Diapositive 1ψs
4 ( , )[1 ( , )] ( ) DOS
t s s t bias ts t s bias t s
eI f E T f E eV T M E E eVπ δ→ = − + − −∑
occupied sample states
unoccupied tip states
DOS: Density of State
Assume tip and sample described by two infinite plate
electrodes
Sample work function
ψt Wave function
t s bias ts t s bias t s
→ →= − =
− + − −∑
dE E dE Eρ ρ→ →∑ ∑∫ ∫
−∞
2
0
4 ( ) ( ) ( ) eV
S F t F eI E eV E M dπ ρ ε ρ ε ε ε= + + +∫
And for small Vbias
eI V E E Mπ ρ ρ=
2 24 ( ) ( )S F T F
dI e E eV E M dV
π ρ ρ= +
Moving to the continuous
The tunneling current is a function of the tip and sample density
of state close to the Fermi level
( ) ( ) ( )2 0 ,
t t
( ) ( ) ( )0 s
d z s z e κψ ψ − −=
Assuming the electron wave functions for tip and sample described
by plane waves, the quantum mechanics predicts an exponential decay
in the vacuum gap following
2( ) ( ) z S F T FI V E E e κρ ρ −∝
STM vertical resolution
Φ in [eV] d in [Å]
1.025 1 100dI e I
ΦΔΔ ≈ − ≈
1.025( ) ( ) d S F T FI V E E eρ ρ − Φ∝
N.B.: if the tip (sample) is an insulator -> ρt(EF) (ρs(EF)) = 0
implying I = 0
The STM is useless in this case
Typically Φ = 5 eV
Easy to measure ΔI/I = 0.1 -> Δd = Φ-0.5 ΔI/I = 0.05 Å
Atomic step Δd = 2 Å
J. Tersoff and D.R. Hamann. Phys. Rev. B. 31, 805 (1985).
For small kT, f(E) −> step function; for small V −> ρ =
constant, one gets
( ) ( )0 0
r r κ κ
ρ ψ δ = −
Local density of states
2 2 0 0( , ) ( ) ( , ) z
t F s FI r V R E r E Ve κρ ρ −≈
The tunneling current depends on the tip position (ρs(r0,EF))
respect to the surface
Tersoff calculated a resolution function which is a Gaussian with
an RMS width of
2 R dL κ +
= atomic tip R = 2 Å d = 2-3 Å L ~ 1 Å
Actually can be better depending on the tip wave function
>0.5 M€
10 K€
Table-top STM
UHV STM
Tip preparation
After electro-erosion
1) Imaging and topography
Constant current image of a stepped Pt surface covered by 1
monolayer Ag
(I = 2.7 nA, Vbias = 10 mV) P. Gambardella, EPFL PhD thesis
(2000)
Basic concepts of STM imaging. Tunneling current is exponentially
dependent on distance between tip and sample (< 1 nm). As the
tip is moved from x1 to x2, the current increases as the
tip-to-sample distance decreases due to the change in sample
height. The increase in current causes the control loop to move the
tip away from the sample until the error signal is again zero.
Recording the value of the tip height (z) as a function of position
(x,y) allows the 3-dimensional topography to be
reconstructed.
feedback
Å
Å
Å
Å
Warning !!!
1.025( ) ( ) d S F T FI V E E eρ ρ − Φ∝
Remember: the tunneling current is a measure of the Local Density
of States (at the Fermi level) of tip and sample
Technique enabling atomic-scale imaging of insulating surfaces. The
tip is mounted on a cantilever and is brought into contact with the
sample surface. The force on the cantilever is related to its
deflection via Hooke’s law: F = -kx, where k is the spring constant
of the cantilever and x is the deflection.
Atomic Force Microscopy (AFM)
(1) Non-contact AFM Van der Waals attraction 10-100 Å tip-surface
separation (2) Contact AFM e--e- repulsion <5 Å tip-surface
separation (3) Intermittent contact AFM (tapping mode AFM) 5-20 Å
tip-surface separation
The attractive or repulsive force between the tip and the sample
causes a deflection of the cantilever towards or away from the
sample. The deflection is measured by a laser beam directed at the
back of the cantilever. As the cantilever deflects, the angle of
the reflected beam changes, and the spot falls on a different part
of the photodetector. The signals from the four quadrants of the
detector are compared to calculate the deflection signal.
Piezo motor for x,y mapping
Optical detection
(1) In contact mode: F(x) = -k × x Hooke's Law Spring constant of
cantilever is less than surface, cantilever bends. Typical
atom-atom k ~ 10 N·m-1, typical cantilever k 0.1-1 N·m-1
Total force on sample 10-6 to 10-8 N If spring constant of
cantilever is greater than surface, surface deformed. This mode can
be used for very high resolution imaging, such as atomic
resolution
(3) In intermittent contact (tapping) mode: Similar to non-contact
AFM using vibrating cantilever except at one extent tip "taps" into
contact mode Useful for soft surfaces - less prone to external
vibration/noise than non-contact Less destructive than contact AFM
and can image rougher samples
DNA acquired in tapping mode
(2) In non-contact mode: Very small force on surface (~ 10-12 N) -
tip-surface distance 10-100 Å - best for soft or elastic surfaces -
least contamination - least destructive - long tip life In
non-contact mode the cantilever oscillates close to the sample
surface, but without making contact with the surface : AC driven
oscillating cantilever (100-1000 Hz frequency, 10-100 Å amplitude)
- resonant frequency ν =1/(2 π) sqrt(k/m) - k varies with external
force gradient (dF(x)/dx) so frequency changes with external force
- electronics adjust tip-surface distance to keep resonant
frequency constant -> constant tip force Contact and non-contact
show similar topography except for soft/deformable materials
First, the tip is vibrated at its resonant frequency (ω0) using a
piezoelectric element while far from the surface (assuming no
interaction). As the tip is moved towards the surface, the presence
of a force gradient modifies the spring constant of the cantilever
so that keff = k – F’ where k is the spring constant of the
cantilever in the absence of a tip-sample interaction. The key
point is that this modification of the spring constant will produce
a shift of the resonant frequency of the cantilever given by:
k F
k F
m k
m Fk
m keff
ω
ω
where ω is the new resonant frequency of the cantilever due to its
interaction with the surface. In practice, the user first selects
an operating resonant frequency, ωsp . As the tip moves towards the
surface both the shift in resonance frequency and the damping of
the cantilever’s oscillation due to tip-surface interactions will
produce a corresponding change in the amplitude of oscillation at
ωsp. To generate a NC- AFM image the user chooses a set-point
amplitude, Asp. As the tip is scanned across the surface the
feedback loop controls the tip-sample separation so as to maintain
the oscillation amplitude constant at Asp. A NC-AFM image therefore
represents a map of a constant force gradient (defined by Asp) due
to the tip- sample interaction.
Details of the non contact mode (NC-AFM)
The method in theory:
Contact mode topography (left) and non contact mode image (right)
of a two-phase block copolymer.
The result in image:
SEM images Dimensions
Spatial resolution = 50 nm
The diffraction limit of spatial resolution. It is not possible to
spatially resolve details that are located closer together than
approximately half the probing wavelength. For optical microscopy,
typically operating at a wavelength of 500 nm (the visible spectrum
ranges from 400 nm to 700 nm), the lateral resolution is thus
limited to about 250 nm.
Scanning Near-Field Optical Microscopy (SNOM), brings a small
optical probe very close to the sample surface, in the region
called "near-field“ (opposed to conventional microscopy which are
far-field microscopes). Here, at distances smaller than the
wavelength away from the surface, also those waves can be probed
that do not propagate, but rather decay exponentially perpendicular
to the surface. In this evanescent field the k-vectors parallel to
the surface can be fairly large, corresponding to small lateral
(spatial) dimensions. Broadly speaking, if the aperture-specimen
separation is kept roughly less than half the diameter of the
aperture, the source does not have the opportunity to diffract
before it interacts with the sample and the resolution of the
system is determined by the aperture diameter as oppose to the
wavelength of light used. An image is built up by raster-scanning
the aperture across the sample and recording the optical response
of the specimen through a conventional far-field microscope
objective
SNOM is a technique based on the STM. In a SNOM experiment, a fiber
tip is scanned in close proximity across a sample, and optical
information like reflectivity, fluorescence, luminescence or
polarization can be derived with sub- wavelength resolution. In
addition, topographical information can usually be obtained, since
a local interaction (e.g. the lateral force between tip and sample
surface) is used for control of the tip-sample distance, in a
similar way as in an atomic force microscope (AFM)
A typical instrument consists illumination (laser, fiber coupler)
and collection optics (objectives, filters, photomultipliers for
moderate light levels or photon counters of very low intensities),
fiber tip holder with shear force feedback (oscillator and lock-in
amplifier), an approach scheme (mechanical or motorized), and a
scanner (piezo tubes or stacks, it is often advantageous to scan
the sample rather than the probe). Digital data acquisition and
anti-vibration damping (optical tables, actively or passively
dampened) completes the equipment. The microscope shown here sits
on a conventional (inverted) light microscope, which allows to
localize the sample with low resolution prior to SNOM
operation.
Most common today is the use of aperture probes for transmission
microscopy, either in illumination (a) or in collection (b).
However, many samples or substrates are opaque, so that working in
reflection is necessary (c). The reflected light can be collected
by optics close to the tip, or by the fiber probe itself, in which
case often uncoated fiber tips are used (d). The probe tip acts as
a scatterer of the evanescent field, leading to homogeneous waves
which can be easily detected.. Of high interest is this arrangement
with inverted light path, Tunnel Near-Field Optical Microscope
(TNOM) or forbidden light near-field optical microscope. Light can
be scattered from the evanescent field by other probe tips, such as
a force microscope tip on a cantilever (e). In the Plasmon
Near-Field microscope, surface plasmons are generated at the
surface of a sample on a thin film metallic substrate, and
scattered by a probe tip (f).
See also FSB/IPMC/LPMV or FSB/IPMC/LPRX/LPFM
SNOM ways of operation