+ All Categories
Home > Documents > 2271-003-PD-I UviLux Datasheet 2141systems.fastwave.com.au/media/1413/uvilux-datasheet.pdf · IP...

2271-003-PD-I UviLux Datasheet 2141systems.fastwave.com.au/media/1413/uvilux-datasheet.pdf · IP...

Date post: 03-Aug-2020
Category:
Upload: others
View: 4 times
Download: 0 times
Share this document with a friend
4
UviLux Miniature In-Situ UV Fluorometer APPLICATIONS Polycyclic Aromatic Hydrocarbon (PAH) Crude and refined oil spill monitoring Hydrocarbon detection around drilling rigs Sub-sea pipeline leak detection Airport apron pollution runoff Coastal pollution monitoring Biogeochemical oceanography Environmental impact assessment Biomass indicator studies (CDOM/Tryptophan) Waste/recycled water quality monitoring Effluent/sewage misconnect detection Main Image: UviLux shown in use with PDA Inserts: Flow Through Chamber, Roamer Wireless Data-Link FEATURES Small, low cost in-situ fluorometer Low power consumption Digital output in calibrated engineering units Range of wavelengths available Low noise, high sensitivity Internal referencing for long term calibration stability Low turbidity breakthrough High rejection of ambient daylight User adjustable dynamic range User selectable sampling rate, 0.1Hz - 3Hz Acetal C housing Easy integration with RS232, 4-20mA, 0-5V, RS422 and SDI-12 interfaces DESCRIPTION The UviLux is an innovative, sensitive, low cost, in-situ digital UV fluorometer providing many unique features. Variants of the UviLux enable real-time monitoring of the following parameters: Polycyclic Aromatic Hydrocarbon (PAH); Coloured Dissolved Organic Matter (CDOM); Tryptophan-like fluorescence, which is a surrogate marker for Biological Oxygen Demand (BOD) and is associated with bacterial contamination in waste, recycled and natural water supplies; and Optical Brighteners used for detecting household wastewater misconnections.
Transcript
Page 1: 2271-003-PD-I UviLux Datasheet 2141systems.fastwave.com.au/media/1413/uvilux-datasheet.pdf · IP rating: IP65 (splash proof) Connector: Power Jack Cable lengths available 1-25m instrument

UviLux Miniature In-Situ UV Fluorometer

APPLICATIONS

• Polycyclic Aromatic Hydrocarbon (PAH)

• Crude and refined oil spill monitoring

• Hydrocarbon detection around drilling rigs

• Sub-sea pipeline leak detection

• Airport apron pollution runoff

• Coastal pollution monitoring

• Biogeochemical oceanography

• Environmental impact assessment

• Biomass indicator studies (CDOM/Tryptophan)

• Waste/recycled water quality monitoring

• Effluent/sewage misconnect detection

Main Image: UviLux shown in use with PDA Inserts: Flow Through Chamber, Roamer Wireless Data-Link

FEATURES

• Small, low cost in-situ fluorometer

• Low power consumption

• Digital output in calibrated engineering units

• Range of wavelengths available

• Low noise, high sensitivity

• Internal referencing for long term calibration stability

• Low turbidity breakthrough

• High rejection of ambient daylight

• User adjustable dynamic range

• User selectable sampling rate, 0.1Hz - 3Hz

• Acetal C housing

• Easy integration with RS232, 4-20mA, 0-5V, RS422 and SDI-12 interfaces

DESCRIPTION The UviLux is an innovative, sensitive, low cost, in-situ digital UV fluorometer providing many unique features. Variants of the UviLux enable real-time monitoring of the following parameters: Polycyclic Aromatic Hydrocarbon (PAH); Coloured Dissolved Organic Matter (CDOM); Tryptophan-like fluorescence, which is a surrogate marker for Biological Oxygen Demand (BOD) and is associated with bacterial contamination in waste, recycled and natural water supplies; and Optical Brighteners used for detecting household wastewater misconnections.

Page 2: 2271-003-PD-I UviLux Datasheet 2141systems.fastwave.com.au/media/1413/uvilux-datasheet.pdf · IP rating: IP65 (splash proof) Connector: Power Jack Cable lengths available 1-25m instrument

The UviLux uses a deep UV LED light source and a miniature Photomultiplier detector to provide state-of-the-art measurement sensitivity at the parts-per-trillion level. Fluorometer sensitivity and hence measurement dynamic range is fully user configurable for different applications. Sophisticated electronic signal processing, combined with cowl design principles employed in our UV AquaTracka, enables the UviLux to operate successfully in high levels of ambient light. The high quality optical filtration used in both the excitation and emission optical paths gives excellent turbidity rejection, while internal referencing of the LED light source intensity provides long-term calibration stability.

The UviLux can be deployed in a wide range of coastal, oceanographic and fresh water applications. The UviLux has a chemically inert acetal resin housing suitable for deploying independently as well as from submersible vehicles, moored or profiling systems. This robust, compact, lightweight fluorometer has low power consumption, is easy to use and gives accurate and repeatable measurements.

Two signals are obtained from the standard UviLux: a digital RS232 serial output in engineering units and a calibrated analogue voltage between 0.5V to 5V. Single RS422, SDI-12 and 4-20mA outputs are also available as options. This flexibility makes the UviLux ideally suited for integrating into many different systems and platforms.

The Windows based Tplot software supplied with the fluorometer allows the user to both plot and record time stamped data when operating the UviLux directly from a PC and gives control over many instrument parameters, including: sampling rate, detector sensitivity and calibration factors. Although small and low cost, the UviLux maintains the

performance standards and build quality associated with CTG’s range of in-situ fluorometers.

SPECIFICATIONS

PAH CDOM Tryptophan

Optical Brighteners

Excitation 255nm 255nm 280nm 350nm

Emission 360nm 450nm 360nm 450nm

Sensitivity 0.005µg/l

5ppt

(Carbazole)

0.002µg/l

2ppt

(Perylene)

0.02µg/l

20ppt

(Tryptophan)

0.1mg/l

100ppb (Commercial Detergent – sensitivity equivalent to 1:20,000 dilution of typical load)

Calibrated Range 0.005 - 200ug/l 0.002 - 15ug/l 0.02 - 800µg/l 0.01 – 2500mg/l

Depth Rating 600m

Interfaces Digital Output RS232 or

SDI-12

Analogue Output 0.5 – 5V or 4 – 20 mA

0.5 – 5V or 4 – 20 mA

or RS422

Input Voltage 9 - 36V

Power Requirement <1Watt @ 12V

Operating Temperature - 2ºC to + 40ºC

Storage Temperature - 40ºC to + 70ºC

Size Ø70mm x 149mm

Housing Material Acetal C

Weight 800g in Air / 150g in Water

Connector Subcon MCBH6M

Detector Gain Fixed (factory set) or Auto (selectable)

Analogue Modes Linear or Logarithmic

Data Validation Data validation flags provided

Page 3: 2271-003-PD-I UviLux Datasheet 2141systems.fastwave.com.au/media/1413/uvilux-datasheet.pdf · IP rating: IP65 (splash proof) Connector: Power Jack Cable lengths available 1-25m instrument

ACCESSORIES

The standard UviLux fluorometer is supplied with transit case, handbook and Windows based Tplot software. There is now a range of accessories available to compliment the UviLux:

Flow Through Chamber (2260-019)

The Flow Chamber has been designed for in-line measurement applications. Incoming flow is directed across the face of the measurement window to minimise the build-up of silt while also ensuring good sample mixing. It is rated to 10Bar.

Lux Cable (2176-023), connects Roamer or USB Dongle to Lux fluorometer (terminated, 5m)

A standard 5m cable is available to provide a connection between the UniLux/TriLux/UviLux fluorometers and the Roamer or USB dongle. This cable comprises a 600mm length of neoprene cable with moulded connector and locking sleeve for attachment to the UniLux/TriLux/UviLux fluorometers spliced to a thinner polyurethane clad cable with a connector suitable for both the Roamer and the USB Dongle. Cable lengths up to 25m can be supplied.

Lux Cable (2176-026), connects Lux fluorometer to user’s system (un-terminated, 5m)

An un-terminated 5m cable is available to provide a connection between the UniLux/TriLux/UviLux fluorometers and user defined system. This cable comprises a 600mm length of neoprene cable with moulded connector and locking sleeve for attachment to the UniLux/TriLux/UviLux fluorometers spliced to a thinner polyurethane clad un-terminated cable for configuration to user defined system. Cable lengths up to 25m can be supplied.

Flying Lead (600mm) with locking sleeve (2217-015)

A standard 600mm cable with moulded connector and locking sleeve at one end and bare conductors at the other is available for integrating the Lux fluorometers into user defined system.

Page 4: 2271-003-PD-I UviLux Datasheet 2141systems.fastwave.com.au/media/1413/uvilux-datasheet.pdf · IP rating: IP65 (splash proof) Connector: Power Jack Cable lengths available 1-25m instrument

USB Interface Dongle for PC power/communications (2176-007) The Lux USB Interface Dongle connects the UviLux to a USB port on a computer. The interface uses the 5V supply from the USB port and converts it to 13.7V to power the fluorometer. Using the drivers supplied the sensor output is then accessible to the Tplot software on the host PC as a serial port.

Mechanical

Size: Ø26.5mm x 145mm

Weight: 100g Housing: Acetal C Connector A: USB A plug Connector B: Fischer quick fit connector

Electrical

Input voltage: 5Vdc (from USB port) Input current: 500mA dc (max) Power requirements: <2.5 Watt @ 5V Voltage output: 13.7Vdc Power output <2 Watt Comms (instrument): RS232 Comms (PC): USB

Roamer Wireless Data Link (2176-032) The Roamer supplies the Lux fluorometers with power as well as acting as an interface to send and receive data over a Bluetooth wireless link to either a PDA (operating a CTG Windows Mobile application) or a PC (operating the Tplot software). When used in conjunction with a PDA, the software provided with the Roamer enables users to record and display (in either tabular or graphical form) Lux data. If a GPS enabled PDA is used, all data can be position stamped.

Performance

Wireless range: 50m in line of sight

Battery Life: 4hrs (continuous)

Recharge time: 1hr

Mechanical

Size: 145mm x 94mm x 26mm

Weight in Air: 340g

IP rating: IP65 (splash proof) Connector: Power Jack

Cable lengths available

1-25m instrument cable

Electrical

Input voltage: 9 to 12Vdc

Data Output: Bluetooth Serial Stream

Power requirements: <0.5 Watt @ 12 volt

Software

PDA: Windows Mobile 5 application enabling recording and display of data to PDA

PC: Tplot software compatible with Microsoft Windows.

Personal Digital Assistant (PDA) The PDA receives the UviLux data over Bluetooth and displays and logs it. CTG can supply the PDA complete with software, or if client wishes to use his own PDA/PC, the application software is provided. The client’s own PDA needs to operate with Windows Mobile 5, have Bluetooth connectivity and a touch screen. The PC requires a Bluetooth connection and Windows XP/Vista.

In view of our policy of continual improvement, the designs and specifications of our products may vary from those described (2271-003-PD-I UviLux)

Chelsea Technologies Group Ltd 55 Central Avenue West Molesey Surrey KT8 2QZ United Kingdom Tel: +44 (0)20 8481 9000 Fax: +44 (0)20 8941 9319 [email protected] www.chelsea.co.uk


Recommended