EXPAIR software - Ascencia · 2020. 10. 15. · 120515_Bodybox_EN (Volumetric option) p• The...

Post on 09-Feb-2021

0 views 0 download

transcript

  • 0029

    Cert. N°: 09/BE/1814ISO 13485 (2003)

    BODYBOX 5500®Barometric total bodyplethysmograph

    EXPAIR software

    The MediSoft factory is a state of the art modern facility with clinical research, precision engineering and computer design departments.

    GEnERAl SPEcIfIcAtIonS

    Dim. Standard Double Door/XL Paediatric (HxWxD)cm 176x87x71 179x120x83 140x66x55 Weight ±130kg ±150Kg ±105Kg Internal Volume 960L ±1250L ±450L Patient chair Pneumaticadjustement BodyBox closing door Sturdyclosingandinternalhandle Power requirements 230/115VAC50/60Hz Power Consumption 100VA(module) Warm up Time 20min(minimum)Conformtoelectricalsafetyreq.IEC60601/1andCE0029

    BODYBOX 5500 PAtIEnt VAlVE

    Pneumotachograph Lillycone Range 0,03to15L/secor20L/sec Résistance 0,4cmH2O/L/sec Relativeaccuracy Error<3%Volumeconv.toBTPSintegratedthermometer(optionalbarometer)Automaticzeroshiftcorrectionofmeasuringelements

    Softwarecomputerisedlinearization Patient valve Pneumatic(TimeO/C:30mS) DeadSpace <60ml/30ml(paediatric).DisinfectionSimpleDismantlingforcoldcleaning Valve Support Arm Moveablearmwith3joints

    PRESSuRE tRAnSducERS

    Piezoresistivesensorsprotectedfromoverload Sensitivity Resolution Calibration Boxpres ±0,5cmH2O BoxPres 0,05cmH2O BoxPres Integrated30mlpump Mouthpres ±50,0cmH2O MouthPres.&Flow 0,01cmH2O MouthPres. Watercolumn Pres.MIP/MEPSNIP ±280,0cmH2O Linearity Error<0,1% Pneumotachograph semi.auto.with1to3L Mouthflow ±5,0cmH2O RelativeAccuracy Error<0,01%syringewithERS/ATSqualitycontrolindicator

    GAS AnAlyzERS

    Helium Thermalconductivity Range 0to15%He Relativeaccuracy ±0,1% Responsetime ±200msecFastHe ±10secHeSTD Carbon monoxide infrared/FastorFuelCell Range 0to0,350%CO Relativeaccuracy ±0,1% Responsetime ±150msecFastCO ±20secCOFuelCell O2 ChemicalFuelCell Range 0to100%O2 Relativeaccuracy ±0,02% Responsetime ±150msec CO2 infrared Range 0to10%CO2 Relativeaccuracy ±0,01% Responsetime ±120msec

    oPtIonAl GAS AnAlyzERS

    Multigas analyser Infraredspectrometer(CO,CH4,CO2,C2H2) Range 0to0,350%,CO2:0-10% Relativeaccuracy ±0,1% Responsetime <20sec(10-90%FS) Nitric oxide Chemicalfuelcell Range 0-450ppm Relativeaccuracy ±0,1% Responsetime <10sec(10-90%Fs)

    GAS SuPPly SPEcIfIcAtIonS

    BodyBox Compressedair TLCOHeoption 0,3%CO,14%He,21%O2,remN2 TLNOoption ±450ppmNo,remN2 TLCOCH4option 0,3%CO,0,3%CH4,21%O2,remN2 Pressureregulator 0-8Bars/15m3/h TLCOss 0,08CO,21%O2remN2 FRCN2 100%O2

    AnAlySER cIRcuIt

    Automatic,rapidandaccuratecalibrationwithqualitycontrol

    VolumEtRIc BodyBoX (oPtIon)

    Pneumotachograph doubleGridLillycone Range 0,01to15L/sec Resistance 0,1cmH2O/L/sec Relativeaccuracy Error<3%

    comPutER IntERfAcE

    Type SerialRS232orUSBConversion 12&16bit. Acq.frequency 100Hz/chanels(Multigas3500Hz) Transmissionspeed 115,200baud Isolation Systemfullyisolated byoptocoupling Computer PCIntel,19’’monitor, PrinterA4colourDeskjet Operatingsystem Windows7®32or64bit

    AmBIEnt condItIonS

    Temp.10-40°c Relativehumidity25to80% noncondensed

    mEASuREd And cAlculAtEd PARAmEtERS

    •RAW(Insp.exp.tot.),SRAW,GAW,SGAW,...•TGV,VC,IRV,ERV,RV,TLC,...•SlowSpirometry:VC,ERV,IRV,IC,EC•FocedSpirometry:FEV1,FIV1,FVC,FEV1/FVC,FEV1/VC,•FEV6,PEF,F25,F50,F75,MEF,MVV,...•Bronchodilationandchallengetest,dose-responsecurves,

    reactivitythreshold,...•VComp.,P.Alv.(option)•TLCO:AV,TLCO/AV,TLCO-NO:Dm,Vc;TLCOss(option)•Compliancestat./dyn.,RLstat./dyn.,CLstat./dyn.,ELdyn.,Wvis.

    (option)•MIP/Mep,SNIP,...(option)•NEP,RNep,Exp.FlowLim,...(option)•FRCN2-RVN2-TLCN2(NEW)•ClosingVolume•LostVolume,IntrathoracicPressure&

    PartialFLow/Volumecurve

    oPtIonS

    •Mixed(VolumetricandBarometricbodyplethysmograph)•TLCOHe,TLCONO,TLCOCH4,TLCOsteadystate/TLCOrebreathing•MIP/MEP,SNIP,NEP,Rint•StaticanddynamicCompliance•Integratedautomaticnebulizer,PROVO4.•ExternalcompensationBox•Computerintegrationtrolleywithelectricalisolationtransformer•DoubleDoorforwheelchairaccess•Medisoftnetwork•Datatransfer&reception(HL7,…)•Automaticdatabackup•Disabledhandrailsupport•ColourLazerPrinter

    Yourofficialdealer:

    BELGIUM HEADQUARTERP.A.E. de SorinnesRoute de la Voie cuivrée 1B-5503 Sorinnes – Belgiumtel: +32 (0)82 22 30 20fax: +32 (0)82 22 33 34info@medisoft.bewww.medisoft.be

    MEDISOFT FRANCERue du maréchal delattre de tassigny, 859000 lilletel: + 33 (0)825 280 030fax: +33 (0)320 787 609info@medisoft.frwww.medisoft.fr

    MEDISOFT CHINAWest tower, Shun tak centre168 connaught Road centralHong Kongtel: +852 2267 8933fax: +852 2267 8983postmaster@medisoft-china.com

    MEDISOFT ITALIAVia Germania, 1435127 cAmIn (Pd)ItAlytel: +39 049 87 03 355fax: +39 049 87 03 388info@medisoft-ram.itwww.medisoft-ram.it

    Medisofts.a.reservestherighttochangeandimprovetheabovespecificationswithoutpriornotice

    MEDISOFT GERMANY GmbHuniversitätsstraße, 136d-44799 BochumGERmAny tel.: +49 (0) 234 976-58410 fax: +49 (0) 234 976-69507info@medisoft-pft.dewww.medisoft-pft.de 12

    0515

    _Bod

    ybox

    _EN

    (Volumetric option)

    p The reference instrument of a PFT laboratory

    p An unrivaled customisation(The largest selection of optional measures)

    p Environment comfortable for the patient

    p High tec and high quality design

    p Fast and accurate measurements

    p Low running cost & easy of maintenance

    p Intuitive Software

    Lost Volume, IntraTh. Pressure&

    Partial Flow/Volume curve analysis(See specific brochure)

    NEW

    for more details, visit our Website

    Bodybox (with handrail support option)

    The most Intuitive, userfriendly and complete sofware basic version

    •Asophisticatedandpowerfuldata-basefunctionandelectronicstorage

    •TrendsReport(Historicfunction)•Interpretationfunction•Commentfunction•OffLineinputandonlinedatatransfer•Reportdesigner•Predictedvalueeditor•Choiceoflanguages

    •Choiceofunitsforthemeasuredparameters•Bronchialtestgeneration•Bloodgaseswithbloodchemistryanalysis

    frommanualentry•UsersUnitscapability•Measurementsequencingconfiguration•Fullcalculationfunction:displayof

    calculationpointswithmanualcorrectioncapability

    •Technicaltoolboxtoenablediagnosticfunctionandfullprogramcontrol

    •Inbuiltqualitycontrolwithcalibrationmarkersforperformance

    •TeleassistanceorVPNassist

  • FRC N2

    LOST VOLUME

    D800

    AIRWAYS RESISTANCE and CONDUCTANCE (RAW & GAW)

    Theairwaysresistancecanbemeasuredeasilyandquickly.ThesubjectestablishesarestingFRCleveloftidalbreathingandtheoperatorcanthenselecttocollectflowagainstboxdataforanalysis.Thetestcanbeperformedatbothrestingventilationandwiththepantingmethod.Thecorrelationbetweenmethodsisexcellent.Whentheoperatorissatisfiedwiththequalityofthecaptureddatatheocclusionvalvecanbeoperatedtosettheactualthoracicgasvolumeatwhichthemeasurementsaremade.ThisisessentialforameaningfulmeasurementofSgaw.Thesoftwarecontrolallowsmanualfittingoftheslopetothedata,selectionandde-selectionofeffortsforthemean,manualselectionofTidalvolumeandVitalcapacityminimumandmaximumpointsaswellasadjustmentoftheFRClevel.

    THORACIC GAS VOLUME, TOTAL LUNG CAPACITY AND LUNG SUB-DIVISIONS

    ThetestcallsforthesubjecttoachieveastableFRClevel,theoperatorcanthenselectocclusionmode.Thepowerfulsoftwarethenclosesthevalveatanendtidalpointandtheresultingtangentofthemouthpressureversustheboxpressureisthethoracicgasvolume(TGV,FRC).AssubjectschangetheirFRClevelduringtheeffort,thesoftwaremonitorsthebaselineshiftandthenaddsorsubtractsthevolumedifferencefromthemeasuredthoracicgasvolumetoshowthe‘TrueFRC’astheTGV.Afullvitalcapacityeffortthenprovidesthecalculationsofthelungsub-divisionsrelatingtotheTGVmeasurement.Completecontroloverreportedparametersandinformation.Selectionofdata,movementofthetidalvolume,vitalcapacityandFRClevelsaswellasdeletionofpoorefforts.Fivesetsofmeasurementscanberecordedatthesameordifferenttimestoidentifyimprovementsinpatientmeasurements.Threedifferentocclusionmodescanbeselected:1.Insp.cycle-2.Completecycle-3.Multiplecycles

    FRC MULTIBREATH N2 WASHOUT (OPTION)

    IntegratedorwithHypAirArtsMeasurementofTLCbyNitrogenwashout.TheNitrogenwashoutusestheaccumulatedvolumeofoxygentowashnitrogenfromthelungsasthe‘true’FRCvalue.Onscreenvisualisationguidestheuserthroughthefulltestprocedure,makingtheunderstandingandqualitycontrolofthetestakeyfeature.ImportingaVCfromaseparateeffortispossiblewhenthesubjectmaybeunabletoperformthebesteffortwithinthetest.ThemeasuringcircuitusesO2andCO2analysersbysubtractiontomeasurethenitrogen;thepatientcircuitconsistsofanautomatictwo-channelvalvewithautomaticdeliveryof100%O2withalowresistancedemandvalve.TheclosingvolumeiscalculatedfromN2slopeduringamaximalexpiration

    SLOW AND FORCED SPIROMETRY (TOTAL OR PARTIAL)

    TheBodyBox5500allowsfullmeasurementofspirometrywithfacilitiesformultiplelevelspirometryaswellasbronchialchallengetesting.Additionally,incentivespirometryassiststhemeasurementsofchildren.Postvisualizationandposttreatmentprovideaconvenienceofuseparticularlyappreciated:Choiceofgraphicalrepresentation,selectionofcalculationpoints,alignementofflowcurvesonRV,timeevolutionofmeasuredparametersbynumericalandgraphicalrepresentation,deletionofpooreffort,trendgraphinganddoseresponsecurves.

    MEASUREMENT OF DIFFUSION CAPACITY TLCO (OPTION)

    UNIQUE - the only body plethysmograph offering the range of 5 diffusion methods.

    Single Breath using the helium trace gas He

    ThewellknowntechniquedescribedbyJ.E.CotesbasedontheJonesMeademethod.Usingabagcollectionsystemthesubjectcanbecontrolledforinspiratoryvolume,washout(discard)volumeandSamplevolume.Thismethodhasprovenrepeatabilityandthemethodwasthesameasthatusedtocollectthepredictedvaluesweusetoday.

    TLCO-NO (trace gas He) NEW & EXCLUSIVE

    Membranediffusionandcapillarybloodvolume(Dm&Qc)measurement.Regardedasthe‘True’diffusioncharacteristicandthemostusefulindicatorofmembranethickening,thismeasurementtakesonanewleaseoflife.ThepowerfulExp’Airsoftwaremakesthecalculationspainless,thecombinationofNOandCOfollowstheworkofProf.GuenardofBordeaux.Simpletoperformandrapidresultsaddthistest

    tothelistoffavouritestudiesundertakeninroutinepractice.AdditionallythismeasurementcanbeperformedinconjunctionwithNEP,thisthenallowstheevaluationofpulmonarybloodflowanditsabilitytorecruitfromthecapillaryreserve.

    Single Breath & Intra Breath using Methane trace gas (CH4)

    Usingfastgasanalysis,thismethodcollectstheexhaledbreathdirectlyasahighresolutiondataarray.Thishastheadvantagethatposttestanalysiscanbeperformedaligningthestartofsampleontothealveolarplateauaftertheclearancesofallthedeadspaces.Asampleaslowas50mlcanbeusedtocalculatethediffusion.Thiscanovercomethevolumelimitationofothersystemsandmethods.

    Single Breath & Intra Breath using trace gas (He)

    Themostsignificantimprovementtodiffusionmeasurementsisthelast10years.PreviouslyonlypossiblewithamassspectrometerandrapidIRcarbonmonoxideanalyzer.ThenewMedisoftrapidgasanalyzertechnologyallowsthemeasurementsin‘realtime’ofexhaleddiffusiongases.Thisisbothintraditionalsingebreathandintrabreathmodes.Withtypicalresponsesintherangeof100-200msfortwodiscreteanalyzersthesystemisperfectlymatched.AsthismethodisfaithfultothestandardbywhichthePredictedValueswerecollected,thedatayieldsthemostreproducibledataforsubjectsrelatedtonormaldata.Themethodextendseasilyto‘steadystate’diffusionin‘realtime’andsohasnovolumelimitationforthemeasurementsofdiffusion.ThepropertiesofHeliumasthetracesgasisbetterpresentedforgasmixingthanalternative,leadingtomoveaccurateestimatesoftotalsurfacearearepresentedbysinglebreathtotallungCapacity(TLC)thealveolarVolume(Va)

    Intra Breath diffusion He & CH4

    Thesampleistakenduringaslowandconstantexhalationintherangeof200–500ml/sec.Applyinglinearregressionstothedataarrayoftheexpiratorygas,thealveolarconcentrationsarecalculated.Asthismethodrequiresnobreathholding,itwillgreatlybenefitsomesubjectgroups.

    Steady State diffusion Tlco ss EXCLUSIVE

    Medi-softhastakenanewlookatthismethod,usingfastgastechnologyandreplacingtheolderbulkyinstrumentation.Thisisa“NEW”methodforanewageofdiffusionmeasurement.Requiringminimalsubjecteffort,thismethodisespeciallyhelpfulforobtainingmeasurementswithchildrenandreluctantsubjects.Performedatasteadystatebreathingcondition,themeasurementisvalidassoonasthesubjectsventilationisuniformandstabilized.

    NEP (OPTION) EXCLUSIVE

    Anewandverysensitivetestthatisspecificandreproduciblefordeterminingthedegreeofexpiratoryflowlimitationbothatrestandduringexercise,particularlywithsubject’sknowntohaveobstructivelungdisease.Thetestappliesanegativepressuretothemouthpieceduringtheexpiratoryphase,thisallowsthecomparisonoftheflowvolumeloopwiththetidaleffortswhenreviewedasaflowvolumeloopdisplay.Thismethodalsoallowstheindirectmeasuremtoftheresistance(RNep)whichprovidesagoodalternativetothestandardscreeningmethod.

    VENTILATION MECHANICS (OPTION)

    •MIP–MEP:MaximalInspiratoryandExpiratoryPressuresasanassessmentofrespiratorymusclestrength,alsousefulforweaningsubjectsfromventilators,…•SNIP:measurementofthemaximalnasalinspiratorypressureusinganasalcanular.Anon-invasiveestimateofmusclefatigue.•STATICANDDYNAMICCOMPLIANCEANDRESISTANCE:Theuseofanoesophagealballooncatheterinsertedintothesubjectallowsthemeasurementsofbothstaticanddynamiccompliance.Thecomponentsofcomplianceandresistancearefundamentaltothemechanicsofthelung

    PROVO 4 (OPTION)

    Equipmentforfullyautomatedbronchochallengetestwithautomaticnebulizer.Thisoptionprovidesfullcontrolofproductsanddosesusedandoftestperformingcriteriaforbronchochallengetesting

    LOST VOLUME (OPTION)

    seespecificbrochure

    INTRATHORACIC PRESSURE (OPTION)

    seespecificbrochure

    p The perfect unit for :p Pulmonary Function Test with both

    children and adultsp Ventilation mechanicsp Bronchodilatation studiesp Hypersensitivity research

    (challenge tests)

    BODYBOX 5500

    New trolley «Standard V2» (With Wide Screen 24" option)

    Customized Bodybox: – Metallic gray – Acrylic seat

    from basic to advanced version, medisoft is the only manufacturer able to offer you customisation at a standard price!

    p Ergonomically adjustable to your needs :p Bring assistance to disabled or infirm subjects with

    the handrail support and the Double door version.p No stress : pleasant environment for the patient.p Add an interactive dimension and improve the quality of

    your tests by choosing the incentive screen.p Customise your box by choosing the size*,

    the color and type of seat you want.

    *(Standard, XL, Double door, Paediatric or Customised)