2nd Annual Middle East Sulphur Plant
Operations Network Forum
Abu Dhabi
Practical Information from
SRU-TGTU On-Line Analyzers
Stephane CAPPE / Randy HAUER
1
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Agenda
Summary…Eight SRU-TGTU analyzer tags with detailed discussion of:
Acid Gas Feed Forward (Total HC, BTEX, H2S, NH3,CO2, H2O)
Tail gas (H2S, SO2)
After acid gas is introduced when can the analyzer be put in control ?
Control signal, over range, best practices
TGTU (H2, H2S, COS, CS2)
Quench or Absorber where to place the H2
Absorber outlet; H2S/COS/CS2 ranges
Sulfur Pit Gas Analyzer (H2S, SO2)
What constitutes abnormal operation conditions
O2 (start up) Analyzer (O2, CO)
2
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
3
SRU-TGTU Analyzer Tags
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Acid Gas Analyzer
Feed Forward Control
New Analysis & Control Schemes
possibilities
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Field Data / HC Process Upset
H2S swings from 83 to 87%
THC =0.2%
THC = 2.5%
12 fold > in THC
from 0.2 to 2.5%
lasting ~3 minutes
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Acid Gas Analyzer
Real time analysis of acid gas using a UV-IR Analyzer
IR is to quantify the Total HC (as methane equivalent)
UV to quantify the H2S
CO2, H2O as required (ABC+)
BTEX to know when to co-fire for higher reaction furnace temperatures
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Feed Forward Analysis
Actual Control Response to Hydro-Carbons (Courtesy of Jacobs Comprimo)
7
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Acid Gas Analysis - Benefits
Improved control, fewer upsets, mitigation of upsets
Demonstrated improvement of ~0.5% recovery efficiency (ER)but…
More to the point it is the loss of 5% ER for 1 hour that is the benefit
Prevention of SO2 breakthrough to the TGTU
Can (should) be used for mass balance recovery efficiency calculation
The objective is to place the analyzer in cascade control
THC/BTEX Speciation is the new focus especially in ME
8
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
The Heated Acid Gas (HAG) Probe
Membrane for liquid ingress, no water removal, no pump
Maintained at process temp, spent sample returned back to process,
Integral process isolation valves, safety is the paramount feature
To be “ready”…Install the HAG at
Turnaround
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Tail Gas Analyzer
Feed Back Control
Air Demand vs Ratio
Start Up and Shutdowns
When to put the analyzer on line
Tail Gas Analyzer
Top of the Pipe Analyzer
880 NSL
Sample Line Analyzer
900 ADA
Insulated steam
jacketed nozzle
Weather cover
for ASR
probe and jacket
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Ratio vs Excess H2S vs Air Demand
Ratio is the non linear expression of 2 H2S / SO2
Excess H2S is the linear expression of
[H2S] - 2[SO2] = -1%…0…+1% H2S
Air Demand uses a scaling factor
AD{[H2S] - 2[SO2]} = -5%...0...+5% AIR
REMINDER
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Controlling “Off Ratio”
To favor the TGTU with H2S
The tail gas analyzer simply measures [H2S] and [SO2]
And calculates a control algorithm….
A dumb analyzer we ask to do a smart thing
Running “Off ratio”
Typically 4:1 or 5:1 but this is just a notional value…
Perhaps better to express it as -1% air demand (~+0.2% xs H2S)
Let’s not be slave to 2:1
13
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Tail Gas: Process Upset & Over-Range Event
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
“After start up when is the
analyser put into control
mode ?” Catalyst preference for H2S over
SO2 can skew true air control.
Sulphur Experts recommends
control mode after 8 hours.
15
Putting the Analyzer in Control After Start-up
“Do you consider tail gas
analysers to be reliable ?” Thankfully yes was common
response
Responses capture the evolution
of the tail gas products.
Time for a 3rd generation.
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
SRU Start-up O2 /CO Measurement
• For start-ups and shut-downs
Requires excess O2 to near
stochiometric conditions of 0.1% xs air
Tunable Diode Laser (TDL) process O2
analyzer
Can monitor CO also
• A fixed (permanent) system or portable
16
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
TGTU Analyzer
Where to install ?
COS/CS2 analysis on top of Absorber
SO2 breakthrough analysis
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Analytical Measurements on a TGTU
H2 is the critical parameter ( so please no GC - ESD analyzer)
H2S can be measured at the Quench outlet or Absorber outlet
[H2S] at Quench outlet is a measure of SRU recovery efficiency
[H2S] at Absorber outlet is a measure of absorber efficiency
COS & CS2 are secondary measurements
[COS] is indication of CoMo catalyst activity (first to breakthrough)
[CS2] is also indication of catalyst activity and contribution to emissions
SO2 can be measured at the quench inlet or outlet (not common)
pH is measured in the quench water (not addressed here)
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Where to install TGTU analyzer
Installed Base of ~175 Analyzers
~140 have sample point at top of Absorber (H2/H2S)
25 of these also measure COS
8 of these also measure COS + CS2
~30 have sample point at Quench outlet (H2/H2S only)
~5 have sample point at Quench inlet (SO2 breakthrough)
Finding right balance between Analyzer complexity
and Benefit of measurement for Operations
19
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
SO2 0 – 100
ppm
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
TGTU (COS & H2S) Steady State & Upset
21
COS
H2S
H2S
COS
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Sulfur at Absorber Outlet After SO2 Upset
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Sulphur Pit Analyzer
The neglected one…
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Sulfur Pit: Solubility of H2S in Sulfur
• Produced sulfur has 200-350 ppm
of dissolved H2S
• Spontaneous degassing and
concentration in the gas phase can
increase to explosive levels (3.5%)
• Pit gas analyzer requires same
sample integrity as tail gas
analyzer ( Sulphur Vapor )
• H2S is measured to warn of LEL,
SO2 is measured to warn sulfur fire
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
H2S / SO2 values
S-Tank sweep gas
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Sulphur Pit High H2S & SO2 Values
(Just prior to an incident)
26
H2S
SO2 CFD modeling confirms
H2S can exist ~16x
The mean value
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
SRU Emission Analyzer
Mass Emission
TGTU bypass situation
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Emission Analyzers
“The Color of Plumes” (courtesy of S-Experts)
White: Steam plume caused by water condensation (cold, Canada)
Bluish white: SO3 plume usually caused by high SO2 emissions
Ask analyzer techs to watch for “green slime” (chromium from 316SS + S-acid)
Orange: NOx plume
Brown: Unburned hydrocarbon / soot plume
Green: Burning H2S plume
Dual Range of SO2 if TGTU can be bypassed (Habshan 5)
Mass Emission if straight Claus SRU (no TGTU)
28
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Major Problems Faced
&
Solutions
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
What are the common problems that Operations
have with Analyzers ? • Frequent choking problems !!!
• Sampling system failure
• Temperature Impacts??
– Is it controlled at each locations with respect to sulfur dewpoint?
30
No heating/
Jacketing
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Common problems …
• Frequent downtime due to maintenance issues
– Lack of SRU Analyzer trainings
– Not much experience on SRU process
– Frequent shifting/changing of site engineers
• High H2S exposure…. SAFETY : People and Environment
• Unpredicted behavior of Analyzer & readings
ex. Unit 52 at Gasco Habshan
31
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
GASCO Habshan DCS
Tail Gas Trends
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Remedies…
• Regular control visits on SRU analyzers by expert
->if problem persist more than 2 times call Manufacturer
• Sites have “generalist” analyzer tech not SRU experts (Do you have?)
• Need SRU Process Trained Analyzer Service Engineers
• Need experienced hands on CBA & other SRU process
• Your Service Staff requires periodically training and evaluation by OEM…
33
CLEAR VISION SOUND STRATEGIES SOLID PERFORMANCE
Solutions for Ametek SRU analyzers in Middle East
• AIMS is our local partner having Focussed SRU experience …
• Big service team with extensive knowledge on all SRU process
analyzers
• First in kind analyzer successfully commissioned for
H2S in Lean Amine & BTEX Analyzer at GASCO Habshan
• AIMS involved with all EPC and System Integrators for
Commissioning and Start Up on AMETEK for Middle East
• Target is 100% uptime of our Analyzers at all sites in GCC
AIMS is our Sulphur Police squad, call them if problems !