+ All Categories
Home > Documents > Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices...

Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices...

Date post: 11-Jul-2020
Category:
Upload: others
View: 1 times
Download: 0 times
Share this document with a friend
7
Advanced charge injection devices for space instruments Kyle Miller and Kathy Doughty, Mission Research Corporation, 5017 N. 30 th Street, Colorado Springs, CO, USA 80919 ABSTRACT Charge Injection Devices (CIDs) have historically played a niche role in visible imager technologies, mainly for applications requiring high radiation tolerance. They have not exhibited the radiometric performance of competing visible- imaging technologies such as CCDs, and so have not been widely applied to space instrument systems. Recent advances in CIDs have demonstrated much higher radiometric performance as well as lower noise operation, without compromising the radiation tolerance of the devices, making the devices suitable for a wide range of space instruments. We present radiometric, noise, and radiation response data for several of the newest CID designs that are candidate technologies for visible space telescope systems. 1. INTRODUCTION Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge- coupled device (CCD), but have not received the same acceptance in the scientific imaging community. This is primarily due to the higher read noise inherent in the design in which the output amplifier must ”see” the capacitance of the entire device, whereas the CCD has a very low read noise due to a small readout capacitor. However, CID’s have also historically demonstrated a vast superiority to the CCD in radiation tolerance, with some designs demonstrating capabilities in the mega-rad range. CCDs are well known to be susceptible to both ionizing and non-ionizing radiation damage, which can limit their performance in a space-based application by dramatically decreasing the charge transfer efficiency whereby charge is moved from one pixel to the next. We discuss a new development in CID technology which focuses on devices that can be used in a stellar reference unit (star tracker). This development effort was initiated by the Jet Propulsion Laboratory, under the Europa Orbiter program, who which demonstrated a need for a highly radiation- tolerant imager that could withstand the high-energy electron environment around Jupiter. The high-energy electron environment produces lifetime damage effects (total ionizing dose and silicon displacement damage effects) as well transient noise generated caused by the high-energy electrons impinging on the pixel storage units and generating spurious signal. The device is also designed needed to meet the radiometric requirements for use in a moderately accurate star tracker. Two devices were developed as a result of this program: the CID816, which features a preamplifier-per-row design and the CID817, which is a preamplifier-per-pixel design. In addition, a third device is under development (delivered wafers) which is a preamplifier-per-pixel design that is thinned and backside illuminated. The thinning is done to reduce the per-charged-particle noise on the detector by reducing the volume of active material that charged particles can interact with, and Backside thinning illumination is used to increase the fill factor and effective device responsivity. The CID816s have been available for approximately 9 months, and have been had preliminary radiometric characterization performed as well as some radiation characterization. The CID817 and CID818 are recent developments and are still in fabrication. Radiation and radiometric characterization is due to occur during the fall of 2003. 2. CID OPERATION The basic CID operation has been described before 1234 and the new imagers follow the same basic concepts. The CID pixel structure utilizes two overlapping MOS capacitors and senses the charge transfer between the two storage sites. The output can be sensed by an amplifier on each device, amplifiers in the row or column, or by in-pixel
Transcript
Page 1: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

Advanced charge injection devices for space instruments

Kyle Miller and Kathy Doughty,Mission Research Corporation, 5017 N. 30th Street, Colorado Springs, CO, USA 80919

ABSTRACT

Charge Injection Devices (CIDs) have historically played a niche role in visible imager technologies, mainly forapplications requiring high radiation tolerance. They have not exhibited the radiometric performance of competingvisible- imaging technologies such as CCDs, and so have not been widely applied to space instrument systems.Recent advances in CIDs have demonstrated much higher radiometric performance as well as lower noise operation,without compromising the radiation tolerance of the devices, making the devices suitable for a wide range of spaceinstruments. We present radiometric, noise, and radiation response data for several of the newest CID designs thatare candidate technologies for visible space telescope systems.

1. INTRODUCTION

Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled device (CCD), but have not received the same acceptance in the scientific imaging community. This isprimarily due to the higher read noise inherent in the design in which the output amplifier must ”see” thecapacitance of the entire device, whereas the CCD has a very low read noise due to a small readout capacitor.However, CID’s have also historically demonstrated a vast superiority to the CCD in radiation tolerance, with somedesigns demonstrating capabilities in the mega-rad range. CCDs are well known to be susceptible to both ionizingand non-ionizing radiation damage, which can limit their performance in a space-based application by dramaticallydecreasing the charge transfer efficiency whereby charge is moved from one pixel to the next.

We discuss a new development in CID technology which focuses on devices that can be used in a stellar referenceunit (star tracker). This development effort was initiated by the Jet Propulsion Laboratory, under the EuropaOrbiter program, who which demonstrated a need for a highly radiation- tolerant imager that could withstand thehigh-energy electron environment around Jupiter. The high-energy electron environment produces lifetime damageeffects (total ionizing dose and silicon displacement damage effects) as well transient noise generated caused by thehigh-energy electrons impinging on the pixel storage units and generating spurious signal. The device is alsodesigned needed to meet the radiometric requirements for use in a moderately accurate star tracker. Two deviceswere developed as a result of this program: the CID816, which features a preamplifier-per-row design and theCID817, which is a preamplifier-per-pixel design. In addition, a third device is under development (deliveredwafers) which is a preamplifier-per-pixel design that is thinned and backside illuminated. The thinning is done toreduce the per-charged-particle noise on the detector by reducing the volume of active material that chargedparticles can interact with, and Backside thinning illumination is used to increase the fill factor and effective deviceresponsivity.

The CID816s have been available for approximately 9 months, and have been had preliminary radiometriccharacterization performed as well as some radiation characterization. The CID817 and CID818 are recentdevelopments and are still in fabrication. Radiation and radiometric characterization is due to occur during the fallof 2003.

2. CID OPERATION

The basic CID operation has been described before1234 and the new imagers follow the same basic concepts. TheCID pixel structure utilizes two overlapping MOS capacitors and senses the charge transfer between the two storagesites. The output can be sensed by an amplifier on each device, amplifiers in the row or column, or by in-pixel

Page 2: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

amplifiers. Because the readout mechanism is non-destructive, i.e., the charge is not removed during a readoutcycle, the temporal noise can be reduced by averaging multiple reads of the same charge and averaging. Thisfeature also allows multiple frame integration which creates the ability for differing dynamic ranges across theimager. The pre-amplifier per pixel design allows random addressing of individual pixels or windows withoutreading out the entire imager, allowing for unique scientific applications to be designed. Figure 1 shows the basicpixel operation of the devices.

Figure 1: CID Operation

1. Integration

RowLow

ColumnLow

Transfer GateLow

Potential Wellsare formed underthe photogates

RowHigh

ColumnLow

Transfer GateLow

2. Back transfer

Charge istransferred fromthe row (sense)electrode to thecolumn (storage)electrode bycollapsing thesensing potentialwell

RowLow

ColumnLow

Transfer GateLow

3. Row Referencing

A deep potential wellis established underthe sensing well

ColumnLow

TransferGateLow

Row Out 1

4. Readout before transfer

The sensing gateis left floating andit’s potential ismeasured

ColumnHigh

Transfer GateLow

Charge is transferredfrom the column(storage) electrode tothe row (sensing)electrode potential well

5. Forward transfer

Row OutTransfer Gate

Low Column High The potential of the

sensing gate ismeasured again. Thedifference between thefirst read and secondread is the signal.

Row Out 2

6. Readout after transfer

RowHigh

Transfer GateHigh

7. Injection

ColumnLow

The potential wellsare all collapsed andthe holes diffuse tothe substrate

Page 3: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

3. CID DESCRIPTIONSDescriptions of the three devices are shown in the following table:

Device Parameter CID816 CID817 CID818Substrate Material n+ epi on p silicon n+ epi on p silicon n+ epi on p siliconArchitecture All PMOS All PMOS All PMOSPixels 512 x 512 512 x 512 512 x 512Pixel Pitch 27 micron square 27 micron square 27 micron squareAmplifier Pre-amp per row Pre-amp per pixel Pre-amp per pixelRandom Access Yes Yes YesThinned No No Yes (nominal 7 microns)Backside Illuminated No No Yes

Device Design Goal SpecificationsRead Noise < 200 electrons at 50

kHz< 50 electron/pixel @

50 kHz< 50 electron/pixel @ 50 kHz

Linearity < 2% pixel capacity < 2% pixel capacity < 2% pixel capacityBright row defects 0 0 0Bright column defects 0 0 0Cold Pixels < 16 < 16 <16Dark fixed patternnoise

< 100 electrons < 100 electrons < 100 electrons

Mean dark current < 0.025 nA/cm2 < 0.025 nA/cm2 < 0.025 nA/cm2Threshold Shift (inradiation)

-5 mV/krad(Si) -5 mV/krad(Si) -5 mV/krad(Si)

4. RADIOMETRIC DATA

Radiometric and noise data was taken for a number of CID816 imagers. Figure 2 shows the measured quantumefficiency for four devices along with the mean measured for a total of 10 devices. Figure 3 shows the mean per-pixel read noise measured at -40C to eliminate contributions from dark current for 10 different devices. Figure 4shows the mean per-pixel dark current for the same 10 devices measured at -20C, 0C, and +20C. Figure 5 shows thefull-well capacity of the same 10 devices.

0

5

10

15

20

25

30

35

40

400 450 500 550 600 650 700 750 800 850 900 950 1000

Wavelength (nm)

Per

cent

(%

)

wfr02-23

wfr09-14

wfr18-14

wfr15-06

Mean

Figure 2: CID816 Quantum Efficiency

200

210

220

230

240

250

260

270

280

290

300

0 1 2 3 4 5 6 7 8 9 10

Device

Mea

n R

ead

Noi

se @

40C

(el

ectr

ons)

Figure 3: CID816 Mean Read Noise

Page 4: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

5. RADIATION DATA

Proton Testing was performed at the Crocker Nuclear Laboratory, University of California, Davis. Protons of 45MeV and 63 MeV were used which, when translated through the window of the enclosing dewar, correspond toapproximately 17 and 60 MeV protons at the device. Two effects were measured: the device noise characteristics asa function of total proton fluence, and the transient noise effects due to the proton flux.

5.1 TOTAL PROTON FLUENCE CHARACTERIZATION

Noise characteristics for the device were taken from 0 to 50Krads. Above 50Krads the variation between the per-column amplifiers’ radiation compensation set-points became large enough that the columns could not be broughtsimultaneously into range. Work on addressing this issue though setting of epi-layer voltage is underway.

5.1.1 MEAN READ NOISE

Mean Read Noise is a measure of the frame-to-frame temporal variation of any individual pixel value. The test isrun with no light on the device. The temperature was held at a relatively cold value of -30C to eliminate darkcurrent contribution. Data was measured on 10 full imager frames (512x512 pixels). Mean Read Noise remainedfairly constant at 300-400 e-/read as shown in Figure 6.

500000

510000

520000

530000

540000

550000

560000

570000

580000

590000

600000

1 2 3 4 5 6 7 8 9 10

Device

Full

Wel

l Cap

acity

(el

ectr

ons)

Figure 5: CID816 Full Well Capacity

10

100

1000

10000

1 2 3 4 5 6 7 8 9 10

Device

Mea

n D

ark

Cur

rent

(el

ectr

ons/

seco

nd)

-20C0C20C

Figure 4: CID816 Mean Dark Current

Page 5: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

5.1.2 READ FIXED PATTERN NOISE (RFPN)

RFPN is a measurement of the spatial variation of read noise measured across all pixels. Again, the data was takenwith no incident optical signal, and was measured at –30C to eliminate dark current contribution. Data wasmeasured on 10 full imager frames (512x512 pixels). There was a definite increase in RFPN with dose as shown inFigure 7, although this is probably largely due to variation in amplifier set-point, as shown in Figure 7.

0

20

40

60

80

100

120

140

0 10 20 30 40 50 60

Total Dose [kRads]

rms

pxl-p

xl v

aria

tion

in re

ad n

oise

[e-]

(0C

) )

Figure 7: Read Fixed Pattern Noise

Figure 6: Mean Read Noise as a function of total proton dose

0

50

100

150

200

250

300

350

400

450

0 10 20 30 40 50 60

Total Dose [KRads]

Mea

n R

ead

Noi

se [e

-]

Page 6: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

0%

5%

10%

15%

20%

25%

0 10 20 30 40 50 60

Total Dose [kRads]

ST

DV

e/e

5.1.3 LINEARITY

Linearity is shown as the standard deviation of the signal at 2/3 well-fill. The initial values were under 5%,increasing with total fluence. Again, column-to-column variation accounts for the bulk of the increase.

Figure 8: Linearity as function of total proton fluence

5.2 TRANSIENT PROTON RADIATION RESPONSE

The response of the device to proton transients was also measured as a part of the characterization procedure. Asprotons travel through the device, electron/hole pairs are generated that are indistinguishable from those created byphoton absorption. This introduces a noise source that is dependent on the proton flux. To measure this, data wastaken in no-light conditions under a constant proton flux. Full-frame data was read out column by column. Becausethe columns were read off sequentially, the amount of time a column was exposed to the radiation flux depended onwhere that column was in the read-out order. This method therefore generated a series of data points in which thecolumn-to-column difference depends on the proton flux times the column read-out time. Example data is shown inFigure 9. A fit is made to the linear region of response. Taking into account the proton flux and column read-outtime gave a mean per-proton signal of 7500 e-/proton for the 17 MeV protons and 5200 e-/proton for the 60 MeVprotons. In a device, this transient radiation response would contribute an additional per-pixel noise of

Where N is the number of radiation events/pixel/frame and PH is the mean per-proton signal. These values areexpected to drop markedly for the thinner devices (CID 818).

2 N⋅ PH⋅

Page 7: Advanced charge injection devices for space instruments · 2009-05-20 · Charge injection devices (CIDs) were developed at approximately the same time as the nearly -ubiquitous charge-coupled

0

200

400

600

800

1000

1200

0 100 200 300 400 500 600

Column #

Mea

n C

olu

mn

Res

po

nse

mean 1-pt

saturated points

Linear (mean 1-pt)

Figure 9: Mean column response for 17 MeV protons

6. CONCLUSIONS

The new CID imagers entering the market show considerable improvement over previous versions. Presentamplifier-per-column devices (CID816s) show a measured noise floor of around 250 electrons, with a factor of 5 orbetter expected for the amplifier-per-pixel devices currently in production. The CID816s show a 30 to 35% peakQE, with a ½ million electron full-well. Back-side illumination of the CID818s is expected to increase the QE to80% or above. The measured values hold out to 50Krads Total Dose, and the tolerable dose is expected to risesharply with adjustments to the compensation circuits. The transient response to radiation is consistent with thatexpected for the active region size, and is similar to that seen for other unthinned devices of other types (e.g.: CCDs,APSs, hybridized PN junctions). Thinning the devices will reduce this value by a factor of 2 or more. These resultsindicate that the emerging CID devices will not only be radiation tolerant but also demonstrate radiometricperformance that will enable them to compete in the visible space instrument market.

REFERENCES 1 Zarnowski, J., et. al., “Selectable One to Four Port, Very High Speed 512 x 512 Charge Injection Device”, Proc.SPIE, Vol. 1447, pg. 191-201, 1991.2 Wentink, R., et. al., “CID Detectors for X-Ray Imaging”, SPIE, San Diego, 1994.3 Carbone, J., et. al., “Application of Low Noise CID Imagers in Scientific Instrumentation Cameras”, SPIE Vol.1447, Feb 25-27,1991, pg. 229-242.4 J. Carbone, et. al., “Megarad and Scientific CIDs”, SPIE IS&TS Symposium on Electronic Imaging andTechnology, 1996.


Recommended