Screenshots January 27 SP-AMS
Laser
Menus
Menu 1, 2, 3
Menu 4 Laser On high m/z
Voltages For V mode (positive ion mode!!)
Analog/Digital Tab (laser off)
DIO5 =1, DIO6=0
Analog/Digital Tab Laser on
DIO5 =0, DIO6=0
Pumps Inlet Valve Closed
Lens P = 0
Back P = 0.78
Pumps Inlet Valve Open
Lens P = 1.28
Back P = 4.99
Heater and General Info
• HB V mode = 40.5 in the menu, 38 on readout
• Heater T Readout = 490C, 4.3V, 1.28A
• Filament emission = 1.25 mA
• Flow rate 1.35
TPS V menus(last saved after Bill retuned it)
120126_v_41.set_f3.set
120126_v_41.set_f3.set
Bitwise V
TD settings
Laser on - V
Laser off
Vmode m/z calibration
W mode first tuning
Jan 28, 6 pm – N winds, fullerenes from the power plant ?
Servo check