+ All Categories
Home > Documents > Adaptive Control of Foot Orthosis

Adaptive Control of Foot Orthosis

Date post: 09-Apr-2018
Category:
Upload: niteen-wakulkar
View: 226 times
Download: 0 times
Share this document with a friend

of 40

Transcript
  • 8/8/2019 Adaptive Control of Foot Orthosis

    1/40

    Technical report, IDE0704, January 2007

    Adaptive Control of Foot Orthosis

    Masters Thesis in Electrical Engineering

    Ali Manzoor, Hesham Elkhbai, Ziad Ekwaneen

    Supervisor: Ulf Holmberg

    School of Information Science, Computer and Electrical

    Engineering

    Halmstad University

  • 8/8/2019 Adaptive Control of Foot Orthosis

    2/40

    Adaptive Control of Foot Orthosis

    Masters thesis in Electrical Engineering

    School of Information Science, Computer and Electrical Engineering

    Halmstad University

    Box 823, S-301 18 Halmstad, Sweden

    January 2007

  • 8/8/2019 Adaptive Control of Foot Orthosis

    3/40

    Preface

    First and foremost we wish to thank our parents for their utmost moral support and freedom we

    needed to move on and to achieve this milestone. We would also like to thank our advisors Ulf

    Holmberg and Wolfgang Svensson, who were always supportive since the days we begin

    working on this project. They have supported us not only by providing a research assistantship,but also academically and emotionally through the rough road to finish this thesis.

    Ali Manzoor, Hesham Elkhbai & Ziad Ekwaneen

    Halmstad University, January 2007

  • 8/8/2019 Adaptive Control of Foot Orthosis

    4/40

    Abstract

    Major problems of the Foot Drop treatment are expensive and complex solutions. This work

    presents the performance of a new inexpensive method named as Semi-Active Ankle Foot

    Orthosis (SAAFO). The concept of this approach is to use inexpensive sensors to detect foot step

    movement. The signals from the sensors afterwards will be fed to a control system of SAAFO inruntime for a smooth foot movement of a drop foot patient while walking. Different sensors have

    been studied in detail along with comparison to the proposed sensor system and mechanical

    design. The signals from the sensors are used to detect different phases of human walking. These

    sensors are placed at different positions on an orthosis and their signals are studied in detail.

    Experiments have been done in different conditions to get a realistic picture either this assembly

    can be implemented commercially. Signals are plotted and discussed yielding that the human

    walking phases can be easily and accurately detected using inexpensive sensor assembly.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    5/40

    Tables of Contents

    1 Introduction.................................................................................................................................1

    1.1 Foot Drop...............................................................................................................................1

    1.2 Aim of the Project.................................................................................................................31.2.1 Motivation.....................................................................................................................3

    1.3 Background...........................................................................................................................4

    1.3.1 Human Gait..................................................................................................................4

    1.3.2 Passive Ankle-Foot Orthosis (P-AFO).......................................................................5

    1.3.3 The Active Ankle Foot Orthosis (AAFO)..................................................................6

    1.3.4 Semi-Active AFO (DACS-AFO).................................................................................7

    2 System Overview.........................................................................................................................9

    2.1 Sensors....................................................................................................................................9

    2.1.1 Sensor Selection.............................................................................................................92.1.2 Sensors used in AFO.....................................................................................................9

    2.1.3 Strain guage sensors...................................................................................................10

    2.2 Sensor placement.................................................................................................................13

    2.2.1 How strain guage used in project..............................................................................14

    2.3 SSAFO Design......................................................................................................................15

    2.3.1 Physical componant of SSAFO..................................................................................15

    2.3.2 SSAFO principle of work...........................................................................................18

    3 Experimental Methods..............................................................................................................20

    3.1 Walking horizontal..............................................................................................................203.2 Walking up-hill....................................................................................................................21

    3.3 Walking down-hill...............................................................................................................21

    4 Results and Discussion...............................................................................................................23

    5 Conclusion..................................................................................................................................32

    References

  • 8/8/2019 Adaptive Control of Foot Orthosis

    6/40

    Adaptive Control of Foot Orthosis

    1

    Introduction

    There are millions of individuals with gait disabilities, requiring either rehabilitation or

    permanent assistance. In the US alone, approximately 866,000 people use an orthosis on a lowerextremity. Current assistive methods are usually labor intensive, highly variable, or require

    constant tuning. Applying robotic technology to gait could provide a means to quantify lowerbody kinematics and kinetics, improve and monitor locomotion during normal and clinic uses andhelps reduce the high-energy cost of disabled locomotion. Further, a robotic assistive device

    could adapt to human gait, which is consistently changing in both speed and form. In initial work,

    adaptive hip and knee orthosis have been able to provide more natural gait for the disabled [1].

    1.1 FOOT DROP

    Definition

    Foot drop is a weakness of muscles that are involved in flexing the ankle and toes. It

    describes an abnormal neuromuscular (nerve and muscle) disorder that affects the patient's

    ability to raise their foot at the ankle. Drop foot is further characterized by an inability topoint the toes toward the body (dorsiflexion) or move the foot at the ankle inward or

    outward. Pain, weakness, and numbness may accompany loss of function [2]. Walking

    becomes a challenge due to the patient's inability to control the foot at the ankle. The footmay appear floppy and the patient may drag the foot and toes while walking.

    Figure 1-1: Drop FootTreatment

    The type of treatment is dependent on the underlying cause of drop foot. Foot drop thatcannot be treated by surgery is often treated using the stimulation of the affected nerves.

    The stimulation is applied as the foot is raised during a stride and is stopped when the foot

    touches down on the ground. Some patients may be fitted with an Ankle Foot Orthosis(AFO), brace, or splint that fits into the shoe to stabilize the ankle/foot.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    7/40

    Introduction

    2

    Recovery and rehabilitation

    Depending on the nature of the cause of foot drop, recovery can be partial or complete.

    Physical therapy and an ankle foot orthosis device worn in the shoe are important aspects ofrehabilitation. [3]

  • 8/8/2019 Adaptive Control of Foot Orthosis

    8/40

    Adaptive Control of Foot Orthosis

    3

    1.2 Aim of the project

    This project is aimed to use a special orthosis device. This device will help to provide normalrange of motion to the foot and ankle during walking.

    Motivation

    The major motivation behind this project is to assist those individuals with gait disabilities. The

    goal is to build lightweight, portable and easy controllable adaptive assistive device for Foot Drop

    disability. To achieve this two main parts are suggested.

    1. Detecting the human walking gait phases. It will be achieved by using simple sensorstructure, strain gauge sensor, to measure the force exerted by foot while walking.

    2. Control signals depending on these sensor signals will be used to control the mechanicalstructure and to keep the walking as smooth as possible.

    3. Magneto-Rheological (MR) Damper technology with controllable dampingcharacteristics will be used in this project, to generate the braking strategy for the orthosisstructure in the suitable position, which has not been used so far for this purpose. It will

    help to operate the Ankle Foot Orthosis (AFO) as a semi-active ankle foot Orthosis [SA-

    AFO].

  • 8/8/2019 Adaptive Control of Foot Orthosis

    9/40

    Introduction

    4

    1.3 Background (Related Works)

    1.3.1 Human Gait

    Definitions:

    To explain the idea of the human gait, some technical terms are needed to be explained

    first.

    Stride Cycle is the time interval of occurrence of repetitive events of walking, i.e. frominitial heel lift to the following initial heel lift.

    Dorsiflexion/Plantar Flexion is act of bending a joint; especially a joint between the bonesof a limb so that the angle between them is decreased. [12] In this case bending of the limb

    with respect to the foot.

    Gait is a particular way or manner of moving on foot. Walking and running are the two

    basic human gaits [13]

    Gait Cycle is defined as the time interval between two successive occurrences of one of therepetitive events of walking. Although any event could be chosen to define the gait cycle, it

    is generally convenient to use the instant at which one foot contacts the ground [14]. The

    gait cycle consists of Stance and Swing. Stance is the period in which the foot is in contactwith the ground and Swing is the period in which the foot is in the air. Stance comprises

    60% of the walking cycle. While 40% is the swing.

    The following terms are used to identify major events during the gait cycle:

    1. Initial contact2. Opposite toe off3. Heel rise4. Opposite initial contact5. Toe off6. Feet adjacent7. Tibia vertical

    These seven events subdivide the gait cycle into seven periods, four of which occur in the

    stance phase and three in the swing phase. [14]

  • 8/8/2019 Adaptive Control of Foot Orthosis

    10/40

    Adaptive Control of Foot Orthosis

    5

    Figure 1-2

    The duration of a complete gait cycle is known as the cycle time, which is divided into

    stance time and swing time [14].

    There are four major objectives for the design of any Orthosis [1]:

    Restoration of normal function and ability to:control of motion

    correction of deformity

    compensation for weakness

    Make it as comfortable to wear as possible.Minimize the abnormal appearance of the Orthosis.

    Lightweight, thickness of sensor, accuracy, range, and speed.

    Most advanced Ankle Foot Orthosis (AFO) has not been able to improve on all four

    objectives from the basic AFO.

    There can be many different ways to treat Drop foot problem as mentioned before. Beloware explained main methods, which have been used so far.

    Since the project is divided into two parts, sensor part and actuating part , previous projectswill be mentioned on these two criteria , the actuating part will consist of : passive , active

    and semi active actuating system and the sensor part will consist of : Force sensors ,

    pressure sensors , special sensors ( which depends on each research method) and the straingauge sensors .

    1.3.2 Passive Ankle-Foot Orthosis (P-AFO)

    Most orthosis devices provide passive support to the Ankle Foot Orthosis joint, PassiveAnkle-Foot Orthosis (P-AFOs) constitute a special class of ankle braces designed to

  • 8/8/2019 Adaptive Control of Foot Orthosis

    11/40

    Introduction

    6

    enhance gait function by providing natural support to the lower limb as it progresses over

    the stance foot. However, a fundamental design characteristic of all P-AFOs is that they are

    not capable of replicating all dynamic characteristics of the natural ankle complex.

    Standard Ankle-Foot Orthosis (AFO)

    An ankle foot Orthosis (AFO) can be defined as a medical mechanical device to supportand align the ankle and foot, to suppress spastic and overpowering ankle and foot muscles,

    to assist weak and paralyzed muscles of the ankle and foot, to prevent or correct ankle and

    foot deformities, and to improve the functions of the ankle and foot. [5]. However, themodel currently used is still a basic mold made out of polypropylene or metal-alloy made

    to fit the ankle of the patient (Fig.1-3).

    The major advantages of polypropylene AFOs are that they distribute pressure over a larger

    surface area of the limb resulting in less discomfort. They are lighter than metal braces,

    more cosmetically appealing, and can be worn with a variety of shoe types.

    With this AFO design there is no ability of the ankle-foot to [plantar-flex] or [dorsi-flexion]

    from heel-strike through gait cycle. The ankle-foot is held only in a rigid neutral position

    during swing phase.

    Figure 1-3: Standard Ankle Foot Orthosis (AFO)

    1.3.3 The Active Ankle Foot Orthosis (AAFO)

    The Active Ankle Foot Orthosis (AAFO) developed in [7]. It consists of apolypropylene

    ankle-foot orthosispowered by a Series Elastic Actuator (SEA). They demonstrated that anadaptive impedance controller has good potential for aiding patients with drop-foot.

    The main advantage is that the SEA exhibits stable behavior while in contact with all

    environments, including in parallel with a human. However, the further complication is thatthe system used to control the AAFO needs to be on the AFO being worn by the subject.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    12/40

    Adaptive Control of Foot Orthosis

    7

    The main disadvantage is that the minimum thickness for a load cell used to measure theGround Reaction Force (GRF) and centre of pressure (COP) is 9.6 mm (0.375 in), which

    does not allow this system to be worn within a sneaker. A rubber pad could be placed under

    the metal plate to act as a sole. This, however, would make the system cumbersome.

    Force sensitive resistors (FSR) are polymer thick film devices that exhibit a decrease inresistance with an increase in the force applied to the active surface this technology allows

    for extremely thin and flexible sensors; however, they lack accuracy and repeatability.Capacitance Force Transducers are more accurate and durable than FSRs, but are also

    larger.

    Before the variable-impedance, AAFO can have broad utility for individuals suffering from

    drop foot gait, system hardware and software must be advanced. The series-elastic actuator

    used in this investigation is too heavy and power intensive to be practical in a commercially

    available active ankle foot system, and would therefore have to be redesigned. Further,control strategies and sensing architectures specifically suited for the ascent and descent of

    stairs and ramps would be necessary.

    Artificial Pneumatic Muscles active Ankle Foot orthosis:

    The basic idea was to provide plantar flexion and dorsiflexion torque during walking to thepatient. It was achieved by computer software, which adjusted air pressure in each artificial

    muscle while walking. It was a Prototype (Not portable), Heavy (mass 1.4 kg), Complex.

    [17]

    Functional Electrical Stimulation (FES)

    Functional Electrical Stimulation (FES) is a methodology that uses shorts bursts ofelectrical pulses to generate muscle contraction [4]. These pulses generate action potentials

    in motor neurons attached to a muscle, using self-adhesive skin surface electrodes placed on

    the side of the leg, which cause that muscle to contract.

    The FES system must produce pulses at a minimum of 20 Hz; otherwise the muscle

    twitches continuously and does not generate a steady output force. Both monophonic andbiphasic current or voltage pulses can be used to stimulate motor neurons. The common

    belief, however, is that the injected charge should be removed from the body and notallowed to accumulate. Most surface electrodes use biphasic current pulses, which changes

    the positions of the anode and cathode during stimulation.

    1.3.4 Semi Active Ankle Foot Orthosis

    Dorsiflexion Assist Controlled by Spring AFO (DACS-AFO)

    This type of AFO Generates a dorsiflexion assist moment during plantar flexion and no

    moment during dorsiflexion using a spring located at the calf. The initial dorsiflexion angle

    of the ankle joint is adjustable and three springs with different moments are available [6].

  • 8/8/2019 Adaptive Control of Foot Orthosis

    13/40

    Introduction

    8

    The DACS-AFO, consists of body and sole portions, which are made from polypropylene

    and are connected with two joints and a cylinder. The cylinder generates a dorsiflexion-assist moment with an angle of plantar flexion at heel strike.

    Unfortunately, no significant difference was found between the DACS-AFO and basic AFOand most subjects of their study preferred their old AFO.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    14/40

    Adaptive Control of Foot Orthosis

    9

    2. System Overview

    2.1 Sensors:

    As mentioned before, basic idea of this project is to develop semi active ankle foot orthosis,

    which is actuated by MR damped. To achieve and build this system, accurate signals for the gaitcycle and human walking should be measured in order to use these signals to operate the

    actuating system (The MR damper). And to choose the most suitable sensors, which can measure

    all gait cycle cases, several options are available such as: Force sensors and Pressure sensors,which have been used before in previous projects. Below, is the description of the procedures

    used to choose sensor type in this project.

    2.1.1 Sensor Selection:

    Under coming text is the description of the main and general criteria to choose any type of sensor

    for an application.

    Criteria for Sensor Selection:

    Range; the sensor should be able to measure the full-scale range of the event. Sufficientmargin should remain at the upper end of the range so the measured event should not

    exceed the range of the sensor and cause damage.

    Frequency response; of the sensor is the range of frequencies over which the sensor gives

    an accurate response. It should be able to measure the full range of frequencies expected

    from the experiment.

    Sensitivity; is defined as the ratio of the change in sensor output to a change in the input tobe measured.

    Accuracy; refers to how close the output of sensor is to the actual event.

    Environmental conditions; such as temperature, humidity, water and other should beconsidered while choosing a sensor.

    Cost; is an important factor and depends on the standard of the project. [15]

    2.1.2 Sensor used in AFO:

    Force Sensors:

    As its name implies, force sensors are used to measure the force. Depending on its

    application, it can also be used to measure the weight or mass.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    15/40

    System Overview

    10

    Force sensors can be integrated into an orthosis or prosthesis to measure ankle/wrist forces,

    moments and loads. It can also be used in gait analysis to measure ground reaction forces

    and loads exerted by a strap or pad in an orthosis.

    A Force Sensing Resistor (FSR) is a thin polymer device that exhibits decreasing electrical

    resistance in response to increasing force-applied perpendicular to the surface of the device.A force sensing resistor is made up of two parts. The first is a resistive material applied to afilm. The second is a set of digitizing contacts applied to another film [15].

    The force versus resistance relationship of an FSR is almost logarithmic, meaning it is non-linear. With no load on the FSR, there is no contact between the electrodes and the resistive

    material and the FSR is seen as an open circuit.

    Due to the mechanical nature, FSRs have a very slow rise time. One of the major

    limitations of FSRs is their creep response. Creep is the deformation or strain of a

    mechanical member over time when exposed to a constant force. It is important to make

    note of the fact that FSRs are not appropriate for accurate measurements of force due to thefact that parts might exhibit as much as 15% to 25% variation between each other. In FSRs

    this creep manifests itself as a decrease in resistance over time. If the FSR is used to

    measure the absolute value of a force, this creep can be a source of error, [15]. Also the sizeof FSR is considered big comparing to other sensors, and due to the Foot orthosis

    application which need a very thin sensor in order to be able to build in the foot structure

    without any difficulties. These sensors have been used in some of previous foot orthosisprojects, and they face the same problems as mentioned above.

    Pressure Sensors:

    To measure pressure, an electrical signal must be generated in response to a pressure input.Pressure is measured either by deflect or strain. This strain can then be measured in a

    variety of ways, using capacitive, PR and PE as well as other techniques to measure

    displacement.

    Pressure is defined as the force exerted over a surface per unit area of surface. Pressure is

    measured in the same units as mechanical stress N/in2 (Pascals, Pa). [15]

    The use of these sensors was limited, due to there size, same as the force sensors, non

    linearity signals , which need some signals analysis before it will be entered into the control

    system. And the use of this sensor in foot orthosis will be limited for measuring some

    points under the foot, which maybe changed from one foot to another.

    Wireless In-shoe Force System:

    The WIFS is a portable system, intended for ambulatory use in a variety of locations. It

    developed to measure times of foot contact occurrence, the approximate weight on each foot

    , and the center of pressure (COP) on each foot (as a function of time) while being reliable,long-lived, easily mounted or installed and cosmetically unobtrusive. This was done using a

  • 8/8/2019 Adaptive Control of Foot Orthosis

    16/40

    Adaptive Control of Foot Orthosis

    11

    set of thick-film force sensors mounted in multi-layer shoe insoles fitted inside the shoe.Information obtained by this insole is transmitted from small, battery powered transmitters

    mounted on the shoe to a receiver. This system consider very expensive.

    2.1.3 Strain Gauge Sensors:

    Definition

    In this section we will describe the sensors which used in our project, strain gauge sensors.

    These sensors normally used to measure the stress or strain in materials depends on there

    amount of bending due to applied force. So this bending will be proportional to the changein sensor resistance, and voltage value applied.

    Stress and strain are very closely related and it is possible to calculate one if the other is

    known. Strain, is defined as the fractional change in length per unit length (delta L/L) or thechange in length divided by the original length. Strain can be tensile (positive) or

    compressive (negative) and is a dimensionless quantity since length units cancel oneanother.

    Strain Gauge senor previously was made up of a thin wire bonded to a dimensionally stable

    substrate in a zigzag pattern. Advances in the design of the device have resulted in a photo-etched foil strain gage. This new construction technique has improved the reliability and

    sensitivity of the device while reducing its size and cost.

    These photo-etched strain gages consist of a very thin foil sheet (commonly Constantan

    alloy or nickel-chromium alloy) bonded to a dimensionally stable substrate such aspolyimide film. The foil is then etched to produce a single conductor which zigzags across

    the substrate. The gage is then attached to the component to be tested usually by bonding orgluing the strain gage to the component with cyanoacrylate or an epoxy adhesive. As the

    component is stressed, it undergoes some dimensional change or deformation. Since thestrain gage is bonded to the surface of the component, it also will undergo the same

    deformation as the component.

    Figure 2-1As the thin foil wire changes dimensions, the resistance of the wire changes. This resistancechange is very small and needs to be converted to voltage and amplified before it can be

  • 8/8/2019 Adaptive Control of Foot Orthosis

    17/40

    System Overview

    12

    measured. To convert the small changes in resistance of a strain gage into a voltage, a

    Wheatstone bridge is used.

    A Wheatstone bridge is a circuit consisting of four resistors and a power supply. One or all

    of these resistors can be strain gages. With voltage supplied to the circuit, a small change in

    resistance of any one of the four resistors will unbalance the bridge and cause a voltagechange at the output points. This voltage can then be amplified and read by a volt meter.

    The sensitivity of strain gages, called the gage factor, is the ratio of the fractional change in

    resistance to the fractional change in length along the axis of the gage. Foil strain gagestypically have gage factors near two.

    Strain gages can be useful in measuring many properties of orthosis and prosthesis. [15]

    The advantages of using strain gauge senor in this project are:

    The Size of strain gauge is very small and very thin, and available in different shapes andsizes. So it can be easily fitted inside the foot orthosis.

    The signal obtained from these sensors is a voltage signal, which can be easily and directlyused in the control system. The output signals from the strain sensors are considerably very

    small, but it can be easily amplified by using an amplification circuit.

    Sensitivity of strain gauge sensors is very high; any small change in the materials attached

    to the sensor will be directly sensed by the strain gauge sensor.

    Accuracy and repeatability of strain gauge sensors are very high also. And it was seen in

    the experiments done in this project.

    Inexpensive; these sensors are very cheap comparing to other sensors, such as force or

    pressure sensors. There cost vary and depends on the application they are used for.

    The limitation of using strain gauge sensors occurs because of there sensitivity to ambient

    temperature variations, but that can be compensated by using some certain arrangements in

    sensor circuits design. But in this project the influence of temperature variations on thestrain gauge sensor was very small, and can be neglected without using additional

    components.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    18/40

    Adaptive Control of Foot Orthosis

    13

    2.2 Sensor placement:

    The second most important task is the sensor placement on the foot in order to reflect thegait cycle cases in the right and accurate way.

    As mentioned before, gait cycle contain three phases: Heel phase, stance phase and swingphase. And the target was to determine these three phases by using the minimum number of

    strain gauge sensors with high accuracy and repeatability.

    For this purpose four main positions on the foot orthosis were chosen to measure the

    bending during the gait cycle, and analyze the signal obtained from each sensor and try to

    extract these phases from them. We numbered the sensors depends on there position as: 1,

    2, 3 and 4. See figure 2-2.

    2-2(a) 2-2(b)

    2-2(c) 2-2(d)

    Figure 2-2: Sensor placement

    Figure 2-2(a) shows four different sensor positions on the lower side of the orthosis, out ofwhich only two positions i.e. 1 & 4 were used. Figure 2-2(b) shows the discarded sensor

  • 8/8/2019 Adaptive Control of Foot Orthosis

    19/40

    System Overview

    14

    positions. Figure 2-2(c) shows the upper two sensor positions which were used in place of the

    discarded ones. Figure 2-2(d) shows the electrical equipment mounted on the orthosis.

    2.2.1 How to use these sensors in project:

    The strain sensor will be used to measure the bending in the foot during the walking, sincethe bending on foot will occur because of the force applied by the human body on the footends (heel and foot front) while walking. So by applying the force on the front fingers (F1),

    multiplying it with the distance from the front fingers to the heel (r1), a Torque (M1) will be

    generated on the foot during walking from the stance phase( Dorsiflexion direction). Andthe same happen when the heel phase happened, the force (F2) from the body and the

    distance (r2) from the heel to the end of the foot generate a torque (M2) in the other

    direction (Planterflesxion). See figure 2-3.

    Figure 2-3One of the aim of this project is to find a way to balance these two torques during the gait

    cycle from the sensor signals.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    20/40

    Adaptive Control of Foot Orthosis

    15

    2.3 Semi-Active AFO Design:

    This project is about building a Semi active AFO hardware, and test it on a real Foot drop

    patient. But due to the fact that most of the time was spent on the sensor part to find theright signals and a logical strategy to control our orthosis as per project target, not much

    was done along the mechanical side. But to compensate that, a sketch design along with theproposed working principle of the orthosis is also explained. In later paragraph, orthosisdesign, components, and working is mentioned. See figure 2-4.

    Figure 2-4: Schematic drawing of proposed semi-AAFO [SAAFO]

    The actuating system of the ankle foot Orthosis will be A Magneto-theological (MR)Damper, with controllable viscosity, which helps in controlling the damper extension in

    exact position to support the foot during the gait cycle. Figure 2-4 shows a proposal

    drawing of the semi-active ankle foot Orthosis that will be build.

    2.3.1 Physical component of AFO:

    The Semi-Active Ankle Foot orthosis (SAAFO) consist of:

    Foot orthosis structure: This is the mechanical part of the project. Due to some reasons,the mechanical building was revised and a new structure was supposed to be build. A new

    foot orthosis that will be able to hold all components as per new design. Below is defined

    the proposed structure for that orthosis, which is not that much different from the normalorthosis but with extra supports and holders.

    Sensors: for detecting and navigating the gait cycle phases, as mentioned before, these are

    strain gauge sensors built in the foot structure. These sensors are flexible and easy to installso it can be used with any type of orthosis material.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    21/40

    System Overview

    16

    MR damper: used is a compact, Magneto-Rheological [LORD RD-1005-3] damper. This

    (MR) fluid damper is suitable for industrial suspension Applications, with continuously

    variable damping which controlled by the increase in yield strength of the MR fluid inresponse to magnetic field strength. So it was fitted in AFO system in a certain design to

    control the ankle joint angle as what will be described later in the principle of work of the

    proposed semi active ankle foot orthosis.

    This new technology has been used for its great features and benefits such as:

    Fast Response Time: responds in less than 15 milliseconds to changes in the magnetic field.

    Easy to Use: provides simple electronics and straightforward controls. So no need for any

    complicated circuits or signals conversions.

    Durable: provides excellent long-term stability.

    Low power consumption: The Input Voltage to operate the MR damper is 12 Volt DC, and

    Input Current if operated in continuous mode for 30 seconds is 1 Amp max. That means wedo not need for a huge size of battery to operate it, and make the foot orthosis lighter.

    In the next section below, we will describe the MR technology in more details.

    Electronic Components: the strain gauge sensors will be used through Wheatstone bridge

    circuit. Each sensor will be connected with one bridge, known as a quarter bridge, see figure

    2-5. Then the output of the bridge will be the input of the amplifying circuit figure 2-6. This

    circuit has been designed and fabricated in Halmstad University Labs. And also filter circuithas been used to avoid the noise in sensor signals.

    Figure 2-5: Strain Gauge Figure 2-6: Amplifying Circuit

    sensors in a wheatstone bridge circuit

  • 8/8/2019 Adaptive Control of Foot Orthosis

    22/40

    Adaptive Control of Foot Orthosis

    17

    Control Circuit: the sensors signals then will be transferred to the control kit (this kit hasbeen designed and fabricated in Halmstad University). Which consist of a PIC

    Microcontroller, analog and digital input/output ports, power supply.

    The control program will be downloaded to this controller, and it will control the SSAFO

    during walking.

    All these components; the Wheatstone bridge circuits, amplifying and filtering circuits,Microcontroller kit and power supply for electronic circuits and MR damper will be

    mounted inside one box on the foot orthosis. The patient will wear the foot orthosis and be

    able to walk with it anywhere.

    Magneto-Rheological (MR) Technology

    In this part MR Technology is described in more details.

    Magneto-Rheology is a branch of Rheology that deals with the flow and deformation of thematerials under an applied magnetic field. The discovery of MR fluids is credited to JacobRabinow in 1949 [2]. Magneto-Rheological (MR) fluids are suspensions of non-colloidal

    (~0.05-10 m), multi-domain, and magnetically soft particles in organic or aqueous liquids.

    Magneto Rheological (MR) materials constitute examples of controllable(smart) fluids,

    whose Rheological properties vary in response to an applied magnetic field. These materials

    typically consist of micron-sized, magnet sable particles dispersed in an insulating medium.The essential characteristic of MR materials is that they may be continuously and reversibly

    varied from a state of free flowing liquids in the absence of an applied magnetic field to that

    of stiff semi-solids in a moderate field. This dipolar interaction is responsible for the chain

    like formation of the particles in the direction of the field, see Figure 2-7.

    Figure 2-7: Schematic of the formation of chain-like formation of magnetic particles in MR fluids in

    the direction of an applied magnetic field

  • 8/8/2019 Adaptive Control of Foot Orthosis

    23/40

    System Overview

    18

    MR technology offers: Real-time, continuously variable control of Damping, motion and

    position control, High dissipative force independent of velocity, Greater energy density,

    Simple design, Quick response time, Consistent efficacy across extreme temperaturevariations, minimal power usage.

    MR fluids can be operated directly from low-voltage power supplies and are significantly

    less sensitive to contaminants and extremes in temperature.

    Applications of MR fluids

    In the marketplace today, state-of-the-art MR fluids are becoming increasingly important inapplications concerning active control of vibrations or torque transfer. Shock absorbers,

    vibration dampers, seismic vibration dampers, clutches and seals are the most exciting

    applications of MR fluid.

    The Lord Corporation [9] has currently manufactured MR fluid devices for commercialapplications including heavy-duty vehicle seat suspensions, rotary brakes that provide a

    tunable resistance for exercise equipment and vibration dampers for various industrialapplications such as dampers for washing machines. The damper that contains MR fluids

    for commercial Applications such as seat suspensions is presented in Figure 2-8.

    Figure 2-8: MR fluid damper 1) Plastic shaft, 2) Sponge saturated with MR fluid, 3) Coil,

    4) Steel tube, 5) Wire supplying current

    The biggest challenge of MR fluid is to have high turn up ratio, temperature stability anddurability. The second biggest challenge of MR fluids is the materials science oriented

    studies such as surface chemistry, polymer physics, in synthesizing stable and redispersal

    MR fluids.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    24/40

    Adaptive Control of Foot Orthosis

    19

    2.3.2 SSAFO principle of work:

    The principle of work for our SSAFO will be as follow:

    During the gait cycle, there are three main phases: Heel phase, stance phase and swing

    phase, as mentioned before. So in each phase the ankle angle will change in a way that issuitable to ground level so the human body remain balanced during walking. In case of the

    patient with foot drop, he cannot change this ankle angle during the walking. So the SSAFOshould do this mission for him.

    In stance phase, the SSAFO will be in the zero position, when the MR damper and the bar(a) are parallel to each other, see figure 2-9(a). So the MR damper will be max compressed

    by the human body weight and the ankle angle is 90 degree (corresponding to ground level).

    When the person start to walk, the ankle angle will start to decrease because of the body

    weight and the force generated on the front foot fingers, and generate the momentum torquein Dorsflexsion direction, so the position of the MR damper will be changed corresponding

    to the bar 2-9(b) position, and make a triangle with it. In this moment the controller willgive the command to the MR damper to freeze, and fix the foot ankle angle in that position,depends on the sensor signals, and this will be the end of stance phase.

    After finishing the stance phase, the swing phase will start, and signals from the sensors willdefine it to the controller. In this phase the controller will command the MR damper to

    release and make it self free moving, so by using the body weight again and the mechanical

    joints shown in the figure 2-9(c), the SSAFO will change the ankle angle to increase,because of the force applied on the heel and the momentum torque generated in

    Planterflexsion direction. The angle then returns to the zero position. After that thecontroller will give the command again to the MR damper to Freeze and keep the foot in

    that position.

    2-9(a) 2-9(b) 2-9(c)

    Gait Phases using new SSAFO

    By repeating these procedures, the SSAFO will be able to control the ankle angle and

    change it depending on the body weight and MR damper.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    25/40

    Adaptive Control of Foot Orthosis

    19

  • 8/8/2019 Adaptive Control of Foot Orthosis

    26/40

    Experimental methods

    20

    3.Experimental methods:

    The Experiments that have been done in this project depends on the human walking, so threedifferent procedures were done in order to recognize different signals from the sensors that are

    used, and also try to define each gait phase from these signals. Below are the descriptions for

    each procedure in details:

    Walking horizontal:

    The Experiment have been done with respect to the human cycle different criteria such as:

    speed (normal speed equal to 1.4 m/s , slow speed equal to 1 m/s and fast speed equal to 1.6

    m/s ), walking uphill ( + 5 degrees ), and walking downhill ( - 5 degrees ).

    The picture below shows the human gait steps for one gait cycle.

    ..... Step ...... ....... Step ........

    ........................ Stride ........................

    Figure 3-1: this figure showing the one cycle.

    Step Length means the distance from one foot hit to the next (left to right or right to left) -with reference to 0.75 m for normal adults.

    Stride Length (one gait cycle) means the two following steps (by both left and right feet -

    with reference to 1.5 m for normal adults.

    Swing means the period when the foot is off the ground (38% of gait cycle in normal

    walking).

    Then the time and distance have been measured in order to analysis the signals obtainedfrom the sensor for each cycle. See figure 3-2 below.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    27/40

    Adaptive Control of Foot Orthosis

    21

    Figure 3-2:

    Walking up hill:

    Experiments on walking up hill were done using 5 degree inclination with a distance 1.5

    meter each. Data was collected 100 samples per second with different walking speeds, i.e.fast and slow walking speeds. In walking up hill, heel comes in contact with the ground first

    and hard, this can be seen in the results also, which will be discussed later in chapter 4.

    Figure 3-3

    Walking down hill:

    Experiments on walking down hill were done using 5 degree declination with a distance 1.5

    meter each. Data was collected 100 samples per second with different walking speeds, i.e.

    fast and slow walking speeds. In walking down hill, heel comes in contact with the groundfirst but the force is applied more on toe.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    28/40

    Experimental methods

    22

    Figure 3-4

    Further details will be discussed in detail in later chapter.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    29/40

    Adaptive Control of Foot Orthosis

    23

    4. Results and Discussion:

    This chapter enlists the experimental results done using strain gauge sensors at four differentpositions. Hence to determine which sensor positions are most appropriate for this kind of

    sensing equipment.

    Different sensor positions have been shown in section 2.1.3.1. Sensor 1&4 are on the lower side

    of the orthosis while 2&3 are on the top. Four different sensor combinations were used to

    determine gait steps, these are;

    1-42-4

    1-3

    2-3

    Signals from these sensors are in milli-volts, hence they need to be amplified. These signals are

    discrete and out of phase, but for easiness heel sensor signals are inverted while being plotted todifferentiate different phases of the gait cycle. Results from these will be discussed below in

    detail.

    Sensor 1&4 (Normal Walking):

    The idea behind these sensor positions was that to detect heel down and front down phase

    without any delay and it was also seen that sensors respond fairly good at these positions.These were tested on walking with different speeds and angels as mentioned in chapter 3.

    Figure 4-1: Sensor 1&4 (Normal Walking)

    VO

    L

    T

    Number of samples

    Red: Front SensorBlue: Back SensorGreen: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: GaitBoarderRed Line: Threshold

  • 8/8/2019 Adaptive Control of Foot Orthosis

    30/40

    Result and Discussion

    24

    Above mentioned figure shows the signals from walking on a normal speed on a horizontal

    plane with shoes. Red curve describes front sensor and blue curve shows heel sensor signal.For each sensor a threshold is configured to differentiate between noise/unwanted signal

    and the actual step. Red line shows this threshold. For sensor 1 it is Zero and for sensor 4 it

    is -1.25. These values remain constant for these sensor positions.

    One assumption taken is that, the gait cycle will always start from heel down. Analyzing the

    figure, blue values below -1.25 are the actual heel down (which have been marked as light

    green color). Blue square shows the flat foot phase. Red curve above 0 displays the frontdown. Yellow square shows the swing phase.

    In phase 1 (heel down) control circuit for the damper starts decreasing the voltage and ineffect the damper starts increasing in length from size A to size B. In phase 2 (flat foot),

    damper remains on position B. In phase C (front down), the damper will decrease in length

    from size B to size C, in effect lifting the heel. In phase 4 (swing phase), damper will move

    from position C to position A, as discussed in section 2.2.2. This way the gait cyclecontinues.

    There are two ways of detecting what next phase is going to be.

    1. A clipper/clamper can be used to get rid of the noise/unwanted signal below threshold. Inthis case next case will be assumed, i.e. after heel down phase it will always be a flat footand so on. The advantage of this approach is a simpler design of the orthosis system.

    2. Signal processing can be used to determine the different phases. The advantage of thisapproach is that gait phases can be determined accurately. After a heel is detected, in

    accordance with the figure, signal from sensor 1 is increasing while signal from sensor 2

    is decreasing. It shows flat foot phase of the gait cycle. Opposite is the case for swingphase. If the signal from sensor 1 is decreasing and from sensor 2 is increasing, it is

    considered to be a swing phase. This approach is accurate but time consuming.

    Sensor positions 1&4 both give pretty good signals and different gait phases can be detectedquite easily.

    It is seen that while walking normal with sensor positions 1&4, the swing phase and flatfoot phase are almost have the same time interval.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    31/40

    Adaptive Control of Foot Orthosis

    25

    Sensor 2&4 (Normal Walking):

    Sensor position 2 lies under the toe which is the maximum bending point and relatively high

    signal was expected from this position.

    Same measurements were taken with these sensor positions as done on the previous ones.

    Figure 4-2: Sensor 2&4 (Normal Walking)

    It can be seen that by using these positions accurate phases in gait cycle can be determined.

    The only disadvantage if compared to the sensor position 1 is that it gives more bending in

    the opposite direction thus giving us more noise/unwanted signal. Due to this fact thethreshold for sensor position 2 is also changed from 0 to 1.20.

    Number of

    VO

    LT

    Red: Front SensorBlue: Back SensorGreen: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: Gait

    BoarderRed Line: Threshold

    Number of samples

  • 8/8/2019 Adaptive Control of Foot Orthosis

    32/40

    Result and Discussion

    26

    Sensor 2&3 (Normal walking):

    Position 3 is right beneath the heel and idea for this placement was that while in heel phase

    a fairly good signal can be attained.

    Figure 4-3: Sensor 2&3 (Normal Walking)

    It can be seen that by using sensor position 3, the gait cycle phases cannot be determined.

    As the signal from the heel decreases very slowly absorbing the next phases. Due to thisreason the threshold for sensor 3 also cannot be determined.

    Sensor position 3 therefore is not a good sensor position.

    Number of samples

    VO

    LT

    Red: Front SensorBlue: Back SensorGreen: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: Gait

    BoarderRed Line: Threshold

  • 8/8/2019 Adaptive Control of Foot Orthosis

    33/40

    Adaptive Control of Foot Orthosis

    27

    Sensor 1&4 (walking uphill):

    Figure 4-4: Sensor 1&4 (walking uphill)

    It can be seen that while walking uphill the signal from front sensor is quite large as

    compared to the heel signal. It can also be seen that the front down phase is noticeably

    larger if compared to the heel down.

    Number of samples

    V

    OL

    T

    Red: Front SensorBlue: Back SensorGreen: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: Gait

    BoarderRed Line: Threshold

  • 8/8/2019 Adaptive Control of Foot Orthosis

    34/40

    Result and Discussion

    28

    Sensor 2&4 (walking uphill):

    Figure 4-5: Sensor 2&4 (walking uphill)

    Each sensor is calibrated at 0.1 V while in swing phase. This calibration gives an advantage

    that when front/toe sensor bends; it gives signal in positive while heel sensor operates

    opposite to it. According to the measurements following table is deducted.

    Number of samples

    V

    O

    L

    T

    Red: Front SensorBlue: Back SensorGreen: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: Gait

    BoarderRed Line: Threshold

  • 8/8/2019 Adaptive Control of Foot Orthosis

    35/40

    Adaptive Control of Foot Orthosis

    29

    Sensor 1&4 (walking downhill):

    Figure 4-6: Sensor 1&4 (walking downhill)

    It can be seen in the figure that we can easily differentiate all gait cycles. It is to be noted

    that heel down phase is larger than front down in this case.

    Number of samples

    V

    O

    LT

    Red: Front SensorBlue: Back Sensor

    Green: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: Gait

    BoarderRed Line: Threshold

  • 8/8/2019 Adaptive Control of Foot Orthosis

    36/40

    Result and Discussion

    30

    Sensor 2&4 (walking downhill):

    Figure 4-7: Sensor 2&4 (walking downhill)

    Table 1 below shows signal strength on different sensor positions according to different walkingconditions.

    Sensor No. Up Hill Signal Down Hill Signal Normal Signal

    1 > 2V < 1.5V 1.5 < V < 2

    2 > 2V < 1.5V 1.5 < V < 2

    3 > 1.5V > 1.5V < 1.5V

    4 < 1.5V > 2V 1.25 < V < 1.5

    Table 1

    Table 1 shows different values from sensors while experimenting with different conditions. The

    table is self explanatory and this information can further be integrated with signal gait analysis to

    get a more accurate state.

    Number of samples

    VO

    L

    T

    Red: Front SensorBlue: Back Sensor

    Green: Heel DownNavy Blue: Flat footGrey: Front downYellow: SwingGrey line: Gait

    BoarderRed Line: Threshold

  • 8/8/2019 Adaptive Control of Foot Orthosis

    37/40

    Adaptive Control of Foot Orthosis

    31

    As described previously that walking uphill Front Down phase is fairly larger than the heel down

    phase. Integrating this analysis with table entry col. 2, row 1 a more accurate decision boundary

    can be made. All these values in the table are a mean value of multiples of foot step values while

    experimenting.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    38/40

    Conclusion

    32

    5.Conclusion:

    This project studies the successful implementation of the proposed sensor system to detect

    different phases of gait cycle. Four gauge sensors were used to determine these phases. These

    sensors were placed at different orthosis points for better detection of the bending forces while

    walking.

    A lot of work has been done in this area using many type of sensors and orthosis equipment. Such

    as pressure, force, capacitive sensors combined with rotary angel potentiometer to determine

    different gait phases. Our contribution to the work is that low cost, high accuracy, simple design

    sensor system consisting of four strain gauge sensors only, along with a design of a new

    technology MR Damper has been proposed.

    Experiments show that by using only one kind of a sensor without any rotary angel

    potentiometer, we were able to successfully determine the gait phases accurately. We can also

    differentiate between horizontal, uphill and downhill walking. It can clearly be seen that while

    walking uphill, the front-down phase is fairly larger as compared to the heel-down phase.Opposite is in the case of walking downhill. While walking horizontal, front-down and heel-down

    phases are almost the same. Further is it also concluded that sensor positions 1, 2&4 are most

    appropriate for these kinds of sensors. Sensor combination 1&4 and 2&4 give somewhat similar

    results, so either one of these can be used for measurements. For more accurate readings these

    three sensors can be combined along with the values from the table 1.

  • 8/8/2019 Adaptive Control of Foot Orthosis

    39/40

    References:

    1. Joaquin A. Blaya, Force-Controllable Ankle Foot Orthosis (AFO) to Assist DropFoot Gait. : Master thesis in Mechanical Engineering at the MASSACHUSETTSINSTITUTE OF TECHNOLOGY - February 2003

    2. Seval Gen, SYNTHESIS AND PROPERTIES OF MAGNETORHEOLOGICAL(MR) FLUIDS. Master thesis at University of Pittsburgh 2002

    3. Ronthal, Michael. Gait Disorders. Boston: Butterworth-Heinemann, 2002.4. M. Goldfrab, A control brake Orthosis for FES aided gait. Mechanical Engineering.

    Cambridge, MA: MIT Press, 1994, p. 272.

    5. J. Michael Wheatley, Ph.D. Ankle Foot Orthosis (AFO) what is it?http://www.mhcop.com/articles/ankle.htm . 10-09-2006.

    6. Sumiko Yamamoto, PhD, Masahiko Ebina, CPO, Shigeru Kubo, Eng., Takeo Hayashi,Yoshiyuki Akita, CPO, Yasuyuki Hayakawa, CPO, Development of an Ankle-Foot

    Orthosis with Dorsiflexion Assist. Part 2. Journal of Prosthetics and Orthotics 1999;

    Vol 11, Num 2, p 24.

    URL: http://www.oandp.org/jpo/library/1999_02_024.asp

    7. Joaquin A. Blaya and Hugh Herr, Adaptive Control of a Variable-Impedance Ankle-Foot Orthosis to Assist Drop-Foot Gait. IEEE TRANSACTIONS ON NEURALSYSTEMS AND REHABILITATION ENGINEERING, VOL. 12, NO. 1, MARCH 2004

    8. Daniel Grafstrm, GAIT ANALYSES USING PORTABLE TACTILE SENSORS.Master thesis in Mechatronical Engineering , Halmstad university - 2005

    9. Lord Company , http://www.lord.com/tabid/3318/Default.aspx , 10 09 200610.AKAZAWA Yasushi, NAKAGAWA Akio, MATSUBARA Hiroyuki, NAKAMURA

    Toshiya NOMURA Tsuyoshi, TANAKA Masao, Introducing Mechatronical

    Technology into Ankle-Foot Orthosis Joint. Experimental gait wearing an AFO withVariable Joint Viscosity GRADATE school of Osaka university - 2005

    11.D. P. Ferris, J. M. Czerniecki, and B. Hannaford, An ankle foot Orthosis (AFO)powered by artificial muscles JOURNAL OF APPLIED BIOMECHANICS, 2005, 21,

    189-197.

    12.http://www.thefreedictionary.com/dorsiflexion , 10 09 200613.http://en.wikipedia.org/wiki/Gait , 10 09 2006

  • 8/8/2019 Adaptive Control of Foot Orthosis

    40/40

    14.Michael W. Whittle, Gait Analysis an introduction.15. Robert M. Havey, Methodology Measurments, Part II: Instrumentation and

    Apparatus.

    16. http://www.cs.princeton.edu/sound/learning/tutorials/sensors/node8.html) , 10 09 2006

    17. Daniel P. Ferris, University of Washington,1998.


Recommended