Safety InstructionsProline Promass 300Class I, Zone 1, AEx/Ex de ia IIC/IIB T6…T1Class I, Zone 1, AEx/Ex d ia IIC/IIB T6…T1Zone 21, AEx/Ex tb IIIC Txx °C
C US
Document: XA01472DSafety instructions for electrical apparatus for explosion-hazardous areas classified according to the National ElectricalCode (NEC) and Canadian Electrical Code (CEC) → 3
XA01472D/06/EN/01.1671327890
Proline Promass 300 XA01472D
Endress+Hauser 3
Proline Promass 300
Table of contents
Associated documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Manufacturer's certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Extended order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety instructions: General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Safety instructions: Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety instructions: Zone 21, Class II, Class III . . . . . . . . . . . . . . . . . . . . . . 10
Temperature tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Connection data: Signal circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
XA01472D Proline Promass 300
4 Endress+Hauser
Associateddocumentation
All documentation is available:• On the CD-ROM supplied (not included in the delivery for all device
versions).• Available for all device versions via:
– Internet: www.endress.com/deviceviewer– Smart phone/tablet: Endress+Hauser Operations App
• In the Download Area of the Endress+Hauser web site:www.endress.com → Download
This document is an integral part of the following OperatingInstructions:
Measuringdevice
Documentation
HART FOUNDATIONFieldbus
PROFIBUS PA ModbusRS485
Promass A 300 BA01482D BA01515D BA01504D BA01493D
Promass E 300 BA01484D BA01517D BA01506D BA01495D
Promass F 300 BA01485D BA01518D BA01507D BA01496D
Promass H 300 BA01486D BA01519D BA01508D BA01497D
Promass I 300 BA01487D BA01520D BA01509D BA01498D
Promass O 300 BA01488D BA01521D BA01510D BA01499D
Promass P 300 BA01489D BA01522D BA01511D BA01500D
Promass Q 300 BA01490D BA01523D BA01512D BA01501D
Promass S 300 BA01491D BA01524D BA01513D BA01502D
Promass X 300 BA01492D BA01525D BA01514D BA01503D
Additional documentation:
Document type Contents Documentation code
Special documentation Remote display and operatingmodule DKX001
SD01763D
Safety instructions Remote display and operatingmodule DKX001:Class I, Zone 1, Zone 21
XA01499D
Control drawing As wanted on thenameplate.
Please note the documentation associated with the device.
Proline Promass 300 XA01472D
Endress+Hauser 5
Manufacturer'scertificates
Certificate numberCSA 16.70087388
Notified bodyCSA: Canadian Standards Association
Extended ordercode
The extended order code is indicated on the nameplate, which is affixedto the device in such a way that it is clearly visible. Additionalinformation about the nameplate is provided in the associatedOperating Instructions.
Structure of the extended order code
* * * * * * _ * * * * * … * * * * * + A*B*C*D*E*F*G*…
Device type Basic specifications Optional specifications
* = Spaceholder: At this position, an option (number or letter) selected fromthe specification is displayed instead of the placeholders.
• Device typeThe device and the device design is defined in the "Device type" section(Product root).
• Basic specificationsThe features that are absolutely essential for the device (mandatoryfeatures) are specified in the basic specifications. The number ofpositions depends on the number of features available. The selectedoption of a feature can consist of several positions.
• Optional specificationsThe optional specifications describe additional features for the device(optional features). The number of positions depends on the numberof features available. The features have a 2-digit structure to aididentification (e.g. JA). The first digit (ID) stands for the featuregroup and consists of a number or a letter (e.g. J = test, certificate).The second digit constitutes the value that stands for the featurewithin the group (e.g. A = 3.1 material (wetted parts), inspectioncertificate).
More detailed information about the device is provided in the followingtables. These tables describe the individual positions and IDs in theextended order code which are relevant to hazardous locations.
XA01472D Proline Promass 300
6 Endress+Hauser
Device type
Item Order code for Selected option Description
1 Instrument family 8 Coriolis flowmeter
2 Sensor A, E, F, H, I, O, P, Q, S, X Sensor type
3 Transmitter 3 Transmitter type:4-wire, compact version
4 Generation index B Platform generation
5, 6 Nominal diameter DN 1 to 350DN 1: 01DN 2: 02toDN 350: 3E, 3F, 3R
Nominal diameter of sensor
Basic specifications
Position 1, 2Order code for"Approval"Optionselected
Position 4, 5Order code for"Output, input 1"Option selected
Type of protection
Transmitter Sensor
C1 BA, BB, GA, LA, NA,RA, SA, MA
Cl.I, Zone 1, AEx/Ex de ia IIB T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db
Cl.I, Zone 1, AEx/Ex ia IIB T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex de ia [ia Ga] IIB
T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db
C2 BA, BB, GA, LA, NA,RA, SA, MA
Cl.I, Zone 1, AEx/Ex de ia IIC T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db
Cl.I, Zone 1, AEx/Ex ia IIC T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex de ia [ia Ga] IIC
T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db
C3 BA, BB, GA, LA, NA,RA, SA, MA
Cl.I, Zone 1, AEx/Ex d ia IIB T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db
Cl.I, Zone 1, AEx/Ex ia IIB T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex d ia [ia Ga] IIB
T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db
C4 BA, BB, GA, LA, NA,RA, SA, MA
Cl.I, Zone 1, AEx/Ex d ia IIC T6...T1 GbZone 21, AEx/Ex tb IIIC T** °C Db
Cl.I, Zone 1, AEx/Ex ia IIC T6...T1GbZone 21, AEx/ Ex ia tb IIIC T** °CDbCA, CB, HA, TA Cl.I, Zone 1, AEx/Ex d ia [ia Ga] IIC
T6...T1 GbZone 21, AEx/Ex tb [ia Da] IIIC T** °C Db
Proline Promass 300 XA01472D
Endress+Hauser 7
Position Order code for Option selected Description
4, 5 Output, input 1 BA 4-20mA HART
CA 4-20mA HART Ex-i
GA PROFIBUS PA
HA PROFIBUS PA Ex-i
MA Modbus RS485
SA FOUNDATION Fieldbus
TA FOUNDATION Fieldbus Ex-i
6 Output, input 2 A W/o;
B 4-20mA
C 4-20mA Ex-i
D Configurable I/O initial setting off
E Pulse/freq./switch output
F Pulse output, phase-shifted
G Pulse/freq./switch output Ex-i
H Relay
I 4-20mA input
J Status input
7 Output, input 3 A W/o;
B 4-20mA
C 4-20mA Ex-i
D Configurable I/O initial setting off
E Pulse/freq./switch output
F Pulse output, phase-shifted
G Pulse/freq./switch output Ex-i
H Relay
I 4-20mA input
J Status input
8 Display, Operation A W/o; via communication
F 4-line, illuminated; touch control
G 4-line, illuminated; touch control + WLAN
M W/o; prepared for remote display DKX001 1)
XA01472D Proline Promass 300
8 Endress+Hauser
Position Order code for Option selected Description
O Separate, with remote display DKX001, 4-line, illuminated; 10m/30ft cable; touch control
17, 18 Device model A1 1
1) DKX001 is approved according to CSA 160686-70030937.
Optional specifications
ID Order code for Option selected Description
Jx Test, Certificate JP Ambient temperature of measuring device –50 °C
Safetyinstructions:General
• Staff must meet the following conditions for mounting, electricalinstallation, commissioning and maintenance of the device:– Be suitably qualified for their role and the tasks they perform.– Be trained in explosion protection.– Be familiar with national regulations (e.g. CEC or NEC).
• Install the device according to the manufacturer's instructions andnational regulations.
• Do not operate the device outside the specified electrical, thermal andmechanical parameters.
• Only use the device in media to which the wetted materials havesufficient durability.
• Refer to the temperature tables for the relationship between thepermitted ambient temperature for the sensor and/or transmitter,depending on the range of application, and the temperature classes.
• Modifications to the device can affect the explosion protection andmust be carried out by staff authorized to perform such work byEndress+Hauser.
• When using in hybrid mixtures (gas and dust occurringsimultaneously), additional measures should be taken. Please seeapproval body.
• Open the device only if one of the following conditions is met:– An explosive atmosphere is not present.– A waiting time of 10 minutes is observed after switching off the
power supply.• Observe all the technical data of the device (see nameplate).
LWARNINGSubstitution of components is not permitted.‣ Substitution of components may impair intrinsic safety.
Proline Promass 300 XA01472D
Endress+Hauser 9
Safetyinstructions:Installation
In the event of potentially explosive vapor/air mixtures, only operatethe device under atmospheric conditions.• Temperature: –20 to +60 °C• Pressure: 80 to 110 kPa (0.8 to 1.1 bar)• Air with normal oxygen content, usually 21 % (V/V)If no potentially explosive mixtures are present, or if additionalprotective measures have been taken, the device may also be operatedunder non-atmospheric conditions in accordance with themanufacturer's specifications.• Continuous service temperature of the connecting cable:
–40 to +80 °C (–50 to +80 °C for optional specification, ID Jx (Test,Certificate) = JP); but at least according to the service temperaturerange of the application taking into account additional influences ofthe process conditions (Ta,min and Ta,max + 20 K).
• Only use certified cable entries suitable for the application. Observeselection criteria as per CEC or NEC.
• The following applies when connecting the transmitter with aconnection compartment in Ex d:Only use separately certified cable and wire entries (Ex d IIC) whichare suitable for operating temperatures up to 85 °C and for IP 67protection. If using conduit entries, the associated sealingmechanisms must be mounted directly on the housing.
• The following applies when connecting the transmitter with aconnection compartment in Ex e:Only use separately certified cable and wire entries and sealing plugs(Ex e II) which are suitable for operating temperatures up to 85 °Cand for IP 67 protection. The cables must be routed such that they aresecurely seated, and sufficient strain relief must be ensured.
• When the measuring device is connected, attention must be paid toexplosion protection at the transmitter.
• In potentially explosive atmospheres:– Do not disconnect the electrical connection of the power supply
circuit when energized.– Do not open the connection compartment cover when energized.
• Install the transmitter circuit wiring according to Canadian ElectricalCode (CEC) respective National Electrical Code (NEC) using threadedconduit or other wiring methods in accordance with articles 500 to510.
• When connecting through a conduit entry approved for this purpose,mount the associated sealing unit directly at the housing.
• Seal unused entry glands with approved sealing plugs that correspondto the type of protection. The plastic transport sealing plug supplieddoes not meet this requirement and must therefore be replacedduring installation.
• Only use certified sealing plugs. The metal sealing plugs suppliedmeet this requirement.
XA01472D Proline Promass 300
10 Endress+Hauser
Intrinsic safety• Observe the guidelines for interconnecting intrinsically safe circuits
(e.g. , Proof of Intrinsic Safety).• When the intrinsically safe Ex ia circuits of the device are connected
to certified intrinsically safe Ex ib circuits with explosion group IIC orIIB:The type of protection changes to Ex ib IIC or Ex ib IIB.
• The device can be connected to the remote display and operatingmodule DKX001 with Ex ia explosion protection; refer to the SpecialDocumentation and Ex documentation.
If the remote display and operating module DKX001 is used, theinternal display unit must be removed from transmitter!
Potential equalization• Integrate the device into the local potential equalization .• If the ground connection has been established via the pipe as
specified, it is also possible to integrate the sensor into the potentialequalization system via the pipe.
Safetyinstructions: Zone21, Class II, ClassIII
• To ensure dust-tightness, securely seal the transmitter housing, cableentries and sealing plugs.
• Only open the transmitter housing briefly, ensuring that no dust ormoisture enters the housing.
• Only use certified cable entries. The metal cable entries supplied meetthis requirement.
• When connecting the transmitter to the remote display andoperating module DKX001: Connection values → 21
Temperaturetables
Ambient temperatureMinimum ambient temperature:• Ta = –40 °C• Optional specification, ID Jx (Test, Certificate) = JP
Ta = –50 °CMaximum ambient temperature:Ta = +60 °C depending on the medium temperature and temperatureclass
Proline Promass 300 XA01472D
Endress+Hauser 11
Medium temperatureMinimum medium temperature:• Promass A, F, H, I, P, Q, S, X:
Tm = –50 °C• Promass E, O: Tm = –40 °C• Promass F, Q with cryogenic temperature version (order code for
"Measuring tube material", Option LA):Tm = –200 °C
Maximum medium temperature:• Tm for T6…T1 depending on the maximum ambient temperature Ta• ( ) = The maximum permitted medium temperatures in brackets only
apply if the sensor is installed in such a way that the transmitter isnot mounted above the sensor and free convection can occur on allsides.
Compact version
Without thermal insulation
Promass A
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
1…4 50 205 50 95 130 150 205 205
60 – 95 130 150 205 205
Promass E
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8…50 50 150 50 100 130 130 150 150
55 – 80 100 130 150 150
60 – (80) (100) (130) (150) (150)
80 50 150 50 75 110 150 150 150
55 – 75 110 150 150 150
60 – (75) (110) (150) (150) (150)
XA01472D Proline Promass 300
12 Endress+Hauser
Promass F
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
08…15 50 150 50 95 130 150 150 150
60 – 95 130 150 150 150
50 240 50 95 130 160 240 240
60 – 95 130 160 (240) (240)
50 150 2) 50 95 100 150 150 150
60 – 95 100 150 (150) (150)
25…50 50 150 50 95 130 150 150 150
60 – 95 130 150 150 150
50 240 50 95 130 160 240 240
60 – 95 130 160 (240) (240)
50 150 2) 50 95 100 150 150 150
60 – 95 100 150 (150) (150)
50 350 45 85 115 175 275 350
60 – 85 115 175 275 350
80 50 350 45 85 115 175 275 350
60 – 85 115 175 275 350
80…250 50 150 50 75 110 150 150 150
60 – 75 110 150 150 150
50 240 50 75 110 170 240 240
60 – 75 110 170 (240) (240)
50 150 2) 50 75 110 150 150 150
60 – 75 110 150 150 150
1) Maximum temperature range see nameplate2) Cryogenic temperature version: Tm = -196…150 °C
Promass H
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 50 150 1) 50 65 100 150 150 150
60 – 65 100 150 150 150
Proline Promass 300 XA01472D
Endress+Hauser 13
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 50 205 2) 50 65 100 160 205 205
60 – 65 100 160 205 205
15…50 50 150 1) 50 75 115 150 150 150
60 – 75 115 150 150 150
15…50 50 205 2) 50 75 115 180 205 205
60 – 75 115 180 205 205
1) Measuring tube material tantalum2) Measuring tube material zirconium
Promass I
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8, 15,15FB, 25
50 150 50 95 130 150 150 150
60 – 95 120 (150) (150) (150)
25FB, 40,40FB, 50
50 150 50 85 120 150 150 150
60 – 85 120 (150) (150) (150)
50FB, 80 50 150 50 85 120 150 150 150
60 – 85 120 (150) (150) (150)
FB = Full bore
Promass O
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
80 … 250 50 205 50 75 110 170 205 205
55 – 75 110 170 205 205
60 – 75 110 170 (205) (205)
XA01472D Proline Promass 300
14 Endress+Hauser
Promass P
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 45 150 45 65 100 150 150 150
60 – 65 100 150 150 150
45 205 45 65 100 160 205 205
60 – 65 100 160 205 205
15…50 50 150 50 75 115 150 150 150
60 – 75 115 150 150 150
50 205 50 75 115 180 205 205
60 – 75 115 180 205 205
1) Maximum temperature range see nameplate
Promass Q
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
25 … 100 50 205 50 75 110 160 205 205
60 – 75 110 160 205 205
25 … 100 50 150 2) 50 75 110 150 150 150
60 – 75 110 150 150 150
1) Maximum temperature range see nameplate2) Cryogenic temperature version: Tm = -196…150 °C
Promass S
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 45 150 45 65 100 150 150 150
60 – 65 100 150 150 150
15…50 50 150 50 75 115 150 150 150
60 – 75 115 150 150 150
Proline Promass 300 XA01472D
Endress+Hauser 15
Promass X
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
350 50 180 50 90 120 170 180 180
55 – 90 120 170 180 180
60 – (90) (120) (170) (180) (180)
With thermal insulation according to Endress+Hauser specificationsFor information on the thermal insulation of the device, see the"Thermal insulation" section of the Operating Instructions
Promass A
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
1…4 50 205 50 95 130 150 205 205
55 – (95) (130) (150) (205) (205)
Promass E
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8…50 50 150 50 100 130 130 150 150
55 – (100) (130) (130) (150) (150)
80 45 150 50 75 110 150 150 150
50 – 75 110 150 150 150
55 – (75) (110) (150) (150) (150)
Promass F
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
08…15 50 150 50 95 130 150 150 150
60 – 95 110 (150) (150) (150)
50 240 50 95 130 160 240 240
XA01472D Proline Promass 300
16 Endress+Hauser
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
55 – (95) (130) (160) (240) (240)
60 – 95 110 110 110 110
25…50 50 150 50 95 130 150 150 150
60 – 95 110 (150) (150) (150)
50 240 50 95 130 160 240 240
55 – (95) (130) (160) (240) (240)
60 – 95 110 110 110 110
50 350 45 85 115 175 275 350
60 – 85 115 175 275 350
80 50 350 45 85 115 175 275 350
60 – 85 115 175 275 350
80…250 50 150 50 75 110 150 150 150
60 – 75 110 (150) (150) (150)
50 240 50 75 110 170 240 240
55 – (75) (110) (170) (240) (240)
60 – 75 110 110 110 110
1) Maximum temperature range see nameplate
Promass H
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 50 150 1) 50 65 100 150 150 150
55 – (65) (100) (150) (150) (150)
60 – 65 100 100 100 100
8 50 205 2) 50 65 100 160 205 205
55 – (65) (100) (160) (205) (205)
60 – 65 100 100 100 100
15…50 50 150 1) 50 75 115 150 150 150
55 – (75) (115) (150) (150) (150)
60 – 75 115 115 115 115
Proline Promass 300 XA01472D
Endress+Hauser 17
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
15…50 50 205 2) 50 75 115 180 205 205
55 – (75) (115) (180) (205) (205)
60 – 75 115 115 115 115
1) Measuring tube material tantalum2) Measuring tube material zirconium
Promass I
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8, 15,15FB, 25
50 150 50 95 130 150 150 150
60 – 95 120 (150) (150) (150)
25FB, 40,40FB, 50
50 150 50 85 120 150 150 150
60 – 85 120 (150) (150) (150)
50FB, 80 50 150 50 85 120 150 150 150
60 – 85 120 (150) (150) (150)
FB = Full bore
Promass O
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
80…250 50 205 50 75 110 170 205 205
55 – (75) (110) (170) (205) (205)
Promass P
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 45 150 45 65 100 150 150 150
50 – 65 100 150 150 150
60 – 65 100 125 (150) (150)
45 205 45 65 100 160 205 205
XA01472D Proline Promass 300
18 Endress+Hauser
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
50 – 65 100 160 205 205
60 – 65 100 115 (205) (205)
15…50 50 150 50 75 115 150 150 150
60 – 75 115 125 (150) (150)
50 205 50 75 115 180 205 205
60 – 75 115 (150) (150) (150)
1) Maximum temperature range see nameplate
Promass Q
DN Ta[°C]
Tm, max 1)
[°C]Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
25…100 50 205 50 75 110 160 205 205
55 – (75) (110) (160) (205) (205)
25…100 50 150 2) 50 75 110 150 150 150
55 – (75) (110) (150) (150) (150)
1) Maximum temperature range see nameplate2) Cryogenic temperature version: Tm = -196…150 °C
Promass S
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
8 45 150 45 65 100 150 150 150
50 – 65 100 150 150 150
60 – 65 100 125 (150) (150)
15…50 50 150 50 75 115 150 150 150
60 – 75 115 125 (150) (150)
Proline Promass 300 XA01472D
Endress+Hauser 19
Promass X
DN Ta[°C]
Tm, max[°C]
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
350 50 180 50 90 120 170 180 180
55 – (90) (120) (170) (180) (180)
With thermal insulation without Endress+Hauser specifications
1
A B
A0031198
1 Position of the reference point for temperature measurement
A Standard versionB Extended temperature version, cryogenic temperature version, high-
temperature version1 Reference point
Tm [°C]
T6[85 °C]
T5[95 °C]
T4[130 °C]
T3[195 °C]
T2[290 °C]
T1[440 °C]
59 72 75 76 77 77
XA01472D Proline Promass 300
20 Endress+Hauser
Connection data:Signal circuits
The following tables contain specifications which are dependent on thetransmitter type and its input and output assignment. Compare thefollowing specifications with those on the nameplate of the transmitter.
Terminal assignment
Transmitter: supply voltage, input/outputs
HART
Supply voltage Input/output 1 Input/output 2 Input/output 3
1 (+) 2 (–) 26 (+) 27 (–) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminalcover.
FOUNDATION Fieldbus
Supply voltage Input/output 1 Input/output 2 Input/output 3
1 (+) 2 (–) 26 (A) 27 (B) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminalcover.
PROFIBUS PA
Supply voltage Input/output 1 Input/output 2 Input/output 3
1 (+) 2 (–) 26 (B) 27 (A) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminalcover.
Modbus RS485
Supply voltage Input/output 1 Input/output 2 Input/output 3
1 (+) 2 (–) 26 (B) 27 (A) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminalcover.
Proline Promass 300 XA01472D
Endress+Hauser 21
Remote display and operating module DKX001
1
2
3
81
Vcc
82
Gn
d
83
A
84
B
81
Vcc
82
Gn
d
83
A
84
B
A0027518
1 Remote display and operating module DKX0012 Connecting cable3 Measuring device
Safety-related values
Order code for"Output; input 1"
Output type Safety-related values"Output; input 1"
26 (+) 27 (–)
Option BA Current output4 to 20 mA HART
UN = 30 VUM = 250 V
Option GA PROFIBUS PA UN = 32 VUM = 250 V
Option MA Modbus RS485 UN = 30 VUM = 250 V
Option SA FOUNDATIONFieldbus
UN = 32 VUM = 250 V
Order code for"Output; input 2";"Output; input 3"
Output type Safety-related values
Output; input 2 Output; input 3
24 (+) 25 (–) 22 (+) 23 (–)
Option B Current output4 to 20 mA
UN = 30 VUM = 250 V
Option D User configurableinput/output
UN = 30 VUM = 250 V
XA01472D Proline Promass 300
22 Endress+Hauser
Order code for"Output; input 2";"Output; input 3"
Output type Safety-related values
Output; input 2 Output; input 3
24 (+) 25 (–) 22 (+) 23 (–)
Option E Pulse/frequency/switch output
UN = 30 VUM = 250 V
Option F Double pulse output UN = 30 VUM = 250 V
Option H Relay output UN = 30 VIN = 100 mA DC/500 mA ACUM = 250 V
Option I Current input4 to 20 mA
UN = 30 VUM = 250 V
Option J Status input UN = 30 VUM = 250 V
Intrinsically safe values
Order code for"Output; input 1"
Output type Intrinsically safe values"Output; input 1"
26 (+) 27 (–)
Option CA Current output 4-20mA HART Ex i
Ui = 30 Vli = 100 mAPi = 1.25 WLi = 0Ci = 0
Option HA PROFIBUS PA Ex i Ex iaUi = 30 Vli = 570 mAPi = 8.5 WLi = 10 µHCi = 5 nF
Option TA FOUNDATIONFieldbus Ex i
Ex iaUi = 30 Vli = 570 mAPi = 8.5 WLi = 10 µHCi = 5 nF
Proline Promass 300 XA01472D
Endress+Hauser 23
Order code for"Output; input 2";"Output; input 3"
Output type Intrinsically safe values
Output; input 2 Output; input 3
24 (+) 25 (–) 22 (+) 23 (–)
Option C Current output 4-20mA Ex i
Ui = 30 Vli = 100 mAPi = 1.25 WLi = 0Ci = 0
Option G Pulse/frequency/switch output Ex i
Ui = 30 Vli = 100 mAPi = 1.25 WLi = 0Ci = 0
Remote display DKX001
Basic specification,position 1, 2Approval
Terminalassignment
Basic specification, position 8Display; OperationOption O
Option C2, C4 81, 82, 83, 84
A connecting cable with the value L/R ≤24 µH/Ω and Ccable ≤ 1 000 nF must be usedfor the version for connecting to the remotedisplay DKX001 or ODKX001.