7/30/2019 DTM User Manual
1/58
E086-69-880
Issue B
Manor Royal, Crawley, West Sussex, RH10 9LW, UK
Telephone: +44 (0) 1293 528844 Fax: +44 (0) 1293 533453
http://www.bocedwards.com
FTM7 Film Thickness Monitor
Description Item Number
FTM7 Film Thickness Monitor E086-69-000
Instruction Manual
7/30/2019 DTM User Manual
2/58
7/30/2019 DTM User Manual
3/58
CONTENTS
Section Title Page
1 INTRODUCTION 11.1 Scope and definitions 1
1.2 Description 2
1.3 Controls and displays 21.4 Rear panel 4
1.5 Principle of operation 41.5.1 Introduction 4
1.5.2 Deposition/etch parameters 51.5.3 Principle of measurement 6
1.5.4 Single crystal sensor/shutter operation 71.5.5 Two crystal sensors/shutters operation 7
1.5.6 Operating modes 71.5.7 Layer sequence control 8
1.6 Typical installation 8
2 TECHNICAL DATA 10
2.1 General 102.2 Performance 10
2.3 Electrical data 122.4 Electrical connectors 12
2.5 RS232 interface 12
3 INSTALLATION 13
3.1 Unpack and inspect 133.2 Configure the electrical supply voltage 133.3 Locate the FTM7 14
3.4 Make the electrical connections 163.5 Test the installation 18
4 OPERATION 194.1 Introduction 19
4.2 Switch on the FTM7 194.3 Data display and entry 19
4.3.1 Select data for display 20
4.3.2 General data entry procedure 204.3.3 Enter layer data 214.3.4 Select backup or layer select mode 22
4.3.5 Configure the analogue output 224.4 Calculate parameter data 22
4.4.1 Density calibration 224.4.2 Calculate the tooling factor 23
4.4.3 Determine the acoustic impedance 26
4.5 Deposition/etch processes 264.5.1 Prepare and enter data 26
4.5.2 Start/stop deposition/etch 27
FTM7 Film Thickness Monitor i
Ipsitech
8068-
03
7/30/2019 DTM User Manual
4/58
Section Title Page
4.6 Failures during deposition/etch 274.6.1 Crystal sensor failures 27
4.6.2 Electrical supply failures 284.7 Switch off the FTM7 28
5 MAINTENANCE 29
5.1 Check the electrical connections 29
5.2 Run the test mode 295.3 Fault finding 30
5.3.1 General fault finding 305.3.2 Display fault messages 33
6 STORAGE AND DISPOSAL 34
6.1 Storage 34
6.2 Disposal 34
7 SERVICE AND ACCESSORIES 35
7.1 Introduction 357.2 Service 35
7.3 Accessories 35
APPENDIX A1 - MEASUREMENT THEORY 37
A1.1 Crystal sensor operation 37A1.2 Measurement in first generation instruments 37
A1.3 Measurement in second generation instruments 38A1.4 Measurement in the FTM7 38
APPENDIX A2 - ANALOGUE OUTPUT CONFIGURATION 39
APPENDIX A3 - RS232 INTERFACE 41
A3.1 Connector pin functions 41A3.2 Communications protocol 41
A3.3 Message protocol 42A3.4 Command messages 42
A3.4.1 Introduction 42
A3.4.2 Check the analogue output mode 43A3.4.3 Check the current layer data 44
A3.4.4 Check the deposition/etch rate 44A3.4.5 Check the shutter status 44
A3.4.6 Check the current thickness 45A3.4.7 Check the FTM7 software version 46
A3.4.8 Check the current crystal sensor 46A3.4.9 Disable the FTM7 keyboard 46
A3.4.10 Select the analogue output mode 46A3.4.11 Program layer parameters 47
A3.4.12 Reset the FTM7 47A3.4.13 Select layer (material) 48
A3.4.14 Close the shutter 48
ii FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
5/58
Section Title Page
A3.4.15 Change terminal thickness 48A3.4.16 Open the shutter 48
A3.5 Return messages 48A3.6 Programming example 50
A3.6.1 RS232 connections 50A3.6.2 Sample program 50
RETURN OF BOC EDWARDS EQUIPMENT
Illustrations
Figure Title Page
1 Front panel controls and indicators 3
2 Rear panel 53 Typical installation 9
4 Dimensions (mm) 115 Fit the FTM7 in a panel or rack 15
A3-1 QuickBASIC programming example 51
Tables
Table Title Page
1 Sequence layers 82 Default values 10
3 Checklist of components 134 Bulk densities and acoustic impedances of common deposition materials 24
5 Parameter data for a typical two-layer deposition26
6 Fault finding 307 Display fault messages 33
A2-1 Analogue output modes 40A3-1 Pins in the RS232 connector mating-half 41
A3-2 Commands 43A3-3 RS232 layer data 45A3-4 Return codes 49
A3-5 RS232 connections to an IBM PC/XT or PC/AT 50
FTM7 Film Thickness Monitor iii
7/30/2019 DTM User Manual
6/58
(This page deliberately left blank)
iv FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
7/58
1 INTRODUCTION
1.1 Scope and definitions
This manual provides installation, operation and maintenance instructions for the BOC
Edwards FTM7 Film Thickness Monitor (abbreviatedto FTM7 in theremainder of this manual).
You must use the FTM7 as specified in this manual. Read this manual before you install andoperate the FTM7.
Important safety information is highlighted as WARNING and CAUTION instructions; youmust obey these instructions. The use of WARNINGS and CAUTIONS is defined below.
WARNING
Warnings are given where failure to observe the instruction could result in injury or deathto people.
CAUTION
Cautions are given where failure to observe the instruction could result in damage to theequipment, associated equipment and process.
In accordance with the requirements of IEC1010, the following symbols appear on the FTM7:
Caution - refer to accompanying documents
Caution - risk of electric shock
Protective conductor terminal
The units used throughout this manual conform to the SI international system of units of
measurement.
FTM7 Film Thickness Monitor 1
7/30/2019 DTM User Manual
8/58
1.2 Description
The FTM7 is designed to measure and display the deposition rate and thickness of films
deposited in vacuum coating systems. You can also use the FTM7 to monitor material removal
during etch processes.
You must use the FTM7 in conjunction with one or two water cooled crystal holders, quartzcrystals, and oscillator units. Note that a crystal holder, crystal and oscillator unit are together
referred to as a crystal sensor throughout this manual.
The FTM7 uses the crystal microbalance technique to measure film thickness and deposition
rate. Film thickness and deposition rate are calculated from your data on the density and theacoustic impedance of the film material.
Data for up to 11 layers (materials) can be stored in non-volatile memory in the FTM7, and youcan specify 'Tooling Factors' to compensate for different detector/substrate geometries so that
the FTM7 may be used on a number of different deposition systems.
You can configure the FTM7 to automatically stop deposition at a specified terminal film
thickness. The FTM7 will close a shutter between the deposition source and the substrate onwhich the film is being deposited. Two shutters can be controlled by the FTM7.
The FTM7 can show both positive and negative frequency changes. You can, therefore, observeflash-heatingeffects in thecrystal sensor, or monitor materialsbeing removed from a pre-loaded
crystal. TheFTM7 canthereforebe used to monitor both thedepositionandetching of thin films.
Without the ability to show negative frequency shifts, you would be unaware of crystal sensorwarm-up or flash heating effects which can cause large errors during film monitoring. To retain
the same number of displayed significant digits when measuring negative frequency changes,the negative sign is suppressed once thickness loss is greater than 99.9 nm.
If required, you can connect your control equipment (for example,a PC) to the FTM7 through anRS232 communications link, and use your control equipment to control and monitor the
operation of the FTM7.
1.3 Controls and displays
Refer to Figure 1. The controls and displays on the FTM7 are used as follows:
Data/Rate
display (1)
This is a seven-segment, four-digit LED display which shows deposition
rate, material or FTM7 status information.
Thickness display
(2)
This is a seven-segment, four-digit LED display which shows a constantly
updated reading of deposit thickness, or shows the terminal film thickness.The display is auto-ranging and the frequency of display updates is once a
second.
Units LEDs (3) One of these LEDs will be on to indicate the current thickness units.
Status LEDs (11) One of these LEDs will be on to indicate the parameter currently shown onthe Data/Rate and Thickness displays: see See Section 4.3.
2 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
9/58
Sensor LEDs (12) One of these LEDs will be on to indicate the current crystal sensor.
Decrement andIncrement
buttons (9, 10)
Use these buttons to decrement or increment the value of the currentparameter on the Data/Rate display, or the terminal thickness on the
Thickness display: see Section 4.3.
Data button (6) Use thisto selecttheparameter tobedisplayedormodified:see Section 4.3.
Test button (7) Use this button to select or deselect the test mode: see Section 5.2.
Test LED (8) This LED flashes when test mode is selected: see Section 5.2.
Run button (5) Use thisbutton to manuallyopen and closethe shutter. Itwillonlyoperate
when a correctly operating crystal sensor is connected to the FTM7.
Shutter status
LEDs (4)
One of these LEDs will be on to identify the status of the shutter (open or
closed); the LED(s) will flash to indicate a fault: refer to Section 5.3.
FTM7 Film Thickness Monitor 3
Figure 1 - Front panel controls and indicators
1. Data/Rate display
2. Thickness display
3. Units LEDs
4. Shutter status LEDs
5. Run button
6. Data button
7. Test button
8. Test LED
9. Decrement button
10. Increment button
11. Status LEDs
12. Sensor LEDs
7/30/2019 DTM User Manual
10/58
1.4 Rear panel
Refer toFigure2 which shows the rearpanel oftheFTM7. Items onthe rearpanel are asfollows:
Earth (ground)stud (1)
Use this to earth (ground) the FTM7: refer to Section 3.4.
Voltageselector/fuse
holder (2)
Use this to selectthecorrect voltage foryour electrical supply: refer toSection 3.2.
On/Off switch (4) Use this to switch the FTM7 on and off.
Shutters terminalblock (5)
Use these to connect one or two shutters to the FTM7: refer toSection 3.4.
RS232 connector(6)
If required, use this to connect the FTM7 to your control equipment(for example, a PC): refer to Section 3.4.
Analogue outputconnector (7)
If required, use this to connect an analogue device such as a chartrecorderto theFTM7. Youcan selectwhatparameteris outputon this
connector: see Section 4.3.
Sensor 1 and 2
connectors (9, 8)
Use these to connect primary and secondary crystal sensors to the
FTM7.
1.5 Principle of operation
1.5.1 Introduction
The FTM7 is a microprocessor-based frequency counter which converts frequency changes into
deposition rates and thickness information for a range of deposition materials. It is used inconjunction with a quartz crystal which is placed in the deposition field, and an oscillator unit
whose output frequency is controlled by the crystal.
The FTM7 samples the output of the oscillator unit and detects frequency changes caused by
material being deposited on the face of the crystal. The FTM7 uses the change in frequency to
calculate the rate and thickness of the material being deposited. The frequency data is modifiedaccording to any data you enter which relates to the type of material being deposited, and the
geometric relationship between the deposition source, the target substrate and the crystalsensor.
UsetheRunbuttononthefrontoftheFTM7toopentheshutterandstartadepositionprocess.
You can then use the Run button to close the shutter, to shield the substrate from further
deposition. Alternatively,youcan program theFTM7 toautomatically close theshutterandstopthe deposition process when a user specified thickness is reached.
TheparameterswhichyoucanmonitorandchangeintheFTM7aredescribedinSection1.5.2.
4 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
11/58
1.5.2 Deposition/etch parameters
Layer You can store deposition film data for up to 11 different layers/materials
(layers 1 to 11) into the FTM7. (Note that when you program the FTM7,layers 12 to 19 are used to select layer sequences of deposition: see
Sections 4.3.3 to 4.3.5.)
Density This is the density of materials you intend to evaporate (in g cm-3).
z-value This is theacoustic impedanceof thedepositionmaterial (in105 cm2 g-1s-1).
FTM7 Film Thickness Monitor 5
Figure 2 - Rear panel
1. Earth (ground) stud
2. Voltage selector / fuse holder3. Electrical supply plug
4. On/Off switch
5. Shutters terminal block
6. RS232 connector
7. Analogue output connector8. Sensor 2 connector
9. Sensor 1 connector
7/30/2019 DTM User Manual
12/58
Terminate The terminal thickness of the film required. On reaching this thickness the
FTM7 can automatically close the shutter to prevent further deposition.
Tooling Tooling factor: each layer can have an associated tooling factor, which
allows you to calibrate the FTM7 for differences between source-substrateand source-sensor distance and attitude (if the sensor is further from the
source than the substrate, it will receive a thinner deposit than thesubstrate). You can calculate the tooling factor assuming a square law
variation of thickness with distance, and a cosine distribution with angle.
However, we recommend that you calculate the required tooling factor byexperimental deposition processes. Note that:
A tooling factor of 1.00 implies that the sensor and substrate receive thesame deposit thickness.
A tooling factor> 1.00 implies that thesensorreceives a thinner deposit
than the substrate. A tooling factor < 1.00 implies that the sensor is more heavily loaded
than the substrate.
xtal In layer-selected mode (see Section 1.5.6), this parameter (1 or 2) determines
which crystal sensor is used to monitor the currently selected layer.
Usage This represents the difference between the measured crystal sensor
frequency and the 6.0 MHz frequency of a new, perfect crystal sensor. Youcannot change this value.
In practice, a new crystal sensor, when installed, will usually have an
oscillating frequency of approximately 5.995 MHz. The starting 'usage'value for such a crystal sensor will be approximately 5.0 kHz (as shown intheData/Ratedisplay). As thecrystal sensor is used to measure deposition
layers and it becomes coated with deposition material, its frequencydecreases and the usage value increases.
The FTM7 processor assumes that the crystal sensor is defective if thefrequencyfallsbelow5.1Mhz(whichwillgiveausagevalueof900.0kHz).
Note: Some brittle deposition materials can cause an interface fracture between the crystal and the
deposited layer, at a much lower 'usage' value. This will cause immediate crystal sensor failure.
We therefore recommend that you determine acceptable limits of 'usage' by experiment for anygiven deposition material.
1.5.3 Principle of measurement
Complex equations for the conversion of frequency change into deposition rate and thicknessdata are applied by the FTM7 to the detected output of the oscillator unit. These equations take
account of thedensity and acoustic impedance of thecrystal sensorand thedeposition material.Refer to Appendix 1 which describes the equations used and how they were developed.
6 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
13/58
1.5.4 Single crystal sensor/shutter operation
Whenyou connect a single crystal sensorand shutter to the FTM7, you can use the Run button to
open the shutter and start the deposition process.
You can use the Run button again tomanually stopdeposition, oryou can configure the FTM7 to
automatically stop deposition at thespecified termination thickness. At this thickness, theFTM7will automatically close the shutter to stop the deposition process.
If the crystal sensor fails, the FTM7 automatically closes the shutter, however the thickness dataat the time of the failure is stored in memory. You can therefore replace the failed crystal sensor,
then continue the deposition/etch process, without loss of thickness data.
1.5.5 Two crystal sensors/shutters operation
When you connect two crystal sensors and two shutters to the FTM7, you must configure thesensors and shutters so that:
Shutter1isbetweenthedepositionsourceandboththetargetsubstrateandcrystalsensor1.
Shutter 2 is between the deposition source and crystal sensor 2.
1.5.6 Operating modes
You can operate the FTM7 in one of two modes: backup mode and layer selected mode, as
described below.
Backup mode
In this mode, shutter 1 is used to control deposition/etch (either by manual operation, orautomatically at therequiredterminal thickness). If crystal sensor1 fails, shutter 2 will open and
the FTM7 will automatically use crystal sensor 2 to determine thickness data; deposition willcontinue without interruption.
Layer-selected mode
In this mode, the appropriate shutter is opened/closed depending on which crystal sensor is
assigned to the layer:
If crystal sensor 1 is assigned, shutter 1 opens/closes. If crystal sensor 2 is assigned, both shutters open/close.
Note that there is no back-up in this mode. If either crystal sensor fails, shutter 1 is automatically
closed to protect the target substrate.
FTM7 Film Thickness Monitor 7
7/30/2019 DTM User Manual
14/58
1.5.7 Sequence layer control
Youcan request theFTM7 to perform a predefined sequenceof up tofive layer depositions: to do
this, select layer sequences identified by two numbers '12 1' to '19 6', as shown in Table 1; in
this Table, the first two characters identify the sequence selected, and the last character specifiesthe first layer which will be used in the sequence.
Layer Sequence of layers
deposited
12 1
13 114 1
15 116 3
17 518 7
19 6
1, 2
1, 2, 31, 2, 3, 4
1, 2, 3, 4, 53, 4
5, 67, 8
6, 7, 8
Table 1 - Sequence layers
For example, if you select layer '19 6':
When you initially press the Run button, the FTM7 will open the shutter and use the dataentered for layer 6 to measure the deposition rate and thickness.
When the terminal thickness for layer 6 is reached, or if you press the Run button again, theshutter is closed.
If you then press the Run button again, the FTM7 will open the shutter and use the dataentered for layer 7 to measure the deposition rate and thickness.
When the terminal thickness for layer 7 is reached, or if you press the Run button again, theshutter is closed.
If you then press the Run button again, the FTM7 will open the shutter and use the dataentered for layer 8 to measure the deposition rate and thickness.
When the terminal thickness for layer 8 is reached, or if you press the Run button again, theshutter is closed.
IfyouthenpresstheRunbuttonagain,thesequenceisrepeatedfortheselectedlayer196.
1.6 Typical installation
This manual assumes that you will use the FTM7 in conjunction with the following accessories,as shown in Figure 3:
Water cooled crystal holder. Electromagnetic shutter actuator *.
Oscillator unit. Electromagnetic shutter controller *.
Accessories suitable for use with the FTM7 are listed in Section 7.3.
8 FTM7 Film Thickness Monitor
* Optional.
7/30/2019 DTM User Manual
15/58
FTM7 Film Thickness Monitor 9
Figure 3 - Typical installation
1. Crystal (in holder)
2. Earth (ground) wire
3. On/Off switch
4. FTM7
5. Shutter 2 connections (if fitted)
6. Control/monitoring equipment (optional)
7. Analogue device (for example, chart recorder)
8. Shutter controller panel
9. Cooling-water connections
10. Short coaxial cable
11. Shutter actuator
12. Source
13. Shutter
14. Bell jar
15. Substrate
16. Oscillator unit
17. Long coaxial cable
18. Long coaxial cable to oscillator unit 2 (if fitted)
7/30/2019 DTM User Manual
16/58
2 TECHNICAL DATA
2.1 General
Dimensions See Figure 4
Mass 2.8 kg
Operating temperature range 5 to 40 oC
Maximum operating humidity 80% up to 31 oC, 50% up to 40 oC
Maximum operating altitude 2000 m
IEC1010 installation category II
IEC1010 pollution category I
Degree of protection IP20
Operating environment Suitable for indoor use only
2.2 Performance
Material layer memories 11
Crystal sensor frequency 6 MHz unloaded
Thickness measurement range 0.0 nm to 999.9 m, auto-rangingwith LED indication of the scale factor(nm or m)
Thickness resolution 0.1 nm
Rate of deposition measurement range 0.00 to 999.9 nm s-1
Resolution of deposition measurement range 0.1 nm s-1
Thickness sampling frequency 1 Hz
Density 0.01 to 99.99 g cm-3
Acoustic impedance 1.0 to 99.9 x 105 g cm-2 s-1
Terminal thickness measurement range 0.1 nm to 99.99 m with LEDindication of scale factor (nm or m)
Tooling factor range 0.01 to 99.9
Default values See Table 2
Parameter Default value
Density 1.00
Acoustic impedance (z-value) 8.834 (g cm-2 s-1)
Terminal thickness 990 (nm)
Tooling factor 1.00
Table 2 - Default values
10 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
17/58
FTM7 Film Thickness Monitor 11
Figure 4 - Dimensions (mm)
A Front view
B Side view
1. Hinged leg
2. Rubber foot
3. Clearance required for electrical supply cable
7/30/2019 DTM User Manual
18/58
2.3 Electrical data
Electrical supply voltage 100/120 V or 220/240 V selectable
Electrical supply voltage tolerance 10%
Electrical supply frequency 50/60 Hz
Power consumption 10 W
Electrical supply fuse ratings
220/240 V 100 mA anti-surge
100/120 V 100 mA anti-surge
Analogue output
Voltage 0 to 1 V, 8-bit resolution
Maximum load 1 k
Shutter relay switch
Maximum rating 50 mA at 240 V a.c. or 200 V d.c.
Maximum switching power 10 W
Sensor head output voltage 10 V
2.4 Electrical connectors
Electrical supply connector IEC plug
Analogue output connector BNC socket
Sensor connectors BNC socket
RS232 connector fitted to FTM7 9-way D-type socket, female
Shutter terminals Push-in, self-lock
2.5 RS232 interface
Transfer rate 4800 baud
Protocol 1 start bit, 7 data bits, even parity, 1 stop bit
12 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
19/58
3 INSTALLATION
3.1 Unpack and inspect
Remove all packing materials and protective covers and check the FTM7. If the FTM7 is
damaged, notify your supplier and the carrier in writing within three days; state the Item
Numberof theequipmenttogetherwith your order numberand your suppliersinvoice number.Retain all packing materials for inspection. Do not use the FTM7 if it is damaged.
Check that your package contains the items listed in Table 3. If any item is missing, notify yoursupplier in writing within three days.
If the FTM7 is not to be used immediately, replace the protective covers. Store it in suitableconditions, as described in Section 6.
Qty Description Check ()
1 FTM7 Film Thickness Monitor
1 Electrical supply cable
1 Mounting-plate
4 Screws
2 Spare fuses
Table 3 - Checklist of components
3.2 Configure the electrical supply voltage
1. Refer to Figure 2. Look at the voltage selector/fuse holder (2):
If the voltage indicated by the white mark is correct for your electrical supply, continueat Section 3.3.
If the voltage indicated by the white mark is not correct for your electrical supply, con-tinue at Step 2 to reconfigure the FTM7.
2. Remove the voltage selector/fuse holder (2) from the FTM7.
3. Turn the voltage selector/fuse holderaround, so that the correct voltage will be indicated bythe white mark, then refit the voltage selector/fuse holder to the FTM7.
FTM7 Film Thickness Monitor 13
7/30/2019 DTM User Manual
20/58
3.3 Locate the FTM7
WARNING
Do not place containers of liquids on or near the FTM7. Liquids spilled into the FTM7 cancause electrical short circuits and the risk of injury by electric shock. Solvents or corrosive
liquids spilled into or onto the FTM7 can damage it.
Notes: If you want to fit the FTM7 in a rack or panel, we recommend that you make all of the electrical
connections to the rear of the FTM7 (see Section 3.4) before you secure the FTM7 in the rack or
panel. We also recommend that you leave the On/Off switch in the 'on' position, and use the rack
isolator switch to switch the FTM7 on and off.
Refer to Figure 4. The FTM7 has hinged legs (1)and rubber feet (2) and can be free-standing on a
table or bench, can be fitted in the AUTO 306, or can be fitted in a rack or panel.
If the FTM7 will be free-standing on a table or bench, if required swing down the hinged legs (1)and place the FTM7 in its required operating position.
To fit the FTM7 in the AUTO 306, refer to the AUTO 306 instruction manual.
To fit the FTM7 in a rack or panel; use the following procedure:
1. Refer toFigure 5. Place the FTM7(3)againsttherearof the mounting-plate(5), sothat the lipon the front of the FTM7 fits into the slot on the rear of the mounting-plate, as shown in
detail B.
2. Secure the mounting-plate to the FTM7:
Fit two of the screwssupplied (1) through the bracket (7) at the top rear of the mounting-plate and into the FTM7.
Fit the other two screws supplied (4) through the bracket at the bottom rear of themounting-plate and into the FTM7.
3. Use suitablescrews(6) through the fixing holes onthe mounting-plate(5) tosecuretheFTM7
and the mounting-plate in the rack or panel (2).
14 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
21/58
FTM7 Film Thickness Monitor 15
Figure 5 - Fit the FTM7 in a panel or rack
1. Screws (2 off)
2. Panel
3. FTM7
4. Screws (2 off)
5. Mounting-plate
6. Screws (4 off)
7. Bracket
8. Lip
9. Slot
7/30/2019 DTM User Manual
22/58
3.4 Make the electrical connections
WARNING
Route all cables and wires away from hot or cold surfaces. If you do not, the cables orwires may be damaged, and there may be a risk of injury by electric shock, or of damage to
the FTM7.
CAUTION
If you install the FTM7 inside an Auto 306 cabinet or other control cabinet, a 0 to 250 Vsignal may be applied to the shutter connector (5).
If you use the FTM7 mounted outside a cabinet, that is on a bench, for your own safety you
must not apply more than 24 V to the shutter connector.
Note: Whenyou install the crystal sensor(s) and shutter(s), ensure that theyare correctly configured: if
you only install a single crystal sensor, connect it to sensor 1 connector; if you only install a
single shutter, connect it to the shutter 1 terminals on the shutter terminal-block.
Use the following procedure to make the electrical connections. Refer toFigure 3 which shows aschematic diagram of the installation. The connectors on the rear of the FTM7 are shown in
Figure 2.
1. Connect a suitable earth (ground) wire (2) between the FTM7 earth (ground) stud and the
baseplate earth (ground) point. Note that if you have fitted the FTM7 in the AUTO 306, youwill have disconnected an earth (ground) wire from the blank panel removed from the
AUTO 306 control cabinet: connect this wire to the FTM7 earth (ground) stud.
2. Fit the two wires from the shutter controller panel (8) tothe shutter 1 terminals on the FTM7
shutter terminal block; to fit each wire:
Gently push a small screwdriver into the upper square slot of the terminal block, thenpush the end of the wire into the lower circular hole in the terminal block.
Ensure that no bare wire is visible outside the terminal block, then remove the screw-driver to secure the wire in place.
3. If necessary, fit the two wires (5)fromthe secondshutter controller to the shutter2 terminalson the FTM7 shutter terminal block: use the method in Step 2.
4. Connect the short coaxial cable (10) from the connector on the end of the oscillator unit (16)marked 'XTAL' to the crystal (1).
5. Connect the long coaxial cable (17) from the connector on the end of the oscillator unit (16)
marked 'FTM' to the sensor 1 connector on the FTM7.
6. If required, connect the second crystal sensor: use the method in Steps 4 and 5, but connect
the long coaxial cable (18) to the sensor 2 connector on the FTM7.
16 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
23/58
7. If required, connect a suitable analogue device (7) to the analogue output connector on the
FTM7.
8. If required, connect the RS232 cable (supplied as an optional extra) to the mating socket
(female) on the FTM7. Connect the other end of the cable to your control monitoringequipment (6). Refer to Appendix A3 for details of operation using the RS232 interface.
9. Fit the IEC socketon the end ofthe electrical supplycableto the electricalsupply plug onthe
rear of the FTM7 (Figure 2, item 3).
10. Connecttheother end ofthe cable totheelectricalsupply; use one ofthe following methods:
Fit a suitable three-pin fused plug to the end of the cable, and fit the plug to an electricalsupply socket.
Connect the wires in the cable to a suitable electrical supply, as follows:
Connect the brown wire to the live electrical supply.
Connect the blue wire to the neutral electrical supply.
Connect the green/yellow wire to earth (ground).
To connect the FTM7 electrical supply cable to an AUTO 306 system, refer to the AUTO306 instruction manual for the location of terminal block TB1 and:
Connect the brown wire in the cable to terminal 7 on terminal block TB1.
Connect the blue wire in the cable to terminal 14 on terminal block TB1.
Connect the green/yellow wire in the cable to an earth (ground) terminal on termi-nal block TB1.
FTM7 Film Thickness Monitor 17
7/30/2019 DTM User Manual
24/58
3.5 Test the installation
Note: Werecommend thatyou takenote ofthe version ofsoftware installed inthe FTM7; you may need
to supply this information if a fault develops in the FTM7: see Section 5.3.2.
1. Refer to Figure 2. Move the On/Off switch (4) to the 'on' position (position '1').
Alternatively, if you have fitted the FTM7 in a rack, switch on the rack isolator to turn on theFTM7.
2. The Data/Rate display should then show F7, and the Thickness display should then show
the version of the software installed in the FTM7. If the displays are incorrect, refer to
Section 5.3 for fault finding.
3. Refer to Figure 1. Check that the Data/Rate and Thickness displays (1, 2) then change toshow values close to zero, typically in the form:
0.0 -0.1
If the displays are incorrect, refer to Section 5.3 for fault finding.
4. Check thatthextal1 LED (1) ison. IftheLED isnoton, refer toSection 5.3 for fault finding.
5. Select test mode: refer to Section 5.2.
6. Iftest modewas successful(thatis,no faultswerefound),if you donotwantto use the FTM7
immediately, switch off the On/Off switch on the rear of the FTM7 (Figure 2, item 4), orswitch off the rack isolator if you have fitted the FTM7 in a rack. Alternatively, continue at
Section 4 to operate the FTM7.
18 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
25/58
4 OPERATION
WARNING
Do not place containers of liquids on or near the FTM7. Liquids spilled into the FTM7 cancause electrical short circuits and the risk of injury by electric shock. Solvents or corrosive
liquids spilled into or onto the FTM7 can damage it.
Note: The following sections describe manual operation and monitoring of the FTM7. Refer to
Appendix A3 if you will use your own control equipment (connected through the RS232
connector) to control and monitor the operation of the FTM7.
4.1 Introduction
If you require accurate and repeatable deposition results, you must enter calibrated parameter
data into the FTM7; the parameters you can enter into the FTM7 are listed in Section 1.5.2.
Section 4.2 describes how to switch on the FTM7. Section 4.3 describes how to display and
change data. Section 4.4 describes how to calculate parameter data, and Section 4.5 describeshow to control deposition processes.
For less accurate deposition processes, you can go directly to Section 4.5 and ignore theinstructions to calculate and enter parameter data. If you do, the FTM7 will use any previously
entered parameter data, or use the default data (see Table 2).
4.2 Switch on the FTM7
Move the On/Off switch on the rear of the FTM7 (Figure 2, item 4) to the 'on' position
(position '1').
Alternatively, switch on the rack isolator if you have fitted the FTM7 in a rack.
4.3 Data display and entry
Note: Data entered into the FTM7is storedin a non-volatile memory. When the FTM7 isswitched off,
the data will be retained in the memory for several years. When you switch on the FTM7, if the
memory is faulty, error E3 will be displayed and default values will be used: see Table 2.
FTM7 Film Thickness Monitor 19
7/30/2019 DTM User Manual
26/58
4.3.1 Select data for display
RefertoFigure1. Thestatus LEDs (11) identify theparameter currentlyshownon thedisplays.
To select a different parameter for display, press and release the Data button (6); in general:
Each time you press the Data button, the next lower status LED will go on, and the currentvalue of the corresponding parameter will be shown on the displays.
If you press the Data button twice when the 'mode' LED is on, the 'mode' LED goes off, the'rate' LED goes on, and the current rate data is shown on the displays.
If you press and hold in the Data button, the FTM7 will automatically cycle through thedisplay parameters (that is, show one parameter after another). Release the Data buttonwhen the required parameter is displayed.
4.3.2 General data entry procedure
Notes: If you do not press a button for 10 seconds: if the FTM7 is in sequence layer control (see
Section 1.5.7), the display will revert to show the layer display; otherwise the display will revert
to show deposition rate and thickness data.
You cannot enter data if a shutter is open, if a shutter has beenclosed due tocrystal sensor failure,
or if the FTM7 is in power fail recovery (the shutter status LEDs flash) after an electrical supply
failure while the shutter was open: if you press the Increment or Decrement buttons, they will be
ignored in these conditions.
If you press and hold in the Increment or Decrement button, the displayed data will
automatically increment/decrement rapidly. In addition, data values for layer/sequence orcrystal 'scroll'; that is, if the display shows '19 6' during layer/sequence entry (as in
Section 4.3.3) and you press the Increment button, the display will scroll to show '1'. Similarly,
if the display shows' 1' and you press the Decrement button, the display will scroll to show
'19 6'. Other data values increment to their maximum value or decrement to their minimum
value. These facilities allow you to easily make large changes to displayed data.
Specific data entry procedures are described in Sections 4.3.3 to 4.3.5.
Refer to Figure 1 and use the following procedure to enter data into the FTM7:
1. Display the parameter to be changed: refer to Section 4.2.1.
2. Change the displayed data:
Press and release the Decrement button (9) to reduce the displayed data by one in theleast significant (right-hand) digit.
Press and release the Increment button (10) to increase the displayed data by one in theleast significant (right-hand) digit.
20 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
27/58
3. When the display shows the new data to be entered, press the Data button (6) to enter the
data into the FTM7 and to display a different parameter (see Section 4.2.1). If you do nottouch any of the buttons for 10 seconds: the new data will be lost and the display will revert
to show rate and thickness data: see the Note above.
4.3.3 Enter layer data
Note: Toconfigurethe FTM7formanual operation, enter a terminal thickness of0. You must then use
the Run button to close the shutter; the shutter will not close automatically.
1. Refer to Figure 1. Press the Data button (6) until the 'layer' status LED (11) is on.
2. Use the Increment and Decrement buttons (10, 9) to select the required layer
number (1 to 11)orsequencelayernumber (12 1 to19 6)ontheData/Ratedisplay(1).
3. Press and release the Data button (6); the 'density' LED (11) will then go on, and the last
entered or default value of density is shown on the display.
4. Use the Increment and Decrement buttons (10, 9) to change the displayed data to therequired value.
5. Press and release the Data button (6); the 'z-value' LED (11) will then go on, and the last
entered or default z-value will be displayed.
6. Use the Increment and Decrement buttons (10, 9) to change the displayed data to the
required value.
7. Press and release the Data button (6); the 'terminate' LED (11) will then go on, and the lastentered or default termination thickness value will be displayed.
8. Use the Increment and Decrement buttons (10, 9) to change the displayed data to the
required value.
9. Press and release the Data button (6); the 'tooling' LED (11) will then go on, and the last
entered or default tooling factor value will be displayed.
10. Use the Increment and Decrement buttons (10, 9) to change the displayed data to therequired value.
11. Press and release the Data button (6); the 'xtal' LED (11) will then go on, and the last enteredor default crystal sensor value will be displayed.
12. Use the Increment and Decrement buttons (10, 9) to change the displayed data to therequired value (1 or 2, for crystal sensor 1 or 2).
13. Press and release the Data button (6); the 'usage' LED (11) will then go on, and the currently
selected crystal usage factor will be displayed.
14. Press and release the Data button (6) four times; the 'layer' LED (11) will then go on, and thelayer (1to11) orlayer sequence (12 1 to19 6)willbeshown onthe Data/Rate display(1).
The FTM7 is now ready for deposition for the currently displayed layer.
FTM7 Film Thickness Monitor 21
7/30/2019 DTM User Manual
28/58
4.3.4 Select backup or layer select mode
1. Press and release the Data button until the 'mode' LED is on; the last entered or default
crystal sensor mode (1 or 2) will be shown on the display
2. If necessary, use the Increment and Decrement buttons to change the mode:
Select '1' for backup mode.
Select '2' for layer select mode.
3. Press and release the Data button three times; layer information will then be shown on the
display again.
4.3.5 Configure the analogue output
1. Press and release the Databutton until the 'mode'LEDis on; the current crystal sensormode
(1 or 2) will then be shown on the display.
2. Press and release the Data button; the Data/Rate display will then show the last entered ordefault analogue mode (0 to 39).
3. Use the Increment and Decrement buttons to select the required analogue output mode:
refer to Appendix 1.
4. Press and release the Data button to display rate/thickness data again.
4.4 Calculate parameter data
4.4.1 Density calibration
Use multiple beam interferometry or a stylus measuring device to calculate the results of testdepositions. To calibrate the density:
1. Fit a new crystal.
2. Set the density tothe bulk value orestimatedvalue: refer to Section 4.3.3. Notethat the bulkdensities of common deposition materials are given in Table 4.
3. Set the z-value to 8.834 x 105 g cm-2 s-1 (or to your predetermined or estimated value).
4. Set the tooling factor to 1.0 (or to your predetermined or estimated value).
5. Place several test substrates, properly masked for thin film thickness measurement, as closeto the crystal as possible and make a trial deposition onto a new crystal. The thickness of the
deposit should be approximately 80 nm.
6. Take note of the indicated thickness values and then use multiple beam interferometry or a
stylus measuring device to measure the actual thickness on the substrates; calculate theaverage value.
22 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
29/58
7. Use the following equation to calculate the material density:
D DT
Ta e
ind
m
= .
Where :
Da = actual density
De = estimated density (bulk value)
T ind = indicated thickness
Tm = measured thickness
8. Record the actual density and enter it into the FTM7 as described in Section 4.3.3.
4.4.2 Calculate the tooling factor
Usea tooling factorof 1.0andtheknown or previously calibrated density, theknown z-value (ora z-value of 8.834 x 105 g cm-2 s-1 ) :
1. Fit a new crystal.
2. Place several test substrates, properly masked for thin film thickness measurement, as closeto the crystal as possible and make a trial deposition onto the new crystal. The thickness of
the deposit should be approximately 80 nm.
3. Take note of the indicated thickness, measure the thickness of the deposit on each sample,
then calculate the average film thickness.
4. Use the following equation to calculate the new tooling factor:
Tooling factorT
Tm
ind
=
5. Record the actual tooling factor and enter it into the FTM7 as described in Section 4.3.3.
FTM7 Film Thickness Monitor 23
7/30/2019 DTM User Manual
30/58
Material Symbol Density(g cm-3)
Acoustic impedance(x 105 g cm-2 s-1)
Aluminium Al 2.70 8.17
Aluminium oxide Al2O2 3.97 *
Antimony Sb 6.62 11.49Arsenic As 5.73 8.83
Barium Ba 3.50 *
Beryllium Be 1.85 16.25
Bismuth Bi 9.80 *
Bismuth oxide Bi2O2 8.90 *
Boron B 2.54 22.69
Cadmium Cd 8.64 12.94
Cadmium selenide CdSe 5.81 *
Cadmium sulphide CdS 4.83 8.66
Cadmium telluride CdTe 5.85 9.00
Calcium Ca 1.55 *
Calcium fluoride CaF2 3.18 11.39Carbon (graphite) C 2.25 2.71
Cerium (III) fluoride CeF3 6.16 *
Cerium (IV) oxide CeO2 7.13 *
Chromium Cr 7.20 28.94
Chromium (III) oxide Cr2O3 5.21 *
Cobalt Co 8.71 25.73
Copper Cu 8.93 20.20
Copper (I) sulphide (Alpha) Cu2S (Alpha) 5.60 *
Copper (I) sulphide (Beta) Cu2S (Beta) 5.80 *
Copper (II) sulphide CuS 4.60 *
Cryolite Na3AlF6 * *
Erbium Er 9.05 *
Gadolinium Gd 7.89 *
Gallium Ga 5.93 14.88
Gallium arsenide GaAs 5.31 5.55
Germanium Ge 5.35 17.10
Gold Au 19.3 23.17
Hafnium Hf 13.09 *
Hafnium oxide HfO2 9.63 *
Indium In 7.30 10.49
Indium antimonide InSb 5.76 10.98
Indium oxide In2O2 7.18 *
Iridium Ir 22.4 68.40
Iron Fe 7.86 25.29
Lanthanum La 6.17 *
Lanthanum fluoride LaF3 5.94 *
Lanthanum oxide LaO3 6.51 *
Lead Pb 11.3 7.81
Lead sulphide PbS 7.50 15.59
Lithium Li 0.53 *
Lithium fluoride LiF 2.64 11.40
Magnesium Mg 1.72 12.18
Magnesium fluoride MgF2 3.00 *
Magnesium oxide MgO 3.58 21.47
Manganese Mn 7.20 23.41
24 FTM7 Film Thickness Monitor
Table 4 - Bulk densities and acoustic impedances for common deposition materials
7/30/2019 DTM User Manual
31/58
Material Symbol Density(g cm-3)
Acoustic impedance(x 105 g cm-2 s-1)
Manganese (II) sulphide MnS 3.99 *
Mercury Hg 13.46 *
Molybdenum Mo 10.2 34.34Neodymium fluoride NdF3 6.51 *
Neodymium oxide Nd2O3 7.24 *
Nickel Ni 8.91 26.66
Niobium Nb 8.57 17.90
Niobium pentoxide Nb2O3 4.47 *
Palladium Pd 12.0 24.72
Platinum Pt 21.4 36.06
Potassium chloride KCl 1.98 4.30
Rhenium Re 21.04 *
Rhodium Rh 12.41 *
Rubidium Rb 1.53 *
Samarium Sm 7.54 *Scandium Sc 3.00 *
Selenium Se 4.82 10.21
Silicon Si 2.32 12.39
Silicon dioxide (fused quartz) SiO2 2.20 8.25
Silicon monoxide SiO 2.13 *
Silver Ag 10.5 16.68
Silver bromide AgBr 6.47 7.48
Silver chloride AgCl 5.56 6.68
Sodium Na 0.97 *
Sodium chloride NaCl 2.17 5.62
Sulphur S 2.07 *
Tantalum Ta 16.6 33.68
Tantalum (V) oxide Ta2O3 8.20 *
Tellurium Te 6.25 9.80
Terbium Tb 8.27 *
Thallium Tl 11.85 *
Thorium (IV) fluoride ThF4 6.32 *
Tin Sn 7.30 12.19
Titanium Ti 4.50 14.05
Titanium (IV) oxide TiO2 4.26 *
Titanium oxide TiO 4.90 *
Tungsten W 19.3 54.14
Tungsten carbide WC 15.6 58.44
Uranium U 18.7 37.08
Vanadium V 5.96 16.65
Ytterbium Yb 6.95 *
Yttrium Y 4.34 10.57
Yttrium oxide Y2O3 5.01 *
Zinc Zn 7.04 17.17
Zinc oxide ZnO 5.61 15.87
Zinc selenide ZnSe 5.26 12.22
Zinc sulphide ZnS 4.09 *
Zirconium Zr 6.51 *
Zirconium oxide ZrO2 5.60 *
FTM7 Film Thickness Monitor 25
Table 4 - Bulk densities and acoustic impedances for common deposition materials (continued)
7/30/2019 DTM User Manual
32/58
4.4.3 Determine the acoustic impedance
Determinetheacoustic impedanceafter thedensity andtooling factorhave been calibrated. The
acoustic impedance only becomes important for heavily loaded crystals, and frequently the
valueforquartz(8.834x105 g cm-2 s-1) isadequate(seeAppendix A1). Table 4 shows the acousticimpedance of some of the common deposition materials; where "*" is shown in the table, no datafor the material is available: calculate the necessary value, or use the default value.
Calculate the acoustic impedance for the material as described below:
1. Fit a new crystal.
2. Deposit a layer of the materialuntil the crystal sensorceasesto oscillate and notethe material
thickness at which failure occurred.
3. Fit another new crystal and load it to 80% of the value noted in Step 1.
4. Place a masked sample in the substrate holder and make a second, short deposition (with athickness of approximately 80 nm).
5. Take note of the indicated value.
6. Use multiple beam interferometry or a stylus measuring device to measure the correct
deposition thickness.
7. Modify the acoustic impedance (z-value), as described in Section 4.3.3, until the correctthickness is displayed.
4.5 Deposition/etch processes
4.5.1 Prepare and enter data
1. Prepare a table of the parameter data to be entered. Table 5 shows parameter data for a
typical two-layer deposition.
2. Enter the data for each layer into the FTM7: refer to Section 4.3.3.
3. Select and enter the crystal sensor mode: refer to Section 4.3.4.
4. Select and enter the analogue output mode: refer to Section 4.3.5.
Layer Material Density z-value Terminal
Thickness
Tooling
factor
xtal
1 Cr 7.2 28.94 20 m 1 1
2 Cu 8.93 20.20 500 m 1 1 or 2
Table 5 - Parameter data for a typical two-layer deposition
26 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
33/58
4.5.2 Start/stop deposition/etch
Note: During sequence layer operation, when the shutter is closed, the Data/Rate display identifies the
sequence layer and the material layer. For example, if the Data/Rate display shows '16 4', this
indicates that sequence 16 has been selected, and that the deposition/etch parameters for layer 4will be used in the next deposition/etch cycle.
Use the procedure below to start a deposition/etch process. During deposition/etch, the
deposition/etch rate will be shown on the Data/Rate display, and the film thickness will beshownontheThickness display. Thextal1and xtal2LEDs will identify thecrystalsensorin use.
Note that if youselectsequencelayeroperation(see Section 1.5.7)andyouhave entered terminalthicknesses for each of the layers in the sequence, you must press the Run button at the end of
layer deposition/etch process to start the deposition/etch process for the next layer in thesequence.
1. Select the required layer: refer to Section 4.3.3.
2. Refer to Figure 1. Press and release the Run button (5); the Data/Rate and Thickness
displays will then show zero and the appropriate shutter will open to start thedeposition/etch.
3. If you haveconfigured the FTM7 for manual operation, press the Run button when required
to close the shutter and stop deposition.
If youhave configured theFTM7 forautomaticoperation, theFTM7 will automatically closethe shutter at the specified termination thickness.
4. If you haveselectedsequence layer operation, continueat Step2 tostart the deposition/etchprocess for the next layer in the sequence.
5. Repeat Steps 1 to 4 as necessary, for further depositions.
4.6 Failures during deposition/etch
4.6.1 Crystal sensor failures
Single crystal
installations
If you have a single crystal sensor and shutter connected to the FTM7, if the
crystal sensor fails, the shutter will automatically close and both the openand close shutter LEDs will flash alternately, however the FTM7 will
assume that the process is still in progress.
To continue the process without loss of thickness information:
Replace the failed crystal sensor.
Press the Run button to open the shutter again and continuedeposition/etch.
FTM7 Film Thickness Monitor 27
7/30/2019 DTM User Manual
34/58
Two crystal
installations
In backup mode, if crystal sensor 1 fails during a deposition/etch process,
the FTM7 will automatically switch to use crystal sensor 2.
In layer selected mode, if either crystal sensor fails, the shutter will
automatically close, however the FTM7 will assume that the process is stillin progress.
To continue the process without loss of thickness information:
Replace the failed crystal sensor.
Press the Run button to open the shutter again and continuedeposition/etch.
4.6.2 Electrical supply failures
If the electrical supply fails during a deposition/etch process, the shutter will automaticallyclose, however the FTM7 will assume that the process is still in progress.
When the electrical supply is restored, press the Run button to open the shutter again and
continue the deposition/etch without loss of thickness information.
4.7 Switch off the FTM7
Move the On/Off switch on the rear of the FTM7 (Figure 2, item 4) to the 'off' position(position '0').
Alternatively, switch off the rack isolator if you have fitted the FTM7 in a rack.
28 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
35/58
5 MAINTENANCE
5.1 Check the electrical connections
Regularly do the following checks:
1. Inspect all of the electrical connections to the FTM7 and check that they are secure. Tightenany loose connections.
2. Inspect all of the electrical cables and wires and check that they are not damaged and havenot overheated. Replace any cable or wire which is damaged or which has overheated.
5.2 Run the test mode
Refer toFigure1. Atany timewhenno shutteris open, you can press the Testbutton(7) toinitiate
test mode. When you initiate test mode, the following should occur:
The Test LED (8) will go on.
For approximately two seconds, all of the LEDs on the front panel will go on and theData/RateandThicknessdisplays (1,2) will both show 8.8.8.8. Youcanthereforecheck that
there are no failed LEDs or display segments.
All of the LEDs will then go off for approximately 2 seconds.
Both shutter relay contacts will then open for one second, and then close again; the shutterstatusLEDs (4) will goon and off toreflect the shutter relay status. This sequencewill repeatthree times. This allows you to check for correct operation of the shutter(s) connected to the
FTM7.
The current crystal sensor frequency will be shown on the Data/Rate and Thicknessdisplays (1, 2), updated once a second. If a crystal sensor and oscillator have been correctlyconnected, the displays will typically show a frequency of approximately 5999 900. If the
displays show a different frequency, refer to Section 5.3 for fault finding. The Test LEDflashes on and off while xtal sensor data is shown and you may press the shutter button to
open the shutter, start a deposition and observe the frequency change. When the xtalfrequency is shown, the analogueoutput is driven: the outputvoltage ramps from 0 to 1 V in
5 seconds then returns to 0 V before repeating the cycle to enable you to calibrate your
analogue output device.
Press the Test button (7) again to deselect test mode. When you deselect test mode, the TestLED (8) will go off.
If any of the above checks fail, refer to Section 5.3 for fault finding.
FTM7 Film Thickness Monitor 29
7/30/2019 DTM User Manual
36/58
5.3 Fault finding
WARNING
Disconnect the FTM7 from the mains supply before you remove the cover. If you do not,there may be a risk of injury by electric shock, or of damage to the FTM7.
WARNING
Disconnect the shutter controller cables from the FTM7 before you remove the cover. Ifyou do not, there may be a risk of injury by electric shock.
5.3.1 General fault finding
Refer to Table 6 for general fault finding.
Symptom Check Action
No LEDs are on and thedisplays are blank.
Is the FTM7 disconnected from theelectrical supply ?
Ensure that the electrical supply cableis correctly fitted to the FTM7 and to
the external electrical supply.
Is the FTM7 switched off ? Ensure that the On/Off switch on therear of the FTM7 is switched on, and
that the external electrical supply is on.
Is the FTM7 configuredincorrectly ?
Check that the FTM7 is correctlyconfigured for your electrical supply.
Has the electrical supply fuse
failed ?
Inspect the fuse and replace if
necessary.
The LEDs are on, but the
displays are blank.
Is a crystal connected ? Check that there is a crystal in the
crystal holder. Switch the unit off then
on again.
Is a crystal connected to the sensor
input ?
Connect the crystal. Switch the unit off
then on again.
Is there a fault in the FTM7 ? Run the test mode to check for correctFTM7 operation. If the displays remain
blank, contact your supplier or BOCEdwards.
In test mode, a crystal
sensor frequency of 0 to400 is displayed.
Are the electrical connections
incorrect ?
Check for open circuits between the
FTM7, the oscillator unit, the crystalholder and the crystal. Rectify any
problem found.
30 FTM7 Film Thickness Monitor
Table 6 - Fault finding
7/30/2019 DTM User Manual
37/58
Symptom Check Action
The displays all show000, 000.
Is the oscillator unit connected thewrong way round ?
Check that the oscillator unit iscorrectly connected to the crystal.
In test mode, a crystal
sensor frequency of 400 to510 000 is displayed.
Is the oscillator unit connected the
wrong way round ?
Check that the oscillator unit is
correctly connected to the crystal.
Is there a poor connection in thecrystal holder, or is it faulty ?
Inspect the crystal holder and crystal.Replace the crystal holder and crystal
as necessary.
Is the crystal overloaded ? Inspect the crystal and replace it if
necessary.
In test mode, a crystalsensor frequency of 510
000 to 599 900 is displayed.
Is the crystal nearly overloaded ? The crystal is probably heavily coated,but is still operating at this time. We
recommend that you inspect the crystaland replace it as soon as possible.
In the test mode, the
shutter status LEDs do notoperate correctly.
Is the FTM7 faulty ? If the LEDs do not operate correctly,
but go on at the start of the test mode,the FTM7 is faulty: contact your
supplier or BOC Edwards.
The xtal1 LED flashes. Is the crystal sensor disconnectedor mis-connected ?
Check the electrical connectionsbetween the crystal holder, the
oscillator unit and the FTM7; correctany problem found.
The xtal1 LED flashes(continued). Is the crystal sensor faulty ? If the electrical connections are correct,the crystal sensor is probably faulty:replace the crystal sensor.
Is a cable damaged ? If you replace the crystal sensor and the
problem still exists, check thecontinuity of the coaxial cables and
replace any damaged cable.
A shutter does not operate. Has the crystal sensor failed ? Run the test mode and refer to theprevious symptoms in this table, to
determine if the crystal sensor hasfailed, or whether the FTM7 is faulty,
and to identify the actions to take.
Is there an electrical connectionfault ?
If the FTM7 shutter status LEDsoperate correctly, inspect the
connections between the FTM7 and theshutter; rectify any problem found.
Is the shutter faulty ? Inspect the shutter and replace if
necessary.
FTM7 Film Thickness Monitor 31
Table 6 - Fault finding (continued)
7/30/2019 DTM User Manual
38/58
Symptom Check Action
None of the buttons haveany effect.
Was there a transient electricalproblem, or is there an earth
(ground) fault ?
Switch off the FTM7, then switch it onagain and run the test mode. If there is
no apparent problem, check the earth(ground) connections and rectify any
problem found.
Are the FTM7 controls disabled ? If you have connected the FTM7 toyour control equipment through the
RS232 connector, ensure that you havenot inadvertently disabled the FTM7
controls; if so, send a reset command:refer to Appendix A3.
The Increment and
Decrement buttons haveno effect.
Is the shutter open ? If so, this is not a fault: you cannot
change deposition data when theshutter is open during a deposition
process.
Has the shutter closed due to acrystal sensor failure ?
If so, this is not a fault: you mustreplace the crystal and restart and
finish the deposition process beforeyou can change deposition data.
- - If you have made all of the checks
above and still cannot identify thecause of a fault, or if you cannot rectify
a fault, contact your supplier or BOC
Edwards.
32 FTM7 Film Thickness Monitor
Table 6 - Fault finding (continued)
7/30/2019 DTM User Manual
39/58
5.3.2 Display fault messages
Fault messages are displayed on the Data/Rate display. The messages and their meanings are
shown in Table 7.
Note that these meanings are supplied for information only. If any of these messages are
displayed, you must contact your supplier or BOC Edwards for advice; you cannot service orrepair the FTM7 yourself.
When you read Table 7, take note of the following:
The232message will only appearduring initial switch-on(refer to Section 3.5); if shown,themessage will also appear on the Thickness display.
Messages E3 to E7 will only appear during initial switch-on (refer to Section 3.5).
Messages E101 to E106 and Pxxx will only appear during normal operation.
Message Pxxx means a message which starts with P and is followed by up to three digits. Ifyou contact your supplier or BOC Edwards after this message has been displayed, you will
be requested to supply the following information:
The actual fault message (Pxxx) displayed.
The software version number (see Section 3.5).
Fault
message
Meaning
232 The RS232 hardware has failed.
E3 The non-volatile memory test has failed.
E4 The main counter chip is faulty.
E6 A front panel button is permanently 'on'.
E7 The digital to analogue output chip is faulty.
E101 The main counter chip has failed to program three times.
E102 As E101.
E103 As E101.
E105 The digital to analogue chip has failed to program three times.
E106 As E105.
Pxxx There is a software fault.
Table 7 - Display fault messages
FTM7 Film Thickness Monitor 33
7/30/2019 DTM User Manual
40/58
6 STORAGE AND DISPOSAL
6.1 Storage
When you want to store the FTM7, return it to its protective packaging and store it in clean, dry
conditions.
When required for use, install the FTM7 as described in Section 3.
6.2 Disposal
Dispose of the FTM7 and any components safely in accordance with all local and national safetyand environmental requirements.
34 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
41/58
7 SERVICE AND ACCESSORIES
7.1 Introduction
BOC Edwards products, spares and accessories are available from BOC Edwards companies in
Belgium, Brazil, China, France, Germany, Israel, Italy, Japan, Korea, Singapore, United
Kingdom, U.S.A, and a world-wide network of distributors. The majority of these centresemploy ServiceEngineerswhohaveundergonecomprehensiveBOCEdwardstraining courses.
Order spare parts and accessories from your nearest BOC Edwards company or distributor.When you order, please state for each part required:
Model and Item Number of your equipment
Serial number (if any)
Item Number and description of part
7.2 Service
BOC Edwards products are supported by a world-wide network of BOC Edwards ServiceCentres. Each Service Centre offers a wide range of options including: equipment
decontamination; service exchange; repair; rebuild and testing to factory sepcifications.Equipment which has been serviced, repaired or rebuilt is returned with a full warranty.
Your local Service Centre can also provide BOC Edwards engineers to support on-sitemaintenance, service or repair of your equipment.
For more information about service options, contact your nearest Service Centre or other BOCEdwards company.
7.3 Accessories
Accessory Item Number
Oscillator unit E086-66-000
Crystals (pack of 5) E086-68-000
Water cooled crystal holder E086-67-000
FTM7 Film Thickness Monitor 35
7/30/2019 DTM User Manual
42/58
(This page deliberately left blank)
36 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
43/58
APPENDIX A1- MEASUREMENT THEORY
A1.1 Crystal sensor operation
The sensor element for the FTM7 is a quartz crystal oscillator. The crystal itself is a plano-convex
plate approximately 14 mm in diameter and 3 mm thick, which is excited into thickness
shear-mode vibrations by an oscillator unit at a frequency of approximately 6 MHz.
The frequency of oscillation is determined by the mass of material which is deposited on the
crystal surface, when the crystal is exposed to the deposition source. As the deposit builds up,the oscillations slow down. BOC Edwards use a plano-convex crystal since this shape induces
most of the oscillations to occur in the central exposed region of the crystal. This increases thesensitivity of thecrystal todepositedmaterial, andmeans that thecrystal maybe firmlyclamped
around the rim; this increases the effectiveness of the cooling mechanism.
A1.2 Measurement in first generation instruments
When quartz crystal resonators were first used to monitor film deposition, it was assumed thatthedepositedfilm wouldhave thesameeffectas an equivalent distributedmass ofquartz, thus:
[1] TN D f f f
Dfq q q
f
=. . . ( )2
where:
Tf
= film thickness (cm)
Dq = density of quartz (g cm-3)
Df= density of film (g cm-3)
Nq = frequency constant for the quartz crystal oscillating in thickness mode (Hz cm-1)
f= frequency of load crystal (Hz)
fq = frequency of unloaded crystal (Hz)
Early crystal detectors were unable to withstand heavy loading and so it was reasonable to
assume fapproximately = fq, in which case equation [1] shows that the deposit thickness isproportional to frequency change.
Note, however, from equation [1] that the 'proportionality constant' is notin fact a constant sincethe term
N D f
D
q q
f
. . 2
explicitly contains the current crystal frequency, which of course is changing.
FTM7 Film Thickness Monitor 37
7/30/2019 DTM User Manual
44/58
A1.3 Measurement in second generation instruments
Second generation film thickness monitors resulted from improvements in crystals, oscillator
circuits and compact, low cost integrated circuits. Crystals could withstand much heavier
loading. These second generation instruments used the following equation:
[2] T ND
DT T
f q
q
f
q= . . ( )
where:
T= period loaded crystal
Tq = period of unloaded crystal
This equation is still an approximation, but is much more accurate than equation [1]. Note that,provided 'period' is thebasic measurement, theequation is linear. This measurement technique
proved to be a considerable improvement and was widely accepted.
A1.4 Measurement in the FTM7
The original assumption - that the effect of loading a quartz crystal with a film material isequivalent to loading it with an equivalent mass of quartz - is of course questionable. Further
advances in instrumentation showed that for very thick films, equation [2] was not valid.
A more rigorous acoustic analysis of the film material, and of the loaded crystal as a one
dimensional composite resonator of quartz, leads to the following equation:
[3] TD
DN T
Z
Z
Z
Z
T
Tfq
f
q
f
q
q
f
q=
. . .
.. tan . tan .
314314 11
where:
Zf= acoustic impedance of the film material
Zq = acoustic impedance of quartz.
This equation now takes account of the acoustic impedance of the material relative to quartz(acoustic impedance ratio).
Equation [3] reduces to equation [2] if Zf = Z. The ability of equation [3] to describe more
accurately the behaviour of a heavily loaded crystal has been demonstrated. The use ofmicroprocessing technology in the FTM7 allows equation [3] to be solved economically. It
should be noted, however, that the deviation of behaviour from that predicated by equation [2]only becomes appreciable for heavily loaded crystals, and for the majority of applications
equation [2] is adequate.
38 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
45/58
7/30/2019 DTM User Manual
46/58
Mode Process Parameter Full scale deflection
0 Deposition Deposition/etch rate 10 nm s-1
1 Etch Deposition/etch rate 10 nm s-1
2 Deposition Film thickness 10 nm3 Etch Film thickness 10 nm
4 Deposition Film thickness Terminal thickness range
5 Etch Film thickness Terminal thickness range
6 Deposition Film thickness modulo 10 nm
7 Etch Film thickness modulo 10 nm
8 Deposition Deposition/etch rate 100 nm s-1
9 Etch Deposition/etch rate 100 nm s-1
10 Deposition Film thickness 100 nm
11 Etch Film thickness 100 nm
12 Deposition Film thickness Terminal thickness range
13 Etch Film thickness Terminal thickness range
14 Deposition Film thickness modulo 100 nm
15 Etch Film thickness modulo 100 nm
16 Deposition Deposition/etch rate 1 m s-1
17 Etch Deposition/etch rate 1 m s-1
18 Deposition Film thickness 1 m
19 Etch Film thickness 1 m
20 Deposition Film thickness Terminal thickness range
21 Etch Film thickness Terminal thickness range
22 Deposition Film thickness modulo 1 m
23 Etch Film thickness modulo 1 m
24 Deposition Deposition/etch rate 10 m s-1
25 Etch Deposition/etch rate 10 m s-1
26 Deposition Film thickness 10 m
27 Etch Film thickness 10 m
28 Deposition Film thickness Terminal thickness range
29 Etch Film thickness Terminal thickness range
30 Deposition Film thickness modulo 10 m
31 Etch Film thickness modulo 10 m
32 Deposition Deposition/etch rate 100 m s-1
33 Etch Deposition/etch rate 100 m s-1
34 Deposition Film thickness 100 m
35 Etch Film thickness 100 m
36 Deposition Film thickness Terminal thickness range
37 Etch Film thickness Terminal thickness range
38 Deposition Film thickness modulo 100 m
39 Etch Film thickness modulo 100 m
Table A2-1 - Analogue output modes
40 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
47/58
APPENDIX A3 - RS232 INTERFACE
A3.1 Connector pin functions
The functions of the pins in the RS232 connector are shown in Table A3-1.
Pin Signal Function
2 TxD Transmit data
3 RxD Receive data
4 RTS Request To Send
5 CTS Clear To Send
7 Ground (earth) -
Table A3-1 - Pins in the RS232 connector mating-half
Use the signals as shown below:
TxD The FTM7 uses this pin to transmit serial data. The data is transmitted at4800 baud, with 1 start bit, 7 data bits, even parity, and 2 stop bits.
RxD The FTM7 uses this pin to receive serial data. This data must be in the sameform as the data transmitted from the FTM7 (see above).
RTS This output from the FTM7 is 'on' (high) when the FTM7 can receive data.
The output is 'off' (low) when the FTM7 is busy. Data transmitted to theFTM7 when this output is off may not be received correctly.
CTS The FTM7 can only transmit data when this input is on.
A3.2 Communications protocol
TheRTSandCTSsignalscanbeusedtoregulatetheflowofcharactersthroughtheseriallink.
The master device can use the CTS signal to control the transmission of data by the FTM7:
When CTS is off (low), the FTM7 transmissions are inhibited. When CTS is on (high), transmissions by the FTM7 are enabled.
IfyourmasterdevicecannotprovideasuitableCTSsignal,youshouldlinktheCTSandRTSpins
in the RS232 connector mating-half.
The FTM7is designedsuchthatyou will not normally needtouse the RTS signal toregulatedata
transmissionby themaster device. However, if theFTM7reports overrun errorson theinterface,you must use the RTS signal.
FTM7 Film Thickness Monitor 41
7/30/2019 DTM User Manual
48/58
A3.3 Message protocol
For reliable operation, use the following message protocol between the master device in your
control system and the FTM7:
The master device transmits a command message (with any parameters) to the FTM7.
The FTM7 interprets the message, then transmits a reponse message to the master device.
The master device receives the FTM7 reponse message.
The master device must wait until it has received the FTM7 response message before it attemptsto transmit another command message.
Refer to Section A3.4 for descriptions of the command messages, and refer to Section A3.5 fordescriptions of the return messages.
A3.4 Command messages
A3.4.1 Introduction
Each command message to the FTM7 consists of a string of characters which define thecommand, terminated by the (carriage return) character (ASCII decimal 13). Some
command messages are then followed by command parameters, each of which is alsoterminated by .
Only thefollowing characters will berecognised:alphanumerics (upperor lowercase),".", "+", "-"and ; all other characters will be ignored.
The commands available are summarised in Table A3-2; note that in a command message, youcan either send the full command name, or the specified two-character abbreviated command.
We recommend that you use abbreviated commands, as these require less processing by the
FTM7.
Note that commands are not always valid; when invalid, a command will be ignored by the
FTM7. The 'validity' column of Table A3-2 identifies command validity as follows:
Validity 'A' commands are always valid.
Validity 'PS' commandsareonly validwhen thereis nodeposition/etch process inprogress.
Validity 'PR' commands are only valid if a deposition/etch process is in progress.
The uses of the command messages are fully defined in Sections A3.4.1 to A3.4.16.
42 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
49/58
Command name Abbreviated
command
Use Validity
CHKAOMODE CA Check the analogue output mode. A
CHKMATERIAL CM Check the layer (material) parameters. A
CHKRATE CR Check the deposition/etch rate. A
CHKSHUTTER CS Check the shutter status. A
CHKTHICKNESS CT Check the thickness. A
CHKVERSION CV Check the software version. A
CHKXTAL CX Check the crystal used. A
KEYBOARD KB Disable the FTM7 controls. A
PROGAOMODE PA Program the analogue output mode. PSPROGMATERIAL PM Program material parameters. PS
RESET RS Reset the FTM7. A
SELMATERIAL SM Select layer (material). PS
SHUTTERCL SC Close the shutter. A
THICKCHG TC Change the terminal thickness. PR
THICKLOOK TL Open the shutter. PS
Table A3-2 - Commands
A3.4.2 Check the analogue output mode
Command message: CHKAOMODE
Return message: {return code}{analogue mode}
Use this message to check the analogue output mode. In the return message:
{analogue mode} is in the range 0 to 39: refer to Appendix A2.
FTM7 Film Thickness Monitor 43
7/30/2019 DTM User Manual
50/58
A3.4.3 Check the current layer data
Command message: CHKMATERIAL
Return message: {return code}{layer}{density}{z-value}{terminal thickness}{tooling factor}{xtal}
Use this message to check the parameters for the current layer. In the return message:
{layer} is in the range 1 to 35 and defines the current layer or sequence
layer: see Table A3-3.
{density} is in the range 0.01 to 99.99 and defines the density of the
deposition material for the layer.
{z-value} is in the range 1.0 to 99.9 and defines the acoustic impedance for
the layer.
{terminal thickness} is in the range 0.0 to 99990.0 and defines the terminal thickness
(in nm).
{tooling factor} is in the range 0.01 to 99.90 and defines the tooling factor.
{xtal} is in the range 1 to 3 and specifies the crystal sensor: if {xtal} = 3,this specifies that the FTM7 is operating in backup mode.
A3.4.4 Check the deposition/etch rate
Command message: CHKRATE
Return message: {return code}{rate}
Use this message to check the current deposition/etch rate. In the return message:
{rate} is in the range 0.0 to 999.90 and defines the current rate (in nm s-1).
A3.4.5 Check the shutter status
Command message: CHKSHUTTER
Return message: {return code}{status}
Use this command to check the shutter status. In the return message:
{status} is 0 (shutter closed) or 1 (shutter open).
44 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
51/58
RS232 {layer} Equivalent Data/Rate
display
Applicable material
layer
1 to 11
1213
1415
1617
18
1920
2122
2324
2526
2728
2930
31
3233
34
35
1 to 11
12 113 1
14 115 1
16 317 5
18 7
19 9- (not used)
12 213 2
13 314 2
14 314 4
15 215 3
15 415 5
16 4
17 618 8
19 7
19 8
1 to 11
11
11
35
7
9- (not used)
22
32
34
23
45
4
68
7
8
Table A3-3 - RS232 layer data
A3.4.6 Check the current thickness
Command message: CHKTHICKNESS
Return message: {return code}{thickness}
Use this message to check the current thickness. In the return message:
{thickness} is in the range 0.0 to 99990.0 and defines the thickness (in nm).
FTM7 Film Thickness Monitor 45
7/30/2019 DTM User Manual
52/58
A3.4.7 Check the FTM7 software version
Command message: CHKVERSION
Return message: {return code}{s/w version}
Use this message to check the version of the software in the FTM7. In the return message:
{s/w version} is an ASCII string which defines the software identification and itsversion.
A3.4.8 Check the current crystal sensor
Command message: CHKXTAL
Return message: {return code}{xtal}{frequency}
Use this message to check the current crystal sensor. In the return message:
{xtal} is in the range 1 or 2 and defines the crystal sensor number.
{frequency} defines the crystal frequency (in Hz).
A3.4.9 Disable the FTM7 controls
Command message: KEYBOARD
Return message: {return code}
Use this message to disable the FTM7 controls. After you have sent this command, the controls
on the FTM7 are ignored. You must send a RESET command to re-enable the FTM7 controls (seeSection A3.4.12).
A3.4.10 Select the analogue output mode
Command message: PROGAOMODE{analogue mode}
Return message: {return code}
Use this message to select the analogue output mode. In the command message:
{analogue mode} is in range 0 to 39 to specify the required mode: see Appendix A2.
Note that the 'layer' and 'mode' LEDs on the front of the FTM7 will flash when this command
message is accepted.
46 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
53/58
A3.4.11 Program layer parameters
Command message: PROGMATERIAL{layer}{density}{z-value}{terminal thickess}{tooling factor}{xtal}
Return message: {return code}
Use this message to enter the deposition/etch parameters for a layer into the FTM7. In thecommand message:
{layer} is in the range 1 to 11 and specifies the layer.
{density} is in the range 0.01 to 99.99 and defines the density.
{z-value} is in the range 1.0 to 99.9 and defines the acoustic impedance.
{terminal thickness} is in the range 0.1 to 99990.0 and specifies the required terminalthickness (in nm).
{tooling factor} is in the range 0.01 to 99.90 and specifies the tooling factor.
{xtal} is in the range 1 to 3 and specifes the crystal sensor: set {xtal} to 3 toselect backup mode.
Note that the parameter LEDsonthe front of the FTM7will flashas the parameter data is enteredinto the FTM7.
A3.4.12 Reset the FTM7Command message: RESET
Return message: {return code}
Use this message toreset the FTM7. Whenyou sendthis command message, this is equivalent to
switchingoffthe FTM7, then switchingit onagain. Note that theFTM7 controls areenabledafterthis command message.
After you have sent a RESET command, wait for three seconds or more before you send anothercommand message. Note that if the command is accepted, the {return code} cannot be
guaranteed to be sent to your control equipment, and during re-initialisation, correct operationof the RS232 control signals are not guaranteed. We therefore recommend that you empty and
ignore any input queue in your master device.
FTM7 Film Thickness Monitor 47
7/30/2019 DTM User Manual
54/58
A3.4.13 Select layer (material)
Command message: SELMATERIAL{layer}
Return message: {return code}
Use this message to select the layer. In the command message:
{layer} is in the range 1 to 35 and specifies the layer: see Table A3-3.
Note that the 'layer' LED on the front of the FTM7 will flash when this command message is
accepted.
A3.4.14 Close the shutter
Command message: SHUTTERCL
Return message: {return code}
Use this message to close the shutter.
A3.4.15 Change terminal thickness
Command message: THICKCHG{terminal thickness}
Return message: {return code}
Use this message to change the terminal thickness. In the command message:
{terminal thickness} is in the range 0.1 to 99990.0 and specifies the required terminal
thickness (in nm).
A3.4.16 Open the shutter
Command message: THICKLOOK
Return message: {return code}
Use this message to open the shutter. When you send this message, it is equivalent to pressingthe Run button on the FTM7.
A3.5 Return messages
A return message consists of a {return code} terminated by a character (as for command
messages), together with return parameters where appropriate: refer to the command messagesdescribed in Section A3.4.
{return code} is an ASCII coded integer and its value specifies whether the command wassuccessfully acted upon: refer to Table A3-4.
48 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
55/58
{return code} Meaning
0 Command received successfully - no fault.
1 Unknown command received.
2 Invalid command: deposition/etch process in progress
3 Invalid command: deposition/etch process not in progress.
4 Conversion error: a command parameter was not converted to a real value.
5 Parameter error: a command parameter was outside the valid range.
6 Command too long: the number of characters in the command was > 15.
7 Parameter too long: the number of characters in a command parameter was > 15.
8 Program error: there was an internal FTM7 program error.
9 Buffer full: the command was ignored because the input buffer was full.
> 15 Communications error: {return code} will be 15 + {comms error}, where {comms error} is
one of (or a summation of) the following communications error codes:
Code Meaning
1 Overrun error
2 Parity error
4 Framing error
8 Received Break
For example, if {return code} = 17 (15+2), a parity error has been detected.If {return code} = 18 (15+2+1), an overrun error and a parity error have been detected.
Table A3-4 - Return codes
FTM7 Film Thickness Monitor 49
7/30/2019 DTM User Manual
56/58
A3.6 Programming example
A3.6.1 RS232 connections
Use the pins in the FTM7 RS232 connector mating-half to connect toan IBM PC/XT or PC/AT asshown in Table A3-5. You must use screened, multicore cable to make the connections. The
connector that is fitted to the FTM7 is a 9-way, female, D-type connector.
FTM7 connector mating-half IBM PC XT
DB25 connector
pin numbers
IBM PC XT/AT
signal
IBM PC AT
DB9 connector
pin numbersPin number Signal
2 TxD 3 RxD 2
3 RxD 2 TxD 3
4 RTS 5 CTS 8
5 CTS 4 RTS 7
7 Ground (earth) 7 Ground (earth) 5
20, 6* DTR, DSR 4, 6*
* Link these pins in the connector.
Table A3-5 - RS232 connections to an IBM PC/XT or PC/AT
A3.6.2 Sample program
The example software shown in Figure A3-1 is written in QuickBASIC Version 4.0 or later and
shows how to check the FTM7 software version, how to read the material parameters for thecurrently selected layer, and how to select the analogue output mode.
50 FTM7 Film Thickness Monitor
7/30/2019 DTM User Manual
57/58
' FTM7 QuickBASIC RS232 Programming Example
' Clear the computer screen and display a headingCLS
PRINT "FTM7 RS232 Test": PRINT
' Open up a serial port for communicating with the FTM7' com2 = serial port 2, 4800 = baud rate, E = even parity
' 7 = number of data bits, 2 = number of stop bits' Change com2 to com1 for serial port 1
comfil$ = "com2:4800, E, 7, 2"OPEN comfil$ FOR RANDOM AS #1
' Set up the command ChkVersion, use the short form "cv"comand$ = cv
' The print command sends the data to the FTM7 and adds' the Carriage Return code
PRINT #1, comand$
' Line Input reads the data back from the FTM7LINE INPUT #1, retcode$' Display the command and return code
PRINT "Command sent - " + comand$,
PRINT "Return Code - " + retcode$' Check that the FTM7 did not return an error
IF retcode$ "0" GOTO FTM7error' As there was no error, read in the Version number
LINE INPUT #1, Verno$' and display it
PRINT "Unit ID - " + Verno$: PRINT' So far we have sent cv and received back
' 0 FTM7 VX.XX which is 13 characters at 2 msec' each. The FTM7 updates at once a second so there
' is no point in continually communicating with it.' The next instructions are a simple delay
FOR I% = 1 TO 1000
NEXT I%' Set up the command ChkMaterial and use the short form "cm"
comand$ = "cm"PRINT #1, comand$
LINE INPUT #1, retcode$PRINT "Command sent - " + comand$,
PRINT "Return Code - " + retcode$IF retcode$ "0" GOTO FTM7error
' No error so read in the material parametersLINE INPUT #1, Mat$
LINE INPUT #1, Den$LINE INPUT #1, Zval$
LINE INPUT #1, Term$LINE INPUT #1, Tool$
LINE INPUT #1, Xtal$
FTM7 Film Thickness Monitor 51
Figure A3-1 - QuickBASIC programming example
7/30/2019 DTM User Manual
58/58
' Display the parameters along a line
PRINT "Material - " + Mat$;PRINT "Density - " + Den$;
PRINT "Z-value - " + Zval$;
PRINT "Terminate - " + Term$;PRINT "Tooling - " + Tool$
PRINT "Xtal - " + Xtal$: PRINT' The next instructions are a simple delay
FOR I% = 1 TO 1000NEXT I%
' An example to program a value in the FTM7' ProgAOMode in short form "pa"
comand$ = "pa"AOMode$ = "8"
PRINT #1, comand$
PRINT #1, AOMode$LINE INPUT #1, retcode$PRINT "Command Sent - " + comand$, " Value Sent - " + AOMode$,