LIGO-G040541-00-D
LIGO calibration during the S3 science run
Michael Landry
LIGO Hanford Observatory
Justin Garofoli, Luca Matone, Hugh Radkins (LHO), Rana Adhikari, Peter Fritschel (MIT),
Patrick Sutton (Caltech), Brian O’Reilly (LLO), Xavier Siemens (UWM), Gabriela González, Andres Rodriguez (LSU), Martin Hewitson (GEO)
GWDAW9December 19, 2004
Annecy, France
GWDAW9, 19 Dec 2004 2LIGO-G040541-00-D
Overview
1. Calibration basics
2. Calibration improvements for S31. Line strengths in all IFOs
2. Accurate DC calibrations
3. Checks on systematics
3. S4 photon calibrator
GWDAW9, 19 Dec 2004 3LIGO-G040541-00-D
IFO review
1 ( ) ( ) ( )( , ) ( , ) ( , ) ( , )
( ) ( )
t t G fh f t R f t ASQ f t ASQ f t
t C f
)()()()()(),( fDtfCtfAtfG
GWDAW9, 19 Dec 2004 4LIGO-G040541-00-D
Calibration method
• Measure open loop gain G, input unity gain to model• Extract sensing function C=G/AD from the model• Produce response function at time of the calibration, R=(1+G)/C• Now, to extrapolate for future times, monitor single calibration
line in AS_Q error signal, plus any changes in gain beta, and form alphas
• Can then produce R at any later time t, given alpha and beta at t• Independent time-domain method (spearheaded by Xavier
Siemens with input from Martin Hewitson) uses frequency domain filters as input. Calibrated frames available for use by analysis groups
• Xavi also demodulates lines, which allows for a nice check on alphas and betas (frequency domain method assumes alpha and beta real, whereas demodulated lines are complex)
GWDAW9, 19 Dec 2004 5LIGO-G040541-00-D
Sample error budget: S2 H1
GWDAW9, 19 Dec 2004 6LIGO-G040541-00-D
Calibration improvements I
Effect of lineamplitude on
error in productof alpha and beta
GWDAW9, 19 Dec 2004 7LIGO-G040541-00-D
S3 V3 , coefficients: L1
Error:0.5%
Over all of S3:
variation: 15% variation: 4%
GWDAW9, 19 Dec 2004 8LIGO-G040541-00-D
S3 Calibration errors
• Errors from reference models in S3 ~ errors in S2 (5-10%)• Random variations, errors in alpha, beta (60 sec integration time):
error
S3 variation
S3 variation
L1 0.5% 15% 4%
H1 0.3% 4% 1%
H2 0.7% 6% 2%
In S2: 0.7% (L1), ~3% (H1, H2).
GWDAW9, 19 Dec 2004 9LIGO-G040541-00-D
Calibration improvements IIDC calibration
GWDAW9, 19 Dec 2004 10LIGO-G040541-00-D
Calibration improvements II
ETMx
Asymmetric Michelson
Lock configuration by feedingBack to ETM
Instead of ~6% error on theETM calibration, ~2%
Lots of checks on actuationstrength calibration: toggling,Michelson swinging, freeswinging AS_Q, tidal actuators
Important for hardware injections
GWDAW9, 19 Dec 2004 11LIGO-G040541-00-D
Calibration improvements IIIunderstanding systematics
• Use of true dc calibration requires good understanding of systematics
• Assess systematic error in digital filter compensation, e.g. dewhitening filters
DigitalControl signal
Digital Anti-dewhitening
filter DAC
Analogdewhitening
filter
GWDAW9, 19 Dec 2004 12LIGO-G040541-00-D
S3 response function R comparison
• V2: assumes digital compensation is perfect• V3: uses hardware measurement
GWDAW9, 19 Dec 2004 13LIGO-G040541-00-D
S4: Photon calibrator
• previously installed and tested – laser troubles with rotating polarization• recently reinstalled, awaiting laser safety approval• expect limited test for S4: independent check on magnitude of calibration (and timing, too)
Oddvar Spjeld, LLO
GWDAW9, 19 Dec 2004 14LIGO-G040541-00-D
Conclusions
• LIGO frequency domain calibration improving in accuracy with each science run
• S3 calibration makes use of several new measurements• Anticipate independent check with photon calibrator for S4
GWDAW9, 19 Dec 2004 15LIGO-G040541-00-D
Contributions to the Error
Magnitude: 2
4
22
4
22
4
2222
1
1
1
1
1
1G
G
G
G
G
G
G
G
G
D
D
A
A
R
R
Phase: 2
4
22222
4
2222
1
1
1
1GDAR
G
GGG
G
G
R(fi,t) = 1 + (t) (t) G(fi)
(t) C(fi)
GWDAW9, 19 Dec 2004 16LIGO-G040541-00-D
News on calibrationsince Aug LSC meeting
Not enough!…• Coefficients: validation studies of “new” method (using
demodulated line)• Models:
» Codes were succesfully reviewed (P. Fritschel), no errors found. – V2 model version reviewed; V3 mods will need to be assessed
» Work in progress on LHO models (to incorporate hardware/digital filters)» Work in progress: systematic model/measurement comparison for different
calibration runs in L1. • Validation (use of!) X. Siemen’s h(t) frames has started…
GWDAW9, 19 Dec 2004 17LIGO-G040541-00-D
Hardware measurements
GWDAW9, 19 Dec 2004 18LIGO-G040541-00-D
S3 V3 , coefficients
• V2 (from P. Sutton’s SenseMon)» from SenseMon averaging input matrix» from SenseMon’s line amp, , and G0(f0)
• V3: use Xavier Siemens’s code to generate demodulated lines in ASQ, DARM, EXC
» Complex = ( 1/D0)*(DARM-EXC)/ASQ» Complex = -(D0/G0)*ASQ/DARM» Complex = -(1/G0)*(DARM-EXC)/DARM» Non-zero mean of imaginary part indicates errors in reference functions D0,G0 (at cal
freq).» Standard deviations of imaginary parts are error estimates in real part (and depend on
sampling frequency).
• Compare consistency of Xavi’s output and existing model
GWDAW9, 19 Dec 2004 20LIGO-G040541-00-D
“New” method: validation steps
H1, reference time
G0 phase “error”:1.5 deg
GWDAW9, 19 Dec 2004 24LIGO-G040541-00-D
S3 V3 coefficient: H1
GWDAW9, 19 Dec 2004 25LIGO-G040541-00-D
S3 V3 coefficient: L1
GWDAW9, 19 Dec 2004 26LIGO-G040541-00-D
S3 V3 , coefficients: H1
Error:0.3%
Over all of S3:
variation: 4% variation: 1%
GWDAW9, 19 Dec 2004 27LIGO-G040541-00-D
H1