MAIN FEATURES AND BENEFITS
> Purity PAR-H2 99.9999%, O2 < 0.1ppm, H20 < -70°C.
> Purity WM-H2 99.99999%, O2 < 0.1 ppm, H20 < -70°C.
> No maintenance to replace the dessicant cartridge.
> 12 bar pressure available, 16 bar in option.
High level of operator safety• Unique 9 stage, fail safe, explosion protection system.• Automatic internal/external H2 leak detection.• Patented gas/water separator electronically controlled.• H2 sensor in option for carrier gas use.
Good user interfaceLCD touch screen showing in real time H2 pressure, H2 fl ow rate, water quality, water level, and status of the system with auto-diagnostic in case of alarms.
Class leading PEM cell• Continuous monitoring of vital parameters.• Unique cell construction and water quality management ensure reliability and longevity of the cell.• 2 years cell warranty.
Unique features• Canister fi lling mode in serie to fi ll a large canister with the
pressure of 16 bar to allow to have a cartridge that can be used for portable GC in the fi eld.
• USB port available allowing the units to be connected to the PC in order to upgrade or train the user by internet.
• Auto refi ll water tank in standard for most of the range.• Remote PC monitoring in standard via RS232 or RS485 to
interface the unit with customer’s PC software. • Capabilities allowing to work in parallel mode.
THE BENEFITS TO USEHYDROGEN AS CARRIER GAS
> Improved chromatograph result• Using H2 as a GC carrier gas increases analysis
speed and sensitivity when compared to helium and nitrogen.
• When H2 is used as a carrier gas, as it requires lower elution temperatures and thus improves the column.
• Using hydrogen as GC carrier gas signifi cantly increases analysis speed and separation effi ciency compared to Helium. Gas costs and downtime are also reduced due to the change from a cylinder based supply.
> Get all the benefi ts of the hydrogen generator as carrier gas... without taking any risk!
The sensor can be connected directly with the Alliance hydrogen generators. It continuously monitors the atmosphere in the GC oven and transmits the data to the hydrogen generator. When the hydrogen concentration detected exceeds the threshold of safety, the hydrogen generator will be automatically blocked and reported an audible and visual alarms. The oven is immediately shut off and the cooling fl ap opens.
THE PAR-H2 AND WM-H2 series provide a higher purity grade of hydrogen and can be used for GC and GC/MS carrier gas application as well as ICP-MS reaction gas.
THE PAR-H2 uses a single column dryer with programmable automatic regeneration via an integrated clever calendar.
THE WM-H2 uses an exclusive double gas column dryer regeneration system thus eliminating all down time for maintenance that is typical of other systems on the market, guaranteeing the best hydrogen purity at all times.
HIGH PURITY HYDROGEN GENERATORSFOR GC CARRIER GAS AND GCMS APPLICATIONS
2
SERIE PAR-H2 / WM-H2
H2 Sensor
APPLICATIONS : GC-FID / NPD / FPD / TCD / GC-carrier gas / GCMS / ICPMS / colision cell / THA
MODELS Flowrateml/min Purity % H2 dyrer Pressure Water reservoir capacity
with auto refi lled included Communication VoltageVAC
Electrical Consumption
DimensionsW x H x D cm
WeightKg
PAR-H2 120120
99.9999 Programmable autoregeneration cartridge
12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 280 W 30 x43 x 43
15
WM-H2 120 99.99999 Auto-drying technology 15
PAR-H2 180180
99.9999 Programmable autoregeneration cartridge
12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 280 W 30 x43 x 43
15
WM-H2 180 99.99999 Auto-drying technology 15
PAR-H2 260260
99.9999 Programmable autoregeneration cartridge
12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 280 W 30 x43 x 43
15
WM-H2 260 99.99999 Auto-drying technology 15
PAR-H2 400400
99.9999 Programmable autoregeneration cartridge
12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 280 W 30 x43 x 43
15
WM-H2 400 99.99999 Auto-drying technology 15
PAR-H2 500500
99.9999 Programmable autoregeneration cartridge
12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 450 W 30 x43 x 43
25
WM-H2 500 99.99999 Auto-drying technology 25
PAR-H2 650650
99.9999 Programmable autoregeneration cartridge
12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 450 W 30 x43 x 43
25
WM-H2 650 99.99999 Auto-drying technology 25
WM-H2 800 800 99.99999 Auto-drying technology 12 bar 2.3 L RS232, RS485,USB port in serie
90 - 240 450 W 30 x43 x 43 25
WM-H2 900 900 99.99999 Auto-drying technology 12 bar 2.3 LRS232, RS485,
USB port in serie90 - 240 560 W 30 x43 x 43 25
WM-H2 1000 1000 99.99999 Auto-drying technology 12 bar 2.3 L RS232, RS485,USB port in serie 90 - 240 560 W 30 x43 x 43 25
WM-H2 1200 1200 99.99999 Auto-drying technology 12 bar 2.3 L RS232, RS485,USB port in serie
90 - 240 560 W 30 x43 x 43 25
3
One step ahead !One step ahead !
H2
MAIN FEATURES AND BENEFITS
> Purity 99.9995%, O2 < 1ppm, H20 < -55°C.> 7 bar pressure for LC-H2, 10 bar pressure for ND-H2.> Limited routine maintenance : Replacing / Regenerating of desiccant cartridge, water fi lter and deioniser bag.> H2 cartridge saturation sensor correlated with the fl ow produced.> Stackable with Zero Air generator 2 solutions in 1! Forming a FID gas station.
High level of operator safety• Unique 9 stage, fail safe, explosion protection system.• Automatic internal/external H2 leak detection.• Patented gas/water separator electronically controlled.• H2 sensor in option for carrier gas use.
Class leading PEM cell• Continuous monitoring of vital parameters.• Unique cell construction and water quality management ensure reliability and longevity of the cell.• 2 years cell warranty.
Good user interfaceLCD touch screen showing in real time H2 pressure, H2 fl ow rate, dessicant cartridge saturation %, water quality, water level, and status of the system with auto-diagnostic in case of alarms.
Unique features• USB port available allowing the units to be connected to the
PC in order to upgrade or train the user by internet.• Auto refi ll water tank in standard for most of the range.• Remote PC monitoring in standard via RS232 or RS485 to
interface the unit with customer’s PC software.• Capabilities allowing to work in parallel mode.
HYDROGEN GENERATORS FOR GC FUEL GAS APPLICATIONS
4
Internal water tank
Valve block Pump Filter
Gas liquid separatorSecurity
valve
Pressure regulatorr
O2 VENT H2 VENT H2 OUTLET
H2 RELIEF
External water
H2O+H2
H2+H2OO2+H2O
THE BENEFIT TO USE H2AS FUEL GAS
• Eliminates inconvenient and dangerous hydrogen cylinders from the laboratory.
• Increases the accuracy of analyses and reduces the cleaning requirement of the detector.
THE LC-H2 AND ND-H2 series are ideal for supplying hydrogen fuel gas to all known combustion detectors used routinely in GC and THA. These series use a desiccant cartridge which can be refi ll or replaced by a new one.
SERIE LC-H2 / ND-H2
LC-H2 SerieND-H2 SerieND-H2
Series
Stackablewith Zero Air Generator :
2 solution in 1!Forming a FID gas station.
O2 VENT H2 VENT
LCND-H2 SerieND-H2
Serieses
StackableStacAir Generator :Air Generator :ero Aero Ation in 1!soluID gas station.g a F
APPLICATIONS : GC-FID / NPD / FPD / TCD / THA
MODELS Flowrateml/min Purity % H2 dyrer Pressure Water reservoir capacity
with auto refi lled included Communication VoltageVAC
Electrical Consumption
DimensionsW x H x D cm
WeightKg
LC-H2 100 100 99.9995 Dessicant cartridge 7 bar 2.1 L RS232 in serieRS485 in option 90 - 240 80 W 25 x 30 x 32 10
LC-H2 140 140 99.9995 Dessicant cartridge 7 bar 2.1 L RS232 in serieRS485 in option 90 - 240 100 W 25 x 30 x 32 10
LC-H2 180 180 99.9995 Dessicant cartridge 7 bar 2.1 L RS232 in serieRS485 in option 90 - 240 120 W 25 x 30 x 32 10
ND-H2 120 120 99.9995 Dessicant cartridge 10 bar 2.3 L RS232 in serieRS485 in option
90 - 240 100 W 30 x43 x 43 15
ND-H2 180 180 99.9995 Dessicant cartridge 10 bar 2.3 L RS232 in serieRS485 in option
90 - 240 125 W 30 x43 x 43 15
ND-H2 260 260 99.9995 Dessicant cartridge 10 bar 2.3 L RS232 and RS485in serie 90 - 240 185 W 30 x43 x 43 15
ND-H2 400 400 99.9995 Dessicant cartridge 10 bar 2.3 L RS232 and RS485in serie
90 - 240 220 W 30 x43 x 43 15
ND-H2 500 500 99.9995 Dessicant cartridge 10 bar 2.3 L RS232 and RS485in serie 90 - 240 240 W 30 x43 x 43 25
5
One step ahead !One step ahead !
H2
APPLICATIONS : GC-FIDFID GAS STATION FOR GC FUEL GAS APPLICATIONS
FID STATION* H2/ Zero Airfl ow rate cc/min H2/AIR purity H2 dyrer
H2 / AIRDeliverypressure
Air inlet pressureAnd air quality required
Water reservoircapacity Communication Voltage
VACElectrical
ConsumptionDimensionsW x H x D cm
WeightKg
FID-H2/AIR-100-1500 100/1500
> 99.9995%H20 Dewpoint
< -55°CO2 < 1 ppm
CH4 < 0,05 ppm
Dessicantcartridge
7 barg / Upto 6.5 barg
Maximum 7 barMax.inlet hydrocarbons
content < 100 ppmwater dewpoint < -20°C
1.2 LRS232 in serie
RS485 in option90 - 240 180 W 28 x 43 x 31 12
FID-H2/AIR-140-1500 140/15007 barg / Upto 6.5 barg
1.2 L RS232 in serieRS485 in option
90 - 240 200 W 28 x 43 x 31 12
FID-H2/AIR-180-1500 180/1500 7 barg / Upto 6.5 barg
1.2 L RS232 in serieRS485 in option
90 - 240 220 W 28 x 43 x 31 12
6
MAIN FEATURES AND BENEFITS
> Purity FID-H2/AIR 99.9995%, H20 Dewpoint < -55°C, O2 < 1 ppm. CH4 < 0.05 ppm. CO < 0.05 ppm.
> External clean and dry air compressor required at maxi. 7 bar.
> Save space on the bench.
IMPROVED CHROMATOGRAPH RESULT
• The reduction of hydrocarbons on zero air part, including methane to < 0.05 ppm decreases the background noise level and gives the baseline much better stability, considerably increasing detector sensitivity and ensuring precise analytical results.
• The use of hydrogen as Fuel Gas increases the accuracy of analysis and reduces the cleaning requirement of the detector.
THE H2/AIR FID STATION combines the hydrogen serie LC-H2 and Zero air serie ZA generators in one unit; Hydrogen gas is produced from deionised water using a Proton Exchange Membrane Technology. Zero air is produced by purifying compressed air sourced from the air network to a total hydrocarbon concentration of < 0.05 ppm (measured as methane).
SERIE FID-H2 / AIR
*ZA serie needs to be connected to an external clean and dry compressed Air source.
7
One step ahead !One step ahead !RACKABLE H2 GENERATORFOR GC FUEL GAS AND CARRIER GAS APPLICATIONS
THE BENEFIT TO USE H2AS FUEL GAS
Eliminates inconvenient and dangeroushydrogen cylinders from the laboratory. Increases the accuracy of analyses and reduces the cleaning requirement of the detector.
THE BENEFIT TO USE H2AS CARRIER GAS
The use of hydrogen as a carrier gas allows lower temperature elution, thus extending the life of the chromatograph column.Hydrogen as a carrier gas is faster and more sensitive than the more-expensive helium. Run time savings of 25% to 35% without a decline in resolution. Save space on the bench.
MAIN FEATURES AND BENEFITS
TWO MODELS AVAILABLE :
> RACK-ND-H2 Simple column dryer located at the rear of the unit.> RACK-WM-H2 No desiccant cartridge maintenance is required : used a double column dryer with automatic regeneration.
> Unique Features• Patented gas/water separator electronically controlled.• LCD touch screen showing in real time H2 pressure, H2 fl ow rate, dessicant
cartridge saturation %, water quality, water level, and status of the system with auto-diagnostic in case of alarms.
• Canister fi lling mode in serie to fi ll a large canister with the pressure of 16 bar to allow to have a cartridge that can be used for portable GC in the fi eld.
• USB port available allowing the units to be connected to the PC in order to upgrade or train the user by internet.
• Auto refi ll water tank in standard for most of the range.• Remote PC monitoring in standard via RS232 or RS485 to interface the unit
with customer’s PC software.• Capabilities allowing to work in parallel mode.• Internal water tank level of 1,2L electronically controlled (when needed it
is fi lled from external tank automatically by only a single tube).
APPLICATIONS : GC-FID / NPD / FPD/ TCD / THA AND CARRIER GAS
MODELS H2 fl ow ratecc/min H2 purity H2 dyrer
H2Deliverypressure
Waterreservoircapacity
ExternalWater tankoptionnal
Communication VoltageVAC
ElectricalConsumption
DimensionsW x H x D cm
WeightKg
RACK-ND-H2 Séries 120,180, 260> 99.9995%
H20 Dewpoint< -55°C, / O2 < 1 ppm
Dessicantcartridge 10 bar 1.2 L 5,10 L RS232 in serie
RS485 in option 90 - 240From 200W
to 800Wdepend
of the model
Standard 19”Rack 4U-deep
40 cmFrom 15
to 25depends
of themodelRACK-WM-H2 Series
120,180, 260,400, 500, 650,800, 900, 1000
> 99.99999%H20 Dewpoint
< -70°C, / O2 < 0,1 ppm
Exclusive doublegas column dryer
regeneration
12 bar(16 bar onrequest)
1.2 L 5,10 L RS-232, RS-485,USB, LAN in serie 90 - 240
Standard 19”Rack 4U-deep
40 cm
SERIE RACK-H2 H2
H2-SENSOR FOR YOUR GC-MS / GC! GET ALL THE BENEFITS OF THE HYDROGEN GENERATOR AS CARRIER GAS...
8
TWO OPERATING MODES :
#1 > The sensor can be connected directly with the Alliance hydrogen generators. It continuously monitors the atmosphere in the GC oven and transmits the data to the hydrogen generator.
When the hydrogen concentration detected exceeds the threshold of safety, the hydrogen generator will be automatically blocked and reported an audible and visual alarms.
It can automatically switch to an inert gas if the gas switching valve is used.
Eliminate the risks in using Hydrogen as a Carrier Gas in GC AnalysisHydrogen has long been considered as the best carrier gas for gas chromatography. In many cases hydrogen has become the carrier gas of choice since it results in fast analysis, high effi ciency, reduced costs, and prolonged column life.
Fast analysisThe diffusion rate of hydrogen and helium are roughly the same (both 3-4 times faster than N2), but hydrogen is half as viscous as helium and therefore the linear gas velocity is higher and retention times are shorter in isothermal analysis.
High effi ciencyHydrogen has the fl attest Van Deemter curve. Compared to helium and nitrogen, hydrogen needs the lowest plate number to achieve the same resolution over a large range of linear velocity
Prolonged column lifeFor some applications, a temperature program can be used to speed up the analysis when using helium as a carrier gas, but this may result in a shorter lifetime of the column due to higher temperatures.
Reduced costsHelium, a rival to Hydrogen has its advantages as a carrier gas for GC; but it also has key disadvantages which are cost and availability. A tank of GC quality hydrogen is approximately 3 times less expensive than its helium equivalent. The price disparity between the two will not improve as the existing helium reserves are drying up and the demand is increasing across different industries. The use of a hydrogen generator also provides long term cost savings. A hydrogen generator allows for the production of the gas on an as needed basis; which avoids the costs associated with storing gas.
The advantages of hydrogen are clear; but it does have one really big disadvantage! Hydrogen is an explosive gas! An undetected gas leak can occur with a broken column or a leaking connection. The danger is that an undetected gas leak could result in an explosion in the GC oven placing laboratories and their personnel at risk.
The ability to safely detect hydrogen leaks in the GC oven is critical to any laboratory using hydrogen as a carrier gas. The new H2 Sensor ensures the safe use of hydrogen in GC analysis. It does this by constantly monitoring the H2 concentrations in the GC oven and automatically switching to an inert gas when typically 25% LEL is reached; this important feature eliminates the risks and at the same time ensures safety.
Alliance Hydrogen Sensor
INER
T GAS
H2 G
AS
SENSOR CALIBRATIONThe H2 sensor needs to be calibrated
annualy by using a calibration gas bottle
Gas Switching ValveOPT.H2.SENS. VB
Alliance
... WITHOUT TAKING THE RISK!
9
One step ahead !One step ahead !
#2 > The sensor can work as a standalone mode by using an external box and a gas switching valve.
The sensor is installed in the GC oven for continuous monitoring of H2 concentrations in the oven. The measured gas concentration is shown on the LCD screen of the external controller. When the hydrogen concentration reaches the user defi ned level, typically between 25% and 50% LEL (equal to 1%-2% by Vol H2), the LCD screen starts fl ashing, an acoustic signal is transmitted and the carrier gas will automatically be switched to an inert gas. A SMS alarm text message can be sent as well.
The external box is able to support up to 4 sensors.
H2