson, 1999). This
theory was confirmed also for immuno-SLM, as
shown in Fig. 3 for the extraction of 5 Ag/l atrazineat different flow rates. It can be seen that Ee
app
increased with donor flow rate up to approximately
450 Al min� 1, while Eapp decreased within the tested
range of flow rates.
Since the Kp (partition coefficient between organic
and aqueous phases) for atrazine was high, the ex-
traction of atrazine is limited by mass transfer in the
donor (donor-controlled extraction) (Jonsson and
Mathiasson, 1999), suggesting that the highest donor
flow rate should be used (Knutsson et al., 1996).
However, a high donor flow rate reduces the lifetime
of the organic membrane, which is why a compromise
was made, and a flow rate of 300 Al min� 1 was
chosen for all further experiments.
Another factor influencing the enrichment factor is
the extraction time, i.e., the time that the analyte is
passed through the donor at a constant flow rate (300
Ag/ml). Eeapp increases with the extraction time until a
sufficient concentration of atrazine in the acceptor
channel has been reached and equilibrium between
the analyte concentrations of all the three phases has
gical Methods 284 (2004) 107–118 113
nt extraction efficiency (E app) on the donor flow rate. Conditions:
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118114
been established (acceptor–organic–donor). This
occurs after 15 min and in all further experiments,
an extraction time of 15 min was used.
The nature of the organic solvent has an important
influence on the rate of mass transfer and selectivity in
SLM extraction. Low viscosity, low volatility and low
solubility in water are physical properties that must
characterize the organic solvent immobilized in the
SLM membrane (Audunsson, 1986). Di-n-hexyl ether
was previously found to be the best organic solvent
for s-triazine extraction by traditional SLM (Chimuka
et al., 1997) and was thus also used in this work. The
stability of the SLM membrane (PTFE impregnated
with di-n-hexyl ether) was tested by immuno-SLM
extractions of 10 and 100 Ag/l atrazine. After about 50extractions, the Eapp decreased with about 25% for
each concentration. This is mainly due to the fact that
the organic solvent is lost from the SLM membrane
during the washing step (the donor channel was
washed with 300 Al of 0.5 M H2SO4 after each
extraction), which is performed to eliminate the mem-
ory effect in the membrane. It was thus found that one
liquid membrane could be used for approximately
thirty extractions before the membrane had to be
exchanged.
A good reproducibility of the immuno-SLM ex-
traction was obtained when testing four different
Fig. 4. Reproducibility of immuno-SLM—FFIA extraction for atrazine
Conditions: as in Fig. 2.
membranes (see Fig. 4), considering that each new
SLM membrane differed from the old one not only in
terms of a fresh organic liquid, but also with a new
PTFE support. Therefore, immuno-SLM extraction
can be reproduced in a satisfactory way even when
the SLM membrane is replaced.
3.1.2.2. Type of antibody and antibody concentra-
tion. The efficiency of the immuno-SLM extraction
is governed by two factors: the affinity of the antibody
for the analyte and the antibody concentration (Eq.
(7)). As stated above, a conflict appears since the
sensitivity of the FFIA used to quantify the immuno
extraction will change as soon as the Ab concentration
or the Ab affinity changes (Wild, 1994; Diamandos
and Chistopoulos, 1996). A compromise between
high efficiency of the immuno extraction and high
sensitivity of the FFIA must thus be made in order to
attain the maximum performance of immuno-SLM–
FFIA system, as will be discussed below.
Three different antibodies (Ab I–III) were com-
pared for extraction of atrazine, using the same tracer
at a fixed concentration. Ab III and Ab I–II differ
with respect to their source (see experimental section)
and in the fact that Ab III is affinity purified using an
analyte affinity column, so the antiserum contains
antibody clones which to different degrees all have
using four different SLM membranes with the same composition.
Table 1
The influence of the type of antibody on the immuno-SLM
extraction parameters
Type of
antibody
[Ab]25%(Ag/ml)
Eapp
(%)
Eeapp LOD
(Ag/l)
Ab I 150 2.6 12 1
Ab II 300 1.0 4.6 4
Ab III 1.3 4.3 19 2
Conditions: as in Fig. 2. (The values in the table are the mean values
of duplicate measurements.)
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118 115
affinity for the analyte. The other two are original
antisera simply purified by ammonium sulfate precip-
itation, meaning that only a small fraction of the
antibody clones have affinity for the analyte.
Working antibody concentrations, leading to 25%
binding of the tracer ([Ab]25), was chosen for all three
cases, followed by analyte calibrations at the fixed
antibody concentrations. In Table 1, the working
antibody concentrations, apparent extraction parame-
ters (Eapp and Eeapp) and sensitivities (LOD) are shown
for the three antibody systems.
As seen in Table 1, significantly lower [Ab]25 was
needed for Ab III compared to the other two, and this,
in turn, led to the highest Eapp and Eeapp. On the other
hand, when looking at the sensitivities of the three
assays, Ab I led to the lowest LOD. Due to the large
Fig. 5. Determination of atrazine by FFIA (.) and immuno-SLM–FFIA (noff-line with the Ab before injection into the acceptor in the system in Fi
amount of Ab I needed to perform each assay all
further experiments were performed with Ab III, an
antiserum that performed well and could be used in
minimum amounts.
3.2. Immuno-SLM–FFIA versus FFIA
Keeping the optimum conditions as specified
above, the immuno-SLM–FFIA and the FFIA were
compared (see Fig. 5) in terms of sensitivity, preci-
sion of measurements and the possibility of applica-
tions in different sample matrixes. The LOD and IC50
for atrazine decreased from 20F 10 to 2.0F 1.1 Ag/l and 36F 16 to 16.0F 1.4 Ag/l, respectively, moving
from the FFIA to the immuno-SLM–FFIA format.
The higher LOD and IC50 obtained for the present
FFIA system compared to a similar system presented
previously (Onnerfjord et al., 1998b) is due to the
fact that the amount of antibody is substantially
higher and the tracer is added in a secondary flow
and mixed in a mixing coil with the eluting acceptor
(Fig. 1).
The ‘‘within-assays’’ and ‘‘between-assays’’ var-
iances were calculated for immuno-SLM–FFIA, con-
sidering the assays in Fig. 5. The ‘‘between-assays’’
variance was higher than the ‘‘within-assays’’ vari-
). In FFIA, the analyte is not enriched, i.e., the analyte is incubated
g. 1. Conditions: as in Fig. 2.
Table 2
[Atrazine]
(Ag/l)Tap
water (%)
Orange
juice (%)
River
water (%)
(a) Recoveries of atrazine from real samples using FFIA
50 99 nd 34
100 99 nd 36
500 95 nd 23
(b) Recoveries for atrazine from real samples using
immuno-SLM–FFIA
5 115 104 111
10 137 153 127
50 145 78 102
100 100 97 96
nd: not detectable.
Conditions: as in Fig. 2. (The values in the table are the mean values
of triplicate measurements.)
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118116
ance giving a calculated F = 19.2 compared with the
one tailed critical F (3.587) for 95% confidence,
suggesting that the repeatability of the extraction
was better than the reproducibility.
3.2.1. Application to real samples
Different sample matrices such as tap water, river
water and orange juice were spiked with atrazine and
the recoveries obtained with FFIA alone and with
immuno-SLM–FFIA were compared with those for
reagent water spiked with atrazine (standard solutions
Fig. 6. Comparison of immuno-SLM–FFIA of atrazine in spiked
of atrazine), as presented in Table 2a and b, respec-
tively. Different spiking levels for the two systems had
to be chosen due to the difference in sensitivity of the
systems, although, all were within the dynamic range
of the assays. For the FFIA (Table 2a), the recovery of
atrazine at different concentrations in the case of tap
water was close to 100%, while the recovery of
atrazine in river water and orange juice was less
satisfactory, which means that the sample matrices
had a strong impact on the FFIA under the conditions
used. For the immuno-SLM–FFIA system, the recov-
eries were in general much better (Table 2b) in almost
all cases within the acceptable limits (between 70%
and 120%) as recommended by the guidelines pub-
lished by US EPA for analysis of environmental
samples (Krotzky and Zeeh, 1995), demonstrating
the reduction of matrix effects and the enrichment of
the analyte by immuno extraction of the analyte
through the SLM membrane.
Finally, the analysis of atrazine from orange juice
was the topic of three inter-laboratory experiments.
Juice samples spiked with unknown atrazine concen-
trations were analysed by immuno-SLM–FIIA. The
values experimentally determined by immuno-SLM–
FIIA were compared with the expected values, as
shown in Fig. 6, in which a relatively high correlation
coefficient was obtained (R = 0.94). Comparing the
constants in the equation of the regression line
orange juice with the real values. Conditions: as in Fig. 2.
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118 117
(a = 0.59F 18.42 and b = 0.78F 0.10, 95% confi-
dence interval) with the ‘‘ideal’’ situation (a = 0 and
b = 1), it can be seen that the effect of immuno-SLM–
FFIA led to an underestimation of the real concentra-
tion value of the analyte by approximately 20%.
3.3. Conclusions
A new immuno extraction–detection technique,
immuno-SLM–FFIA, was investigated for the analyte
atrazine in terms of extraction efficiency, enrichment
factor, flow rate, extraction time, sensitivity and selec-
tivity in different sample matrices. The immuno-
SLM–FFIA results were compared with a FFIA
system without extraction, demonstrating that this
new technique results in automatic and simultaneous
enrichment and sample clean up of the analyte in
sample matrices such as tap and river water and
orange juice with a total analysis time of 27 min
(compared to 60 min for a conventional SLM-LC ana-
lysis of atrazine).
The presented immuno-SLM–FFIA system (unla-
belled antibodies in the acceptor for analyte extraction
and titration of residual free antibodies of the acceptor
with an analyte tracer) is at this initial stage, a
relatively complex assay system in which an increase
in antibody concentration leads to increased extraction
efficiency and enrichment in the immuno-SLM part of
the system, but where the sensitivity of the FFIA used
for quantifying the extraction is counteracted by this
parameter, thus making it difficult to optimize the
conditions of the immuno SLM extraction. A more
suitable flow immunoassay format in which some of
these counteracting effects can be eliminated is a
system that makes use of labeled antibodies in the
acceptor and an antigen support to trap the residual
excess of labeled antibodies.
Acknowledgements
Financial support is kindly acknowledged from the
European Commission (EC contracts: ENV4-CT97-
0476 (INExSPORT), IC15-CT98-0138 (BIOTOOLS),
IC15CT98-0910 (MEBFOOD), ICA2-CT-2000-
10033 (BIOFEED) and INTAS contract 99-0995),
the Swedish Council for Forestry and Agricultural
Research (SJFR), the Swedish Research Council
(Vetenskapsradet) and the Swedish Environmental
Protection Agency (NVV-Naturvardsverket). The
authors are also very grateful for the kind supply of
antibodies and analyte hapten derivatives from Dr. R.
Abuknesha (Kings College, London, UK), M.-P.
Marco (CID-CSIC, Barcelona, Spain) and Dr. S.
Eremin (M.V. Lomonosov Moscow State University,
Russia) as well as restricted access material from Dr.
D. Lubda (Merck) and Prof. K.-S. Boos.
References
Audunsson, G., 1986. Aqueous/aqueous extraction by means of a
liquid membrane sample cleanup and preconcentration of amines
in a flow system. Anal. Chem. 58, 2714–2723.
Bachas, L.G., Lewis, P.F., Meyerhoff, M.E., 1984. Cooperative in-
teraction of immobilised folate binding protein with enzyme–
folate conjugates: an enzyme-linked assay for folate. Anal.
Chem. 56, 1723–1728.
Barbarakis, M.S., Qaisi, W.G., Daunert, S., Bachas, L.G., 1993.
Observation of ‘‘Hook Effects’’ in the inhibition and dose– re-
sponse curves of biotin assays based on the interaction of bio-
tinylated glucose oxidase with (strept)avidin. Anal. Chem. 65,
457–460.
Battista, M., Di Corcia, A., Marchetti, M., 1988. High-performance
liquid chromatographic method for determining triazine herbi-
cide residues in soil. J. Chromatogr. 454, 233–242.
Chimuka, L., Nindi, M.M., Jonsson, J.A., 1997. Supported liquid
membrane enrichment studies of natural water samples applied
to liquid chromatographic determination of triazine herbicides.
Int. J. Environ. Anal. Chem. 68, 429–445.
Chimuka, L., Megersa, N., Norberg, J., Mathiasson, L., Jonsson,
J.A., 1998. Incomplete trapping in supported liquid membrane
extraction with a stagnant acceptor for weak bases. Anal. Chem.
70, 3906–3911.
Diamandos, E.P., Chistopoulos, T.K., 1996. Immunoassay. Acade-
mic Press, San Diego, CA.
Jonsson, J.A., Mathiasson, L., 1999. Liquid membrane extraction in
analytical sample preparation: I. Principles. TrAC 18, 318–325.
Jonsson, J.A., Mathiasson, L., 2000. Membrane-based techniques
for sample enrichment. J. Chromatogr. A 902, 205–225.
Jonsson, J.A., Lovkvist, P., Audunsson, G., Nilve, G., 1993. Mass
transfer kinetics for analytical enrichment and sample prepara-
tion using supported liquid membranes in a flow system with
stagnant acceptor liquid. Anal. Chim. Acta 277, 9–24.
Knutsson, M., Nilve, G., Mathiasson, L., Jonsson, J.A., 1996. Sup-
ported liquid membranes for sampling and sample preparation
of pesticides in water. J. Chromatogr. A 754, 197–205.
Krotzky, A., Zeeh, B., 1995. Applied chemistry division on agro-
chemicals. Pure Appl. Chem. 67, 2065–2088.
Lee, H.B., Stokker, Y.D., 1986. Pesticide and industrial chemical
residues—analysis of eleven triazines in natural waters. J. As-
soc. Off. Anal. Chem. 69, 568–572.
M. Tudorache et al. / Journal of Immunological Methods 284 (2004) 107–118118
Megersa, N., Solomon, T., Chandravanshi, B.S., Jonsson, J.A.,
2000. Sample clean-up, enrichment and determination of s-tria-
zine herbicides from southern ethiopian lakes using supported
liquid membrane extraction. Bull. Chem. Soc. Ethiop. 14, 9–24.
Muir, D.C.G., Baker, B.E., 1978. Method for routine semi-quan-
titative determination of hydroxy-s-triazine in soils. J. Agric.
Food Chem. 26, 420–424.
Onnerfjord, P., Eremin, S., Emneus, J., Marko-Varga, G., 1998a.
Fluorescence polarisation for immunoreagent characterisation.
J. Immunol. Methods 123, 31–39.
Onnerfjord, P., Eremin, S.A., Emneus, J., Marko-Varga, G.,
1998b. High sample throughput flow immunoassay utilising
restricted access columns for the separation of bound and free
label. J. Chromatogr. A 800, 219–230.
Pichon, V., Bouzige, M., Hennion, M.-C., 1998. New trends in
environmental trace-analysis of organic pollutants: class-selec-
tive immunoextraction and clean-up in one step using immuno-
sorbents. Anal. Chim. Acta 376, 21–35.
Pou, K., Ong, H., Adam, A., Lamothe, P., Delahaut, P., 1994.
Combined immunoextraction approach coupled to a chemilumi-
nescence enzyme immunoassay for the determination of trace
levels of salbutamol and clenbuterol in tissue samples. Analyst
119, 2659–2662.
Pourfazaneh, M., White, G.W., Landon, J., Smith, D.S., 1980. Cor-
tisol directly determined in serum by fluoroimmunoassay with
magnetizable solid phase. Clin. Chem. 26, 730–733.
Rivasseau, C., Hennion, M.-C., 1999. Potential of immunoextraction
coupled to analytical and bioanalitycal methods (liquid chroma-
tography, ELISA kit and phosphatase inhibition test) for an im-
proved environmental monitoring of cyanobacterial toxins. Anal.
Chim. Acta 399, 75–87.
Sun, F., Littlejohn, D., David Gibson, M., 1998. Ultrasonication
extraction and solid phase extraction clean-up for determination
of US EPA 16 priority pollutant polycyclic aromatic hydrocar-
bons in soils by reversed-phase liquid chromatography with
ultraviolet absorption detection. Anal. Chim. Acta 364, 1–11.
Thomas, D.H., Beck-Westermeyer, M., Hage, D.S., 1994. Determi-
nation of atrazine in water using tandem high-performance im-
munoaffinity chromatography and reversed-phase liquid chroma-
tography. Anal. Chem. 66, 3823–3831.
Thordarson, E., Jonsson, J.A., Emneus, J., 2000. Immunological
trapping in supported liquid membrane extraction. Anal. Chem.
72, 5280–5284.
Trocewicz, J., 1996. Determination of herbicides in surface water
by means of a supported liquid membrane technique and high-
performance liquid chromatography. J. Chromatogr. A 725,
121–127.
Wild, D., 1994. The Immunoassay Handbook. Stockton Press, N.Y.
USA.
Yook, K.-S., Hong, S.-M., Kim, J.-H., 1994. Comparison of
liquid– liquid extraction and solid-phase extraction coupled with
GC/MS for determination of priority pollutants in water. Anal.
Sci. Technol. 7, 441–453.
Yrieix, C., Gonzalez, C., Deroux, J.M., Lacoste, C., Leybros, J.,
1996. Countercurrent liquid/liquid extraction for analysis of or-
ganic water pollutants by GC/MS. Water Res. 30, 1791–1800.