Modification to Log-Amp position measurement in TT2-TT10

Post on 03-Jan-2016

27 views 0 download

Tags:

description

Modification to Log-Amp position measurement in TT2-TT10. BDI day 2003 T Bogey AB/BDI-EM. Outline. TTPOS principle Recall of problems Solutions implemented in 2003 Things to be done for 2004 Conclusions. Principle. BMU10T. PS. Trigger. TT2. FT16UDC151. FT16UDC211. - PowerPoint PPT Presentation

transcript

Modification to Log-Ampposition measurement in TT2-TT10Modification to Log-Ampposition measurement in TT2-TT10

BDI day 2003BDI day 2003T Bogey AB/BDI-EMT Bogey AB/BDI-EM

BDI day 2003 2

OutlineOutline

TTPOS principle Recall of problems Solutions implemented in 2003 Things to be done for 2004 Conclusions

BDI day 2003 3

PrinciplePrinciple

PS

SPS

TT2

TT10

BPCL100618

BPCL100509

FT16UDC151

FT16UDC211

FT16UDC251

FT16UDC311

FT16UDC346

FT16UDC368

BPCL100407

BPCL100208

BPCL101302

BPCL101502

BPCL101902

BPCL102073

BPCL102302

BPCL102402

BPCL102573

BPCL102739

BPCL102911

BPCL102963

BMU02T

BMU10T

Trigger

Trigger

BDI day 2003 4

PrinciplePrinciple

Pickup

BP FilterLogAmp

BP FilterLogAmp

BP FilterLogAmp

BP FilterLogAmp

Up-Down

Up

Down

Left

Right

Left-Right

100 / 600 m

Op Amp Integrator ADC

Op Amp Integrator ADC

VMECrate

Tunnel Auxiliary Building

Vertical

Horizontal

Auto Trigger

BDI day 2003 5

Recall of problemsRecall of problems

Auto trigger BMU10T● Position● Lead ions (stripped)

Radiation damage (102963) Position signal with no beam Loading settings when SSC changes Remote diagnostics

BDI day 2003 6

Auto trigger BMU10Twith lead ions (stripped)Auto trigger BMU10Twith lead ions (stripped)

2002: Auto trigger works with Pb53+

(unstripped) but does not work with Pb82+ (stripped)

2003: A new electronics for the auto trigger

module solved the problem.

BDI day 2003 7

Autotrigger position for BMU10TAutotrigger position for BMU10T

1013

1015

1019

1020

1023

1024

1025

1027

1029

1030

BMU10T

2003 Problem

2004

Solution

BDI day 2003 8

Radiations damages (102963)Radiations damages (102963)

The monitor is very close to the dump

During 2002 run ~ 23 E+3 gray on the electronics module

We moved the electronics away from the BPM

No more radiation damage during 2003 (but still a problem on the vertical plane)

BDI day 2003 9

Position signal with no beamPosition signal with no beam

Without beam Log-Amp still gives an output signal

No information of intensity

Autotrigger module gives a hardware interrupt to validate the data

A software algorithm check the data

BDI day 2003 10

Loading settingsLoading settings

When SSC changes MCR have to call experts to load settings

Stephen Jackson has implemented an option on the PCR Operational Software - BI applications web page to help operation

For a new SSC specialist still needed

BDI day 2003 11

BDI day 2003 12

BDI day 2003 13

Remote timing diagnosticRemote timing diagnostic

Already exists but requires expert

Solution has to be implemented for operators

BDI day 2003 14

ConclusionsConclusions

TT2/10 experience helps us for TT40/41 CNGS electronics

Each Log-Amp will have its own autotrigger

Position and Intensity available Possibility to use the same electronics

in TT2/10 upgrade