Post on 20-Apr-2015
transcript
TechnicalPublications
2262352Revision 1
LOGIQt200 PRO SeriesInSite Option Installation Instructions
CopyrightE 1999, 2000 By General Electric Co.used by GEMS- -Europe, GEMS- -Asia and GEMS- -Americas
Operating Documentation
GE Medical Systems
SAMSUNG GE MEDICAL SYSTEMS, LTD.SERVICE TEAM
64--3, Sangdaewon--dong, Chungwon--ku, Sungnam--siKyunggido, KOREA
REPORT ON TECHNICAL PUBLICATION -- RT(OMISSIONS,ERRORS,SUGGESTIONS)
PRODUCT PUBLICATION #
PAGE # PAGE REV DATA REPORTED
(DO NOT TEAR) FOLD HERE AND SEAL (DO NOTTEAR)
SERVICE RECORD SUBMITTED YES NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS
FOLD FOLD
IF ADDITIONAL PAGES ARE INCLUDED, FOLD AND STAPLE TO LOWER PART OF THISSHEET BEFORE MAILING.
SERVICE REPPHONEFAXCOUNTRYCITY
REV0
Stick along this line
2184996
R & D Support Group Of SGMSFAX Number: 82--342--742--0436
REV. 02184997
GE MEDICAL SYSTEMS
2262352
LOGIQTM 200 InSite Option--Installation Instructions
A
REV 1
LIST OF EFFECTIVE PAGES
REV DATE PRIMARY REASON FOR CHANGE
0 NOV. 22, 1999 Initial release. . . . . . .1 MAR. 1, 2000 System V3.1 release. . . . . . . .
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page 1. . . . . . . . . .GE Logo page --. . . . . .P9030CB 2. . . . . . . . . . .P9030CC 2. . . . . . . . . .P9030CD 0. . . . . . . . . .
A 1. . . . . . . . . . . . . . . . . .1 to 16 1. . . . . . . . . . . . .
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION1
REV 1
1. InSite Option Installation
1--1 INTRODUCTION
These Service Instructions describe the installation of the InSite Service Capability for LOGIQ 200 PRO system.
1--2 TOOLS REQUIRED
Standard field Service Tool kitLOGIQ 200 PRO Service Manual (2235374)LOGIQ 200 PRO Proprietary Manual (2242594)
1--3 TIME REQUIRED
Approximately 2 hours plus travel time
1--4 PROCEDUREREAD and UNDERSTAND these instructions thoroughly before proceeding with the installation. Perform each step insequence and check it off when completed. If a problem occurs after completion, go back through the procedure andcheck for implementation errors before contacting your Ultrasound Region Engineer.Check each step as it is completed.
Functional Check--outPerform functional checks to verify that the system is functioning properly. Refer to all Functional Checks inChapter 4 of the LOGIQ 200 PRO Service Manual, 2235374.
NoteThe LOGIQ 200 PROsystem supports 2 kinds of modem -- US Robotics and MultiTech.Installation procedures are described separately in the following chapters of this insturction.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION2
REV 1
2. InSite Installation for US Robotics Modem
2--1 PARTS REQUIRED
ITEM PART NAME QTY PART # REMARKS1. Modem 1 2135816--2 US Robotics V.Everything2. RS232C Cable 1 local3. LOGIQ 200 PRO InSite Option 1 2262352
Installation Instructions
NoteThe LOGIQ 200 PRO Modem includes the preparatory RS232C cable to validate the InSitefunction. Never use this cable when scanning patient to prevent the patient injury.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION3
REV 1
2--2 Setting-Up Modem
1. Make sure that the DIP switches located at the rear side of the modem are set as shown in NO TAG.
1 2 3 4 5 6 7 8 9 10
quadOFF
ON
DIP SWITCH SETTINGSILLUSTRATION 1--1
2. Connect the power cord with the modem and plug it with a wall outlet. Refer to NO TAG.
3. Power the modem ON.
4. Power the modem OFF to connect the modem to LOGIQ 200 PRO.system.
2--3 Connecting Modem
1. Make sure that both the LOGIQ 200 PRO system and the modem are powered OFF.
2. Plug the D--sub 25--pin Connector (System side) into the Service port of the LOGIQ 200 PRO Rear Panel.
3. Plug the D--sub 25--pin Connector (Modem side) into the RS232 port of the US Robotics V.EverythingModem.
4. Connect the Modular Audio Cable between the phone jack ( ) of the modem and the dedicated analogtelephone line. Refer to ILLUSTRATION 1--1.
5. Connect one side of the power cord (supplied with the Modem Set) to the Modem as shown in NO TAG.
6. Connect another side of the power cord to a wall outlet.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION4
REV 1
2--3 Connecting Modem (Continued)
Rear Panel ofLOGIQ 200
RS232C Connector[System side](supplied with RS232C Cable Set)
RS232C Connector[Modem side](supplied with RS232C Cable Set)
Rear Panel OfUS Robotics V. Everything
Modular Audio Cable(supplied with Modem Set)
AC/DC Adaptor
Dedicated Analog Telephone Line
Wall Outlet
NoteThe type of power plug is different depending on the region.
MODEM CONNECTIONILLUSTRATION 1--2
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION5
REV 1
2--4 Setting-Up LOGIQ 200 PRO Series
1. Power the LOGIQ 200 PRO system ON.
2. Power the modem ON after completing the LOGIQ 200 PRO power-up sequence.
3. Press Setup Menu key to go directly to the Setup/Top Menu.
4. In the Setup Top menu, Using the Trackball or arrow keys, choose General System 2 Setup and press Set.Refer to NO TAG .
Set to go to General System Setup Menu
SETUP MENU : Top Menu 01/13/99 11:12:35CATEGORY : OBSTETRICS PRESET :1 Ver. 3.0COMMAND : EXIT SAVE DELETE
General System 1 Setup*General System 2 SetupProbe Parameter 1 SetupProbe Parameter 2 SetupImage Display & Application SetupBody Pattern SetupComment SetupMeasurement SetupPatient Entry SetupUser Define Category & Key SetupUser Utility
Activating InSite Option--Service Port Setting (Setup Top Menu)ILLUSTRATION 2--1
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION6
REV 1
2--4 Setting-Up LOGIQ 200 PRO Series (continued)
5. The General System 2 Setup Menu will appear on the monitor as shown in NO TAG.
6. In the General System 2 Setup Menu, Using the Track ball or Arrow Up/Down keys choose “Service Port”.
7. Choose “4:Insite”by inputing the number and press Set .
SETUP MENU : General System 2 01/13/99 11:14:16
COMMAND : EXIT SAVE DELETE
Record 1 :NoneRecord 2 :NonePrinter-A Port :1 (1.None 2.Printer)Printer-B Port :1 (1.None 2.Printer 3.DMC)
* Service Port :4 (1.None 2.Printer 3.DMC 4.InSite)Start OB Report page :1 (1.Summary Page 2.Graph Page)OB Trend Graph Page :1 (1.1graph 2.5graphs)EFW GP :1 (1:Hadlock 2:William 3:Brenner)EFW USA :AutoEFW Europe :1 (1.Shp/Wrs 2.Rch/Brk 3.Hdlck 4.Grm 5.Merz)EFW Tokyo :1 (1.TYO 2.TYO-S1 3.TYO-S2 4.TYO-S3)OB Report Form :1 (1.Tokyo 2.Osaka 3.USA 4.Europe)Hadlock table :1 (1:’84 2:’82)OB User Table : Sub menuBSA Type :1 (1.Oriental 2.Occidental)Display Body pattern Menu :1 (1.Yes 2.No)Display Comment Menu :1 (1.Yes 2.No)Auto Patient data Search :1 (1.Yes 2.No)New Patient Menu Display :1 (1.Yes 2.No)
Activating InSite Option Setting (General System 2 Setup)ILLUSTRATION 2--2
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION7
REV 1
2--5 Activating InSite Option
1. Enter the service software after completing the power--up sequence. For the service software, refer to LO-GIQ 200 PRO Series Proprietary Manual, 2242594. The Service Utility Menu:Top Menu will be displayedas shown in NO TAG. Refer to the latest service note for the password.
2. Move the cursor to the item “Option Set”and press Set key. Option Set Menu will be displayed.
Service Utility MENU : Top Menu 99/03/12 09:16:37Current Software Version is 3.0
System DiagnosisBoard/Program VersionSystem ResetKey Operation History
* Option SetInitialize :1 (1:America 2:Europe 3:Asia 4:Japan)System Clock AdjustSoftware/Probe data upgradeExit
SERVICE TOP MENUILLUSTRATION 2--3
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION8
REV 1
2--5 Activating InSite Option (continued)
3. The Option Set Menu will appear on the monitor as shown in the NO TAG. The status of each option will beindicated.
Service UtilitY MENU : Option Set 99/03/12 09:16:37Current Software Version is 3.0
CINE :1 (1:Not Install 2:Install)ATO :1 (1:Not Install 2:Install)MOD Driver :1 (1:Not Install 2:Install)Biopsy Guideline :1 (1:Not Install 2:Install)MGOB Calculation :1 (1:Not Install 2:Install)Cardiac Calculation :1 (1:Not Install 2:Install)Urology Calculation :l (1:Not Install 2:Install)Serial Printer :l (1:Not Install 2:Install)Paralled Printer :1 (1:Not Install 2:Install)Data Management Center:l (1:Not Install 2:Install)
* InSite :2 (1:Not Install 2:Install)
Exit
OPTION SET MENUILLUSTRATION 2--4
4. Using the Arrow up/down keys or Trackball , select the InSite option item and then Input the number 2:Install.
5. To activate the Option, exit the menu and reset the System.
6. This completes the InSite option installation of the LOGIQ 200PRO Series.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION9
REV 1
3. InSite Installation for MultiTech Modem
3--1 PARTS REQUIRED
ITEM PART NAME QTY PART # REMARKS1. Modem 1 2245773 MultiTech2. RS232C Cable 1 local3. LOGIQ 200 PRO InSite Option 1 2262352
Installation Instructions
NoteThe LOGIQ 200 PRO Modem includes the preparatory RS232C cable to validate the InSitefunction. Never use this cable when scanning patient to prevent the patient injury.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION10
REV 1
3--2 Setting-Up Modem
1. Connect the power cord with the modem and plug it with a wall outlet. Refer to NO TAG.
2. Power the modem ON.
3. Power the modem OFF to connect the modem to LOGIQ 200 PRO.system.
3--3 Connecting Modem
1. Make sure that both the LOGIQ 200 PRO system and the modem are powered OFF.
2. Plug the D--sub 25--pin Connector (System side) into the Service port of the LOGIQ 200 PRO Rear Panel.
3. Plug the D--sub 25--pin Connector (Modem side) into the RS232 port of the MultiTech Modem.
4. Connect the Modular Audio Cable between the phone jack of the modem and the dedicated analogtelephone line. Refer to ILLUSTRATION 1--3.
5. Connect one side of the power cord (supplied with the Modem Set) to the Modem as shown inILLUSTRATION 1--3.
6. Connect another side of the power cord to a wall outlet.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION11
REV 1
3--3 Connecting Modem (Continued)
Rear Panel ofLOGIQ 200
RS232C Connector[System side](supplied with RS232C Cable Set)
RS232C Connector[Modem side](supplied with RS232C Cable Set)
Rear Panel OfMultiTech Modem
Modular Audio Cable(supplied with Modem Set)
AC/DC Adaptor
Dedicated Analog Telephone Line
Wall Outlet
NoteThe type of power plug is different depending on the region.
MODEM CONNECTION .ILLUSTRATION 1--3
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION12
REV 1
3--4 Setting-Up LOGIQ 200 PRO Series
1. Power the LOGIQ 200 PRO system ON.
2. Power the modem ON after completing the LOGIQ 200 PRO power-up sequence.
3. Press Setup Menu key to go directly to the Setup/Top Menu.
4. In the Setup Top menu, Using the Trackball or arrow keys, choose General System 2 Setup and press Set.Refer to NO TAG .
Set to go to General System Setup Menu
SETUP MENU : Top Menu 01/13/99 11:12:35CATEGORY : OBSTETRICS PRESET :1 Ver. 3.0COMMAND : EXIT SAVE DELETE
General System 1 Setup*General System 2 SetupProbe Parameter 1 SetupProbe Parameter 2 SetupImage Display & Application SetupBody Pattern SetupComment SetupMeasurement SetupPatient Entry SetupUser Define Category & Key SetupUser Utility
Activating InSite Option--Service Port Setting (Setup Top Menu)ILLUSTRATION 2--5
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION13
REV 1
3--4 Setting-Up LOGIQ 200 PRO Series (continued)
5. The General System 2 Setup Menu will appear on the monitor as shown in NO TAG.
6. In the General System 2 Setup Menu, Using the Track ball or Arrow Up/Down keys choose “Service Port”.
7. Choose “4:Insite”by inputing the number and press Set .
SETUP MENU : General System 2 01/13/99 11:14:16
COMMAND : EXIT SAVE DELETE
Record 1 :NoneRecord 2 :NonePrinter-A Port :1 (1.None 2.Printer)Printer-B Port :1 (1.None 2.Printer 3.DMC)
* Service Port :4 (1.None 2.Printer 3.DMC 4.InSite)Start OB Report page :1 (1.Summary Page 2.Graph Page)OB Trend Graph Page :1 (1.1graph 2.5graphs)EFW GP :1 (1:Hadlock 2:William 3:Brenner)EFW USA :AutoEFW Europe :1 (1.Shp/Wrs 2.Rch/Brk 3.Hdlck 4.Grm 5.Merz)EFW Tokyo :1 (1.TYO 2.TYO-S1 3.TYO-S2 4.TYO-S3)OB Report Form :1 (1.Tokyo 2.Osaka 3.USA 4.Europe)Hadlock table :1 (1:’84 2:’82)OB User Table : Sub menuBSA Type :1 (1.Oriental 2.Occidental)Display Body pattern Menu :1 (1.Yes 2.No)Display Comment Menu :1 (1.Yes 2.No)Auto Patient data Search :1 (1.Yes 2.No)New Patient Menu Display :1 (1.Yes 2.No)
Activating InSite Option Setting (General System 2 Setup)ILLUSTRATION 2--6
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION14
REV 1
3--5 Activating InSite Option
1. Enter the service software after completing the power--up sequence. For the service software, refer to LO-GIQ 200 PRO Series Proprietary Manual, 2242594. The Service Utility Menu:Top Menu will be displayedas shown in NO TAG. Refer to the latest service note for the password.
2. Move the cursor to the item “Option Set”and press Set key. Option Set Menu will be displayed.
Service Utility MENU : Top Menu 99/03/12 09:16:37Current Software Version is 3.0
System DiagnosisBoard/Program VersionSystem ResetKey Operation History
* Option SetInitialize :1 (1:America 2:Europe 3:Asia 4:Japan)System Clock AdjustSoftware/Probe data upgradeExit
SERVICE TOP MENUILLUSTRATION 2--7
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION15
REV 1
3--5 Activating InSite Option (continued)
3. The Option Set Menu will appear on the monitor as shown in the NO TAG. The status of each option will beindicated.
Service UtilitY MENU : Option Set 99/03/12 09:16:37Current Software Version is 3.0
CINE :1 (1:Not Install 2:Install)ATO :1 (1:Not Install 2:Install)MOD Driver :1 (1:Not Install 2:Install)Biopsy Guideline :1 (1:Not Install 2:Install)MGOB Calculation :1 (1:Not Install 2:Install)Cardiac Calculation :1 (1:Not Install 2:Install)Urology Calculation :l (1:Not Install 2:Install)Serial Printer :l (1:Not Install 2:Install)Paralled Printer :1 (1:Not Install 2:Install)Data Management Center:l (1:Not Install 2:Install)
* InSite :2 (1:Not Install 2:Install)
Exit
OPTION SET MENUILLUSTRATION 2--8
4. Using the Arrow up/down keys or Trackball , select the InSite option item and then Input the number 2:Install.
5. To activate the Option, exit the menu and reset the System.
6. This completes the InSite option installation of the LOGIQ 200PRO Series.
GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions
2262352
INSTALLATION16
REV 1
This page is left blank intentionally.