PARAMETER RANGE
Nitric Oxide (NO) <5 - 4,000 ppb
Standard Nitrogen Dioxide (NO2)
<5 - 4,000 ppb
Ozone (O3) <5 - 1,800 ppb
Par cle Count 1 - 30 μm (size)
Op onal Carbon Monoxide (CO) 5 - 6,000 ppb
Sulphur Dioxide (SO2)
5 - 10,000 ppb
0
500
1000
1500
2000
2500
3/7/2015 3/17/2015 3/27/2015 4/6/2015 4/16/2015 4/26/2015 5/6/2015 5/16/2015
CO p
pb
CO
Reference
AQmesh
0
5
10
15
20
25
30
35
07-03-15 0:00 17-03-15 0:00 27-03-15 0:00 06-04-15 0:00 16-04-15 0:00 26-04-15 0:00 06-05-15 0:00 16-05-15 0:00
NO2 p
pb
NO2
Reference
AQMesh
0
10
20
30
40
50
60
3/7/2015 3/17/2015 3/27/2015 4/6/2015 4/16/2015 4/26/2015 5/6/2015 5/16/2015
O3 p
pb
O3
Reference
AQMesh
0
20
40
60
80
100
120
140
160
3/7/2015 3/17/2015 3/27/2015 4/6/2015 4/16/2015 4/26/2015 5/6/2015 5/16/2015
NO p
pb
NO
Reference
AQmesh
R² = 0.9509
0
50
100
150
200
250
300
0 50 100 150 200 250 300
FIDA
S
AQMesh
Particle Count
FIDA
S
Gases measured Up to fi ve gases - NO, NO2, O3, CO and SO2
Other measured parameters Pod temperature, atmospheric pressure, rela ve humidity, par cle count, PM2.5#O, PM10
#O
Measurement Range Parameter Units of Measure Range#1 Accuracy#2
NO ppb or μg/m3 <5 to 4,000 ppb + 5 ppb
NO2 ppb or μg/m3 <5 to 4,000 ppb + 5 ppb
O3 ppb or μg/m3 <5 to 1,800 ppb + 5 ppb
CO ppb or μg/m3 <5 to 6,000 ppb + 5 ppb
SO2 ppb or μg/m3 <5 to 10,000 ppb + 5 ppb
Pod Temperature oC -20 to 100 oC + 2 oC
Pressure mbar 500 to 1500 mbar + 5 mb
Humidity %RH 1 to 100 %RH + 5 %RH
Par cle Count Par cles/cm3 0.5 - 30 μm NA
PM2.5 μg/m3
Site Specifi cPM10 μg/m3
Gas Sensor Life Up to 2 years
Power Op ons Primary lithium metal ba eries#3 providing up to 2 years opera on (depending on measurement strategy), mains power op on, external 12V DC power op on, rechargeable ba ery and charger op on
Communica ons Raw data sent via wireless communica ons to remote server for processing - data access contract required
Physical Enclosure: Polyurethane moulded or ABS
Approx. Size: L160 mm x W220 mm x H230 mm (not inc. antenna 180 mm)
Weight: <2 kg
Protec on: IP65
Environmental Opera onal Ranges Temperature: -20 to +40 oC
Humidity: #4 15 to 85% RH
Moun ng Various moun ng accessories available
Measurement Period User changeable variable from 1 minute to 1 hour (1 hour default)
Data Management Web browser based, processing of raw date to give indica ve concentra on, database storage on secure server data access - sta c tables, sta c graphs, data download, mul -user access, password controlled