Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry

Post on 22-Feb-2016

31 views 0 download

description

Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry. Jerry Dulude and Scott Bridger Glass Expansion. Assist Automated Productivity Accessory. Assist components. Niagara II Plus ™ Rapid Rinse accessory Sample syringe plus drive Built in switching valve - PowerPoint PPT Presentation

transcript

Advantages of a Syringe Driven Sample Introduction Accessory for ICP SpectrometryJerry Dulude and Scott Bridger

Glass Expansion

Glass Expansion www.geicp.com

Assist Automated Productivity Accessory

Glass Expansion www.geicp.com

Assist components

Niagara II Plus™ Rapid Rinse accessory Sample syringe plus drive Built in switching valve Built in peristaltic pump Optional internal standard syringe plus drive

Glass Expansion www.geicp.com

Niagara II Plus

Glass Expansion www.geicp.com

Niagara II Plus Flow Diagram

Glass Expansion www.geicp.com

Niagara II Plus Flow Diagram-2

Glass Expansion www.geicp.com

Screen display of Niagara II Plus software

Glass Expansion www.geicp.com

ICP-MS Carryover Comparison

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

Blank1

Blank2

Blank3

Blank4

Blank5

Blank6

ppb

Bor

on

Standard

w ith Niagara

Boron

0.00

0.50

1.00

1.50

2.00

2.50

3.00

3.50

ppb

Mo

Standard

w ith Niagara

Moly

0.00

0.20

0.40

0.60

0.80

1.00

1.20

Blank1 Blank2 Blank3 Blank4 Blank5 Blank6

ppb

Zn Standard

with Niagara

Zinc

0.00

0.20

0.40

0.60

0.80

1.00

1.20

1.40

Blank1 Blank2 Blank3 Blank4 Blank5 Blank6

ppb

Tl Standard

with Niagara

Thallium

Glass Expansion www.geicp.com

Niagara II Plus Selectable sample loop Optional internal standard lines Positive displacement pump Multi-stage valve Sample does not contact pump tubing

Glass Expansion www.geicp.com

Standard operation Niagara II PlusCycle time Rinse time Cycle time Rinse time

Probe to sample 5 0 5 5Uptake delay 15 0 5 5Stabilization 10 0 10 0Read 15 0 15 0Probe to rinse 5 0 0 0Rinse 30 30 0 0TOTAL 80 30 35 10

Time Saving for Niagara II Plus

Glass Expansion www.geicp.com

Additional Assist Features

Precision syringe and drive for sample Optional precision syringe and drive for

internal standard

Glass Expansion www.geicp.com

Syringe vs. peri pump

Glass Expansion www.geicp.com

Advantages of Syringe Sample Delivery Computer control of delivery speed Broad selection of sample/IS ratio High precision for scanning systems

Glass Expansion www.geicp.com

e.g. 20x dilution with peri pump @ 0.5ml/min uptake 0.38mm ID (orange/green) 1.75mm ID (blue/green) Pump speed=2rpm

Very Noisy Signal

Glass Expansion www.geicp.com

Assist Performance on a Elan9000 ICP-MS

Uptake rate: 0.5ml/minSample: .025ml/minIS: .475ml/minDilution = 20x

* Internal standards

*

*

Glass Expansion www.geicp.com

Assist Productivity Enhancement

Without Assist With AssistCycle time Rinse time Cycle time Spray chamber

rinseProbe/uptake line rinse

Probe to sample

5 0 5 5 0

Uptake delay 15 0 5 5 0

Stabilization 10 0 4 0 4

Read 15 0 15 0 15

Probe to rinse 5 0 0 0 0

Rinse 30 30 0 0 0

TOTAL 80 30 29 10 19

Glass Expansion www.geicp.com

Productivity comparison*

Standard operation Niagara II (w/o FIA) Niagara II Plus Assist

19% 27% 43% 52%

* % of cycle devoted to measuring signal