Date post: | 26-May-2015 |
Category: |
Technology |
Upload: | kate-jones |
View: | 162 times |
Download: | 0 times |
Benefits of ICP-qqq-MS
in MS/MS mode for
challenging clinical
trace element
applications
Dr Raimund Wahlen
Agilent Technologies
07920 -466 161
Transforming ICP-MS
Technology
1 October 15, 2013
Presentation Overview
• Addressing interferences in ICP-MS – collision mode
• Reaction mode in quadrupole ICP-QMS
• Benefits of ICP-qqq-MS or MS/MS mode
• Examples of clinical applications:
• Trace Ti in serum + urine
• Se in presence of Gd – MRI patients
• Mn in whole blood
• Conclusions
2 October 15, 2013
Some Common Polyatomic Spectral Interferences
for Analysis of Clinical Samples by ICP-MS
Ion Interfering ion 51V+ 35Cl16O+ 52Cr+ 40Ar12C+ 59Co+ 43Ca16O+ 60Ni+ 44Ca16O+ 63Cu+ 40Ar23Na+ 66Zn+ 32S16O18O 75As+ 40Ar35Cl+ 78Se+ 38Ar40Ar+ 80Se+ 40Ar40Ar+ 95Mo+ 79Br16O+ 98Mo+ 81Br16O1H+ 103Rh+ 40Ar63Cu+ 111Cd+ 95Mo16O+
In addition you can get isobaric interferences (eg 40Ar+ on 40Ca+) and doubly-
charged interferences (eg 160Gd++ on 80Se) on target isotopes
Page 4
Energy Energy
Cell Entrance
Cell Exit
Energy loss from each collision with a He atom is the same for analyte and polyatomic ion, but polyatomics are bigger and so collide more often
At cell entrance, analyte and polyatomic ion energies overlap. Energy spread of both groups of ions is narrow, due to ShieldTorch System
Polyatomic ions
Analyte ions
Polyatomic ions
Analyte ions
Energy distribution of analyte and interfering polyatomic ions with the same mass
Bias voltage rejects low energy (polyatomic) ions
By cell exit, ion energies no longer overlap; polyatomics are rejected using a bias voltage “step”. Analyte ions have enough residual energy to get over step; polyatomics don’t (energy discrimination)
Principle of Helium Collision Mode and Kinetic Energy Discrimination (KED)
Efficient interference removal in He mode He collision mode is well-accepted for accurate multi-element analysis of unknown,
variable and high-matrix samples (enviro, food, clinical, pharma…)
He mode on 7700 is effective for polyatomic interferences at low- and sub-ppb levels
All elements in routine clinical matrices can be measured with a single set of
instrument parameters
Easy set-up, fast validation, reliable results
But … limitations for isobaric or doubly-charged interferences (e.g. Gd++)
Complex matrix in no gas (above) and
He mode (right, with 10ppb std inset)
7700 removes ALL polyatomics
7700 - He mode for Polyatomic Interferences
No gas mode (left) and He mode (below)
5 October 15, 2013
1. On-Mass Measurement: Unreactive analyte does not react with
chosen cell gas, remains at original m/z and so can be separated from
reactive interferences
Reactive interferences are converted to product ions at a new mass – can
be rejected by analyzer quad, which is set to original analyte mass
How Reaction Mode Works in ICP-QMS
Reaction
product ion
On-mass
interference
Analyte
Reaction
gas
Interferenc
e M+
MR+
Analyte and
interfering ions
enter reaction
cell
Quad set to original
analyte mass – rejects
interference product
ion(s)
Analyte
Interference
reacts to form
product ion
6 October 15, 2013
2. Mass-Shift Measurement: Reactive analyte reacts with chosen cell
gas, is moved to a new product ion mass and can be separated from
unreactive interferences
Reactive analyte is converted to product ions at a new mass –
interferences remain at original mass and are rejected by analyzer quad
How Reaction Mode Works in ICP-QMS
Original
interfering
ion
On-mass
interference
Analyte
Reaction
gas
Analyte
M+ MR+
Analyte and
interfering ions
enter reaction
cell
Quad set to analyte
product ion mass –
rejects original
interfering ions
Analyte
product ion
Analyte reacts
to form product
ion
7 October 15, 2013
Limitations of Reaction Mode in ICP-QMS
Limitations of reactive cell gases in quadrupole ICP-MS are well-
documented:
All ions enter the cell, affecting reaction processes and product ions formed.
Gives variable results when sample type/matrix or co-existing analytes change
Product ions from matrix or other elements can create new overlaps on analytes
Analyte product ions can be overlapped by other analytes/matrix elements
Can tandem MS configuration (ICP-MS/MS) address the variability
caused by co-existing elements and changing matrix components?
Agilent 8800 ICP-MS/MS matches 7700
performance in He mode, but biggest
benefit should be controlled and consistent
MS/MS operation in reaction mode.
8 October 15, 2013
New 8800 ICP Triple Quad MS – unique capabilities
Low flow
sample
introduction
system
High matrix
introduction
(HMI)
technology
Fast, frequency-
matching 27MHz
RF generator Efficient twin-turbo
vacuum system
Dual conical Extraction and
Omega lens focuses ions
across the mass range
9 orders
dynamic range
electron
multiplier (EM)
detector
Analyzer quad Q2:
High frequency
hyperbolic quadrupole
– selects ions that
pass to detector
High-transmission,
matrix tolerant interface
First quad Q1: High frequency
hyperbolic quadrupole mass filter –
selects ions that enter the cell
3rd generation collision/
reaction cell (ORS3)
with 4 cell gas lines
Peltier-
cooled
spray
chamber
Robust, high-temperature
plasma ion source
9 October 15, 2013
1. On-Mass Measurement: Unreactive analyte does not react with
chosen cell gas, remains at original m/z and so can be separated from
reactive interferences. No new cell-formed interferences can occur at
the analyte mass, since all non-target masses are rejected by Q1
With ICP-MS/MS, Q1 rejects all non-target masses, ensuring no new
analyte/matrix product ions can form new overlaps on original analyte mass
How Reaction Mode Works in ICP-MS/MS
Reaction
product ion
On-mass
interference
Analyte
Off-mass
interference
Reaction
gas
Interferenc
e M+
MR+
Q1 set to analyte
mass – rejects
all non-target
masses
Q2 set to original
analyte mass – rejects
any off-mass product
ion(s)
Analyte
Interference
reacts to form
product ion
All non-target
masses
10 October 15, 2013
2. Mass-Shift Measurement: Reactive analyte reacts with chosen cell
gas, is moved to a new product ion mass and can be separated from
unreactive interferences. No existing ions can overlap new analyte
product ion, as all non-target masses are rejected by Q1
With ICP-MS/MS, Q1 rejects all non-target masses, ensuring no existing
ions (analyte, matrix, or polyatomic) can overlap new analyte product ion
How Reaction Mode Works in ICP-MS/MS
Original
interfering
ion
On-mass
interference
Analyte
Off-mass
interference
Reaction
gas
Analyte
M+ MR+
Q2 set to analyte
product ion mass –
rejects original
interfering ions
Analyte
product ion
Analyte reacts
to form product
ion
Q1 set to analyte
mass – rejects
all non-target
masses
All non-target
masses
11 October 15, 2013
MS/MS Product Ion Scan - Titanium
Ammonia gas
introduced into
the cell
Q1 set to m/ƶ 48
Q2 scans
spectrum
Spectrum
displays all
resulting ions
from m/ƶ 48
ONLY
October 15, 2013
Confidentiality Label
12
TiNH(NH3)3
TiNH2(NH3)3
Ti(NH3)4
TiNH(NH3)4
TiNH2(NH3)4
Ti(NH3)5
TiNH2(NH3)5
TiNH(NH3)5
Ti(NH3)6
“Ti” TiNH
TiNH(NH3) TiNH(NH3)2
Resulting spectrum might look complicated but all peaks are
only from m/ƶ 48 single transitions can be selected…
Peaks corresponding to TiNH2(NH3)4 and Ti(NH3)6 selected for
Neutral Gain Scans
Q1 set to let in only m/ƶ 46, 47, 48, 49, 50 INDIVIDUALLY
Q2 set to transition masses:
Q1 +84amu [TiNH2(NH3)4]
Q1 +102amu [Ti(NH3)6]
Therefore transitions used:
46 130 46 148
47 131 47 149
48 132 48 150
49 133 49 151
50 134 50 152
MS/MS Neutral Gain Scan – Titanium with
Ammonia Adducts at Two Mass Transitions
October 15, 2013
Confidentiality Label
13
Q2 set to Q1 +84amu [TiNH2(NH3)4] & +102amu [Ti(NH3)6]
Serum and Urine Check
Based upon results from this HNO3 standard the system was run using basic
preparation for clinical samples (dilution into NH4OH, EDTA, Triton-X, BuOH)
• Standards were prepared in the basic preparation medium
• Standard addition NOT used
• Both diluted serum and urine (10x) run within same batch against same
calibration
• Data compared to no cell gas and other gas modes
Ti in clinical matrices
Target 47 -> 131 Ti [ NH3 ] 48 -> 132 Ti [ NH3 ] 49 -> 133 Ti [ NH3 ] 50 -> 134 Ti [ NH3 ]
Sample Name Conc. [ ug/l ] Conc. [ ug/l ] Conc. [ ug/l ] Conc. [ ug/l ]
Urine Blank 4.6 (2.2-7.0) 2.80 2.79 2.92 2.49
Urine Blank 4.6 (2.2-7.0) 3.50 2.93 3.33 2.66
Urine Trace Elements 14.81 15.27 14.42 15.13
Urine Trace Elements 14.99 15.49 15.50 15.05
Serum L1 1.28 (0.86-1.80) 1.21 1.15 1.14 0.80
Serum L1 1.28 (0.86-1.80) 1.27 1.18 1.09 0.89
Serum L2 1.76 1.92 1.61 1.13
Serum L2 1.82 1.64 1.76 1.22
Multiple calibrations for qualifier isotopes
Allows measurement of SeO+ at product ion mass, after removal of original
Ar2+/REE++ interference, and existing ions at SeO+ product ion mass.
Se by ICP-MS/MS Mass-Shift with O2 Cell Gas
Same reaction with O2 cell gas is also used for Se on 8800 ICP-MS/MS:
80Se+ + O2 <cell gas> 96SeO+
40Ar2+, Gd++, Dy++ + O2 no reaction
BUT Q1 of 8800 rejects other ions that would overlap SeO+ at mass 96
80Se16O+
40Ar40Ar+ 160Gd++, 160Dy++
80Se+
96Zr+, 96Mo+, 96Ru+
96Zr+, 96Mo+, 96Ru+
Reaction
gas (O2)
80Se+ 80Se16O+
Q1 (m/z 80) – rejects all
ions apart from m/z 80
Q2 (m/z 96) – rejects all
ions apart from m/z 96
40Ar40Ar+
160Gd++,160Dy++
16 October 15, 2013
Gd++ Interference on Selenium
• Gd is used as a contrast agent for MRI. Concentration in
patient can be as high as 1ppm (sometimes even higher)
• This causes problems with obtaining accurate Se
measurements for those patients
• MS/MS and O2 cell gas effectively avoids the Gd++ bias in the
data, giving consistent recovery for Se (target 89ng/mL)
No Gas Optimized O2
Serum + Gd 250 ppb 99.87 91.38
Serum + Gd 500 ppb 112.12 91.70
Serum + Gd 1000 ppb 121.07 91.78
17 October 15, 2013
8800 Abundance Sensitivity in Single-Quad Mode
Trace 55Mn is overlapped by 56Fe in 1000ppm Fe
Abundance Sensitivity (AS) is
the degree of peak “tailing” – the
contribution a peak makes to the
adjacent (-1 and +1amu) masses
7700 specification is 5 x 10-7 on the
low mass side
Note: log intensity scale
18 October 15, 2013
Other Sample Types – Manganese in Whole Blood
Mn is difficult to measure by ICP-
QMS at natural (sub-ppb) levels in
whole blood, due to “tail” of 56Fe
(and 54Fe) peak across 55Mn
8800 MS/MS ensures 55Mn is
completely separated from 54/56Fe
Overlaid spectra show blank whole
blood (10x dil) & 500ppt Mn spike
19 October 15, 2013
Conclusions
• Efficient He KED mode suitable for many routine clinical labs
– 8800 can be used in He SQ mode for these elements
• MS/MS reaction cell modes can overcome situations where KED mode is not sufficient:
– Lower LOD requirements (similar to HR-ICP-MS)
– Multiple spectral interferences other than polyatomics on same isotope
• Unique capabilities for addressing abundance sensitivity issues
• Ease of use is similar to 7700 single-quad ICP-MS • Same MassHunter software
• Many of the same hardware components
• Capital and running costs significantly lower than other high-end techniques such as SF-ICP-MS
• ICP-qqq-MS already established in many leading R+D labs, and one system soon to be installed in NHS lab
20 October 15, 2013
Agilent 8800 QQQ
ICP-MS
The Market Leader in Atomic Spectroscopy
Agilent MP-AES
Agilent 7700 ICP-MS Agilent ICP-OES Agilent AAS
21
Thank You for your attention
Questions?