1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards and Calibration Laboratory (SCL) The Government of the Hong Kong Special Administrative Region
Transcript
Page 1: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

1Standards and Calibration Laboratory, SCL

Evaluation of Flicker Severity in Standards and Calibration

Laboratory (SCL)

Speaker: Chung Yin, PoonStandards and Calibration Laboratory (SCL)The Government of the Hong Kong Special

Administrative Region

Page 2: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

2Standards and Calibration Laboratory, SCL

Learning Objectives

• To introduce the meaning of flickers in power line

• To describe a method for calibrating flicker source and flickermeter

Page 3: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

3Standards and Calibration Laboratory, SCL

Introduction

• Flicker is the change in brightness of a lamp due to rapid fluctuations in the voltage of the power supply

• Cause disturbance to human visual perception system

• Voltage change characteristic at the terminals of the equipment under test, such as microwave ovens, washing machines

• All types of voltage fluctuations may be assessed by direct measurement using a flickermeter

Page 4: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

4Standards and Calibration Laboratory, SCL

Introduction

Source: Figure B.1, IEC 61000-4-15-2010 Flickermeter - Functional and design specifications

Page 5: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

5Standards and Calibration Laboratory, SCL

Evaluation of flicker parameters

• Flickermeter (Functional and design specification in IEC 61000-4-15)

• Flicker source (Limits of voltage changes produced by an equipment in IEC 61000-3-3)

• SCL developed a system to evaluate the accuracy of flicker source and flicker measurement functions of flickermeter

Page 6: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

6Standards and Calibration Laboratory, SCL

Operating Principle• Flicker waveform is modeled as an amplitude

modulated (AM) signal in power line (120 V / 230 V)

Page 7: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

7Standards and Calibration Laboratory, SCL

Operating Principle

• Relative voltage change / Voltage fluctuation

• Vupper: Amplitudes of upper sideband

• Vlower: Amplitude of lower sideband

• Vcarrier: Amplitude of the carrier

Page 8: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

8Standards and Calibration Laboratory, SCL

Calibration of flicker source - Setup

Page 9: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

9Standards and Calibration Laboratory, SCL

Resistive voltage divider

Page 10: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

10Standards and Calibration Laboratory, SCL

Calibration of flicker source - Setup

Page 11: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

11Standards and Calibration Laboratory, SCL

Calibration of flicker source

• Voltages of lower sideband (Vlower), carrier (Vcarrier) and upper sideband (Vupper) in dB are measured by signal analyzer

• Accuracy of the voltage fluctuation of standard flicker source can be determined

Page 12: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

12Standards and Calibration Laboratory, SCL

Calibration of flickermeter

• Flickermeters under test are calibrated by direct comparison with the standard flicker source

• Calibration in terms of voltage fluctuation and modulating frequency

• Instantaneous flicker sensation (Pinst) will be compared against the theoretical values as illustrated in IEC 61000-4-15

Page 13: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

13Standards and Calibration Laboratory, SCL

Calibration of flickermeter - Setup

Page 14: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

14Standards and Calibration Laboratory, SCL

What to be measured?

Page 15: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

15Standards and Calibration Laboratory, SCL

Measurement uncertainties

• Depends on– Amplitude accuracy of signal analyzer– Performance of standard flicker source– Calibration setup

Page 16: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

16Standards and Calibration Laboratory, SCL

Maximum uncertainty of voltage fluctuation measurement

Page 17: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

17Standards and Calibration Laboratory, SCL

Measurement uncertainties

• Flicker source: Measurement uncertainty < 0.03% at 95% confidence level

• Flickermeter: Measurement uncertainty < 0.3% for determination of flicker severities at 95% confidence level

• Type testing of Class F1 flickermeters according to IEC 61000-4-15– Tolerance of 8% in Pinst for sinusoidal voltage changes

Page 18: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

18Standards and Calibration Laboratory, SCL

Ratio between the specifications of standard flicker source and maximum measurement uncertainty (TUR)

Page 19: 1 Standards and Calibration Laboratory, SCL Evaluation of Flicker Severity in Standards and Calibration Laboratory (SCL) Speaker: Chung Yin, Poon Standards.

19Standards and Calibration Laboratory, SCL