+ All Categories
Transcript
Page 1: Enhancing Gas Chromatography Performance - Energy

W. T. Shmayda, N. P. Redden, R. F. EarleyUniversity of RochesterLaboratory for Laser Energetics

Enhancing Gas Chromatography Performance

1

Tritium Focus Group-SandiaAlbuquerque, NM

22–25 October 2018

Page 2: Enhancing Gas Chromatography Performance - Energy

E28000

2

• Thermal conductivity and tritium monitor detector measurements are complementary and predict similar atom fractions in H/D/T mixtures within 1%

• Miniaturization of the pressure sensing detector and the effluent isolation valve will reduce DT injection quantities for assay by 60%

• LLE has analyzed two of the three inactive gas samples provided by Sandia National Laboratories as part of the round robin exercise to compare assay results across the complex

Outline/Summary

Page 3: Enhancing Gas Chromatography Performance - Energy

Filling cryogenic targets in the permeation cell is a multistep process

G6218r

5

DT High-Pressure System (DTHPS) Glovebox

Tertiary containment

Syringe pump

Cryostat

FTS Glovebox

To tritium removal system

TFS Glovebox

USB 2 Assay

USB 1

DTHPScondensation

tube

High-pressure

guage Permeation cell

Condensationtube

P/Ddiffuser

TFS: Tritium Fill StationUSB: uranium storage bedFTS: Fill/Transfer Station

Page 4: Enhancing Gas Chromatography Performance - Energy

The lighter isotopic species concentrate whilepressurizing the permeator in the Tritium Fill Station

E25066c

6

Medium-pressuresystem

High-pressuresystem

T/D = 1.62

T/D = 1.60T/D = 1.62

PermeatorDiaphragmcompressor

TFS

T/D = 2.33

Composition in %: H/D/T

Variation in T/D downstream of the TFS = 1.2%

Page 5: Enhancing Gas Chromatography Performance - Energy

Gas chromatography uses iron-doped alumina at 77 K to separate the hydrogen isotopologues

E26599b

3

Minimum aliquot size: 80 nLError:

– systematic 1.2% – relative (peak to peak) <0.5% – reproducibility <0.2%

Fe-dopedMS 5 ÅLN2

Helium

Helium

MS 5 Å80°C

NeonTCD

Sample

Vacuum

Neon

P

TM

TCD

MS: molecular sieve TM: tritium monitor TCD: thermal conductivity detector

Page 6: Enhancing Gas Chromatography Performance - Energy

The assaying process requires four steps; first step: evacuate the injection volume

E26599c

4

Fe-dopedMS 5 ÅLN2

Helium

Helium

MS 5 Å80°C

NeonTCD

SampleVacuum

Neon

P

TM

TCD

To effluent capture

To effluent capture

Pump

Page 7: Enhancing Gas Chromatography Performance - Energy

Fe-dopedMS 5 ÅLN2

Helium

Helium

MS 5 Å80°C

NeonTCD

SampleVacuum

Neon

P

TM

TCD

To effluent capture

To effluent capture

Pump

Second step: charge the loop with a predefined aliquot of gas

E26599d

5

Minimum aliquot size: 80 nL

Page 8: Enhancing Gas Chromatography Performance - Energy

VacuumTo effluent capture

Pump

Fe-dopedMS 5 ÅLN2

Helium

Helium

MS 5 Å80°C

NeonTCD

Sample

Neon

P

TM

TCD

To effluent capture

Third step: inject the aliquot into the separation column and detection circuit

E26599e

6

Page 9: Enhancing Gas Chromatography Performance - Energy

Fourth step: sample analysis

E28005

7

00.0

2.5

5.0

7.5

2 4Time (min)

O2 CH4

COSign

al (m

V)

6 8

Fe-dopedMS 5 ÅLN2

Helium

Helium

MS 5 Å80°C

NeonTCD

Sample

Vacuum

Neon

P

TM

TCD

To effluent capture

To effluent capture

0

3

6

9

Sig

nal

(V)

Time (min)0 20 40

50% T91% T

T2

HTDT

Time (min)0

0

1

2

20

H2T2

HT

HD

5.45% H60.62% D33.93% T

40

Det

ecto

r sig

nal (n

V)

60

D2 DT

Page 10: Enhancing Gas Chromatography Performance - Energy

The 4-cm3 tritium monitor and thermal conductivity agree within 1% over all T/D ratios

E28001

8

1.04

1.03

1.02

1.01

1.00

0.99

0.98

0.97

0.960 10 20 30 40 50 60 70 80 90 100Tritium fraction from thermal conductivity detector

TCD error: !1.5%

Triti

um m

onito

r/the

rmal

con

duct

ivity

The minimum aliquot required for analysis by the TM is 0.01 nL.The gas inventory of a typical target is ~20 nL!

Page 11: Enhancing Gas Chromatography Performance - Energy

Sample

Vacuum

Neon

PDiaphragm

MKSbaratrongauge

Hand weld Microweld

Microweld

0.125

Hand weld

SS-4-VCR-1

Hand weld

6LV-4-VCR-3S-4TB7

SS-4BK-TW-1Cwith Cu STEM

60% of the gas delivered to the gas chromatograph is not used for analysis

E28002

9

Injection loop volume: 1.5 mLPressure transducer andisolation valve volume: ~2.24 mLVolume injected for assay: 3.74 mL (4.84 Ci)

Page 12: Enhancing Gas Chromatography Performance - Energy

Selector valvePressuretransducer

Backup 4B safetyisolation valve

0Pressure (Torr)

200 400 600 800 10000

20

40

60

80

Volta

ge (m

VDC)

Voltage (mVDC) = 0.0762 P(Torr) –2.369R2 = 0.9997

The volume of the pressure detector and isolation valve assembly has been miniaturized ~200 fold

E28003

10

Injection loop volume: 1.5 mLPressure transducer andisolation valve volume: <10 nLVolume injected for assay: ~1.5 mL (1.9 Ci)

Page 13: Enhancing Gas Chromatography Performance - Energy

Tritium interacts with stainless steel to generate CH4 and CO

E26600a

11

Impurity growth rateVessel ID CH4 CO677 21.2 14.3624 1.9 6.0

* Time after filling the sample vessel

00

30

60

90

120

150

2Time (min)

Sig

nal

(m

V/s

cc)

4

544 ppm 307 ppm265 ppm

60 ppmCO CO390 ppm

64 ppm169 ppm33 ppm

Vessel ID 677 Vessel ID 625

6 8 00

30

60

90

120

150

2Time (min)

4 6 8

51 h*601 h

66 h*602 h

CH4 CH4

Page 14: Enhancing Gas Chromatography Performance - Energy

0

2

1

3

Pro

tiu

m c

on

cen

trat

ion

in f

uel

(%)

0 600

Duration since purifying fuel (days)

1200

H2O

H2O

H2O

H2O

H2O

H2O

H2O

H2O

H2OHO

HO

Evacuated

Absorbedwater layer

Bulkmetal

Baked

Tritiumconditioned

HOH

M

M

M

M

M

HH

H

H

HO

HO

HO

HO

HO

HO

HO

H2O

H2O

HTOTO

HO

TO

TO

HOHTO

HM

M

M

M

M

HH

H

H

TM

M

M

M

M

HT

H

T

Assay vessel1.88 %/mo

Permeation cell

Contamination caused by

sample vessel

“D” and “T” exchange with adsorbed water layers on the inner surfaces of the sample vessel and the process loop

E26601b

12

• Direct measurement of the protium content in a gas contained in unconditioned vessels leads to 300% errors in the protium fraction

Page 15: Enhancing Gas Chromatography Performance - Energy

LLE is participating in the national lab round robin gas-analysis exercise

E28004

13

SampleID Composition Nominal

mixture (%)Measured

valueRSD*(%)

1 H2 35 42.95 0.11HD – 3.01 0.80D2 65 54.04 0.05H/D 53.8 80.02 0.16

3 H2 20 30.02 0.02HD – 44.30 0.06D2 20 25.50 0.01H/D 1 1.10 0.06

* Relative standard deviation

Page 16: Enhancing Gas Chromatography Performance - Energy

The isotopologue ‘HD’ dominates the H2/D2 spectrum in the Mass Cal Gas sample

2525 202015 1510 105 5

300

150

10

5

Time (min)Time (min)

Sign

al (m

V) H2

H2

D2

D2

HD

HD

Sample #1 (H2/D2) Sample #3 (Mass Cal Gas)

Page 17: Enhancing Gas Chromatography Performance - Energy

Need to run a calibration gas to confirm the locations of the non-hydrogenic species

1

0.5

Sign

al (V)

642Time (min)

H2

N2

CH4COorAr

This column can not identify CO2

Composition of Mass Cal Gas• H2• D2• N2• Ar• CH4• CO2

Page 18: Enhancing Gas Chromatography Performance - Energy

E28000

14

Outline/Summary

• Thermal conductivity and tritium monitor detector measurements are complementary and predict similar atom fractions in H/D/T mixtures within 1%

• Miniaturization of the pressure sensing detector and the effluent isolation valve will reduce DT injection quantities for assay by 60%

• LLE has analyzed two of the three inactive gas samples provided by Sandia National Laboratories as part of the round robin exercise to compare assay results across the complex


Top Related