+ All Categories
Transcript
Page 1: BIOCREDIT CoviFlu Ag Duo...SARS-CoV-2 and Influenza virus (type A and type B) antigen test BIOCREDIT CoviFlu Ag Duo Insert the swab specimen and swirl he swab 5~10 times. Remove the

Intended Use

Package

Storage

Specimen Type

Shelf life

Time to result

For the qualitative detection of SARS-CoV-2 and Influenza virus (type A and type B) antigen

20 Tests / kit

2~ 30℃

Nasopharyngeal swab

24 months from the date of manufacture

5~ 8 minutes

BIOCREDIT CoviFlu Ag Duorapid test kit

www.rapigen-inc.com

BIOCREDIT CoviFlu Ag DuoSARS-CoV-2 and Influenza virus (type A and type B) Antigen test» One-sample generates Three-results» Rapid Test time (5~ 8 min)» User friendly Dual Color System (Control line: Red, Test lines: Black)» Simple detection of both SARS-CoV-2 and Influenza virus (type A and type B) antigen

Core Technology Black Gold Particle

Dual ColorSystem

Performance Characteristics

Differentiated Black Gold Conjugate Technology

Black Gold Particle

BIOCREDIT COVID-19 Ag

RT-PCR (after symptoms occur) Total Sensitivity SpecificityPositive Negative

Positive 46 0 46

90.2% 100.0%Negative 5 136 141

Total 51 136 187

BIOCREDIT InfluenzaA&B Ag

RT-PCR (after symptoms occur) Total Sensitivity SpecificityA B Negative

A Positive 52 0 0 52 94.5%

B Positive 0 27 0 27 90.0%

Negative 3 3 80 86100.0%

Total 55 30 80 165

1023_CoviFlu Ag Duo .indd 1 2020. 10. 30. �� 3:23

Page 2: BIOCREDIT CoviFlu Ag Duo...SARS-CoV-2 and Influenza virus (type A and type B) antigen test BIOCREDIT CoviFlu Ag Duo Insert the swab specimen and swirl he swab 5~10 times. Remove the

SARS-CoV-2 and Influenza virus (type A and type B) antigen test BIOCREDIT CoviFlu Ag Duo www.rapigen-inc.com

➎ Insert the swab specimen and swirl he swab 5~10 times.

➏ Remove the swab while gently squeezing the head of the swab.

➐ Close the assay diluent tube with a filter cap securely.

➊ Tilt the patient's head slightly backwards

➋ Insert a nasopharyngeal swab horizontally into the nasal cavity until resistance is met at the level of the turbinate.

➌ Rotate the swab gently against the nasopharyngeal mucosa for 10 – 15 seconds.

➍ Remove the swab while making sure that the tip of the swab is wet.

Invert the assay diluent tube and gently squeeze it to draw 3~4 drops (90~150㎕) into a sample well on the device.

Read the result within 5-8 minutes.

One red line "C" within the result window.

•COVID-19 positive: Two bands appear; one red control line (C) and one black test line (T).

•Influenza A positive: Two bands appear; one red control line (C) and one black test line(A)

•Influenza B positive: Two bands appear; one red control line (C) and one black test line (B)

•Mixed infection: Red bands on the control line (C) and black bands on the multiple test lines (A,B and/or T).

No "C" line within the result window. It is recommended that the

specimen be retested.

NEGATIVE POSITIVE INVALID

X 5-10

3-4F, 16, LS-ro 91beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do 14119, Republic of Korea

Tel +82-31-427-4677 | Fax +82-31-427-4678 | Email [email protected]

PROCEDURESpecimen collection Analysis

INTERPRETATION

SARS-CoV-2 and Influenza virus (type A and type B) antigen Rapid Diagnostic Test

Read result within

5-8 minCOVID-19 Ag

Influenza A&B Ag

90°

90°

1023_CoviFlu Ag Duo .indd 2 2020. 10. 30. �� 3:23