UMDNS Information
This Product Comparison covers the followingdevice terms and product codes as listed inECRI Institute’s Universal Medical DeviceNomenclature System™ (UMDNS™):
Audiometers [10-228] Audiometric Booths [10-229]
Audiometers; Audiometric Booths
Scope of this Product Comparison
This Product Comparison covers pure-tone audiometers used in routine testing and screening for hearing
threshold, as well as pure-tone audiometers with speech audiometry capabilities. Also covered are portable and
nonportable single- and double-room audiometric booths that have either single- or double-wall construction.
Tympanometers, middle-ear analyzers, auditory evoked-potential units, and newborn hearing screening
devices are excluded. For more information, see the Product Comparisons titled Recorders, Graphic, Evoked-
Potential; Electromyographs; Electronystagmographs and Auditory Function Screening Devices, Newborn. These devices are also called: audiometric evoked-potential
units, audio spectrum analyzers, tinnitus masker analyzers,
acoustic casewalls, acoustic chambers, sound-treated rooms.
Purpose
Audiometers measure and characterize hearing loss by
determining an individual’s hearing threshold (the lowest
audible sound to which the individual will respond 50% of the
time) for pure tones and speech and then comparing that
threshold with a standard range of normal threshold values.
These normal audiometric threshold values represent the mean
lowest level at which a group of young adults can detect a
selected sound.
Hearing loss can result from lesions anywhere along the
auditory tract. When the outer and/or middle ear is affected, the hearing loss is termed conductive; when the
inner ear or auditory nerves are affected, it is termed sensorineural. Abnormal conditions that can cause a
conductive hearing loss include obstruction of the outer ear canal (e.g., foreign mass, swelling), neoplasms,
congenital malformation, perforated tympanic membrane, disruption or fixation of the ossicles (bones of the
middle ear), and fluid and scarring in the middle ear. Abnormal conditions that can cause a sensorineural hearing
loss include congenital malformation, head trauma, exposure to intense noise, drug-induced toxicity, genetic
factors, aging, infectious diseases, and neural tumors. Audiometry aids in diagnosis of otologic and/or neurologic
hearing loss and selection of therapy.
Pure-tone audiometry measures hearing sensitivity (hearing
threshold) for a series of single-frequency signals chosen from the
normal range of human hearing. Although this method can assess the
degree of hearing loss, it cannot characterize the extent of impaired
speech discrimination (understanding of speech). Speech audiometry
measures hearing sensitivity for spoken material. In addition, spoken
material can be used to test for speech discrimination.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 2/116
Audiometers; Audiometric Booths
2 ©2007 ECRI Institute. All Rights Reserved.
Figure 1. Examples of pure-tone air-conduction audiograms: ( A ) X-Y graph ( ● = right ear, ▲ = left ear), ( B ) bar graph (hearing
threshold level in dB, across; frequency in Hz, down), and ( C ) table format (L/dB, R/dB = hearing threshold level in dB of left andright ears, respectively). Bar graph (B) and table (C) courtesy of Tremetrics, Inc.
Audiometric booths provide a consistent and controlled acoustic environment and keep background noise at
an acceptably low level to provide a suitable test area for clinical audiometry or research.
Principles of operation
Audiometers
In the normal ear, sound waves travel through air in the outer ear and cause bones in the middle ear to vibrate.The inner ear converts these mechanical vibrations to hydraulic energy and ultimately to neural impulses that are
transmitted along the auditory nerve to the brain, where it is perceived as sound.
When a patient has a suspected hearing loss, a physician may order an audiometric analysis; screening and
diagnostic testing are then directed by the audiologist. Baseline and annual screenings may be performed for
workers at risk of hearing loss from occupational noise exposure as part of a hearing conservation program and in
compliance with the Occupational Safety and Health Administration (OSHA) standards. The conventional
procedure for determining hearing threshold begins with familiarizing the patient with the stimulus. The test
proceeds according to a protocol recommended by the American Speech-Language-Hearing Association; the
protocol requires a fixed number of positive responses to stimuli (pure tones or speech) presented in sequences of
increasing intensity.
The weakest sound heard at a selected frequency is the hearing level in decibels (dB HL) for that particularfrequency. This is a relative value; the intensity reference is 0 dB HL, or audiometric 0, which corresponds to the
average threshold response of a group of 18- to 25-year-olds with no otologic pathology. The sensitivity of the
normal ear varies with frequency; therefore, 0 dB HL represents different levels of sound pressure at different
frequencies. (Negative dB HL readings indicate that hearing sensitivity is greater than normal at that p articular
frequency.)
Pure-tone audiometers typically generate a series of frequencies between 125 and 8,000 Hz, depending on the
type of test and audiometer. The American National Standards Institute (ANSI) specifies a number of categories
of audiometers based on the range of frequencies, sound-pressure levels, and testing capabilities available.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 3/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 3
In a pure-tone audiometer, single-frequency tones are produced by an audio oscillator. An attenuator control is
used to adjust the intensity changes within a specified range, typically from 0 to 90 dB HL. Signals can be
delivered in a continuous or pulsed (interrupted) mode; pulsed tones appear to be more useful when testing
children or patients with tinnitus (ringing in the ear). Microprocessor-based units give the examiner the option of
delivering the signal manually or automatically, using a programmed presentation sequence. The examiner
directs the signal to earphones or to a bone-conduction vibrator. The patient may use hand signals or a
handswitch to respond to a tone that has been heard. Both air- and bone-conduction pure-tone testing must be
performed to determine whether hearing loss is conductive, sensorineural, or both.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 4/116
Audiometers; Audiometric Booths
4 ©2007 ECRI Institute. All Rights Reserved.
Figure 3. Single-room configuration, double-wall constructionaudiometric booth.
Figure 4. Double-room configuration (control and exam),single-wall construction audiometric booth.
Pure tones are directed to earphones to test air-
conduction hearing, which relies on the action of the
outer, middle, and inner ear and neural pathways. The
electrical signal from the audiometer is transformed
into vibrations within an earphone’s diaphragm that
generate a sound of a frequency and amplitude (dB)proportional to the electrical signal. The examiner can
test the left ear, right ear, or both. The threshold search
protocol is performed for a fixed series of frequencies
in each ear. Results are recorded in a graphic and/or
tabular format called an audiogram (see Figure 1).
Bone-conduction hearing relies solely on the action
of the inner ear and neural pathways. Pure tones are
directed to a bone-conduction vibrator placed on the
forehead or the mastoid process behind the ear. The
electrical signal from the audiometer creates
vibrations within a plastic shell in the vibrator. As the
vibrations are transferred to the skull, the inner ear is
stimulated directly as a result of bone distortion and
indirectly as a result of inputs received from outer-
and middle-ear vibrations. Some instruments record
air- and bone-conduction data on the same graph or
table, while others show the data on separate audiograms.
The procedure for bone-conduction testing is basically the same as that for air-conduction testing. Ideally, both
should be performed in an audiometric booth; however, when this kind of room is unavailable, testing should be
performed in quiet surroundings with minimal background noise. Some instruments feature a two-way
microphone built into the earphone headset to permit communication between the patient and the examiner
when the patient is isolated in an audiometric booth.
In conductive hearing loss, air-conduction thresholds
are higher (indicating more hearing loss) than bone-
conduction thresholds. In sensorineural hearing loss,
both air- and bone-conduction thresholds are equal and
elevated (hearing is below normal). When hearing loss
is “mixed” (both conductive and sensorineural), air- and
bone-conduction thresholds are again elevated;
however, air-conduction thresholds are higher than
bone-conduction thresholds. Bone-conduction testing is
sometimes used in lieu of air-conduction testing for
young children, when reliable pure-tone thresholds
cannot be obtained.Speech audiometry measures the sound levels at
which speech is detected and individual words are
distinguished from one another. Spoken material is
introduced as an input signal to the audiometer from a
microphone, phonograph, tape recorder, or compact
disc (CD). The signal is intensified in an amplifier and
monitored by a volume-units meter. Intensity range is
calibrated in dB HL and is controlled with an attenuator
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 5/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 5
as in pure-tone audiometry. The examiner directs the signal to earphones or through a second amplifier to
loudspeakers. An ear-selector control may also be used to direct the signal to the left or right earphone only.
Most testing is done through air-conduction
earphones. Sensitivity for speech is determined using a
speech-reception threshold (SRT) test. Threshold search
protocols are similar to those used for pure-toneaudiometry. The second component of speech
audiometry is the speech-discrimination test (SDT),
which assesses the patient’s ability to understand
monosyllabic words at suprathreshold (above
threshold) intensities. Patients with severe inner ear
disease cannot distinguish between similar words, even
at these high sound levels.
When earphones are not practical (e.g., when young
children are being tested), free-field testing is used. The
patient is placed equidistant between two loudspeakers
in a sound-isolation room. Warble tones (tones with
specific modulation rates), narrow-band noise (tones of
several frequencies of equal intensity), or spoken material is used as the auditory signal.
The shape of audiometric curves can be very useful in diagnosing certain types of hearing loss when other
methods of clinical and laboratory analysis are unsuccessful. For example, patients with hereditary dominant
mid- and low-frequency sensorineural loss show no evidence of pathology aside from their characteristic
audiograms (see Figure 2). Audiometric testing of siblings provides confirmation of this diagnosis.
When the nontest ear is suspected of responding to the test signal, it must be masked. One of three types of
masking is applied to the earphone corresponding to the nontest ear: white noise, pink noise, or narrow-band
noise. White noise consists of a wide range of frequencies with uniform loudness and is used to mask tones or
speech. Pink noise is similar to white noise except that it consists of a higher proportion of lower frequencies and
is used to mask speech alone. Narrow-band noise consists only of frequencies close to the ones being tested and is
used to mask pure tones. Proper application of masking is complex, especially when the hearing loss is mixed or
if only one ear has a severe sensorineural loss. Masking intensity is usually quantified in decibels sound-pressure
level (dB SPL).
Audiometric booths
The walls of audiometric booths consist of steel panels on the outside, sound-absorbing perforated steel panels
on the inside, and an incombustible acoustic insulating material (e.g., fiberglass) in between.
Single-wall booths are constructed from a single wall and a layer of sound-absorbing material and are used
primarily for pure-tone air-conduction testing. Double-wall booths employ “booth inside a booth” construction
in which the walls of the inner examination booth are separated from the outer panels by an air space; this design
further reduces outside noise interference and is preferred for bone-conduction, sound-field, and speech testing.Single- or double-wall construction can be used in both single- and double-room configurations.
Single-room booths (see Figure 3) isolate the subject from the environment, but not the examiner. Most
audiometry can be performed in a single-room booth; the exception is live-voice speech testing. Double-room
booths (see Figure 4) consist of two connecting rooms (a control room and an exam room) that isolate the
examiner and subject from the environment and also from each other. Either a single or double wall (depending
on the booth construction) separates the two rooms and incorporates a window consisting of several layers of
glass. Windows should provide visual contact between examiner and subject but obscure the subject’s view of
any distracting activities in the control room. Windows are also present in the walls of a single-room booth and
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 6/116
Audiometers; Audiometric Booths
6 ©2007 ECRI Institute. All Rights Reserved.
may be present in the external walls of a double-room booth to allow the inside of the booth to be viewed from
the outside.
Single-wall booths have one door, which opens outward; double-wall booths have a second door, which opens
inward and is mounted on the same frame. Some double-room configurations offer an optional single or double
door between the control and test rooms. Vibration isolators underneath the booth minimize floor vibrations and
sound transmission. A specially designed forced-air system can minimize ventilation noise.Portable booths are single-wall structures and move on locking casters. They may also come with vibration
isolators and ventilation systems. Larger, nonportable booths come in kits that include prewired wall, floor, and
roof sections; wall and corner joiners; acoustic filling material; and a ventilation system—many portable booths
come completely assembled. Some smaller portable booths are designed to be disassembled for transport from
one testing location to another.
Reported problems
Reliable audiometric measurement depends on equipment calibration, the test environment, patient
cooperation, and examiner performance. All audiometric equipment must be calibrated according to international
standards. However, calibration and test results are invalid if the test is not performed in the proper environment.
Background noise levels must not exceed those prescribed by organizations such as ANSI and OSHA, and visualdistractions, ventilation, and temperature must be controlled. Finally, the room should be set up so that the
patient cannot see what the examiner is doing with the audiometer while the test is being performed.
ECRI Institute has received no reports of safety problems unique to audiometric booths. Single-wall booths can
be constructed of optional fireproof panels and doors; the inner panels and doors of double-wall booths can also
be made of fireproof material.
Purchase considerations
ECRI Institute recommendations
The accompanying comparison charts include ECRI Institute’s recommendations for minimum performancerequirements for audiometers. The recommended specifications have been divided into two categories:
audiometers used for screening purposes and audiometers used in clinical environments. Each of these categories
is further broken down into two subcategories—one for minimum requirements based on ANSI S3.6-1996, which
will provide the user with a basic audiometer, and the other for recommendations intended for more advanced
functions. These recommendations will be most appropriate for users who commonly perform comprehensive
studies. (See ANSI type designations.)
Screening audiometers should be type 4 as described in ANSI 3.6-1996. Audiometers with these capabilities
should provide the user with all necessary functions to perform under most screening conditions. Selecting a unit
with greater capability may result in more difficult operation and/or higher costs. Ease of operation is important,
particularly in screening programs in which operators may have minimal training and experience. Screening
programs generally do not use speech audiometry, and OSHA does not require it for annual audiograms;
therefore, this feature is not recommended for audiometers used for screening purposes. For screening, pure-tone
audiometers, at minimum, should be capable of testing at frequencies of 500 to 6,000 Hz.
Audiometers used for clinical testing and diagnosis should be at least type 3 as specified by ANSI S3.6-1996.
Type 1 audiometers may be required under certain conditions and are recommended for clinical environments in
which a full range of testing applications must be met. Clinical audiometers may also require speech-testing
capabilities. For this purpose, the audiometer should be designated at least type C. Standardized, specially
prepared speech material from tape or CD input is preferred to microphone input. This ensures consistent testing
and results among subjects.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 7/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 7
Standard clinical audiometry testing is performed over a range of 250 to 8,000 Hz. OSHA does not require
bone-conduction testing for annual or baseline audiograms; however, it can be useful in determining diagnosis
(e.g., discovering where a hearing problem may lie) in a clinical setting. For most testing, frequency accuracy
should be within 3%; accuracy within 1% is recommended for some clinical applications. It is also recommended
that half-octave or octave steps be available for both clinical and screening applications. Testing at half-octave
frequencies is often used to verify that the subject’s hearing level does not change significantly within an octave.
Masking noise should be available for clinical testing to ensure that one ear is not compensating for the other
during the test.
Certain considerations for earphones
must be made for any type of audiometric
testing. Those used to deliver the signal to
the subject should include one earphone for
each ear. Although supra-aural earphones
are commonly used, availability of insert
earphones is advantageous, since these
earphones are generally lighter and provide
a better fit than supra-aural earphones. They
also offer greater protection against the
signal being heard by the opposite ear. Free-
field testing is not normally recommended
for screening purposes. However, they are a
good alternative to earphones when testing
small children who may not tolerate
earphones. Free-field testing is also used in
certain hearing-aid assessments.
Earphones should be in proper condition: pads should be soft, pliable, and free of cracks. Cords should also be
checked to make sure that the left and right earphones are connected to the correct earjacks. Earphones should be
calibrated for use with the audiometer to produce the required sound-pressure levels.
Audiometric booth requirements depend on individual circumstances. In some cases, testing may be
performed without a booth; however, when the test area is too noisy, a booth is required. Industry standards for
noise levels should be considered. The amount of sound reduction required will depend on the noise level in the
area, taking into account the noisiest periods of time. Other booth requirements will also depend on individual
circumstances. Size requirements depend on the available space, but the booth should be able to comfortably
accommodate the subject, the examiner, and all necessary equipment. In audiometric booths with two rooms,
there must be a window between the rooms.
Booths with fluorescent lighting must have the light’s starter mounted on the outside of the examination room
because starters produce a hum that may interfere with testing. In addition, windows must have moisture-
absorbing materials between the panes to prevent fogging. If the booth has a step, a ramp must be provided for
subjects in wheelchairs or on stretchers. Because doors must be solid and close with an acoustic seal, they are
often very heavy.
Other considerations
Some audiometers feature software that assesses the hearing level by comparing it to age-related values from
the general population. Software is available that allows the user to design a format that can accommodate a large
amount of patient (and company) data or to customize test protocols to fit the individual patient’s needs and
capabilities. Some audiometers provide a graphic output of results for threshold testing, and several units can
store a limited number of audiograms in nonvolatile memory.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 8/116
Audiometers; Audiometric Booths
8 ©2007 ECRI Institute. All Rights Reserved.
Multichannel audiometers are available that can examine several people at once. Each channel of a group
audiometer is programmed to administer a complete series of tests to each individual; all channels are linked to a
central control channel through which the examiner can set the test parameters and initiate testing. Special
software programs compare the individual’s hearing threshold to a normal range of values (adjusted for age and
exposure to noise)—for example, from an employee’s occupational group. (Many companies require their
employees to undergo regular audiometric testing as part of a hearing-conservation program. Multichannelaudiometers enable companies to perform time- and cost-effective on-site testing of the workforce.)
Various suppliers include optional parts such as external power amplifiers, computer cables, microphones,
software, memory cards, and service manuals. Check to see if the warranty includes a hotline service for any
technical problems and/or user difficulty.
Stage of development
Microprocessor-based units have automated most test procedures and hearing-threshold calculations. Certain
computerized units have validity checks within their operational programs that repeat the signal for a given
frequency and compare the two responses. Personal-computer-based audiometers that can be used with
notebook, desktop, or networked systems are also available. These systems may have features such as graphical
displays of results, warnings, and referral levels; use of disk drives for storage; and connection to a printer toobtain additional records.
Tympanometers, middle-ear analyzers, and auditory-evoked response units have been developed that test
certain parts of the ear and auditory nerve to gather more specific information about the nature of a hearing
problem. More sensitive and specialized tests have been created for these instruments to enhance diagnosis and
monitor the results of rehabilitative therapy.
Audiometric booths are a widely accepted component of the instrumentation used in clinical diagnosis of
hearing disorders. There have been no recent changes in the technology of these rooms, and none are foreseeable
in the near future.
Bibliography
Burgess RC, ed. Technology and equipment review: audiometers. J Clin Neurophysiol 1994 Sep;11(5):525-8.
Caple GE. Effective hearing programs require assessment, testing, enforcement. Occup Health Saf 1989
Jan;58(1):48-9, 52.
Durrant JD, Lovrinic JH. Bases of hearing sciences. 3rd ed. Baltimore: Williams & Wilkins; 1994.
Erdreich J. Threshold shift in hearing level stems from non-occupational cause. Occup Health Saf 1989 Jan;58(1):44-
7, 60.
Frank T. ANSI update: specification of audiometers. Am J Audiol 1997 Nov;6(3):29-32.
Kratz IC. Using equipment in unfamiliar clinical settings: audiology screening. J Pediatr Nurs 1997 Oct;12(5):307-
10.
Martin FN. Introduction to audiology. 3rd ed. Englewood Cliffs (NJ): Prentice-Hall; 1986.
Rintelmann W, ed. Hearing assessment. Baltimore: University Park Press; 1979.
Sawyer C. Audiometric monitoring may reduce hearing loss, compensation awards. Occup Health Saf 1989
Jan;58(1):50-2.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 9/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 9
Supplier information
Chart A: Audiometers
AMBCO
Ambco Electronics [102475]
15052 Redhill Ave Suite DTustin, CA 92680-6525
Phone: (714) 259-7930, (800) 345-1079 Fax: (714) 259-1688
Internet: http://www.ambco.com
E-mail: [email protected]
AMPLIFON
Amplaid Biomedical [408757]
via Gaetano Donizetti 12
Assago MI I-20090
Italy
Phone: 39 (02) 457901 Fax: 39 (02) 48822102
Internet: http://www.amplimedical.com
E-mail: [email protected]
AMPLIVOX
Amplivox Ltd [178837]
29-30 Station Approach
Kidlington OX5 1JD
England
Phone: 44 (1865) 842411 Fax: 44 (1865) 841853
Internet: http://www.amplivox.ltd.uk
E-mail: [email protected]
CASTLE
Castle Group Ltd [324542]
Salter Road Cayton Low Road Industrial Estate
Scarborough YO11 3UZ
England
Phone: 44 (1723) 584250 Fax: 44 (1723) 583728
Internet: http://www.castlegroup.co.uk
E-mail: [email protected]
DANPLEX
GN Otometrics A/S [391083]
2 Dybendalsvaenget Postboks 119
Taastrup DK-2630
Denmark
Phone: 45 72111555 Fax: 45 72111548Internet: http://www.gnotometrics.com
E-mail: [email protected]
GN Otometrics North America Inc [426355]
125 Commerce Dr
Schaumburg, IL 60173-5329
Phone: (847) 534-2150, (800) 289-2150 Fax: (847) 534-2151
Internet: http://www.gnotometrics.com
E-mail: [email protected]
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 10/116
Audiometers; Audiometric Booths
10 ©2007 ECRI Institute. All Rights Reserved.
DECOS
Decos Medical Systems bv [391295]
Bonnikeplein 24-26
Noordwijk NL-2201 XA
The Netherlands
Phone: 31 (71) 3640712 Fax: 31 (71) 3620833
Internet: http://www.decosmedical.nl
E-mail: [email protected]
ELECTRONIC PATZIG
Electronic Patzig Co for Electronic mbH [434007]
Niedersedlitzer Platz 6b
Dresden D-01259
Germany
Phone: 49 (351) 448640 Fax: 49 (351) 4486420
Internet: http://www.electronicpatzig.de
E-mail: [email protected]
ENTOMED
Entomed AB [157537]
Bariumgatan 29
Malmo S-213 64
Sweden
Phone: 46 (40) 210090 Fax: 46 (40) 210099
Internet: http://www.entomed.se
E-mail: [email protected]
EYMASA
EYMASA [354740]
Maresme 115-117
Barcelona E-08019
SpainPhone: 34 (933) 034356 Fax: 34 (933) 661848
Internet: http://www.eymasa.com
E-mail: [email protected]
FRYE
Frye Electronics Inc [101276]
9826 SW Tigard St PO Box 23391
Tigard, OR 97281-3391
Phone: (503) 620-2722, (800) 547-8209 Fax: (503) 639-0128
Internet: http://www.frye.com
E-mail: [email protected]
GM INSTRUMENTS
GM Instruments Ltd [177196]
Unit 6 Ashgrove Workshops Ashgrove Road
Kilwinning KA13 6PU
Scotland
Phone: 44 (1294) 554664 Fax: 44 (1294) 551154
Internet: http://www.gm_instruments.com
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 11/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 11
GRASON STADLER
Grason-Stadler [102263]
5225 Verona Rd Bldg 2
Madison, WI 53711-4495
Phone: (608) 441-2323, (800) 700-2282 Fax: (608) 441-2234
Internet: http://www.grason-stadler.com
INTERACOUSTICS
Interacoustics A/S [150984]
Drejervaenget 8
Assens DK-5610
Denmark
Phone: 45 63713555 Fax: 45 63713522
Internet: http://www.interacoustics.com
E-mail: [email protected]
Interacoustics USA [361066]
7625 Golden Triangle Dr Unit F
Eden Prairie, MN 55344Phone: (952) 941-4200, (800) 947-6334 Fax: (952) 903-4100
Internet: http://www.interacoustics.com
E-mail: [email protected]
MADSEN
Madsen Electronics (Denmark) [139621]
Markaervej 2a Postboks 119
Taastrup DK-2630
Denmark
Phone: 45 72111555 Fax: 45 72111348
Internet: http://www.madsen.com
E-mail: [email protected]
MAICO
Maico Diagnostic GmbH [236957]
Salzufer 13/14
Berlin D-10587
Germany
Phone: 49 (30) 70714650 Fax: 49 (30) 70714699
Internet: http://www.maico-diagnostic.com
E-mail: [email protected]
Maico Diagnostics [348943]
7625 Golden Triangle Dr
Eden Prairie, MN 55344
Phone: (952) 941-4200, (888) 941-4201 Fax: (952) 903-4200, (888) 941-4200
Internet: http://www.maico-diagnostics.com
E-mail: [email protected]
MEDIROLL
Mediroll Medico-technical Ltd [287831]
Segner ter 1
Debrecen H-4025
Hungary
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 12/116
Audiometers; Audiometric Booths
12 ©2007 ECRI Institute. All Rights Reserved.
Phone: 36 (52) 533737 Fax: 36 (52) 534446
Internet: http://www.mediroll.hu
E-mail: [email protected]
MICRO AUDIOMETRICS
Micro Audiometrics Corp [108789]
655 Keller RdMurphy, NC 28906-5889
Phone: (828) 644-0771, (866) 327-7226 Fax: (828) 644-0772
Internet: http://www.microaud.com
E-mail: [email protected]
MONITOR
Monitor Instruments Inc [106938]
437 Dimmocks Mill Rd
Hillsborough, NC 27278-2300
Phone: (919) 732-5400, (800) 853-6785 Fax: (919) 732-6153
Internet: http://www.monitorinstrumentsinc.com
E-mail: [email protected]
NAGASHIMA
Nagashima Medical Instruments Co Ltd [155887]
24-1 Hongo 5-chome Bunkyo-ku
Tokyo 113
Japan
Phone: 81 (3) 38121271 Fax: 81 (3) 38162824
Internet: http://www.nagashima-medical.co.jp
E-mail: [email protected]
QUALITONE
Qualitone [103650]
6415 Flying Cloud Dr
Eden Prairie, MN 55344-3305
Phone: (952) 917-7161, (800) 328-3897 Fax: (952) 828-9262
Internet: http://www.qualitonehearing.com
SIBELMED
Sibel SA [177708]
Rossello 500
Barcelona E-08026
Spain
Phone: 34 (93) 4360007 Fax: 34 (93) 4361611
Internet: http://www.sibelmed.com
E-mail: [email protected]
SINGER
Singer Medical Products Inc [106865]
790 Maple Ln
Bensenville, IL 60106-1513
Phone: (630) 860-6500 Fax: (630) 860-3672
E-mail: [email protected]
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 13/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 13
STARKEY
Starkey Australia Pty Ltd [418605]
74-96 Pyrmont Bridge Road
Camperdown 2050
Australia
Phone: 61 (2) 95502077
Internet: http://www.starkey.com.au
Starkey Colombia [418609]
Calle 98 No 22-64
Santa Fe de Bogota
Colombia
Phone: 57 (1) 6360960
Internet: http://www.starkey.com
Starkey France Sarl [418610]
23 rue Claude Nicolas le Doux Europarc Batiment 14
Creteil F-94000
FrancePhone: 33 (1) 49807474
Internet: http://www.starkey.com
Starkey Laboratories Inc [107526]
6700 Washington Ave S
Eden Prairie, MN 55344
Phone: (952) 941-6401, (800) 328-8602 Fax: (952) 828-9262
Internet: http://www.starkey.com
Welch Allyn
Welch Allyn GmbH & Co KG [178366]
Zollerstrasse 2-4
Jungingen D-72417Germany
Phone: 49 (7477) 92710 Fax: 49 (7477) 927190
Internet: http://www.welchallyn.de
E-mail: [email protected]
Welch Allyn Inc [101850]
4341 State Street Rd PO Box 220
Skaneateles Falls, NY 13153-0220
Phone: (315) 685-4100, (800) 535-6663 Fax: (315) 685-4091
Internet: http://www.welchallyn.com
E-mail: [email protected]
Chart B: Audiometric Booths
ACOUSTAIR
Acoustair [321007]
Postbus 1
Moerkapelle NL-2750 AA
The Netherlands
Phone: 31 (79) 5931341 Fax: 31 (79) 5931317
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 14/116
Audiometers; Audiometric Booths
14 ©2007 ECRI Institute. All Rights Reserved.
Internet: http://www.acoustair.com
E-mail: [email protected]
ACOUSTIC
Acoustic Systems [104980]
415 E St Elmo Rd
Austin, TX 78745Phone: (512) 444-1961, (800) 749-1460 Fax: (512) 444-2282
Internet: http://www.acousticsystems.com
E-mail: [email protected]
AMPLIFON
Amplaid Biomedical [408757]
via Gaetano Donizetti 12
Assago MI I-20090
Italy
Phone: 39 (02) 457901 Fax: 39 (02) 48822102
Internet: http://www.amplimedical.com
E-mail: [email protected]
ECKEL
Eckel Industries Inc [104016]
155 Fawcett St
Cambridge, MA 02138
Phone: (617) 491-3221 Fax: (617) 547-2171
Internet: http://www.eckelacoustic.com
E-mail: [email protected]
EYMASA
EYMASA [354740]
Maresme 115-117
Barcelona E-08019
Spain
Phone: 34 (933) 034356 Fax: 34 (933) 661848
Internet: http://www.eymasa.com
E-mail: [email protected]
IAC
Industrial Acoustics Co [102440]
1160 Commerce Ave
Bronx, NY 10462
Phone: (718) 430-4523 Fax: (718) 863-1138
Internet: http://www.industrialacoustics.com
E-mail: [email protected]
Industrial Acoustics Co GmbH [235118]
Solweg 17
Niederkruechten D-41372
Germany
Phone: 49 (2163) 99910 Fax: 49 (2163) 999123
Internet: http://www.iac-gmbh.de
E-mail: [email protected]
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 15/116
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 16/116
Audiometers; Audiometric Booths
16 ©2007 ECRI Institute. All Rights Reserved.
Policy Statement
The Healthcare Product Comparison System (HPCS) is published by ECRI Institute, a nonprofit organization.
HPCS provides comprehensive information to help healthcare professionals select and purchase diagnostic and
therapeutic capital equipment more effectively in support of improved patient care.
The information in Product Comparisons comes from a number of sources: medical and biomedical
engineering literature, correspondence and discussion with manufacturers and distributors, specifications from
product literature, and ECRI Institute’s Problem Reporting System. While these data are reviewed by qualified
health professionals, they have not been tested by ECRI Institute’s clinical and engineering personnel and are
largely unconfirmed. The Healthcare Product Comparison System and ECRI Institute are not responsible for the
quality or validity of information derived from outside sources or for any adverse consequences of acting on such
information.
The appearance or listing of any item, or the use of a photograph thereof, in the Healthcare Product Comparison
System does not constitute the endorsement or approval of the product’s quality, performance, or value, or of
claims made for it by the manufacturer. The information and photographs published in Product Comparisons
appear at no charge to manufacturers.
Many of the words or model descriptions appearing in the Healthcare Product Comparison System are
proprietary names (e.g., trademarks), even though no reference to this fact may be made. The appearance of any
name without designation as proprietary should not be regarded as a representation that is not the subject of
proprietary rights.
ECRI Institute respects and is impartial to all ethical medical device companies and practices. The Healthcare
Product Comparison System accepts no advertising and has no obligations to any commercial interests. ECRI
Institute and its employees accept no royalties, gifts, finder’s fees, or commissions from the medical device
industry, nor do they own stock in medical device companies. Employees engage in no private consulting work
for the medical device industry.
About ECRI Institute
ECRI Institute, a nonprofit organization, dedicates itself to bringing the discipline of applied scientific research
in healthcare to uncover the best approaches to improving patient care. As pioneers in this science for nearly 40
years, ECRI Institute marries experience and independence with the objectivity of evidence-based research.
More than 5,000 healthcare organizations worldwide rely on ECRI Institute’s expertise in patient safety
improvement, risk and quality management, healthcare processes, devices, procedures, and drug technology.
ECRI Institute is one of only a handful of organizations designated as both a Collaborating Center of the World
Health Organization and an Evidence-based Practice Center by the U.S. Agency for Healthcare Research and
Quality. For more information, visit http://www.ecri.org.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 17/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 17
Product Comparison Charts
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 18/116
Audiometers; Audiometric Booths
18 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
Clinical Audiometers,
Recommended
Clinical Audiometers,
Required
Screening Audiometers,
Recommended
Screening Audiometers,
RequiredWHERE MARKETED
FDA CLEARANCE
CE MARK (MDD)
TYPE Type 1 pure tone Type 3 pure tone Type 4 pure tone Type 4 pure toneSpeech Type C
CONFIGURATION
PURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125, 250, 500, 750, 1,000,
1,500, 2,000, 3,000, 4,000,6,000, 8,000
250, 500, 1,000, 2,000,3,000, 4,000, 6,000, 8,000
500, 1,000, 2,000, 3,000,4,000, 6,000
500, 1,000, 2,000, 3,000,4,000, 6,000
Bone conduction 250, 500, 750, 1,000, 1,500,2,000, 3,000, 4,000, 6,000
250, 500, 1,000, 2,000,3,000, 4,000
Continuous or step Step StepFrequency accuracy 1% 3% 3% 3%
Intensity, dB HLAir conduction
250 Hz 90 70 90 701,000 Hz 120 @ 500-4,000 Hz 100 @ 500-4,000 Hz8,000 Hz 100 80Graduations 5 5 5 5
Bone conduction250 Hz 70 351,000 Hz 70 @ 1,000-3,000 Hz 60 @ 1,000-3,000 Hz4,000 Hz 60 50Graduations 5 5
Masking, noise type Narrow band, white Narrow bandIntensity, dB SPL ≤115 ≤115
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 19/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 19
Chart A: Audiometers
MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
Clinical Audiometers,Recommended
Clinical Audiometers,Required
Screening Audiometers,Recommended
Screening Audiometers,Required
SPEECH AUDIOMETRYInputs Tape, CDIntensity, dB HLMasking, noise type Speech spectrum noise
Intensity, dB SPL ≤115 ≤115 EARPHONE TYPE TDH-39, TDH-49, ER3-A
Cushion type MX-41-AR, insertearphones
BONE VIBRATOR YesType
SPECIAL TESTS Hearing-aid testingcapabilities when thesetests are likely to beperformed
POWER REQUIREMENTSLine, VACBattery
Type, Volts
W x D x H, cm (in)
WEIGHT, kg (lb)
PURCHASEINFORMATION
List priceWarrantyDelivery time, AROFiscal year
This is the second of three
pages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 20/116
Audiometers; Audiometric Booths
20 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
Clinical Audiometers,Recommended
Clinical Audiometers,Required
Screening Audiometers,Recommended
Screening Audiometers,Required
OTHER SPECIFICATIONS Other recommendedfeatures: audible monitoringof test signal; operator-to-subject speechcommunication; signalindicator/VU meter; subjectresponse system.
Other recommendedfeatures: subject responsesystem.
Other recommendedfeatures: subject responsesystem.
LAST UPDATED
Supplier Footnotes 1These recommendationsare the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on
this data.
1These specifications arethe opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on
this data.
1These recommendationsare the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on
this data.
1These recommendationsare the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on
this data.Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 21/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 21
Chart A: Audiometers
MODEL AMBCO AMBCO AMBCO AMPLIFON1000 + Otoscreen 2500 Automatic Storage 650 A Amplaid A137
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified YesCE MARK (MDD) Not specified Not specified Not specified YesTYPE Wide range Wide range Limited range Limited range
Speech No No No NoCONFIGURATION Portable Tabletop/portable Portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 250-8,000 (octave,
interoctave bands)250-8,000 (octave,interoctave bands)
250-8,000 (octave,interoctave bands)
125-8,000
Bone conduction NA NA NA 250-8,000Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified >1%
Intensity, dB HLAir conduction
250 Hz 0 to +90 0 to +95 0 to +90 -10 to +100
1,000 Hz 0 to +90 0 to +95 0 to +90 -10 to +1208,000 Hz 0 to +90 0 to +95 0 to +90 -10 to +100Graduations 5 5 5 5
Bone conduction250 Hz NA NA NA -10 to +501,000 Hz NA NA NA -10 to +804,000 Hz NA NA NA -10 to +80Graduations NA NA NA 5
Masking, noise type NA NA NA Narrow band, whiteIntensity, dB SPL NA NA NA -10 to +95 dB HL
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 22/116
Audiometers; Audiometric Booths
22 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL AMBCO AMBCO AMBCO AMPLIFON1000 + Otoscreen 2500 Automatic Storage 650 A Amplaid A137
SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA
Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-49, EARtone 3A
Cushion type Supra-aural Supra-aural Supra-aural MX-41BONE VIBRATOR
Type NA NA NA B-71SPECIAL TESTS Sequential screening,
accurate, fast screening,digital, printouts
None specified None specified AC-BC steady and pulsedpure tones, warble,automatic threshold,synchronous masking, LCD,noise enclosures
POWER REQUIREMENTSLine, VAC 120/220 120/220 105-125/220 110/240, 50/60 HzBattery None None AB None
Type, Volts NA NA 9 NAW x D x H, cm (in) 24.1 x 29.2 x 16.5 (9.5 x
11.5 x 6.5)24.1 x 29.2 x 16.5 (9.5 x11.5 x 6.5)
22.9 x 33 x 15.2 (9 x 13 x 6) 36 x 28 x 12 (14 x 11 x 5)
WEIGHT, kg (lb) 3.2 (7) 3.2 (7) 1.8 (4) 1.9 (4.3)PURCHASEINFORMATION
List price $795 $1,975 $735 €1,700 (US $2,150) Warranty 5 years 5 years 5 years 1 yearDelivery time, ARO 1 week 1 week 1 week 30 daysFiscal year May to April May to April May to April January to December
This is the second of threepages covering the abovemodel(s). These
specifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 23/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 23
Chart A: Audiometers
MODEL AMBCO AMBCO AMBCO AMPLIFON1000 + Otoscreen 2500 Automatic Storage 650 A Amplaid A137
OTHER SPECIFICATIONS Digital LED display; front-panel membrane switches;printer available. Meetsrequirements of ANSI.
Microprocessor-controlled;stores 540 tests; fullyautomatic, semiautomatic,or manual testing; computercompatible. Meetsrequirements of ANSI;OSHA compliant.
Carrying case. Meetsrequirements of ANSI.
Built-in talk-forward mic;single insert phone formasking; noise enclosures;LCD, 4 rows x 20characters; external printerand carrying case; datastorage and transfer; built-inmemory for 128 patients,
AC-BC both ears; RS232port; transfer to Amplaid
ASA database (Windows);NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.
LAST UPDATED January 2004 January 2004 January 2004 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 24/116
Audiometers; Audiometric Booths
24 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON Amplaid A177 Plus Amplaid A321 Amplaid A321 HF Amplaid A724
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range, clinical, ANSI
1A-EWide range, clinical, ANSI1A-E
Wide range, clinical, ANSIS3.39
Speech Yes Yes Yes NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 125-8,000 125-20,000 500-4,000Bone conduction 250-8,000 250-8,000 250-8,000 NoContinuous or step Step Step Step StepFrequency accuracy >1% >1% >1% >1%
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +100 No
1,000 Hz -10 to +120 -10 to +120 -10 to +120 -50 to +1208,000 Hz -10 to +100 -10 to +100 -10 to +100 NoGraduations 1, 2, and 5 1, 2, and 5 1, 2, and 5 5
Bone conduction250 Hz -10 to +50 -10 to +50 -10 to +50 NA1,000 Hz -10 to +80 -10 to +80 -10 to +80 NA4,000 Hz -10 to +80 -10 to +80 -10 to +80 NAGraduations 1, 2, and 5 1, 2, and 5 1, 2, and 5 NA
Masking, noise type Narrow band, white Narrow band, white Narrow band, white NAIntensity, dB SPL -10 to +95 dB HL -10 to +95 dB HL -10 to +95 dB HL NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 25/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 25
Chart A: Audiometers
MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON Amplaid A177 Plus Amplaid A321 Amplaid A321 HF Amplaid A724
SPEECH AUDIOMETRYInputs Mic, tape, CD Mic, tape, CD (2 separate
EXT inputs)Mic, tape, CD (2 separateEXT inputs)
NA
Intensity, dB HL -10 to +105 -10 to +105 -10 to +105 NAMasking, noise type Speech, white Speech, white Speech, white NA
Intensity, dB SPL -10 to +95 dB HL -10 to +95 dB EM -10 to +95 dB EM Not specifiedEARPHONE TYPE TDH-49, EARtone 3A TDH-49, EARtone 3A TDH-49, EARtone 3A Not specified
Cushion type MX-41 MX-41 MX-41 Not specifiedBONE VIBRATOR
Type B-71 B-71 B-71 Not specifiedSPECIAL TESTS AC-BC-FF pure tones and
speech, warble, automaticthreshold, Stereo Speech,
ABLB, synchronousmasking, LCD, noiseenclosures
AC-BC-FF pure tones andspeech, warble, StereoSpeech, multifrequency,
ABLB (Fowler), MLB(Reger), DLI (Lusher), andSISI (Jerger), tone decay,automatic threshold, tone
and speech Stenger, MHA
AC-BC-FF pure tones andspeech, warble, StereoSpeech, multifrequency,
ABLB (Fowler), MLB(Reger), DLI (Lusher) andSISI (Jerger), tone decay,automatic threshold, tone
and speech Stenger, MHA,MLD (Masking leveldifference), high frequencyaudiometry (2 frequencyranges), Bekesy (sweepand fixed frequencies,continuous, pulsed and LOTstimuli, continuous and fixedmasking noise)
Contralateral audiometry,226 Hz tympanometry,acoustic reflex tests,sequential tests includingdecay
POWER REQUIREMENTSLine, VAC 100/240, 50/60 Hz 100/240, 50/60 Hz 100/240, 50/60 Hz 100/240, 50/60 HzBattery None None None None
Type, Volts NA NA NA NAW x D x H, cm (in) 36 x 28 x 12 (14 x 11 x 5) 36 x 28 x 18 (14 x 11 x 7) 36 x 28 x 18 (14 x 11 x 7) 49 x 40 x 20 (19.3 x 15.7 x
7.9)WEIGHT, kg (lb) 1.9 (4.3) 3 (7.7) 3 (7.7) 8.5 (18.8)
PURCHASEINFORMATIONList price €2,400 (US $3,036) €3,900 (US$4,933) €4,850 (US$6,135) €4,850 (US$6,135) Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 30 days 30 days 30 days 30 daysFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 26/116
Audiometers; Audiometric Booths
26 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON Amplaid A177 Plus Amplaid A321 Amplaid A321 HF Amplaid A724
OTHER SPECIFICATIONS Built-in talk-forward mic;goose-neck mic; patientmic; single insert phone formasking; free-field stereooutputs; noise enclosures;LCD, 4 rows x 20characters; external printerand carrying case; datastorage and transfer; built-inmemory for 128 patients,
AC-BC both ears; RS232port; transfer to Amplaid
ASA database (Windows);NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.
Twin channels; LCDdisplaying right and leftaudiograms; built-in printerand output to externalprinters, PLC5 languagecompatible; built-in talk-forward mic; goose-neckmic; patient mic; singleinsert phone for masking;free-field stereo outputs;noise enclosures; carryingcase; data storage andtransfer; built-in memory forall tests for one patient;RS232 and USB serialports; transfer to Amplaid
ASA database (Windows);
NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.
Twin channels; LCDdisplaying right and leftaudiograms; built-in printerand output to externalprinters, PLC5 languagecompatible; built-in talk-forward mic; goose-neckmic; patient mic; singleinsert phone for masking;free-field stereo outputs;noise enclosures; carryingcase; data storage andtransfer; built-in memory forall tests for one patient;RS232 and USB serialports; transfer to Amplaid
ASA database (Windows);
NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.
Diagnostic and highfrequency impedanceaudiometer; built-in thermalprinter; LCD; RS232 datatransfer to Amplaid ASA(Windows); also availableas A724 clinical impedanceaudiometer. Meetsrequirements of EMC andIEC 601-1 Class I Type B.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 27/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 27
Chart A: Audiometers
MODEL AMPLIFON AMPLIFON AMPLIVOX AMPLIVOX Amplaid A728 Amplaid A756 116 240
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Not specified Not specifiedCE MARK (MDD) Yes Yes Yes YesTYPE Clinical HF, Type 1 ANSI
S3.39 (1987)Screening, Type 3 ANSIS3.39 (1987)
Limited range Wide range
Speech No No No NoCONFIGURATION Tabletop Tabletop/portable Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-12,000 500-4,000 250-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands)
Bone conduction No No NA 250-8,000 (octave,interoctave bands)
Continuous or step Step Step Step StepFrequency accuracy >1% >1% Not specified Not specified
Intensity, dB HL
Air conduction250 Hz -10 to +85 No -10 to +80 -10 to +901,000 Hz -10 to +120 50 to 120 -10 to +100 -10 to +1208,000 Hz -10 to +120 No -10 to +80 -10 to +90Graduations 5 5 5 5
Bone conduction250 Hz NA NA NA -10 to +301,000 Hz NA NA NA -10 to +604,000 Hz NA NA NA -10 to +60Graduations NA NA NA 5
Masking, noise type Narrow band NA NA Narrow band, broadIntensity, dB SPL -10 to +95 dB HL NA NA -10 to +100
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 28/116
Audiometers; Audiometric Booths
28 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL AMPLIFON AMPLIFON AMPLIVOX AMPLIVOX Amplaid A728 Amplaid A756 116 240
SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA
Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-49 TDH-49 TDH-39 TDH-49
Cushion type MX-41 MX-41 Circumaural or supra-aural Circumaural or supra-auralBONE VIBRATOR
Type NA NA NA B-71SPECIAL TESTS Automatic and manual AC
threshold without masking,automatic and manualtympanometry (226, 678,800, and 1,000 Hz),eustachian-tube testing(normal and perforated TC),manual and automated
reflexometry, reflex growth,nonacoustic reflex,sequential tymp plus reflexprotocols
Contralateral audiometry,226 Hz tympanometry,acoustic reflex tests,sequential tests
None None
POWER REQUIREMENTSLine, VAC 110/240, 50/60 Hz 110/240, 50/60 Hz 100-110/220-240 100-110/220-240Battery None None Alkaline Alkaline
Type, Volts NA NA 6 6W x D x H, cm (in) 49 x 40 x 20 (19.3 x 15.7 x
7.9)37 x 29 x 18 (14 x 11 x 7) 25 x 17 x 7 (10 x 7 x 2.5) 25 x 17 x 7 (10 x 7 x 2.5)
WEIGHT, kg (lb) 8.5 (18.8) 3.5 (7.7) 0.73 (1.6) 0.85 (1.82)PURCHASEINFORMATION
List price €6,186 (US $7,825) €3,000 (US $3,795) $825 $1,912Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 30 days 30 days 4 weeks 4 weeks
Fiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 29/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 29
Chart A: Audiometers
MODEL AMPLIFON AMPLIFON AMPLIVOX AMPLIVOX Amplaid A728 Amplaid A756 116 240
OTHER SPECIFICATIONS Diagnostic and high-frequency impedanceaudiometer; built-in thermalprinter; LCD; RS232 datatransfer to Amplaid ASA(Windows); also availableas A724 clinical impedanceaudiometer; ample choice of
AR tests, includingnonacoustic. Meetsrequirements of EMC andIEC 601-1 Class I Type B.
Screening, Type 3 ANSIS3.39 (1987) withsequential tests and severalreflex tests comparing reflexgrowth.
Carrying case. Meetsrelevant BS and IECrequirements.
Carrying case. Meetsrelevant BS and IECrequirements.
LAST UPDATED July 2005 July 2005 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 30/116
Audiometers; Audiometric Booths
30 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL AMPLIVOX AMPLIVOX CASTLE DANPLEX260 CA850/3 GA1001 Excalibur DA65
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Limited range Pure tone Wide range
Speech Yes No No YesCONFIGURATION Tabletop/portable Tabletop/portable Portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)500-8,000 (octave,interoctave bands)
125-12,000 250-12,500 (12 frequencies)
Bone conduction 250-8,000 (octave,interoctave bands)
NA NA 250-8,000 (8 frequencies)
Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +90 NA -20 to +120 -10 to +1101,000 Hz -10 to +120 -10 to +90 -20 to +120 -10 to +1208,000 Hz -10 to +90 -10 to +90 -20 to +120 -10 to +100Graduations 5 2-5, 5 5 1, 2.5, 5 dB
Bone conduction250 Hz -10 to +30 NA NA -10 to +401,000 Hz -10 to +60 NA NA -10 to +804,000 Hz -10 to +60 NA NA -10 to +75Graduations 5 NA NA 1, 2.5, 5 dB
Masking, noise type Narrow band, broad NA NA Narrow band, whiteIntensity, dB SPL -10 to +100 NA NA -10 dB to maximum (for
pure tones, -20 dB)
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 31/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 31
Chart A: Audiometers
MODEL AMPLIVOX AMPLIVOX CASTLE DANPLEX260 CA850/3 GA1001 Excalibur DA65
SPEECH AUDIOMETRYInputs Mic, tape NA NA Tape, CD, micIntensity, dB HL -10 to +100 NA NA -10 to +80Masking, noise type Wide band NA NA Speech, white, ipsilateral,
contralateralIntensity, dB SPL -10 to +100 NA NA -10 to +80
EARPHONE TYPE TDH-49 TDH-39 Telephonics TDH-39Cushion type Circumaural or supra-aural Circumaural or supra-aural TDH-39 Circumaural
BONE VIBRATORType B-71 NA NA B-71
SPECIAL TESTS ABLB, Stenger, pulse andcontinuous tone,synchronous masking
Automatic Bekesy andHughson-Westlake graphics
None Stenger, Fowler (ABLB),SISI, master hearing aid
POWER REQUIREMENTSLine, VAC 100-110/220-240 100-110/220-240 220/110 110/220/240 ±10%, 50/60
HzBattery Alkaline None None None
Type, Volts 6 NA NA NAW x D x H, cm (in) 25 x 17 x 7 (10 x 7 x 2.5) 33.5 x 26.5 x 8.5 (13.2 x
10.4 x 3.3)27.5 x 28 x 7.5 (10.8 x 11 x3)
45 x 29 x 8.5 (17.7 x 11.4 x3.3)
WEIGHT, kg (lb) 0.85 (1.85) 2.9 (6.4) 1.6 (3.5) 4.5 (9.9)PURCHASEINFORMATION
List price $2,100 $3,675 £2,540 (US$3,750) $2,350Warranty 1 year 1 year 4 years 2 yearsDelivery time, ARO 4 weeks 4 weeks ~2 weeks Not specifiedFiscal year January to December January to December Not specified January to December
This is the second of threepages covering the abovemodel(s). These
specifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 32/116
Audiometers; Audiometric Booths
32 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL AMPLIVOX AMPLIVOX CASTLE DANPLEX260 CA850/3 GA1001 Excalibur DA65
OTHER SPECIFICATIONS Carrying case. Meetsrelevant BS and IECrequirements.
Internal and PC database.Meets relevant BS and IECrequirements.
Unit works with mostnotebook, desktop, ornetworked systems runningon Windows 3.1, 95, 98, or2000; unlimited storage onNT disk; statistical analysisand report letter generationfacility; printout of results in
A4 format. Meetsrequirements of BS 5966and 601-1.
Optional carrying case, highfrequency to 16,000 Hz.Meets requirements of IECPub. 645. Complies withIEC 601-1.
LAST UPDATED January 2007 January 2007 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 33/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 33
Chart A: Audiometers
MODEL DECOS ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG Audiology Workstation MA 12 MA 14 MA 15
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No Not specified Not specified Not specifiedCE MARK (MDD) Yes Not specified Not specified Not specifiedTYPE Wide range Narrow range Wide range Wide range
Speech Yes No No YesCONFIGURATION Patient interface, TDH36
stereo, TDH39 contra, B71bone conductor, patientresponse button
Portable Portable Portable
PURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-20,000 250-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands)
125-8,000 (octave,interoctave bands)
Bone conduction 250-8,000 NA 125-8,000 (octave,interoctave bands)
125-8,000 (octave,interoctave bands)
Continuous or step Step Step Step Step
Frequency accuracy 1 Hz Not specified Not specified Not specifiedIntensity, dB HL
Air conduction250 Hz -10 to +100 -10 to +90 -10 to +100 -10 to +1001,000 Hz -10 to +120 -10 to +90 -10 to +100 -10 to +1008,000 Hz -10 to +110 -10 to +90 -10 to +100 -10 to +100Graduations 1 5 5 5
Bone conduction250 Hz -10 to +45 NA -10 to +60 -10 to +601,000 Hz -10 to +75 NA -10 to +70 -10 to +704,000 Hz -10 to +75 NA -10 to +60 -10 to +60Graduations 1 NA 5 5
Masking, noise type Band, white, pink, speech,user-definable
NA Narrow band Narrow band
Intensity, dB SPL -10 to +110 NA -10 to +80 -10 to +80
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 34/116
Audiometers; Audiometric Booths
34 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL DECOS ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG Audiology Workstation MA 12 MA 14 MA 15
SPEECH AUDIOMETRYInputs CD, mic, hard disk NA NA Mic, tape, CDIntensity, dB HL -10 to +120 NA NA -10 to +110Masking, noise type White, pink, speech, band,
user-definableNA NA Speech
Intensity, dB SPL -10 to +110 NA NA -10 to +110EARPHONE TYPE TDH-39 8105 B2 8105 B2 8106
Cushion type Supra-aural Not specified Not specified Not specifiedBONE VIBRATOR
Type Radioear B71 NA KH 70 KH 70SPECIAL TESTS Tone audiometry, high
frequency, speechaudiometry, ABLB, tonedecay, Weber, speech innoise, directional hearing,visual reinforcementaudiometry, tinnitus, CAT,
tympanometer link
None SISI SISI
POWER REQUIREMENTSLine, VAC 100-110/220-240 230 230 230Battery None 3/R14 None None
Type, Volts NA 1.5 NA NAW x D x H, cm (in) 29 x 25 x 6 (11.4 x 9.8 x
2.4)30 x 22.5 x 5 (11.8 x 8.9 x2)
30 x 22.5 x 5 (11.8 x 8.9 x2)
30 x 22.5 x 5 (11.8 x 8.9 x2)
WEIGHT, kg (lb) 3 (6.6) 1.5 (3.3) 2 (4.4) 2 (4.4)PURCHASEINFORMATION
List price $8,900 $1,218 $1,867 $2,038Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 4 weeks Not specified Not specified Not specifiedFiscal year January to December Not specified Not specified Not specified
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 35/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 35
Chart A: Audiometers
MODEL DECOS ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG Audiology Workstation MA 12 MA 14 MA 15
OTHER SPECIFICATIONS PC-controlled audiometer;network base links to HL7,SQL server, and Oracle withopen database definition.Custom integration of everyelectronic patient file. Meetsrequirements of ISO 389,645, 4870, 6189, and 8253.
Carrying case included. Carrying case included; freefield via loudspeaker.
Carrying case included; freefield via loudspeaker.
LAST UPDATED January 2007 December 20021 December 20021 December 20021 Supplier Footnotes
Model Footnotes
Data Footnotes 1Model listed is currentlymarketed.
1Model listed is currentlymarketed.
1Model listed is currentlymarketed.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 36/116
Audiometers; Audiometric Booths
36 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIGMA 22-1 MA 22-2 MA 22-3 MA 32
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Wide range Wide range Wide range Wide range
Speech Yes No Yes YesCONFIGURATION Portable/tabletop Portable/tabletop Portable/tabletop TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements
125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements
125-8,000 (octave,interoctave bands)
Bone conduction 125-8,000 (octave,interoctave bands)
125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements
125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements
125-8,000 (octave,interoctave bands)
Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to 0 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +30 -10 to +130 -10 to +130 -10 to +1308,000 Hz -10 to +10 -10 to +110 -10 to +110 -10 to +110Graduations 5 5 5 5
Bone conduction250 Hz -10 to +60 -10 to +60 -10 to +60 -10 to +601,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +60 -10 to +60 -10 to +60 -10 to +60Graduations 5 5 5 5
Masking, noise type Wide/narrow band Wide/narrow band Wide/narrow band Narrow band, whiteIntensity, dB SPL -10 to +100, -10 to +110 -10 to +100, -10 to +110 -10 to +100, -10 to +110 -10 to +100, -10 to +110
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 37/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 37
Chart A: Audiometers
MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIGMA 22-1 MA 22-2 MA 22-3 MA 32
SPEECH AUDIOMETRYInputs Mic, tape, CD NA Mic, tape, CD Mic, tape, CDIntensity, dB HL -10 to +110 NA -10 to +110 -10 to +110Masking, noise type Speech NA Speech White, speech
Intensity, dB SPL -10 to +110 NA -10 to +110 -10 to +110EARPHONE TYPE 8106 HDA-200 8106, HDA-200 Not specified
Cushion type Not specified Not specified Not specified EARtone 3ABONE VIBRATOR
Type KH 70 KH 70 KH 70 KH 70SPECIAL TESTS SISI, Stenger, Fowler,
LombardNone specified SISI, Stenger, Fowler,
LombardSISI, Stenger, Luscher,Rainville, Fowler,Langenbeck, Carhart,Lombard, Kingsbury
POWER REQUIREMENTSLine, VAC 230 230 230 230Battery None None None None
Type, Volts NA NA NA NA
W x D x H, cm (in) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5)WEIGHT, kg (lb) 4 (8.8) 4 (8.8) 4 (8.8) 4 (8.8)PURCHASEINFORMATION
List price $2,773 $2,831 $3,130 $3,863Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 38/116
Audiometers; Audiometric Booths
38 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIGMA 22-1 MA 22-2 MA 22-3 MA 32
OTHER SPECIFICATIONS RS232 port for PCconnection; CENTRONICSfor printer connection;optional carrying case;monaural and dichoticspeech. Meets requirementsof IEC 645 and and ISO389.
RS232 port for PCconnection; CENTRONICSfor printer connection;optional carrying case.Meets requirements of IEC645 and ISO 389.
RS232 port for PCconnection; CENTRONICSfor printer connection;monaural and dichoticspeech; optional carryingcase. Meets requirements ofIEC 645 and ISO 389.
RS232 port for PCconnection; monaural,dichotic, ipsilateral talkoverspeech system. Meetsrequirements of IEC 645and ISO 389.
LAST UPDATED December 20021 December 20021 December 20021 December 20021 Supplier Footnotes
Model Footnotes
Data Footnotes 1Model listed is currentlymarketed.
1Model listed is currentlymarketed.
1Model listed is currentlymarketed.
1Model listed is currentlymarketed.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 39/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 39
Chart A: Audiometers
MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ENTOMED ENTOMEDMA 33 MA 33-1 GSI 67 GSI 68
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Not specified Not specified Yes YesTYPE Wide range Wide range Wide range Wide range
Speech Yes Yes No YesCONFIGURATION Tabletop Tabletop Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands)
125-8,000 125-8,000
Bone conduction 125-8,000 (octave,interoctave bands)
250-8,000 (octave,interoctave bands)
250-8,000 250-8,000
Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +130 -10 to +130 -10 to +110 -10 to +1108,000 Hz -10 to +110 -10 to +110 -10 to +100 -10 to +100Graduations 5 5 5 5
Bone conduction250 Hz -10 to +60 -10 to +60 -10 to +35 -10 to +351,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +60 -10 to +60 -10 to +80 -10 to +80Graduations 5 5 5 5
Masking, noise type Narrow band, white Narrow band, white Noise band Noise band, speechIntensity, dB SPL -10 to +100, -10 to +110 -10 to +110 0 to +95 0 to +95
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 40/116
Audiometers; Audiometric Booths
40 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ENTOMED ENTOMEDMA 33 MA 33-1 GSI 67 GSI 68
SPEECH AUDIOMETRYInputs Mic, tape, CD Mic, tape, CD NA CD, tape and micIntensity, dB HL -10 to +110 -10 to +110 NA -10 to +100Masking, noise type White, speech White, speech NA Noise band, speech
Intensity, dB SPL -10 to +110 -10 to +110 NA -10 to +100EARPHONE TYPE Not specified Not specified TDH-39, EARtone 5A TDH-39, EARtone 5A
Cushion type EARtone 3A EARtone 3A Supra-aural Supra-auralBONE VIBRATOR
Type KH 70 KH 70 B-71 B-71SPECIAL TESTS SISI, Stenger, Luscher,
Rainville, Fowler,Langenbeck, Carhart,Lombard, Kingsbury
SISI, Stenger, Luscher,Rainville, Fowler,Langenbeck, Carhart,Lombard, Kingsbury
Free-field option Free-field option
POWER REQUIREMENTSLine, VAC 230 230 115/230 115/230Battery None None None None
Type, Volts NA NA NA NA
W x D x H, cm (in) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5) 38.4 x 26.9 x 7.4 (15.1 x10.6 x 2.9)
48 x 26.9 x 7.4 (18.9 x 10.6x 2.9)
WEIGHT, kg (lb) 4 (8.8) 4 (8.8) 2 (4.4) 2.6 (5.7)PURCHASEINFORMATION
List price $4,439 $4,726 Not specified Not specifiedWarranty 1 year 1 year 2 year 2 yearDelivery time, ARO Not specified Not specified Stock, low quantities Stock, low quantitiesFiscal year Not specified Not specified January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue onto
the next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 41/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 41
Chart A: Audiometers
MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ENTOMED ENTOMEDMA 33 MA 33-1 GSI 67 GSI 68
OTHER SPECIFICATIONS RS232 port for PCconnection; matrix display;monaural, dichotic, oripsilateral talkover speechsystem. Meets requirementsof IEC 645 and ISO 389.
RS232 port for PCconnection; matrix display;monaural, dichotic, oripsilateral talkover speechsystem. Meets requirementsof IEC 645 and ISO 389.
Automatic Type 3audiometer; built-in real-time clock and talk-forwardmicrophone; pulsed tone;masking noise tracking thetone level; softwarecalibration; memory for 50audiograms; patient IDnumber entered by numerickeypad; RS232 interface toPC or printer; optionalcarrying case, optional free-field; patient responseswitch supplied; user-programmable setup menu.Meets requirements of ANSIand IEC.
Automatic Type 3audiometer; built-in real-time clock and talk-forwardmicrophone; pulsed tone;masking noise tracking thetone level; softwarecalibration; memory for 50audiograms; patient IDnumber entered by numerickeypad; RS232 interface toPC or printer; optionalcarrying case, optional free-field; patient responseswitch supplied; talk-backmicrophone supplied;monitor headset supplied;optional live speech monitor
headset; user-programmable setup menu;LED-VU meter in speechsection. Meets requirementsof ANSI and IEC.
LAST UPDATED December 20021 December 20021 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes 1Model listed is currentlymarketed.
1aModel listed is currentlymarketed.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 42/116
Audiometers; Audiometric Booths
42 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ENTOMED ENTOMED ENTOMED EYMASASA 201 SA 202/GSI 66 SA 50 Audios 310
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Wide range
Speech No No No YesCONFIGURATION Tabletop/portable Tabletop/portable Handheld TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 125-8,000 500-4,000 125-8,000 (octave bands)Bone conduction NA NA NA 250-6,000 (octave bands)Continuous or step Step Step Step Continuous, step, warbleFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 NA -10 to +1001,000 Hz -10 to +110 -10 to +110 +20 to +60 -10 to +120
8,000 Hz -10 to +100 -10 to +100 NA -10 to +90Graduations 5 5 10 0.5, 1, 5
Bone conduction250 Hz NA NA NA -10 to +451,000 Hz NA NA NA -10 to +704,000 Hz NA NA NA -10 to +70Graduations NA NA NA 0.5, 1, 5
Masking, noise type NA NA NA Narrow band, white, speechIntensity, dB SPL NA NA NA Not specified
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 43/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 43
Chart A: Audiometers
MODEL ENTOMED ENTOMED ENTOMED EYMASASA 201 SA 202/GSI 66 SA 50 Audios 310
SPEECH AUDIOMETRYInputs NA NA NA Mic, CDIntensity, dB HL NA NA NA -10 to +90Masking, noise type NA NA NA Narrow band, white, speech
Intensity, dB SPL NA NA NA -10 to +90EARPHONE TYPE TDH-39 or Sennheiser
HDA-200, EARtone 5Aseparated outputs
TDH-39 or SennheiserHDA-200, EARtone 5Aseparated outputs
TDH-39 TDH-39
Cushion type Supra-aural (HDA-200circumaural)
Supra-aural (HDA-200circumaural)
Circumaural Circumaural
BONE VIBRATORType NA NA NA B-71
SPECIAL TESTS None None None Weber, SISIPOWER REQUIREMENTS
Line, VAC 115/230 115/230 NA 220, 50/60 Hz; 110, 50/60Hz
Battery None None 6/F22 Not specified
Type, Volts NA NA 9 Not specifiedW x D x H, cm (in) 38.4 x 26.9 x 7.4 (15.1 x
10.6 x 2.9)38.4 x 26.9 x 7.4 (15.1 x10.6 x 2.9)
15.2 x 7.6 x 2.5 (6 x 3 x 1) 46 x 43 x 17 (18.1 x 16.9 x6.6)
WEIGHT, kg (lb) 2 (4.4) 2 (4.4) 0.2 (0.4) 12 (26.4)PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 2 year 2 year 1 year 1 yearDelivery time, ARO Stock, low quantities Stock, low quantities Stock, low quantities Not specifiedFiscal year January to December January to December January to December Not specified
This is the second of threepages covering the abovemodel(s). These
specifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 44/116
Audiometers; Audiometric Booths
44 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL ENTOMED ENTOMED ENTOMED EYMASASA 201 SA 202/GSI 66 SA 50 Audios 310
OTHER SPECIFICATIONS Built-in talk-forwardmicrophone; pulsed tone;software calibration;memory for 1 audiogram;optional carrying case,patient response switchsupplied. Meetsrequirements of ANSI andIEC.
Automatic Type 3audiometer; built-in real-time clock and talk-forwardmicrophone; pulsed tone;software calibration;memory for 50 audiograms;patient ID number enteredby numeric keypad; RS232interface to PC or printer;optional carrying case,patient response switchsupplied; user-programmable setup menu.Meets requirements of ANSIand IEC.
Automatic switchoff; low-battery indicator; standardheadset; optional patientresponse switch. Meetsrequirements of ANSI andIEC.
Air and bone conduction;vocal audiometry; automatictests; modulated signal;patient response control;examiners, headphones,and microphones.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model FootnotesData Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 45/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 45
Chart A: Audiometers
MODEL EYMASA EYMASA EYMASA EYMASA Audios 320 Audios 330 REDUS 75 REDUS 85
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Limited range
Speech Yes Yes No NoCONFIGURATION Tabletop Tabletop Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-20,000 (octave bands) 250-20,000 (octave bands) 250-8,000 (octave bands) 250-8,000 (octave bands)Bone conduction 250-6,000 (octave bands) 250-6,000 (octave bands) NA NAContinuous or step Continuous, step, warble Continuous pulse, warble Continuous, step Continuous, stepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +120 -10 to +120 -10 to +100 -10 to +100
8,000 Hz -10 to +90 -10 to +90 -10 to +100 -10 to +100Graduations 0.5, 1, 5 0.5, 1, 5 5 5
Bone conduction250 Hz -10 to +45 -10 to +45 NA NA1,000 Hz -10 to +70 -10 to +70 NA NA4,000 Hz -10 to +70 -10 to +70 NA NAGraduations 0.5, 1, 5 0.5, 1, 5 NA NA
Masking, noise type Narrow band, white, speech Narrow band, white, speech NA NAIntensity, dB SPL Not specified Not specified NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 46/116
Audiometers; Audiometric Booths
46 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL EYMASA EYMASA EYMASA EYMASA Audios 320 Audios 330 REDUS 75 REDUS 85
SPEECH AUDIOMETRYInputs Mic, CD Mic, CD NA NAIntensity, dB HL -10 to +110 -10 to +110 NA NAMasking, noise type Narrow band, white, speech Narrow band, white, speech NA NA
Intensity, dB SPL -10 to +90 NA NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type Circumaural Circumaural Circumaural CircumauralBONE VIBRATOR
Type B-71 B-71 NA NASPECIAL TESTS Weber, SISI, DLI Weber, SISI, DLI, SAL,
Stenger, tone decayNone Screening and abbreviated
ascendant automatic testsPOWER REQUIREMENTS
Line, VAC 220, 50/60 Hz; 110, 50/60Hz
220, 50/60 Hz; 110, 50/60Hz
220, 50/60 Hz; 110, 50/60Hz
220, 50/60 Hz; 110, 50/60Hz
Battery Not specified Not specified Not specified Not specifiedType, Volts Not specified Not specified Not specified Not specified
W x D x H, cm (in) 46 x 43 x 17 (18.1 x 16.9 x
6.6)
46 x 43 x 17 (18.1 x 16.9 x
6.6)
42.9 x 22.1 x 8.4 (16.9 x 8.7
x 3.3)
42.9 x 22.1 x 8.4 (16.9 x 8.7
x 3.3)WEIGHT, kg (lb) 12 (26.4) 12 (26.4) 6 (13.2) 6 (13.2)PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 47/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 47
Chart A: Audiometers
MODEL EYMASA EYMASA EYMASA EYMASA Audios 320 Audios 330 REDUS 75 REDUS 85
OTHER SPECIFICATIONS Air and bone conduction;vocal audiometry; automatictests; modulated signal;patient response control;examiners, headphones,and microphones.
Air and bone conduction;vocal audiometry; free field;automatic tests; modulatedsignal; patient responsecontrol; examiners,headphones, andmicrophones; optional highfrequency, tinnitus, insertearphones, externalCD/tape, and vocal.
Optional RS232 interface,printer, carrying case, andPC software. Meetsrequirements of ANSI, IEC601 and 645, and ISO 389.
Optional RS232 interface,printer, carrying case, andPC software. Meetsrequirements of ANSI, IEC601 and 645, and ISO 389.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 48/116
Audiometers; Audiometric Booths
48 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL EYMASA FRYE FRYE GM INSTRUMENTSREDUS 95 FA-10 FA-12 ASRA
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Limited range Wide range Wide range Wide range
Speech No Yes Yes NoCONFIGURATION Tabletop/portable Tabletop Tabletop PortablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 250-8,000 (octave bands) 125-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands)
125-12,000
Bone conduction 250-6,000 250-6,000 (octave,interoctave bands)
250-6,000 (octave,interoctave bands)
NA
Continuous or step Continuous, step Step Step StepFrequency accuracy Not specified Within 1% Not specified ±0.1%
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +90 -10 to +90 -20 to +901,000 Hz -10 to +100 -10 to +110 -10 to +120 -20 to +1208,000 Hz -10 to +100 -10 to +90 -10 to +90 -20 to +100Graduations 5 5 5 10
Bone conduction250 Hz -10 to +40 -10 to +45 -10 to +45 NA1,000 Hz -10 to +60 -10 to +60 -10 to +70 NA4,000 Hz -10 to +60 -10 to +60 -10 to +60 NAGraduations 5 5 5 NA
Masking, noise type NA Narrow band, white, speech Narrow band, white, speech NAIntensity, dB SPL NA 110 white; others vary 110 white; others vary NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 49/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 49
Chart A: Audiometers
MODEL EYMASA FRYE FRYE GM INSTRUMENTSREDUS 95 FA-10 FA-12 ASRA
SPEECH AUDIOMETRYInputs NA Mic, external (L, R) Mic, external (L, R) NAIntensity, dB HL NA +95 +95 NAMasking, noise type NA Speech Speech NA
Intensity, dB SPL NA -10 to +95 -10 to +105 NAEARPHONE TYPE TDH-39 TDH only, insert earphones
only, or dual calibrationTDH only, insert earphonesonly, or dual calibration
TDH-39-51
Cushion type Circumaural Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type B-71 Radioear B-71 Radioear B-71 NASPECIAL TESTS Screening and abbreviated
ascendant automatic testsStenger, ABLB, MLB, SISIare free upon request
ABLB, MLB, SISI, Stengerare free upon request
Manual/semiautomatic andautomatic Hughson-Westlake; pre-/postcomparisons andstatistics
POWER REQUIREMENTSLine, VAC 220, 50/60 Hz; 110, 50/60
Hz
105-132 105-240 100-110/220-240
Battery Not specified None None NoneType, Volts Not specified NA NA NA
W x D x H, cm (in) 42.9 x 22.1 x 8.4 (16.9 x 8.7x 3.3)
46.4 x 34.3 x 14 (18.3 x13.5 x 5.5)
46.4 x 34.3 x 14 (18.3 x13.5 x 5.5)
27 x 6 x 28 (10.63 x 2.36 x11.02)
WEIGHT, kg (lb) 6 (13.2) 5 (11) 5 (11) 1 (2.2)PURCHASEINFORMATION
List price Not specified $3,295 TDH only; $3,400inserts only; $3,575 dualcalibration
$3,295 TDH only; $3,400inserts only; $3,575 dualcalibration
$2,900
Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified 2 weeks 2 weeks 2 weeksFiscal year Not specified April to March April to March January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 50/116
Audiometers; Audiometric Booths
50 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL EYMASA FRYE FRYE GM INSTRUMENTSREDUS 95 FA-10 FA-12 ASRA
OTHER SPECIFICATIONS Optional RS232 interface,printer, and carrying case.Meets requirements of
ANSI, IEC 601 and 645, andISO 389.
Frequency-specific free-fieldcalibration; master hearing-aid curves; optional RS232interface, carrying case andfree-field speakers; optional
ABLB, SISI, MLB package;dual calibration availablewith insert earphones.Meets requirements of ANSI3.42.
Frequency-specific free-fieldcalibration; master hearing-aid curves; optional RS232interface, carrying case andfree-field speakers; optional
ABLB, SISI, MLB package;dual calibration availablewith insert earphones.Meets requirements of ANSIand ISO.
Baseline and age-relatedassessments; interfaceswith any IBM or compatiblePC running on Windows3.1, 95, 98, NT, 2000, orXP. Meets requirements ofIEC 601-1 and UL.
LAST UPDATED January 2007 July 2005 July 2005 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 51/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 51
Chart A: Audiometers
MODEL GRASON-STADLER GRASON-STADLER GRASON-STADLER GRASON-STADLERGSI 17 GSI 38 GSI 61 GSI 66
WHERE MARKETED Worldwide Worldwide Worldwide Worldwide, except USAFDA CLEARANCE Yes Yes Yes NoCE MARK (MDD) Yes Yes Yes YesTYPE Limited range Limited range Wide range Wide range
Speech No No Yes NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)125-8,000 125-12,000 (octave,
interoctave bands)125-8,000
Bone conduction NA NA 250-6,000 (octave,interoctave bands)
NA
Continuous or step Continuous, pulsed Step Both StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +70 -10 to +70 -10 to +90 -10 to +1001,000 Hz -10 to +90 -10 to +90 -10 to +120 -10 to +1108,000 Hz -10 to +70 -10 to +70 -10 to +100 -10 to +100Graduations 5 5 1, 2, 5 5
Bone conduction250 Hz NA NA -10 to +45 NA1,000 Hz NA NA -10 to +70 NA4,000 Hz NA NA -10 to +60 NAGraduations NA NA 1, 2, 5 NA
Masking, noise type NA NA Narrow band, white NAIntensity, dB SPL NA NA -10 to +110 (HL) NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 52/116
Audiometers; Audiometric Booths
52 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL GRASON-STADLER GRASON-STADLER GRASON-STADLER GRASON-STADLERGSI 17 GSI 38 GSI 61 GSI 66
SPEECH AUDIOMETRYInputs NA NA Mic, tape, CD NAIntensity, dB HL NA NA -10 to +105 NAMasking, noise type NA NA White, speech NA
Intensity, dB SPL NA NA -10 to +115 NAEARPHONE TYPE TDH-39 TDH-39 TDH-50P TDH-39, EARtone 3A
Cushion type Supra-aural Supra-aural Supra-aural Circumaural or supra-auralBONE VIBRATOR
Type NA NA B-71 NASPECIAL TESTS None specified None specified ABLB, Doerfler-Stewart,
SISI, Stenger Automatic test
POWER REQUIREMENTSLine, VAC 100, 120, 240 100, 120, 240 100, 120, 240 115/230Battery Optional Ni-Cd None None None
Type, Volts NA NA NA NAW x D x H, cm (in) 33.7 x 35.6 x 9.5 (13.3 x 14
x 3.8)33.4 x 36.8 x 10.9 (13.2 x14.5 x 4.3)
48.3 x 45.7 x 20.3 (19 x 18x 8)
38 x 27 x 7 (15 x 10.6 x 2.8)
WEIGHT, kg (lb) 2.5 (5.6) 4.5 (10) 10 (22) 2 (4.4)PURCHASEINFORMATION
List price $810-929 $2,750-3,550 $5,710-8,000 Not specifiedWarranty 1 year from date of
shipment1 year from date ofshipment
1 year from date ofshipment
1 year
Delivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December Not specified
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 53/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 53
Chart A: Audiometers
MODEL GRASON-STADLER GRASON-STADLER GRASON-STADLER GRASON-STADLERGSI 17 GSI 38 GSI 61 GSI 66
OTHER SPECIFICATIONS Battery pack. Meetsrequirements of ANSI, IEC,and ISO 9001.
1 sec tympanometry; built-inprinter; ipsilateral andcontralateral reflex testing;many versions available.Meets requirements of
ANSI, IEC, and ISO 9001.
Optional free-field speakers,built-in amplifiers, computer-interface printer port, high-frequency audiometry andinsert phones, HP DeskJetprinter, sound-field speakerplus booster amplification.Meets requirements of
ANSI, IEC, and ISO 9001.
Built-in talk-forward mic;optional carrying case, 50-audiogram memory withpatient ID, softwarecalibration, and RS232 port.
LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 54/116
Audiometers; Audiometric Booths
54 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL GRASON-STADLER GRASON-STADLER INTERACOUSTICS INTERACOUSTICSGSI 67 GSI 68 AA220/AA222 AC33
WHERE MARKETED Worldwide, except USA Worldwide, except USA Worldwide WorldwideFDA CLEARANCE No No Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Wide range
Speech No Yes Yes, AA222 YesCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 125-8,000 125-8,000 125-8,000Bone conduction 250-8,000 250-8,000 250-8,000 250-8,000Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified 1% >±2%
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +110 -10 to +1101,000 Hz -10 to +110 -10 to +110 -10 to +120 -10 to +120
8,000 Hz -10 to +100 -10 to +100 -10 to +110 -10 to +110Graduations 5 5 1, 2, 5 1 or 5
Bone conduction250 Hz -10 to +35 -10 to +35 -10 to +45 -10 to +451,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +80 -10 to +80 -10 to +80 -10 to +80Graduations 5 5 1 or 5 1 or 5
Masking, noise type Narrow band Narrow band Narrow band, white (AA222,speech)
Narrow band, white
Intensity, dB SPL 0 to +95 0 to +95 0 to +110 (dB HL) 0 to +110 (dB HL)
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 55/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 55
Chart A: Audiometers
MODEL GRASON-STADLER GRASON-STADLER INTERACOUSTICS INTERACOUSTICSGSI 67 GSI 68 AA220/AA222 AC33
SPEECH AUDIOMETRYInputs NA Mic, tape, CD (2-channel) AA222, CD, Mic, tape CD, mic, tapeIntensity, dB HL NA -10 to +100 AA222, -10 to +120 -10 to +120Masking, noise type NA Speech AA, speech noise Speech noise
Intensity, dB SPL NA 0-100 AA222, 0 to +110 (dB HL) 0 to +110 (dB HL)EARPHONE TYPE TDH-39, EARtone 3A TDH-39, EARtone 3A TDH-39 TDH-39
Cushion type Circumaural or supra-aural Circumaural or supra-aural MX-41/AR MX-41/ARBONE VIBRATOR
Type B-71 B-71 B-71 B-71SPECIAL TESTS Automatic test Automatic test SISI, ABLB, automatic
threshold (HW and Bekesy),Stenger (binaural pure-tonestimulation), AA222 alsoincludes binaural speech
SISI, ABLB, MLD, automaticthreshold (HW), Stenger(binaural pure-tonestimulation), binauralspeech
POWER REQUIREMENTSLine, VAC 115/230 115/230 100-120 or 200-240, 50/60
Hz100-120 or 200-240, 50/60Hz
Battery None None None NoneType, Volts NA NA NA NA
W x D x H, cm (in) 38 x 27 x 7 (15 x 10.6 x 2.8) 48 x 27 x 7 (18.9 x 10.6 x2.8)
35 x 26 x 10 (14 x 10.2 x 4) 50 x 47 x 20 (20 x 19 x 8)
WEIGHT, kg (lb) 2 (4.4) 2.7 (6) 3.3 (7.1) 9 (20)PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 1 year 1 year 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified January to December January to December
This is the second of threepages covering the above
model(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 56/116
Audiometers; Audiometric Booths
56 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL GRASON-STADLER GRASON-STADLER INTERACOUSTICS INTERACOUSTICSGSI 67 GSI 68 AA220/AA222 AC33
OTHER SPECIFICATIONS Built-in talk-forward mic;optional carrying case, 50-audiogram memory withpatient ID, softwarecalibration, and RS232 port.
Built-in talk-forwardmicrophone; optionalcarrying case, 50-audiogram memory withpatient ID, and softwareRS232 port.
Diagnostic impedanceaudiometer; built-in thermalprinter; different headsetand probe holder solutions;clinical probe system;optional carrying case,warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking;RS232C or USBconnection; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring and
printing; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.
Warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking;RS232C or USBconnection; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; NOAH compatible.Meets requirements of ANSIand IEC.
LAST UPDATED January 2004 January 2004 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 57/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 57
Chart A: Audiometers
MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AC40 AC440/Equinox AD226 AD229e
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Wide range Wide range
Speech Yes Yes Yes YesCONFIGURATION Tabletop Black box Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-20,000 125-8,000 125-8,000 125-8,000Bone conduction 250-8,000 250-8,000 250-8,000 250-8,000Continuous or step Step Step Step StepFrequency accuracy > ±1% >±1% >±1% >±1%
Intensity, dB HLAir conduction
250 Hz -10 to +110 -10 to +110 -10 to +110 -10 to +1101,000 Hz -10 to +120 -10 to +120 -10 to +120 -10 to +120
8,000 Hz -10 to +110 -10 to +110 -10 to +110 -10 to +110Graduations 1 or 5 1 or 5 1 or 5 1 or 5
Bone conduction250 Hz -10 to +45 -10 to +45 -10 to +45 -10 to +451,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +80 -10 to +80 -10 to +80 -10 to +80Graduations 1 or 5 1 or 5 1 or 5 1 or 5
Masking, noise type Narrow band, white Narrow band, white Narrow band, white Narrow band, whiteIntensity, dB SPL 0 to +110 (dB HL) 0 to +110 (dB HL) 0 to +110 (dB HL) 0 to +110 (dB HL)
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 58/116
Audiometers; Audiometric Booths
58 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AC40 AC440/Equinox AD226 AD229e
SPEECH AUDIOMETRYInputs CD, mic, tape CD, mic, tape NA CD, mic, tapeIntensity, dB HL -10 to +120 -10 to +120 NA -10 to +120Masking, noise type Speech noise Speech NA Speech
Intensity, dB SPL 0 to +110 (dB HL) 0 to +110 (dB HL) NA 0 to +110 (dB HL)EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type MX-41/AR MX-41/AR MX-41/AR MX-41/ARBONE VIBRATOR
Type B-71 B-71 B-71 B-71SPECIAL TESTS SISI, ABLB, MLD, automatic
threshold (HW and Bekesy),Stenger (binaural pure-tonestimulation), binauralspeech, HF and volumescaling
ABLB, Stenger (binauralpure-tone stimulation),binaural speech, HF, MF,SISI, MLD,, Weber, MasterHearing Aid, Hearing LossSimulation
SISI, ABLB, automaticthreshold (HW and Bekesy),Stenger (binaural pure-tonestimulation)
SISI, ABLB, automaticthreshold (HW and Bekesy),Stenger (binaural pure-tonestimulation), binauralspeech, MHA
POWER REQUIREMENTSLine, VAC 100-120 or 200-240, 50/60
Hz
100-120 or 200-240, 50/60
Hz
100-120 or 200-240, 50/60
Hz
100-120 or 200-240, 50/60
HzBattery None None None None
Type, Volts NA NA NA NAW x D x H, cm (in) 50 x 47 x 20 (20 x 19 x 8) 37 x 43.5 x 7.7 (14.5 x 17 x
3)35 x 26 x 10 (14 x 10 x 4) 48 x 40 x 15 (19 x 16 x 6)
WEIGHT, kg (lb) 13 (29) 5.1 (11.3) 2.1 (4.6) 2.5 (5)PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 2 years 2 years 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December
This is the second of three
pages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 59/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 59
Chart A: Audiometers
MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AC40 AC440/Equinox AD226 AD229e
OTHER SPECIFICATIONS Warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking;RS232C or USBconnection, internal printeror output for laser printer;Windows 98 software;IaBaseII database,PrintView for online PCmonitoring and printing;NOAH compatible. Meetsrequirements of ANSI andIEC.
PC audiometer to beoperated via PC keyboardor audiometric keyboard;warble, talk-forward 0-110dB SPL, talk-back, EARtone5A insert phones (withoutEQFF in speech), HDA280(with EQFF in speech),HDA200 separatelycalibrated, synchronousmasking, and USBconnection; Windows 98software; IaBaseIIdatabase; NOAHcompatible. Meetsrequirements of ANSI andIEC.
Optional carrying case,warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking,RS232C or USBconnection, and output forMTP10 printer, laser printer,and inkjet; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; NOAH andCONNEX compatible.Meets requirements of ANSI
and IEC.
Optional carrying case,warble, talk-forward 0-110dB SPL, talk back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking,RS232C or USBconnection, and output forMTP10 printer, laser printer,and inkjet; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; NOAH andCONNEX compatible.Meets requirements of ANSI
and IEC.LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 60/116
Audiometers; Audiometric Booths
60 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AS208 AT235 AT235h AZ26
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Limited range Limited range Limited range Limited range
Speech No No No NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 125-8,000 125-8,000 125-8,000Bone conduction NA NA NA NAContinuous or step Step Step Step StepFrequency accuracy >±2% 1% 1% 2%
Intensity, dB HLAir conduction
250 Hz -10 to +90 -10 to +110 -10 to +110 -10 to +901,000 Hz -10 to +100 -10 to +120 -10 to +120 -10 to +100
8,000 Hz -10 to +100 -10 to +100 -10 to +110 -10 to +100Graduations 5 1, 2, 5 1, 2, 5 1, 2, 5
Bone conduction250 Hz NA NA NA NA1,000 Hz NA NA NA NA4,000 Hz NA NA NA NAGraduations NA NA NA NA
Masking, noise type NA NA NA NAIntensity, dB SPL NA NA NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 61/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 61
Chart A: Audiometers
MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AS208 AT235 AT235h AZ26
SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA
Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type MX-41/AR MX-41/AR MX-41/AR MX-41/ARBONE VIBRATOR
Type NA NA NA NASPECIAL TESTS None Automatic threshold (HW),
tympanometry, automaticand manual reflexometry,decay test, eustachian-tubefunction test
Automatic threshold (HW),normal and high probe-tonetympanometry, automaticand manual reflexometry,decay test, eustachian-tubefunction test
Not specified
POWER REQUIREMENTSLine, VAC Optional 100-120 or 200-
240, 50/60 Hz
100-120 or 200-240, 50/60
Hz
100-120 or 200-240, 50/60
Hz
100-120 or 200-240, 50/60
HzBattery AA batteries (4) None None None
Type, Volts 9-12 NA NA NAW x D x H, cm (in) 23 x 18 x 8 (9 x 7 x 3) 36 x 26 x 10 (14.2 x 10.2 x
4)36 x 26 x 10 (14.2 x 10.2 x4)
48 x 40 x 16 (19 x 16 x 6)
WEIGHT, kg (lb) 1.3 (2.9); includes batteriesand headset
2.8 (6.2) 2.8 (6.2) 9.5 (21)
PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 2 years 2 years 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 62/116
Audiometers; Audiometric Booths
62 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AS208 AT235 AT235h AZ26
OTHER SPECIFICATIONS Pure, warble, and pulsedtones available; patientsignal button; batteryindicator for testingreliability; optional carryingcase, power supply (100-120 or 200-240 V). Meetsrequirements of ANSI andIEC.
Diagnostic impedanceaudiometer; optionalcarrying case; built-inthermal printer; differentheadset and probe holdersolutions; clinical probesystem; EARtone 3A insertphotos; phones; Windows98 software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.
Diagnostic impedanceaudiometer; optionalcarrying case; built-inthermal printer; differentheadset and probe holdersolutions; clinical probesystem; EARtone 3A insertphones; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.
Clinical impedanceaudiometer; built-in thermalprinter; clinical probesystem; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.
LAST UPDATED January 2007 January 2007 January 2007 January 2007
Supplier FootnotesModel Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 63/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 63
Chart A: Audiometers
MODEL MADSEN MADSEN MADSEN MADSENConera ITERA MM 304 MM 602
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Wide range
Speech Yes High frequency No YesCONFIGURATION Portable/desktop, computer
drivenTabletop/portable Portable Tabletop/portable
PURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)250-12,500 (12 frequencies) 250-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands)
Bone conduction 250-8,000 (octave,interoctave bands)
250-8,000 (8 frequencies) NA 250-6,000 (octave,interoctave bands)
Continuous or step Continuous/step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HL
Air conduction250 Hz -10 to +100 -10 to +110 -10 to +90 -10 to +1001,000 Hz -10 to +125 -10 to +120 -10 to +90 -10 to +1258,000 Hz -10 to +100 -10 to +100 -10 to +90 -10 to +100Graduations 1, 5 1, 2.5, 5 5 5
Bone conduction250 Hz -10 to +45 -10 to +40 NA -10 to +451,000 Hz -10 to +75 -10 to +80 NA -10 to +754,000 Hz -10 to +75 -10 to +75 NA -10 to +75Graduations 1, 2, 3, 5, 10 1, 2.5, 5 NA 5
Masking, noise type White, speech Narrow band, white NA Narrow bandIntensity, dB SPL +40 to +90 -10 to +110 NA -10 to +115
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 64/116
Audiometers; Audiometric Booths
64 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MADSEN MADSEN MADSEN MADSENConera ITERA MM 304 MM 602
SPEECH AUDIOMETRYInputs Mic, tape, CD Tape, CD, mic NA Mic, tape 1, tape 2Intensity, dB HL -10 to +115 -10 to +80 NA -10 to +115Masking, noise type Narrow band, white, speech White, speech, ipsilateral,
contralateralNA White, speech
Intensity, dB SPL -10 to +115 -10 to +80 NA -10 to +115EARPHONE TYPE TDH-39, ER-3A TDH-39 TDH-39 TDH-39
Cushion type Not specified Circumaural Circumaural Supra-auralBONE VIBRATOR
Type B-71 B-71 NA B-71SPECIAL TESTS Automatic threshold, SISI,
ABLB, sound-level check,Stenger, Bekesy
Fowler (ABLB), SISI,Stenger, hearing instrumentstimulator
None SISI, Stenger, Fowler
POWER REQUIREMENTSLine, VAC 100/115/127/220/240 110-240, 50/60 Hz NA 100-120/200-240Battery None None Yes None
Type, Volts NA NA 9 NA
W x D x H, cm (in) 38.1 x 35 x 13 (15 x 13.8 x5.1)
45 x 29 x 8.5 (17 x 11.4 x3.3)
30.5 x 21.6 x 6.4 (12 x 8.5 x2.5)
38.1 x 39.4 x 10.9 (15 x15.5 x 4.3)
WEIGHT, kg (lb) 4.6 (10.1) 4.5 (9.9) 1.3 (2.8) 3.9 (8.6)PURCHASEINFORMATION
List price $4,995 $3,995 $795 $2,995Warranty 2 years 2 years 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue onto
the next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 65/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 65
Chart A: Audiometers
MODEL MADSEN MADSEN MADSEN MADSENConera ITERA MM 304 MM 602
OTHER SPECIFICATIONS Computer required. Optional carrying case,extended high-frequencyoption to 16,000 Hz.
Standard ME70 headset;optional AC adapter. Meetsrequirements of ANSI andIEC.
Computer memory,calibration, and test setups;audible patient response.Meets requirements of ANSIand IEC.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 66/116
Audiometers; Audiometric Booths
66 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MADSEN MADSEN MAICO MAICOMM 622 OB 922 MA 25 Series II MA 27
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes No NoTYPE Wide range Wide range, high frequency Limited range Limited range
Speech Yes Yes No NoCONFIGURATION Tabletop/portable Tabletop Portable PortablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)125-20,000 (octave,interoctave bands)
250-8,000 (octave bands) 250-8,000 (octave bands)
Bone conduction 250-6,000 (octave,interoctave bands)
250-8,000 (octave,interoctave bands)
NA NA
Continuous or step Step Step/continuous Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +90 -15 to +901,000 Hz -10 to +125 -10 to +125 -10 to +90 -15 to +908,000 Hz -10 to +100 -10 to +100 -10 to +90 -15 to +90Graduations 1, 5 1, 5 5 5
Bone conduction250 Hz -10 to +45 -10 to +45 NA NA1,000 Hz -10 to +75 -10 to +75 NA NA4,000 Hz -10 to +75 -10 to +75 NA NAGraduations 1, 5 1, 2, 3, 5, 10 NA NA
Masking, noise type Narrow band Narrow band, speech,white, external
NA NA
Intensity, dB SPL -10 to +115 -10 to +115 NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 67/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 67
Chart A: Audiometers
MODEL MADSEN MADSEN MAICO MAICOMM 622 OB 922 MA 25 Series II MA 27
SPEECH AUDIOMETRYInputs Mic, tape 1, tape 2 Mic, tape, CD NA NAIntensity, dB HL -10 to +115 -10 to +115 NA NAMasking, noise type White, speech Narrow band, white,
speech, externalNA NA
Intensity, dB SPL -10 to +115 -10 to +115 NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type B-71 B-71 NA NASPECIAL TESTS SISI, Stenger, Fowler ABLB, DLI, SISI, Stenger,
autothreshold,suprathreshold, Rainville,tone decay, Bekesy
None specified None specified
POWER REQUIREMENTSLine, VAC 100-120/200-240 110/115/127/220/240 Optional 120 adapter 117; 220 optionalBattery None None Yes None
Type, Volts NA NA Not specified NAW x D x H, cm (in) 38.1 x 39.4 x 10.9 (15 x
15.5 x 4.3)52.1 x 41.9 x 15.2 (20.5 x16.5 x 6)
21.6 x 19.1 x 9.1 (8.5 x 7.5x 3.6)
35.6 x 43.2 x 17.8 (14 x 17x 7)
WEIGHT, kg (lb) 3.9 (8.6) 9.1 (20) 1.8 (4) 2.7 (6)PURCHASEINFORMATION
List price $4,500 $5,995 $825 $915Warranty 2 years 2 years 1 year 1 yearDelivery time, ARO Not specified Not specified 10 days 10 daysFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). These
specifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 68/116
Audiometers; Audiometric Booths
68 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MADSEN MADSEN MAICO MAICOMM 622 OB 922 MA 25 Series II MA 27
OTHER SPECIFICATIONS Computer memory,calibration, and test setups;audible patient response;masking and test signalsgiven simultaneously. Meetsrequirements of ANSI andIEC.
Computer interface;bidirectional software; 2screen languages; clock;12-test memory; optionalprinter. Meets requirementsof ANSI and IEC.
Battery operated withoptional 120 VAC adapter;carrying case. Meetsrequirements of ANSI andIEC.
Built-in carrying case. Meetsrequirements of ANSI andIEC.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 69/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 69
Chart A: Audiometers
MODEL MAICO MAICO MAICO MAICOMA 39 MA 40 MA 41 MA 42
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) No No No NoTYPE Limited range Wide range Wide range Wide range
Speech No No Yes YesCONFIGURATION Portable Portable Portable Portable/tabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave bands) 125-8,000 (octave bands) 125-8,000 (octave bands) 125-8,000Bone conduction NA 125-4,000 (octave bands) 125-4,000 (octave bands) 125-4,000Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +110 -10 to +110 -10 to +110 -10 to +110
8,000 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +100Graduations 5 5 5 5
Bone conduction250 Hz NA -10 to +40 -10 to +40 -10 to +401,000 Hz NA -10 to +70 -10 to +70 -10 to +704,000 Hz NA -10 to +70 -10 to +70 -10 to +70Graduations NA 5 5 5
Masking, noise type NA Narrow band Narrow band, speech Narrow band, speechIntensity, dB SPL NA -10 to +100 -10 to +100 -10 to +100
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 70/116
Audiometers; Audiometric Booths
70 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MAICO MAICO MAICO MAICOMA 39 MA 40 MA 41 MA 42
SPEECH AUDIOMETRYInputs NA NA Mic, tape, CD Mic, tape, CDIntensity, dB HL NA NA -10 to +100 -10 to +100Masking, noise type NA NA Narrow band, speech Narrow band, speech
Intensity, dB SPL NA NA -10 to +100 -10 to +100EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type NA B-71 B-71 B-71SPECIAL TESTS None specified None specified None specified None specifiedPOWER REQUIREMENTS
Line, VAC 117/234 117/234 117/234 117/234Battery None None None None
Type, Volts NA NA NA NAW x D x H, cm (in) 32 x 40 x 16 (12.5 x 15.5 x
6.3)32 x 40 x 16 (12.5 x 15.5 x6.3)
32 x 40 x 16 (12.5 x 15.5 x6.3)
32 x 40 x 16 (12.5 x 15.5 x6.3)
WEIGHT, kg (lb) 7.5 (16.5) 7.5 (16.5) 7.5 (16.5) 7.5 (16.5)
PURCHASEINFORMATION
List price $1,235 $1,725 $2,100 $2,495Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 10 days 10 days 10 days 10 daysFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 71/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 71
Chart A: Audiometers
MODEL MAICO MAICO MAICO MAICOMA 39 MA 40 MA 41 MA 42
OTHER SPECIFICATIONS Digital display. Meetsrequirements of ANSI andIEC.
Digital display. Meetsrequirements of ANSI andIEC.
Digital display. Meetsrequirements of ANSI andIEC.
Master hearing aid with 4commonly used frequencyslopes to imitate hearing-aidacoustics; 2-channel speechtesting; built-in amplifier.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 72/116
Audiometers; Audiometric Booths
72 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MAICO MAICO MEDIROLL MEDIROLLMA 790 MA 800 Rollingtone AT-61 SA-51
WHERE MARKETED Worldwide Worldwide Not specified Not specifiedFDA CLEARANCE Yes Yes No NoCE MARK (MDD) No No Not specified YesTYPE Limited range Limited range Not specified Wide range
Speech No No Optional NoCONFIGURATION Tabletop Tabletop Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 500-8,000 (octave bands) 500-8,000 (octave bands) 125-10,000 250-8,000Bone conduction NA NA 250-8,000 NAContinuous or step NA NA Step StepFrequency accuracy Not specified Not specified <1% <1%
Intensity, dB HLAir conduction
250 Hz NA NA -10 to +90 -10 to +801,000 Hz 0 to +95 0 to +95 -10 to +120 -10 to +100
8,000 Hz 0 to +95 0 to +95 -10 to +100 -10 to +80Graduations 5 5 5 or 1 5
Bone conduction250 Hz NA NA -10 to +45 NA1,000 Hz NA NA -10 to +70 NA4,000 Hz NA NA -10 to +60 NAGraduations NA NA 5 or 1 NA
Masking, noise type NA NA NBN; WN; SN NAIntensity, dB SPL NA NA -10 to +100(AC); -10 to
+60(BC); NBC -10 to+110(AC); -10 to +70(BC);WN -10 to +110(AC);SN -10to +70(BC) SN
NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 73/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 73
Chart A: Audiometers
MODEL MAICO MAICO MEDIROLL MEDIROLLMA 790 MA 800 Rollingtone AT-61 SA-51
SPEECH AUDIOMETRYInputs NA NA CD or built-in speech test NAIntensity, dB HL NA NA -10 to +110 (AC); -10 to +65
(BC)NA
Masking, noise type NA NA NBN; WN; SN NAIntensity, dB SPL NA NA -10 to +100 (AC); -10 to +60
(BC); NBC -10 to +110(AC); -10 to +70(BC) WN; -10 to +110-(AC); -10 to +70(BC) SN
NA
EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39Cushion type Supra-aural Supra-aural Not specified Not specified
BONE VIBRATORType NA NA B-71 NA
SPECIAL TESTS None specified None specified TTDT; SBLB; SISI; ABLB;Lucher; Bekesy; automatictest; built-in speech test;
"Melody" test
Automatic test
POWER REQUIREMENTSLine, VAC 117/234, 50/60 Hz 117/234, 50/60 Hz 110-240 V; 50-60 Hz 230, 50 HzBattery None None NA None
Type, Volts NA NA NA NAW x D x H, cm (in) 37.5 x 32.5 x 10 (15 x 13 x
4)37.5 x 32.5 x 10 (15 x 13 x4)
27.5 x 19.5 x 5 (10.8 x 5.9 x2)
19.5 x 15 x 5 (7.7 x 5.9 x 2)
WEIGHT, kg (lb) 4.5 (10) 4.5 (10) 1 (2) 0.45 (1)PURCHASEINFORMATION
List price $3,225 $3,795 €4,100 EUR 1,396Warranty 1 year 1 year 2 years 2 yearsDelivery time, ARO 10 days 10 days 60 days 40 daysFiscal year January to December January to December Not specified Not specified
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 74/116
Audiometers; Audiometric Booths
74 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MAICO MAICO MEDIROLL MEDIROLLMA 790 MA 800 Rollingtone AT-61 SA-51
OTHER SPECIFICATIONS Automatic computer ormanual testing; LCD;thermal printouts; RS232 ormodem interface. Meetsrequirements of ANSI andIEC.
Automatic computer ormanual testing; LCD;thermal printouts; RS232 ormodem interface; stores upto 500 audiograms andbaselines in memory. Meetsrequirements of ANSI andIEC.
Aplitutde and frequencymodulation; syncronousmasking; 3 seperatechanels, LCD, free-fieldaudiometry with specialamplifier, computerinterface; mains adapter,carrying case; extra smallsize; Meets requirements ofISO 389, EN 60645, andIEC 60601.
Noise-excluding headset;stores audiograms for 99patients; carrying case;LCD; computer interface;built-in mic; built-in watch.Meets requirements of IEC645-4.
LAST UPDATED July 2005 July 2005 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 75/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 75
Chart A: Audiometers
MODEL MEDIROLL MICRO AUDIOMETRICS MICRO AUDIOMETRICS MICRO AUDIOMETRICSSA-6 Mini Earscan 3 Automatic
Threshold Audiometer
(ES3)
Earscan 3 Manual Audiometer (ES3M)
Earscan 3 Screening Audiometer (ES3S)
WHERE MARKETED Not specified Worldwide Worldwide WorldwideFDA CLEARANCE No Yes Yes YesCE MARK (MDD) Yes No No NoTYPE Wide range Wide range Limited range Wide range
Speech No No No NoCONFIGURATION Portable Handheld, tabletop, portable Handheld, tabletop, portable Handheld, tabletop, portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 250-8,000 125-8,000 250-8,000 125-8,000Bone conduction NA NA NA NAContinuous or step Step Step Step StepFrequency accuracy <1% <1% <1% <1%
Intensity, dB HLAir conduction
250 Hz -10 to +70 -10 to +90 -10 to +80 -10 to +801,000 Hz -10 to +100 -10 to +100 -10 to +80 -10 to +808,000 Hz -10 to +90 -10 to +90 -10 to +80 -10 to +80Graduations 5 5 5 5
Bone conduction250 Hz NA NA NA NA1,000 Hz NA NA NA NA4,000 Hz NA NA NA NAGraduations NA NA NA NA
Masking, noise type NA No No NoIntensity, dB SPL NA NA NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 76/116
Audiometers; Audiometric Booths
76 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MEDIROLL MICRO AUDIOMETRICS MICRO AUDIOMETRICS MICRO AUDIOMETRICSSA-6 Mini Earscan 3 Automatic
Threshold Audiometer
(ES3)
Earscan 3 Manual Audiometer (ES3M)
Earscan 3 Screening Audiometer (ES3S)
SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA
Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-39 TDH-39, EAR-5A TDH-39, EAR-5A TDH-39, EAR-5A
Cushion type Not specified Supra-aural, inserts Supra-aural Supra-aural, insertsBONE VIBRATOR
Type NA NA NA NASPECIAL TESTS Automatic test; "melody"
test for children (patented) Automatic threshold,binaural presentations
Binaural presentations Automatic screening,binaural presentations
POWER REQUIREMENTSLine, VAC 230, 50 Hz 100-240 100-240 100-240Battery None AA alkaline (4) AA alkaline (4) AA alkaline (4)
Type, Volts NA 5 5 5
W x D x H, cm (in) 12.3 x 7.1 x 3.5 (4.8 x 2.8 x1.4)
9.1 x 3.3 x 14.7 (3.6 x 1.3 x5.8)
9.1 x 3.3 x 14.7 (3.6 x 1.3 x5.8)
9.1 x 3.3 x 14.7 (3.6 x 1.3 x5.8)
WEIGHT, kg (lb) 0.2 (0.4) 0.28 (0.6125) 0.28 (0.6125) 0.28 (0.6125)PURCHASEINFORMATION
List price EUR 1,099 $1,995 $795 $995Warranty 2 years 3 years 3 years 3 yearsDelivery time, ARO 40 days 7-10 days 7-10 days 7-10 daysFiscal year Not specified June to May June to May June to May
This is the second of threepages covering the abovemodel(s). Thesespecifications continue onto
the next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 77/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 77
Chart A: Audiometers
MODEL MEDIROLL MICRO AUDIOMETRICS MICRO AUDIOMETRICS MICRO AUDIOMETRICSSA-6 Mini Earscan 3 Automatic
Threshold Audiometer
(ES3)
Earscan 3 Manual Audiometer (ES3M)
Earscan 3 Screening Audiometer (ES3S)
OTHER SPECIFICATIONS Noise-excluding headset;stores audiograms for 50patients; carrying case;LCD. Meets requirements ofIEC 645-4.
Talkover; RS-232, USB;optional printer, datastorage; STS and PBIcalculations; batteries andcarrying case included.Meets requirements of ANSIS3.6, IEC, and ISO. MeetsOSHA requirements foraudiometers
Talkover; Batteries andcarrying case included.Meets requirements of ANSIS3.6, IEC, and ISO.
Talkover; RS-323, USB;batteries and carrying caseincluded. Meetsrequirements of ANSI S3.6,IEC, and ISO.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 78/116
Audiometers; Audiometric Booths
78 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MONITOR MONITOR NAGASHIMA NAGASHIMAMI-5000B MI-7000 AD-98 MT-3A
WHERE MARKETED Canada, USA Canada, USA Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Limited range Limited range Not specified Wide range
Speech No No Yes YesCONFIGURATION Tabletop/portable Tabletop/portable Tabletop Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 500-8,000 500-8,000 125-8,000 (octave bands) 125-8,000 (octave bands)Bone conduction NA NA 250-8,000 250-6,000Continuous or step Step Step Step StepFrequency accuracy 1% 1% Not specified Not specified
Intensity, dB HLAir conduction
250 Hz NA NA -10 to +90 -10 to +901,000 Hz 0 to +90 0 to +90 -10 to +110 -10 to +100
8,000 Hz 0 to +90 0 to +90 -10 to +90 -10 to +90Graduations 5 5 5 5
Bone conduction250 Hz NA NA -10 to +55 -10 to +501,000 Hz NA NA -10 to +70 -10 to +604,000 Hz NA NA -10 to +60 -10 to +50Graduations NA NA 5 5
Masking, noise type No No Narrow band, SPN Narrow band, SPNIntensity, dB SPL NA NA -10 to +110 +40 to +100
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 79/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 79
Chart A: Audiometers
MODEL MONITOR MONITOR NAGASHIMA NAGASHIMAMI-5000B MI-7000 AD-98 MT-3A
SPEECH AUDIOMETRYInputs NA NA Auxiliary AuxiliaryIntensity, dB HL NA NA 0 to +110 -10 to +100Masking, noise type NA NA Narrow band, SPN, white Narrow band, SPN, white
Intensity, dB SPL NA NA -10 to +110 +40 to +100EARPHONE TYPE TDH-49 TDH-49 TDH-49 TDH-49
Cushion type MX-41/AR MX-41/AR 51 51BONE VIBRATOR
Type NA NA BR-572 BR-572SPECIAL TESTS None None SISI, ABLB NonePOWER REQUIREMENTS
Line, VAC 120/240 100-250 100-120, 220-240, 50/60 Hz 100-120, 220-240, 50/60 HzBattery None None None None
Type, Volts NA NA NA NAW x D x H, cm (in) 18.4 x 21.6 x 7 (7.3 x 8.5 x
2.8)24 x 29 x 7.5 (9.5 x 11.5 x3)
42 x 43 x 31.7 (16.5 x 17 x12.5)
21 x 23 x 8.5 (8.3 x 9.1 x3.3)
WEIGHT, kg (lb) 1.4 (3) 2.3 (5) 7.5 (16.5) 3.8 (8.3)
PURCHASEINFORMATION
List price $1,995 $2,795 Not specified Not specifiedWarranty 2 years 2 years 1 year 1 yearDelivery time, ARO 3 weeks 3 weeks 30 days 30 daysFiscal year January to December January to December July to June July to June
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 80/116
Audiometers; Audiometric Booths
80 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL MONITOR MONITOR NAGASHIMA NAGASHIMAMI-5000B MI-7000 AD-98 MT-3A
OTHER SPECIFICATIONS RS232 interface. RS232 interface; talkover;internal printer.
Type 1 diagnosticaudiometer. Conforms toJIS T1201-1982.
Type 1 diagnosticaudiometer. Conforms toJIS T1201-1982.
LAST UPDATED July 2005 July 2005 April 2001 April 2001Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 81/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 81
Chart A: Audiometers
MODEL NAGASHIMA QUALITONE QUALITONE QUALITONENSA-100S AAJ-P : C-AAJ AS-110 Screener AS-90 Screener
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Yes Yes YesCE MARK (MDD) Not specified Yes Yes YesTYPE Not specified Wide range, ANSI Type 2C Wide range, ANSI Type 4 Wide range, ANSI Type 4
Speech No Yes No NoCONFIGURATION Tabletop/portable Portable Portable PortablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 500-4,000 (octave bands) 125-8,000 (octave,
interoctave bands)250-8,000 (octave,interoctave bands)
250-8,000 (octave,interoctave bands)
Bone conduction No 250-4,000 (octave,interoctave bands)
NA NA
Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz NA -10 to +100 -10 to +100 -10 to +901,000 Hz -10 to +70 -10 to +110 -10 to +110 -10 to +908,000 Hz NA -10 to +100 -10 to +100 -10 to +90Graduations 5 2.5, 5 5 5
Bone conduction250 Hz NA -10 to +70 NA NA1,000 Hz NA -10 to +70 NA NA4,000 Hz NA -10 to +70 NA NAGraduations NA 2.5, 5 NA NA
Masking, noise type No Narrow band, white, speech NA NAIntensity, dB SPL NA -10 to +120 NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 82/116
Audiometers; Audiometric Booths
82 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL NAGASHIMA QUALITONE QUALITONE QUALITONENSA-100S AAJ-P : C-AAJ AS-110 Screener AS-90 Screener
SPEECH AUDIOMETRYInputs No Mic, tape, auxiliary NA NAIntensity, dB HL No -10 to +110 NA NAMasking, noise type No Narrow band, white, speech NA NA
Intensity, dB SPL NA -10 to +120 NA NAEARPHONE TYPE TDR-30 TDH-39 TDH-39 TDH-39
Cushion type 41 Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type NA B-71A, B-72 NA NASPECIAL TESTS None None None NonePOWER REQUIREMENTS
Line, VAC 100-120, 220-240, 50/60 Hz 110/220 110/220 110/220Battery None None None Alkaline
Type, Volts NA NA NA 6; 55 hr life (maximum @ 8hr/day)
W x D x H, cm (in) 26.5 x 22 x 10 (10.4 x 8.7 x4)
48.3 x 35.6 x 22.9 (19 x 14x 9) : 48.3 x 35.6 x 14 (19 x
14 x 5.5)
33 x 33 x 15.2 (13 x 13 x 6) 22.9 x 20.3 x 10.2 (9 x 8 x4)
WEIGHT, kg (lb) 3 (6.6) 11.8 (26) : 9.1 (20.1) 4.1 (9) 1.8 (4)PURCHASEINFORMATION
List price Not specified $3,950; $3840 $865 $790Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 30 days Not specified Not specified Not specifiedFiscal year July to June January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 83/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 83
Chart A: Audiometers
MODEL NAGASHIMA QUALITONE QUALITONE QUALITONENSA-100S AAJ-P : C-AAJ AS-110 Screener AS-90 Screener
OTHER SPECIFICATIONS Screening audiometer.Conforms to JIS TR01-1982.
Free-field capability; masterhearing aid; 2 channels;built-in cassette deck. Meetsrequirements of ANSI Type2C.
Continuous pulsed-tonewarble (FM); patient signal;reversible tone-interrupterswitch; group-presentationcapability. Meetsrequirements of ANSI Type4.
Continuous, pulsed-tonewarble; patient signal;reversible tone-interrupterswitch. Meets requirementsof ANSI Type 4.
LAST UPDATED April 2001 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 84/116
Audiometers; Audiometric Booths
84 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL QUALITONE QUALITONE QUALITONE SIBELMEDCD 2 : CD 3 LWAAJ-P WR-A : WR-B : WR-C AC50-C
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes NoCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range, ANSI Type 2C Wide range Wide range
Speech Yes Yes Yes Yes (optional)CONFIGURATION Portable Portable Portable TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)125-8,000 (octave,interoctave bands)
125-8,000 (octave,interoctave bands)
125-8,000
Bone conduction 250-4,000 (octave,interoctave bands)
250-4,000 (octave,interoctave bands)
250-4,000 (octave,interoctave bands)
250-4,000
Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to [+110 : +120] -10 to +110 -10 to [+110 : +120] -10 to +1208,000 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +100Graduations 2.5, 5 2.5, 5 2.5, 5 5
Bone conduction250 Hz -10 to +70 -10 to +70 -10 to +45 -10 to +501,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +70Graduations 2.5, 5 2.5, 5 2.5, 5 5
Masking, noise type Narrow band, white, speech Narrow band, white, speech Narrow band, white, speech Narrow band, whiteIntensity, dB SPL -10 to +120 -10 to +120 -10 to +120 -10 to +100
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 85/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 85
Chart A: Audiometers
MODEL QUALITONE QUALITONE QUALITONE SIBELMEDCD 2 : CD 3 LWAAJ-P WR-A : WR-B : WR-C AC50-C
SPEECH AUDIOMETRYInputs Mic, CD, auxiliary Mic, auxiliary Mic, auxiliary Tape, CD, mic (optional)Intensity, dB HL -10 to +110 -10 to +110 -10 to +110 -10 to +100Masking, noise type Narrow band, white, speech Narrow band, white, speech White, speech, narrow band Speech
Intensity, dB SPL -10 to +120 -10 to +125 0 to +125 : 0 to +120 : 0 to+120
-10 to +100
EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39Cushion type Supra-aural Supra-aural Supra-aural Supra-aural
BONE VIBRATORType B-71A, B-72 B-71A, B-72 B-71A, B-72 B-71
SPECIAL TESTS None SISI optional None ABLB, Luscher, Stenger,warble, SISI (optional)
POWER REQUIREMENTSLine, VAC 110/220 110/220 110/220 220 (others optional)Battery None None None Not specified
Type, Volts NA NA NA Not specifiedW x D x H, cm (in) 48.3 x 35.6 x 22.9 (19 x 14
x 9)
48.3 x 35.6 x 12.7 (19 x 14
x 5)
33 x 33 x 15.2 (13 x 13 x 6) 47 x 25 x 9.5 (18.5 x 9.8 x
3.7)WEIGHT, kg (lb) 12.2 (27) 7.8 (17.2) 4.5 (10) 4.5 (9.9)PURCHASEINFORMATION
List price $4,750 : $4,995 $3,375 $1,285 : $1,750 : $1,920 $1,400Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified 7 daysFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 86/116
Audiometers; Audiometric Booths
86 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL QUALITONE QUALITONE QUALITONE SIBELMEDCD 2 : CD 3 LWAAJ-P WR-A : WR-B : WR-C AC50-C
OTHER SPECIFICATIONS Free-field capability; masterhearing aid; 2 channels;channel reverse; patientsignal; electronic frequencyswitching; channel mixing;talkover, talk-back. Meetsrequirements of ANSI.
Free-field capability; masterhearing aid; 2 channels.Meets requirements of ANSIType 2C.
1 channel; reversible toneinterrupter; master hearingaid; patient signal warble.Meets requirements of
ANSI.
External printer; internaldatabase, hearing-losscalculation; AM; FM; 2channels; optional PCdatabase and free field.Meets requirements of IEC601-1 and 645 and ISO9001.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 87/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 87
Chart A: Audiometers
MODEL SIBELMED SIBELMED SIBELMED SIBELMED AC50-D AS5-A AS5-AOM SIBELSOUND 400 A
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Limited range Limited range
Speech Yes No No Yes (optional)CONFIGURATION Tabletop Portable Portable TabletopPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 250-8,000 250-8,000 125-8,000Bone conduction 250-4,000 NA 250-4,000 250-4,000Continuous or step Step Step Step BothFrequency accuracy Not specified Not specified Not specified 2%
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +90 -10 to +90 -10 to +1001,000 Hz -10 to +120 -10 to +105 -10 to +105 -10 to +120
8,000 Hz -10 to +100 -10 to +80 -10 to +80 -10 to +100Graduations 5 5 5 1 or 5 (selectable)
Bone conduction250 Hz -10 to +50 NA -10 to +40 -10 to +501,000 Hz -10 to +70 NA -10 to +50 -10 to +704,000 Hz -10 to +70 NA -10 to +50 -10 to +70Graduations 5 NA 5 1 or 5 (selectable)
Masking, noise type Narrow band, white White White Narrow band, white, speechIntensity, dB SPL -10 to +100 +30 to +80 +30 to +80 -10 to +100
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 88/116
Audiometers; Audiometric Booths
88 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL SIBELMED SIBELMED SIBELMED SIBELMED AC50-D AS5-A AS5-AOM SIBELSOUND 400 A
SPEECH AUDIOMETRYInputs Tape, CD, mic NA NA Tape, CD, mic (optional)Intensity, dB HL -10 to +100 NA NA -10 to +100Masking, noise type Speech NA NA Speech
Intensity, dB SPL -10 to +100 NA NA -10 to +100EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type B-71 NA B-71 B-71SPECIAL TESTS ABLB, Luscher, Stenger,
warble, SISINone None ABLB, Luscher, Stenger,
warble, SISI (optional)POWER REQUIREMENTS
Line, VAC 220 (other options) 220 (others optional) 220 (others optional) 110-220Battery Not specified Not specified Not specified No
Type, Volts Not specified Not specified Not specified NoW x D x H, cm (in) 47 x 25 x 9.5 (18.5 x 9.8 x
3.7)32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)
32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)
47 x 25 x 9.5 (18.5 x 9.8 x3.7)
WEIGHT, kg (lb) 4.5 (9.9) 2.5 (5.5) 2.5 (5.5) 4.5 (9.9)PURCHASEINFORMATION
List price $1,600 $750 $1,100 $1,400Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 7 days 7 days 7 days 7 daysFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 89/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 89
Chart A: Audiometers
MODEL SIBELMED SIBELMED SIBELMED SIBELMED AC50-D AS5-A AS5-AOM SIBELSOUND 400 A
OTHER SPECIFICATIONS External printer; internaldatabase; hearing-losscalculation; AM; FM; 2channels; optional PCdatabase and free field.Meets requirements of IEC601-1 and 645 and ISO9001.
Optional masking. Meetsrequirements of ANSI 53.6,IEC 601-1, and ISO 9001.
Meets requirements of ANSI53.6, IEC 601-1, and ISO9001.
USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1-1, ISO 9001, and ISE13485.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 90/116
Audiometers; Audiometric Booths
90 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL SIBELMED SIBELMED SIBELMED SIBELMEDSIBELSOUND 400 AM SIBELSOUND 400 AO SIBELSOUND 400 AOM SIBELSOUND 400 AOM+
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesTYPE
Speech Yes (optional) Yes (optional) Yes (optional) YesCONFIGURATION Tabletop Portable Portable PortablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 125-8,000 125-20,000 125-20,000Bone conduction 250-4,000 NA 250-4,000 250-4,000Continuous or step Both Both Both BothFrequency accuracy 2% 2% 2% 2%
Intensity, dB HLAir conduction
250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +120 -10 to +120 -10 to +120 -10 to +120
8,000 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +100Graduations 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable)
Bone conduction250 Hz -10 to +50 -10 to +50 -10 to +50 -10 to +501,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +70Graduations 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable)
Masking, noise type Narrow band, white, speechnoise
Narrow band, white, speechnoise
Narrow band, white, speechnoise
Narrow band, white, speechnoise
Intensity, dB SPL -10 to +100 -10 to +100 -10 to +100 -10 to +100
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 91/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 91
Chart A: Audiometers
MODEL SIBELMED SIBELMED SIBELMED SIBELMEDSIBELSOUND 400 AM SIBELSOUND 400 AO SIBELSOUND 400 AOM SIBELSOUND 400 AOM+
SPEECH AUDIOMETRYInputs Tape, CD, mic (optional) Tape, CD, mic (optional) Tape, CD, mic (optional) Tape, CD, micIntensity, dB HL -10 to +100 -10 to +100 -10 to +100 -10 to +100Masking, noise type Speech Speech Speech Speech
Intensity, dB SPL -10 to +100 -10 to +100 -10 to +100 -10 to +100EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39
Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type B-71 B-71 B-71 B-71SPECIAL TESTS ABLB, Luscher, Stenger,
warble, SISI (optional) ABLB, Luscher, Stenger,warble, SISI (optional)
ABLB, Luscher, Stenger,warble, SISI (optional)
ABLB, Luscher, Stenger,warble, SISI (optional)
POWER REQUIREMENTSLine, VAC 110-240 110-240 110-240 110-240Battery No No No No
Type, Volts No No No NoW x D x H, cm (in) 47 x 25 x 9.5 (18.5 x 9.8 x
3.7)32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)
32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)
32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)
WEIGHT, kg (lb) 4.5 (9.9) 2.5 (5.5) 2.5 (5.5) 2.5 (5.5)PURCHASEINFORMATION
List price $1,600 $750 $1,100 $1,100Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 7 days 7 days 7 days 7 daysFiscal year January to December January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 92/116
Audiometers; Audiometric Booths
92 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL SIBELMED SIBELMED SIBELMED SIBELMEDSIBELSOUND 400 AM SIBELSOUND 400 AO SIBELSOUND 400 AOM SIBELSOUND 400 AOM+
OTHER SPECIFICATIONS USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.
USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.
USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM,FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.
USB connection andprinting; internal database,hearing-loss calculation; 2channels (AM, FM); optionalPC database and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 93/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 93
Chart A: Audiometers
MODEL SIBELMED SINGER SINGER SINGERSIBELSOUND 400 SUPRA MD-1 MD-3 MD-4
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No Yes Yes YesCE MARK (MDD) Yes No No NoTYPE Limited range Limited range Wide range
Speech Yes No No NoCONFIGURATION Portable Portable Tabletop Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-20,000 500-6,000 (octave plus
6,000)500-6,000 (octave,interoctave bands)
125-8,000 (octave,interoctave bands)
Bone conduction 250-4,000 NA NA NAContinuous or step Both Step Step StepFrequency accuracy 2% Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +100 NA NA -10 to +100
1,000 Hz -10 to +120 0 to +100 0 to +100 -10 to +1108,000 Hz -10 to +100 NA NA -10 to +100Graduations 1 or 5 (selectable) 5 5 5
Bone conduction250 Hz -10 to +50 NA NA NA1,000 Hz -10 to +70 NA NA NA4,000 Hz -10 to +70 NA NA NAGraduations 1 or 5 (selectable) NA NA NA
Masking, noise type Narrow band, white, speechnoise
NA NA NA
Intensity, dB SPL -10 to +100 NA NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 94/116
Audiometers; Audiometric Booths
94 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL SIBELMED SINGER SINGER SINGERSIBELSOUND 400 SUPRA MD-1 MD-3 MD-4
SPEECH AUDIOMETRYInputs Tape, CD, mic NA NA NAIntensity, dB HL -10 to +100 NA NA NAMasking, noise type Speech NA NA NA
Intensity, dB SPL -10 to +100 NA NA NAEARPHONE TYPE TDH-39 TDH-50 TDH-50 TDH-49
Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type B-71 NA NA NASPECIAL TESTS ABLB, Luscher, Stenger,
warble, SISI (optional)None None Pulse
POWER REQUIREMENTSLine, VAC 110-240 105-125; optional 220 105-125; optional 220 105-125; optional 220Battery No Not specified Not specified Not specified
Type, Volts No Not specified Not specified Not specifiedW x D x H, cm (in) 32.4 x 24.6 x 9.5 (12.8 x 9.7
x 3.7)45.7 x 33 x 14 (18 x 13 x5.5)
48.3 x 25.4 x 27.9 (19 x 10x 11)
38.1 x 17.8 x 6.4 (15 x 7 x2.5)
WEIGHT, kg (lb) 2.5 (5.5) 7.7 (17) 11.3 (25) 1.6 (3.5)PURCHASEINFORMATION
List price $1,100 $2,295 $2,695 $695Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 7 days Not specified Not specified Not specifiedFiscal year January to December Not specified Not specified Not specified
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 95/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 95
Chart A: Audiometers
MODEL SIBELMED SINGER SINGER SINGERSIBELSOUND 400 SUPRA MD-1 MD-3 MD-4
OTHER SPECIFICATIONS USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM), optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.
Includes automatictympanometer; price withoptional printer is $2,700.Meets requirements of
ANSI.
Includes automatictympanometer; price withoptional printer is $2,995.Meets requirements of
ANSI.
Model draws audiograms;price with optional printer is$995. Meets requirementsof ANSI.
LAST UPDATED January 2007 January 2004 January 2004 January 2004Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 96/116
Audiometers; Audiometric Booths
96 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL STARKEY TREMETRICS TREMETRICS TREMETRICS AA 50 RA 300 : 3000 RA 500 RA 650
WHERE MARKETED Not specified Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes No No NoTYPE ANSI Type 1A Not specified Limited range Limited range
Speech Yes No No NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable Tabletop/mobile vanPURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 500-8,000 (octave,
interoctave bands)500-8,000 (octave,interoctave bands)
500-8,000 (octave,interoctave bands)
Bone conduction 250-4,000 NA NA NAContinuous or step Step Both Both BothFrequency accuracy Not specified Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +110 NA NA NA
1,000 Hz -10 to +120 0 to +100 -10 to +100 -10 to +1008,000 Hz -10 to +110 0 to +100 -10 to +100 -10 to +100Graduations 1 to 5 5 5 5
Bone conduction250 Hz -10 to +45 NA NA NA1,000 Hz -10 to +70 NA NA NA4,000 Hz -10 to +80 NA NA NAGraduations 1 to 5 NA NA NA
Masking, noise type Narrow band, white, speech NA NA NAIntensity, dB SPL 0 to +100 NA NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 97/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 97
Chart A: Audiometers
MODEL STARKEY TREMETRICS TREMETRICS TREMETRICS AA 50 RA 300 : 3000 RA 500 RA 650
SPEECH AUDIOMETRYInputs Mic, CD, auxillary NA NA NAIntensity, dB HL 0 to +100 NA NA NAMasking, noise type Speech NA NA NA
Intensity, dB SPL 0 to +100 NA NA NAEARPHONE TYPE TDH-39, 3A, 5A TDH-39 TDH-39 TDH-39
Cushion type Circumaural/insert Supra-aural Supra-aural Supra-auralBONE VIBRATOR
Type B-71 NA NA NASPECIAL TESTS ABLB, Stenger, SISI,
automatic threshold,Langenbeck, dual intensity
None specified : 1-buttonexpress test mode
None None
POWER REQUIREMENTSLine, VAC 100-240, automatic
switching power supply90-240, selectable 120/240 90-240, selectable
Battery None None None NoneType, Volts NA NA NA NA
W x D x H, cm (in) 44 x 51 x 17 (17.3 x 20 x6.3)
33 x 28 x 11.5 (13 x 11 x4.5)
44.5 x 31.1 x 10.2 (17.5 x12.3 x 4)
16.7 x 31.8 x 22.9 (6.6 x12.5 x 9)
WEIGHT, kg (lb) Not specified 2.7 (6.0) 4.3 (9.5) 2.9 (6.4)PURCHASEINFORMATION
List price $6,500 $1,895 $4,295 Depends on number ofchannels
Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified January to December January to December January to December
This is the second of threepages covering the abovemodel(s). These
specifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 98/116
Audiometers; Audiometric Booths
98 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL STARKEY TREMETRICS TREMETRICS TREMETRICS AA 50 RA 300 : 3000 RA 500 RA 650
OTHER SPECIFICATIONS Master hearing aid;frequency field; printingoption; hearing losssimulator; automaticscoring; talk-forward, talk-back; patient signal; twinmics; external audio input.
Additional frequencies: 200,2500, 3500, 4500, 5000 Hz.NOAH certified. Meetsrequirements of ANSI S3.6Type 1 and EN 60601-1.
Patented multimediacapability for computer-voice-instructed testing;RS232 port; optionaltalkover. Meetsrequirements of ANSI andISO 9001.
Automatic, semiautomatic,and manual test; 25customizable test questions;memory for 1,200audiograms;telecommunicationssoftware; dual computer-interface ports. Meetsrequirements of ANSI andISO 9001.
Windows 95, 98, NTcontroller; group testingsystem; optional 1-4channels/module with up to8 channels per PCcontroller; audio/visual errorcodes; built-in talkover; rawdata transfer;telecommunicationssoftware; computer voiceinstructions in 5 languages;patented multimediacapability. Meetsrequirements of ANSI, ISO9001, and OSHA.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model FootnotesData Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 99/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 99
Chart A: Audiometers
MODEL Welch Allyn Welch Allyn Welch Allyn AM232 AudioScope 3 TM262
WHERE MARKETED Worldwide Worldwide WorldwideFDA CLEARANCE Yes Not specified YesCE MARK (MDD) Yes Not specified YesTYPE Limited range Not specified Limited range
Speech No Not specified NoCONFIGURATION Tabletop/portable Not specified Tabletop/portablePURE-TONEAUDIOMETRY
Frequency, HzAir conduction 125-8,000 (octave,
interoctave bands)500, 1,000, 2,000, 4,000 125-8,000
Bone conduction NA NA NAContinuous or step Continuous, pulsed Continuous StepFrequency accuracy Not specified Not specified Not specified
Intensity, dB HLAir conduction
250 Hz -10 to +70 Not specified -10 to +70
1,000 Hz -10 to +90 20, 25, 40 -10 to +908,000 Hz -10 to +70 NA -10 to +70Graduations 5 Not specified 5
Bone conduction250 Hz NA NA NA1,000 Hz NA NA NA4,000 Hz NA NA NAGraduations NA NA NA
Masking, noise type NA NA NAIntensity, dB SPL NA NA NA
This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 100/116
Audiometers; Audiometric Booths
100 ©2007 ECRI Institute. All Rights Reserved.
Chart A: Audiometers
MODEL Welch Allyn Welch Allyn Welch Allyn AM232 AudioScope 3 TM262
SPEECH AUDIOMETRYInputs NA NA NAIntensity, dB HL NA NA NAMasking, noise type NA NA NA
Intensity, dB SPL NA NA NAEARPHONE TYPE TDH-39 NA TDH-39
Cushion type Supra-aural NA Supra-auralBONE VIBRATOR
Type NA NA NASPECIAL TESTS None specified None specified None specifiedPOWER REQUIREMENTS
Line, VAC 100, 120, 240 110, 220, 240, 240/260 100, 120, 240Battery Optional Ni-Cad Ni-Cad No
Type, Volts Not specified 3.5 V NAW x D x H, cm (in) 33.7 x 35.6 x 9.5 (13.3 x 14
x 3.8)3.7 x 21.6 (1.4 x 8.5) 33.4 x 36.8 x 10.9 (13.2 x
14.5 x 4.3)WEIGHT, kg (lb) 2.5 (5.6) 0.25 (0.55) 4.5 (10)
PURCHASEINFORMATION
List price $1,083 $678 $3,398-3,942Warranty 1 year 1 year 1 yearDelivery time, ARO 5 days 5 days 5 daysFiscal year January to December January to December January to December
This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 101/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 101
Chart A: Audiometers
MODEL Welch Allyn Welch Allyn Welch Allyn AM232 AudioScope 3 TM262
OTHER SPECIFICATIONS Meets requirements of ANSI, IEC, and ISO 9001.
Meets requirements of ANSI, CSA, ETL, IEC.
One tympanometry, built-inprinter, ipsilateral reflextesting, two versionsavailable. Meetsrequirements of ANSI, IEC,and ISO 9001.
LAST UPDATED July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 102/116
Audiometers; Audiometric Booths
102 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL ACOUSTAIR ACOUSTAIR ACOUSTAIR ACOUSTAIR A-1 A-2 A-3 Audicel
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specified
TYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type Optional Optional Optional OptionalVENTILATION, m³/min(ft³/min)
2.1 (75) 2.1 (75) 2.1 (75) 2.1 (75)
LIGHT PLS-2xgw PLS-2xgw PLS-2xgw PLS-2xgwNumber 1 1 2 1
AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Optional Optional Optional OptionalELECTRICAL OUTLET Optional Optional Optional OptionalW x D x H, m (ft)
Inside, exam 1 x 1 x 2 (3.3 x 3.3 x 6.8) 1.6 x 1 x 2 (5.1 x 3.3 x 6.8) 1.8 x 1.6 x 2 (5.9 x 5.1 x6.8)
1 x 1 x 2 (3.3 x 3.3 x 6.8)
Inside, control Not specified Not specified Not specified Not specified
Outside, overall 1.2 x 1.2 x 2.3 (4 x 4 x 7.6) 1.8 x 1.2 x 2.3 (5.8 x 4 x7.6) 2 x 1.8 x 2.3 (6.6 x 5.8 x7.6) 1.1 x 1.1 x 2.3 (3.6 x 3.6 x7.6)WINDOW
W x H, cm (in) 60 x 60 (24 x 24) 60 x 60 (24 x 24) 60 x 60 (24 x 24) 60 x 60 (24 x 24)WEIGHT, kg (lb) 700 (1,550) 900 (2,000) 1,200 (2,600) 600 (1,300)LIST PRICE Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO On request On request On request On requestFISCAL YEAR Not specified Not specified Not specified Not specifiedOTHER SPECIFICATIONS Radio shielding available;
other sizes possible.Radio shielding available;other sizes possible.
Radio shielding available;other sizes possible.
Radio shielding available;other sizes possible.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 103/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 103
Chart B: Audiometric Booths
MODEL ACOUSTIC ACOUSTIC ACOUSTIC ACOUSTIC142 Delta 143 Delta RE-125 : RE-126 RE-130
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Optional OptionalJACK PANEL
Number, type 10, stereo phono; 3- and 4-pin CJ; also 2.54 cm (1 in)pass-through
10, stereo phono; 3- and 4-pin CJ; also 2.54 cm (1 in)pass-through
4, stereo phono, pass-through, folding shelf
6, stereo phono
VENTILATION, m³/min(ft³/min)
5.7 (200) 5.7 (200) 2.8 (100) 2.8 (100)
LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2 2 1 1
AUDIOMETER SHELF Optional Optional Yes YesCARPETING Yes Yes Yes YesELECTRICAL OUTLET 6-plug strip mount, external 6-plug strip mount, external None NoneW x D x H, m (ft)
Inside, exam 6 x 6 x 6.5 (15.2 x 15.2 x16.5) (triangular shape)
2.3 x 2.3 x 2 (7.4 x 7.4 x6.5) (triangular shape)
0.7 x 0.9 x 1.5 (2.3 x 3 x 5) 0.6 x 0.9 x 1.6 (2 x 3 x 5.2)
Inside, control NA NA NA NAOutside, overall 6.6 x 6.6 x 7.8 (17.6 x 17.6
x 19.8)2.4 x 2.4 x 2.3 (8 x 8 x 7.8) 0.8 x 1 x 1.8 (2.7 x 3.3 x 6) 1 x 0.8 x 2.01 (3.3 x 2.7 x
6.6)WINDOW
W x H, cm (in) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 45.7 x 61 (18 x 24) 45.7 x 61 (18 x 24)WEIGHT, kg (lb) 1,590 (3,500), shipping 1,633 (3,600), shipping 376 (830) 376 (830), shippingLIST PRICE $9,700 $12,425 $3,200 : $3,200 $4,500WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30-45 days 30-45 days 30 days 30 daysFISCAL YEAR October to September October to September October to September October to SeptemberOTHER SPECIFICATIONS Triangular unit designed for
corner fit in small host room;intended for sound-fieldtesting. Meets requirements
of ANSI.
Unit is triangular anddesigned to fit in a cornerfor sound-field testing in asmall host room. Meets
requirements of ANSI.
All units preassembled;attenuation values for 125-8,000 Hz (octaves)available. Meets
requirements of ANSI.
All units preassembled; 3"thick walls; attenuationvalues for 125-8,000 Hz(octaves) available. Meets
requirements of ANSI.LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 104/116
Audiometers; Audiometric Booths
104 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL ACOUSTIC ACOUSTIC ACOUSTIC ACOUSTICRE-132 RE-140 : RE-141 RE-142 RE-240
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Double wallCONFIGURATION Single room Single room Double room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
VENTILATION, m³/min(ft³/min)
5.7 (200) 2.8 (100) 5.7 (200) 2.8 (100)
LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2 1 2 1
AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 1 None 2; 1 duplex 6-plug strip 6-plug strip mounted
externally under windowW x D x H, m (ft)
Inside, exam 1.8 x 2 x 2.1 (6 x 6.5 x 7) 1 x 0.9 x 2 (3.3 x 3 x 6.5) :1.2 x 0.9 x 2 (4 x 3 x 6.5)
1.9 x 1.8 x 2 (6.3 x 6 x 6.5) 1 x 0.9 x 2 (3.3 x 3 x 6.5)
Inside, control NA NA NA NAOutside, overall 2 x 2.1 x 2.4 (6.5 x 7 x 7.8) 1.2 x 1.1 x 2.3 (4 x 3.7 x
7.7) : 1.4 x 1.2 x 2.3 (4.6 x 4x 7.7)
2.1 x 2 x 2.1 (6.9 x 6.7 x6.9)
1.6 x 1.5 x 2.5 (5.3 x 5 x8.1)
WINDOWW x H, cm (in) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24)
WEIGHT, kg (lb) 1,089 (2,400), shipping 748 (1,650), shipping 1,651 (3,640), shipping 1,814 (4,000), shippingLIST PRICE $8,650 $5,675 $10,700 $14,950WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30 days 30 days 30-45 days 30-45 daysFISCAL YEAR October to September October to September October to September October to SeptemberOTHER SPECIFICATIONS Meets requirements of
ANSI.
Meets requirements of
ANSI.
7 other models available in
varied sizes; optional radio-frequency shielding. Meetsrequirements of ANSI.
8 other models available in
varied sizes; optional radio-frequency shielding; largermodels have 2 to 4 lightsand 2 duplex electricaloutlets. Meets requirementsof ANSI.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 105/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 105
Chart B: Audiometric Booths
MODEL ACOUSTIC ACOUSTIC ACOUSTIC ACOUSTICRM-182 RM-282 RS-142 RS-242
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Double wall Single wall Double wallCONFIGURATION Single room with 4
examination stationsDouble room with 4examination stations
Double room Double room
VIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type NA1 NA1 10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
VENTILATION, m³/min(ft³/min)
5.7 (200) 5.7 (200) 5.7 (200), each room 5.7 (200), each room
LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2 2 2/room 2/room
AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET None None 2/room 2/roomW x D x H, m (ft)
Inside, exam 2.2 x 5.3 x 2 (7.3 x 5 x 6.5) 2.2 x 5.3 x 2 (7.3 x 5 x 6.5) 1.8 x 1.9 x 2 (6 x 6.3 x 6.5) 1.8 x 1.9 x 2 (6 x 6.3 x 6.5)Inside, control NA NA 1.2 x 1.9 x 2 (4 x 6.3 x 6.5) 1.2 x 1.9 x 2 (4 x 6.3 x 6.5)Outside, overall 2.4 x 1.7 x 2.2 (8 x 5.6 x
7.3)2.8 x 2.1 x 2.4 (9.3 x 6.9 x8)
3.5 x 2.1 x 2.3 (11.6 x 7 x7.4)
4 x 2.5 x 2.4 (13 x 8.3 x 8)
WINDOWW x H, cm (in) 61 x 33 (24 x 13) 61 x 45.7 (24 x 18) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24)
WEIGHT, kg (lb) 1,706 (3,760), shipping 3,302 (7,280), shipping 2,989 (6,590), shipping 5,489 (12,100), shippingLIST PRICE $13,750 $22,000 $19,975 $35,425WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30-45 days 30-45 days 30-45 days 30-45 daysFISCAL YEAR October to September October to September October to September October to SeptemberOTHER SPECIFICATIONS Group testing; can seat/test
4 people simultaneously;metal stools and divider
curtains available; othermodels seat 2, 6, 8, or 10people. Meets requirementsof ANSI.
Group testing; can seat/test4 people simultaneously;metal stools and divider
curtains included; othermodels seat 2, 6, 8, or 10people. Meets requirementsof ANSI.
7 other models available invaried sizes; optional radio-frequency shielding. Meets
requirements of ANSI.
7 other models available invaried sizes; optional radio-frequency shielding. Meets
requirements of ANSI.
LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes
Model Footnotes
Data Footnotes 1Each test station has anindividual jack box markedfor left and rightheadphones and aresponse button wired to anexternal patchbox.
1Each test station has anindividual jack box markedfor left and rightheadphones and aresponse button wired to anexternal patchbox.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 106/116
Audiometers; Audiometric Booths
106 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL ACOUSTIC AMPLIFON AMPLIFON AMPLIFONRS-252 AF100 AF111 AF210
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Yes Yes YesTYPE Single-wall control room,
double-wall examinationroom
Single wall Single wall Single wall
CONFIGURATION Double room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through
Jack and RCA for allaudiometers
Jack and RCA for allaudiometers
Jack and RCA for allaudiometers
VENTILATION, m³/min(ft³/min)
5.7 (200), each room Not specified Not specified Not specified
LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2/room 1 1 1
AUDIOMETER SHELF Optional Optional Optional Optional
CARPETING Yes Yes Yes YesELECTRICAL OUTLET 2/room Yes Yes YesW x D x H, m (ft)
Inside, exam 1.9 x 1.8 x 2 (6.3 x 5.9 x6.5)
0.9 x 1 x 1.9 (3 x 3.3 x 6.2) 1.1 x 1.1 x 2.1 (3.6 x 3.6 x6.9)
2.3 x 1.1 x 2.1 (7.4 x 3.6 x6.9)
Inside, control 1.8 x 2.2 x 2.2 (5.9 x 7.2 x7.2)
NA NA NA
Outside, overall 4.5 x 2.4 x 2.4 (14.7 x 8 x 8) 1 x 1.1 x 2.2 (3.3 x 3.6 x 7) 1.3 x 1.3 x 2.3 (4.1 x 4.1 x7.4)
2.4 x 1.3 x 2.3 (7.8 x 4.1 x7.4)
WINDOWW x H, cm (in) 76.2 x 61 (30 x 24) 71 x 61 (28 x 24) 63 x 54 (25 x 21) 63 x 54 (25 x 21)
WEIGHT, kg (lb) 4,695 (10,350), shipping 300 (662) 260 (573) 380 (838)LIST PRICE $29,650 Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30-45 days 60 days 60 days 60 daysFISCAL YEAR October to September January to December January to December January to December
OTHER SPECIFICATIONS 6 other models available invaried sizes; optional radio-frequency shielding. Meetsrequirements of ANSI.
None specified. None specified. None specified.
LAST UPDATED July 2005 December 2002 December 2002 December 2002Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 107/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 107
Chart B: Audiometric Booths
MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON AF220 AF90 G1x1 G2x2
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Yes Yes Yes YesTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type Jack and RCA for allaudiometers
Jack and RCA for allaudiometers
Jack and RCA for allaudiometers
Jack and RCA for allaudiometers
VENTILATION, m³/min(ft³/min)
Not specified Not specified Not specified Not specified
LIGHT Incandescent Incandescent Incandescent IncandescentNumber 1 1 1 1
AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET Yes Yes Yes YesW x D x H, m (ft)
Inside, exam 2.3 x 2.3 x 2.1 (7.4 x 7.4 x6.9)
0.8 x 0.8 x 1.8 (2.7 x 2.7 x5.9)
1.1 x 1.1 x 2 (3.6 x 3.6 x6.6)
2.2 x 2.2 x 2 (7.3 x 7.3 x6.6)
Inside, control NA NA NA NAOutside, overall 2.4 x 2.4 x 2.3 (7.8 x 7.8 x
7.4)0.9 x 0.9 x 1.9 (3 x 3 x 6.3) 1.3 x 1.3 x 2.3 (4.1 x 4.1 x
7.4)2.4 x 2.4 x 2.3 (7.8 x 7.8 x7.4)
WINDOWW x H, cm (in) 63 x 54 (25 x 21) 56 x 56 (22 x 22) 55 x 70 (22 x 28) 55 x 70 (22 x 28)
WEIGHT, kg (lb) 560 (1,235) 150 (331) 470 (1,036) 1,310 (2,889)LIST PRICE Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 60 days 60 days 60 days 60 daysFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. Optional custom sizes with
built-in Faraday cage.Optional custom sizes withbuilt-in Faraday cage.
LAST UPDATED December 2002 December 2002 December 2002 December 2002Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 108/116
Audiometers; Audiometric Booths
108 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL AMPLIFON ECKEL ECKEL ECKELMini Silent Cabin AB-150 AB-200 AB-2000
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Yes Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes No Yes YesJACK PANEL
Number, type Jack and RCA for allaudiometers
4, stereo phono; RS232 port 4, stereo phono 4, stereo phono
VENTILATION, m³/min(ft³/min)
Not specified Passive 30 changes/hr 30 changes/hr
LIGHT Incandescent None Optional OptionalNumber 1 NA Not specified Not specified
AUDIOMETER SHELF Optional None Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET Yes No None NoW x D x H, m (ft)
Inside, exam 0.6 x 0.9 x 1.6 (1.9 x 3 x5.2)
11 x 9.8 x 19.5 (36 x 32 x64)
11 x 8.5 x 17.7 (36 x 28 x58)
11.6 x 7.9 x 20.7 (38 x 26 x68) : 11.6 x 11 x 20.7 (38 x36 x 68)
Inside, control NA NA NA NAOutside, overall 0.7 x 1 x 1.7 (2.3 x 3.3 x
5.6)1 x 0.9 x 1.7 (40 x 36 x 66) 12.2 x 9.8 x 19.8 (40 x 32 x
65)12.8 x 9.1 x 22.9 (42 x 30 x75) : 12.8 x 12.2 x 22.9 (42x 40 x 75)
WINDOWW x H, cm (in) 55 x 70 (22 x 28) 91.4 x 167.6 (36 x 66) 58.4 x 58.4 (23 x 23) 58.4 x 58.4 (23 x 23)
WEIGHT, kg (lb) 220 (485) 99.8 (220), shipping 272 (600), shipping 318 (700), shippingLIST PRICE Not specified $2,300 $2,790 $3,025WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 60 days 4-6 weeks 2-4 weeks 2-4 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS Can be installed on test
vehicles.None specified. None specified. None specified.
LAST UPDATED December 2002 January 2004 January 2004 January 2004
Supplier FootnotesModel Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 109/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 109
Chart B: Audiometric Booths
MODEL ECKEL ECKEL ECKEL ECKEL AB-4230 : AB-4240 ABC-100 C-11 C-12
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Double wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes No Yes YesJACK PANEL
Number, type 4, stereo phono Pass-through 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
VENTILATION, m³/min(ft³/min)
30 changes/hr Passive 30 changes/hr 30 changes/hr
LIGHT Optional None Single SingleNumber Not specified NA 2 2
AUDIOMETER SHELF Optional No Optional OptionalCARPETING Yes No Yes YesELECTRICAL OUTLET No No 2 duplex 2 duplexW x D x H, m (ft)
Inside, exam 1 x 0.7 x 1.7 (3.2 x 2.2 x5.7) : 1 x 0.9 x 1.7 (3.2 x 3 x5.7)
10.1 x 6.4 x 4.6 (33 x 21 x15)
0.9 x 1.2 x 2 (2.9 x 3.8 x6.6)
1.2 x 1.4 x 2 (3.8 x 4.7 x6.6)
Inside, control NA NA NA NAOutside, overall 1.1 x 0.8 x 1.9 (3.5 x 2.5 x
6.3) : 1.1 x 1 x 1.9 (3.5 x 3.3x 6.3)
11 x 7.3 x 5.5 (36 x 24 x 18) 1.1 x 1.3 x 2.3 (3.5 x 4.4 x7.5)
1.3 x 1.6 x 2.3 (4.3 x 5.3 x7.5) : 1.3 x 2.2 x 2.4 (4.3 x7.1 x 7.8)
WINDOWW x H, cm (in) 58.4 x 58.4 (23 x 23) 38.1 x 76.2 (15 x 30) 61 x 76.2 (24 x 30), double
glazed61 x 76.2 (24 x 30), doubleglazed
WEIGHT, kg (lb) 295 (650) : 344.7 (760),shipping
22.7 (50), shipping 711 (1,568) 1,026 (2,263)
LIST PRICE $3,195 : $3,695 $2,150 $5,900 $6,500WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 2-4 weeks 2-4 weeks 4-6 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to December
OTHER SPECIFICATIONS None specified. None specified. None specified. None specified.LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 110/116
Audiometers; Audiometric Booths
110 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL ECKEL ECKEL ECKEL ECKELC-13 C-14 C-15 C-16 : C-17
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
VENTILATION, m³/min(ft³/min)
30 changes/hr 30 changes/hr 30 changes/hr 30 changes/hr
LIGHT Single Single Single SingleNumber 2 2 2 2
AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 2 duplex 2 duplex 2 duplex 2 duplexW x D x H, m (ft)
Inside, exam 1.5 x 1.7 x 2 (4.8 x 5.6 x6.6)
1.7 x 1.9 x 2 (5.6 x 6.2 x6.6)
1.9 x 2.3 x 2 (6.5 x 7.5 x6.6)
2.5 x 2.8 x 2 (8.3 x 9.2 x6.6)
Inside, control NA NA NA NAOutside, overall 1.6 x 1.9 x 2 (5.3 x 6.2 x
6.6)1.9 x 2.2 x 2.3 (6.2 x 7.1 x7.5)
2.2 x 2.4 x 2.3 (7.1 x 8 x7.5)
2.4 x 2.5 x 2.3 (8 x 8.1 x7.5) : 2.5 x 2.8 x 2.3 (8.1 x9.1 x 7.5)
WINDOWW x H, cm (in) 61 x 76.2 (24 x 30), double
glazed61 x 76.2 (24 x 30), doubleglazed
61 x 76.2 (24 x 30), doubleglazed
61 x 76.2 (24 x 30), doubleglazed
WEIGHT, kg (lb) 1,163 (2,565) 1,412 (3,114) 1,641 (3,618) 2,036.7 (4,491) : 2,440.8(5,382)
LIST PRICE $8,700 $9,600 $12,200 $13,825 : $14,625WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. None specified. None specified.
LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 111/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 111
Chart B: Audiometric Booths
MODEL ECKEL ECKEL ECKEL ECKELC-24 : C-25 C-26 : C-27 CL-66 : CL-88 S-121 : S-122
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Double wall Double wall Single wall Single-wall control room,
double-wall examinationroom
CONFIGURATION Single room Single room Single room with cornerbooth
Suite
VIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
VENTILATION, m³/min(ft³/min)
30 changes/hr 30 changes/hr 30 changes/hr 30 changes/hr
LIGHT Single Single Bulb/track SingleNumber 2 2 2 2
AUDIOMETER SHELF Optional Optional Optional Optional
CARPETING Yes Yes Yes YesELECTRICAL OUTLET 2 duplex 2 duplex 2 duplex 2 duplexW x D x H, m (ft)
Inside, exam 1.7 x 1.9 x 2 (5.6 x 6.5 x6.6) : 1.9 x 2.2 x 2 (6.5 x 7.4x 6.6)
2.3 x 2.5 x 2 (7.4 x 8.3 x6.6) : 2.5 x 2.8 x 2 (8.3 x 9.2x 6.6)
1.6 x 1.6 x 2 (5.1 x 5.1 x6.6) : 2.3 x 2.3 x 2 (7.4 x 7.4x 6.6)
2.3 x 1.7 x 2.4 (7.5 x 5.6 x7.9) : 2.5 x 2 x 2.4 (8.2 x 6.6x 7.9)
Inside, control NA NA Not specified 1.8 x 1.9 x 2 (6 x 6.2 x 6.6) :1.8 x 2.3 x 2 (6 x 7.5 x 6.6)
Outside, overall 2.3 x 2.6 x 2.5 (7.5 x 8.5 x8.1) : 2.6 x 2.9 x 2.5 (8.5 x9.4 x 8.1)
2.9 x 3.1 x 2.5 (9.4 x 10.3 x8.1) : 3.1 x 3.4 x 2.5 (10.3 x11.2 x 8.1)
1.9 x 1.9 x 2.3 (6.5 x 6.5 x7.7) : 2.4 x 2.4 x 2.3 (8 x 8 x7.7)
1.9 x 3.9 x 2.3 (6.2 x 12.9 x7.5) : 2.7 x 5 x 2.4 (8.9 x16.4 x 7.9)
WINDOWW x H, cm (in) 61 x 76.2 (24 x 30), double
glazed61 x 76.2 (24 x 30), doubleglazed
61 x 76.2 (24 x 30), doubleglazed
61 x 76.2 (24 x 30), doubleglazed
WEIGHT, kg (lb) 2,923 (6,446) : 3,370.4(7,432)
3,730.5 (8,226) : 4,326.4(9,540)
975 (2,150 ) : 1,292.5(2,850)
4,021.2 (8,865) : 5,041.3(11,114)
LIST PRICE $17,100 : $21,000 $24,150 : $26,500 $7,425 : $9,000 $27,475 : $29,850WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. None specified. None specified.LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 112/116
Audiometers; Audiometric Booths
112 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL ECKEL ECKEL ECKEL EYMASAS-123 : S-124 S-34 : S-45 S-56 : S-57 CI/CL 40
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified YesTYPE Single-wall control room,
double-wall examinationroom
Single wall Single wall Single wall
CONFIGURATION Suite Suite Suite Not specifiedVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through
6, stereo phono
VENTILATION, m³/min(ft³/min)
30 changes/hr 30 changes/hr 30 changes/hr Assisted
LIGHT Single Single Single IncandescentNumber 2 2 2 1
AUDIOMETER SHELF Optional Optional Optional YesCARPETING Yes Yes Yes YesELECTRICAL OUTLET 2 duplex 2 duplex 2 duplex 1W x D x H, m (ft)
Inside, exam 2.8 x 2.3 x 2.4 (9.2 x 7.5 x7.9) : 3 x 2.5 x 2.4 (10.1 x8.2 x 7.9)
1.7 x 1.5 x 2 (5.6 x 4.8 x6.6) : 1.9 x 1.7 x 2 (6.5 x 5.6x 6.6)
2.3 x 1.9 x 2 (7.5 x 6.2 x6.6) : 2.5 x 2.3 x 2 (8.2 x 7.5x 6.6)
1.3 x 1.2 x 2.1 (4.2 x 3.9 x6.8), optional custom sizes
Inside, control 2.2 x 2.6 x 2 (7.1 x 8.6 x6.6) : 2.7 x 2.9 x 2 (8.9 x 9.5x 6.6)
1.7 x 1.9 x 2 (5.6 x 6.5 x6.6) : 1.9 x 2.3 x 2 (6.5 x 7.4x 6.6)
2.3 x 1.9 x 2 (7.5 x 6.2 x6.6) : 2.5 x 2.3 x 2 (8.2 x 7.5x 6.6)
Not specified
Outside, overall 2.4 x 5 x 2.3 (8 x 16.4 x 7.5): 2.5 x 5.5 x 2.3 (8.2 x 18 x7.5)
1.8 x 3.9 x 2.2 (6.2 x 12.9 x7.5) : 2.1 x 4.5 x 2.2 (7.1 x14.6 x 7.5)
2.4 x 5 x 2.3 (8 x 16.4 x 7.5): 2.5 x 5.5 x 2.3 (8.2 x 18.2x 7.5)
1.3 x 1.2 x 2.2 (4.3 x 3.9 x7.2)
WINDOWW x H, cm (in) 61 x 76.2 (24 x 30), double
glazed61 x 76.2 (24 x 30), doubleglazed
61 x 76.2 (24 x 30), doubleglazed
41 x 75 (16 x 30)
WEIGHT, kg (lb) 5,827.9 (12,848) : 6,583.6(14,514)
2,575.4 (5,679) : 3,053(6,732)
3,678.2 (8,109) : 4,478.4(9,873)
460 (1,014.3)
LIST PRICE $37,325 : $42,800 $17,550 : $20,000 $24,150 : $27,800 $2,949WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks 2 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. None specified. Optional left- or right-hand
swing door; attenuationvalues available for 125-8,000 Hz (octaves). Meetsrequirements of ANSI.
LAST UPDATED January 2004 January 2004 January 2004 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 113/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 113
Chart B: Audiometric Booths
MODEL EYMASA EYMASA EYMASA IACCI-35 CI-40 CI-45 : CI-50 120A-1
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Yes Yes Yes Not specifiedTYPE Single wall Single wall Single wall : Double wall Double wallCONFIGURATION Not specified Not specified Not specified Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL
Number, type 6, stereo phono 6, stereo phono 6, stereo phono 10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through
VENTILATION, m³/min(ft³/min)
Free Free Assisted 2.8 (100)
LIGHT Incandescent Incandescent Incandescent FluorescentNumber 1 1 1 1
AUDIOMETER SHELF Yes Yes Yes OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 1 1 1 1W x D x H, m (ft)
Inside, exam Not specified Not specified 1.3 x 1.2 x 2.1 (4.2 x 3.9 x6.8)
1.2 x 1.2 x 2 (3.8 x 3.8 x6.5)
Inside, control Not specified Not specified Not specified NAOutside, overall 0.9 x 0.7 x 2 (3 x 2.3 x 6.6) 1.1 x 1.1 x 2.2 (3.6 x 3.6 x
7.2), optional custom sizes1.1 x 1.1 x 2.3 (3.6 x 3.6 x7.5), optional custom sizes
1.8 x 1.8 x 2.5 (5.8 x 5.8 x8.2)
WINDOWW x H, cm (in) 41 x 75 (16 x 30) 41 x 75 (16 x 30) 41 x 75 (16 x 30) 61 x 76.2 (24 x 30)
WEIGHT, kg (lb) 250 (551.2) 360 (793.8) 460 (1,014.3) 1,814 (4,000)LIST PRICE $2,349 $2,599 $3,999 : $4,799 $15,400WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 2 weeks 2 weeks 2 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS Optional left- or right-hand
swing door; attenuationvalues available for 125-8,000 Hz (octaves). Meets
requirements of ANSI.
Optional left- or right-handswing door; attenuationvalues available for 125-8,000 Hz (octaves). Meets
requirements of ANSI.
Optional left- or right-handswing door; attenuationvalues available for 125-8,000 Hz (octaves). Meets
requirements of ANSI.
Optional varied room sizesavailable; optional radio-frequency shielding. Meetsrequirements of ANSI.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 114/116
Audiometers; Audiometric Booths
114 ©2007 ECRI Institute. All Rights Reserved.
Chart B: Audiometric Booths
MODEL IAC IAC IAC IAC120ACT-1 140ACT-1 160ACT-1 40A-1
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Double wall Single-wall control room,
double-wall examinationroom
Single-wall control room,double-wall examinationroom
Single wall
CONFIGURATION Double room Double room Double room Single roomVIBRATION ISOLATORS Yes Exam room Yes YesJACK PANEL
Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through
10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through
10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through
6, stereo phono; 1, Cinch303; 1, Cinch 304
VENTILATION, m³/min(ft³/min)
5.7 (200), each room 5.7 (200), each room 5.7 (200), each room 2.8 (100)
LIGHT Fluorescent Fluorescent Fluorescent FluorescentNumber 2, each room 2, each room 2, each room 1
AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 6, control room 6, control room 6, control room 1W x D x H, m (ft)
Inside, exam 1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)
1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)
1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)
0.8 x 1.1 x 2 (2.7 x 3.8 x6.5)
Inside, control 1.9 x 1.2 x 2 (6.2 x 3.8 x6.5)
2.3 x 2 x 2.4 (7.5 x 6.7 x7.8)
2.3 x 1.9 x 2 (7.5 x 6.3 x6.5)
NA
Outside, overall 2.5 x 3.9 x 2.5 (8.2 x 12.8 x8.2)
2.5 x 4.6 x 2.5 (8.2 x 15.2 x8.2)
2.5 x 4.7 x 2.5 (8.2 x 15.5 x8.2)1
1 x 1.4 x 2.3 (3.3 x 4.5 x7.6)
WINDOWW x H, cm (in) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30)
WEIGHT, kg (lb) 5,398 (11,900) 4,627 (10,200) 5,080 (11,200) 816 (1,800)LIST PRICE $33,900 $29,100 $35,500 $5,680WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks From stock, 9 modelsFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS Other models available in
the 120ACT series, whichrange in room size; optionalradio-frequency shielding.Meets requirements of
ANSI.
Other models available in
the 140ACT series, whichrange in room size; optionalradio-frequency shielding.Meets requirements of
ANSI.
Other models available in
the 160ACT series, whichrange in room size; optionalradio-frequency shielding.Meets requirements of
ANSI.
Other models available in
the 40A series, which rangein room size; optional radio-frequency shielding. Meetsrequirements of ANSI.
LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes 1The 2 rooms have differentheights; specifications arefor the larger room.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 115/116
Audiometers; Audiometric Booths
©2007 ECRI Institute. All Rights Reserved 115
Chart B: Audiometric Booths
MODEL IAC IAC NAGASHIMA SIBELMED40ACT-1 Mini Series 250 SN-4A S40
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Triple wall Single wallCONFIGURATION Double room Single room Single room Single roomVIBRATION ISOLATORS Yes Optional Yes YesJACK PANEL
Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through
6, stereo phono 4, 3c30 3-pin plug; 1, 3-pinconnector; 1, 10-pin spare;1, relay cord
6, stereo phono
VENTILATION, m³/min(ft³/min)
5.7 (200), each room Not specified Not specified Not specified
LIGHT Fluorescent Fluorescent Incandescent FluorescentNumber 2, each room 1 1 (tungsten), includes
replacement1
AUDIOMETER SHELF Optional Yes Optional YesCARPETING Yes Yes Yes YesELECTRICAL OUTLET 6, control room None 1 1W x D x H, m (ft)
Inside, exam 1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)
0.9 x 0.6 x 1.7 (2.9 x 2 x5.5)
0.7 x 0.7 x 1.8 (2.3 x 2.3 x5.9)
Not specified
Inside, control 1.9 x 1.2 x 2 (6.2 x 3.8 x6.5)
NA NA NA
Outside, overall 2.1 x 3.6 x 2.3 (6.8 x 11.7 x7.6)
1 x 0.7 x 1.9 (3.3 x 2.3 x6.2)
0.8 x 0.8 x 1.9 (2.6 x 2.6 x6.2)
0.9 x 0.9 x 2.15 (3 x 3 x 7.1)
WINDOWW x H, cm (in) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 53 x 53 (20.9 x 20.9) 45 x 55 (17.7 x 21.7)
WEIGHT, kg (lb) 2,767 (6,100) 295 (650) 290 (639) 265 (584.3)LIST PRICE $21,300 $3,250 Not specified $2,275WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks, standard models From stock Not specified Not specifiedFISCAL YEAR January to December January to December Not specified January to DecemberOTHER SPECIFICATIONS Other models available in
the 40ACT series, which
range in room size; optionalradio-frequency shielding.Meets requirements of
ANSI.
Casters; preassembled;optional intercom system.
Meets requirements of ANSI.
Radio-frequency shielded;attenuation values available
for 125-8,000 Hz (octaves).
Right- or left-hand swingdoor; attenuation values
available for 125-8,000 Hz(octaves). Meetsrequirements of ANSI.
LAST UPDATED January 2007 January 2007 April 2001 January 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Audiometers audiometric booths.PDF
http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 116/116
Audiometers; Audiometric Booths
Chart B: Audiometric Booths
MODEL TREMETRICS AR 901
WHERE MARKETED WorldwideCE MARK (MDD) Not specifiedTYPE Not specifiedCONFIGURATION Single roomVIBRATION ISOLATORS OptionalJACK PANEL
Number, type 6, stereo phonoVENTILATION, m³/min(ft³/min)
Not specified
LIGHT Not specifiedNumber Not specified
AUDIOMETER SHELF 16 x 27 in, foldingCARPETING YesELECTRICAL OUTLET YesW x D x H, m (ft)
Inside, exam 0.9 x 0.7 x 1.7 (3 x 2.3 x5.6)
Inside, control Not specifiedOutside, overall 1 x 0.8 x 2 (3.3 x 2.6 x 6.6)
WINDOWW x H, cm (in) 35.6 x 152.4 (17 x 60)
WEIGHT, kg (lb) 347.7 (765)LIST PRICE $3,275WARRANTY 1 yearDELIVERY TIME, ARO Not specifiedFISCAL YEAR January to DecemberOTHER SPECIFICATIONS Portable; casters included;
right- or left-hand doorswing.
LAST UPDATED July 2005Supplier Footnotes
Model Footnotes
Data Footnotes